K241400 · Crown Aesthetics · QAI · Aug 29, 2024 · General, Plastic Surgery
Device Facts
Record ID
K241400
Device Name
SkinPen Precision Elite System
Applicant
Crown Aesthetics
Product Code
QAI · General, Plastic Surgery
Decision Date
Aug 29, 2024
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 878.4430
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The SkinPen® Precision Elite System is a microneedling device and accessories intended to be used as a treatment to improve the appearance of wrinkles of the neck for Fitzpatrick skin types II - IV and to improve the appearance of facial acne scars in adults with all Fitzpatrick skin types aged 22 years and older.
Device Story
Microneedling device for aesthetic use; consists of handpiece, sterile single-use needle cartridge, charging base, and BioSheath. Used in clinical settings by trained healthcare professionals. Device creates controlled micro-injuries in skin to improve appearance of neck wrinkles and facial acne scars. Subject device adds LCD screen and NFC chip for cartridge authentication/verification to ensure correct installation and prevent cartridge reuse. Operation involves reciprocating needle motion (105-128.3 Hz; 1470-1797 punctures/sec) with adjustable protrusion (0-2.5 mm). Output is physical skin perforation. Clinical benefit derived from mechanical stimulation of skin tissue.
Clinical Evidence
No clinical data was conducted as part of this submission; bench testing only.
Technological Characteristics
Microneedling handpiece with 14-needle round cartridge. Reciprocating rate 105-128.3 Hz; puncture rate 1470-1797 punctures/sec. Needle protrusion 0-2.5 mm. Standards: IEC 60601-1 (3.2), IEC 60601-1-2 (4.1), ISO 10993-1 (2018), IEC 62304 (1.1), ISO 11135 (2014), IEC 60601-1-6 (3.2). Biocompatible materials. NFC-based cartridge authentication. LCD user interface.
Indications for Use
Indicated for adults aged 22+ with Fitzpatrick skin types II-IV for neck wrinkles and all Fitzpatrick skin types for facial acne scars.
Regulatory Classification
Identification
A microneedling device for aesthetic use is a device using one or more needles to mechanically puncture and injure skin tissue for aesthetic use. This classification does not include devices intended for transdermal delivery of topical products such as cosmetics, drugs, or biologics.
Special Controls
The technical specifications and needle characteristics must be identified, including needle length, geometry, maximum penetration depth, and puncture rate. Non-clinical performance data must demonstrate that the device performs as intended under anticipated conditions of use. The following performance characteristics must be tested: (i) Accuracy of needle penetration depth and puncture rate; (ii) Safety features built into the device to protect against cross-contamination, including fluid ingress protection; and (iii) Identification of the maximum safe needle penetration depth for the device for the labeled indications for use. Performance data must demonstrate the sterility of the patient-contacting components of the device. Performance data must support the shelf life of the device by demonstrating continued sterility, package integrity, and device functionality over the intended shelf life. Performance data must demonstrate the electrical safety and electromagnetic compatibility (EMC) of all electrical components of the device. Software verification, validation, and hazard analysis must be performed for all software components of the device. The patient-contacting components of the device must be demonstrated to be biocompatible. Performance data must validate the cleaning and disinfection instructions for reusable components of the device. Labeling must include the following: (i) Information on how to operate the device and its components and the typical course of treatment; (ii) A summary of the device technical parameters, including needle length, needle geometry, maximum penetration depth, and puncture rate; (iii) Validated methods and instructions for reprocessing of any reusable components: (iv) Disposal instructions; and (v) Shelf life. Patient labeling must be provided and must include: (i) Information on how the device operates and the typical course of treatment; (ii) The probable risks and benefits associated with use of the device; and (iii) Post-operative care instructions.
*Classification.* Class II (special controls). The special controls for this device are:(1) The technical specifications and needle characteristics must be identified, including needle length, geometry, maximum penetration depth, and puncture rate.
(2) Non-clinical performance data must demonstrate that the device performs as intended under anticipated conditions of use. The following performance characteristics must be tested:
(i) Accuracy of needle penetration depth and puncture rate;
(ii) Safety features built into the device to protect against cross-contamination, including fluid ingress protection; and
(iii) Identification of the maximum safe needle penetration depth for the device for the labeled indications for use.
