Advisor™ HD Grid X Mapping Catheter, Sensor Enabled™
Applicant
ABBOTT MEDICAL
Product Code
MTD · Cardiovascular
Decision Date
Jul 9, 2024
Decision
SESE
Submission Type
Special
Regulation
21 CFR 870.1220
Device Class
Class 2
Indications for Use
The Advisor™ HD Grid X Mapping Catheter, Sensor Enabled™, is indicated for multiple electrode electrophysiological mapping of cardiac structures in the heart. i.e., recording or stimulation only. This catheter is intended to obtain electrograms in the atrial and ventricular regions of the heart.
Device Story
Sterile, single-use, irrigated, high-density mapping catheter; 7.5F shaft, 8F distal deflectable section; 110 cm length. Features paddle-shaped distal tip with 16 electrodes, two paddle magnetic sensors, two distal shaft ring electrodes, and two shaft magnetic sensors. Operated by physicians in clinical settings; used with EnSite™ X EP System, connecting cables, and irrigation pumps. Insertion tool facilitates passage through introducer sheath hemostasis valves. Device records cardiac electrograms or provides stimulation; data processed by EnSite™ X system to visualize cardiac structures. Enables high-density mapping to assist in diagnosis/treatment of cardiac arrhythmias.
Clinical Evidence
Bench testing only. Verification activities included dimensional, visual, mechanical integrity, simulated use, electrical, packaging, and shelf-life testing. Biocompatibility testing performed per ISO 10993 for external communicating devices with limited (≤24 hour) circulating blood contact; all endpoints met.
Technological Characteristics
7.5F/8F catheter; polymer braided shaft; 16-electrode paddle tip; magnetic sensors in shaft and paddle; irrigated design. Biocompatibility per ISO 10993. Connectivity: EnSite™ X EP System. Sterilization: Not specified.
Indications for Use
Indicated for patients requiring electrophysiological mapping of cardiac structures (atrial and ventricular regions) for recording or stimulation purposes.
Regulatory Classification
Identification
An electrode recording catheter or an electrode recording probe is a device used to detect an intracardiac electrocardiogram, or to detect cardiac output or left-to-right heart shunts. The device may be unipolar or multipolar for electrocardiogram detection, or may be a platinum-tipped catheter which senses the presence of a special indicator for cardiac output or left-to-right heart shunt determinations.
Predicate Devices
Advisor™ HD Grid Mapping Catheter, Sensor Enabled™ (K172393)
Reference Devices
Advisor™ HD Grid Mapping Catheter, Sensor Enabled™ (K202066)
Submission Summary (Full Text)
{1}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food & Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
July 9, 2024
Abbott Medical Samantha Stretar Senior Regulatory Affairs Specialist 5050 Nathan Lane North Plymouth, Minnesota 55442
Re: K241372
Trade/Device Name: Advisor™ HD Grid X Mapping Catheter, Sensor Enabled™ Regulation Number: 21 CFR 870.1220 Regulation Name: Electrode Recording Catheter Or Electrode Recording Probe Regulatory Class: Class II Product Code: MTD Dated: July 2, 2024 Received: July 2, 2024
### Dear Samantha Stretar:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
{2}------------------------------------------------
Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review. the OS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
## Marco Cannella -S for
Aneesh Deoras Assistant Director Division of Cardiac Electrophysiology, Diagnostics and Monitoring Devices Office of Cardiovascular Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{3}------------------------------------------------
## Indications for Use
Form Approved: OMB No. 0910-0120 Expiration Date: 07/31/2026 See PRA Statement below.
Submission Number (if known)
K241372
Device Name
Advisor™ HD Grid X Mapping Catheter, Sensor Enabled™
Indications for Use (Describe)
The Advisor™ HD Grid X Mapping Catheter, Sensor Enabled™, is indicated for multiple electrode electrophysiological mapping of cardiac structures in the heart. i.e., recording or stimulation only. This catheter is intended to obtain electrograms in the atrial and ventricular regions of the heart.
