eZSURE™ Empty Fluid Container with ProSeal™ Injection Site
K241442 · Epic Medical Pte. , Ltd. · KPE · Jun 21, 2024 · General Hospital
Device Facts
Record ID
K241442
Device Name
eZSURE™ Empty Fluid Container with ProSeal™ Injection Site
Applicant
Epic Medical Pte. , Ltd.
Product Code
KPE · General Hospital
Decision Date
Jun 21, 2024
Decision
SESE
Submission Type
Special
Regulation
21 CFR 880.5025
Device Class
Class 2
Indications for Use
The Empty Fluid Container is used to hold an admixture of compatible fluids for intravenous administration to a patient. Medication transfer in and out of the container is done using aseptic technique.
Device Story
Empty, sterile, non-pyrogenic flexible IV container (100-500 mL) for holding/administering compatible fluids. Features two ports: a spiking/administration port for standard IV sets and an additive/filling port integrated with a ProSeal™ Injection Site. Filling occurs via transfer set/syringe with ProSeal™ Injector/Injector Plus. ProSeal™ system uses double-membrane septum design; cannula perforates membranes for liquid transfer; retraction seals membranes to prevent drug/vapor escape and microbial ingress. Used in clinical settings by healthcare professionals/pharmacists. Enhances safety by minimizing exposure to drug vapors, aerosols, and spills during medication transfer.
Clinical Evidence
Bench testing only. No clinical data. Performance verified via ISO 15747 (leakage, pressure, temperature, drop, microbial impermeability), ISO 8536-4, ISO 80369-7, and NIOSH 2016 vapor containment protocols. Biocompatibility leveraged from K223674 and K240433 (ISO 10993-1:2018).
Technological Characteristics
Flexible IV bag (Nexcel® Polyolefin film). Two-port configuration: spiking port (thermoplastic elastomer) and additive port (ProSeal™ Injection Site with polyisoprene rubber membrane). Sterilization: Ethylene Oxide (SAL 10^-6). Shelf life: 3 years (ASTM F1980-16). Connectivity: Manual/automated filling via ProSeal™ Injector/Injector Plus (male Luer lock).
Indications for Use
Indicated for patients requiring intravenous administration of compatible fluids. Used by healthcare professionals or pharmacists in clinical settings.
Regulatory Classification
Identification
An I.V. container is a container made of plastic or glass used to hold a fluid mixture to be administered to a patient through an intravascular administration set.
ToxiSeal™ Vial Adaptor with External Balloon (K222929)
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logos of the U.S. Food and Drug Administration (FDA) and the Department of Health & Human Services. The FDA logo features the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue text. The Department of Health & Human Services logo is a symbol with the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" surrounding it.
June 21, 2024
Epic Medical Pte. Ltd. Freddie Lee CEO/md 105 Cecil Street #20-04, The Octagon Singapore, SG 069534 Singapore
Re: K241442
Trade/Device Name: eZSURE™ Empty Fluid Container with ProSeal™ Injection Site Regulation Number: 21 CFR 880.5025 Regulation Name: I.V. Container Regulatory Class: Class II Product Code: KPE, ONB Dated: May 22, 2024 Received: May 22, 2024
Dear Freddie Lee:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
{1}------------------------------------------------
Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely, Porsche Bennett
Porsche Bennett For David Wolloscheck, Ph.D. Assistant Director DHT3C: Division of Drug Delivery and General Hospital Devices, and Human Factors OHT3: Office of Gastrorenal, ObGyn, General Hospital, and Urology Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
{2}------------------------------------------------
## Indications for Use
510(k) Number (if known) K241442
Device Name
eZSURE™ Empty Fluid Container with ProSeal™ Injection Site
Indications for Use (Describe)
The Empty Fluid Container is used to hold an admixture of compatible fluids for intravenous administration to a patient. Medication transfer in and out of the container is done using aseptic technique.
