AGILIS - Multi-function Physical Therapy Table (Model: Plus, Pro and Basic.)
K241193 · Marzzan Lucas Jorge · JFB · Jan 31, 2025 · Physical Medicine
Device Facts
Record ID
K241193
Device Name
AGILIS - Multi-function Physical Therapy Table (Model: Plus, Pro and Basic.)
Applicant
Marzzan Lucas Jorge
Product Code
JFB · Physical Medicine
Decision Date
Jan 31, 2025
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 890.5880
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
AGILIS is a Multi-function physical therapy tables intended for medical purposes that consists of a motorized table which can be equipped to provide patients with powered flexion, and muscle relaxation therapy. Neither device provides heat therapies.
Device Story
AGILIS is a multi-function physical therapy table for clinical use by professionals. It provides powered distraction, powered flexion, and muscle relaxation therapy via 24Vdc electric motor actuators. The device features articulated modules for manual or electric adjustments, including horizontal elevation, lower limb flexion, and linear distraction. It includes manual drop systems, adjustable headpieces, and lateral flexion. The operator controls the device via a touchscreen interface and foot pedals. An emergency stop mechanism is accessible to both patient and operator; activation triggers a visual alarm and halts operation, requiring manual reset by the operator. The device is designed to improve patient outcomes through controlled mechanical therapy. It is available in Plus, Pro, and Basic configurations, with the Plus and Pro models featuring motorized lifting systems.
Clinical Evidence
No clinical studies were performed. Evidence is based on non-clinical bench testing, including static and dynamic load application, functional analysis, and electrical safety/EMC testing (EN/IEC 60601-1 and IEC 60601-1-2).
Technological Characteristics
Multi-function physical therapy table with vinyl upholstery. Powered by 24Vdc electric motor actuators. Features include manual drop systems, adjustable headpieces, and lateral flexion. Connectivity includes a touchscreen control interface. Safety features include dual emergency stops and visual alarm systems. Complies with EN/IEC 60601-1 and IEC 60601-1-2 standards.
Indications for Use
Indicated for patients requiring powered distraction, powered flexion, and muscle relaxation therapy. Contraindicated for patients under 22 years or <100 lbs; severe osteoporosis; recent spinal surgery/fractures; spinal tumors/infections; severe psychiatric disorders; liver/renal dysfunction; cardiac failure; pregnancy; history of CVA or myocardial infarction within 6 months.
Regulatory Classification
Identification
A multi-function physical therapy table is a device intended for medical purposes that consists of a motorized table equipped to provide patients with heat, traction, and muscle relaxation therapy.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo for the U.S. Food and Drug Administration (FDA). The logo consists of two parts: on the left, there is a symbol representing the Department of Health & Human Services - USA, and on the right, there is the text "FDA U.S. FOOD & DRUG ADMINISTRATION" in blue. The FDA logo is a recognizable symbol of the agency's role in regulating and overseeing food and drug products in the United States.
January 31, 2025
Marzzan Lucas Jorge % Jose Huerta Consultant Sanrubio. LLC 8345 Park Blvd, Apt 6310. Mailbox 289 Miami, Florida 33126
Re: K241193
Trade/Device Name: AGILIS - Multi-function Physical Therapy Table (Model: Plus, Pro and Basic.) Regulation Number: 21 CFR 890.5880 Regulation Name: Multi-Function Physical Therapy Table Regulatory Class: Class II Product Code: JFB Dated: January 3, 2025 Received: January 3, 2025
Dear Jose Huerta:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device"
{1}------------------------------------------------
(https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30. Design controls; 21 CFR 820.90. Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review. the OS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rue"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advicecomprehensive-regulatory-assistance/unique-device-identification-system-udi-system.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
{2}------------------------------------------------
Sincerely,
# AmberT. Ballard -S
Amber Ballard, PhD Assistant Director DHT5B: Division of Neuromodulation and Physical Medicine Devices OHT5: Office of Neurological and Physical Medicine Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{3}------------------------------------------------
# Indications for Use
510(k) Number (if known) K241193
Device Name
AGILIS - Multi-function Physical Therapy Table (Model: Plus, Pro and Basic.)
