K241331 · Springbok, Inc. · LNH · Oct 1, 2024 · Radiology
Device Facts
Record ID
K241331
Device Name
MuscleView
Applicant
Springbok, Inc.
Product Code
LNH · Radiology
Decision Date
Oct 1, 2024
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 892.1000
Device Class
Class 2
Attributes
AI/ML, Software as a Medical Device, Pediatric
AI Performance
Output
Algorithm
Acceptance
Observed
Dev DS
Dev Readers
Test DS
Test Readers
Muscle and bone structure segmentation
Convolutional networks (CNNs)
Dice similarity coefficient or volume difference below interobserver variability
Mean better than or equal to the acceptance criteria threshold
Training data: 1658 unique scans from 1294 unique subjects
—
Validation data: 148 unique scans from 148 unique subjects
>1 (expert readers)
Muscle volume, fat infiltration, and asymmetry metrics
Convolutional networks (CNNs)
Dice similarity coefficient or volume difference below interobserver variability
Mean better than or equal to the acceptance criteria threshold
Training data: 1658 unique scans from 1294 unique subjects
—
Validation data: 148 unique scans from 148 unique subjects
>1 (expert readers)
Indications for Use
MuscleView is used in adults and pediatric patients aged 18 and older to automatically segment muscle and bone structures of the lower extremities from magnetic resonance imaging using a machine learning-based approach. After segmentation, it can provide derived metrics including muscle volume, bone volume, intramuscular fat percentage, and left/right asymmetry. It is intended to be used by physicians who are trained to interpret MRI images, and serves as an initial method to segment muscle and bone structures from one or more study series. The segmentation results need to be reviewed and edited using appropriate software. It is intended to only provide the segmentation and derived metrics for muscle and bone structures and cannot serve as direct guidance for diagnosis of any diseases. This device is not intended for use with patients who have tumors in lower limb.
Device Story
MuscleView is a software-only device for automated segmentation of 80 musculoskeletal structures (muscles and bones) in the lower extremities from MRI. It uses a machine learning-based approach, specifically convolutional neural networks (CNNs) trained on expert-contoured cases. Input consists of DICOM-compliant MRI data; the system preprocesses images to create 3D volumes, performs segmentation, and calculates metrics including volume, intramuscular fat percentage, and limb asymmetry. The device is used in a clinical setting by physicians trained in MRI interpretation. It operates as a service on a workstation with a GPU, accessed via a web-based interface. The output is a 3D visualization with quantitative metrics, intended as an initial segmentation method requiring physician review and editing. It does not provide diagnostic guidance. Benefits include automated quantification of musculoskeletal structures to support clinical assessment.
Clinical Evidence
Bench testing only. Validation performed on 148 scans from 148 subjects, independent of the 1,658-scan training set. Performance evaluated against manual expert segmentation using Dice Similarity Coefficient (DSC) and Volume Difference (VDt). Subgroup analyses included healthy vs. patient populations (amputees, muscular dystrophy), MRI manufacturers (GE, Siemens, Philips, Canon, Toshiba), and age/sex demographics. All ROIs met predetermined acceptance criteria based on interobserver repeatability.
Technological Characteristics
Software-only device; runs on PC-compatible workstation with GPU. Processes DICOM-compliant MRI data. Uses CNN-based machine learning for automated segmentation. Supports 80 musculoskeletal structures (72 muscles, 8 bones) in lower extremities. Web-based interface for data management, segmentation, and 3D visualization. No manual editing supported within the device.
Indications for Use
Indicated for adults and pediatric patients aged 18 and older for automatic segmentation of lower extremity muscle and bone structures from MRI. Contraindicated for patients with lower limb tumors.
Regulatory Classification
Identification
A magnetic resonance diagnostic device is intended for general diagnostic use to present images which reflect the spatial distribution and/or magnetic resonance spectra which reflect frequency and distribution of nuclei exhibiting nuclear magnetic resonance. Other physical parameters derived from the images and/or spectra may also be produced. The device includes hydrogen-1 (proton) imaging, sodium-23 imaging, hydrogen-1 spectroscopy, phosphorus-31 spectroscopy, and chemical shift imaging (preserving simultaneous frequency and spatial information).
Special Controls
*Classification.* Class II (special controls). A magnetic resonance imaging disposable kit intended for use with a magnetic resonance diagnostic device only is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 892.9.
{1}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: a symbol on the left and the FDA name on the right. The symbol on the left is a stylized image of a human figure, while the FDA name on the right is written in blue letters. The words "U.S. FOOD & DRUG ADMINISTRATION" are written in a clear, sans-serif font.
October 1, 2024
Springbok, Inc. Scott Magargee Chief Executive Officer 110 Old Preston Ave Charlottesville, Virginia 22902
Re: K241331
Trade/Device Name: MuscleView Regulation Number: 21 CFR 892.1000 Regulation Name: Magnetic Resonance Diagnostic Device Regulatory Class: Class II Product Code: LNH Dated: August 19, 2024 Received: August 19, 2024
Dear Scott Magargee:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
{2}------------------------------------------------
Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review. the OS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-device-advicecomprehensive-regulatory-assistance/unique-device-identification-system-udi-system.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
{3}------------------------------------------------
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
D. G. K.
Daniel M. Krainak, Ph.D. Assistant Director DHT8C: Division of Radiological Imaging and Radiation Therapy Devices OHT8: Office of Radiological Health Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{4}------------------------------------------------
# Indications for Use
510(k) Number (if known) K241331
Device Name MuscleView
#### Indications for Use (Describe)
MuscleView is used in adults and pediatics aged 18 and older to automatically segment muscle and bone structures of the lower extremities from magnetic resonance imaging using a machine learning-based approach. After segmentation, it can provide derived metrics including muscle volume, intramuscular fat percentage, and left/right asymmetry.
