K241628 · Medical Electronic Systems , Ltd. · POV · Nov 29, 2024 · Hematology
Device Facts
Record ID
K241628
Device Name
YO Home Sperm Test
Applicant
Medical Electronic Systems , Ltd.
Product Code
POV · Hematology
Decision Date
Nov 29, 2024
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 864.5220
Device Class
Class 2
Indications for Use
The YO Sperm Test (YO 3.0) is a smartphone-based home test for semen analysis performed by lay users. The parameters reported by the YO Home Sperm Test (YO 3.0) are: 1. Total Sperm Concentration / Sperm Concentration, M/mL 2. Total Motile / Motility (PR + Non Progressive [NP]), % 3. Progressive Motility (PR), % (combines Rapidly and Slowly Progressive, %) 4. Motile Sperm Concentration (MSC), M/mL 5. Progressively Motile Sperm Concentration (PMSC), M/mL (combines Rapidly and Slowly Motile Sperm Concentration, M/mL) The YO Home Sperm Test (YO 3.0) does not provide a comprehensive evaluation of a male's fertility status and is intended for in vitro, over the counter only.
Device Story
Smartphone-based semen analysis system for lay users; consists of a desktop unit with magnification lens, LED light source, and WiFi camera; connects to user's smartphone via WiFi. User prepares semen sample using provided collection cup, pipette, and fixed coverslip slide; inserts slide into device. App guides user through step-by-step preparation and testing; captures HD video of semen sample. Proprietary algorithm analyzes light fluctuations from sperm movement to calculate concentration and motility parameters. Results and video are stored and displayed on the smartphone app. Device provides quantitative results for 5 parameters; recommends medical consultation if results are below normal cut-offs. Benefits include accessible, home-based screening for male fertility factors; reduces need for immediate clinical visits for initial assessment.
Clinical Evidence
Clinical method comparison study (n=309) at 3 US sites comparing lay-user results on YO 3.0 vs. trained-operator results on SQA-V. Passing-Bablok regression showed strong correlation (r > 0.88) for all parameters. Slopes for concentration, MSC, and PMSC were 0.86, 0.92, and 1.03 respectively. Exit questionnaires confirmed lay users understood test parameters and follow-up procedures.
Technological Characteristics
Smartphone-based optical system. Components: plastic collection cup, fixed coverslip slide, liquefaction media (chymotrypsin), magnification lens, LED light source, WiFi camera. Connectivity: USB cable to smartphone, cloud-based software. Analysis: proprietary algorithm for video-based sperm tracking and light fluctuation analysis. Standards: CLSI EP05-A3, EP07, EP17-A2, EP06, ASTM D4169-16, IEC 61010-1.
Indications for Use
Indicated for lay users performing home semen analysis to measure sperm concentration, motility, and progressive motility. Not for comprehensive fertility evaluation.
Regulatory Classification
Identification
An automated differential cell counter is a device used to identify one or more of the formed elements of the blood. The device may also have the capability to flag, count, or classify immature or abnormal hematopoietic cells of the blood, bone marrow, or other body fluids. These devices may combine an electronic particle counting method, optical method, or a flow cytometric method utilizing monoclonal CD (cluster designation) markers. The device includes accessory CD markers.
Special Controls
*Classification.* Class II (special controls). The special control for this device is the FDA document entitled “Class II Special Controls Guidance Document: Premarket Notifications for Automated Differential Cell Counters for Immature or Abnormal Blood Cells; Final Guidance for Industry and FDA.”
{1}
FDA
U.S. FOOD & DRUG
ADMINISTRATION
# 510(k) SUBSTANTIAL EQUIVALENCE DETERMINATION DECISION SUMMARY
INSTRUMENT ONLY
## I Background Information:
A 510(k) Number
K241628
B Applicant
Medical Electronic Systems Ltd.
