VUNO Med-Chest X-ray Triage/VUNO Med-CXR Link Triage
K241439 · Vuno, Inc. · QFM · Nov 15, 2024 · Radiology
Device Facts
Record ID
K241439
Device Name
VUNO Med-Chest X-ray Triage/VUNO Med-CXR Link Triage
Applicant
Vuno, Inc.
Product Code
QFM · Radiology
Decision Date
Nov 15, 2024
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 892.2080
Device Class
Class 2
Attributes
AI/ML, Software as a Medical Device, Real-World Evidence
Real-World Evidence
Submission
Device
Sponsor
RWD Sources
RWE Use Summary
Key Tags
K241439 · Nov 15, 2024
VUNO Med-Chest X-ray Triage/VUNO Med-CXR Link Triage
Vuno, Inc.
Retrospectively collected clinical chest X-ray images from medical imaging centers and hospitals
Retrospective clinical data was used to evaluate the classification accuracy and performance of the AI algorithm for triaging pneumothorax and pleural effusion.
Retrospective clinical data; Performance validation; Chest X-ray; AI algorithm
Clinical Evidence
Study Design
Population
Comparator
Key Endpoints
Retrospective performance evaluation
Adult patients (22 years and older) undergoing chest X-rays; Sample Size: 1,190 scans for pneumothorax (716 with, 474 without); 1,997 scans for pleural effusion (1,200 with, 797 without); Number of Sites: Multiple sites across the US
Not applicable for this study
Classification accuracy (ROC AUC, Sensitivity, Specificity) for pneumothorax and pleural effusion
Clinical performance study: 1,200 scans with pleural effusion and 797 scans without pleural effusion.
3 (ABR radiologists)
Indications for Use
VUNO Med-Chest X-ray Triage/VUNO Med-CXR Link Triage is a radiological computer-assisted triage and notification software that analyzes adult chest X-ray images for the presence of pre-specified suspected critical findings (pleural effusion and/or pneumothorax). VUNO Med-Chest X-ray Triage/VUNO Med-CXR Link Triage uses an artificial intelligence algorithm to analyze images for features suggestive of critical findings and provides case-level output available in the PACS/workstation for worklist prioritization or triage. As a passive notification for prioritization-only software tool within standard of care workflow, VUNO Med-Chest X-ray Triage/VUNO Med-CXR Link Triage does not send a proactive alert directly to the appropriately trained medical specialists. VUNO Med-Chest X-ray Triage/VUNO Med-CXR Link Triage is not intended to direct attention to specific portions of an image or to anomalies other than pleural effusion and/or pneumothorax. Its results are not intended to be used on a stand-alone basis for clinical decision-making.
Device Story
VUNO Med-Chest X-ray Triage/VUNO Med-CXR Link Triage is a radiological computer-assisted triage and notification software; analyzes frontal chest X-ray images (DICOM, JPG, PNG) for pleural effusion and pneumothorax. Device uses a locked convolutional neural network (CNN) to identify features suggestive of critical findings; provides case-level output to PACS/workstation worklists. Operates as a passive notification tool; does not alert specialists directly; no image markup provided. Used by radiologists and healthcare professionals in clinical settings; results integrated into standard-of-care workflows. Output assists in prioritizing worklists; does not replace clinical decision-making. Benefits include faster triage of critical findings.
Clinical Evidence
Retrospective clinical performance study using 1,200 scans for pleural effusion and 716 scans for pneumothorax. Pneumothorax results: AUC 0.9883, sensitivity 95.45%, specificity 96.41%. Pleural effusion results: AUC 0.9900, sensitivity 96.53%, specificity 95.11%. Subgroup analysis confirmed robustness across age, gender, view position, and X-ray vendors. Ground truth established by 3 ABR radiologists (min 5 years experience). Performance is comparable to predicate device.
Technological Characteristics
Software-based radiological triage tool; utilizes locked convolutional neural network (CNN) architecture. Compatible with DICOM, JPG, and PNG image formats. Vendor-agnostic; integrates with PACS/workstation via DICOM protocol. Risk management complies with ISO 14971:2019. No hardware components.