It is intended to be used by physicians who are trained to interpret MRI images, and serves as an initial method to segment muscle and bone structures from one or more study series. The segmentation results need to be reviewed and edited using appropriate software.
It is intended to only provide the segmentation and derived metrics for muscle and bone structures and cannot serve as direct guidance for dagnosis of any diseases. This device is not intents who have tumors in lower limb.
| Type of Use (Select one or both, as applicable) |
|-------------------------------------------------|
|-------------------------------------------------|
X Prescription Use (Part 21 CFR 801 Subpart D)
| Over-The-Counter Use (21 CFR 801 Subpart C)
### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{5}------------------------------------------------
#### Device Name: MuscleView
#### Date Summary was Prepared: September 30th, 2024
# 1. Applicant:
Springbok, Inc.
110 Old Preston Ave
Charlottesville, VA 22902 USA
Contact Name: Scott Magargee – Chief Executive Officer
Phone: 1-215-680-9078
Fax: N/A
E-mail: scott.magargee@springbokanalytics.com
#### 2. Device:
Trade Name: MuscleView Common Name: MuscleView Model Number: v1.0 Product Code: LNH Regulation Description: Magnetic Resonance Diagnostic Device Regulation Number: 21 CFR 892.1000 Device Class: II
### 3. Predicate Devices:
Trade Name: AMRA Profiler
Manufacturer: AMRA Medical AB
{6}------------------------------------------------
Address: 68 Southwood Ter SOUTHBURY, CT 06488 Regulation Number: 21 CFR 892.1000 Regulation Name: Magnetic resonance diagnostic device Device Class: Class II Product Code: LNH 510(k) Number: K173749 510(k) Clearance Date: Dec 06, 2018
### 4. Device Description
MuscleView is a software only product that uses a machine learning-based approach for the automatic segmentation of musculoskeletal structures from MRI. Based on the segmentation, metrics such as volume and length of the segmented structures are calculated.
The software has the following modules: user management, data management, image processing, Al segmentation & 3D model viewer and metrics calculation. User management involves authentication and access to the software and its results. Data management involves medical image data and its interactions with the system workflow. Image processing involves Preprocessing the DICOM data to create a combined continuous 3D volume(s) of series with similar settings for use in Al segmentation & 3D model viewer module handles training data and algorithms to obtain the pre-trained models and algorithms to update models. Metric calculation module handles the final calculation of relevant metrics.
Input data is preprocessed and prepared for 3D volume segmentation of the musculoskeletal structures. A library of already contoured expert cases is utilized to train the machine learning algorithms, specifically convolutional networks (CNNs) perform automated segmentation. This process is in an auxiliary module for AI training.
MuscleView is intended to be used by physicians who are trained to interpret MRI images, and serves as an initial method to segment muscle and bone structures from one or more study series. The segmentation results need to be reviewed and edited using appropriate software. This device is not intended for use with patients who have tumors in lower limb. The currently supported anatomical regions for automatic segmentation are 80 different muscles and bones of the lower extremity. The supported musculoskeletal structures for each region are shown below in Table 1.
{7}------------------------------------------------
Upon segmentation, a suite of metrics regarding the segmented 3D volumes is provided. It is intended to only provide the segmentation and derived metrics for muscle and bone structures and cannot serve as direct guidance for diagnosis of any diseases. These metrics include segmentation volume, fat infiltration (if applicable), and limb side asymmetry. The metrics are provided in conjunction with an interactive visualization of the 3D segmentation results.
The software is deployed within a private network on a workstation with an advanced graphic processing unit (GPU) and runs as a service. A web-based interface is used to access the service and manage the data transfer, automatic segmentation, and visualization.
Table 1: List of supported musculoskeletal structures for segmentation. Each ROI is supported for both left and right sides, which are analyzed and validated independently. *Gemelli is a grouping of the superior and inferior gemellus. **Fibulari is a grouping of the fibularis brevis and fibularis longus. ***Phalangeal extensors are a grouping of the extensor digitorum longus, extensor hallucis longus, and fibularis tertius.
| Structure Name | Structure Type |
|--------------------------------|----------------|
| Adductor brevis | Muscle |
| Adductor longus | Muscle |
| Adductor magnus | Muscle |
| Biceps femoris (long head) | Muscle |
| Biceps femoris (short head) | Muscle |
| Fibulari** | Muscle |
| Flexor Digitorum Longus | Muscle |
| Flexor Hallucis Longus | Muscle |
| Gastrocnemius (lateral head) | Muscle |
| Gastrocnemius (medial<br>head) | Muscle |
| Gemelli* | Muscle |
| Gluteus maximus | Muscle |
| Gluteus medius | Muscle |
| Gluteus minimus | Muscle |
| Gracilis | Muscle |
| Iliacus | Muscle |
| Obturator externus | Muscle |
| Obturator internus | Muscle |
| Pectineus | Muscle |
| Phalangeal extensors*** | Muscle |
| Piriformis | Muscle |
| Popliteus | Muscle |
| Psoas major | Muscle |
| Quadratus femoris | Muscle |
| Quadratus lumborum | Muscle |
| Rectus femoris | Muscle |
| Sartorius | Muscle |
| Semimembranosus | Muscle |
| Semitendinosus | Muscle |
| Soleus | Muscle |
| Tensor fasciae latae | Muscle |
| Tibialis anterior | Muscle |
| Tibialis Posterior | Muscle |
| Vastus intermedius | Muscle |
| Vastus lateralis | Muscle |
| Vastus medialis | Muscle |
| Pelvis | Bone |
| Femur | Bone |
| Tibia | Bone |
| Fibula | Bone |
{8}------------------------------------------------
# 5. Indications for Use Statement
MuscleView is used in adults and pediatric patients aged 18 and older to automatically segment muscle and bone structures of the lower extremities from magnetic resonance imaging using a
{9}------------------------------------------------
machine learning-based approach. After segmentation, it can provide derived metrics including muscle volume, bone volume, intramuscular fat percentage, and left/right asymmetry.
lt is intended to be used by physicians who are trained to interpret MRI images, and serves as an initial method to segment muscle and bone structures from one or more study series. The segmentation results need to be reviewed and edited using appropriate software.