(3) Performance data must demonstrate the sterility of the patient-contacting components of the device.
(4) Performance data must support the shelf life of the device by demonstrating continued sterility, package integrity, and device functionality over the intended shelf life.
(5) Performance data must demonstrate the electrical safety and electromagnetic compatibility (EMC) of all electrical components of the device.
(6) Software verification, validation, and hazard analysis must be performed for all software components of the device.
(7) The patient-contacting components of the device must be demonstrated to be biocompatible.
(8) Performance data must validate the cleaning and disinfection instructions for reusable components of the device.
(9) Labeling must include the following:
(i) Information on how to operate the device and its components and the typical course of treatment;
(ii) A summary of the device technical parameters, including needle length, needle geometry, maximum penetration depth, and puncture rate;
(iii) Validated methods and instructions for reprocessing of any reusable components;
(iv) Disposal instructions; and
(v) A shelf life.
(10) Patient labeling must be provided and must include:
(i) Information on how the device operates and the typical course of treatment;
(ii) The probable risks and benefits associated with use of the device; and
(iii) Postoperative care instructions.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food & Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA logo on the right. The FDA logo features the letters 'FDA' in a blue square, followed by the words 'U.S. FOOD & DRUG ADMINISTRATION' in blue text.
August 29, 2024
Crown Aesthetics Marie Fogartie Sr. Director of Regulatory Affairs 5005 Lyndon B. Johnson Frwy. Suite 370 Dallas, Texas 75244
Re: K241400
Trade/Device Name: SkinPen Precision Elite System Regulation Number: 21 CFR 878.4430 Regulation Name: Microneedling Device For Aesthetic Use Regulatory Class: Class II Product Code: QAI Dated: August 1, 2024 Received: August 1, 2024
Dear Marie Fogartie:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}------------------------------------------------
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30. Design controls; 21 CFR 820.90. Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Image /page/1/Picture/7 description: The image shows the name "Julie A. Morabito -S" in a simple, sans-serif font. The name is arranged vertically, with "Julie A." on the top line and "Morabito -S" on the bottom line. The text is black against a white background, making it easily readable.
Julie Morabito, Ph.D. Assistant Director DHT4B: Division of Plastic and Reconstructive Surgery Devices
{2}------------------------------------------------
OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{3}------------------------------------------------
# Indications for Use
Submission Number (if known)
K241400
Device Name
SkinPen Precision Elite System
Indications for Use (Describe)
The SkinPen® Precision Elite system is a microneedling device and accessories intended to be used as a treatment to improve the appearance of wrinkles of the neck for Fitzpatrick skin types II IV and to improve the appearance of facial acne scars in adults with all Fitzpatrick skin types aged 22 vears and older.
Type of Use (Select one or both, as applicable)
> Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
# CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
## *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{4}------------------------------------------------
# 510(K) SUMMARY
Date Prepared: May 1, 2024
## 1. Submitter and Owner:
Bellus Medical LLC, dba Crown Aesthetics 5005 Lyndon B. Johnson Freeway, Suite 370 Dallas, Texas 75244, USA
- 2. Official Correspondent: Marie Fogartie, Sr. Director of Regulatory Affairs Phone: 423-630-2296 Email: mfogartie@crownlaboratories.com
- 3. Submission Type: Traditional 510(k)
- 4. Proprietary Name: SkinPen® Precision Elite System
#### 5. Device Classification Information:
Classification Name: Microneedling Device for Aesthetic Use Regulation Number: 878.4430 Product Code: QAI Device Class: 2
## 6. Predicate Device: K220506 SkinPen Precision System
#### 7. Device Description
The SkinPen® Precision Elite System consists of a microneedling pen handpiece (SkinPen® Precision Elite) and a sterile needle cartridge (SkinPen® Precision Elite Cartridge). The accessories are a charging base with power adaptor and a BioSheath to cover the handpiece.
#### 8. Indications/Intended use:
The SkinPen® Precision Elite System is a microneedling device and accessories intended to be used as a treatment to improve the appearance of wrinkles of the neck for Fitzpatrick skin types II - IV and to improve the appearance of facial acne scars in adults with all Fitzpatrick skin types aged 22 years and older.
