Unity Total Knee System

K241570 · Corin U.S.A. Limited · JWH · Sep 23, 2024 · Orthopedic

Device Facts

Record IDK241570
Device NameUnity Total Knee System
ApplicantCorin U.S.A. Limited
Product CodeJWH · Orthopedic
Decision DateSep 23, 2024
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 888.3560
Device ClassClass 2
AttributesTherapeutic

Intended Use

The Unity Total Knee System is intended for use in total knee arthroplasty in skeletally mature patients, to provide increased mobility and reduce pain by replacing the damaged knee joint articulation where there is evidence of sufficient sound bone to seat and support the components. The General total knee arthroplasty indications include: Painful, disabling joint disease of the knee resulting from: degenerative arthritis, rheumatoid arthritis or post-traumatic arthritis. Post-traumatic loss of knee joint configuration and function. Moderate varus, valgus, or flexion deformity in which the ligamentous structures can be returned to adequate function. Revision of previous unsuccessful knee replacement or other procedure, where soft tissue stability is adequate. Fracture of the distal femur and/or proximal tibia that cannot be stabilized by standard fracture management techniques. The posterior stabilized variant is also indicated for PCL instability requiring implant bearing surface geometries with increased anterior-posterior constraint and absent or non-functioning posterior cruciate ligament. The Unity Total Knee System is intended for cemented use, single use only.

Device Story

Fixed-bearing total knee replacement system; includes cobalt chromium femoral component, UHMWPE or Vitamin E-enriched (ECiMa) tibial insert, titanium alloy keel/augments, and UHMWPE patellar component. Used in primary or revision total knee arthroplasty. Surgeon replaces damaged knee joint articulation to restore mobility and reduce pain. System variants include Cruciate Retaining (CR) and Posterior Stabilized (PS); new Medial Constrained (MC) tibial insert provides high medial conformity and lower lateral conformity to improve stability. Device is cemented and single-use. Surgeon selects components based on patient anatomy and ligamentous stability. Benefits include pain relief and improved joint function.

Clinical Evidence

Bench testing only. Testing included constraint determination (ASTM F1223-20), range of motion CAD assessment (ISO 21536:2023), contact pressure analysis (ISO 21536:2023), dynamic disassociation testing (ASTM F2723-21), and usability testing. Micromotion and material characterization data were leveraged from previously cleared Unity and Apex Knee systems.

Technological Characteristics

Materials: Cobalt chromium alloy (femoral), UHMWPE or Vitamin E-enriched UHMWPE (tibial/patellar), titanium alloy (keel/augments). Design: Fixed-bearing, cemented, single-use. Standards: ASTM F1223-20 (constraint), ISO 21536:2023 (ROM/contact pressure), ASTM F2723-21 (disassociation). Connectivity: None (mechanical implant).

Indications for Use

Indicated for skeletally mature patients undergoing total knee arthroplasty for painful, disabling joint disease (degenerative, rheumatoid, or post-traumatic arthritis), post-traumatic loss of function, moderate varus/valgus/flexion deformity, revision of failed knee replacement, or distal femur/proximal tibia fractures. PS variant indicated for PCL instability/absence.

Regulatory Classification

Identification

A knee joint patellofemorotibial polymer/metal/polymer semi-constrained cemented prosthesis is a device intended to be implanted to replace a knee joint. The device limits translation and rotation in one or more planes via the geometry of its articulating surfaces. It has no linkage across-the-joint. This generic type of device includes prostheses that have a femoral component made of alloys, such as cobalt-chromium-molybdenum, and a tibial component or components and a retropatellar resurfacing component made of ultra-high molecular weight polyethylene. This generic type of device is limited to those prostheses intended for use with bone cement (§ 888.3027).