It is intended to only provide the segmentation and derived metrics for muscle and bone structures and cannot serve as direct guidance for diagnosis of any diseases. This device is not intended for use with patients who have tumors in lower limb.
# 6. Summary of Technoloqical Characteristics Comparison
The similarities and differences between the technological characteristics of the two products are shown in Table 2. The key difference is the detailed implementation of the automated segmentation algorithms. Testing demonstrates that the differences do not raise new questions of safety or effectiveness.
| Topic | AMRA Profiler (510k Number:<br>K173749) | MuscleView |
|------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------|
| Physical<br>Characteristics | A service that is provided with a<br>cloud-based service using an<br>automated image-analysis pipeline<br>(with manual quality control for scan<br>preprocessing and label quality<br>control) | Software package that operates on a<br>virtual machine within off-the-shelf<br>hardware |
| Computer | Not applicable | PC Compatible |
| DICOM<br>Standard<br>Compliance | The service processes DICOM<br>compliant image data in accordance<br>with a required MRI protocol. | The software processes DICOM<br>compliant image data |
| Modalities | MRI | MRI |
| MRI<br>Parameters of<br>Importance | Collected Series: 3D Dixon water and<br>fat phase.<br>Region of Interest: Lower extremity,<br>Upper body as well<br>Supported Field Strength: | The same except that we support a<br>larger variability in scan properties<br>and coverage |
Table 2. Summary of technological characteristic comparison.
{10}------------------------------------------------
| | Direction of Capture: Axial<br>Supported In-Plane Resolution:<br>Supported Slice Spacing:<br>Note: Rigid conformity to MRI protocol is needed. | |
|---------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| User Interface | Concierge Service. Data is provided and analyzed, and results returned | The software is designed for use on a workstation with a web-based user interface. Functionalities are largely the same. |
| Segmentation Structures | 12 individual muscles and 6 muscle groups across the lower and upper extremity and liver | Eighty (8 bones, 72 muscles) individual structures on both the left and right side for the lower extremity focused region. |
| Segmentation Metrics | Muscle Volume, Fat Fraction | Structure volume, muscle fat infiltration, and derived metrics including asymmetry, muscle length, and cross-sectional area |
| Overall Segmentation Method | Labeled muscle segmentations automatically generated using non-rigid image registration to atlases utilizing 2D analysis techniques | AI segmentation model-based approach using a library of expert contours for training, MuscleView uses a machine learning-based method to train CNNs from expert contours to perform segmentation on the target images to generate contours |
| Support of Manual Editing by customer | No | No |
### 7. Performance Data
The safety and performance of MuscleView have been evaluated and verified in accordance with software specifications and applicable performance standards through software verification and validation testing. Non-clinical verification and validation test results, including model performance and software usability, established that the device meets its design requirements and intended use, that it is as safe and as effective as the AMRA Profiler (510k Number: K173749), and that no new issues of safety and effectiveness were raised.
{11}------------------------------------------------
Further, during development of the software, potential hazards were controlled by a risk management plan including risk analysis, risk mitigation, and validation.
### 8. Summary of Al Validation
The artificial intelligence (AI) responsible for muscle segmentation within MuscleView was trained and tested on MRI scans generated across varying patients, demographics (including – but not limited to - gender, age and ethnicity), scan sites, and MRI parameters (including manufacturer, magnetic field strength, series settings/types, matrix size, field of view, and scan resolution). See Table 3 below detailing variation in the data.
Table 3. Summary of Al training and validation dataset's demographics. *Demographic data were not available for some scans.
| Factor | Groups | Training data | Validation data |
|--------------------------------------------------|-----------------------------------------|---------------|-----------------|
| Number of unique scans | n/a | 1658 | 148 |
| Number of unique subjects | n/a | 1294 | 148 |
| Gender | Male (%) | 1192* | 102 |
| | Female (%) | 374* | 46 |
| Age* | Mean | 29 | 31.7 |
| | Standard Deviation | 13.4 | 15.5 |
| Ethnicity<br>(based on regional<br>demographics) | % Non-Hispanic White | 52 | 52 |
| | % Hispanic/Latino | 18 | 18 |
| | % Black/African American | 14 | 14 |
| | % Asian | 10 | 10 |
| | % Australian | 2 | 2 |
| | % American Indian /<br>Alaska Native | <1 | <1 |
| | % Native Hawaiian /<br>Pacific Islander | <1 | <1 |
| | % Australian Aboriginal | <1 | <1 |
| | % Other | 2 | 2 |
{12}------------------------------------------------
| | Siemens | 680 | 74 |
|------------------|-------------------------|-----|----|
| | GE Medical Systems | 435 | 28 |
| MRI Manufacturer | Philips Medical Systems | 82 | 29 |
| | Canon | 1 | 4 |
| | Other (Toshiba) or | 460 | |
| | Unknown | | |
Performance testing of segmentation and associated metrics was performed by comparing the results obtained to a reference standard developed by manual segmentation performed by experts. Validation Tests met the acceptance criteria if the dice similarity coefficient (DSC) or volume difference (VDt) was below interobserver variability.