Type of Use (Select one or both, as applicable)
X Prescription Use (Part 21 CFR 801 Subpart D)
| Over-The-Counter Use (21 CFR 801 Subpart C)
## CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
## *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
## K241442 – 510(k) Summary
#### I. Submitter
Epic Medical Pte. Ltd. 105 Cecil Street #20-04, The Octagon, Singapore 069534. Phone: +65 9635 2618 / +66 81 761 5292
Contact Person: Mr. Freddie LEE, Chief Executive Officer/ Managing Director Date Prepared: June 20, 2024 Content and Format: Prepared in accordance with 21 CFR 807.92 Type of Submission: Special
#### II. Subject Device
| 510(k) Number: | K241442 |
|---------------------------------------------|------------------------------------------------------------|
| Proprietary/<br>Trade/ Device Name: | eZSURE™ Empty Fluid Container with ProSeal™ Injection Site |
| Common/Usual/Generic<br>Name: | Empty I.V. bag |
| Classification Name/<br>Regulation Name: | I.V. container |
| Device Classification<br>Regulation Number: | 21 CFR 880.5025 |
| Regulatory/ Device Class: | Class: II |
| Classification Product<br>Code: | KPE (primary) and ONB (secondary) |
#### III. Predicate
| 510(k) Number: | K223674 |
|----------------------------------------------------------|------------------------------------------------------------------|
| Proprietary/<br>Trade/ Device Name: | eZSURE™ Empty Fluid Container<br>(models 426030, 426040, 426110) |
| Common/Usual/Generic<br>Name: | Empty I.V. bag |
| Classification Name/<br>Regulation Name/<br>Description: | I.V. container |
| Device Classification<br>Regulation Number: | 21 CFR 880.5025 |
| Classification Product<br>Code: | KPE |
| Regulatory/ Device Class: | Class: II |
{4}------------------------------------------------
#### IV. Device Description
The purpose of this Special 510(k) Submission is to seek FDA clearance prior to the introduction in the US, of a modified design of the existing eZSURE™ Empty Fluid Container - the Subject eZSURE™ Empty Fluid Container with ProSeal™ Injection Site. This device is an addition to the ProSeal™ CSTD system of devices (cleared K240433). The existing device, and also the nominated Predicate device, is the eZSURE™ Empty Fluid Container (cleared K223674).
The eZSURE™ Empty Fluid Container (EFC) devices are empty single-use, sterile, nonpyrogenic flexible IV container devices/ bags. These are discarded after use. The Subject EFC is made of non-PVC materials. The Nexcel Film for IV bag of the Subject EFC device is composed of a flexible plastic film bag and the device is provided in a two-port configuration with closures.
A closed system inlet-/ entry-/ additive- port is used for filling one or more compatible fluid(s) into the bag by a transfer set/ syringe without needle with the ProSeal™ Injector or ProSeal™ Injector Plus (cleared K240171) attached, and another port, the spiking/ administration port, is used for accessing the infusate in the bag with a standard bag spike in an IV therapy administration from the EFC. The transfer device with a male Luer lock attached with the ProSeal™ Injector (or ProSeal™ Injector Plus) is used to connect to the filling-/ additive- port for filling. The additive port incorporates a ProSeal™ Injection Site (cleared K240433) as its integrated subcomponent; hence no other injection needle/ cannula is needed. The transfer device is removed at the end of the preparation step, and the self-sealing additive-/ injection-/ filling- port secures the admixture contents until their administration.
The closed transfer of liquid that takes place with the use of the ProSeal™ CSTD system (K240433) as follows:
- · A double membrane septum design utilizing self-sealing elastomeric membranes tightly fit together when the system components engage. A cannula within the ProSeal™ Injector (or ProSeal™ Injector Plus) housing perforates the double membrane for the transfer of liquid. When the cannula is retracted, the membranes seal off the transfer of environments into the system and/or escape of drug or vapor concentrations outside the system, thereby minimizing the individual and environmental exposure to drug vapor, aerosols, and spills, and also minimizing the risk of microbial contamination.