Indications for Use
Indicated for adult patients (22+ years) to triage chest X-ray images for suspected pleural effusion and/or pneumothorax. Intended for use by radiologists and healthcare professionals as a passive notification tool for worklist prioritization within standard clinical workflows. Not for stand-alone diagnostic use.
Regulatory Classification
Identification
Radiological computer aided triage and notification software is an image processing prescription device intended to aid in prioritization and triage of radiological medical images. The device notifies a designated list of clinicians of the availability of time sensitive radiological medical images for review based on computer aided image analysis of those images performed by the device. The device does not mark, highlight, or direct users' attention to a specific location in the original image. The device does not remove cases from a reading queue. The device operates in parallel with the standard of care, which remains the default option for all cases.
Special Controls
Radiological computer aided triage and notification software must comply with the following special controls: 1. Design verification and validation must include: i. A detailed description of the notification and triage algorithms and all underlying image analysis algorithms including, but not limited to, a detailed description of the algorithm inputs and outputs, each major component or block, how the algorithm affects or relates to clinical practice or patient care, and any algorithm limitations. ii. A detailed description of pre-specified performance testing protocols and dataset(s) used to assess whether the device will provide effective triage (e.g., improved time to review of prioritized images for pre-specified clinicians). iii. Results from performance testing that demonstrate that the device will provide effective triage. The performance assessment must be based on an appropriate measure to estimate the clinical effectiveness. The test dataset must contain sufficient numbers of cases from important cohorts (e.g., subsets defined by clinically relevant confounders, effect modifiers, associated diseases, and subsets defined by image acquisition characteristics) such that the performance estimates and confidence intervals for these individual subsets can be characterized with the device for the intended use population and imaging equipment. iv. Standalone performance testing protocols and results of the device. v. Appropriate software documentation (e.g., device hazard analysis; software requirements specification document; software design specification document; traceability analysis; description of verification and validation activities including system level test protocol, pass/fail criteria, and results). 2. Labeling must include the following: i. A detailed description of the patient population for which the device is indicated for use. ii. A detailed description of the intended user and user training that addresses appropriate use protocols for the device. iii. Discussion of warnings, precautions, and limitations must include situations in which the device may fail or may not operate at its expected performance level (e.g., poor image quality for certain subpopulations), as applicable. iv. A detailed description of compatible imaging hardware, imaging protocols, and requirements for input images. v. Device operating instructions. vi. A detailed summary of the performance testing, including: test methods, dataset characteristics, triage effectiveness (e.g., improved time to review of prioritized images for pre-specified clinicians), diagnostic accuracy of algorithms informing triage decision, and results with associated statistical uncertainty (e.g., confidence intervals), including a summary of subanalyses on case distributions stratified by relevant confounders, such as lesion and organ characteristics, disease stages, and imaging equipment.
*Classification.* Class II (special controls). The special controls for this device are:(1) Design verification and validation must include:
(i) A detailed description of the notification and triage algorithms and all underlying image analysis algorithms including, but not limited to, a detailed description of the algorithm inputs and outputs, each major component or block, how the algorithm affects or relates to clinical practice or patient care, and any algorithm limitations.
(ii) A detailed description of pre-specified performance testing protocols and dataset(s) used to assess whether the device will provide effective triage (
*e.g.,* improved time to review of prioritized images for pre-specified clinicians).(iii) Results from performance testing that demonstrate that the device will provide effective triage. The performance assessment must be based on an appropriate measure to estimate the clinical effectiveness. The test dataset must contain sufficient numbers of cases from important cohorts (
*e.g.,* subsets defined by clinically relevant confounders, effect modifiers, associated diseases, and subsets defined by image acquisition characteristics) such that the performance estimates and confidence intervals for these individual subsets can be characterized with the device for the intended use population and imaging equipment.(iv) Stand-alone performance testing protocols and results of the device.