Type of Use (Select one or both, as applicable)
Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{4}------------------------------------------------
## K241372
# 510(K) SUMMARY
This 510(k) summary is prepared in accordance with the requirements of 21 CFR §807.92.
| 510(k) Information | |
|---------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| 510(k) Number | K241372 |
| 510(k) Type | Special 510(k) |
| Date Prepared | 14-May-2024 |
| Submitter Information | |
| Manufacturer Name and<br>Address | Abbott Medical<br>5050 Nathan Lane North<br>Plymouth, MN 55442<br>USA |
| Phone Number | 651-756-5833 |
| Contact Person | Samantha Stretar<br>Senior Regulatory Affairs Specialist |
| Device Information | |
| Trade Name | AdvisorTM HD Grid X Mapping Catheter, Sensor EnabledTM |
| Common Name | Catheter, Intracardiac Mapping, High-Density Array |
| Class | II |
| Classification Name | 21 CFR 870.1220 Electrode recording catheter or electrode recording probe |
| Product Code | MTD - Catheter, Intracardiac Mapping, High-Density Array |
| Predicate Device | AdvisorTM HD Grid Mapping Catheter, Sensor EnabledTM (K172393, cleared 23-<br>Apr-2018) |
| Reference Device | AdvisorTM HD Grid Mapping Catheter, Sensor EnabledTM (K202066, cleared 25-<br>Nov-2020) |
| Device Description | The AdvisorTM HD Grid X Mapping Catheter, Sensor EnabledTM, is a sterile,<br>single use, irrigated, high-density mapping catheter with a 7.5F shaft and an 8F<br>distal shaft deflectable section. It is available in a D-F bi-directional curve model<br>that is deflected using the actuator located on the catheter handle. The catheter<br>working length is 110 cm. The device consists of a paddle-shaped distal tip with<br>16 electrodes, two paddle magnetic sensors, two distal shaft ring electrodes,<br>two shaft magnetic sensors, polymer braided shaft, handle, fluid lumen<br>extension with a luer, and an electrical connector. The catheter also has an<br>insertion tool intended to compress and guide the distal paddle into, and<br>withdraw from, the hemostasis valve of an introducer sheath.<br>The catheter is intended to be used with the EnSiteTM X EP System and other<br>accessories, including the connecting cable and commercially available<br>irrigation pumps. |
| 510(k) Information | |
| Indications for Use | The Advisor™ HD Grid X Mapping Catheter, Sensor Enabled™, is indicated for<br>multiple electrode electrophysiological mapping of cardiac structures in the<br>heart, i.e., recording or stimulation only. This catheter is intended to obtain<br>electrograms in the atrial and ventricular regions of the heart. |
| Predicate Comparison | |
| Comparison of<br>Technological<br>Characteristics | The subject device has the same intended use, indications for use, and<br>principles of operation as the predicate device. The subject device has<br>substantially equivalent technological characteristics to the predicate device.<br>Like the predicate, the subject device incorporates two magnetic sensors in the catheter shaft. The subject device also includes two magnetic sensors in the outer splines of the distal tip paddle assembly. Additional catheter components were modified from the predicate to accommodate the addition of the magnetic sensors within the paddle, improve manufacturability, and improve material stability for continuous improvement purposes. The predicate and subject device have the same working length, shaft French size, and electrode spacing. |
| Non-Clinical Testing<br>Summary | Bench design verification activities were performed and met their respective<br>acceptance criteria to ensure that the device in scope of this submission is<br>substantially equivalent to the predicate. Testing included dimensional, visual,<br>mechanical integrity, simulated use, electrical, packaging, and shelf-life testing<br>to assess substantial equivalence. Additionally, biocompatibility testing was conducted per the voluntary consensus standards under ISO 10993 and met all the biocompatibility endpoints for a medical device category for an external communicating medical device with limited (≤24 hour) circulating blood contact. |
| Statement of Equivalence | The subject device has the same intended use, indications for use, and<br>principles of operation as the predicate device. The subject device has similar<br>technological characteristics to the predicate device and the completed bench<br>design verification testing and biocompatibility testing provides objective<br>evidence that the subject device is at least as safe and effective as the<br>predicate, and therefore, is substantially equivalent to the predicate. |
{5}------------------------------------------------
## K241372
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.