K241492 · JJGC Indústria e Comércio de Materiais Dentários S.A. · KCT · Aug 23, 2024 · General Hospital
Device Facts
Record ID
K241492
Device Name
Guided Surgery Kit Cases
Applicant
JJGC Indústria e Comércio de Materiais Dentários S.A.
Product Code
KCT · General Hospital
Decision Date
Aug 23, 2024
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 880.6850
Device Class
Class 2
Indications for Use
Zirconia Guided Surgery Kit Cases are intended to be used to enclose other medical devices that are to be sterilized by a health care provider. Zirconia Guided Surgery Kit Cases are intended to allow sterilization of the enclosed medical devices. Zirconia Guided Surgery Kit Cases requre the use of FDA-cleared wrap to maintain the sterility of the enclosed devices. The kits are to be enclosed in a sterlization wrap that is FDA-cleared for the indicated cycles, and moist heat (steam) sterilized using one of the following cycles: Dynamic Air Removal (pre-vacuum) – Exposure at 132°C for 4 minutes, 20-minute dry time. Gravity displacement – Exposure at 132°C for 15 minute dry time. Ziroonia Guided Surgery Kit Cases are intended for sterilization of non-porous loads. The Zirconia Guided Surgery Kit Case and the associated instruments is 342.2 g. The weight of the empty Kit is 236.8 grams. Zirconia Guided Surgery Kit Case are recommended not to be stacked during sterilization. Nuvo CF Guided Surgery Kit Cases are intended to be used to enclose other medical devices that are to be sterilized by a health care provider. Nuvo CF Guided Surgery Kit Cases are intended to allow sterilization of the enclosed medical devices. Nuvo CF Guided Surgery Kit Cases require the use of FDA-cleared wrap to maintain the sterility of the enclosed devices. The kits are to be enclosed in a sterilization wrap that is FDA-cleared for the indicated cycles, and moist heat (steam) sterilized using one of the following cycles: Fractionated vacuum): exposure at 132°C for 4 minutes, drying for 20 minutes. Gravity displacement: exposure at 132°C for 15 minutes, drying for 30 minutes. Nuvo CF Guided Surgery Kit Cases are intended for sterilization of non-porous loads. The combined weight of the Nuvo CF Guided Surgery Kit Cases and the associated instruments is approximately 693 g. The weight of the empty case is 487 grams. Nuvo CF Guided Surgery Kit Cases are not recommended to be stacked during sterilization.
Device Story
Reusable rigid containers (case bottom, inner tray, lid) for dental instrument organization, sterilization, and protection; composed of autoclavable PPSU polymer with silicone retention grommets. Used by healthcare providers in clinical settings. Instruments are placed in the tray, enclosed in FDA-cleared sterilization wrap, and processed via steam autoclave. Output is a sterilized, organized kit for surgical procedures. Benefits include secure storage and maintenance of instrument sterility during transport and surgical preparation.
Clinical Evidence
No clinical data included. Bench testing only: steam sterilization validation (ISO 17665-1), manual cleaning validation (AAMI TIR30:2011), life cycle testing (100 cycles), and biocompatibility (ISO 10993-5).
Technological Characteristics
Materials: Polyphenylsulfone (PPSU) base/lid, medical-grade silicone grommets. Design: Rigid, perforated, reusable (up to 100 cycles). Sterilization: Moist heat (steam) to SAL 10^-6. Dimensions: Zirconia (195x90x67mm), Nuvo CF (264x163x58mm).
Indications for Use
Indicated for healthcare providers to organize, store, and sterilize dental instruments (specifically for Bone Level ceramic implants or Nuvo CF implants) via moist heat (steam) sterilization within an FDA-cleared sterilization wrap.
Regulatory Classification
Identification
A sterilization wrap (pack, sterilization wrapper, bag, or accessories, is a device intended to be used to enclose another medical device that is to be sterilized by a health care provider. It is intended to allow sterilization of the enclosed medical device and also to maintain sterility of the enclosed device until used.
Neodent Implant System – Helix Short Surgical Kit Cases (K223662)
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logos of the Department of Health & Human Services and the Food and Drug Administration (FDA). The Department of Health & Human Services logo is on the left, and the FDA logo is on the right. The FDA logo is a blue square with the letters "FDA" in white, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue.