C Proprietary and Established Names
YO Home Sperm Test
D Regulatory Information
| Product Code(s) | Classification | Regulation Section | Panel |
| --- | --- | --- | --- |
| POV | Class II | 21 CFR 864.5220 - Automated Differential Cell Counter | HE - Hematology |
## II Submission/Device Overview:
A Purpose for Submission:
Clearance of a new device
B Type of Test:
Quantitative analysis of the following semen parameters:
Total Sperm Concentration / Sperm Concentration, M/mL
Total Motile / Motility (PR + Non Progressive [NP]), %
Progressive Motility (PR), % (combines Rapidly and Slowly Progressive, %)
Motile Sperm Concentration (MSC), M/mL
Progressively Motile Sperm Concentration (PMSC), M/mL (combines Rapidly and Slowly
Motile Sperm Concentration, M/mL)
Food and Drug Administration
10903 New Hampshire Avenue
Silver Spring, MD 20993-0002
www.fda.gov
{2}
K241628 - Page 2 of 12
## III Intended Use/Indications for Use:
### A Intended Use(s):
See Indications for Use below.
### B Indication(s) for Use:
The YO Sperm Test (YO 3.0) is a smartphone-based home test for semen analysis performed by lay users.
The parameters reported by the YO Home Sperm Test (YO 3.0) are:
1. Total Sperm Concentration / Sperm Concentration, M/mL
2. Total Motile / Motility (PR + Non Progressive [NP]), %
3. Progressive Motility (PR), % (combines Rapidly and Slowly Progressive, %)
4. Motile Sperm Concentration (MSC), M/mL
5. Progressively Motile Sperm Concentration (PMSC), M/mL (combines Rapidly and Slowly Motile Sperm Concentration, M/mL)
The YO Home Sperm Test (YO 3.0) does not provide a comprehensive evaluation of a male's fertility status and is intended for in vitro, over the counter only.
### C Special Conditions for Use Statement(s):
OTC - Over The Counter
## IV Device/System Characteristics:
### A Device Description:
The YO Home Sperm Test (YO 3.0) is a smartphone-based, over-the-counter, home test for semen analysis performed by lay users that provides a quantitative analysis of total sperm concentration, total motile/motility, progressive motility, motile sperm concentration (MSC) and progressively motile sperm concentration (PMSC). The YO Home Sperm Test (YO 3.0) utilizes proprietary algorithms to both conduct semen analysis and present and store the results and videos on the user's smartphone and in the cloud. The YO application (app) is downloaded onto the user's own smartphone (iPhone/Android) and is controlled by the user through a proprietary graphical interface (GUI). The GUI guides the user through the process step by step on the App's screen and operates the YO device.
The YO Home Sperm Test kit provides the supplies necessary to test up to six semen samples and contains:
- Plastic cups for collecting the semen sample
- Fixed coverslip slides
- Vials containing 5 mg chymotrypsin liquefaction media
- Disposable plastic pipettes for sample aspiration and delivery to the fixed coverslip slide
- YO Device
- USB cable for connecting the YO device to the user's smartphone/PC
- Indications for Use
{3}
B Principle of Operation:
The plastic YO device contains a fixed coverslip slide insertion channel, magnification lens, lens holder, WiFi camera and an LED that lights up the optical path. A specimen-filled slide is inserted into the OTC device unit and illuminated by the light source. The image is magnified by a magnification lens in the YO device and projected to the image sensors (YO camera). The camera transmits the video to the smartphone. The YO software captures a video in HD (high definition) mode and implements a unique software algorithm to identify sperm and analyze the light fluctuations resulting from sperm movement to report semen values. The algorithm recognizes when the YO autofocus function has the best image and then defines the optimal area of the video for analysis. MSC/PMSC are measured by capturing a video of sperm cells in high definition (HD) mode, and analyzing the moving and total spermatozoa, expressed in millions per milliliter (M/mL). Sperm concentration is calculated by counting the objects from the same sperm video and classifying them as sperm cells.
C Instrument Description Information:
1. Instrument Name:
YO Home Sperm Test
2. Specimen Identification:
Specimen identification is not required because the YO Home Sperm Test is intended for a single user at home.
3. Specimen Sampling and Handling:
The users are instructed to wash their hands with soap and water before and after handling their semen sample or test components. If the user is running the test for someone else, the user will wear exam gloves. When instructed by the YO app, the user collects the sample directly into the sample container provided in the YO kit. The sample should be placed at room temperature. Once the user places the liquefaction powder into the sample as directed in the application, the user should wait ten minutes before testing as instructed. The App instructs the user to not wait longer than one hour from collection to test the sample.
4. Calibration:
Not applicable.
5. Quality Control:
Internal controls are performed automatically and internally. The software checks that the video signal is stabilized and running smoothly for analysis and that the video is not black or blank.
