NUVISION™ Ultrasound Catheter; NUVISION™ NAV Ultrasound Catheter
K241540 · Biosense Webster, Inc. · OBJ · Jun 28, 2024 · Cardiovascular
Device Facts
Record ID
K241540
Device Name
NUVISION™ Ultrasound Catheter; NUVISION™ NAV Ultrasound Catheter
Applicant
Biosense Webster, Inc.
Product Code
OBJ · Cardiovascular
Decision Date
Jun 28, 2024
Decision
SESE
Submission Type
Special
Regulation
21 CFR 870.1200
Device Class
Class 2
Attributes
Pediatric
Indications for Use
The NUVISION™ Ultrasound Catheter is intended for intra-cardiac and intra luminal visualization of cardiac and great vessel anatomy and physiology as well as visualization of other devices in the heart. The NUVISION™ NAV Ultrasound Catheter is intended for intra-cardiac and intra-luminal visualization of cardiac and great vessel anatomy and physiology as well as visualization of other devices in the heart.
Device Story
Sterile, single-use intracardiac echocardiography (ICE) catheters; distal tip features multi-element 2D phased array-on-ASIC ultrasound transducer. Inputs: ultrasound signals; outputs: real-time 2D, 3D, and multiplane imaging. NUVISION compatible with GE Vivid S70N/E95 systems for 4D imaging. NUVISION NAV integrates with CARTO 3 EP Navigation System via 3D location sensor for location mapping. Operated by physicians in clinical settings during percutaneous procedures. Transducer array rotates independently of catheter shaft deflection. Output provides visual guidance for anatomy and device placement; assists clinical decision-making during interventions. Benefits: real-time visualization of cardiac structures and interventional devices.
Clinical Evidence
Bench testing only. No clinical data presented. Performance verified through non-clinical bench testing, confirming substantial equivalence in design, operation, and imaging capabilities.
Technological Characteristics
10F/10.5F diameter catheters; multi-element 2D phased array-on-ASIC ultrasound transducer at distal tip. Imaging modes: B-Mode (2D/3D), M-Mode, Pulsed/Continuous Wave Doppler, Color Doppler (2D/3D), Power Doppler. Connectivity: compatible with GE Vivid S70N/E95 systems and CARTO 3 EP Navigation System. Sterilization: EtO.
Indications for Use
Indicated for intracardiac and intra-luminal visualization of cardiac and great vessel anatomy, physiology, and other devices in the heart of adult and pediatric patients (NUVISION) or adult patients (NUVISION NAV). Intended for imaging guidance only, not treatment delivery, during cardiac interventional percutaneous procedures.
Regulatory Classification
Identification
An intravascular diagnostic catheter is a device used to record intracardiac pressures, to sample blood, and to introduce substances into the heart and vessels. Included in this generic device are right-heart catheters, left-heart catheters, and angiographic catheters, among others.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA text logo on the right. The text logo has the FDA acronym in a blue square, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue.
June 28, 2024
Biosense Webster, Inc. Sheba Chacko Senior Regulatory Affairs Program Lead 31 Technology Drive, Suite 200 Irvine, California 92618
Re: K241540
Trade/Device Name: NUVISION™ Utrasound Catheter; NUVISION™ NAV Ultrasound Catheter Regulation Number: 21 CFR 870.1200 Regulation Name: Diagnostic intravascular catheter Regulatory Class: Class II Product Code: OBJ Dated: May 30, 2024 Received: May 31, 2024
Dear Sheba Chacko:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}------------------------------------------------
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90. Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Aneesh S. Deoras -S
Aneesh Deoras Assistant Director Division of Cardiac Electrophysiology, Diagnostics and Monitoring Devices
{2}------------------------------------------------
Office of Cardiovascular Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{3}------------------------------------------------
# Indications for Use
510(k) Number (if known) K241540
Device Name NUVISION™ Ultrasound Catheter NUVISION™ NAV Ultrasound Catheter
### Indications for Use (Describe)
The NUVISION™ Ultrasound Catheter is indicated for intracardiac and intra-luminal visualization of cardiac and great vessel anatomy and physiology as well as visualization of other devices in the heart of adult and pediatric patients. The catheter is intended for imaging guidance only, not treatment delivery, during cardiac interventional percutaneous procedures.
