← Product Code [QEH](/productcode/QEH) · K241385

# Omnifix Syringe NRFit (K241385)

_B.Braun Medical, Inc. · QEH · Jul 13, 2024 · General Hospital · SESE_

**Canonical URL:** https://fda.innolitics.com/device/K241385

## Device Facts

- **Applicant:** B.Braun Medical, Inc.
- **Product Code:** [QEH](/productcode/QEH.md)
- **Decision Date:** Jul 13, 2024
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 880.5860
- **Device Class:** Class 2
- **Review Panel:** General Hospital
- **Attributes:** Therapeutic

## Indications for Use

The B. Braun Omnifix Syringe NRFit is intended to inject fluid into, or withdraw fluids from, the body.

## Device Story

Sterile, single-use piston syringe; 1mL, 3mL, 10mL, 20mL capacities; lock and slip configurations. Features ISO 80369-6 compliant NRFit small-bore connector to prevent misconnections with non-neuraxial devices. Plunger rod colored yellow for visual identification of neuraxial compatibility. Operated manually by healthcare professionals in clinical settings; plunger rod depressed to inject or pulled to aspirate fluids. Used for neuraxial applications involving potential indirect contact with cerebral spinal fluid. Design modification from standard Luer (ISO 594) to NRFit (ISO 80369-6) limits device to neuraxial use. Benefits include reduced risk of medication errors via misconnection.

## Clinical Evidence

Bench testing only. No clinical data. Performance validated via ISO 80369-6 functional tests (dimensional, leakage, pressure decay, stress cracking, separation force, unscrewing torque, overriding resistance), biocompatibility (cytotoxicity, sensitization, systemic toxicity, genotoxicity, pyrogenicity, hemolysis), and neurotoxicity assessment. Packaging integrity verified via ASTM F2096 and ASTM F88/F88M.

## Technological Characteristics

Materials: Polypropylene (barrel), Polyisoprene (stopper), Silicone (lubricant/stopper). ISO 80369-6 compliant NRFit small-bore connector. Manual operation. Sterilization: Ethylene Oxide (SAL 10^-6). Shelf life: 5 years. Complies with ISO 7886-1 (hypodermic syringe function) and ISO 10993-1 (biocompatibility).

## Regulatory Identification

A piston syringe is a device intended for medical purposes that consists of a calibrated hollow barrel and a movable plunger. At one end of the barrel there is a male connector (nozzle) for fitting the female connector (hub) of a hypodermic single lumen needle. The device is used to inject fluids into, or withdraw fluids from, the body.

## Predicate Devices

- B. Braun Piston Syringes ([K071459](/device/K071459.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food & Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.

July 12, 2024

B. Braun Medical Inc. Kimberly Smith Technical Manager I - Regulatory Affairs 824 Twelfth Avenue Bethlehem, Pennsylvania 18018

## Re: K241385

Trade/Device Name: Omnifix Syringe NRFit Regulation Number: 21 CFR 880.5860 Regulation Name: Piston Syringe Regulatory Class: Class II Product Code: QEH Dated: May 15, 2024 Received: May 15, 2024

## Dear Kimberly Smith:

We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).

{1}------------------------------------------------

Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review. the OS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

# ForGang Peng)-S

Shruti Mistry Assistant Director Division of Drug Delivery and General Hospital Devices Office of Gastrorenal, ObGyn, General Hospital and Urology Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

{2}------------------------------------------------

## Indications for Use

510(k) Number (if known) K241385

Device Name Omnifix Syringe NRFit

Indications for Use (Describe)

The B. Braun Omnifix Syringe NRFit is intended for neuraxial use for aspiration/injection of fluids used by healtheare professionals in healthcare facilities.

| Type of Use (Select one or both, as applicable) |
|-------------------------------------------------|
|-------------------------------------------------|

X Prescription Use (Part 21 CFR 801 Subpart D)

| Over-The-Counter Use (21 CFR 801 Subpart C)

#### CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

# 510(k) Summary

## Submitter Information

| Name:           | B. Braun Medical Inc.                         |
|-----------------|-----------------------------------------------|
| Address:        | 824 Twelfth Avenue<br>Bethlehem, PA 18018 USA |
| Contact Person: | Kimberly Smith                                |
| Title:          | Technical Manager I – Regulatory Affairs      |
| Tel:            | (610) 596-2326                                |
| Email:          | kim.smith@bbraunusa.com                       |
| Date Prepared:  | July 12, 2024                                 |