K241628 - Page 3 of 12
{4}
V Substantial Equivalence Information:
A Predicate Device Name(s):
Sqa V, Sperm Quality Analyzer, YO Home Sperm Test
B Predicate 510(k) Number(s):
K021746, K161493
C Comparison with Predicate(s):
| Device & Predicate Device(s): | K241628 | K161493 | K021746 |
| --- | --- | --- | --- |
| Device Trade Name | YO Home Sperm Test | YO Home Sperm Test | SQAV, Sperm Quality Analyzer |
| General Device Characteristic Similarities | | | |
| Intended Use/Indications For Use | The YO Home Sperm Test (YO 3.0) is a smartphone-based home test for semen analysis performed by lay users. The parameters reported by the YO Home Sperm Test (YO 3.0) are:
1. Total Sperm Concentration/Sperm Concentration, M/mL
2. Total Motile/Motility (PR + Non Progressive [NP], %
3. Progressive Motility (PR), % (combines Rapidly and Slowly Progressive, %)
4. Motile Sperm Concentration (MSC), M/mL
5. Progressively Motile Sperm Concentration (PMSC), M/mL (combines Rapidly and Slowly Motile Sperm Concentration, | The YO Home Sperm Test is a smartphone-based home test which provides a qualitative assessment of motile sperm concentration (MSC) in human semen. MSC is one aspect of a male semen examination. The YO Home Sperm Test does not provide a comprehensive evaluation of a male’s fertility status and is intended for over-the-counter, in vitro use only. | The SQAV is a point-of-care, easy-to-use, electro-optical device with on-screen visualization and image freezing capabilities for semen analysis. The SQQ V provides direct and calculated measurements for:
• Total sperm concentration (TSC, millions/mL)
• Percent motility (%MOT) and % progressive motility (%PMOT)
• % normal morphology (%MORPH)
• Motile sperm concentration (MSC, millions/mL) and progressive MSC (PMSC)
• Functional sperm concentration |
K241628 - Page 4 of 12
{5}
| | M/mL)
The YO Home Sperm Test (YO 3.0) does not provide a comprehensive evaluation of a male’s fertility status and is intended for in vitro, over-the-counter (Home Use) use only. | | (FSC, millions/mL) |
| --- | --- | --- | --- |
| Intended User | Lay Users | Same | Healthcare Professionals (trained lab technicians) |
| WHO compliance | WHO 6^{th} | WHO 5^{th} | WHO 5^{th} and WHO 6th |
| Sample Type | Human semen | Same | Same |
| Sample delivery method | YO Fixed Cover Slip Slide which is inserted into the YO device | Same | SQA-V testing capillary which is inserted into the SQA-V device |
| Semen parameters | • Total Sperm Concentration/Sperm Concentration, M/mL
• Percent Motility/Total Motile (PR+NP), %
• Motile Sperm Concentration (MSC), M/mL
• Progressive Motility (PR), % (combines Rapidly and Slowly Progressive, %)
• Progressively Motile Sperm Concentration (PMSC), M/mL (combines Rapidly and Slowly Motile Sperm Concentration, M/mL) | • Motile Sperm Concentration (MSC), millions/mL above and below the 6M/mL cut-off for normal/abnormal. Results are presented as LOW or Moderate/Normal Range | • Sperm concentration/Total Sperm Concentration, M/mL
• Percent Motility/Total Motile (PR+NP), %
• Motile Sperm Concentration (MSC), M/mL
The SQA-V also reports other parameters. |