Indications for Use (Describe)
AGILIS is a Multi-function physical therapy tables intended for medical purposes that consists of a motorized table which can be equipped to provide patients with powered flexion, and muscle relaxation therapy. Neither device provides heat therapies.
| Type of Use (Select one or both, as applicable) | |
|-------------------------------------------------|---------------------------------------------|
| × Prescription Use (Part 21 CFR 801 Subpart D) | Over-The-Counter Use (21 CFR 801 Subpart C) |
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{4}------------------------------------------------
Image /page/4/Picture/0 description: The image shows the text 'K241193' in a simple, sans-serif font. The text is arranged horizontally, with the letter 'K' followed by the numbers '241193'. The text is black against a white background. The numbers are evenly spaced.
ALTH
# 510(K) SUMMARY
January 31, 2025
# Administrative information
| Applicant | MARZZAN LUCAS JORGE<br>Gdor. Crespo 1333 - Rosario – CP2000 –<br>Santa Fe - Argentina. |
|------------------|---------------------------------------------------------------------------------------------------------------------------------------------------|
| Official Contact | Lucas Jorge Marzzan<br>Tel: +54 -3415 70 0725<br>e-mail: lmarzzan@gmail.com ; info@althery-tables.com |
| Consultant | Jose Huerta<br>SANRUBIO, LLC<br>8345 Park Blvd, Apt 6310. Mailbox 289,<br>Miami, FL 33126 USA<br>Tel. (786) 557-8049<br>e-mail: info@sanrubio.com |
# Device name and classification
| Type of Submission: | Traditional |
|-----------------------|------------------------------------------------|
| Device's Trade Name: | AGILIS - Multi-function Physical Therapy Table |
| Device's Common Name: | AGILIS - Multi-function Physical Therapy Table |
| Classification Name: | Table, Physical Therapy, Multi-Function |
| 510(k) Submitter: | MARZZAN LUCAS JORGE |
| Regulation Number: | 21CFR 890.5880 |
| Device Class: | II |
| Classification Panel: | Physical medicine |
| Product Code: | JFB |
# Predicate information
| 510(K) number | K063034 |
|-----------------------|-----------------------------------------|
| Applicant | HILL LABORATORIES CO. |
| Model name | AIR-FLEX WITH AUTO-DISTRACTION |
| Classification name | Table, Physical Therapy, Multi-Function |
| Classification Panel: | Physical medicine |
| Regulation Number: | 21CFR 890.5880 |
| Device Class: | II |
| Product Code: | JFB |
{5}------------------------------------------------
Image /page/5/Picture/0 description: The image shows the logo for ALTHERY. The logo consists of a stylized flame-like symbol above the word "ALTHERY" in a modern, sans-serif font. The flame symbol is composed of two curved, parallel lines that converge at the top, creating a sense of upward movement. The overall design is clean and minimalist, conveying a sense of sophistication and innovation.
## Indications for Use
AGILIS is a multi-function physical therapy tables intended for medical purposes that consists of a motorized table which can be equipped to provide patients with powered distraction, powered flexion, and muscle relaxation therapy. Neither device provides heat therapies
## Device Description
The AGILIS multifunction physiotherapy table and its line of models (Plus, Pro and Basic) are devices intended to provide patients with powered distraction, powered flexion, and muscle relaxation therapy. Neither device provides heat therapies.
The devices of the AGILIS line are made up of different modules articulated with each other, which provide movements and regulations for the manipulations carried out by the professional. The movements are divided according to the type of operation: manual or electric.
Manual operation systems:
- . Headpiece.
- . Extension of feet.
- Drops. .
- Dorso - lumbar release/breakaway
- Lateral Flexion. ●
Electrical operating systems:
- . Horizontal elevation.
- . Flexion of the lower limb sector.
- Extension or linear movement of the lower limb sector (distraction/traction).
## Physiotherapy table models:
- AGILIS "PLUS" multifunction physiotherapy table, is the most complete model in the line of physiotherapy tables and has the following features:
- Lifting system with 24Vdc electric motor/actuator. .