August 23, 2024
JJGC Indústria e Comércio de Materiais Dentários S.A. % Jennifer Jackson Sr. Director, Regulatory Affairs and Quality Straumann USA, LLC 60 Minuteman Road Andover, Massachusetts 01810
Re: K241492
Trade/Device Name: Zirconia and Nuvo CF Guided Surgery Kit Cases Regulation Number: 21 CFR 880.6850 Regulation Name: Sterilization Wrap Regulatory Class: Class II Product Code: KCT Dated: Mav 28, 2024 Received: May 28, 2024
Dear Jennifer Jackson:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device"
{1}------------------------------------------------
(https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30. Design controls; 21 CFR 820.90. Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review. the OS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Image /page/1/Picture/6 description: This image shows a digital signature and contact information. The signature is from Stephen A. Anisko -S, and it was signed on August 23, 2024, at 12:50:29 -04'00'. The contact information is for Christopher Dugard, Assistant Director at the Division of Infection Control Devices, Office of Surgical and Infection Control Devices, Office of Product Evaluation and Quality, and Center for Devices and Radiological Health.
{2}------------------------------------------------
### K241492 - Jennifer Jackson
Enclosure
{3}------------------------------------------------
### Indications for Use
510(k) Number (if known) K241492
#### Device Name
Zirconia and Nuvo CF Guided Surgery Kit Cases
#### Indications for Use (Describe)
Zirconia Guided Surgery Kit Cases:
Zirconia Guided Surgery Kit Cases are intended to be used to enclose other medical devices that are to be sterilized by a health care provider. Zirconia Guided Surgery Kit Cases are intended to allow sterilization of the enclosed medical devices. Zirconia Guided Surgery Kit Cases requre the use of FDA-cleared wrap to maintain the sterility of the enclosed devices. The kits are to be enclosed in a sterlization wrap that is FDA-cleared for the indicated cycles, and moist heat (steam) sterilized using one of the following cycles: Dynamic Air Removal (pre-vacuum) – Exposure at 132°C for 4 minutes, 20-minute dry time. Gravity displacement – Exposure at 132°C for 15 minute dry time. Ziroonia Guided Surgery Kit Cases are intended for sterilization of non-porous loads. The Zirconia Guided Surgery Kit Case and the associated instruments is 342.2 g. The weight of the empty Kit is 236.8 grams. Zirconia Guided Surgery Kit Case are recommended not to be stacked during sterilization.
#### Nuvo CF Guided Surgery Kit Cases:
Nuvo CF Guided Surgery Kit Cases are intended to be used to enclose other medical devices that are to be sterilized by a health care provider. Nuvo CF Guided Surgery Kit Cases are intended to allow sterilization of the enclosed medical devices. Nuvo CF Guided Surgery Kit Cases require the use of FDA-cleared wrap to maintain the sterility of the enclosed devices. The kits are to be enclosed in a sterilization wrap that is FDA-cleared for the indicated cycles, and moist heat (steam) sterilized using one of the following cycles: Fractionated vacuum): exposure at 132°C for 4 minutes, drying for 20 minutes. Gravity displacement: exposure at 132°C for 15 minutes, drying for 30 minutes. Nuvo CF Guided Surgery Kit Cases are intended for sterilization of non-porous loads. The combined weight of the Nuvo CF Guided Surgery Kit Cases and the associated instruments is approximately 693 g. The weight of the empty case is 487 grams. Nuvo CF Guided Surgery Kit Cases are not recommended to be stacked during sterilization.