| Test Type | Quantitative | Qualitative | Quantitative |
| General Device Characteristic Differences | | | |
| Technology | Desktop unit consists of | Magnification “Clip” | Desktop unit consists of |
K241628 - Page 5 of 12
{6}
K241628 - Page 6 of 12
| | a light source, and image sensors (camera), wirelessly connected to a smartphone, onto which the cloud-based software is downloaded. The software is used to capture and analyze magnified sperm videos using a proprietary algorithm to produce results. A specimen filled slide is inserted into the OTC device unit and illuminated by the light source. The image is magnified by a magnification lens in the YO device and projected to the image sensors (YO camera). The camera transmits the video to a smartphone. | which slides over the top of the Smartphone. Software downloaded onto the phone is used to capture and analyze magnified sperm videos using a proprietary algorithm to produce results. A specimen-filled slide is inserted into the “Clip” and illuminated by the smartphone’s light source. In addition, the “Clip” utilizes the smartphone’s lens and image sensors (camera) to assess semen parameters. The image is magnified by the magnification lens in the “Clip”, and using mirrors projected back to the smartphone’s lens and to the smartphone’s image sensors (camera). | a light source, optical sensors (electro-optical processing, conversion of light fluctuations to analog signals), built-in video microscopy and an internal computer containing algorithms for the assessment of semen parameters. A specimen-filled capillary is inserted into the desktop unit and illuminated by the light source. |
| --- | --- | --- | --- |
| Connection Technology | Wireless to smartphone/USB cable to power the YO device | Physical connection to a smartphone with the use of a clip that slides over the top of the smartphone | RS32 cable to PC |
| Software | Cloud-based software | On-premise software | On-premise software |
VI Standards/Guidance Documents Referenced:
- CLSI EP05-A3: Evaluation of Precision Performance of Quantitative Measurement Methods; Approved Guideline – Third Edition
- CLSI EP07-3rd Edition: Interference Testing in Clinical Chemistry; Approved Guideline – Third Edition
- CLSI EP17-A2: Evaluation of Detection Capability for Clinical Laboratory Measurement Procedures; Approved Guideline – Second Edition
- CLSI EP06: Evaluation of Linearity of Quantitative Measurement Procedures, 2nd Edition
- ASTM D4169-16 Standard Practice for Performance Testing of Shipping Containers and Systems
- IEC 61010-1 Edition 3.1 2017-01 Safety Requirements for Electrical Equipment For Measurement, Control, And Laboratory Use – Part 1: General Requirements
{7}
VII Performance Characteristics (if/when applicable):
## A Analytical Performance:
### 1. Precision/Reproducibility:
#### Repeatability
Seventy-eight lay users at three US sites tested their own samples in triplicate using multiple devices and three different slide lots (one per site). Parameters reported were sperm concentration, motile sperm concentration, progressively motile sperm concentration, motility and progressive motility. The Standard deviations (SDs) and percent coefficients of variation (%CVs) were calculated for each sample. The samples spanned the sperm concentration range of 2–150 M/mL. The results from all samples demonstrated acceptable repeatability of the device used by lay users.