(v) Appropriate software documentation (
*e.g.,* device hazard analysis; software requirements specification document; software design specification document; traceability analysis; description of verification and validation activities including system level test protocol, pass/fail criteria, and results).(2) Labeling must include the following:
(i) A detailed description of the patient population for which the device is indicated for use;
(ii) A detailed description of the intended user and user training that addresses appropriate use protocols for the device;
(iii) Discussion of warnings, precautions, and limitations must include situations in which the device may fail or may not operate at its expected performance level (
*e.g.,* poor image quality for certain subpopulations), as applicable;(iv) A detailed description of compatible imaging hardware, imaging protocols, and requirements for input images;
(v) Device operating instructions; and
(vi) A detailed summary of the performance testing, including: test methods, dataset characteristics, triage effectiveness (
*e.g.,* improved time to review of prioritized images for pre-specified clinicians), diagnostic accuracy of algorithms informing triage decision, and results with associated statistical uncertainty (*e.g.,* confidence intervals), including a summary of subanalyses on case distributions stratified by relevant confounders, such as lesion and organ characteristics, disease stages, and imaging equipment.
{1}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA acronym and full name on the right. The Department of Health & Human Services logo features a stylized human figure, while the FDA part includes the acronym in a blue square and the full name "U.S. Food & Drug Administration" in blue text.
VUNO Inc. % Priscilla Chung RA Associate LK Consulting Group USA, Inc. 18881 Von Karman Ave STE 160 Irvine. California 92612
November 15, 2024
Re: K241439
Trade/Device Name: VUNO Med-Chest X-ray Triage/VUNO Med-CXR Link Triage Regulation Number: 21 CFR 892.2080 Regulation Name: Radiological computer aided triage and notification software Regulatory Class: Class II Product Code: QFM Dated: May 21, 2024 Received: October 9, 2024
Dear Priscilla Chung:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{2}------------------------------------------------
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510/k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change reguires premarket review. the OS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-device-advicecomprehensive-regulatory-assistance/unique-device-identification-system-udi-system.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
{3}------------------------------------------------
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely.
Jessica Lamb
Jessica Lamb, Ph.D. Assistant Director Imaging Software Team DHT8B: Division of Radiological Imaging Devices and Electronic Products OHT8: Office of Radiological Health Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{4}------------------------------------------------
# Indications for Use
Form Approved: OMB No. 0910-0120 Expiration Date: 07/31/2026 See PRA Statement below.
Submission Number (if known)
K241439
Device Name
VUNO Med-Chest X-ray Triage/VUNO Med-CXR Link Triage
Indications for Use (Describe)
VUNO Med-Chest X-ray Triage/VUNO Med-CXR Link Triage is a radiological computer-assisted triage and notification software that analyzes adult chest X-ray images for the presence of prespecified suspected critical findings (pleural effusion and/or pneumothorax). VUNO Med-Chest X-ray Triage/VUNO Med-CXR Link Triage uses an artificial intelligence algorithm to analyze images for features suggestive of critical findings and provides case-level output available in the PACS/ workstation for worklist prioritization or triage.
As a passive notification for prioritization-only software tool within standard of care workflow, VUNO Med-Chest X-ray Triage/VUNO Med-CXR Link Triage does not send a proactive alert directly to the appropriately trained medical specialists. VUNO Med-Chest X-ray Triage/VUNO Med-CXR Link Triage is not intended to direct attention to specific portions of an image or to anomalies other than pleural effusion and/or pneumothorax. Its results are not intended to be used on a stand-alone basis for clinical decision-making.
Type of Use (Select one or both, as applicable)
> Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{5}------------------------------------------------
# K241439
# 510(k) Summary
(K241439)
This summary of 510(k) information is being submitted in accordance with requirements of 21 CFR Part 807.92.