- Flexion system with 24Vdc electric motor/actuator. •
- Lumbar distraction system with 24Vdc electric motor/actuator.
- Manual drops system.
- Manual adjustment of the articulated headpiece.
- . Manual regulation of the dorso-lumbar release/breakaway.
- . Manual regulation of leg length.
- Manual regulation of lateral flexion movement. ●
- . Wheel system for transportation.
{6}------------------------------------------------
- AGILIS "PRO" multifunction physiotherapy table, only differs from the "Plus" model in that the lifting system is compensated by mechanical springs. which allows the time to raise the equipment to be shorter by having to exert less effort.
- AGILIS "BASIC" multifunction physiotherapy table, this configuration is a simplification of the previous configurations since it does not have a lifting system. It has the following features:
- Flexion system with 24Vdc electric motor/actuator.
- . Lumbar distraction system with 24Vdc electric motor/actuator.
- Manual adjustment of the articulated headpiece. ●
- Manual regulation of the dorso-lumbar release/breakaway. ●
- Manual regulation of leg length. ●
- . Manual regulation of lateral flexion movement.
- Manual drop system
- Fixed height .
#### Dimensions:
- . Physiotherapy table: 65" to 73" length, 23" width - optional 26".
- Headpiece: width 18" wide x 13" long. Made up of 3 pieces: 1 (one) central piece to support the patient's face and 2 (two) sides.
- Upper dorsal sector: 23" wide x 6.5" long.
- Lumbar back sector: 23" wide x 9" long. o
- Pelvic sector and leg: 23" wide x 17" long and 23" x 6" long. ●
- . Extendable sector: 23" wide x 9" long.
#### Components:
- Electrical source, model: MEA-250A24C.
## Accessories:
- Set of belts and fastenings.
- Removable Armrest: This optional component consists of two armrests measuring 23.5" . inches x 6 inches each and an additional upholstery section measuring 23 inches' x 6 inches.
- Side supports: Comprising two lateral supports (right and left) and two belts with a fixed ● end and an adjustable end.
## Contraindications:
Use of the chiropractic table is contraindicated in the following situations:
{7}------------------------------------------------
Image /page/7/Picture/0 description: The image shows the logo for Althery. The logo consists of a stylized flame-like symbol above the word "ALTHERY" in a modern, sans-serif font. The flame symbol is composed of two curved lines that converge at the top, creating a sense of upward movement.
- . Patients under 22 years of age or weighing less than 100 lbs., where the force applied may be inappropriate for their body weight or developmental stage.
- . Patients with severe osteoporosis, where bones are highly fragile and susceptible to fractures under mechanical force.
- Recent spinal surgery or spinal fractures, where mechanical adjustments or traction could disrupt healing or cause further injury.
- Tumors, infections, or fractures of the spine, where mechanical force could exacerbate the condition or cause undue stress to affected areas.
- . Severe psychiatric disorders, such as depression where the condition could affect the patient's ability to cooperate during treatment or pose a safety concern.
- Liver and renal dysfunction, which may limit the patient's ability to recover or handle • physical therapy.
- Patients with cardiac failure, where high-force adjustments may pose a risk to • cardiovascular health.
- Pregnancy, especially in later stages, where mechanical traction or significant force on the ● abdominal area should be avoided.
- Patients with a history of cerebrovascular accidents (CVA) or myocardial infarctions . within the last six months, where the application of physical stress may trigger complications.
Additional exclusion criteria include patients with LBP caused by tumors, infections, or fractures; those with previous back surgery; severe osteoporosis; psychiatric disorders (e.g. schizophrenia); liver or renal dysfunction; pregnancy; medication for cardiac failure; a history of CVA or myocardial infarction within six months; and those deemed unsuitable for traction by the attending physician.
As always, consult the patient's overall medical history and perform a thorough evaluation before beginning chiropractic therapy, and adjust treatment parameters accordingly to ensure both safety and efficacy.