#### Reproducibility
Sixteen native semen samples, representing different levels of sperm concentration, motile sperm concentration and progressively motile sperm concentration were assayed in duplicate on three devices at three sites over four timepoints. Due to limited stability of semen samples, each “day” is represented by different times of day with testing at four different timepoints, resulting in 24 replicates per sample for each of the five parameters. Standard deviations (SDs) and percent coefficients of variation (%CVs) were calculated for each sample. The results met the pre-defined acceptance criteria.
| CONCENTRATION, M/ml | | | | | | | | |
| --- | --- | --- | --- | --- | --- | --- | --- | --- |
| Sample # | N (total replicates) | Mean, M/mL | Within-run | | Between-run | | Between-operator/lot/instrument | |
| | | | SD | %CV | SD | %CV | SD | %CV |
| Sample 1 | 24 | 36.2 | 2.6 | 7.3% | 1.6 | 4.4% | 2.9 | 8.1% |
| Sample 2 | 24 | 36.7 | 1.9 | 5.2% | 1.5 | 4.1% | 2.2 | 6.0% |
| Sample 3 | 24 | 65.5 | 3.2 | 4.9% | 3.0 | 4.6% | 3.9 | 6.0% |
| Sample 4 | 24 | 103.5 | 6.9 | 6.5% | 7.7 | 7.4% | 11.6 | 11.2% |
| Sample 5 | 24 | 46.6 | 2.9 | 6.3% | 2.1 | 4.4% | 2.9 | 6.3% |
| Sample 6 | 24 | 7.8 | 1.2 | 15.4 | 1.7 | 21.8% | 1.8 | 23.1% |
| Sample 7 | 24 | 42.9 | 3.4 | 8.1% | 3.0 | 7.0% | 4.2 | 9.7% |
| Sample 8 | 24 | 6.0 | 0.6 | 10.0% | 1.5 | 25.0 | 1.2 | 20.0% |
| Sample 9 | 24 | 38.8 | 2.4 | 6.2% | 2.1 | 5.6% | 3.0 | 7.8% |
| Sample 10 | 24 | 9.8 | 1.2 | 12.2% | 1.7 | 17.3% | 1.5 | 15.3% |
| Sample 11 | 24 | 7.1 | 1.7 | 23.9% | 1.4 | 19.7% | 1.9 | 26.8% |
| Sample 12 | 24 | 104.2 | 17.7 | 17.1% | 6.7 | 6.3% | 16.2 | 15.6% |
| Sample 13 | 24 | 71.2 | 9.1 | 12.5% | 3.8 | 5.3% | 10.0 | 14.1% |
| Sample 14 | 24 | 65.7 | 7.5 | 11.3% | 4.8 | 7.4% | 7.6 | 11.6% |
| Sample 15 | 24 | 4.9 | 0.7 | 14.3% | 0.6 | 12.2 | 0.9 | 18.4% |
| Sample 16 | 24 | 14.8 | 1.1 | 7.4% | 1.8 | 12.4% | 1.6 | 10.5% |
K241628 - Page 7 of 12
{8}
K241628 - Page 8 of 12
| MSC, M/ml | | | | | | | | |
| --- | --- | --- | --- | --- | --- | --- | --- | --- |
| Sample # | N (total replicates) | Mean, M/mL | Within-run | | Between-run | | Between-operator/lot/instrument | |
| | | | SD | %CV | SD | %CV | SD | %CV |
| Sample 1 | 24 | 28.6 | 2.4 | 8.4% | 0.9 | 3.3% | 2.6 | 9.0% |
| Sample 2 | 24 | 32.7 | 1.8 | 5.4% | 1.4 | 4.2% | 2.1 | 6.4% |
| Sample 3 | 24 | 57.9 | 2.9 | 10.0% | 2.6 | 4.5% | 3.5 | 6.0% |
| Sample 4 | 24 | 74.7 | 2.3 | 3.1% | 5.5 | 7.6% | 4.5 | 6.0% |
| Sample 5 | 24 | 39.8 | 4.2 | 10.5% | 1.9 | 4.8% | 3.4 | 8.6% |
| Sample 6 | 24 | 5.4 | 0.7 | N/A | 1.1 | N/A | 1.0 | N/A |
| Sample 7 | 24 | 38.0 | 3.1 | 8.1% | 2.9 | 7.6% | 3.8 | 10.0% |
| Sample 8 | 24 | 3.6 | 0.2 | N/A | 0.9 | N/A | 0.4 | N/A |
| Sample 9 | 24 | 34.0 | 2.1 | 6.3% | 2.0 | 5.9% | 2.6 | 7.5% |