| Type of Use (Select one or both, as applicable) |
|-----------------------------------------------------------------------------------------------------------------------------------------------|
| <span style="unicode-bidi:isolate; direction:ltr;"> <input checked="" type="checkbox"/> Prescription Use (Part 21 CFR 801 Subpart D) </span> |
| <span style="unicode-bidi:isolate; direction:ltr;"> <input type="checkbox"/> Over-The-Counter Use (21 CFR 801 Subpart C) </span> |
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{4}------------------------------------------------
## Zirconia and Nuvo CF Guided Surgery Kit Cases
510(k) Summary - K241492
#### 510(k) Summary 1
### Submitter's Contact Information
| Submitter: | Straumann USA, LLC<br>60 Minuteman Road<br>Andover, MA 01810, USA<br>Registration No.: 1222315 Owner/Operator No.: 9005052 |
|-------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| | On the behalf of: |
| | JJGC Indústria e Comércio de Materiais Dentários S.A. (dba Neodent<br>Av. Juscelino Kubitschek de Oliveira, 3291 - CIC<br>Curitiba, Paraná, Brazil 81270-200<br>Registration No.: 3008261720<br>Owner/Operator No.: 10031702 |
| Contact Person: | Jennifer M. Jackson, MS, RAC<br>Sr. Director, Regulatory Affairs and Quality<br>Email: jennifer.jackson@straumann.com<br>Phone Number: +1 978 747-2509<br>Fax Number: +1 978 747-0023 |
| Prepared By &<br>Alternate Contact: | Leticia Milani<br>Regulatory Affairs Analyst<br>JJGC Indústria e Comércio de Materiais Dentários SA<br>E-mail: leticia.milani@neodent.com<br>Phone Number: +55 41 216-9485 |
| Date Prepared: | August 22, 2024 |
### Name of the Device
| Trade Names: | Zirconia and Nuvo CF Guided Surgery Kit Cases |
|------------------------|---------------------------------------------------------|
| Common Name: | Instrument Sterilization Trays |
| Classification Name: | Sterilization Wrap Containers, Trays, Cassettes & Other |
| Regulation Number: | 21 CFR 880.6850, Class II, Sterilization wrap |
| Device Classification: | II |
| Product Code(s): | KCT |
| Classification Panel: | General Hospital |
{5}------------------------------------------------
### Zirconia and Nuvo CF Guided Surgery Kit Cases
510(k) Summary - K241492
### Predicate Device(s)
#### Primary Predicate:
- K203618 Neodent EasyGuide Kit Cases (JJGC Indústria e Comércio de Materiais . Dentários S.A)
#### Reference Devices:
- K182865 Neodent Instrument Kits (JJGC Indústria e Comércio de Materiais Dentários S.A) .
- . K223662 – Neodent Implant System – Helix Short Surgical Kit Cases (JJGC Indústria e Comércio de Materiais Dentários S.A)
### Device Description
The Zirconia and Nuvo CF Guided Surgery Kit Cases subject to this submission are reusable rigid containers, composed by a case bottom, a removable inner tray and a lid, which are made of autoclavable polymer. They are indicated to hold various dental instruments (not subject of this submission) in order to organize, sterilize and protect them. Sterilization via moist heat (steam) is performed by the healthcare provider.
#### Intended Use
The Zirconia and Nuvo CF Guided Surgery Kit Cases are used to safe storage and securing of instruments during sanitization and surgical procedures. The Zirconia Guided Surgery Kit Cases are designed for storing the instruments used to install Bone Level (BL) ceramic implants. The Nuvo CF Guided Surgery Kit Cases are intended for the installation of Nuvo CF implants.
### Indications for Use
Zirconia Guided Surgery Kit Cases are intended to be used to enclose other medical devices that are to be sterilized by a health care provider. Zirconia Guided Surgery Kit Cases are intended to allow sterilization of the enclosed medical devices. Zirconia Guided Surgery Kit Cases require the use of FDA-cleared wrap to maintain the sterility of the enclosed devices. The kits are to be enclosed in a sterilization wrap that is FDA-cleared for the indicated cycles, and moist heat (steam) sterilized using one of the following cycles: Dynamic Air Removal (pre-vacuum) – Exposure at 132°C for 4 minutes, 20-minute dry time. Gravity displacement – Exposure at 132°C for 15 minutes, 20-minute dry time. Zirconia Guided Surgery Kit Cases are intended for sterilization of non-porous loads. The combined weight of the Zirconia Guided Surgery Kit Case and the associated instruments is 342.2
{6}------------------------------------------------
## Zirconia and Nuvo CF Guided Surgery Kit Cases
510(k) Summary - K241492
g. The weight of the empty Kit is 236.8 grams. Zirconia Guided Surgery Kit Case are recommended not to be stacked during sterilization.
Nuvo CF Guided Surgery Kit Cases are intended to be used to enclose other medical devices that are to be sterilized by a health care provider. Nuvo CF Guided Surgery Kit Cases are intended to allow sterilization of the enclosed medical devices. Nuvo CF Guided Surgery Kit Cases require the use of FDA-cleared wrap to maintain the sterility of the enclosed devices. The kits are to be enclosed in a sterilization wrap that is FDA-cleared for the indicated cycles, and moist heat (steam) sterilized using one of the following cycles: Fractionated vacuum (pre-vacuum): exposure at 132°C for 4 minutes, drying for 20 minutes. Gravity displacement: exposure at 132°C for 15 minutes, drying for 30 minutes. Nuvo CF Guided Surgery Kit Cases are intended for sterilization of non-porous loads. The combined weight of the Nuvo CF Guided Surgery Kit Cases and the associated instruments is approximately 693 g. The weight of the empty case is 487 grams. Nuvo CF Guided Surgery Kit Cases are not recommended to be stacked during sterilization.
