Reprocessed HARMONIC 1100 Shears, 5mm Diameter, 20cm Length (HAR1120); Reprocessed HARMONIC 1100 Shears, 5mm Diameter, 36cm Length (HAR1136)
K241606 · Stryker Sustainability Solutions · NLQ · Jul 3, 2024 · SU
Device Facts
| Record ID | K241606 |
| Device Name | Reprocessed HARMONIC 1100 Shears, 5mm Diameter, 20cm Length (HAR1120); Reprocessed HARMONIC 1100 Shears, 5mm Diameter, 36cm Length (HAR1136) |
| Applicant | Stryker Sustainability Solutions |
| Product Code | NLQ · SU |
| Decision Date | Jul 3, 2024 |
| Decision | SESE |
| Submission Type | Special |
| Device Class | Class U |
| Attributes | Therapeutic, Pediatric |
Indications for Use
The Reprocessed HARMONIC 1100 Shears are indicated for soft tissue incisions when bleeding control and minimal thermal injury are desired. The instruments can be used as an adjunct to or substitute for electrosurgery, lasers and steel scalpels in general, pediatric, gynecologic, urologic, thoracic, sealing and transection of lymphatic vessels, and other open and endoscopic procedures. The instruments allow for the coagulation of vessels up to and including 7 mm in diameter, using the Energy button with Advanced Hemostasis.
Device Story
Reprocessed ultrasonic surgical instrument; used for dissection, grasping, coagulation, and cutting of soft tissue. Input: electrical energy from GEN11 generator; output: mechanical ultrasonic vibration at blade tip. Operation: user-actuated handle triggers clamp arm to compress tissue against curved blade; Adaptive Tissue Technology monitors tissue impedance to modulate power output. Used in OR by surgeons for open/endoscopic procedures. Provides hemostasis for vessels up to 7 mm; reduces thermal injury compared to electrosurgery. Benefits: precise cutting, effective vessel sealing, minimal thermal spread.
Clinical Evidence
Bench testing and supplemental chronic 21-day survival animal studies conducted. Data confirms long-term seal quality for veins up to 7 mm in diameter, supporting the expanded indications for use.
Technological Characteristics
Ultrasonic surgical instrument; 5mm shaft diameter; 20cm or 36cm lengths. Features ergonomic handle, two energy buttons (standard and Advanced Hemostasis), and clamp/curved blade assembly. Utilizes Adaptive Tissue Technology for generator power modulation. Compatible with GEN11 generator (software v2018-1 or later). Sterile, single-patient use.
Indications for Use
Indicated for soft tissue incisions, coagulation, and vessel sealing in general, pediatric, gynecologic, urologic, and thoracic open or endoscopic procedures. Suitable for vessels up to 7 mm in diameter using Advanced Hemostasis mode.
Regulatory Classification
Identification
Single-use reprocessed versions of the ultrasonic devices used in surgical procedures either for fragmentation, emulsification and aspiration of soft tissue and hard tissue or for ligation of vessels (under product code LFL). Reprocessing validation data for this device type must be included in a 510(k) submission.. (70 FR 56911, available at https://www.govinfo.gov/content/pkg/FR-2005-09-29/pdf/05-19510.pdf).
Predicate Devices
- Reprocessed HARMONIC 1100 Shears, 5mm Diameter, 20cm (K233471)
- Harmonic 1100 Shears, 20cm length, Harmonic 1100 36cm length (K200841)
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
July 3, 2024
Stryker Sustainability Solutions Scott English scott.english(@stryker.com 1810 West Drake Drive Tempe, Arizona 85283
Re: K241606
Trade/Device Name: Reprocessed HARMONIC 1100 Shears, 5mm Diameter, 20cm Length (HAR1120); Reprocessed HARMONIC 1100 Shears, 5mm Diameter, 36cm Length (HAR1136) Regulatory Class: Unclassified Product Code: NLO, LFL Dated: June 4, 2024 Received: June 4, 2024
Dear Scott English:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrb/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}------------------------------------------------
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality System (OS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.70) and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE(@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely.
Digitally signed by Long H. Chen Long H. Chen-S ; Date: 2024.07.03 16:05:30 -04'00'
Long Chen. Ph.D. Assistant Director DHT4A: Division of General Surgery Devices
{2}------------------------------------------------
OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{3}------------------------------------------------
## Reprocessed Single-Use Device Models Subject to Clearance:
| OEM | Model<br>Number | Device<br>Name/Description | Shaft<br>Diameter | Shaft Length |
|---------|-----------------|----------------------------|-------------------|--------------|
| Ethicon | HAR1120 | HARMONIC 1100<br>Shears | 5 mm | 20 cm |
| Ethicon | HAR1136 | HARMONIC 1100<br>Shears | 5 mm | 36 cm |
{4}------------------------------------------------
#### Indications for Use
Form Approved: OMB No. 0910-0120 Expiration Date: 07/31/2026 See PRA Statement below.
Submission Number (if known)
K241606
Device Name
Reprocessed HARMONIC 1100 Shears, 5mm Diameter, 20cm Length (HAR1120); Reprocessed HARMONIC 1100 Shears, 5mm Diameter, 36cm Length (HAR1136)
#### Indications for Use (Describe)
The Reprocessed HARMONIC 1100 Shears are indicated for soft tissue incisions when bleeding control and minimal thermal injury are desired. The instruments can be used as an adjunct to or substitute for electrosurgery, lasers and steel scalpels in general, pediatric, gynecologic, urologic, thoracic, sealing and transection of lymphatic vessels, and other open and endoscopic procedures. The instruments allow for the coagulation of vessels up to and including 7 mm in diameter, using the Energy button with Advanced Hemostasis.