| Sample 10 | 24 | 7.4 | 0.9 | 12.2% | 1.5 | 20.3% | 1.3 | 17.6% |
| Sample 11 | 24 | 4.8 | 1.0 | N/A | 0.6 | N/A | 0.9 | N/A |
| Sample 12 | 24 | 72.3 | 7.2 | 9.9% | 2.8 | 3.9% | 6.6 | 9.2% |
| Sample 13 | 24 | 60.5 | 7.7 | 12.7% | 3.3 | 5.5% | 7.3 | 12.1% |
| Sample 14 | 24 | 47.1 | 5.2 | 11.1% | 6.3 | 13.5% | 4.8 | 10.3% |
| Sample 15 | 24 | 4.0 | 0.6 | N/A | 0.6 | N/A | 0.8 | N/A |
| Sample 16 | 24 | 10.4 | 1.1 | 10.7% | 0.8 | 7.8% | 1.2 | 11.5% |
| PMSC, M/ml | | | | | | | | |
| Sample # | N (total replicates) | Mean, M/mL | Within-run | | Between-run | | Between-operator/lot/instrument | |
| | | | SD | %CV | SD | %CV | SD | %CV |
| Sample 1 | 24 | 26.4 | 3.3 | 12.7% | 1.4 | 5.3% | 3.3 | 12.4% |
| Sample 2 | 24 | 31.1 | 1.7 | 5.4% | 1.3 | 4.3% | 2.0 | 6.4% |
| Sample 3 | 24 | 55.0 | 2.8 | 10.0% | 2.4 | 0.0% | 3.3 | 6.1% |
| Sample 4 | 24 | 70.2 | 3.0 | 4.3% | 6.5 | 9.4% | 4.6 | 6.6% |
| Sample 5 | 24 | 37.8 | 4.0 | 10.5% | 1.8 | 4.8% | 3.3 | 8.6% |
| Sample 6 | 24 | 2.4 | 0.8 | N/A | 0.8 | N/A | 1.0 | N/A |
| Sample 7 | 24 | 36.1 | 2.9 | 8.1% | 2.7 | 7.6% | 3.6 | 10.0% |
| Sample 8 | 24 | 0.7 | 0.3 | N/A | 0.9 | N/A | 0.4 | N/A |
| Sample 9 | 24 | 32.3 | 2.0 | 6.3% | 1.9 | 5.9% | 2.4 | 7.5% |
| Sample 10 | 24 | 4.5 | 0.9 | N/A | 1.2 | N/A | 1.1 | N/A |
| Sample 11 | 24 | 2.0 | 0.7 | N/A | 0.5 | N/A | 0.7 | N/A |
| Sample 12 | 24 | 69.0 | 7.5 | 10.9% | 2.3 | 3.3% | 6.2 | 9.0% |
| Sample 13 | 24 | 57.2 | 6.9 | 12.0% | 3.1 | 5.3% | 6.6 | 11.6% |
| Sample 14 | 24 | 43.6 | 5.9 | 13.6% | 7.4 | 17.1% | 5.7 | 13.1% |
| Sample 15 | 24 | 1.8 | 0.7 | N/A | 0.7 | N/A | 0.7 | N/A |
| Sample 16 | 24 | 6.4 | 1.5 | 23.4% | 1.3 | 20.3% | 1.6 | 25.0% |
{9}
| MOTILITY, % | | | | | | | | |
| --- | --- | --- | --- | --- | --- | --- | --- | --- |
| Sample # | N (total replicates) | Mean, M/mL | Within-run | | Between-run | | Between-operator/lot/instrument | |
| | | | SD | %CV | SD | %CV | SD | %CV |
| Sample 1 | 24 | 79.3 | 6.7 | 8.5% | 4.2 | 5.3% | 6.7 | 8.4% |
| Sample 2 | 24 | 89.2 | 1.3 | 1.4% | 1.2 | 1.4% | 1.8 | 2.1% |
| Sample 3 | 24 | 88.4 | 1.8 | 2.0% | 2.6 | 2.9% | 2.0 | 2.3% |
| Sample 4 | 24 | 72.7 | 3.7 | 5.1% | 6.7 | 9.2% | 5.0 | 6.9% |
| Sample 5 | 24 | 85.2 | 5.3 | 6.2% | 1.9 | 2.2% | 4.9 | 5.7% |
| Sample 6 | 24 | 69.4 | 10.0 | 14.4% | 5.3 | 7.6% | 11.0 | 15.9% |
| Sample 7 | 24 | 88.8 | 1.8 | 2.0% | 1.2 | 1.4% | 2.5 | 2.8% |
| Sample 8 | 24 | 61.4 | 6.4 | 10.4% | 11.2 | 18.2% | 10.7 | 17.4% |
| Sample 9 | 24 | 87.7 | 2.9 | 3.3% | 2.3 | 2.6% | 3.2 | 3.6% |
| Sample 10 | 24 | 74.9 | 5.1 | 6.8% | 5.4 | 7.2% | 6.8 | 9.1% |
| Sample 11 | 24 | 69.4 | 7.4 | 10.7% | 6.4 | 9.2% | 8.6 | 12.4% |
| Sample 12 | 24 | 70.3 | 6.0 | 8.5% | 3.3 | 4.8% | 6.4 | 9.2% |
| Sample 13 | 24 | 85.3 | 4.0 | 4.7% | 2.5 | 3.0% | 4.7 | 5.6% |
| Sample 14 | 24 | 71.6 | 3.2 | 4.5% | 6.6 | 9.3% | 5.1 | 7.1% |
| Sample 15 | 24 | 80.1 | 4.6 | 5.7% | 5.2 | 6.5% | 6.5 | 8.1% |
| Sample 16 | 24 | 71.2 | 6.2 | 8.7% | 8.4 | 11.8% | 7.4 | 10.4% |
| PROGRESSIVE MOTILITY, % | | | | | | | | |
| --- | --- | --- | --- | --- | --- | --- | --- | --- |
| Sample # | N (total replicates) | Mean, M/mL | Within-run | | Between-run | | Between-operator/lot/instrument | |