## Device Name and Classification

| Device Trade Name:   | Omnifix Syringe NRFit                                                                                            |
|----------------------|------------------------------------------------------------------------------------------------------------------|
| Common Name:         | Piston Syringe                                                                                                   |
| Classification Name: | Piston Syringe With Neuraxial Connector – Epidural, Peripheral,<br>And/Or Indirect Cerebral Spinal Fluid Contact |
| Regulation Number:   | 21 CFR §880.5860                                                                                                 |
| Regulation Name:     | Piston Syringe                                                                                                   |
| Device Class:        | 2                                                                                                                |
| Product Code:        | QEH                                                                                                              |

## Predicate Device

| 510(k) Number:       | K071459                  |
|----------------------|--------------------------|
| Device Trade Name:   | B. Braun Piston Syringes |
| Common Name:         | Piston Syringe           |
| Classification Name: | Piston Syringe           |
| Regulation Number:   | 21 CFR §880.5860         |
| Regulation Name:     | Piston Syringe           |
| Device Class:        | 2                        |
| Product Code:        | FMF                      |

{4}------------------------------------------------

## Reason for Submission

The purpose of this submission is for the modification of the current B. Braun Medical Inc. Single Use, Piston Syringe (previously cleared under K071459) from the current Luer connections per ISO 594-1 and ISO 594-2 to the new ISO 80369-6 connectors referred to as NRFit. The new syringes with NRFit connectors will be referred to as Omnifix Syringe NRFit and will be supplied in luer slip and luer lock configurations. B. Braun Medical Inc. will continue to market the current B. Braun Medical Inc. Single Use, Piston Syringes with Luer connections for general purpose use, while the new B. Braun Medical Inc. Omnifix Syringe NRFit will be indicated for neuraxial use per ISO 80369-6.

## Device Description

B. Braun Medical Inc. Omnifix Syringe NRFit configurations are sterile, single use syringes with ISO 80369-6 NRFit compliant fittings. They are available in 1mL, 3mL, 10mL, and 20mL lock and slip configurations. The NRFit tips allow for connections of neuraxial specific applications while reducing the likelihood of misconnections to non-neuraxial devices. The syringe assembly consists of a barrel with a printed graduated scale in milliliters (mL), a stopper and a plunger rod. The plunger rod is yellow to designate a device intended to only connect to ISO 80369-6 compatible devices such as spinal or epidural needles. The plunger rod is pulled back to aspirate fluids or depressed to inject or expel fluids.

### Indications for Use

The B. Braun Omnifix Syringe NRFit is intended for neuraxial use for aspiration/injection of fluids used by healthcare professionals in healthcare facilities.

### Intended Use

The B. Braun Omnifix Syringe NRFit is intended to inject fluid into, or withdraw fluids from, the body.

### Technological Characteristics

The B. Braun Medical Inc. Omnifix Syringe NRFit configurations are equivalent to that of the predicate B. Braun Medical Inc. Omnifix Syringe in intended use, materials and performance characteristics.

The difference between the proposed and predicate B. Braun Medical Inc. Omnifix Syringe is the modification of the tip of the current B. Braun Medical Inc. Single Use, Piston Syringe from the current Luer connections per ISO 594-1 and ISO 594-2 to the new ISO 80369-6 connectors referred to as NRFit. This modification therefore revises the Indications for Use for a neuraxial use application but does not modify the intended use. These differences do not impact the statement of substantial equivalence.