Type of Use (Select one or both, as applicable)
> Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{5}------------------------------------------------
| | K241606 | | Page 1/2 | |
|-----------------------------------------------------------|---------|-------------------------------------------------------------------------------------------------------------------------------------------------------|----------|-------------------------|
| | | 510(k) Summary | | Prepared on: 2024-06-04 |
| Contact Details | | | | 21 CFR 807.92(a)(1) |
| Applicant Name | | Stryker Sustainability Solutions | | |
| Applicant Address | | 1810 West Drake Drive Tempe AZ 85283 United States | | |
| Applicant Contact Telephone | | 9014511456 | | |
| Applicant Contact | | Mr. Scott English | | |
| Applicant Contact Email | | scott.english@stryker.com | | |
| Device Name | | | | 21 CFR 807.92(a)(2) |
| Device Trade Name | | Reprocessed HARMONIC 1100 Shears, 5mm Diameter, 20cm Length<br>(HAR1120);<br>Reprocessed HARMONIC 1100 Shears, 5mm Diameter, 36cm Length<br>(HAR1136) | | |
| Common Name | | Reprocessed HARMONIC 1100 Shears | | |
| Classification Name | | Single-Use Reprocessed Ultrasonic Surgical Instruments | | |
| Regulation Number | | N/A | | |
| Product Code(s) | | NLQ | | |
| Legally Marketed Predicate Devices<br>21 CFR 807.92(a)(3) | | | | |
| Predicate # | | Predicate Trade Name (Primary Predicate is listed first) | | Product Code |
| K233471 | | Reprocessed HARMONIC 1100 Shears, 5mm Diameter, 20cm | | NLQ |
| K200841 | | Harmonic 1100 Shears, 20cm length, Harmonic 1100 36cm length | | LFL |
## Device Description Summary
The Reprocessed HARMONIC 1100 Shears instrument is a sterile, single patient used for dissection, grasping, coagulation, and cutting between the blade and clamp arm. It consists of an ergonomic handle with an integrated hand piece and two energy delivery buttons.
21 CFR 807.92(a)(4)
1) Energy button for power levels 1-5, and
2) Energy with Advanced Hemostasis button for large vessel sealing.
The instrument is available in two shaft lengths (20 cm and 36 cm). An integrated audible and tactile mechanism in the grip housing indicates full trigger closure. The instrument has a clamp and coated curved blade that are designed to work through a 5 mm trocar, through a 5 mm reducer cap in a larger diameter trocar, or through an incision without the use of a trocar. The instrument shafts can be rotated continuously to facilitate visualization and access to targeted tissue. The instrument are intended to represent relative vessel size. The Energy button is indiameter. When the Energy button is used, cutting speed is the fastest. The energy button with Advanced Hemostasis is designed for larger vessels and is indicated for vessels up to 7 mm in diameter. When the Energy button with Advanced Hemostasis is reduced and hemostasis is maximized. The instrument utilizes Adaptive Tissue Technology. This provides the generator with the ability to identify and monitor the instrument during use, which enables the generator to modulate and adjust its power output as well as provide audible feedback to the user as appropriate. The HARMONC 1100 Shears instrument is designed for use exclusively with the Generator 11 (GEN11) software version
{6}------------------------------------------------
K241606
Page 2/2
#### 2018-1 or later.
# Intended Use/Indications for Use
The Reprocessed HARMONC 1100 Shears are indicated for soft tissue incisions when bleeding control and minimal thermal injury are desired. The instruments can be used as an adjunct to or electrosurgery, lasers and steel scalpels in general, pediatric, gynecologic, urologic, thoracic, sealing and transection of lymphatic vessels, and other open and endoscopic procedures. The instruments allow for the coagulation of vessels up to and including 7 mm in diameter, using the Energy button with Advanced Hemostasis.
# Indications for Use Comparison
The difference in indications for use does not constitute a new intended use. The predicate indications for use were limited to coagulation of vessels, up to 5 mm in diameter for veins and up to 7 mm in diameter for arteries, using the Energy button with Advanced Hemostasis. The subject device is indicated for for coagulation of vessels up to and including 7 mm in diameter, using the Energy button with Advanced Hemostasis. The inclusion of veins up to and including 7 mm in diameter is supported by additional chronic animal study data per the recommendation of FDA reviewers of K233471.
# Technological Comparison
There are no differences in the technological characteristics (i.e., design, material, chemical composition, principal of operation, energy source, etc.) The subject device is the same device as the predicate. The expansion of the indications for use.
#### Non-Clinical and/or Clinical Tests Summary & Conclusions 21 CFR 807.92(b)
Supplemental chronic 21-day survival studies were completed to assess the long-term seal quality of veins up to and including 7 mm in diameter. The data supports the expansion of the include coagulation of vessels up to and including 7 mm in diameter, using the Energy button with Advanced Hemostasis.
## 21 CFR 807.92(a)(5)
# 21 CFR 807.92(a)(6)
21 CFR 807.92(a)(5)