Independence of test data from training data was ensured in several ways. (1) MRI Data used to train the Al was explicitly separate from validation data (both as MRI scans and as subjects), (2) the majority (70%) of datasets in the validation set came from imaging centers and organizations not used in the train datasets (19 new sites), and (3) the personnel involved in establishing the reference standard for the Al were not involved in the algorithm's development to ensure the independence of training and testing.
The Al segmentation was validated on the 80 musculoskeletal structures, demonstrating their accuracy across many varying scan and patient cases as related to the gold-standard of label generation – interobserver repeatability of manually vetted labels. The segmentation performance and derived metrics of all ROIs passed the predetermined acceptance criteria in both healthy and patient population subgroups, i.e. a one-sample T-test between the results of the validation set and the acceptance criteria threshold (the 95% confidence interval from the interobserver repeatability of manually vetted labels). A desired outcome to "approve" a ROI passed validation was a mean better than or equal to the acceptance criteria (a significant T-test showing the sample differed significantly better than the acceptance criteria threshold). Each musculoskeletal structure either used dice similarity coefficient (DSC) or volume difference (VDt) as its comparison metric, dependent on what best captured the structure's segmentation accuracy. The healthy and patient subgroup analysis split the validation dataset into two groups, healthy (athlete and healthy control) and patients (amputees, muscular dystrophy patient post-surgery). The subgroup analysis across different manufacturers (GE, Siemens, Phillips, Toshiba, Canon) also showed consistent performance, indicating high generalizability of our product. The age/biologic sex subgroup analysis (males 18 – 21 years old, females 18 – 21 years old, males >21 years old, and females >21 years old) also demonstrated that each subgroup passed the predetermined
{13}------------------------------------------------
acceptance criteria. For the 95% confidence interval for all structures and subgroups, see Table 4 (DSC) and Table 5 (VDt).
# Table 4: Dice similarity coefficient 95% confidence interval for all 80 musculoskeletal structures across all subgroup analyses.
| | Descriptive Statistics (DSC) 95% Confidence Interval | | | | | | | | | | |
|------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------|-------------|--------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------|---------|---------|---------|---------------------------------------|----------------------------------------------------------------------------------------------------------|--------------|--------------------------------------------------------------|
| ROI | Pathology Level<br>Subgroup Analysis | | | MRI Manufacturer<br>Subgroup Analysis | | | | Age/Biologic Sex<br>Subgroup Analysis | | | |
| | Healthy | Patient | Canon | GE | Philips | Siemens | Toshiba | Years | Years | years | Male > 21 Male 18-21 Female > 21 Female 18-21<br>years |
| adductor brevis | | | 0.962-0.965 0.909-0.932 0.978-0.984 0.947-0.967 0.937-0.954 0.956-0.96 0.956-0.96 0.95-0.959 | | | | | | | 0.958-0.963 | 0.911 -0.941 |
| adductor longus | | | 0.978-0.979 0.934-0.952 0.982-0.989 0.965-0.98 0.954-0.972 0.962-0.9710.971-0.975 0.97-0.982 | | | | | | 0.968 -0.975 | 0.974 -0.978 | 0.936 -0.96 |
| adductor magnus | | | 0.988-0.989 0.963-0.974 0.994-0.997 0.981-0.992 0.976-0.986 0.979-0.9840.981-0.9850.984-0.991 0.982-0.987 | | | | | | | 0.987 -0.99 | 0.968 -0.978 |