For administration to a patient, the device is then connected to an external IV set /IV line, via a bag spike. The IV bag is piped by inserting the spike point of a bag spike into the spiking-/administration- port of the IV bag, doing this by performing a twisting motion. When the bag is already filled, other medications can be added using the additive/ injection/ filling port, even during administration. Medication transfer in and out of the container is done using aseptic technique. The bags range in volume capacities of 100 mL, 250 mL, and 500 mL.
The EFC sub-components are externally communicating devices with no contact to the blood path. Contact duration is categorized as B-prolonged, (>24h to 30d), per ISO 10993-1: 2018 biocompatibility guidelines.
{5}------------------------------------------------
# 집 미 I
The only modification proposed in this Submission is the change of the fill port design from the needle free valve to the ProSeal™ Injection Site (cleared K240433).
The modified and existing device have the same intended use, and the ProSeal™ Injection Site is a cleared CSTD component device (K220433). The principles of operation, indications for use statement, and the user-applications remain the same.
#### V. Indications for Use Statement
The Empty Fluid Container is used to hold an admixture of compatible fluids for intravenous administration to a patient. Medication transfer in and out of the container is done using aseptic technique.
#### VI. Comparison of Technological Characteristics
The Subject device and the Predicate device share the following key characteristics:
- Same 2-port configuration an additive/ filling port and a spiking/ administration port ●
- Same basic design features clear, flexible, and empty IV fluid bag that allows for ● preparation and administration of IV therapies
- Same means of attachment provided for an IV administration set/ line, for the delivery of IV therapy
An overview table summarizing the comparison between the key characteristics between the Subject and the Predicate device is provided from the next page.
{6}------------------------------------------------
| Characteristic<br>compared | Predicate Device<br>eZSURE™ Empty Fluid Container (K223674) | Subject Device<br>eZSURE™ Empty Fluid Container<br>with ProSeal™ Injection Site (K241442) | Comment/<br>Discussion |
|---------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Indications for Use<br>statement | "The Empty Fluid Container is used to hold an<br>admixture of compatible fluids for intravenous<br>administration to a patient. Medication transfer<br>in and out of the container is done using aseptic<br>technique." | "The Empty Fluid Container is used to hold an<br>admixture of compatible fluids for intravenous<br>administration to a patient. Medication transfer<br>in and out of the container is done using aseptic<br>technique." | Same |
| Primary product code<br>and regulation number | KPE<br>21 CFR 880.5025 | KPE<br>21 CFR 880.5025 | Same |
| Intended user<br>population/ intended<br>use environment | Adequately trained<br>health care professionals<br>or pharmacists/ clinical setting | Adequately trained<br>health care professionals<br>or pharmacists/ clinical setting | Same |
| IV bag<br>components/<br>access<br>components | Device has 2 access points:<br>1. Additive port/ filling port<br>2. Spiking/administration port, IV spike bag<br>access for IV set/ IV line | Device has 2 access points:<br>1. Additive port/ filling port<br>2. Spiking/ administration port, IV spike bag<br>access for IV set/ IV line | Same |
| Characteristic<br>compared | Predicate Device<br>eZSURE™ Empty Fluid Container (K223674) | Subject Device<br>eZSURE™ Empty Fluid Container<br>with ProSeal™ Injection Site (K241442) | Comment/<br>Discussion |
