K241140 · Applied Photonix, LLC · FRA · Aug 1, 2024 · General Hospital
Device Facts
Record ID
K241140
Device Name
ViroZap Indoor Air Purifier, In Duct Model 1008
Applicant
Applied Photonix, LLC
Product Code
FRA · General Hospital
Decision Date
Aug 1, 2024
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 880.6500
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The ViroZap™ Indoor Air Purifier, In Duct Model 1008 is a device intended for medical purposes that is used to destroy microorganisms entrained on the filter through a photochemical reaction in the air by exposure to UV-A radiation.
Device Story
In-duct air purifier for medical facilities; utilizes photo-electrochemical (PEC) technology to destroy airborne microorganisms. System comprises pre-filter, UV-A LED lamps (320-400 nm), and catalytic filter coated with TiO2 and plasmonic nanoparticles. Installed in HVAC ducts; operates by drawing air through filter where UV-A light activates photocatalyst, inducing photochemical reaction to destroy entrained pathogens. Benefits patient/staff by reducing bioaerosol levels in clinical environments. No user-facing software or clinical decision-making output; device functions as passive infrastructure component.
Clinical Evidence
Bench testing only. Efficacy demonstrated via bioaerosol chamber studies using six microorganism species (bacteria, viruses, mold spores, endospores). Testing confirmed >4.0 log reduction in kill kinetics at various temperatures and effective reduction of airborne bioaerosols in a 9.1' x 9.1' x 7' chamber. Accelerated aging studies (equivalent to one year) confirmed performance stability.
Technological Characteristics
In-duct medical air purifier; metal housing; pre-filter; UV-A LED light source (320-400 nm); catalytic filter (TiO2 + plasmonic nanoparticles) on metal wire mesh. Power: 120/220V, up to 1.25A, 240W. Dimensions: 28.5"x12.25" or 24"x12.25". UL compliant electronics.
Indications for Use
Indicated for destruction of microorganisms (Staphylococcus epidermidis, Klebsiella aerogenes, MS2 virus, Phi-X174 virus, Aspergillus brasiliensis mold spores, and Bacillus subtilis endospores) entrained on the filter in medical facilities.
Regulatory Classification
Identification
A medical ultraviolet air purifier is a device intended for medical purposes that is used to destroy bacteria in the air by exposure to ultraviolet radiation.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which consists of the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG" in blue, with the word "ADMINISTRATION" underneath.
August 1, 2024
Applied Photonix, LLC % Jeffrey Shapiro King & Spalding LLP 1700 Pennsylvania Avenue, NW Suite 900 Washington DC 20006
Re: K241140
Trade/Device Name: ViroZap™ Indoor Air Purifier, In Duct Model 1008 Regulation Number: 21 CFR 880.6500 Regulation Name: Medical Ultraviolet Air Purifier Regulatory Class: Class II Product Code: FRA Dated: May 16, 2024 Received: May 16, 2024
Dear Jeffrery Shapiro:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
{1}------------------------------------------------
Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
# Katharine Segars -S
Katharine Segars, PhD Assistant Director DHT4C: Division of Infection Control Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known)
#### K241140
Device Name
ViroZap™ Indoor Air Purifier, In Duct Model 1008
Indications for Use (Describe)
The ViroZap™ Indoor Air Purifier, In Duct Model 1008 is a device intended for medical purposes that is used to destroy microorganisms entrained on the filter through a photochemical reaction in the air by exposure to UV-A radiation.