#### 1. Date: 10/08/2024
#### 2. Applicant / Submitter
VUNO Inc. 9F, 479, Gangnam-daero, Seocho-gu Seoul, 06541, Republic of Korea Tel : +82-2-515-6646 Fax : +82-2-515-6647
#### 3. U.S. Designated Agent
Priscilla Chung LK Consulting Group USA, Inc. 18881 Von Karman Ave. STE 160 Irvine, CA 92612 Tel: 714.202.5789 Fax: 714.409.3357 Email: juhee.c@LKconsultingGroup.com
#### 4. Trade/Proprietary Name:
VUNO Med-Chest X-ray Triage/VUNO Med-CXR Link Triage
#### 5. Common Name:
Radiological Computer Assisted Prioritization Software for Lesions
#### 6. Classification:
21 CFR 892.2080 QFM (Class II) Radiological computer aided triage and notification software
#### 7. Device Description:
VUNO Med-Chest X-ray Triage/VUNO Med-CXR Link Triage is an automated computerassisted triage and notification software that analyzes adult chest X-ray images for the presence of pleural effusion and pneumothorax. It is based on an artificial intelligence
{6}------------------------------------------------
analysis model, specifically a convolutional network (CNN), which employs deep learning technology to learn features from data.
The training data is sourced from 4 distinct sites of South Korea and India data provider, including medical imaging centers, data partners, and medical hospitals, and over 13 different modality manufacturers such as GE. Philps, FUJI, Canon, Samsung, SIEMENS, etc.
A "locked" algorithm is used, and the same input gives the same results every time. The software receives an image of a frontal chest radiograph and automatically analyzes it for the presence of pre-specified critical findings. If any findings are suspected, the image is flagged, and a passive notification is provided to the user. Subsequently, trained radiologists or healthcare professionals should make the final decision which is the standard of care at present. A user interface is provided for visualization, displaying the loaded image and any detected findings.
The data can be transmitted from Picture Archive and Communications Systems (PACS) using the DICOM protocol.
# 8. Indication for use:
VUNO Med-Chest X-ray Triage/VUNO Med-CXR Link Triage is a radiological computerassisted triage and notification software that analyzes adult chest X-ray images for the presence of pre-specified suspected critical findings (pleural effusion and/or pneumothorax). VUNO Med-Chest X-ray Triage/VUNO Med-CXR Link Triage uses an artificial intelligence algorithm to analyze images for features suggestive of critical findings and provides caselevel output available in the PACS/workstation for worklist prioritization or triage. As a passive notification for prioritization-only software tool within standard of care workflow, VUNO Med-Chest X-ray Triage/VUNO Med-CXR Link Triage does not send a proactive alert directly to the appropriately trained medical specialists. VUNO Med-Chest Xray Triage/VUNO Med-CXR Link Triage is not intended to direct attention to specific portions of an image or to anomalies other than pleural effusion and/or pneumothorax. Its results are not intended to be used on a stand-alone basis for clinical decision-making.
# 9. Predicate Device:
qXR-PTX-PE (K230899) by Oure.ai Technologies
# 10. Substantial Equivalence:
# Comparison Table
The subject device is substantially equivalent to the following predicate device:
- 510(k) number: K230899
- · Device name: qXR-PTX-PE
{7}------------------------------------------------
| | Subject Device | Predicate Device |
|-------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Device Name | VUNO Med-Chest X-ray Triage/VUNO Med-CXR Link Triage | qXR-PTX-PE |
| 510(k) Number | K241439 | K230899 |
| Regulation | 21 CFR 892.2080 | 21 CFR 892.2080 |
| Regulation<br>Description | Radiological computer aided triage and notification software | Radiological computer aided triage and notification software |
| Product Code | QFM | QFM |
| Device Type | Radiological Computer-Assisted Prioritization Software For Lesions | Radiological Computer-Assisted Prioritization Software For Lesions |
| Manufacturer | VUNO Inc. | Qure.ai Technologies |
