K241250 · Davol Inc., Subsidiary of C. R. Bard, Inc. · FTL · Jun 14, 2024 · General, Plastic Surgery
Device Facts
Record ID
K241250
Device Name
3DMax Light Mesh
Applicant
Davol Inc., Subsidiary of C. R. Bard, Inc.
Product Code
FTL · General, Plastic Surgery
Decision Date
Jun 14, 2024
Decision
SESE
Submission Type
Special
Regulation
21 CFR 878.3300
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The 3DMax™ Light Mesh is indicated for use in the reinforcement of soft tissue where weakness exists, in the repair of inguinal hernias.
Device Story
3DMax Light Mesh is an anatomically shaped, preformed surgical mesh designed for laparoscopic inguinal hernia repair. Constructed from knitted polypropylene monofilaments (0.0048 inch diameter), the mesh features sealed edges and an 'M'-shaped orientation marker with an arrow to assist surgeons in correct positioning. The device is implanted by a surgeon to reinforce soft tissue where weakness exists. It is designed to stretch in multiple directions to accommodate tissue defects. The device is identical to its predicate; this submission serves only to notify the FDA of labeling updates to align with European medical device regulations and internal portfolio standards.
Clinical Evidence
No non-clinical or clinical testing was provided in support of this Special 510(k).
Technological Characteristics
Knitted polypropylene monofilament mesh; 0.0048 inch diameter filaments. Features 3D-shaped design with preformed, semi-rigid edge seal. Includes polypropylene orientation marker with Phthalocyaninato(2-) copper colorant. Sterilized via ethylene oxide. Available in Medium, Large, and Extra Large sizes for left or right orientation.
Indications for Use
Indicated for soft tissue reinforcement in patients requiring repair of inguinal hernias.
Regulatory Classification
Identification
Surgical mesh is a metallic or polymeric screen intended to be implanted to reinforce soft tissue or bone where weakness exists. Examples of surgical mesh are metallic and polymeric mesh for hernia repair, and acetabular and cement restrictor mesh used during orthopedic surgery.
{1}
[LOGO]
FDA
U.S. FOOD & DRUG
ADMINISTRATION
June 14, 2024
Davol Inc., Subsidiary of C.R. Bard, Inc.
Richard Spaziano
Staff Regulatory Affairs Specialist
100 Crossings Boulevard
Warwick, Rhode Island 02886
Re: K241250
Trade/Device Name: 3DMax Light Mesh
Regulation Number: 21 CFR 878.3300
Regulation Name: Surgical Mesh
Regulatory Class: Class II
Product Code: FTL
Dated: May 2, 2024
Received: May 3, 2024
Dear Richard Spaziano:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
U.S. Food & Drug Administration
10903 New Hampshire Avenue
Silver Spring, MD 20993
www.fda.gov
{2}
K241250 - Richard Spaziano
Page 2
Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Tek N.
Lamichhane -S
Digitally signed by Tek N.
Lamichhane -S
Date: 2024.06.14 16:19:33 -04'00'
Tek N. Lamichhane, Ph.D.
Assistant Director
DHT4B: Division of Infection Control
and Plastic and Reconstructive Surgery Devices
OHT4: Office of Surgical and Infection Control Devices
Office of Product Evaluation and Quality
Center for Devices and Radiological Health
Enclosure
{3}
DEPARTMENT OF HEALTH AND HUMAN SERVICES
Food and Drug Administration
# Indications for Use
Form Approved: OMB No. 0910-0120
Expiration Date: 07/31/2026
See PRA Statement below.
Submission Number (if known)
K241250
Device Name
3DMax Light Mesh
Indications for Use (Describe)
The 3DMax™ Light Mesh is indicated for use in the reinforcement of soft tissue where weakness exists, in the repair of inguinal hernias.
Type of Use (Select one or both, as applicable)
Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
# CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
# *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
Department of Health and Human Services
Food and Drug Administration
Office of Chief Information Officer
Paperwork Reduction Act (PRA) Staff
PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{4}
K241250
Page 1 of 3
### 510(k) Summary
K241250
# **Submitter:**
Davol Inc., Subsidiary of C.R. Bard, Inc.
100 Crossings Blvd.
Warwick, RI 02886
United States
Contact: Richard Spaziano
Title: Staff Regulatory Affairs Specialist
Email: richard.spaziano@bd.com
Phone: (401) 825-8566
Date of Submission: May 3, 2024
# **Subject Device:**
| Device Trade Name: | 3DMax™ Light Mesh |
| --- | --- |
| Common Name: | Surgical mesh |
| Classification Name: | Mesh, Surgical, Polymeric |
| Regulatory Class: | II |
| Regulation Number: | 21 CFR 878.3300 |
| Product Code: | FTL |
# **Predicate Device:**
| Device Trade Name: | 3DMax™ Light Mesh (K091659), cleared on August 3, 2009 |
| --- | --- |
| Common Name: | Surgical mesh |
| Classification Name: | Mesh, Surgical, Polymeric |
| Regulatory Class: | II |
| Regulation Number: | 21 CFR 878.3300 |
| Product Code: | FTL |
# **Device Description:**
The purpose of this Special 510(k) is to notify FDA of the changes to the 3DMax™ Light Mesh labeling. There are no changes to the device itself. The device description is identical to the device description for the predicate device, 3DMax™ Light Mesh (K091659).