The ViroZap™ Indoor Air Purifier, In Duct Model 1008 has been demonstrated to destroy Staphylococcus epidermidis, Klebsiella aerogenes, MS2 - a non-enveloped ssRNA virus, Phi-X174 - a non-enveloped ssDNA virus, Aspergillus brasiliensis mold spores, and Bacillus subtilis endospores entrained on the filter under the following exposure conditions:
| Organism | Name | Average Maximum Log Reduction / Exposure Time<br>(hours) by Temperature | | |
|------------------------|-------------------------------|-------------------------------------------------------------------------|------------------|------------------|
| | | 45°F | 72°F | 110°F |
| Bacteria | Staphylococcus<br>epidermidis | 4.94 / 12 hours | 4.54 / 4 hours | 4.51 / 4 hours |
| Bacteria | Klebsiella aerogenes | 5.40 / 1 hour | 6.55 / 1 hour | 6.54 / 1 hour |
| Virus | MS2 bacteriophage | 4.32 / 12 hours | 4.51 / 4 hours | 4.58 / 4 hours |
| Virus | Phi-X174 bacteriophage | 4.39 / 12 hours | 5.24 / 4 hours | 5.14 / 4 hours |
| Mold endospore | Aspergillus brasiliensis | 4.52 / 168 hours | 5.61 / 168 hours | 5.61 / 168 hours |
| Bacterial<br>endospore | Bacillus Subtilis | 4.83 / 168 hours | 4.21 / 168 hours | 4.34 / 168 hours |
The ViroZap™ Indoor Air Purifier, In Duct Model 1008 has been demonstrated to destroy Staphylococcus epidermidis, Klebsiella aerogenes, MS2, Phi-X174, Aspergillus brasiliensis, and Bacillus subtilis in a large chamber (9.1 ft x 9.1 ft x 7 ft) at a flow rate of 500 cfm (cubic feet per minute):
| Organism | Name | Average Net Log Reduction<br>in 20 Minutes |
|---------------------|----------------------------|--------------------------------------------|
| Bacteria | Staphylococcus epidermidis | 5.17 |
| Bacteria | Klebsiella aerogenes | 5.56 |
| Virus | MS2 bacteriophage | 5.00 |
| Virus | Phi-X174 bacteriophage | 5.34 |
| Mold endospore | Aspergillus brasiliensis | 4.50 |
| Bacterial endospore | Bacillus Subtilis | 4.58 |
{3}------------------------------------------------
Prescription Use (Part 21 CFR 801 Subpart D)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
FORM FDA 3881 (8/23)
PSC Publishing Services (301) 443-6740 EF
X |Over-The-Counter Use (21 CFR 801 Subpart C)
{4}------------------------------------------------
Image /page/4/Picture/0 description: The image shows the logo for Applied Photonix. The logo consists of a row of five red squares above the words "APPLIED" and "PHOTONIX", which are stacked on top of each other. The words are in a serif font and are black.
# 510(k) SUMMARY
# ViroZap™ Indoor Air Purifier, In Duct Model 1008
# K241140
Submitter:
| Company Name: | Applied Photonix, LLC |
|-------------------------------------|-------------------------------------------------------------|
| Company Address: | 3802 Spectrum Blvd.<br>Suite 202-1<br>Tampa, FL 33612 |
| Contact Person:<br>Phone:<br>Email: | Dr. D. Yogi Goswami<br>352-871-3800<br>Y.Goswami@holtec.com |
| Date Prepared: | August 1, 2024 |
| Device: | |
| Proprietary Name: | ViroZapTM Indoor Air Purifier, In Duct Model 1008 |
| Common or Usual Name: | Air purifier ultraviolet or ultraviolet air purifier |
| Classification Name: | Medical Ultraviolet Air Purifier |
| Regulation Number: | 21 C.F.R. § 880.6500 |
| Product Code: | FRA |
| Device Class: | Class 2 |
| Category: | General Hospital |
{5}------------------------------------------------
Image /page/5/Picture/0 description: The image shows the logo for Applied Photonix. The logo consists of five red squares in a row above the words "APPLIED" and "PHOTONIX", which are stacked on top of each other. The words are in a simple, sans-serif font and are black in color.