Predicate Devices

Reference Devices

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA logo on the right. The FDA logo is in blue and includes the letters "FDA" followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in a smaller font. September 23, 2024 Corin USA Limited Marine De Paoli Senior Regulatory Affairs Specialist 12750 Citrus Park Lane Tampa, Florida 33625 Re: K241570 Trade/Device Name: Unity Total Knee System Regulation Number: 21 CFR 888.3560 Regulation Name: Knee Joint Patellofemorotibial Polymer Semi-Constrained Cemented Prosthesis Regulatory Class: Class II Product Codes: JWH, OIY Dated: August 23, 2024 Received: August 23, 2024 Dear Marine De Paoli: We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. {1}------------------------------------------------ Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download). Your device is also subject to, among other requirements, the Quality System (OS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181). Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050. All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-device-advicecomprehensive-regulatory-assistance/unique-device-identification-system-udi-system. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems. For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatory {2}------------------------------------------------ assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100). Sincerely, Image /page/2/Picture/3 description: The image shows the name "Lixin Liu -S" in a large, bold, sans-serif font. The text is black against a white background. The letters are evenly spaced and easily readable. Lixin Liu, Ph.D Assistant Director DHT6A: Division of Joint Arthroplasty Devices OHT6: Office of Orthopedic Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health Enclosure {3}------------------------------------------------ #### Indications for Use 510(k) Number (if known) K241570 Device Name Unity Total Knee System #### Indications for Use (Describe) The Unity Total Knee System is intended for use in total knee arthroplasty in skeletally mature patients, to provide increased mobility and reduce pain by replacing the damaged knee joint articulation where there is sufficient sound bone to seat and support the components. The General total knee arthroplasty indications include: · Painful, disabling joint disease of the resulting from: degenerative arthritis, theumatoid arthritis or post-traumatic arthritis · Post-traumatic loss of knee joint configuration and function - · Moderate varus, valgus, or flexion deformity in which the ligamentous structures can be returned to adequate function - · Revision of previous unsuccessful knee replacement or other soft tissue stability is adequate - · Fracture of the distal femur and/or proximal tibia that cannot be standard fracture management techniques - · The posterior stabilized variant is also indicated for PCL instability requiring implant bearing surface with - increased anterior-posterior constraint and absent or non-functioning posterior cruciate ligament The Unity Total Knee System is intended for cemented use, single use only. | Type of Use (Select one or both, as applicable) | | |--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------| | <div style="display:inline-block;"><span> <svg fill="none" height="16" viewbox="0 0 16 16" width="16" xmlns="http://www.w3.org/2000/svg"> <path d="M13.6569 2.34315C14.0474 2.73368 14.0474 3.36637 13.6569 3.7569L3.7569 13.6569C3.36637 14.0474 2.73368 14.0474 2.34315 13.6569C1.95262 13.2664 1.95262 12.6337 2.34315 12.2431L12.2431 2.34315C12.6337 1.95262 13.2664 1.95262 13.6569 2.34315Z" fill="black"></path> <path d="M2.34315 2.34315C1.95262 2.73368 1.95262 3.36637 2.34315 3.7569L12.2431 13.6569C12.6337 14.0474 13.2664 14.0474 13.6569 13.6569C14.0474 13.2664 14.0474 12.6337 13.6569 12.2431L3.7569 2.34315C3.36637 1.95262 2.73368 1.95262 2.34315 2.34315Z" fill="black"></path> </svg> </span>Prescription Use (Part 21 CFR 801 Subpart D)</div> | <div style="display:inline-block;"><span> <svg fill="none" height="16" viewbox="0 0 16 16" width="16" xmlns="http://www.w3.org/2000/svg"> <rect height="15" rx="1.5" stroke="black" width="15" x="0.5" y="0.5"></rect> </svg> </span>Over-The-Counter Use (21 CFR 801 Subpart C)</div> | #### CONTINUE ON A SEPARATE PAGE IF NEEDED. This section applies only to requirements of the Paperwork Reduction Act of 1995. #### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: > Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." {4}------------------------------------------------ Image /page/4/Picture/0 description: The image shows the logo for Corin, a company focused on connected orthopedic insight. The word "Corin" is displayed in a bold, sans-serif font with rounded edges, and the phrase "Connected Orthopaedic Insight" is written in a smaller, lighter font below the company name. The logo is simple and modern, with a focus on clarity and readability. The colors used are a dark teal for the company name and a light green for the tagline. ## 510(k) Summary #### 510(k) #: K241570 Prepared on: 2024-09-19 ### Contact Details | Applicant Name: | Corin USA Limited | |------------------------------|-----------------------------------------------------| | Applicant Address: | 12750 Citrus Park Lane Tampa FL 33625 United States | | Applicant Contact Telephone: | +33456609245 | | Applicant Contact : | Ms. Marine De Paoli | | Applicant Contact Email : | marine.depaoli@coringroup.com | #### Device Name | Device Trade Name: | Unity Total Knee System | |----------------------|----------------------------------------------------------------------------------------------| | Common Name: | Knee prosthesis | | Classification Name: | Knee Joint Patellofemorotibial Polymer/Metal/Polymer Semi-Constrained<br>Cemented Prosthesis | | Regulation Number: | 21 CFR 888.3560 | | Product Code(s): | JWH, OIY | ### Legally Marketed Predicate Devices | Predicate # | Predicate Trade Name (Primary Predicate is listed first) | Product Code | |-------------|----------------------------------------------------------|---------------| | K183533 | Unity Total Knee System, Unity Knee CS Tibial Insert | JWH | | K113060 | Unity Total Knee System, Unity Knee CR Tibial Insert | JWH | | K150090 | Zimmer® Persona® Personalized Knee System, MC insert | JWH, OIY, MBH | ### Device Description Summary The Unity Total Knee System is a fixed bearing total knee replacement system that consists of a cobalt chromium alloy (CoCr) femoral component, a UHMWPE (Ultra-High-Molecular-Weight-Polyethylene) or ECiMa (Vitamin E enriched advanced polyethylene) tibial tray with a titanium alloy keel extension and all-polyethylene (UHMWPE) patellar component for use in primary and revision total knee arthroplasty. The Unity Total Knee System femoral component is provided in two variants, cruciate retaining (CR) and posterior stabilised (PS). - The CR femoral component is intended for use in conjunction with the CR tibial insert or Medial Constrained (MC) tibial insert where the posterior cruciate ligament (PCL) is functional or in conjunction with a Condylar Stabilised (CS) tibial insert or MC tibial insert where the PCL is present but is lax or non-functioning or when the PCL is absent. - The PS femoral component is intended for use in conjunction with the PS tibial insert where the posterior cruciate ligament is non-functioning or absent, resulting in joint instability, or in conjunction with the constrained posterior-stabilized (PS-C) tibial insert where is instability of the lateral collateral ligament and/or medial collateral ligament (LCL/MCL). {5}------------------------------------------------ Image /page/5/Picture/1 description: The image shows the logo for Corin, a company focused on connected orthopedic insight. The logo features the company name in a bold, sans-serif font, with a dark teal color scheme. Below the name, there is a tagline in green that reads "Connected Orthopaedic Insight." The Unity Total Knee System patellar component is optional and for use with either the CR or PS femoral variants and is intended for use where replacement of the articular surface of the patella is required. The system also provides titanium alloy augments including femoral augments, tibial augments, stem extensions and offset adapters. The Unity Total Knee System is intended for use in total knee arthroplasty in skeletally mature patients, to provide increased mobility and reduce pain by replacing the joint articulation where there is evidence of sufficient sound bone to seat and support the components. The Unity Total Knee system was originally cleared in K113060. This submission is for the inclusion of an additional range of Unity Knee MC Tibial Inserts. The additional range of tibial inserts has the same intended purpose as those cleared in the original 510(k) submission (K113060). ## Intended Use/Indications for Use The Unity Total Knee System is intended for use in total knee arthroplasty