| System<br>component<br>devices | 1. ProSeal™ Injector (model 421010) (K240433)<br>2. ProSeal™ Injector Plus (model 421050) (K240433)<br>3. ProSeal™ Connector (model 422010) (K240433)<br>4. ProSeal™ Injection Site Extended Male Luer<br>Lock (K240433)<br>5. ProSeal™ Vial Adaptor (3 models 4200X0,<br>for options of vial neck sizes and pressure<br>equalizations) (K240433)<br>6. ToxiSeal™ Vial Adaptor with External<br>Balloon (K222929)<br>7. ProSeal™ Assembly Fixture (Fixture for<br>ProSeal™ Vial Adaptor assembly, model<br>424010) (K240433) | 1. ProSeal™ Injector (model 421010) (K240433)<br>2. ProSeal™ Injector Plus (model 421050) (K240433)<br>3. ProSeal™ Connector (422010) (K240433)<br>4. ProSeal™ Injection Site Extended Male Luer Lock (K240433)<br>5. ProSeal™ Vial Adaptor (3 models 4200X0, for options of vial neck sizes and pressure equalizations) K240433)<br>6. ToxiSeal™ Vial Adaptor with External Balloon (K222929)<br>7. ProSeal™ Assembly Fixture (Fixture for ProSeal™ Vial Adaptor assembly, model 424010) (K240433)<br>8. eZSURE™ Empty Fluid Container with ProSeal™ Injection Site ((K241442) the Subject device) | Different<br>The Subject system<br>component device<br>(#8 in the column<br>Subject Device), is<br>an addition to the<br>existing seven (7)<br>component devices<br>of ProSeal™ CSTD<br>(K240433) and the<br>ToxiSeal™ Vial<br>Adaptor with<br>External Balloon<br>(K222929), indicate<br>in <b>bold face texts</b> .<br>Differences are<br>addressed in<br>Discussion of<br>differences in<br>technological<br>characteristics:<br>Comments #1 and #2<br>under this table, on<br>page 8. |
| Intended<br>drug type | Parenteral drugs | Parenteral drugs | Same |
| R/<br>Prescription use | R only | R only | Same |
| Characteristic<br>compared | Predicate Device<br>eZSURE™ Empty Fluid Container (K223674) | Subject Device<br>eZSURE™ Empty Fluid Container<br>with ProSeal™ Injection Site (K241442) | Comment/<br>Discussion |
| Principles<br>of operation | The empty bags are filled by connecting<br>transfer devices, e.g., through standard<br>male Luer lock syringes (without needle)<br>or automated filling devices<br>The additive port is a self-sealing needle-<br>free female luer lock connector, no<br>different from standard injection site<br>devices of market-cleared IV sets.<br>Silicone valve of the additive port self-<br>seals/ re-seals when mating component<br>devices are disconnected from one another<br><br>Image: Filling Port and Spiking port with protective cap | The empty bags are filled by connecting<br>transfer devices, e.g., through standard<br>male Luer lock syringes (without needle)<br>or automated filling devices, attached to<br>the ProSeal™ Injector (or the ProSeal™<br>Injector Plus)<br>The additive port is a self-sealing closed<br>system ProSeal™ Injection Site.<br>Membrane of the additive port self-seals/<br>re-seals when mating component devices<br>are disconnected from one another<br><br>Image: Filling Port and Spiking port with protective cap | Different<br>See Comment #1 |
| | IV set/ IV line is attached through the<br>spiking/ administration port to dispense IV<br>therapy using a standard spike and tubing | IV set/ IV line is attached through the<br>spiking/ administration port to dispense IV<br>therapy using a standard spike and tubing | Same |
| Technology<br>and design | The bag is manufactured by welding a<br>plastic attachment to the bag body and<br>assembling the bottom with entry and exit<br>connectors/ ports | The bag is manufactured by welding a<br>plastic attachment to the bag body and<br>assembling the bottom with entry and exit<br>connectors/ ports | Same |
| Material<br>of empty fluid<br>container body | Sealed Air Corp's Medical Packaging Film,<br>Nexcel® Film, Grade M315, Polyolefin | Sealed Air Corp's Medical Packaging Film,<br>Nexcel® Film, Grade M315, Polyolefin | Same |
| Characteristic<br>compared | Predicate Device<br>eZSURE™ Empty Fluid Container (K223674) | Subject Device<br>eZSURE™ Empty Fluid Container<br>with ProSeal™ Injection Site (K241442) | Comment/<br>Discussion |
| Material of<br>additive filling port | Polypropylene, Polycarbonate,<br>Liquid Silicone Rubber | Polypropylene, Polyisoprene Rubber (IR) | Different<br>See Comment #1 |
| Material of spiking/<br>administration port | Thermoplastic elastomer | Thermoplastic elastomer | Same |
| Intended Injector for<br>filling the empty fluid<br>container | ProSeal™ Injector (model 421010)<br>(cleared K192075)<br>Male Luer Lock tip | ProSeal™ Injector (model 421010) or<br>ProSeal™ Injector Plus (model 421050) (cleared<br>K241071)<br>both Male Luer Lock tip | Different<br>See Comment #2 |