Predicate Device: The predicate device is shown in the following table:
| Predicate Device | Transformair Indoor Air Purifier |
|---------------------|----------------------------------|
| Manufacturer | Transformair, Inc. |
| 510(k) Number | K161468 |
| Regulation Number | 21 C.F.R. § 880.6500 |
| Regulation Name | Medical Ultraviolet Air Purifier |
| Regulatory Class | II |
| Product Code | FRA |
| Review Panel | General Hospital |
| Combination Product | No |
## Intended Use / Indications for Use:
The ViroZap™ Indoor Air Purifier, In Duct Model 1008 is a device intended for medical purposes that is used to destroy microorganisms entrained on the filter through a photochemical reaction in the air by exposure to UV-A radiation.
The ViroZap™ Indoor Air Purifier, In Duct Model 1008 has been demonstrated to destroy Staphylococcus epidermidis, Klebsiella aerogenes, MS2 - a non-enveloped ssRNA virus, Phi-X174 - a non-enveloped ssDNA virus, Aspergillus brasiliensis mold spores, and Bacillus subtilis endospores entrained on the filter under the following exposure conditions:
| Organism | Name | Average Maximum Log Reduction / Exposure Time<br>(hours) by Temperature | | |
|------------------------|-------------------------------|-------------------------------------------------------------------------|------------------|------------------|
| | | 45°F | 72°F | 110°F |
| Bacteria | Staphylococcus<br>epidermidis | 4.94 / 12 hours | 4.54 / 4 hours | 4.51 / 4 hours |
| Bacteria | Klebsiella aerogenes | 5.40 / 1 hour | 6.55 / 1 hour | 6.54 / 1 hour |
| Virus | MS2 bacteriophage | 4.32 / 12 hours | 4.51 / 4 hours | 4.58 / 4 hours |
| Virus | Phi-X174 bacteriophage | 4.39 / 12 hours | 5.24 / 4 hours | 5.14 / 4 hours |
| Mold endospore | Aspergillus brasiliensis | 4.52 / 168 hours | 5.61 / 168 hours | 5.61 / 168 hours |
| Bacterial<br>endospore | Bacillus Subtilis | 4.83 / 168 hours | 4.21 / 168 hours | 4.34 / 168 hours |
{6}------------------------------------------------
Image /page/6/Picture/0 description: The image shows the logo for Applied Photonix. The logo consists of a row of five red squares above the words "APPLIED" and "PHOTONIX", which are stacked on top of each other. There is a thin black line above the word "PHOTONIX".
The ViroZap™ Indoor Air Purifier, In Duct Model 1008 has been demonstrated to destroy Staphylococus epidermidis, Klebsiella aerogenes, MS2, Phi-X174, Aspergillus brasiliensis, and Bacillus subtilis in a large chamber (9.1 ft x 9.1 ft x 7 ft) at a flow rate of 500 cfm (cubic feet per minute):
| Organism | Name | Average Net Log Reduction<br>in 20 Minutes |
|------------------------|----------------------------|--------------------------------------------|
| Bacteria | Staphylococcus epidermidis | 5.17 |
| Bacteria | Klebsiella aerogenes | 5.56 |
| Virus | MS2 bacteriophage | 5.00 |
| Virus | Phi-X174 bacteriophage | 5.34 |
| Mold endospore | Aspergillus brasiliensis | 4.50 |
| Bacterial<br>endospore | Bacillus Subtilis | 4.58 |
# Device Description:
The ViroZap™ Indoor Air Purifier, In Duct Model 1008 is a medical ultraviolet air purifier, which incorporates patented photo-electrochemical or photo-electrocatalyst (PEC) ultraviolet air purification technology that destroys bacteria in air in medical facilities. It consists of a pre-fiter, UV-A lamps, and a catalytic filter coated with a photocatalyst. The ViroZap™ has metal housing with electronic controls, and it is installed in the ducts of a building's HVAC system.
It will be available in two sizes: one size with outer dimensions of 28.5 inches by 12.25 inches, and a second size with outer dimensions of 24 inches by 12.25 inches. The ViroZap 110 device has a pre-filter and a chamber equipped with 6 low energy ultraviolet LED lights (UV-A), which emit light with wavelengths between 320-400 nm, and a catalytic filter.