| | | | SD | %CV | SD | %CV | SD | %CV |
| Sample 1 | 24 | 73.1 | 9.3 | 12.7% | 5.7 | 7.7% | 9.1 | 12.5% |
| Sample 2 | 24 | 84.7 | 1.2 | 1.4% | 1.2 | 1.4% | 1.7 | 2.1% |
| Sample 3 | 24 | 84.0 | 1.8 | 2.1% | 2.5 | 3.0% | 1.9 | 2.3% |
| Sample 4 | 24 | 68.3 | 4.2 | 6.1% | 7.5 | 11.0% | 5.3 | 7.7% |
| Sample 5 | 24 | 81.0 | 5.1 | 6.3% | 1.8 | 2.2% | 4.7 | 5.8% |
| Sample 6 | 24 | 30.9 | 11.5 | 37.2% | 8.1 | 26.2% | 13.0 | 42.1% |
| Sample 7 | 24 | 84.4 | 1.7 | 2.0% | 1.2 | 1.4% | 2.4 | 2.8% |
| Sample 8 | 24 | 9.8 | 5.9 | N/A | 12.9 | N/A | 6.9 | N/A |
| Sample 9 | 24 | 83.3 | 2.7 | 3.3% | 2.2 | 2.6% | 3.0 | 3.7% |
| Sample 10 | 24 | 45.4 | 7.3 | 16.1% | 4.8 | 10.6% | 8.2 | 18.1% |
| Sample 11 | 24 | 28.6 | 8.0 | 28.0% | 5.0 | 17.5% | 9.7 | 33.9% |
| Sample 12 | 24 | 66.9 | 5.3 | 7.9% | 2.6 | 4.0% | 5.5 | 8.3% |
| Sample 13 | 24 | 80.7 | 3.3 | 4.1% | 2.7 | 3.4% | 4.2 | 5.1% |
| Sample 14 | 24 | 66.3 | 4.5 | 6.9% | 8.6 | 13.2% | 7.0 | 10.5% |
| Sample 15 | 24 | 34.6 | 11.4 | 32.9% | 11.6 | 33.5% | 10.4 | 30.1% |
| Sample 16 | 24 | 44.2 | 10.0 | 22.6% | 11.7 | 26.5% | 11.3 | 25.6% |
K241628 - Page 9 of 12
{10}
K241628 - Page 10 of 12
2. Linearity:
Linearity of the YO Home Sperm Test (YO 3.0) was evaluated for the sperm concentration using three YO 3.0 devices with one replicate per device. Native semen samples were prepared by pooling native samples from ten donors and concentrating sperm via centrifugation and removal of the supernatant. The sperm pellet was aliquoted into seminal plasma and the low-level concentration sample was confirmed by testing with the SQA-V comparator. The sperm concentration was demonstrated to be linear from $2 - 150\mathrm{M / mL}$.
3. Analytical Specificity/Interference:
The potential interference of various substances on YO 3 were evaluated by using two sperm concentration levels (20–60 M/mL and 60–100 M/mL). The parameters of sperm concentration, motile sperm concentration and progressively motile sperm concentration were evaluated. The following 11 interfering substances were tested in the study at the recommended level: yeast, E. coli, Vitamin B12, urine, saliva, red blood cells (RBCs), white blood cells (WBCs), testosterone, $\beta$-estradiol, D-norgestrel and agglutinates. Samples were compared to a native control sample composed of seminal plasma. Results showed that all tested interfering substances met the acceptance criteria, so no significant interference was caused by the tested substances at the evaluated level.
4. Accuracy (Instrument)
Method Comparison
The method comparison study was conducted at three U.S. sites and included 309 native semen samples with approximately 100 semen samples per site, assayed in singleton and in a blinded fashion by the intended users. The intended users analyzed their own samples. In addition, some female intended users were enrolled to test donor semen samples. In parallel, trained operators tested the sample on the SQA-V comparator device. The data were analyzed by the Passing-Bablok regression, and the results are shown in the table below. The results met the pre-defined acceptance criteria.