{8}------------------------------------------------
Image /page/8/Picture/0 description: The image shows the logo for Althery. The logo consists of a stylized flame-like symbol above the word "ALTHERY" in a modern, sans-serif font. The flame symbol is composed of two curved, parallel lines that converge at the top, creating a sleek and abstract design.
| Feature | Proposed Device | Predicate Device | Comparison |
|---------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------|
| General characteristics | | | |
| Name | AGILIS | AIR FLEX with auto distraction | - |
| 510(K) number | K241193 | K063034 | - |
| Class | II | II | Same |
| Product code | JFB | JFB | Same |
| Regulation | 21 CFR 890.5880 | 21 CFR 890.5880 | Same |
| Indications for Use | AGILIS is a multi-function physical therapy tables intended for medical purposes that consists of a motorized table which can be equipped to provide patients with powered distraction, powered flexion, and muscle relaxation therapy. Neither device provides heat therapies | The Air-Flex is a multi-function physical therapy table intended for medical purposes that consists of a motorized table which can be equipped to provide patients with powered distraction, powered flexion, and muscle relaxation therapy. | Same |
| Dimensions | Length 65" to 73" - width 23", optional 26". | Length 68" to 77" - width 24", optional 27".<br>Not declared in version I of this device, the measurements represented here were obtained from version II of the predicate device. | Similar - The difference between the specification does not generate or modify the risks already established for this type of devices. |
| Product weight | 330 lbs. without accessories | 250 lbs. | Similar - The difference between the specification does not generate or modify the risks already established for this type of devices. |
| Materials | Vinyl | Vinyl | Same |
| Technological characteristics | | | |
| Electrical operation | 24Vdc | 24 Vdc | Same |
| Abdominal decompression | Yes | Yes | Similar - The difference between the specification does not generate or modify the risks already established for this type of devices. |
| Headpiece | Manual and mechanical operation system with multiple positions: prone, supine, and lateral position. | Manual or air-drop drive system | Similar - The difference between the specification does not generate or modify the risks already |
| | Range of motion:<br>Flexion: 20 degrees when in<br>the uppermost position.<br><br>Extension: 15 degrees when<br>in the lowest position | Angle: 30° positive or negative.<br>Movements: Drops, lateral<br>flexion, and flexion. | risks already<br>established for this type<br>of devices. |
| Lateral motion | Free movement of 15° to<br>each side. It can also be left<br>fixed every 2.5°.<br>Lateral Flexion with locking<br>mechanisms. | Rotation and Lateral<br>Flexion with locking<br>mechanisms, up to 21°. | Similar - The difference<br>between the<br>specification does not<br>generate or modify the<br>risks already<br>established for this type<br>of devices. |
| Arm rest | Yes | Yes | Same |
| Feet extension | Length 65" a 73", | Length 68" to 77"<br>Not declared in version I of this<br>device, the measurements<br>represented here were obtained<br>from version II of the predicate<br>device. | Similar - The difference<br>between the<br>specification does not<br>generate or modify the<br>risks already<br>established for this type<br>of devices. |
| Emergency stop | 2 emergency stops (user and<br>operator)<br>Visual alarm system | 2 emergency stops (user<br>and operator)<br>Alarm system | Similar - The difference<br>between the<br>specification does not<br>generate or modify the<br>risks already<br>established for this type<br>of devices. |
| Drop system | 6 drops: 2 cervical, 1 upper<br>dorsal, 1 lumbar dorsum, 1<br>sacral, and 1 ankle.<br>Manual operation | 5 drops: thoracic, lumbar,<br>pelvic, cervical.<br>Pneumatic actuation | Similar - The difference<br>between the<br>specification does not<br>generate or modify the<br>risks already<br>established for this type<br>of devices. |
| Traction system | Electric actuators system | Pneumatic actuators<br>system | Similar - the<br>differences are<br>analyzed in the<br>following paragraph<br>"substantial<br>equivalence (SE) -<br>rationale" |
| Flexion system | Electric actuators system | Pneumatic actuators<br>system and manual flexion<br>system | Similar - the<br>differences are<br>analyzed in the<br>following paragraph<br>"substantial |
| | | | equivalence (SE) - rationale" |
| Lifting capacity - with lying patient | 450 pounds | 450 pounds | Same |