The 3DMax™ Light Mesh is designed to fit the inguinal anatomy. The device is anatomically shaped and preformed with sealed edges that allow for easier positioning of the device than a traditional flat sheet of mesh in a laparoscopic inguinal hernia repair. The device is constructed of knitted polypropylene monofilaments 0.0048 inches in diameter. The knit construction allows the mesh to be stretched in both directions in order to accommodate and reinforce tissue defects. The device contains an orientation marker in an "M" shape to designate the medial edge of the mesh and an arrow pointing to the medial aspect. Both the arrow and "M" will aid the user during positioning and placement of the device. The orientation marker is composed of polypropylene monofilament dyed with Phthalocyaninato(2-) copper colorant.
{5}
K241250
Page 2 of 3
### Indications for Use:
The 3DMax™ Light Mesh is indicated for use in the reinforcement of soft tissue where weakness exists, in the repair of inguinal hernias.
### Indications for Use Comparison to Predicate Device:
The subject device and predicate device have identical intended uses, which is for use in soft tissue reinforcement where weakness exists. The subject device indications specify use in the repair of inguinal hernias while the predicate device indications are broad hernia repair indications. The differences between the subject device and predicate device indications do not constitute a new intended use, as both the subject and predicate device are indicated for the same tissue type (soft tissue) and disease entity (repair of hernias).
### Technological Comparison to Predicate Device:
The subject device is identical to the predicate device in terms of design, materials, manufacturing, packaging, sterilization, biocompatibility, and shelf-life.
| Device Characteristic | Subject Device: 3DMax™ Light Mesh (K241250) | | | Predicate Device: 3DMax™ Light Mesh (K091659) | | | Comparison Assessment |
| --- | --- | --- | --- | --- | --- | --- | --- |
| Device Classification | Class II Surgical Mesh 21 CFR 878.3300 Product Code: FTL | | | Class II Surgical Mesh 21 CFR 878.3300 Product Code: FTL | | | Same |
| Intended Use | Soft tissue repair/reinforcement | | | Soft tissue repair/reinforcement | | | Same |
| Indications for Use | The 3DMax™ Light Mesh is indicated for use in the reinforcement of soft tissue where weakness exists, in the repair of inguinal hernias. | | | The 3DMax™ Light Mesh is indicated for use in the reinforcement of soft tissue where weakness exists, such as in the repair of hernia defects. | | | Similar |
| Principles of Operation | • Insertion • Placement/Positioning | | | • Insertion • Placement/Positioning | | | Same |
| Materials | **Mesh:** Polypropylene monofilament **Orientation marker:** Polypropylene monofilament with blue colorant | | | **Mesh:** Polypropylene monofilament **Orientation marker:** Polypropylene monofilament with blue colorant | | | Same |
| Design | Single layer knitted 3D-shaped mesh with preformed, semi-rigid edge seal and orientation marker ("M" and arrow). | | | Single layer knitted 3D-shaped mesh with preformed, semi-rigid edge seal and orientation marker ("M" and arrow). | | | Same |
| Sizes | Size | Orientations | Dimensions | Size | Orientations | Dimensions | Same |
| | Medium | Left or Right | 7.9 cm x 13.4 cm (3.1" x 5.3") | Medium | Left or Right | 7.9 cm x 13.4 cm (3.1" x 5.3") | |
| | Large | Left or Right | 10.3 cm x 15.7 cm (4.1" x 6.2") | Large | Left or Right | 10.3 cm x 15.7 cm (4.1" x 6.2") | |
| | Extra Large | Left or Right | 12.2 cm x 17.0 cm (4.8" x 6.7") | Extra Large | Left or Right | 12.2 cm x 17.0 cm (4.8" x 6.7") | |
| Sterilization Method | Ethylene oxide | | | Ethylene oxide | | | Same |
The purpose of this Special 510(k) is to notify FDA of labeling updates to the 3DMax™ Light Mesh that are being made to comply with the requirements of the European medical device regulation, Regulation (EU) 2017/745, as well as to align the labeling with other hernia mesh devices across the Davol Inc. portfolio. These labeling modifications do not impact the substantial equivalence between the subject and predicate device.
{6}
K241250
Page 3 of 3
# Non-Clinical and/or Clinical Tests Summary & Conclusions:
No non-clinical or clinical testing was provided in support of this Special 510(k).
# Conclusion:
The subject device and predicate device have the same intended use and technological characteristics. The only differences between the subject and predicate device are modifications to the labeling for the subject device. These labeling modifications do not impact the safety or performance of the subject device, nor do they impact the substantial equivalency between the subject and predicate devices. Therefore, the subject device, 3DMax™ Light Mesh, is substantially equivalent to the predicate device, 3DMax™ Light Mesh (K091659).
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.