| biceps femoris: long head | | | 0.981 -0.982 0.946 -0.964 0.976 -0.985 0.964 -0.977 0.967 -0.976 0.975 -0.98 0.972 -0.983 -0.97 -0.978 | | | | | | | 0.979-0.982 | 0.956 -0.974 |
| biceps femoris: short head | | | | | | | | | 0.96-0.969 0.88-0.931 0.827-1.025 0.919-0.979 0.93-0.967 0.936-0.9550.951-0.9570.967-0.972 0.943-0.961 | 0.953-0.96 | 0.887 -0.94 |
| external rotators | | | | | | | | | 0.867-0.876 0.809-0.834 0.884-0.909 0.821-0.8510.846-0.867 0.852-0.8670.869-0.8830.868-0.881 0.861-0.874 | 0.843 -0.861 | 0.809 -0.837 |
| femur | | | | | | | | | 0.987-0.99 0.979-0.984 0.921-0.985 0.983-0.9910.986-0.9880.988-0.989 0.99-0.992 0.984-0.988 | 0.989 -0.991 | 0.981 -0.985 |
| fibula | | | 0.914-0.927 0.797-0.852 0.941-0.958 0.904-0.9190.909-0.924 0.858-0.895 0.856-0.94 0.918-0.931 0.902-0.92 | | | | | | | 0.867 -0.919 | 0.773 -0.86 |
| fibulari | | | | | | | | | 0.961-0.969 0.918-0.946 0.978-0.983 0.962-0.97 0.956-0.968 0.942-0.96 0.922-0.9720.963-0.972 0.957-0.966 | 0.939-0.97 | 0.91 -0.954 |
| flexor digitorum longus | | | | | | | | | 0.863-0.882 0.827-0.851 0.905-0.935 0.855-0.8740.853-0.875 0.875 0.875 0.875 0.875 -0.898 0.869-0.882 | 0.793 -0.864 | 0.822 -0.857 |
| flexor hallucis longus | | | | | | | | | 0.94-0.952 0.869-0.909 0.963-0.976 0.94-0.953 0.932-0.942 0.91-0.936 0.879-0.9560.949-0.956 0.926-0.942 | 0.907 -0.956 | 0.871 -0.927 |
| Gastrocnemius: lateral head 0.972 -0.974 0.982-0.986 0.969-0.976 0.964 0.966-0.97 0.97-0.976 0.953-0.969 | | | | | | | | | | 0.97 -0.974 | 0.938 -0.962 |
| Gastrocnemius: medial head 0.979-0.983 0.928-0.993 0.979-0.985 0.969-0.98 0.951-0.9690.971-0.9810.972-0.985 0.96 0.976 | | | | | | | | | | 0.977 -0.984 | 0.949 -0.969 |
| gluteus maximus | | | | | | | | | 0.993-0.994 0.977-0.985 0.998-0.998 0.99-0.995 0.99-0.993 0.993-0.9940.991-0.995 0.988-0.993 | 0.994 -0.995 | 0.983 -0.988 |
| gluteus medius | | | 0.985-0.987 0.966-0.975 0.993-0.995 0.981-0.9890.977-0.983 0.977-0.9820.983-0.988 0.98 -0.98 -0.98 -0.984 | | | | | | | 0.985 -0.988 | 0.97 -0.979 |
| gluteus minimus | | | | | | | | | 0.964-0.966 0.929-0.946 0.974-0.982 0.949-0.9660.951-0.959 0.951-0.96 0.957-0.9630.965-0.969 0.954-0.963 | 0.959 -0.965 | 0.937 -0.95 |
| gracilis | | | 0.969-0.971 0.907-0.939 0.964-0.986 0.94-0.972 0.956-0.97 0.943-0.9610.955-0.962 0.97 -0.976 | | | | | | 0.95 -0.967 | 0.959-0.965 | 0.927 -0.949 |
| iliacus | 0.974 -0.976 0.94 -0.96 | | | 0.986-0.989 0.967 -0.9790.965 -0.971 0.958-0.97 0.967 -0.971 0.976 -0.98 | | | | | 0.964 -0.974 | 0.971 -0.975 | 0.94 -0.964 |
| obturator externus | | | | | | | | | 0.927-0.934 0.881-0.904 0.945-0.96 0.893-0.9210.901-0.9240.918-0.9290.924-0.9390.916-0.938 0.92-0.931 | 0.919-0.931 | 0.893 -0.913 |
| obturator internus | | | | | | | | | 0.881 -0.889 0.815 -0.855 0.895-0.919 0.836 -0.865 0.863 -0.8850.869-0.8890.889 -0.902 0.875-0.891 | 0.85 -0.865 | 0.803 -0.855 |
| pectineus | | | | | | | | | 0.946-0.952 0.894-0.923 0.965-0.974 0.927 -0.9520.908 -0.933 0.932-0.947 0.943-0.95 0.952 -0.963 0.948 | 0.943 -0.947 | 0.883 -0.924 |
| pelvis | | | | | | | | | 0.978-0.981 0.952-0.964 0.989-0.992 0.968-0.9820.969-0.977 0.968-0.9740.979-0.981 0.98-0.984 0.971-0.977 | 0.978 -0.983 | 0.956 -0.967 |
| phalangeal extensors | | | | | | | | | 0.948-0.958 0.879-0.912 0.967-0.977 0.946-0.9570.942-0.954 0.917-0.9390.896-0.9630.953-0.959 0.932-0.947 | 0.918 -0.96 | 0.892 -0.938 |
| piriformis | | | | | | | | | 0.917 -0.936 0.88 -0.908 0.938 -0.932 0.894 -0.924 0.917 -0.9290.919 -0.9350.883 -0.949 0.923 -0.932 | 0.914 -0.927 | 0.873 -0.912 |
| popliteus | | | | | | | | | 0.863-0.889 0.846-0.865 0.889-0.911 0.852-0.8680.873-0.887 0.849-0.89 0.857-0.8760.895-0.906 0.854-0.895 | 0.84 -0.852 | 0.837 -0.854 |