{5}------------------------------------------------

|                                                                                | Proposed Omnifix Syringe NRFit                                                                                                                                                                                                                                                            | Predicate Device (K071459)<br>B. Braun Piston Syringes                                                                                                               | Differences                                                                                                                                                                                                                           |                                                                                                                                                                                 |
|--------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Indications for Use                                                            | The B. Braun Omnifix Syringe<br>NRFit is intended for neuraxial use<br>for aspiration/injection of fluids used<br>by healthcare professionals in<br>healthcare facilities.                                                                                                                | The Omnifix piston syringes<br>are intended to be used to<br>inject fluid into, or withdraw<br>fluids from, the body.                                                | The proposed device is<br>limited to neuraxial use.                                                                                                                                                                                   |                                                                                                                                                                                 |
| Product Code                                                                   | QEH                                                                                                                                                                                                                                                                                       | FMF                                                                                                                                                                  | The proposed device is<br>limited to neuraxial use.<br>Predicate device was for<br>general use.                                                                                                                                       |                                                                                                                                                                                 |
| Intended Use                                                                   | The B. Braun Omnifix Syringe<br>NRFit is intended to be used to<br>inject fluid into, or withdraw fluids<br>from, the body.                                                                                                                                                               | The Omnifix piston syringes<br>are intended to be used to<br>inject fluid into, or withdraw<br>fluids from, the body.                                                | None, Same                                                                                                                                                                                                                            |                                                                                                                                                                                 |
| Intended User                                                                  | Healthcare Professionals                                                                                                                                                                                                                                                                  | Healthcare Professionals                                                                                                                                             | None, Same                                                                                                                                                                                                                            |                                                                                                                                                                                 |
| Environment of Use                                                             | Healthcare Facilities                                                                                                                                                                                                                                                                     | Healthcare Facilities                                                                                                                                                | None, Same                                                                                                                                                                                                                            |                                                                                                                                                                                 |
| Syringe<br>components/materials                                                | Barrel                                                                                                                                                                                                                                                                                    | Polypropylene                                                                                                                                                        | Polypropylene                                                                                                                                                                                                                         | None, Same                                                                                                                                                                      |
|                                                                                | Plunger                                                                                                                                                                                                                                                                                   | Polyisoprene                                                                                                                                                         | Polyisoprene                                                                                                                                                                                                                          | None, Same                                                                                                                                                                      |
|                                                                                | Stopper                                                                                                                                                                                                                                                                                   | Silicone                                                                                                                                                             | Silicone                                                                                                                                                                                                                              | None, Same                                                                                                                                                                      |
|                                                                                | Lubricant                                                                                                                                                                                                                                                                                 | Silicone                                                                                                                                                             | Silicone                                                                                                                                                                                                                              | None, Same                                                                                                                                                                      |
|                                                                                | Plunger Rod                                                                                                                                                                                                                                                                               | Yellow<br>1mL Polystyrene /<br>3, 5, 10, 20 mL<br>Polypropylene                                                                                                      | Clear<br>1mL Polystyrene /<br>3, 5, 10, 20 mL Polypropylene                                                                                                                                                                           | Base material is the same.<br>Color is different - change<br>to the plunger from clear to<br>yellow to differentiate the<br>NRFit syringes from the<br>general piston syringes. |
| Tip Type<br>smallbore connector                                                | NRFit lock or slip per ISO 80369-<br>6:2016                                                                                                                                                                                                                                               | Lock or Slip per ISO 594-<br>1:1986 and ISO 594-2:1998                                                                                                               | The proposed device<br>contains a NRFit smallbore<br>connector which is                                                                                                                                                               |                                                                                                                                                                                 |
| Functional Testing<br>Summary of ISO<br>80369-6 compliant<br>nonclinical tests | • Dimensional Requirements<br>• Fluid leakage requirement<br>• Leakage by pressure decay<br>• Subatmospheric pressure air<br>leakage<br>• Stress cracking<br>• Resistance to separation from<br>axial load<br>• Resistance to separation from<br>unscrewing<br>• Resistance to overriding | • Gauging<br>• Liquid leakage<br>• Air leakage<br>• Separation force<br>• Unscrewing torque<br>• Ease of assembly<br>• Resistance to overriding<br>• Stress cracking | intended for neuraxial use.<br>ISO 80369-6 testing<br>performed to demonstrate<br>safety and effectiveness of<br>the neuraxial smallbore<br>connector for the neuraxial<br>intended use and to<br>confirm substantial<br>equivalence. |                                                                                                                                                                                 |
| Patient Contact<br>category/ Duration                                          | Externally Communicating,<br>Tissue/Bone/Dentin, prolonged<br>exposure                                                                                                                                                                                                                    | Externally Communicating,<br>Blood Path Indirect, limited<br>exposure                                                                                                | The proposed device is<br>intended for neuraxial use<br>which may have indirect<br>contact with cerebral<br>spinal fluid (CSF). For this<br>reason,<br>Tissue/Bone/Dentin with a<br>longer duration exposure<br>was chosen.           |                                                                                                                                                                                 |
| Biocompatibility                                                               | Conforms to ISO 10993-1                                                                                                                                                                                                                                                                   | Conforms to ISO 10993-1                                                                                                                                              | None, Same                                                                                                                                                                                                                            |                                                                                                                                                                                 |
| Hypodermic Syringe<br>Functional                                               | Conforms to ISO 7886-1                                                                                                                                                                                                                                                                    | Conforms to ISO 7886-1                                                                                                                                               | None, Same                                                                                                                                                                                                                            |                                                                                                                                                                                 |
|                                                                                | Proposed Omnifix Syringe NRFit                                                                                                                                                                                                                                                            | Predicate Device (K071459)<br>B. Braun Piston Syringes                                                                                                               | Differences                                                                                                                                                                                                                           |                                                                                                                                                                                 |
| Neurotoxicity<br>Assessment                                                    | No signs of systemic toxicity or<br>neurological impairment from<br>exposure of leachable compounds<br>from the test article.                                                                                                                                                             | N/A. Not required previously<br>for predicate device.                                                                                                                | The proposed device is<br>intended for neuraxial use<br>which may have indirect<br>contact with CSF. For this<br>reason, a neurotoxicity<br>evaluation was<br>performed.                                                              |                                                                                                                                                                                 |
| Pyrogenicity                                                                   | Spinal limit of 2.15 EU/device                                                                                                                                                                                                                                                            | Limit of 20 EU/device                                                                                                                                                | The proposed device is<br>intended for neuraxial use<br>which may have indirect<br>contact with CSF. The<br>proposed devices are<br>therefore tested for<br>endotoxin at the spinal<br>limit of 2.15 EU/device.                       |                                                                                                                                                                                 |
| Packaging                                                                      | Individual Unit Package                                                                                                                                                                                                                                                                   | Individual Unit Package                                                                                                                                              | None, Same                                                                                                                                                                                                                            |                                                                                                                                                                                 |
| Sterilization Method                                                           | Ethylene Oxide                                                                                                                                                                                                                                                                            | Ethylene Oxide                                                                                                                                                       | None, Same                                                                                                                                                                                                                            |                                                                                                                                                                                 |
| SAL                                                                            | 10-6                                                                                                                                                                                                                                                                                      | 10-6                                                                                                                                                                 | None, Same                                                                                                                                                                                                                            |                                                                                                                                                                                 |
| Shelf Life                                                                     | 5 Years                                                                                                                                                                                                                                                                                   | 5 Years                                                                                                                                                              | None, Same                                                                                                                                                                                                                            |                                                                                                                                                                                 |

