← Product Code [OUJ](/productcode/OUJ) · K241536

# trophon Wireless Ultrasound Probe Holder (K241536)

_Nanosonics Limited · OUJ · Sep 27, 2024 · Radiology · SESE_

**Canonical URL:** https://fda.innolitics.com/device/K241536

## Device Facts

- **Applicant:** Nanosonics Limited
- **Product Code:** [OUJ](/productcode/OUJ.md)
- **Decision Date:** Sep 27, 2024
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 892.1570
- **Device Class:** Class 2
- **Review Panel:** Radiology

## Indications for Use

The trophon Wireless Ultrasound Probe Holder is an accessory to be added to the trophon EPR and trophon2 to allow wireless ultrasound probes to be disinfected. The trophon EPR is designed to provide High-Level Disinfection of ultrasound transducers. The system uses the trophon Disinfectant which is intended to be used exclusively with the trophon EPR device. The trophon Disinfectant is intended for use as a High-Level Disinfectant to be used exclusively with the trophon EPR for the High-Level Disinfection of ultrasound transducers. The trophon EPR is suitable for use in general hospital and health care facilities by trained personnel. The trophon EPR system consists of a multiple use instrument combined with a single use disinfectant, delivered from a multi-dose cartridge. The trophon Disinfectant should be used with the following contact conditions: Minimum Operational Cycle Time: 7 minutes Minimum Concentration: 31.5% Minimum Disinfectant Dose: 1.0 g Minimum Chamber Temperature: 56℃. The trophon2 is designed to provide High-Level Disinfection (HLD) of validated ultrasound transducers. High-Level Disinfection is achieved by surface exposure to a controlled dose of hydrogen peroxide mist delivered to a disinfection chamber contain the ultrasound probe. The trophon2 system consists of a multiple use instrument combined with a single use disinfectant "trophon Sonex-HL", delivered from a multi-dose cartridge. The trophon2 is suitable for use in general hospital and health care facilities by trained personnel. The trophon Sonex-HL should be used with the following contact conditions: Minimum Operational Cycle Time: 7 minutes Minimum Concentration: 31.5% Minimum Disinfectant Dose: 1.0 g Minimum Chamber Temperature: 56°C

## Device Story

Accessory for trophon EPR and trophon2 disinfection systems; enables high-level disinfection (HLD) of wireless ultrasound probes. Device holds wireless probe in place within existing disinfection chamber using cable clamp mechanism. System operates via automated cycle: nebulizer generates hydrogen peroxide mist; mist contacts probe surfaces for specified time, temperature, and dosage; automated aeration removes disinfectant. Used in hospitals/clinics by trained personnel. Output is a disinfected probe ready for immediate clinical use. Benefits include expanded compatibility for wireless transducers while maintaining established HLD efficacy and safety standards.

## Clinical Evidence

No clinical data. Bench testing only, including HLD efficacy (Mycobacterium terrae), residual H2O2 assessment, material compatibility (5000 cycles), temperature monitoring, leak testing, cleaning validation, and human factors validation.

## Technological Characteristics

Accessory for trophon EPR/trophon2. Operates via hydrogen peroxide mist (35% concentration) generated by nebulizer. Includes integrated catalytic destruct system. Software-controlled automated cycle. Complies with IEC 61010-1, IEC 61010-2-040, IEC 61326, IEC 62304, ISO 62366, ISO 10993-1, and ISO 14971.

## Regulatory Identification

A diagnostic ultrasonic transducer is a device made of a piezoelectric material that converts electrical signals into acoustic signals and acoustic signals into electrical signals and intended for use in diagnostic ultrasonic medical devices. Accessories of this generic type of device may include transmission media for acoustically coupling the transducer to the body surface, such as acoustic gel, paste, or a flexible fluid container.

## Predicate Devices

- trophon2 ([K173865](/device/K173865.md))

## Reference Devices

- trophon EPR ([K103059](/device/K103059.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.

September 27, 2024

Nanosonics Limited Nancy Kaiser Regulatory Affairs Manager 7-11 Talavera Road Macquarie Park, NSW 2113 Australia

Re: K241536

Trade/Device Name: trophon® Wireless Ultrasound Probe Holder Regulation Number: 21 CFR 892.1570 Regulation Name: Diagnostic Ultrasonic Transducer Regulatory Class: Class II Product Code: OUJ Dated: May 30, 2024 Received: August 29, 2024

Dear Nancy Kaiser:

We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device"

{1}------------------------------------------------

(https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).

Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30. Design controls; 21 CFR 820.90. Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review. the OS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.

All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rue"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advicecomprehensive-regulatory-assistance/unique-device-identification-system-udi-system.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

{2}------------------------------------------------

Sincerely,

Image /page/2/Picture/3 description: The image shows the name "Katharine Segars" followed by "-S". The text is in a large, clear font, making it easily readable. The background is plain, which helps the text stand out.

Katharine Segars, Ph.D. Assistant Director DHT4C: Division of Infection Control Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{3}------------------------------------------------

# Indications for Use

## 510(k) Number (if known) K241536

### Device Name

trophon® Wireless Ultrasound Probe Holder

### Indications for Use (Describe)

The trophon Wireless Ultrasound Probe Holder is an accessory to be added to the trophon EPR and trophon2 to allow wireless ultrasound probes to be disinfected.

trophon EPR:

The trophon EPR is designed to provide High-Level Disinfection of ultrasound transducers. The system uses the trophon Disinfectant which is intended to be used exclusively with the trophon EPR device.

The trophon Disinfectant is intended for use as a High-Level Disinfectant to be used exclusively with the trophon EPR for the High-Level Disinfection of ultrasound transducers.

The trophon EPR is suitable for use in general hospital and health care facilities by trained personnel.

The trophon EPR system consists of a multiple use instrument combined with a single use disinfectant, delivered from a multi-dose cartridge.

The trophon Disinfectant should be used with the following contact conditions:

Minimum Operational Cycle Time: 7 minutes Minimum Concentration: 31.5% Minimum Disinfectant Dose: 1.0 g Minimum Chamber Temperature: 56℃

trophon2:

The trophon2 is designed to provide High-Level Disinfection (HLD) of validated ultrasound transducers. High-Level Disinfection is achieved by surface exposure to a controlled dose of hydrogen peroxide mist delivered to a disinfection chamber contain the ultrasound probe.

The trophon2 system consists of a multiple use instrument combined with a single use disinfectant "trophon Sonex-HL", delivered from a multi-dose cartridge.

The trophon2 is suitable for use in general hospital and health care facilities by trained personnel.

The trophon Sonex-HL should be used with the following contact conditions: Minimum Operational Cycle Time: 7 minutes Minimum Concentration: 31.5% Minimum Disinfectant Dose: 1.0 g Minimum Chamber Temperature: 56°C Type of Use (Select one or both, as applicable)

| Prescription Use (Part 21 CFR 801 Subpart D)

|X Over-The-Counter Use (21 CFR 801 Subpart C)

## CONTINUE ON A SEPARATE PAGE IF NEEDED.

{4}------------------------------------------------

This section applies only to requirements of the Paperwork Reduction Act of 1995.

## *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{5}------------------------------------------------

Image /page/5/Picture/0 description: The image shows the logo for Nanosonics. The word "nanosonics" is written in a sans-serif font, with "nano" in blue and "sonics" in gray. A blue arc is above the word "nano". Below the word "nanosonics" is the phrase "Infection Prevention. For Life." in a smaller font.

# 510(k) SUMMARY - trophon® Wireless Ultrasound Probe Holder

### I. DATE PREPARED September 24, 2024

II. 510(k) NUMBER K241536

#### III. 510(k) SUBMITTER

Nanosonics Limited 7-11 Talavera Road Macquarie Park NSW 2113 Australia

Contact Person:

Nancy Kaiser, Regulatory Affairs Manager Address: 7-11 Talavera Road Macquarie Park, NSW 2113 Australia Email: n.kaiser@nanosonics.com Telephone: (317) 854-7625

#### IV. DEVICE

| Trade Name of Device: | trophon® Wireless Ultrasound Probe Holder.                                       |
|-----------------------|----------------------------------------------------------------------------------|
| Common or Usual Name: | High Level Disinfection Reprocessing Instrument for Ultrasonic Transducers, Mist |
| Classification:       | II                                                                               |
| Regulation Number:    | 21 CFR 892.1570                                                                  |
| Product Code:         | OUJ                                                                              |

#### V. PREDICATE DEVICE

| Predicate Device |                 |
|------------------|-----------------|
| Trade Name:      | trophon2        |
| 510(k) Number:   | K173865         |
| Company Name:    | Nanosonics Ltd. |

## Reference Device

| Trade Name:    | trophon EPR     |
|----------------|-----------------|
| 510(k) Number: | K103059         |
| Company Name:  | Nanosonics Ltd. |

#### VI. DEVICE DESCRIPTION

The trophon Wireless Ultrasound Probe Holder is an accessory to be added to the trophon EPR and trophon2 to allow wireless ultrasound probes to be disinfected.