| psoas major | | | 0.98-0.982 0.943-0.965 0.989-0.974-0.984 0.976-0.98 0.961-0.9740.975-0.9790.979-0.984 0.969-0.98 | | | | | | | 0.978-0.982 | 0.944 -0.97 |
| quadratus femoris | | | 0.905-0.915 0.774-0.833 0.859-0.962 0.871-0.9030.876-0.908 0.848-0.8870.867-0.9070.914-0.925 0.878-0.907 | | | | | | | 0.884 -0.9 | 0.759-0.833 |
| quadratus lumborum | | | 0.926-0.933 0.865-0.896 0.93-0.951 0.903-0.9240.914-0.9320.895-0.9170.922-0.9310.922-0.944 0.911-0.93 | | | | | | | 0.902 -0.919 | 0.87 -0.898 |
| rectus femoris | | | | | | | | | 0.982 -0.985 0.915 -0.946 0.947 -0.982 0.972 -0.9880.949 -0.976 0.98 -0.983 0.97 -0.988 0.962 -0.976 | 0.983 -0.986 | 0.923 -0.965 |
| sartorius | | | | | | | | | 0.974-0.976 0.928-0.952 0.984-0.987 0.958-0.9780.956-0.974 0.955-0.9670.968-0.971 0.976-0.98 0.963-0.972 | 0.97 -0.974 | 0.927 -0.958 |
| semimembranosus | | | | | | | | | 0.983-0.985 0.912-0.959 0.991-0.993 0.934-0.9940.955-0.976 0.969-0.979 0.975-0.98 0.98-0.986 0.971-0.981 | 0.981 -0.984 | 0.943 -0.967 |
| semitendinosus | | | | | | | | | 0.982-0.984 0.989-0.964 0.988-0.992 0.962-0.9870.969-0.981 0.968-0.9780.979-0.987 0.975-0.982 | 0.977 -0.981 | 0.946 -0.971 |
| soleus | | | 0.982-0.986 0.937-0.958 0.993-0.995 0.983-0.99 0.968-0.981 0.962-0.9750.964-0.984 0.98 -0.985 | | | | | | 0.97 -0.98 | 0.975 -0.989 | 0.943 -0.968 |
| tensor fasciae latae | | | 0.955-0.959 0.907-0.937 0.943-0.974 0.948-0.9560.939-0.958 0.934-0.9560.958-0.965 0.939-0.957 | | | | | | | 0.94 -0.95 | 0.919 -0.942 |
| tibia | | | 0.98 -0.984 0.966 -0.974 0.993 -0.994 0.976 -0.983 0.979 -0.984 0.975 -0.98 0.958 -0.986 -0.98 -0.98 -0.98 -0.98 - | | | | | | | 0.966 -0.983 | 0.966 -0.976 |
| tibialis anterior | | | | | | | | | 0.965-0.972 0.914-0.94 0.979-0.985 0.964-0.9710.959-0.969 0.941-0.96 0.931-0.972 0.96-0.977 0.956-0.967 | 0.944 -0.97 | 0.917 -0.955 |
| tibialis posterior | | | | | | | | | 0.957-0.965 0.938-0.95 0.973-0.98 0.953-0.965 0.955-0.96 0.954-0.9610.912-0.968 0.957-0.962 | 0.932 -0.968 | 0.941 -0.954 |
| vastus intermedius | 0.98 -0.981 | 0.91 -0.955 | | | | | | | 0.99 -0.994 0.925 -0.992 0.958 -0.978 0.961 -0.9710.971 -0.976 0.97 -0.985 0.97 -0.977 | 0.977 -0.982 | 0.942 -0.962 |
| vastus lateralis | | | | | | | | | 0.991 -0.992 0.96 -0.979 0.992 -0.998 0.97 -0.997 0.982 -0.987 0.987 -0.99 0.985 -0.994 0.984 -0.989 | 0.991 -0.994 | 0.977 -0.984 |
| vastus medialis | | | 0.988-0.99 0.947-0.973 0.962-0.995 0.968-0.9980.975-0.986 0.977-0.9870.984-0.993 0.979-0.987 | | | | | | | 0.988 -0.99 | 0.965-0.978 |
{14}------------------------------------------------
| | | Table 5: Volume difference (ml) 95% confidence interval for all 80 musculoskeletal structures across |
|------------------------|--|------------------------------------------------------------------------------------------------------|
| all subgroup analyses. | | |
| | Descriptive Statistics (VDt) ml 95% Confidence Interval | | | | | | | | | | |
|-----------------------------|----------------------------------------------------------|-------------|---------------------------------------|-------------|------------|-----------------------|------------------------------------|---------------------------------------|-------------|--------------|--------------------------------------------------------------|
| ROI | Pathology Level<br>Subgroup Analysis | | MRI Manufacturer<br>Subgroup Analysis | | | | | Age/Biologic Sex<br>Subgroup Analysis | | | |
| | Healthy | Patient | Canon | GE | Philips | Siemens | Toshiba | Years | Years | years | Male > 21 Male 18-21 Female > 21 Female 18-21<br>years |