{6}------------------------------------------------

The intended use of the proposed and predicate devices are the same in that both are intended for use by healthcare professionals for aspiration/injection of fluids. However, due to the ISO 80369-6 connector, the Omnifix Syringe NRFit configurations are limited to neuraxial applications. As such, the neuraxial indication, is a subset of the predicate indications for use, which is for general use. Therefore, the above difference in indications for use is not critical to the intended use as the intended use remains for aspiration of fluids. It further does not affect the safety and effectiveness of the device when used as labeled as demonstrated through functional testing. ISO 10993-1 biocompatibility testing, a neurotoxicological assessment, and ISO 80369-6 functional performance testing were conducted to address the changes in indications for neuraxial use and tip design.

## Nonclinical Performance Tests

Bench testing performed on the proposed devices supports substantial equivalence. No clinical testing was performed as this device does not require clinical studies to demonstrate substantial equivalence with the predicate device. The following ISO 80369-6 testing has been completed for the proposed device.

- · Dimensional Requirements
- · Fluid leakage requirement
- · Leakage by pressure decay
- · Subatmospheric pressure air leakage
- · Stress cracking
- · Resistance to separation from axial load
- · Resistance to separation from unscrewing
- · Resistance to overriding

{7}------------------------------------------------

A biocompatibility evaluation was conducted on the subject device per ISO 10993-1:2018. The contact classification is external communicating, indirect contact with blood and tissue with a prolonged use. Based on the evaluation, the following biological tests were conducted:

- · Cytotoxicity
- · Intracutaneous reactivity
- · Sensitization
- · Acute systemic toxicity
- · Subacute/subchronic systemic toxicity
- · Genotoxicity
- · Material-mediated pyrogenicity
- Hemolysis

Additionally, the following tests were performed:

- · Particulate Matter per USP <788>
- · Neurotoxicity assessment

Packaging: Per ASTM F2096:2011 · Bubble Leak Testing Per ASTM F88/F88M:2015 · Seal Strength Testing

## Summary of Substantial Equivalence

The B. Braun Medical Inc. Omnifix Syringe NRFit configuration is substantially equivalent to the predicate device in intended use, principles of operation, technology, design, materials and performance and are as safe and effective as the predicate devices.

---

**Source:** [https://fda.innolitics.com/device/K241385](https://fda.innolitics.com/device/K241385)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