in skeletally mature patients, to provide increased mobility and reduce pain by replacing the damaged knee joint articulation where there is evidence of sufficient sound bone to seat and support the components. The General total knee arthroplasty indications include: - Painful, disabling joint disease of the knee resulting from: degenerative arthritis, rheumatoid arthritis or post-traumatic arthritis - . Post-traumatic loss of knee joint configuration and function - Moderate varus, valgus, or flexion deformity in which the ligamentous structures can be returned to adequate function - . Revision of previous unsuccessful knee replacement or other procedure, where soft tissue stability is adequate - Fracture of the distal femur and/or proximal tibia that cannot be stabilized by standard fracture management techniques - . The posterior stabilized variant is also indicated for PCL instability requiring implant bearing surface geometries with increased anterior-posterior constraint and absent or non-functioning posterior cruciate ligament The Unity Total Knee System is intended for cemented use, single use only. ### Indications for Use Comparison The subject Unity Total Knee System is identical to the predicate Unity Total Knee System (K183533, K113060) and Zimmer Personalized Knee System (K150090) in terms of intended use. The subject Unity Total Knee System indications for use are identical to the indications for use of the predicate Unity Total Knee System. The Unity Total Knee System indications for use are similar to the indications for use of the Zimmer Persona Personalized Knee System; they fall within the intended use of the predicate devices and the differences do not alter the intended therapeutic use of the subject device nor do they affect the safety and effectiveness of the device relative to the predicate. {6}------------------------------------------------ Image /page/6/Picture/1 description: The image shows the logo for Corin, a company that provides connected orthopedic insights. The word "Corin" is written in a bold, sans-serif font, with the letters in a dark teal color. Below the company name, the tagline "Connected Orthopaedic Insight" is written in a smaller, lighter green font. # Technological Comparison The Unity Knee MC tibial insert features a high conformity on the medial condyle, similar to that of the existing Unity Knee Cruciate Sacrificing (CS) tibial insert (K183533), and a lower conformity on the lateral side, very similar to that of the existing Unity Knee Cruciate Retaining (CR) tibial insert (K113060). Compared to the Unity Knee CS lateral condyle, the subject Unity Knee MC lateral condyle is less conforming. The Unity Knee CR medial condyle is less conforming compared to the subject Unity Knee MC medial condyle. The Unity Knee MC tibial insert is identical to the Unity Knee CS insert and Unity Knee CR insert in terms of connected devices, footprint sizes, size range and thickness. The Zimmer Persona Personalized Knee System Medial Congruent (MC) insert (K150090) has a similar asymmetrical tibial insert condyle shape that provides greater medial conformity and less lateral conformity. ## Non-Clinical and/or Clinical Tests Summary Non-clinical testing conducted to demonstrate substantial equivalence includes: - Constraint determination in accordance with ASTM F1223-20 - - -Range of motion CAD assessment in accordance with ISO 21536:2023 - -Contact pressure analysis in accordance with ISO 21536:2023 - -Dynamic disassociation testing in accordance with ASTM F2723-21 - -Usability testing A wear assessment rationale for the Unity Knee MC tibial insert was also conducted. The Unity Knee MC tibial insert is part of the Unity Total Knee System originally cleared in K113060. The micromotion assessment performed for the Unity Total Knee System is applicable to the Unity Knee MC tibial insert. The Unity Knee MC Tibial Insert is made of the same material as the Apex Knee ECiMa tibial insert (K201611). The material characterization performed for the ECiMa tibial insert is applicable to the Unity Knee MC tibial insert. ### Conclusions The results of the specific mechanical testing performed on the Unity Knee MC Tibial Insert show that the device is substantially equivalent to the predicate devices. A comparison of intended use and indications for use also demonstrated substantial equivalence.
Innolitics
510(k) Summary
Decision Summary
Classification Order
Enter a record ID and click Load to view the document.
100%