| adductor brevis | 1.09 -1.74 | 1.51 -2.29 | 0-0.09 | 0.7 -1.52 | 0.98 -1.98 | 1.54 -2.41 | 0.6 -0.94 | 1.21 -2 | 1.19 -1.77 | 0.09 -2.46 | 1.28-2.57 |
| adductor longus | 1.5 -2.26 | 1.52 -2.59 | 0.08 -2.17 | 1.08 -2.52 | 1.93 -3.18 | 1.48 -2.47 | 0.62 -1.08 | 1.37 -2.54 | 1.69 -2.45 | 0.19 -2.31 | 0.88 -2.79 |
| adductor magnus | 3.43 -5.45 | 3.74 -5.62 | 0.53 -1.84 | 2.96 -5.6 | 2.5 -5.58 | 3.75 -6.24 | 2.37 -6.3 | 2.6 -4.76 | 3.66 -5.42 | 2.16 -9.98 | 2.6 -4.36 |
| biceps femoris: long head | 1.13 -1.72 | 2.28 -4.2 | 0.13 -1.4 | 1.18 -3.66 | 1.41 -2.25 | 1.54 -2.59 | 0.79 -1.47 | 1 -2.38 | 1.52 -2.51 | 0.34 -2.15 | 1.21 -3.44 |
| biceps femoris: short head | 0.79 -2.28 | 1.86 -3.38 | 0-0.09 | 1.07 -6.66 | 0.86 -1.28 | | 1.34 -2.06 | 1.08 -1.99 | 1.28-2.16 | -1.03 -6.24 | 1 -2.11 |
| external rotators | 0.72 -0.98 | 0.73 -1.1 | 0.16 -0.89 | 0.47 -0.92 | 0.7 -1.27 | 0.81 -1.11 | 0.37 -0.84 | 0.76 -1.31 | 0.8 -1.1 | 0.33 -0.74 | 0.43 -0.89 |
| femur | 1.58 -5.2 | 2.07 -3.04 | -3.97 -43.01 0.23 -11.13 | | 1.42 -2.27 | | 1.76 -2.55 | 1.26 -1.99 | 1.49 -4.51 | -0.64 -13.83 | 1.76 -3.08 |
| fibula | 0.52 -0.77 | 2.37 -4.27 | 0.06 -0.22 | 1.18 -3.27 | 0.45 -0.77 | 1.09 -1.77 | 0.28 -0.49 | 0.57 -1.22 | 1.07 -2.14 | 0.29 -0.51 | 0.75 -1.92 |
| fibulari | 1.49 -2.47 | 2.75 -5.28 | 0.09 -1.95 | 1.43 -3.76 | 1.7 -2.53 | 1.91 -3.71 | 1.54 -2.47 | 1.04 -4.11 | 2.04 -3.27 | 1.01 -1.8 | 1.46 -4.35 |
| flexor digitorum longus | 0.65 -1.21 | 0.97 -1.88 | 0.04 -0.54 | 0.66 -1.44 | 0.48 -0.83 | 0.91 -1.86 | 0.39 -0.76 | 0.65 -2.49 | 0.71 -1.2 | 0.41 -0.79 | 0.65 -1.51 |
| flexor hallucis longus | 0.93 -1.57 | 1.81 -3.52 | 0.12 -0.46 | 0.93 -2.27 | 0.69 -2.83 | | 1.32 -2.23 0.56 -1.38 | 0.8-1.48 | 1.36-2.51 | 0.29 -0.88 | 1.09 -3.2 |
| Gastrocnemius: lateral head | 1.35 -2 | 2.41 -4.07 | 0-6.67 | 1.32 -3.03 | 1.34 -1.87 | 1.85 -2.97 | 0.78-1.79 | 1.22 -2.16 | 1.77 -2.88 | 0.68 -1.92 | 1.55 -3.42 |
| Gastrocnemius: medial head | 1.18 -3.03 | 4.05 -9.07 | -2.85 -33.17 1.47 -7.55 | | 1.11 -1.68 | 2.34 -4.27 | 0.2 -3.12 | 1-1.78 | 2.32 -6.16 | 0.46 -3.39 | 2.05 -6 |
| gluteus maximus | 4.2 -6.04 | 8.76 -21.66 | 0.04 -3.28 | 0.23 -14.63 | 4.05 -7.74 | 7.24-12.78 1.22 -3.07 | | 4.69 -7.2 | 5.37 -11.23 | 0.69 -2.42 | 6.05 -12.05 |
| gluteus medius | 1.27 -2.29 | 2.22 -3.81 | 0.02 -0.34 | 0.98 -2.28 | 1.46 -2.61 | 1.93 -3.49 | 0.56 -1.11 | 1.54 -3 | 1.65 -3.12 | -0.08 -2.27 | 1.43 -2.41 |
| gluteus minimus | 1-1.38 | 1.54 -2.4 | 0.53 -1.8 | 0.93 -2.04 | 1.06 -1.7 | 1.27 -1.82 | 0.51 -0.9 | 0.98 -1.55 | 1.23 -1.77 | 0.36 -0.85 | 1.12 -2.27 |
| gracilis | 0.77 -1.3 | 1.63 -3.19 | -0.03 -0.5 | 0.71 -2.27 | 0.74 -1.21 | 1.25 -2.28 | 0.7 -1.24 | 0.69 -1.2 | 1.1 -1.94 | 0.2 -2.43 | 0.96 -2.82 |
| iliacus | 1.49 -2.18 | 2.44 -4.44 | 0.08 -0.82 | 1.23 -3.38 | 1.98 -3.37 | 1.94 -3.13 | 0.53 -0.98 | 1.28 -2.42 | 2.17 -3.2 | 0.09 -1.93 | 1.02 -3.8 |
| obturator externus | 1.78 -2.53 | 1.99 -3.08 | 0.2 -1.5 | 1.3 -2.34 | 1.71 -2.93 | 2.09 -3.09 | 0.74-2.69 | 1.78-3.55 | 1.88 -2.72 | 1 -2.41 | 1.48 -2.9 |
| obturator internus | 0.64 -0.9 | 1.01 -1.77 | -0.04 -0.47 | 0.82 -1.96 | 0.74 -1.28 | 0.72 -1.03 | 0.35 -0.74 | 0.7 -1.23 | 0.85 -1.33 | 0.24 -0.74 | 0.47 -1.11 |
| pectineus | 1.55 -2.71 | 1.62 -2.58 | -0.05 -0.72 | 0.66 -1.84 | 2.47 -5 | 1.54 -2.8 | 0.32 -1.02 | 0.9 -1.87 | 2.06 -3.6 | 0.1 -2.31 | 1.22 -2.25 |
| pelvis | 1.66 -2.38 | 2.53 -3.98 | 0.45 -1.49 | 1.25 -2.75 | 1.47 -2.39 | 2.49 -3.61 | 0.59 -1.26 | 1.16 -2.65 | 2.21 -3.21 | 0.7 -2.35 | 1.75 -3.06 |
| phalangeal extensors | 1.24 -1.59 | 1.84 -3 | 0.18 -0.75 | 1.28 -2.45 | 0.92 -1.49 | | 1.68 -2.28 0.86 -1.31 | 1.19 -1.89 | 1.49 -2.15 | 0.83 -1.19 | 1.25 -2.48 |