#### 9. Indications for Use Comparison:
| | Subject Device | Predicate Device<br>K220506 | Comparison |
|--------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Indication for Use | The SkinPen® Precision Elite System is a microneedling device and accessories intended to be used as a treatment to improve the appearance of wrinkles of the neck for Fitzpatrick skin types II - IV and to improve the appearance of facial acne scars in adults with all Fitzpatrick skin types aged 22 years and older. | The SkinPen® Precision System is a microneedling device and accessories intended to be used as a treatment to improve the appearance of wrinkles of the neck for Fitzpatrick skin types II - IV and to improve the appearance of facial acne scars in adults with all Fitzpatrick skin types aged 22 years and older. | Same. |
| Characteristic | Subject Device | Predicate Device<br>K220506 | Comparison |
| Trade Name | SkinPen® Precision<br>Elite System | SkinPen® Precision<br>System | Same |
| Classification Name | Microneedling device<br>for aesthetic use | Microneedling device<br>for aesthetic use | Same |
| Classification | 2 | 2 | Same |
| Product Code | QAI | QAI | Same |
| Regulation No. | 878.4430 | 878.4430 | Same |
| Prescription Use | Yes | Yes | Same |
| Target Population | Adults age 22 and<br>over | Adults age 22 and<br>over | Same |
| Target User | Healthcare<br>professionals trained<br>in use of the device | Healthcare<br>professionals trained<br>in use of the device | Same |
| Environment | Clinical | Clinical | Same |
| Treatment Area | Face and Neck | Face and neck | Same |
| Shape of needle cartridge | Round | Round | Same |
| Use of needle cartridge | Sterile, single use | Sterile, single use | Same |
| Needles | 14 total | 14 total | Same |
| Needle Protrusion settings | 0–2.5 mm | 0–2.5 mm | Same |
| Max. Needle Length used<br>in the clinical study | 2.5mm | 2.5mm | Same |
| Frequency | 105-128.3 Hz | 105-128.3 Hz | Same |
| Puncture rate | 1470-1797<br>punctures/sec | 1470-1797<br>punctures/sec | Same |
| User Interface | LCD screen and<br>On/Off button | On/Off button | Different. The new<br>LCD screen allows<br>users to see whether<br>the cartridge is<br>correctly installed<br>and verify that the<br>cartridge is not being<br>re-used. Software<br>Verification and<br>Validation was<br>performed for this<br>software change. This<br>function does not<br>impact safety or<br>efficacy of the device. |
| Cartridge<br>authentication/verification | NFC Chip | Manual lock out | Different. Cartridges<br>now can be<br>authenticated with an<br>NFC chip. This<br>function does not<br>impact safety or<br>efficacy of the device. |
{5}------------------------------------------------
# 10. Technological characteristics comparison:
{6}------------------------------------------------
# 510(K) SUMMARY
#### 11. Non-clinical performance testing
Performance testing was conducted on the cartridge's reciprocating rate to ensure that it operates as intended. Testing was also conducted to verify puncture depth of the needle.
#### 12. Clinical Testing Summary
No Clinical testing was conducted as part of this submission.
### 13. Electrical Safety, Risk Management, Sterility and Software
The SkinPen® Precision Elite System was developed and tested according to the following standards:
IEC 60601-1 Edition 3.2 2020-08 IEC 60601-1-2 Edition 4.1 2020-09 ISO 10993-1 Fifth edition 2018-08 IEC 62304 Edition 1.1 2015-06 ISO 11135 Second edition 2014-07-15 IEC 60601-1-6 Edition 3.2 2020-07
## 14. Biocompatibility
All patient contacting components of the SkinPen® Precision Elite System have been tested according to ISO 10993-1:2018 and are considered biocompatible.
#### 15. Statement of Substantial Equivalence
The SkinPen® Precision Elite System, the subject device for this submission, is very similar to the predicate device, the SkinPen® Precision System. Changes to the user interface, cartridge connection points, and the addition of an NFC chip for cartridge verification do not impact the device effectiveness, and safety is increased with the use of the NFC chip. There are no changes to the general technology, mechanism of action, or indications for use. Changes to the device are supported through software verification and validation, IEC, biocompatibility, and usability testing, Therefore, the two devices can be considered substantially equivalent.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.