Onycho Laser V

K241116 · Terasysd&C, Inc. · PDZ · Jul 22, 2024 · General, Plastic Surgery

Device Facts

Record IDK241116
Device NameOnycho Laser V
ApplicantTerasysd&C, Inc.
Product CodePDZ · General, Plastic Surgery
Decision DateJul 22, 2024
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 878.4810
Device ClassClass 2
AttributesTherapeutic

Indications for Use

The Onycho Laser V is indicated for use for the temporary increase of clear nail in patients with onychomycosis (e.g., dermatophytes Trichophyton rubrum and T. mentagrophytes, and/or yeasts, Candida albicans, etc.).

Device Story

Onycho Laser V is a floor-positioned laser device for treating onychomycosis. It emits light at 405 nm and 635 nm wavelengths in a constant wave, electronically scanned line pattern. Operated via a 7.0-inch TFT touch screen, the device allows clinicians to set treatment times (0-12 minutes). The laser energy targets the nail area to promote a temporary increase in clear nail. It is intended for clinical use by healthcare professionals. The device provides a therapeutic light-based intervention; clinical decision-making is supported by the device's ability to deliver controlled laser energy to the affected nail, potentially benefiting patients by improving the appearance of the nail through the temporary clearance of fungal infection.

Clinical Evidence

Bench testing only. The device conforms to IEC 60601-1, IEC 60601-1-2, IEC 60601-2-22, and IEC 60825-1 standards. Biocompatibility testing was performed per ISO 10993-5, ISO 10993-10, and ISO 10993-23. No clinical data provided.

Technological Characteristics

Laser surgical instrument; dual-wavelength (405nm/635nm) laser; constant wave output; electronically scanned line pattern; 7.0-inch TFT touch screen interface; 17kg weight. Complies with IEC 60601-1 (v3.2), IEC 60601-1-2 (v4.1), IEC 60601-2-22 (v4.0), and IEC 60825-1 (v3.0). Biocompatibility per ISO 10993-5, -10, and -23.

Indications for Use

Indicated for patients with onychomycosis (e.g., dermatophytes Trichophyton rubrum, T. mentagrophytes, and/or yeasts like Candida albicans) to achieve a temporary increase of clear nail.

Regulatory Classification

Identification

(1) A carbon dioxide laser for use in general surgery and in dermatology is a laser device intended to cut, destroy, or remove tissue by light energy emitted by carbon dioxide.(2) An argon laser for use in dermatology is a laser device intended to destroy or coagulate tissue by light energy emitted by argon.

Predicate Devices

Submission Summary (Full Text)