| Height<br>regulation: | Agilis Plus:21.5" – 29.5"<br>Agilis Pro: 21.5" – 31" | 21.5"-28" or 22.5"-29" | Similar - The difference<br>between the<br>specification does not<br>generate or modify the<br>risks already<br>established for this type<br>of devices. |
| | Agilis Care: fixed height in<br>20", 25" or 30" | | |
| Display | Touchscreen Control | Touchscreen Control | Similar - The difference<br>between the<br>specification does not<br>generate or modify the<br>risks already<br>established for this type<br>of devices. |
| Pedals | Height Pedal | Corded Height Pedal<br><br>Optional Rocker Foot<br>Pedal<br><br>Power Foot Strips for<br>activating Airdrops<br>(optional)<br><br>Air-Pressure Foot Pedal 1 | Similar - The difference<br>between the<br>specification does not<br>generate or modify the<br>risks already<br>established for this type<br>of devices. |
| Air system | Not include | Air Tank<br>Power and Air-Drop<br>Control Panel | Different - The<br>difference between the<br>specification does not<br>generate or modify the<br>risks already<br>established for this type<br>of devices. |
| Accessories | Set of belts and fastenings.<br><br>Removable Armrest<br><br>Side Supports | Seatbelt<br><br>Ankle straps | Similar - The difference<br>between the<br>specification does not<br>generate or modify the<br>risks already<br>established for this type<br>of devices. |
| | | | |
| Safety and performance testing | | | |
| Non-clinical<br>testing | EN/IEC 60601 -1:2005 +<br>AMD1:2012 + AMD2:2020 -<br>Medical Electrical<br>Equipment - Part I: General<br>Requirements for Basic<br>Safety and Essential | IEC 60601-1 Medical<br>Electrical Equipment- Part<br>1: General Requirements<br>for Safety | Similar |
# Substantial equivalence comparison
{9}------------------------------------------------
Image /page/9/Picture/0 description: The image shows the logo for Althery. The logo consists of the word "ALTHERY" in a stylized font, with a symbol above it. The symbol resembles a flame or a stylized leaf with three curved segments.
{10}------------------------------------------------
Image /page/10/Picture/0 description: The image shows the logo for ALTHERY. Above the text "ALTHERY" is a stylized image of a flame. The flame is black and white and appears to be made of two curved lines.
{11}------------------------------------------------
Image /page/11/Picture/0 description: The image shows the logo for ALTHERY. The logo consists of a stylized flame-like symbol above the word "ALTHERY" in a sans-serif font. The flame symbol is made up of three curved lines that converge at the top, resembling a stylized flame or plume. The word "ALTHERY" is written in all capital letters, with a slightly curved line extending from the right side of the "Y".
| | IEC 60601-1-2:2014<br>+AMD1:2020- Medical<br>Electrical Equipment - Part 1-<br>2: General Requirements for<br>Basic Safety and Essential<br>Performance - Collateral<br>Standard: Electromagnetic<br>Compatibility - Requirements<br>and Tests. |
|--|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
|--|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
#### Substantial equivalence (SE) - rationale
The main features and differences found between the devices are summarized below:
- -The proposed device shares the same or similar basic characteristics and the same general use in physical medicine as the predicate device.
- -The product code (JFB) and Indications for Use between the predicate device and the proposed device is the same: multi-function physical therapy table intended for medical purposes that consists of a motorized table which can be equipped to provide patients with powered distraction, powered flexion, and muscle relaxation therapy.
- The proposed device operates on AC to bring the motors voltage to 24 volts DC in the same manner as the AIR FLEX predicate device cleared under K063064. The power is through an external source for medical use.
- The proposed device, as the predicate, allows the automatic cycle of the distraction (traction) and flexion process in a specific period of time with multiple configurable parameters. These automatic cycles are controlled by a microprocessor on both devices.
- The proposed device, as the predicate device, has an emergency stop mechanism accessible to the patient in case the patient feels discomfort. On both devices this mechanism interrupts the operation, and the patient cannot reactivate the operation by himself, the operation must be reactivated by the qualified operator.
- -The materials used in the construction of the device, including the structure and upholstery, are similar.