| Primary<br>package top web | Medical grade paper and<br>medical plastic film, heat sealed | Medical grade paper and<br>medical plastic film, heat sealed | Same |
| Sterilization method | Ethylene Oxide, EO, SAL 10-6 | Ethylene Oxide, EO, SAL 10-6 | Same |
| Validated<br>Shelf life | 3 years (36 months) | 3 years (36 months) | Same |
| Reuse<br>or single-use | Single-use only | Single-use only | Same |
| Labeling<br>specifications | Meets the requirements<br>specified in 21 CFR 801 | Meets the requirements<br>specified in 21 CFR 801 | Same |
{7}------------------------------------------------
{8}------------------------------------------------
# 집 1 C
{9}------------------------------------------------
{10}------------------------------------------------
## Discussion of differences in technological characteristics
## Comment #1
The membrane material of Subject device's closed system ProSeal™ Injection Site Extended Male Luer Lock (model 422140) (ProSeal™ Injection Site) is made of polyisoprene rubber (IR), the same as the IR present in Epic Medical Pte Ltd's cleared K240433, ProSeal™ Injection Site, in the same design configuration and it had been biologically evaluated to have conformed with ISO 10993-1: 2018. Therefore, biocompatibility testing was leveraged from K240433. Moreover, the functional performance of the assembled ProSeal™ Injection Site in the Subject device, had been adequately evaluated and as such, performance testing was leveraged form K240443. In additional bench top performance verification testing of the subject device over the shelf life was conducted to the standards listed in section VII.A second paragraph, hereunder, to demonstrate that the difference in technological characteristics do not raise different questions of safety and effectiveness.
## Comment #2
The 2024-03-28 cleared ProSeal™ Injector Plus (cleared K240171), a modified version of ProSeal™ Injector, is another option of an Injector available for users to use with the Subject Emptv Fluid Container devices. This difference does not raise new questions of safety and effectiveness of the Subject device.
#### VII. Performance Data Supporting Substantial Equivalence
## A. Functional Performance
The sterile, single-use, non-pyrogenic flexible IV container devices/ bags described in this Summary were evaluated to be in conformance with the following ISO and FDA recognized standards:
- ISO 8536-4: 2019, Infusion equipment for medical use - Part 4: Infusion sets for single use, gravity feed
- ISO 15747: 2018, Plastic containers for intravenous injections ●
- . ISO 22413: 2021, Transfer sets for pharmaceutical preparations – Requirements and test methods
Bench performance verifications and validations performed on the Subject device:
- Resistance to Temperature, Pressure and Leakage test to ISO 15747 on the Subject device ●
- Resistance to Dropping test to ISO 15747 on the Subject device ●
- . Additive Port Liquid Tightness (Air and Liquid Leakages) test to ISO 15747 on the Subject device
- Impermeability to Microorganism test to ISO 15747 on the Subject device ●
{11}------------------------------------------------
The ProSeal™ Injection Site (cleared K240433) incorporated as the Subject device's integrated additive port, described in section IV in this Summary, and its IR membrane subcomponent part, were tested and demonstrated to be in conformance with the following ISO and FDA recognized standards and FDA guidance document:
- ISO 8536-4: 2019, Infusion equipment for medical use - Part 4: Infusion sets for single use, gravity feed
- . ISO 80369-7: 2016, Small-bore connectors for liquids and gases in healthcare application -Part 7, Connectors for intravascular or hypodermic applications
- . US FDA Guidance for Industry and FDA Staff, Intravascular Administration Sets Premarket Notification Submissions [510(k)], Issued on July 11, 2008
Bench performance verifications and validations were performed on clearance of the ProSeal™ Injection Site, cleared under K240433 incorporated in the Subject device's additive port:
- Positive pressure fluid leakage test
- Sub-atmospheric pressure air leakage test .