| Intended use/<br>Indications for<br>Use | VUNO Med-Chest X-ray Triage/VUNO Med-CXR Link Triage is a radiological computer-assisted triage and notification software that analyzes adult chest X-ray images for the presence of pre-specified suspected critical findings (pleural effusion and/or pneumothorax). VUNO Med-Chest X-ray Triage/VUNO Med-CXR Link Triage uses an artificial intelligence algorithm to analyze images for features suggestive of critical findings and provides case-level output available in the PACS/workstation for worklist prioritization or triage. As a passive notification for prioritization-only software tool within standard of care workflow, VUNO Med-Chest X-ray Triage/VUNO Med-CXR Link Triage does not send a proactive alert directly to the appropriately trained medical specialists. VUNO Med-Chest X-ray Triage/VUNO Med-CXR Link Triage is not intended to direct attention to specific portions of an image or to anomalies other than pleural effusion and/or pneumothorax. Its results are not intended to be used on a stand-alone basis for clinical | qXR-PTX-PE is a radiological computer-assisted triage and notification software that analyzes adult chest X-ray images for the presence of pre-specified suspected critical findings (pleural effusion and/or pneumothorax). qXR-PTX-PE uses an artificial intelligence algorithm to analyze images for features suggestive of critical findings and provides case-level output available in the PACS/workstation for worklist prioritization or triage. As a passive notification for prioritization-only software tool within standard of care workflow, qXR-PTX-PE does not send a proactive alert directly to the appropriately trained medical specialists. qXR-PTXPE is not intended to direct attention to specific portions of an image or to anomalies other than pleural effusion and/or pneumothorax. Its results are not intended to be used on a standalone basis for clinical decision-making. |
| | Subject Device | Predicate Device |
| | decision-making. | |
| Intended User | Trained radiologists and healthcare<br>professionals | Radiologists, clinicians, and other<br>appropriately trained medical<br>specialists qualified to read chest<br>radiographs |
| Modality | Chest X-ray | Chest X-ray |
| Target Clinical<br>Conditions | Pneumothorax, Pleural effusion on<br>Chest/Lung Frontal Chest X-rays | Pneumothorax, Pleural effusion on<br>Chest/Lung Frontal Chest X-rays |
| Algorithm for<br>pre-specified<br>critical findings<br>detection | VUNO Med-Chest X-ray Triage/<br>VUNO Med-CXR Link Triage uses<br>an AI algorithm to detect<br>pneumothorax and pleural effusion<br>on chest X-ray images.<br>VUNO Med-Chest X-ray Triage/<br>VUNO Med-CXR Link Triage uses<br>a vendor agnostic algorithm<br>compatible with DICOM chest X-<br>ray images. | qXR-PTX-PE uses an AI algorithm<br>to detect pneumothorax and pleural<br>effusion on chest X-ray images.<br>qXR-PTX-PE uses a vendor<br>agnostic algorithm compatible with<br>DICOM<br>chest X-ray images. |
| Notification<br>only/<br>Parallel<br>workflow | Yes | Yes |
| Input format | DICOM, JPG, PNG | DICOM |
| Device output in<br>case of positive<br>detection | When deployed on other<br>radiological imaging equipment,<br>VUNO Med-Chest X-ray<br>Triage/VUNO Med-CXR Link<br>Triage will automatically run after<br>image acquisition to perform triage.<br>It displays the analysis result<br>through the worklist interface of<br>PACS/workstation.<br>No markup of the conditions will be<br>done on the original image.<br>Secondary capture of the device will<br>indicate the presence of findings<br>suspicious of pneumothorax or<br>pleural effusion.<br>Upon image acquisition from other<br>radiological imaging equipment<br>(e.g. X-ray systems) a passive<br>notification is generated. | When deployed on other<br>radiological imaging equipment,<br>qXR-PTX-PE will automatically run<br>after image acquisition to perform<br>triage. It displays the analysis result<br>through the worklist interface of<br>PACS/workstation.<br>No markup of the conditions will be<br>done on the original image.<br>Secondary capture of the device will<br>indicate the presence of findings<br>suspicious of pneumothorax or<br>pleural effusion.<br>Upon image acquisition from other<br>radiological imaging equipment<br>(e.g. X-ray systems) a passive<br>notification is generated. |
| Notification<br>(i.e., recipient, | Passive notification. Images with | Passive notification. Images with |
| | suspicion of pneumothorax and/or | suspicion of pneumothorax and/or |
| | Subject Device | Predicate Device |