{7}------------------------------------------------
## Zirconia and Nuvo CF Guided Surgery Kit Cases
510(k) Summary - K241492
## Technological Characteristics
The technological characteristics of the subject devices are compared to the primary predicate and reference devices in the following tables:
| FEATURE | SUBJECT DEVICE | PRIMARY PREDICATE DEVICE | REFERENCE DEVICE | EQUIVALENCE DISCUSSION |
|----------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| K Number | K241492<br>Zirconia Guided Surgery Kit Cases | K203618<br>Neodent EasyGuide Kit Cases | K223662<br>Helix Short Surgical Kit Cases | |
| Indications for<br>Use Statement | Zirconia Guided Surgery Kit Cases are<br>intended to be used to enclose other<br>medical devices that are to be sterilized<br>by a health care provider. Zirconia<br>Guided Surgery Kit Cases are intended<br>to allow sterilization of the enclosed<br>medical devices. Zirconia Guided<br>Surgery Kit Cases require the use of<br>FDA-cleared wrap to maintain the<br>sterility of the enclosed devices. The<br>kits are to be enclosed in a sterilization<br>wrap that is FDA-cleared for the<br>indicated cycles, and moist heat<br>(steam) sterilized using one of the<br>following cycles: Dynamic Air Removal<br>(pre-vacuum) - Exposure at 132°C for<br>4 minutes, 20-minute dry time. Gravity<br>displacement - Exposure at 132°C for<br>15 minutes, 20-minute dry time.<br>Zirconia Guided Surgery Kit Cases are<br>intended for sterilization of non-porous<br>loads. The combined weight of the<br>Zirconia Guided Surgery Kit Case and<br>the associated instruments is 342.2 g.<br>The weight of the empty Kit is 236.8<br>grams. Zirconia Guided Surgery Kit<br>Case are recommended not to be<br>stacked during sterilization. | Neodent Instrument Kits are intended to<br>be used to enclose other medical<br>devices that are to be sterilized by a<br>health care provider. Neodent<br>Instrument Kits are intended to allow<br>sterilization of the enclosed medical<br>devices. Neodent Instrument Kits<br>require the use of FDA-cleared wrap to<br>maintain the sterility of the enclosed<br>devices. The Kits are to be enclosed in<br>a sterilization wrap that is FDA-cleared<br>for the indicated cycles, and moist heat<br>(steam) sterilized using one of the<br>following cycles: Fractionated vacuum<br>(pre-vacuum) - Exposure at 132 °C for<br>4 minutes, 20-minute dry time. Gravity<br>displacement - Exposure at 132 °C for<br>15 minutes, 20-minute dry time.<br>Neodent Instrument Kits are intended<br>for sterilization of non-porous loads. The<br>combined weight of the GM EasyGuide<br>Surgical Kit Case Narrow/Regular and<br>the associated instruments is 310,18 g.<br>The weight of the empty Kit Case is<br>263,63 g. Neodent GM EasyGuide Kit<br>Cases should not be stacked during<br>sterilization. | HS Surgical Kit Cases are indicated for the<br>organization of surgical and/or prosthetic<br>instruments during sterilization, storage,<br>and transport. Use of this product facilitates<br>the storage and organization of instruments<br>during and after surgical procedures.<br>Neodent instrument kit cases are intended<br>to allow sterilization of the medical devices<br>included. Neodent instrument kit cases<br>must be wrapped in FDA-approved<br>materials to maintain the sterility of the<br>devices included. The kits should be placed<br>in an FDA-approved sterilization wrap for<br>the indicated cycles and sterilized by moist<br>heat (steam) using one of the following<br>cycles: Dynamic Air Removal (pre-vacuum):<br>exposure at 132°C for 4 minutes, drying for<br>30 minutes. Gravity displacement: exposure<br>at 132°C for 15 minutes, drying for 50<br>minutes. Neodent instrument kit cases are<br>intended for sterilization of non-porous<br>fillers. The combined weight of the HS<br>Surgical Kit case and associated<br>instruments is 302.88 grams. The weight of<br>the empty case is approximately 214.85<br>grams. Neodent instrument kit cases should<br>not be stacked during sterilization. | Similar<br>The subject and primary predicate<br>devices present the same indications<br>for use, being both intended to allow<br>sterilization of the enclosed medical<br>devices. The reference predicate<br>device has a slightly different language,<br>but the indications for use are similar.<br>All devices are intended to enclose<br>other medical devices that are to be<br>sterilized by a health care provider. |