## Technological Characteristics Comparison:
Compared to the predicate device (K161468), the subject device has the same intended use but differs in the catalytic filter, light source, and the organisms demonstrated to be destroyed by the device (identified in the indications for use statement). The table below provides a comparison of the subject device and predicate device.
{7}------------------------------------------------
Image /page/7/Picture/0 description: The image shows the logo for Applied Photonix. The logo consists of a row of five red squares above the words "APPLIED" and "PHOTONIX", which are stacked on top of each other. The text is in a simple, sans-serif font.
Office: (813) 825 - 1790
| Element for<br>Comparison | Predicate:<br>Transformair Indoor<br>Air Purifier; Model<br>16108<br>(K161468) | Subject Device:<br>ViroZapTM; Indoor Air<br>Purifier, In Duct Model<br>1008<br>(K241140) | Comparison |
|---------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Product type | Medical Ultraviolet Air<br>Purifier | Medical Ultraviolet Air<br>Purifier | Identical |
| Regulation<br>Number | 21 C.F.R. § 880.6500 | 21 C.F.R. § 880.6500 | Identical |
| Indications<br>for Use | The Transformair Indoor<br>Air Purifier, In Duct<br>Model 16108 is a device<br>intended for medical<br>purposes that is used to<br>destroy microorganisms<br>entrained on the filter<br>through a photochemical<br>reaction in the air by<br>exposure to UV-A<br>radiation.<br><br>Transformair Indoor Air<br>Purifier, In Duct Model<br>16108 has been<br>demonstrated to destroy<br>Staphylococcus<br>epidermidis - a Gram<br>positive bacterium,<br>Escherichia coli – a Gram<br>negative bacterium, MS2,<br>Phi-X174, Aspergillus<br>Niger mold spores and<br>Bacillus globigii<br>endospores entrained on<br>the filter of the subject<br>device under the exposure<br>conditions described<br>above. | The ViroZapTM Indoor Air<br>Purifier, In Duct Model<br>1008 is a device intended<br>for medical purposes that<br>is used to destroy<br>microorganisms entrained<br>on the filter through<br>photochemical reaction in<br>the air by exposure to UV-<br>A radiation.<br><br>The ViroZapTM Indoor Air<br>Purifier, In Duct Model<br>1008 has been<br>demonstrated to destroy<br>Staphylococcus<br>epidermidis - a Gram<br>positive bacterium,<br>Klebsiella Aerogenes – a<br>Gram negative bacterium,<br>MS2 – a non-enveloped<br>ssRNA virus, Phi-X174 –<br>a non-enveloped ssDNA<br>virus, Bacillus subtilis<br>endospores, and<br>Aspergillus brasiliensis<br>mold spores under the<br>exposure conditions<br>described above. | Similar: The proposed<br>device and the predicate<br>device have the same<br>indications for use, except<br>there are minor differences<br>in the microorganisms<br>demonstrated to be<br>destroyed by the devices.<br>Both devices are<br>demonstrated to destroy a<br>Gram positive bacterium, a<br>Gram negative bacterium,<br>a non-enveloped ssRNA<br>virus, a non-enveloped<br>ssDNA virus, mold spores,<br>and endospores. |
| Element for<br>Comparison | Predicate:<br>Transformair Indoor<br>Air Purifier; Model<br>16108<br>(K161468) | Subject Device:<br>ViroZap™; Indoor Air<br>Purifier, In Duct Model<br>1008<br>(K241140) | Comparison |