| timing and<br>means of<br>notification) | pleural effusion are flagged in<br>PACS/workstation/DICOM viewer. | pleural effusion are flagged in<br>PACS/workstation/DICOM viewer. |
| Where<br>generated<br>results(i.e.,<br>DICOM files)<br>are stored | PACS/Workstation/DICOM viewer | PACS/Workstation/DICOM viewer |
| Performance<br>level<br>- Timing of<br>notification | The average time taken for the<br>notification to travel from VUNO<br>Med-Chest X-ray Triage/VUNO<br>Med-CXR Link Triage to the point<br>at which the result is displayed in<br>the destination Picture Archiving<br>and Communication System(PACS)<br>or workstation/digital radiographic<br>processing system (ex. digital<br>radiography, digital X-ray system<br>etc.) is below 10 seconds. | The average time taken for the<br>notification to travel from qXR-<br>PTX-PE to the point at which the<br>result is displayed in the destination<br>Picture Archiving and<br>Communication System(PACS) or<br>workstation/digital radiographic<br>processing system (ex. digital<br>radiography, digital X-ray system<br>etc.) is 10 seconds. |
| Performance<br>level<br>- Accuracy of<br>classification | Pneumothorax<br>ROC AUC > 0.95<br>AUC: 0.9883 (95% CI: [0.9815,<br>0.9939])<br>Sensitivity 95.45 % (95% CI:<br>[92.01, 97.71])<br>Specificity 96.41% (95% CI: [94.32,<br>97.90])<br>Pleural Effusion<br>ROC AUC > 0.95<br>AUC: 0.9900 (95% CI: [0.9863,<br>0.9932])<br>Sensitivity 96.53% (95% CI: [94.24,<br>98.09])<br>Specificity 95.11% (95% CI: [93.37,<br>96.50]) | Pneumothorax<br>ROC AUC > 0.95<br>AUC: 0.9894 (95% CI: [0.9829,<br>0.9980])<br>Sensitivity 94.53% (95% CI: [90.42,<br>97.24])<br>Specificity 96.36% (95% CI: [94.07,<br>97.95])<br>Pleural Effusion<br>ROC AUC > 0.95<br>AUC: 0.989 (95% CI: [0.9847,<br>0.9944])<br>Sensitivity 96.22% (95% CI: [93.62,<br>97.97])<br>Specificity 94.90% (95% CI: [93.04,<br>96.39]) |
{8}------------------------------------------------
{9}------------------------------------------------
# Substantial Equivalence Discussion
The comparison table shows the description of the subject device and predicate device in each aspect. The comparison rationale and gap between the subject device and the predicate device are provided below.
- Indications for use
{10}------------------------------------------------
The indications for use are the same. Both are intended to detect presence of critical findings(pleural effusion and /or pneumothorax).
The intended user, modality, and target clinical condition are the same as the predicate device.
- · Algorithm for pre-specified critical findings detection Both devices use AI algorithm to detect findings. As referred to in the literature, the predicate device uses a series of convolutional neural networks(CNNs) trained to identify different abnormalities on frontal CXRs. The algorithm first resizes and normalizes CXRs to decrease variations in the acquisition process, then applies modifications in either densenet or resnet network architectures to separate CXRs from radiographs of other anatomies. Subsequently, multiple networks, including densenets and resnets, are applied for individual CXR findings.1 Similarly, the subject device uses convolutional networks(CNNs) to analyze prespecified findings. The algorithm firstly resizes and normalizes the image through a preprocessing step, then proceeds through the classification model to detect findings. The algorithm used for both predicate and subject device is equivalent, although there may be minor differences. However, the accuracy of the output is verified through performance test.
- · Notification only/Parallel workflow Both devices are used for notification/triage. The output of device is not directly used to the diagnosis, and the final decision should be performed by user. It is used parallel to the standard workflow.
- Input format
The predicate device generates DICOM format image, while the subject device generates DICOM, JPG, and PNG format.
Both devices utilize the DICOM format. which is the international standard for medical images and related information. It defines the formats for medical images that can be exchanged with the data and quality necessary for clinical use. Additionally, for use convenience, the subject device accepts JPG and PNG format. When user uploads the image, DICOM, JPG or PNG format is acceptable. This functionality is verified and validated through software testing. The input format is considered to be the equivalent.