| FEATURE | SUBJECT DEVICE | PRIMARY PREDICATE DEVICE | REFERENCE DEVICE | EQUIVALENCE DISCUSSION |
| K Number | K241492<br>Zirconia Guided Surgery Kit Cases | K203618<br>Neodent EasyGuide Kit Cases | K223662<br>Helix Short Surgical Kit Cases | EQUIVALENCE DISCUSSION |
| Material and<br>design | Rigid polyphenylsulfone (PPSU) polymer base and removable inner tray with a polyphenylsulfone (PPSU) lid.<br>Retention grommets of medical grade silicone. | Rigid polysulfone (PSU) polymer base and removable inner tray with a polyphenylsulfone (PPSU) lid. Retention grommets of medical grade silicone. | Rigid polysulfone (PSU) polymer base and removable inner tray with a polyphenylsulfone (PPSU) lid. Retention grommets of medical grade silicone. | Similar<br>The subject and predicate devices are made of the same material. Despite the subject devices presenting a difference in base and tray composition, the PPSU material has already been used before in devices cleared by FDA. Furthermore, this material change does not interfere with the biocompatibility and performance of the new products. |
| Perforated | Yes | Yes | Yes | Same |
| Overall<br>dimensions | 195 L x 90 W x 67 H, mm | 195 L x 90 W x 64 H, mm | 195 L x 90 W x 67 H, mm | Similar<br>The subject, primary predicate and reference devices have the same dimensions. The small difference between the height of subject and primary predicate device does not compromise safety and efficacy as is better discussed along this submission. |
| Reusable | Yes, reusable up to 100 cycles | Yes, reusable up to 100 cycles | Yes, reusable up to 100 cycles | Same |
| Volume to Vento<br>Ratio | 64.8 cm³/ cm² (25.5 in³/ in²) | 63.5 cm³/ cm² (25.0 in³/ in²) | 64.8 cm³/ cm² (25.5 in³/ in²) | Similar<br>The subject and primary predicate device have a slight difference in volume to vent rate, which does not compromise safety and efficacy, as is proved by the presented sterilization validation. |
| Sterilization<br>Method | Moist heat (steam) to a SAL of 10-6 | Moist heat (steam) to a SAL of 10-6 | Moist heat (steam) to a SAL of 10-6 | Same |
| Cycles and<br>parameters | Fractionated vacuum (pre-vacuum):<br>Sterilization temperature: 132 °C;<br>Sterilization time: 4 minutes;<br>Drying time: 20 minutes.<br>Gravity displacement:<br>Sterilization temperature: 132 °C;<br>Sterilization time: 15 minutes;<br>Drying time: 20 minutes. | Fractionated vacuum (pre-vacuum):<br>Sterilization temperature: 132 °C;<br>Sterilization time: 4 minutes;<br>Drying time: 20 minutes.<br>Gravity displacement:<br>Sterilization temperature: 132 °C;<br>Sterilization time: 15 minutes;<br>Drying time: 20 minutes. | Fractionated vacuum (pre-vacuum):<br>Sterilization temperature: 132 °C;<br>Sterilization time: 4 minutes;<br>Drying time: 30 minutes.<br>Gravity displacement:<br>Sterilization temperature: 132 °C;<br>Sterilization time: 15 minutes;<br>Drying time: 50 minutes. | Similar<br>The subject devices have the same cycle parameters already cleared for the predicate devices, however the reference predicate device requires a longer drying cycle. |
| FEATURE | SUBJECT DEVICE | PRIMARY PREDICATE DEVICE | REFERENCE DEVICE | EQUIVALENCE DISCUSSION |
| K Number | K241492<br>Zirconia Guided Surgery Kit Cases | K203618<br>Neodent EasyGuide Kit Cases | K223662<br>Helix Short Surgical Kit Cases | EQUIVALENCE DISCUSSION |
| Sterile Barrier | Sterilization wrap, FDA-cleared<br>for<br>indicated method and cycles | Sterilization wrap, FDA-cleared for<br>indicated method and cycles | Sterilization wrap,<br>FDA-cleared<br>for<br>indicated method and cycles | Same |
| Biocompatibility | The assessment to Biocompatibility<br>was performed per ISO 10993-1 and<br>testing was performed using methods<br>described in AAMI/ANSI/ISO 10993-5.<br>Results indicate that the subject<br>devices are biocompatible. | The assessment to Biocompatibility was<br>performed per ISO 10993-1 and testing<br>performed using methods<br>described in AAMI/ANSI/ISO 10993-5.<br>Results indicate that the subject devices<br>are biocompatible.…
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.