| Parameter | N | Range | Slope (95% CI) | Intercept (95% CI) | Correlation (95% CI) |
| --- | --- | --- | --- | --- | --- |
| Concentration, M/mL | 309 | 2.0 to 150.0 | 0.86 (0.82, 0.91) | 2.29 (1.29, 3.25) | 0.93 (0.92, 0.95) |
| Motility, % | 309 | 0.0 to 90.0 | 1.05 (1.00, 1.11) | 0.0 (0.00, 3.00) | 0.90 (0.88, 0.92) |
| Progressive motility, % | 309 | 0.0 to 86.0 | 1.24 (1.16, 1.31) | -0.47 (-2.78, 0.00) | 0.88 (0.85, 0.90) |
| MSC, M/mL | 309 | 0.0 to 94.5 | 0.92 (0.88, 0.95) | 1.84 (1.50, 2.20) | 0.94 (0.93, 0.95) |
| PMSC, M/mL | 309 | 0.0 to 84.0 | 1.03 (0.98, 1.07) | -0.04 (-0.44, 0.00) | 0.94 (0.92, 0.95) |
{11}
K241628 - Page 11 of 12
# B Expected Values/Reference Range:
The lower limit of the reference range for parameters Total Sperm Concentration, Total Motility $(\%)$ , and Progressive Motility $(\%)$ (including rapidly and slowly progressive motility) were directly reported as the $5^{\text{th}}$ percentile in the WHO Laboratory Manual for the Examination and Processing of Human Semen, $6^{\text{th}}$ Edition (Table 8.3). The lower limit of the reference range for parameters Motile Sperm Concentration and Progressive Motile Sperm Concentration (including Rapid and Slow PMSC) were calculated as the $5^{\text{th}}$ percentile based on the raw data from the study conducted by Campbell et al $^{1}$ .
| Reference Values for Semen Parameters | | |
| --- | --- | --- |
| SEMEN PARAMETER | REFERENCE RANGE | SOURCE |
| TOTAL SPERM CONCENTRATION, M/mL | ≥ 16 | WHO
6^{th} Edition |
| TOTAL MOTILITY (PR+NP), % | ≥ 42 | |
| PROGRESSIVE MOTILITY (PR), %
(RAPIDLY PROGRESSIVE + SLOWLY PROGRESSIVE MOTILITY) | ≥ 30 | |
| MOTILE SPERM CONCENTRATION (MSC), M/mL | ≥ 7.0 | |
| PROGRESSIVELY MOTILE SPERM CONCENTRATION
(PMSC), M/mL | ≥ 5.0 | |
# C Other Supportive Instrument Performance Characteristics Data:
1. **Sample Stability**
A sample stability study was performed to demonstrate stability of labile motility-related semen parameters in different time intervals. Ten random liquified native semen samples were assessed in duplicate by two operators using two YO 3 devices with two lots of testing YO slides and one SQA-V comparator device with two lots of testing capillaries in the following time intervals of sample collection: one hour, two hours, three hours, and four hours. Averages of motility-related semen parameters (total motility, MSC and PMSC) were assessed at the time intervals and their results were compared to their initial (T0) results. The study results support the recommendations as per the WHO 6th Edition to perform semen analysis on the YO 3 within one hour of collection.
2. **Detection Limit**
A study was conducted to determine the detection limits of the sperm concentration analyzed with the YO Home Sperm Test. Blank seminal plasma was used to represent a surrogate for the blank sample. Low level semen samples were prepared by pooling three donor semen samples and concentrating sperm via centrifugation and removal of the supernatant. The cell pellet was resuspended in seminal plasma to a concentration of approximately $2 - 8\mathrm{M / mL}$ as verified by microscope. Results were calculated and the detection limits were determined to be:
1. Campbell MJ, Lotti F, Baldi E, Schlatt S, Festin MP, Bjorndahl L et al. Distribution of Semen Examination Results 2020 - a follow up of data collated for the WHO semen analysis manual 2010. Andrology. 2021.
{12}
Limit of Blank (LoB) = 0.00 M/mL
Limit of Detection (LoD) = 1.74 M/mL
Limit of Quantitation (LoQ) = 1.74 M/mL
3. User Comprehension Study
As part of the method comparison study, the sponsor conducted a user comprehension study by issuing a post-test questionnaire and a user comprehension quiz. The post-test questionnaire consisted of six questions and was completed by 315 participants. The percentage distribution of results for responses of the 5-point Likert scale are reported. The results for all questions were in categories one and two approximately 98% of the time. The user comprehension quiz consisted of nine questions and was completed by 84 participants. The percentage distributions of results for responses are reported. Over 85% correct answers for all questions are reported.
VIII Proposed Labeling:
The labeling supports the finding of substantial equivalence for this device.
IX Conclusion:
The submitted information in this premarket notification is complete and supports a substantial equivalence decision.
K241628 - Page 12 of 12
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.