- Device output in case of positive detection Both devices provide same output in case of positive detection. There is no markup of the conditions, and the secondary capture of result will be sent to the worklist interface of PACS or workstation. Upon image acquisition from other radiological imaging equipment (e.g. X-ray systems) a passive notification is generated.
- · Notification (i.e., recipient, timing and means of notification)
<sup>1</sup> Parisa Kaviani, et al., Performance of a Chest Radiography AI Algorithm for Detection of Mislabeled Findings: A Multicenter Study, Diagnostics 2022, 12(9), 2086;
{11}------------------------------------------------
Recipient, timing and means of notification is the same between subject and predicate device.
A passive notification is provided in both devices. Images with suspicion of findings(pneumothorax and/or pleural effusion) are flagged in PACS/workstation/DICOM viewer. The performance result of notification timing is described below.
- · Where generated results(i.e., DICOM files) are stored The generated results are stored in the PACS/Workstation/DICOM viewer which is the same between the subject and predicated device.
- · Performance level Timing of notification The timing of notification performance is equivalent. The average time for displaying results was 10 seconds for the predicate device, and 7.86 seconds for the subject device.
- Performance level Accuracy of classification The accuracy of classification performance is equivalent. Both Pneumothorax and Pleural Effusion exceed ROC AUC 0.95.
Clinical studies were conducted on retrospectively collected chest X-rays to evaluate the performance of subject device for triaging of pneumothorax and pleural effusion.
The subgroup analysis was performed to demonstrate the robustness of the device performance under diverse conditions. Every subgroup, age, gender, view position, dataset, and vendor outperformed the target performance.
- · Safety
The risks associated with the use of a medical device should be established, implemented, documented, and maintained to ensure the device's safety. The risk management process complies with ISO 14971:2019. Through risk analysis, the device is determined to be safe and beneficial.
VUNO Med-Chest X-ray Triage/VUNO Med-CXR Link Triage is a radiological computerassisted triage and notification software that analyzes adult chest X-ray images for the presence of pre-specified suspected critical findings (pleural effusion and/or pneumothorax). The intended use and other aspects are considered to be the same as the predicate device. However, the algorithm used for data analysis might have difference. Although both use convolutional neural networks(CNNs), which are one of the deep learning models, the detailed mechanisms are different.
Despite the minor differences, the subject device shows reliable output in the performance test. And an analysis results are not intended to be used on a stand-alone basis for clinical decisionmaking.
In conclusion, the subject device is substantially equivalent to the predicate device, and no additional risks are presented in the perspective of safety and effectiveness.
{12}------------------------------------------------
# 11. Performance Data:
# Software Validation
Software Software verification and validation testing were conducted, and documentation was provided as recommended by FDA's Guidance for Industry and FDA Staff, "Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices."
# Performance Testing - Clinical
The accuracy of classification performance and the timing of notification are validated through clinical tests. These tests are conducted on retrospectively collected chest X-rays to evaluate the performance of the device for triaging pneumothorax and pleural effusion. Both pneumothorax and pleural effusion exceed the acceptance criteria, with ROC AUC values of 0.95.
A total dataset of 1,200 scans for pleural effusion and 716 scans for pneumothorax were used in the test dataset to evaluate the standalone performance of VUNO Med-Chest X-ray/VUNO Med-CXR Link Triage for triaging of pleural effusion and pneumothorax, in terms of classification accuracy. The test dataset is independent of the training dataset, with each sourced from a different country.
Each scan sourced from different subjects and was obtained through a general inclusion/exclusion criterion, ensuring comprehensive subgroup representation.