{0}------------------------------------------------ Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: on the left, there is a seal with an eagle emblem, and on the right, there is the text "FDA U.S. FOOD & DRUG ADMINISTRATION" in blue. The text is arranged in three lines, with "FDA" being the largest and boldest, followed by "U.S. FOOD & DRUG" and then "ADMINISTRATION" in a smaller font size. July 22, 2024 TerasysD&C Inc. % Kim Jonghyeon CEO GMSC Co., Ltd. B 612, 66, Cheongcho-ro, Deokyang-gu, Goyang-si Gyeonggi-do, Korea, South Re: K24116 Trade/Device Name: Onycho Laser V Regulation Number: 21 CFR 878.4810 Regulation Name: Laser Surgical Instrument For Use In General And Plastic Surgery And In Dermatology Regulatory Class: Class II Product Code: PDZ Dated: April 8, 2024 Received: April 23, 2024 Dear Kim Jonghyeon: We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of {1}------------------------------------------------ Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download). Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181). Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems. For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). {2}------------------------------------------------ Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/deviceadvice-comprehensive-regulatory-assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100). Sincerely. Digitally signed by TANISHA TANISHA L. HITHE -S L. HITHE -○ Date: 2024.07.22 14:49:18 -04'00' Tanisha Hithe Assistant Director DHT4A: Division of General Surgery Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health Enclosure {3}------------------------------------------------ # Indications for Use 510(k) Number (if known) K241116 Device Name Onycho Laser V Indications for Use (Describe) The Onycho Laser V is indicated for use for the temporary increase of clear nail in patients with onychomycosis (e.g., dermatophytes Trichophyton rubrum and T. mentagrophytes, and/or yeasts, Candida albicans, etc.). Type of Use (Select one or both, as applicable) | <div><span style="font-size: 10pt;"> <svg height="15" width="15"> <rect height="15" style="fill:none;stroke:#000000;stroke-width:1" width="15"></rect> <path d="M2 2 L13 13 M13 2 L2 13" stroke="#000000" stroke-width="1"></path> </svg> Prescription Use (Part 21 CFR 801 Subpart D)</span></div> | |--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------| | <div><span style="font-size: 10pt;"> <svg height="15" width="15"> <rect height="15" style="fill:none;stroke:#000000;stroke-width:1" width="15"></rect> </svg> Over-The-Counter Use (21 CFR 801 Subpart C)</span></div> | #### CONTINUE ON A SEPARATE PAGE IF NEEDED. This section applies only to requirements of the Paperwork Reduction Act of 1995. #### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: > Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." {4}------------------------------------------------ uired by 21 # 1. Date Prepared [21 CFR 807.92(a)(1)] Apr.06, 2024 ### 2. Submitter's Information [21 CFR 807.92(a)(1)] - Name of Manufacturer: TerasysD&C. Inc. ● - Address: Unit 1101~1108, Building N, 249 Sunhwagung-ro, Namyangjusi, Gyeonggi-do, Republic of Korea - Contact Name: SangHyeop Han - +82 2-2293-6667 ● Telephone No.: - Email Address: qc@terasysdnc.com - TBD 0 Registration No.: ## 3. Identification of Proposed Device(s) [21 CFR 807.92(a)(2)] | 510(k) Number | K241116 | |-----------------------------|----------------------------------------------------------------------------------------| | Trade/Device/Model Name | Onycho Laser V | | Product Name | Onycho Laser V | | Common Name | Lasers For Temporary Increase of Clear Nail in Patients with<br>Onychomycosis | | Regulation Name | Laser surgical instrument for use in general and plastic<br>surgery and in dermatology | | Regulation Number | 21 CFR 878.4810 | | Classification Product Code | PDZ | | Device Class | II | | 510(k) Review Panel | General and Plastic Surgery | {5}------------------------------------------------ ## 4. Identification of Predicate Device(s) [21 CFR 807.92(a)(3)[ The identified predicate device within this submission is shown as follow; | 510(k) Number | K221363 | |-----------------------------|----------------------------------------------------------------------------------------| | Trade/Device/Model Name | AF Laser | | Product Name | N/A | | Common Name | Lasers For Temporary Increase of Clear Nail in Patients<br>with Onychomycosis | | Requlation Name | Laser surgical instrument for use in general and plastic<br>surgery and in dermatology | | Regulation Number | 21 CFR 878.4810 | | Classification Product Code | PDZ | | Device Class | II | | 510(k) Review Panel | General and Plastic Surgery | These predicate devices have not been subject to a design-related recall. # 5. Description of the Device [21 CFR 807.92(a)(4)] The `Onycho Laser V' is a compact device that is positioned on the floor and is operated via a 7.0 inch TFT touch screen for ease of use. The device incorporates lasers and outputs light at two wavelengths, 405 nm and 635 nm, in order to achieve its therapeutic effect. #### 6. Indications for Use [21 CFR 807.92(a)(5)] The 'Onycho Laser V' is indicated for use for the temporary increase of clear nail in patients with onychomycosis (e.g., dermatophytes Trichophyton rubrum and T. mentagrophytes, and/or yeasts Candida albicans, etc.). {6}------------------------------------------------ # 7. Technological Comparison [21 CFR 