| Mechanism<br>of Action | UV light of sufficient<br>energy (UV-A) activates<br>photocatalyst that destroys<br>microorganisms entrained<br>on the filter through a<br>photochemical reaction<br>that is enhanced by a metal<br>wire mesh. | UV light of sufficient<br>energy (UV-A LED)<br>activates photocatalyst that<br>destroys microorganisms<br>entrained on the filter<br>through a photochemical<br>reaction that is enhanced<br>by a metal wire mesh. | Identical |
| Use Location | Medical Facilities | Medical Facilities | Identical |
| Installation | In-Duct | In-Duct | Identical |
| Pre-Filter | Standard UL listed air<br>filter | Standard UL listed air<br>filter | Identical |
| Photocatalyst | Catalytic filter coated with<br>TiO2 and held in a metal<br>wire mesh | Catalytic filter coated with<br>TiO2 and plasmonic<br>nanoparticles, and held in<br>a metal wire mesh | Similar: Both the<br>proposed and predicate<br>device use a catalytic filter<br>coated with TiO2. The<br>plasmonic nanoparticle<br>coating was added to the<br>catalytic filter in the<br>proposed device to<br>enhance light capture. |
| Disinfecting | Catalytic filter | Catalytic filter | Similar: Both the<br>proposed and predicate<br>device use a catalytic filter<br>coated with TiO2. The<br>plasmonic nanoparticle<br>coating was added to the<br>catalytic filter in the<br>proposed device to<br>enhance light capture. |
| Element for<br>Comparison | Predicate:<br>Transformair Indoor<br>Air Purifier; Model<br>16108<br>(K161468) | Subject Device:<br>ViroZap™; Indoor Air<br>Purifier, In Duct Model<br>1008<br>(K241140) | Comparison |
| Light source | UV-A fluorescent tube<br>(~320-400 nm) | UV-A LED tube (~320-<br>400 nm) | Similar: The predicate<br>device uses a UV-A<br>fluorescent tube, and the<br>proposed device uses a<br>UV-A LED tube with the<br>same UV-A wavelength<br>and intensity. |
| Voltage | Up to 120/220 Volt (Plugs<br>into standard outlet) | Up to 120/220 Volt (Plugs<br>into standard outlet) | Identical |
| Current | Up to 1.25 amps | Up to 1.25 amps | Identical |
| Power<br>Consumption | Up to 240 Watts | Up to 240 Watts | Identical |
| Air cleaning<br>chamber | Particulate pre-filter, UV-<br>A lamps and a catalytic<br>filter | Particulate pre-filter, UV-<br>A lamps and a catalytic<br>filter | Identical |
| Electronics | UL compliant | UL compliant | Identical |
| Outer<br>Dimensions | 28.5 inches by 21 inches<br>by 12.25 inches | 28.5 inches by 21 inches<br>by 12.25 inches, and 24<br>inches by 24 inches by<br>12.25 inches | Identical and Additional<br>Size:<br>An additional size will be<br>offered for the proposed<br>device to fit inside<br>additional building ducts. |
{8}------------------------------------------------
Image /page/8/Picture/0 description: The image shows the logo for Applied Photonix. The logo consists of five red squares arranged horizontally above the company name. The words "APPLIED" and "PHOTONIX" are stacked vertically, with "APPLIED" on top and "PHOTONIX" below.
Office: (813) 825 - 1790
{9}------------------------------------------------
Image /page/9/Picture/0 description: The image shows the logo for Applied PhotoniX. The logo consists of five red squares arranged horizontally above the words "APPLIED" and "PHOTONIX", which are stacked vertically. The text is in a simple, sans-serif font.
Office: (813) 825 - 1790
{10}------------------------------------------------
Image /page/10/Picture/0 description: The image shows the logo for Applied Photonix. The logo consists of a row of five red squares above the words "APPLIED" and "PHOTONIX", which are stacked on top of each other. The text is in a simple, sans-serif font.