The dataset for pneumothorax included 716 scans with pneumothorax and 474 scans without pneumothorax). It is sourced from various regions of the US: Midwest, West, Northeast, and South. Various demographic and medical characteristics such as gender(males and females), age(22 years and older), race/ethnicity(White. Hispanic and Latino, Black, and Asian), radiographic view position (PA and AP), and vendors(Siemens, Konica Minolta, Samsung Electronics, Canon, GE Medical Systems, Carestream Health, and Fujifilm) are considered to ensure that the device's performance is consistent across the intended population. The dataset consisted of clinical confounders that included opacities, emphysema, scarring, mediastinal widening, pleural thickening, and presence of hardware. The datasets were also obtained from diverse X-ray device vendors, such as GE, Philips, Siemens, Samsung electronics, Konica, Cannon, Fuji film, with an average exposure of 112±13 kVp (range from 49 kVp to 135 kVp) to ensure consistent performance. The predicate device has used 613 scans (201 scans with pneumothorax and 412 scans without pneumothorax) for performance testing of pneumothorax. The ground truth was established by 3 ABR radiologists with a minimum of 5 years of experience.
The AUC of the subject device in triaging scans with findings suspicious of pneumothorax exceeded the acceptance criteria with AUC 98.83 (95%CI, 98.15 - 99.39), Sensitivity 95.45 (92.01 - 97.71), and Specificity 96.41 (94.32 - 97.90).
The predicated device was reported as (AUC 98.94 (95% CI, 98.28 - 99.82)), Sensitivity 94.53 (90.42-97.24) and Specificity 96.36(94.07-97.95).
{13}------------------------------------------------
| AUC (95% CI) | Sensitivity (95% CI) | Specificity (95% CI) |
|-----------------------|-----------------------|-----------------------|
| 98.83 (98.15 - 99.39) | 95.45 (92.01 - 97.71) | 96.41 (94.32 - 97.90) |
The subgroup analysis was performed to demonstrate the robustness of the device performance under diverse conditions. Every subgroup, age, gender, view position, dataset, and vendor outperformed the target performance.
Test dataset for pleural effusion included 1,200 scans with pleural effusion and 797 scans without pleural effusion) It is sourced from various regions of the US: Midwest, West, Northeast, and South. Various demographic and medical characteristics such as gender(males and females), age(22 years and older), race/ethnicity(White, Hispanic and Latino, Black, and Asian), radiographic view position (PA and AP), and vendors(Siemens, Konica Minolta, Samsung Electronics, Canon, GE Medical Systems, Carestream Health, and Fujifilm) are considered to ensure that the device's performance is consistent across the intended population. The dataset consisted of clinical confounders that included opacities, emphysema, scarring, mediastinal widening, pleural thickening, and presence of hardware. The datasets were also obtained from diverse X-ray device vendors, such as GE, Philips, Siemens, Samsung electronics, Konica, Cannon, Fuji film, with an average exposure of 112±11 kVp (range from 49 kVp to 137 kVp) to ensure consistent performance. The predicate device has used 1,070 scans (344 scans with pleural effusion and 726 scans without pleural effusion) for performance testing of pleural effusion.
The AUC of the subject device in triaging scans with findings suspicious of pleural effusion exceeded the acceptance criteria with AUC 99.00 (95%CI, 98.63 - 99.32), Sensitivity 96.53 (94.24 - 98.09), and Specificity 95.11 (93.37 - 96.50).
The predicated device was reported as AUC 98.90 (98.47 -99.44)), sensitivity 96.22 (93.62-97.97) and specificity 94.90 (93.04-96.39).
| AUC (95% CI) | Sensitivity (95% CI) | Specificity (95% CI) |
|-----------------------|-----------------------|-----------------------|
| 99.00 (98.63 - 99.32) | 96.53 (94.24 - 98.09) | 95.11 (93.37 - 96.50) |
[Table 2. Overall results of classification accuracy of Pleural effusion]
The subgroup analysis was performed to demonstrate the robustness of the device performance under diverse conditions. Every subgroup, age, gender, view position, dataset, and vendor outperformed the target performance.
The average time for displaying results was 10 seconds for the predicate device, and also below 10 seconds for the subject device.
{14}------------------------------------------------
### 12. Conclusion:
The subject device is substantially equivalent in the areas of technical characteristics, general function, application, and indications for use. The new device does not introduce a fundamentally new scientific technology, and the device has been validated through performance test. Therefore, we conclude that the subject device described in this submission is substantially equivalent to the predicate device.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.