The NUVISION™ NA V Utrasound Catheter and related accessory devices are indicated for intra-cardiac and intraluminal visualization of cardiac and great vessel anatomy and physiology as well as visualization of other devices in the heart. When used with the compatible CARTO™ 3 EP Navigation System, the catheter provides location information. The catheter is intended for imaging guidance only, not treatment delivery, during cardiac interventional percutaneous procedures.
| Type of Use (Select one or both, as applicable) | |
|----------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------|
| <span style="font-family: DejaVu Sans, sans-serif">☑</span> Prescription Use (Part 21 CFR 801 Subpart D) | <span style="font-family: DejaVu Sans, sans-serif">☐</span> Over-The-Counter Use (21 CFR 801 Subpart C) |
### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW! *
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{4}------------------------------------------------
## 510(k) Summary
This 510(k) Summary is being submitted in accordance with the requirements of SMDA 1990 and 21 CFR 807.92
| Date Summary Prepared | May 30, 2024 |
|------------------------|----------------------------------------------------------------------------------------------------------------------------|
| Applicant | Biosense Webster, Inc.<br>31 Technology Drive, Suite 200<br>Irvine, CA 92618<br>Establishment Registration Number: 9044811 |
| Official Correspondent | Sheba Chacko<br>Senior Regulatory Affairs Program Lead<br>Telephone: (949) 450-6058<br>Fax: (949) 450-6886 |
| Trade Name | NUVISION™ Ultrasound Catheter<br>NUVISION™ NAV Ultrasound Catheter |
| Common Name | Intracardiac Echocardiography Catheter |
| Classification Name | Diagnostic Intravascular Catheters |
| Device Classification | Class II, 21 CFR 870.1200 |
| FDA Product Code | OBJ |
| Predicate device(s) | NuVision ICE Catheter [K201775]<br>NUVISION™ NAV Ultrasound Catheter [K2237661] |
#### 1.1 Submitter | 21 CFR 807.92(a) (1)| and Device Information [21 CFR 807.92 (a) (2)|
#### 1.2 Predicate Device Information [21 CFR 807.92(a) (3)|
| Predicate Device Information | | | |
|--------------------------------------|-----------------------|----------|---------------|
| Predicate Device Name | Manufacturer | 510(k) # | Decision Date |
| NuVision ICE Catheter | Biosense Webster, Inc | K201775 | 05 Mar 2021 |
| NUVISION™ NAV<br>Ultrasound Catheter | Biosense Webster, Inc | K223766 | 16 Feb 2023 |
### 1.3 Description of the Device Subject to Premarket Notification [21 CFR 807.92(a) (4)]
The NUVISION™ Catheters are sterile, single use ultrasound catheters. They intended for intra-cardiac and intra-luminal visualization of cardiac and great vessel anatomy and physiology as well as visualization of other devices in the heart of patients. The catheters are intended for imaging guidance only, not treatment or therapy delivery, during cardiac interventional percutaneous procedures.
{5}------------------------------------------------
#### Intended Use [21 CFR 807.92(a) (5)] 1.4
Intended Use for NUVISION Ultrasound Catheter: The NUVISION™ Ultrasound Catheter is intended for intra-cardiac and intra luminal visualization of cardiac and great vessel anatomy and physiology as well as visualization of other devices in the heart.
Indications for Use for NUVISION™ Ultrasound Catheter: The NUVISION™ Ultrasound Catheter is indicated for intracardiac and intra-luminal visualization of cardiac and great vessel anatomy and physiology as well as visualization of other devices in the heart of adult and pediatric patients. The catheter is intended for imaging guidance only, not treatment delivery, during cardiac interventional percutaneous procedures.
Intended Use for NUVISION NAV Ultrasound Catheter: The NUVISION™ NAV Ultrasound Catheter is intended for intra-cardiac and intra-luminal visualization of cardiac and great vessel anatomy and physiology as well as visualization of other devices in the heart.
Indications for Use for NUVISION™ NAV Ultrasound Catheter: The NUVISION™ NAV Ultrasound Catheter and related accessory devices are indicated for intra-cardiac and intra-luminal visualization of cardiac and great vessel anatomy and physiology as well as visualization of other devices in the heart. When used with the compatible CARTO™ 3 EP Navigation System, the catheter provides location information. The catheter is intended for imaging guidance only, not treatment delivery, during cardiac interventional percutaneous procedures.