# Summary of Non-Clinical Testing:
| Test | Brief Description | Acceptance Criteria | Results (Pass/Fail) |
|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------|
| Study to characterize the<br>efficacy of the subject<br>device to reduce<br>respirable bioaerosol<br>levels for six broad-<br>ranged species of<br>microorganisms from<br>room air. | Six microorganism species<br>were aerosolized in a<br>sealed bioaerosol chamber<br>containing the subject<br>device.<br><br>Bioaerosol samples were<br>taken at multiple time<br>points throughout each<br>trial, in order to quantify<br>the reduction rate<br>capability of the device. | Confirmation that the<br>effectiveness of the<br>device against ssRNA<br>virus, ssDNA virus,<br>Gram negative<br>bacteria, Gram positive<br>bacteria, bacteria<br>endospores, and mold<br>spores was equivalent<br>or superior to the<br>predicate, Transformair<br>(Κ161468). | Pass<br><br>The subject device was<br>effective in reducing a<br>broad range of airborne<br>microorganisms within<br>a short period of time. |
| In vitro study to<br>characterize the<br>temperature variant kill<br>rate kinetics for six<br>different bioaerosols<br>deposited on the surface<br>of the subject device's<br>photocatalytic coated<br>filter media after<br>exposure to UV-A light. | Samples of the subject<br>device's photocatalytic<br>coated filter media were<br>tested against six<br>aerosolized<br>microorganisms.<br><br>The samples were tested at<br>three different<br>temperatures. The kill<br>kinetic testing was<br>assessed at multiple<br>sampling points to create a<br>reduction curve with each<br>species at each<br>temperature. The trials<br>were carried out to<br>quantify at least a 4.0 log<br>reduction | Confirmation that the<br>kill kinetics of the<br>modified catalytic filter<br>was equivalent or<br>superior to the<br>predicate device<br>catalytic filter. | Pass<br><br>As with the predicate<br>device, the filter media<br>achieved more than a<br>4.0 log reduction in<br>each species at each<br>temperature. |
{11}------------------------------------------------
Image /page/11/Picture/0 description: The image shows the logo for Applied Photonix. The logo consists of six red squares arranged horizontally above the company name. The words "APPLIED" and "PHOTONIX" are stacked vertically, with a horizontal line separating them.
Office: (813) 825 - 1790
| Test | Brief Description | Acceptance Criteria | Results (Pass/Fail) |
|---------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Test to evaluate the<br>efficacy of filter media<br>following accelerated<br>aging equivalent to one<br>year. | A microorganism was<br>aerosolized into a sealed<br>environment containing the<br>subject device. Bioaerosol<br>samples were taken at<br>multiple time points<br>throughout each trial to<br>quantify the reduction rate<br>capability of the device. | Confirmation that the<br>subject device had<br>equivalent performance<br>after accelerated aging<br>equivalent to one year. | Pass<br><br>The subject device was<br>effective at reducing<br>the microorganism<br>after undergoing<br>accelerated aging. The<br>results from the aged<br>device were equivalent<br>to trials run on the<br>unaged device. |
## Clinical Testing: N/A
### Substantial Equivalence Discussion:
The proposed device and the predicate device have the same indications for use, except there are minor differences in the microorganisms demonstrated to be destroyed by the devices. The minor differences do not constitute a new intended use because both devices are demonstrated to destroy a Gram positive bacterium, a Gram negative bacterium, a non-enveloped ssRNA virus, a nonenveloped ssDNA virus, mold spores, and endospores.
The subject device has a catalytic filter with a coating consisting of TiO2 mixed with plasmonic nanoparticles, whereas the prior cleared version has only a TiO2 coating. Plasmonic nanoparticles are added to improve light capture and thereby performance of the filter. This is demonstrated by testing the subject device using the same lab and test methods that were used for the predecessor. predicate. This change improves device performance and does not raise different questions of safety or effectiveness.
The change from fluorescent to LED UV-A bulbs does not raise different questions of safety or effectiveness since the identical UV-A light (at 365 nm wavelength) is still being delivered.
The subject device is therefore substantially equivalent to the predicate device.
#### Conclusion:
Based on the intended uses, technological characteristics and non-clinical performance data, the subject device is as safe, as effective, and performs at least as safely and effectively as the legally marketed predicate device (K161468), Class II (21 CFR 880.6500), product code FRA.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.