| piriformis | 1.09 -1.72 | 1.64 -2.8 | -0.1 -1.15 | 1.05 -3.26 | 0.87 -1.53 | 1.4 -2.08 | 0.82 -1.98 | 0.8 -2.82 | 1.15 -1.71 | 1.04 -1.89 | 1.16 -2.73 |
| popliteus | 0.26 -0.64 | 0.4 -0.64 | -0.01 -0.03 | 0.29 -0.64 | 0.27 -0.41 | 0.31 -0.86 | 0.21 -0.36 | 0.34 -0.54 | 0.3 -0.88 | 0.15 -0.27 | 0.26 -0.42 |
| psoas major | 2.29 -3.39 | 2.31 -6.09 | 0.29 -1.45 | 1.57 -4.75 | 2.49 -4.77 | 2.51 -4.75 | 0.58-1.33 | 2.4 -4.77 | 2.41 -4.07 | 0.85 -2.32 | 1.09 -6.89 |
| quadratus femoris | 0.79 -1.1 | 0.99 -1.81 | 0.33 -1.19 | 0.68 -1.21 | 0.72 -1.22 | 1.01 -1.59 | 0.29 -0.58 | 0.79 -1.31 | 0.92 -1.24 | 0.26 -1.34 | 0.59 -1.95 |
| quadratus lumborum | 1.78 -2.25 | 2.35 -5.13 | 1.41 -3.92 | 2.06 -3.75 | 1.63 -2.45 | 1.96 -3.71 | 0.77 -1.58 | 1.4 -2.34 | 1.84 -3.47 | 1.38 -2.5 | 1.88 -3.85 |
| rectus femoris | 1.98 -3.59 | 3.47 -6.14 | -5.19 -20.36 1.29 -5.04 | | 2.41 -3.55 | 2.83 -4.69 | 0.8 -1.66 | 1.59 -4.31 | 2.37 -4.44 | 0.41 -4.4 | 2.71 -5.85 |
| sartorius | 1.37 -2.01 | 2.66 -5.16 | 0.06 -0.73 | 1.27 -4.35 | 1.79 -3.37 | 1.89 -3.07 | 0.45 -1 | 1.05 -3.06 | 1.85 -3.11 | 0.19 -1.86 | 1.83 -4.09 |
| semimembranosus | 1.69 -2.93 | 3.28-6.43 | 0.01 -0.71 | 0.9-3.64 | 1.84 -3.49 | | 2.45 -4.58 1.19 -4.41 | 1.5 -2.62 | 2.21 -4.07 | 0.67 -4.61 | 1.5 -5.07 |
| semitendinosus | 1.33 -1.99 | 2.37 -4.13 | 0.02 -0.57 | 0.5 -1.78 | 1.79 -2.86 | 1.96 -3.14 | 0.92 -1.49 | 0.87 -2.45 | 1.82 -2.75 | 0.61 -1.18 | 1.45 -3.65 |
| soleus | 2.23 -3.76 | | 6.99 -12.81 -1.74 -12.49 | 2.41 -7.78 | 1.98 -3.6 | 3.91 -7.1 | 2.25 -5.58 | 2.1 -5 | 3.37 -6.51 | 0.97 -5.96 | 3.72 -9.58 |
| tensor fasciae latae | 1.06 -1.51 | 1.77 -3.34 | -0.25 -3.22 | 1.35 -3.02 | 0.65 -1.38 | 1.47 -2.33 | 0.39 -0.88 | 0.79 -1.59 | 1.5 -2.37 | 0.46 -1.39 | 0.91 -2.44 |
| tibia | 1.96 -3.16 | 2.87 -5.08 | -3.1 -10.73 | 2.1 -4.12 | 1.4 -4.15 | 2.43 -3.92 | 0.79 -1.68 | 1.66 -3.14 | 2.33 -4.1 | 1.24 -2.65 | 1.96 -5.72 |
| tibialis anterior | 0.82 -1.88 | 1.71 -2.75 | -5.14 -17.75 | 0.68 -1.77 | 0.84 -1.47 | 1.32 -1.92 | 1.11 -1.7 | 0.83 -1.73 | 0.95 -2.69 | 0.71 -1.24 | 1.25 -2.25 |
| tibialis posterior | 1.09 -1.48 | 1.27 -2.16 | 0.24 -0.61 | 0.7 -1.72 | 1.34 -2.03 | | 1.31 -1.85 -0.45 -0.97 | 1.48 -2.34 | 1.18-1.73 | 0.34 -1.09 | 0.85 -1.68 |
| vastus intermedius | 2.05 -2.73 | 2.46 -4.12 | 0.04 -0.54 | 0.97 -3.11 | 2.05 -3.35 | 2.55 -3.48 | 1.81 -3.41 | 2.1 -3.36 | 2.32 -3.22 | 1.04 -2.55 | 1.57 -3.49 |
| vastus lateralis | 3.72 -5.69 | 5.57 -9.23 | 0.19-1.52 | 1.97 -6.94 | 4.54 -6.35 | 5.03 -7.73 | 1.3 -7.45 | 3.39 -7.21 | 3.79 -5.23 | 1.78 -9.16 | 4.81 -8.79 |
| vastus medialis | 2.08 -4.39 | 3.17 -6.29 | -3.01 -11.05 | 1.21 -7.64 | 2.33 -3.48 | | 2.36 -3.87 -0.26 -13.14 1.64 -3.29 | | 2.04 -3.35 | 1.43 -12.16 | 2.2 -5.5 |
| | | | | | | | | | | | |
{15}------------------------------------------------
### 9. Substantial Equivalence Conclusion
In conclusion, MuscleView is intended for use in adults and pediatric patients aged 18 and older to quantify automatically segment muscle and bone structures of the lower extremities from magnetic resonance imaging and provide derived metrics including muscle volume, bone volume, intramuscular fat percentage, and left/right asymmetry and other derived metrics. It has similar intended use and indications for use statement as the AMRA Profiler (510k Number: K173749). This 510(k) submission includes information on the MuscleView technological characteristics, as well as performance data and verification and validation activities demonstrating that MuscleView is as safe and effective as the AMRA Profiler (510k Number: K173749), and does not raise different questions of safety and effectiveness.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.