807.92(a)(6)] Provided below is a table that compares technological characteristics of the Onycho Laser V and the predicate device. | | Proposed Device | Predicate Device | Note | |---------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------| | K Number | K241116 | K221363 | - | | Manufacturer | TerasysD&C Inc. | ShenB Co Ltd. | - | | Trade Name | Onycho Laser V | AF Laser | - | | Product Name | Onycho Laser V | - | - | | Product Code | PDZ | PDZ | Identical | | Regulation Number | 21 CFR 878.4810 | 21 CFR 878.4810 | Identical | | 510(k) Review Panel | General and Plastic Surgery | General and Plastic Surgery | Identical | | Indications for Use | The 'Onycho Laser V' is indicated for use for the temporary increase of clear nail in patients with onychomycosis (e.g., dermatophytes Trichophyton rubrum and T. mentagrophytes, and/or yeasts Candida albicans, etc.). | The AF Laser device is indicated for use for the temporary increase of clear nail in patients with onychomycosis (e.g., dermatophytes Trichophyton rubrum and T. mentagrophytes, and/or yeasts Candida albicans, etc.). | Identical | | Laser<br>Wavelength | 405nm/635nm (±10%) | 405nm/635nm (±10%) | Identical | | Output Energy | 405nm: 23mW ± 1.85mW<br>635nm: 17mW ± 1.35mW | 405nm: 23mW ± 1.85mW<br>635nm: 17mW ± 1.35mW | Identical | | Output area | Line pattern electronically scanned over area of treatment | Line pattern electronically scanned over area of treatment | Both devices scan the output beams over the areas being treated. | | Output Type | Constant Wave | Constant Wave | Identical | | Operating Time | 0-12 minutes with 1minute increment | 0-12 minutes (±5%) with 1minute increment | Identical | | Dimension | 530mm(W) × 410 mm(L) × | 424mm(W) × 308mm(L) × | Similar | | | Proposed Device | Predicate Device | Note | | | 445mm(H) | 352mm(H) | | | Weight | 17kg (1unit) | 17.5kg | Similar | | Screen | TFT touch 7.0 inch(Diagonal) | LCD Touch Screen | Similar | [Table 3. Comparison of Proposed Device to Predicate Devices] {7}------------------------------------------------ The Onycho Laser V device uses the same basic laser technology as that used by the predicate device. The technological parameters of the Onycho Laser V device are either identical or similar to those of the predicate device, and the differences do not raise new types of questions regarding the safety and effectiveness for the proposed indications for use #### 8. Non-Clinical Test Summary The `Onycho Laser V conforms to voluntary standards for electrical safety, electromagnetic compatibility, and other performance. Test results from the following performance standards were provided in support of the safety and effectiveness of the Onycho Laser V device | Standards<br>No. | Standards<br>Organization | Standard Title | Version | Publication<br>Year | |------------------|---------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------|---------------------| | 60601-1 | IEC | Medical electrical equipment Part 1:<br>General<br>Requirements for basic<br>safety and essential performance | 3.2 | 2020 | | 60601-1-2 | IEC | Medical electrical equipment - Part 1-<br>2: General requirements for basic<br>safety and essential performance -<br>Collateral Standard: Electromagnetic<br>disturbances - Requirements and tests | 4.1 | 2020 | | 60601-2-22 | IEC | Medical electrical equipment - Part 2-<br>22: Particular requirements for basic<br>safety and essential performance of<br>surgical, cosmetic, therapeutic and<br>diagnostic laser equipment | 4.0 | 2019 | 1) The Yonycho Laser V' conforms to the following performance standards: {8}------------------------------------------------ | Standards<br>No. | Standards<br>Organization | Standard Title | Version | Publication<br>Year | |------------------|---------------------------|------------------------------------------------------------------------------------|---------|---------------------| | 60825-1 | IEC | Safety of laser products - Part 1:<br>Equipment classification and<br>requirements | 3.0 | 2014 | ### 2) Biocompatibility The `Onycho Laser V′ conforms to the biocompatibility requirements established by the standards. | Standards<br>No. | Standards<br>Organization | Standard Title | Version | Publication<br>Year | |------------------|---------------------------|------------------------------------------------------------------------------------|---------|---------------------| | 10993-5 | ISO | Biological evaluation of medical devices - Part 5: Tests for in vitro cytotoxicity | 3.0 | 2009 | | 10993-10 | ISO | Biological evaluation of medical devices - Part 10: Tests for skin sensitization | 4.0 | 2021 | | 10993-23 | ISO | Biological evaluation of medical devices - Part 23: Tests for irritation | 1.0 | 2021 | {9}------------------------------------------------ #### 3) Software Validation The software was designed and developed according to a software development process and was verified and validated. Software information is provided in accordance with FDA guidance: - The content of premarket submissions for software contained in medical devices, on June 14, 2023 # 9. Conclusion [21 CFR 807.92(b)(3)] In accordance with the Federal Food & Drug and cosmetic Act, 21 CFR Part 807, and based on the information provided in this premarket notification, it is concluded that the `Onycho Laser V′ is substantially equivalent to the predicate device K221363
Innolitics
Guidance Recalls

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...