- Stress cracking test
- Resistance to separation from axial load test ●
- . Resistance to separation from unscrewing test
- . Resistance to overriding
- . Device leakage integrity test
- Vapor containment test per NIOSH 2016 draft protocol ●
- Microbial ingress test per FDA guidance and AAMI CN27: 2021 ●
- B. Biocompatibility
In accordance with ISO 10993-1:2018, the Subject device, just like the existing device, is classified as: Externally Communicating Device, Blood Path Indirect, Prolonged Contact (>24hr to 30d). The following testing are referred to, from testing data on the existing devices, cleared, as listed hereunder :
- Cytotoxicity to ISO 10993-5 under K223674 & K240433 ●
- Sensitization to ISO 10993-10 under K223674 & K240433 ●
- Intracutaneous Reactivity to ISO 10993-10 under K223674 & K240433 .
- Acute Systemic Toxicity to ISO 10993-11 under K223674 & K240433 ●
- 14-day Subacute/ Subchronic Acute Systemic Toxicity to ISO 10993-11 under K223674 & . K240433
- . In-vitro Hemolysis Assessment to ISO 10993-4 under K223674 & K240433
- Material Mediated Pyrogenicity to ISO 10993-11 under K223674 & K240433 .
- Chemical Characterization & Toxicological Risk Assessment to ISO 10993-18 & ISO 10993-17 . under K223674 & K240433
Particulate matter testing was conducted in accordance with ISO 15747: 2018, Plastic containers for intravenous injections and USP <788> Particulate Matter in Injections.
{12}------------------------------------------------
## cidit
- C. Sterility, Shipping, and Shelf-Life
The Subject device, like the existing device cleared under K223674, complies with sterilization requirements of ISO 11135:2014, Sterilization of Health Care Products - Ethylene Oxide -Part I: Requirements for Development, Validation and Routine Control of a Sterilization Process for Medical Devices and the following testing/ evaluations:
- Package Integrity Tests per ASTM F1980-16, Standard guide for accelerated aging of sterile barrier systems for medical devices and Sterile Barrier Packaging Testing performed on the proposed device: Seal strength – ASTM F88/F88M-21, Standard test method for seal strength of flexible barrier materials, ASTM F1929-15, Standard test method for detecting seal leaks in porous medical device packaging by dye penetration, EN 868-5: 2009, Packaging materials and systems for medical devices which are to be sterilized – Part 5: Heat and self-sealable pouches and reels of paper and plastic film construction – Requirements and test methods
- Pyrogen Tests per ANSI/AAMI ST72/ 2019, Bacterial endotoxins Test methods, routing ● monitoring, and alternatives to batch testing, USP 40 <151>, Pyrogen test (USP rabbit test), USP-NF <161>, Medical Devices-Bacterial Endotoxin and Pyrogen Tests, USP-NF <85>, Bacterial Endotoxins Test
- . Shelf-life of 3 years has been validated using the FDA recognized standard, ASTM 1980-16, Standard Guide for Accelerated Aging of Sterile Barrier Systems for Medical Devices.
## VIII. Clinical Tests
Not applicable.
#### IX. Conclusion
The differences between the Predicate and the Subject device do not raise any new or different questions of safety or effectiveness. The Subject, eZSURE™ Empty Fluid Container with ProSeal™ Injection Site (fill port) is substantially equivalent to the Predicate, the eZSURE™ Empty Fluid Container (with needle-free valve fill port), with respect to the indications for use, principles of operation and technological characteristics.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.