#### Summary of Substantial Equivalence [21 CFR 807.92 (a) (6)] 1.5
| Element of Comparison | Proposed/Subject Device | Predicate Device<br>[K201775] |
|-----------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Device Name | NUVISION™ Ultrasound<br>Catheter | NuVision ICE Catheter |
| 510(k) | Subject Device | K201775 |
| Manufacturer | Biosense Webster, Inc. | Biosense Webster, Inc. |
| Regulation | 21 CFR 870.1200 | 21 CFR 870.1200 |
| Classification | Class II | Class II |
| Product code | OBJ | OBJ |
| Indications for Use | The NUVISION™ Ultrasound<br>Catheter is indicated for<br>intracardiac and intra-luminal<br>visualization of cardiac and great<br>vessel anatomy and physiology as<br>well as visualization of other<br>devices in the heart of adult and<br>pediatric patients. The catheter is<br>intended for imaging guidance<br>only, not treatment delivery,<br>during cardiac interventional<br>percutaneous procedures. | The NUVISION™ Ultrasound<br>Catheter is indicated for<br>intracardiac and intra-luminal<br>visualization of cardiac and great<br>vessel anatomy and physiology as<br>well as visualization of other<br>devices in the heart of adult and<br>pediatric patients. The catheter is<br>intended for imaging guidance<br>only, not treatment delivery,<br>during cardiac interventional<br>percutaneous procedures. |
### NUVISION™ Ultrasound Catheter:
{6}------------------------------------------------
### NUVISION™ Ultrasound Catheter NUVISION™ NAV Ultrasound Catheter
| Patient Population | Adult and Pediatric | Adult and Pediatric |
|----------------------------------|-----------------------------|----------------------------------|
| Outside Diameter | 10F | 10 F |
| Imaging Modes | B-Mode (2D and 3D) | B-Mode (2D and 3D) |
| | M-Mode | M-Mode |
| | Doppler Pulsed Wave Doppler | Doppler Pulsed Wave Doppler |
| | Continuous Wave Doppler | Continuous Wave Doppler |
| | Color Doppler (2D and 3D) | Color Doppler (2D and 3D) |
| | Power Doppler | Power Doppler |
| | Acoustic Array | Multi-element 2D phased array- |
| on-ASIC ultrasound transducer at | | on-ASIC ultrasound transducer at |
| distal tip | | distal tip |
| Sterilization Method | EtO | EtO |
## NUVISION NAV Ultrasound Catheter:
| Element of Comparison | Proposed/Subject Device | Predicate Device [K223766] |
|-----------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Device Name | NUVISION™ NAV Ultrasound Catheter | NUVISION™ NAV Ultrasound Catheter |
| 510(k) | Subject Device | K223766 |
| Manufacturer | Biosense Webster, Inc. | Biosense Webster, Inc. |
| Regulation | 21 CFR 870.1200 | 21 CFR 870.1200 |
| Classification | Class II | Class II |
| Product code | OBJ | OBJ |
| Indications for Use | The NUVISION™ NAV Ultrasound Catheter and related accessory devices are indicated for intra-cardiac and intra-luminal visualization of cardiac and great vessel anatomy and physiology as well as visualization of other devices in the heart. When used with the compatible CARTO™ 3 EP Navigation System, the catheter provides location information. The catheter is intended for imaging guidance only, not treatment delivery, during cardiac interventional percutaneous procedures. | The NUVISION™ NAV Ultrasound Catheter and related accessory devices are indicated for intra-cardiac and intraluminal visualization of cardiac and great vessel anatomy and physiology as well as visualization of other devices in the heart. When used with the compatible CARTO™ 3 EP Navigation System, the catheter provides location information. The catheter is intended for imaging guidance only, not treatment delivery, during cardiac interventional percutaneous procedures |
| Patient Population | Adult | Adult |
| Outside Diameter | 10.5 F | 10 F |
| Imaging Modes | B-Mode (2D and 3D)<br>M-Mode<br>Doppler Pulsed Wave Doppler<br>Continuous Wave Doppler<br>Color Doppler (2D and 3D)<br>Power Doppler | B-Mode (2D and 3D)<br>M-Mode<br>Doppler Pulsed Wave Doppler<br>Continuous Wave Doppler<br>Color Doppler (2D and 3D)<br>Power Doppler |
| Acoustic Array | Multi-element 2D phased array-on-ASIC ultrasound transducer at distal tip | Multi-element 2D phased array-on-ASIC ultrasound transducer at distal tip |
| Sterilization Method | EtO | EtO |
{7}------------------------------------------------
### NUVISION™ Ultrasound Catheter NUVISION™ NAV Ultrasound Catheter
#### 1.6 Performance Testing: [21 CFR 807.92(b)(1)]
The NUVISION Ultrasound Catheter and NUVISION NAV Ultrasound Catheter are sterile, single-use, disposable intracardiac echo (ICE) ultrasound imaging catheters. The distal end of the catheter has an ultrasound transducer with 2D acoustic element array on ASIC enabling real time 2D, 3D, and multiplane imaging. NUVISION Ultrasound Catheter is compatible with GE Vivid™ S70N and GE E95 Ultrasound System to enable 4D intra-cardiac imaging. While the NUVISION NAV Ultrasound Catheter's grid transducer combines the processing power of the GE Vivid™ S70N Ultrasound System to enable 4D intra-cardiac imaging and when coupled with a 3D location sensor enables location mapping in a compatible CARTO 3 System. The transducer array can be rotated independently of the deflection plane of the catheter shaft. The NUVISION NAV Ultrasound Catheter adds integration into the CARTO environment. Both catheters met all acceptance criteria in accordance with appropriate test criteria and standards.
#### 1.7 Conclusion on Safety and Effectiveness [21 CFR 807.92(b) (3)|
The NUVISION Ultrasound Catheter and NUVISION NAV Ultrasound Catheter are substantially equivalent to its currently cleared predicate devices based on the successful completion of nonclinical bench testing, as well as the technological comparison exhibiting similar principles of design, operation, and indications for use.