The Nanosonics trophon EPR and trophon2 are software-controlled devices which provide High- Level Disinfection of ultrasound transducers. The device consists of a sealed disinfection chamber and operates in conjunction with a multi-dose cartridge of concentrated hydrogen peroxide disinfectant, supplied as an accessory to the device. Precleaned and dried ultrasound transducers are placed within the trophon EPR/trophon2

{6}------------------------------------------------

Image /page/6/Picture/0 description: The image shows the logo for Nanosonics. The word "nanosonics" is written in a combination of blue and gray letters. The word "nano" is in blue, and the word "sonics" is in gray. Below the word "nanosonics" is the phrase "Infection Prevention. For Life." in a smaller font.

chamber and disinfected by means of an automated disinfection and aeration cycle. The disinfected ultrasound transducer is removed from the chamber and is ready for immediate use.

#### VII. INDICATIONS FOR USE

The trophon Wireless Ultrasound Probe Holder is an accessory to be added to the trophon EPR and trophon2 to allow wireless ultrasound probes to be disinfected.

trophon EPR:

The trophon EPR is designed to provide High-Level Disinfection of ultrasound transducers. The system uses the trophon Disinfectant which is intended to be used exclusively with the trophon EPR device.

The trophon Disinfectant is intended for use as a High-Level Disinfectant to be used exclusively with the trophon EPR for the High-Level Disinfection of ultrasound transducers.

The trophon EPR is suitable for use in general hospital and health care facilities by trained personnel.

The trophon EPR system consists of a multiple use instrument combined with a single use disinfectant, delivered from a multi-dose cartridge.

The trophon Disinfectant should be used with the following contact conditions: Minimum Operational Cycle Time: 7 minutes Minimum Concentration: 31.5% Minimum Disinfectant Dose: 1.0 g Minimum Chamber Temperature: 56°C

## trophon2:

The trophon2 is designed to provide High-Level Disinfection (HLD) of validated ultrasound transducers. High-Level Disinfection is achieved by surface exposure to a controlled dose of hydrogen peroxide mist delivered to a disinfection chamber contain the ultrasound probe.

The trophon2 system consists of a multiple use instrument combined with a single use disinfectant "trophon Sonex-HL", delivered from a multi-dose cartridge.

The trophon2 is suitable for use in general hospital and health care facilities by trained personnel.

The trophon Sonex-HL should be used with the following contact conditions: Minimum Operational Cycle Time: 7 minutes Minimum Concentration: 31.5% Minimum Disinfectant Dose: 1.0 g Minimum Chamber Temperature: 56°C

{7}------------------------------------------------

Image /page/7/Picture/0 description: The image shows the logo for Nanosonics. The word "nanosonics" is written in a sans-serif font, with "nano" in blue and "sonics" in gray. Below the word "nanosonics" is the phrase "Infection Prevention. For Life." in a smaller, sans-serif font. A blue arc is above the word "nanosonics".

### VIII. COMPARISON OF TECHNOLOGICAL CHARACTERISTICS WITH THE PREDICATE DEVICE

The trophon2 and the trophon EPR with the proposed Ultrasound Probe Holder are as safe and as effective in terms of technological characteristics and principles of operation to the predicate device, trophon2 and reference device, trophon EPR without the Ultrasound Probe Holder.

Like the predicate and reference devices, the trophon2 and trophon EPR still achieve HLD by having the probe suspended in their chamber, the Probe Holder is held in place by the cable clamp during high level disinfection to achieve this positioning for a wireless ultrasound probe. During each disinfection cycle, a controlled quantity of hydrogen peroxide mist is generated by a nebulizer and is delivered to the trophon EPR/trophon2 chamber. In the chamber, the mist directly contacts the wired or wireless ultrasound probe covering its surface. Contact occurs for a specified time, temperature, and liquid/mist dosage which determines the germicidal efficacy. The disinfected wired or wireless ultrasound transducer is removed from the chamber and is ready for immediate use. Design verification testing demonstrates that the proposed Wireless Ultrasound Probe Holder meets the product specifications and validation testing confirmed, intended use and user needs continue to be met.