- The predicated device, as the proposed device, has an alarm system that is activated when the emergency stop is activated; none of the devices allow the patient to reactivate the operation of the device without intervention of the professional operator. The predicate device has an audible alarm, while the proposed device has a visual alarm system through the device screen. This difference does not affect the safety and effectiveness of the device.
Main differences between technological features:
- -The predicated device uses pneumatic actuators for traction and flexion of the device components. while the proposed device uses electric motors to achieve the same result. This difference in the proposed device is for the purpose of minimizing the dependence of the actuator systems on each other, minimizing the possibility of a generalized failure occurring in all the systems
{12}------------------------------------------------
in the tables. leaving it unusable and avoiding the potential failure of the established parameters and that could potentially generate sudden movements due to changes in loading conditions. By working with electric actuators, the independence of each actuator system is achieved, avoiding a generalized failure and at the same time guaranteeing that without power, the load cannot generate unexpected movements.
This difference does not affect the safety and effectiveness of the device.
- The predicated device has a system of 5 drops located in the thoracic, lumbar, pelvic and 2 cervical areas, the proposed device includes 6 drops located as follows: 2 cervical, 1 upper dorsal, 1 lumbar dorsum, 1 sacrum, and 1 ankle. The proposed device adds an ankle drop, which has the same operating mechanism as cervical drops, the risks being those already known, analyzed, and minimized for this type of devices.
This difference does not affect the safety and effectiveness of the device.
Therefore, the proposed device is considered by MARZZAN LUCAS JORGE to be Substantially Equivalent to the predicate device.
#### Non-Clinical Testing Summary
The Multi-function Physical Therapy Table - Model AGILIS was tested to evaluate its safety and effectiveness based on the following standards:
#### Electrical Safety, Electromagnetic Compatibility:
- . EN/IEC 60601 -1:2005 + AMD1:2012 + AMD2:2020 - Medical Electrical Equipment - Part I: General Requirements for Basic Safety and Essential Performance.
- . IEC 60601-1-2:2014 +AMD1:2020- Medical Electrical Equipment - Part 1-2: General Requirements for Basic Safety and Essential Performance - Collateral Standard: Electromagnetic Compatibility - Requirements and Tests.
#### Performance Standard
Since there are no mandatory standards for these type of devices, "MARZZAN LUCAS JORGE" decided to adhere to the FDA guidelines outlined in the guidance document 'Recommended content and format of non-clinical bench performance testing in premarket submissions - Guidance for industry personnel and the Food and Drug Administration', and has performance Safety Testing - Load Application (static and dynamic) and Efficacy Testing - Functional Analysis tests on the AGILIS multifunction physiotherapy tables.
Furthermore, in order to guarantee the correct functioning, safety and effectiveness of the physiotherapy tables, an operational control is carried out on 100% of the manufactured tables before their release to the market, according to the procedures described in this submission.
#### Software
After a comprehensive evaluation that considered: the device. its Indications for Use. software analysis, known or foreseeable software hazards, and dangerous situations associated with the device, including those that could result from misuse, whether intentional or not, and the impact that these
{13}------------------------------------------------
could have on the patient or user, it was determined that the Software documentation level evaluation is Basic Documentation.
The software information is provided in accordance with FDA guidance "Content of Premarket Submissions for Device Software Functions".
#### Animal Study
No animal studies were considered necessary and performed.
## Clinical Studies
No clinical studies were considered necessary and performed.
## Conclusion
The proposed device is substantially equivalent to the currently marketed and predicate device in terms of design features, fundamental scientific technology, indications for use, and safety and effectiveness.
Additionally, substantial equivalence was demonstrated through the non-clinical performance, which complied with the requirements specified in the international and FDA-recognized consensus standards, IEC 60601-1, IEC 60601-1-2, and bench testing.
The results of these tests demonstrate that the proposed device meets the acceptance criteria and is adequate for this Indications for Use. The comparison of technological characteristics, non-clinical performance data, electrical safety and electromagnetic compatibility testing demonstrates that the device is substantially equivalent to the previously cleared predicate devices.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.