| Feature                                                           | Table 1. A Comparison between the Subject and Predicate Device<br>Subject Device:<br>trophon2 (K241536)                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                   | Predicate Device:<br>trophon2 (K173865)                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                     | Comparison                                                                                                                              |
|-------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------|
| Manufacturer                                                      | Nanosonics Limited                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                        | Nanosonics Limited                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                          | Same                                                                                                                                    |
| Regulation Number                                                 | 21 CFR 892.1570                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                           | 21 CFR 892.1570                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                             | Same                                                                                                                                    |
| Product Code                                                      | OUJ                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                       | OUJ                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                         | Same                                                                                                                                    |
| Indication for<br>Use/Intended Use                                | The trophon Wireless Ultrasound<br>Probe Holder is an accessory to be<br>added to the trophon EPR and<br>trophon2 to allow wireless ultrasound<br>probes to be disinfected.<br>The trophon2 is designed to provide<br>High-level Disinfection (HLD) of<br>validated ultrasound probes, High-<br>Level Disinfection is achieved by<br>surface exposure to a controlled dose<br>of hydrogen peroxide mist delivered to<br>a disinfection chamber containing the<br>ultrasound probe.<br>The trophon2 is suitable for use in<br>general hospital and health care<br>facilities by trained personnel.<br>The trophon Sonex-HL should be<br>used with the following contact<br>conditions:<br>Minimum Operational Cycle Time: 7<br>minutes<br>Minimum Concentration: 31.5%<br>Minimum Disinfectant Dose: 1.0 g<br>Minimum Chamber Temperature: | The trophon2 is designed to provide<br>High-Level Disinfection (HLD) of<br>validated ultrasound transducers. High-<br>Level Disinfection is achieved by<br>surface exposure to a controlled dose of<br>hydrogen peroxide mist delivered to a<br>disinfection chamber contain the<br>ultrasound probe.<br>The trophon2 system consists of a<br>multiple use instrument combined with a<br>single use disinfectant "trophon Sonex-<br>HL", delivered from a multi-dose<br>cartridge.<br>The trophon2 is suitable for use in<br>general hospital and health care facilities<br>by trained personnel.<br>The trophon Sonex-HL should be used<br>with the following contact conditions:<br>Minimum Operational Cycle Time: 7<br>minutes<br>Minimum Concentration: 31.5%<br>Minimum Disinfectant Dose: 1.0 g<br>Minimum Chamber Temperature: 56°C | Same                                                                                                                                    |
| Operating Principle                                               | Software controlled systems<br>that deliver measured doses of<br>hydrogen peroxide disinfectant<br>to achieve High -Level<br>Disinfection (HLD)                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                           | Software controlled systems that<br>deliver measured doses of<br>hydrogen peroxide disinfectant to<br>achieve High -Level Disinfection<br>(HLD)                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                             | Same                                                                                                                                    |
| Disinfectant                                                      | trophon Sonex-HL (35%<br>hydrogen peroxide in<br>cartridge)                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                               | trophon Sonex-HL (35%<br>hydrogen peroxide in cartridge)                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                    | Same                                                                                                                                    |
| Disinfectant Delivery                                             | Liquid Aerosol Mist                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                       | Liquid Aerosol Mist                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                         | Same                                                                                                                                    |
| Disinfectant Removal<br>Process                                   | Automate Aeration                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                         | Automate Aeration                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                           | Same                                                                                                                                    |
| Process Monitoring                                                | Automated process monitoring<br>in the device                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                             | Automated process monitoring in<br>the device                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                               | Same                                                                                                                                    |
| Chemical Indicator<br>Required?                                   | Yes, trophon chemical<br>indicator (cleared in K103126)                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                   | Yes, trophon chemical indicator<br>(cleared in K103126)                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                     | Same                                                                                                                                    |
| Microbiology/Efficacy<br>AOAC Performance<br>Standards            | Meets AOAC Methods                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                        | Meets AOAC Methods                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                          | Same                                                                                                                                    |
| Device Performance<br>Standards                                   | IEC 601010-1<br>IEC 61010-2-040<br>IEC 61326<br>IEC 62304<br>ISO 62366 -1 and -2<br>ISO 10993-1<br>ISO 14971                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                              | IEC 601010-1<br>IEC 61010-2-040<br>IEC 61326<br>IEC 62304<br>ISO 62366 -1 and -2<br>ISO 10993-1<br>ISO 14971                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                | Same                                                                                                                                    |
| Residue Testing                                                   | Effectively removes residues<br>from disinfected transducers                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                              | Effectively removes residues<br>from disinfected transducers                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                | Same                                                                                                                                    |
| Chamber Design                                                    | Regular shaped chamber                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                    | Irregular shaped chamber                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                    | Same                                                                                                                                    |
| Chamber Temperature                                               | Set point 65°C                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                            | Set point 75°C                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                              | Same                                                                                                                                    |
| Door Lock                                                         | Motor and hook assembly                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                   | Motor and hook assembly                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                     | Same                                                                                                                                    |
| Probe Clamp                                                       | Spring loaded cleats                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                      | Spring loaded cleats                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                        | Same                                                                                                                                    |
| All in One Catalytic<br>Destruct                                  | An integrated catalytic<br>destruct system                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                | An integrated catalytic destruct<br>system                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                  | Same                                                                                                                                    |
| Touch Screen                                                      | Color touch screen panel                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                  | Color touch screen panel                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                    | Same                                                                                                                                    |
| Software/Firmware                                                 | Multiple software/firmware<br>components which are<br>deployed on two PCBAs                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                               | Multiple software/firmware<br>components which are deployed<br>on two PCBAs                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                 | Same                                                                                                                                    |
| Traceability                                                      | An integrated RFID module<br>allowing automated<br>traceability features in the<br>device.<br>Patient information is not<br>received or recorded by the<br>device; therefore, it cannot be<br>accessed via RFID.                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                          | An integrated RFID module<br>allowing automated traceability<br>features in the device.<br>Patient information is not<br>received or recorded by the<br>device; therefore, it cannot be<br>accessed via RFID.                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                               | Same                                                                                                                                    |
| Communication Ports                                               | 3 USB ports to connect to<br>external device (i.e. printer)<br>1 Ethernet port<br>The trophon2 will be able to<br>connect externally to a                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                 | 3 USB ports to connect to<br>external device (i.e. printer)<br>1 Ethernet port<br>The trophon2 will be able to<br>connect externally to a network                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                           | Same                                                                                                                                    |
|                                                                   | network via the Ethernet port                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                             | via the Ethernet port                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                       |                                                                                                                                         |
| Key Accessories                                                   | trophon AcuTrace Operator<br>Card<br>trophon AcuTrace Medical<br>Instrument Tag<br>trophon AcuTrace PLUS<br>activation Card                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                               | trophon AcuTrace Operator Card<br>trophon AcuTrace Medical<br>Instrument Tag<br>trophon AcuTrace PLUS<br>activation Card<br>trophon Sonex-HL                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                | Same<br>An optional<br>Wireless<br>Ultrasound<br>Probe Holder                                                                           |
| Image: Infection Prevention. For Life.                            | trophon Sonex-HL                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                          |                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                             | (provided<br>separately) to<br>provide high<br>level<br>disinfection for<br>wireless GE<br>ultrasound<br>probe (model:<br>Vscan Air CL) |
| Intended to provide high<br>level disinfection to<br>Wired Probes | Yes                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                       | Yes                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                         | Same<br>Wired Probes<br>and Validated<br>Wireless GE<br>Probe (model:<br>Vscan Air CL)                                                  |

# Table 1. A Comparison between the Subject and Predicate Device

{8}------------------------------------------------

Image /page/8/Picture/0 description: The image shows the logo for Nanosonics. The word "nanosonics" is written in a combination of blue and gray colors. A blue arc is above the word. Below the word "nanosonics" is the phrase "Infection Prevention. For Life."

{9}------------------------------------------------

Image /page/9/Picture/0 description: The image shows the logo for "nanosonics". The word "nano" is in blue, and the word "sonics" is in gray. There is a blue arc above the word "nano". Below the word "nanosonics" is the phrase "Infection Prevention. For Life."

# Table 2. A Comparison between the Subject and Reference Device

| Feature                                                           | Subject Device:<br>trophon EPR (K241536)                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                    | Reference Device:<br>trophon EPR (K103059)                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                       | Comparison                                                                                                                                                      |
|-------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Nanosonics Limited                                                | Nanosonics Limited                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                          | Nanosonics Limited                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                               | Same                                                                                                                                                            |
| Regulation Number                                                 | 21 CFR 892.1570                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                             | 21 CFR 892.1570                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                  | Same                                                                                                                                                            |
| Product Code                                                      | OUJ                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                         | OUJ                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                              | Same…

---

**Source:** [https://fda.innolitics.com/device/K241536](https://fda.innolitics.com/device/K241536)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
