VIVIX-M detectors are flat panel detectors used in mammographic applications to acquire digital images for screening and diagnosis.
Device Story
VIVIX-M is a series of flat panel detectors (models FXMD-2430S, FXMD-1008S) for mammography. The device intercepts X-ray photons; a scintillator (thallium-doped cesium iodide) converts these into visible light; an amorphous silicon (a-Si) array converts light into electrical signals. These signals are digitized and processed by VXvue Mammo software. The system integrates with an operating PC and X-ray generator. Used in clinical settings by healthcare professionals to acquire, process, and transmit digital mammographic images for diagnostic review. The output is a digital image displayed on a monitor, facilitating clinical assessment of breast tissue. Benefits include high-resolution digital imaging for screening and diagnosis, replacing traditional film-based systems.
Clinical Evidence
Clinical image evaluation conducted per FDA 'Class II Special Controls Guidance: Full-Field Digital Mammography System'. Study confirmed VIVIX-M provides images of equivalent diagnostic capability to the predicate device (RSM 2430C).
Indicated for use in screening and diagnostic mammography for patients requiring digital imaging.
Regulatory Classification
Identification
A full-field digital mammography system is a device intended to produce planar digital x-ray images of the entire breast. This generic type of device may include digital mammography acquisition software, full-field digital image receptor, acquisition workstation, automatic exposure control, image processing and reconstruction programs, patient and equipment supports, component parts, and accessories.
Special Controls
*Classification.* Class II (special controls). The special control for the device is FDA's guidance document entitled “Class II Special Controls Guidance Document: Full-Field Digital Mammography System.”*See* § 892.1(e) for the availability of this guidance document.
{1}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food & Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
January 16, 2025
Vieworks Co., Ltd. % Priscilla Chung Regulatory Affairs Consultant LK Consulting Group USA Inc. 18881 Von Karman Ave, STE 160 Irvine. California 92612
Re: K241113
Trade/Device Name: VIVIX-M Regulation Number: 21 CFR 892.1715 Regulation Name: Full-Field Digital Mammography System Regulatory Class: Class II Product Code: MUE Dated: October 18, 2024 Received: October 18, 2024
Dear Priscilla Chung:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{2}------------------------------------------------
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30. Design controls; 21 CFR 820.90. Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-device-advicecomprehensive-regulatory-assistance/unique-device-identification-system-udi-system.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatory
{3}------------------------------------------------
assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
D. Ryk
Daniel M. Krainak, Ph.D. Assistant Director DHT8C: Division of Radiological Imaging and Radiation Therapy Devices OHT8: Office of Radiological Health Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{4}------------------------------------------------
## Indications for Use
510(k) Number (if known) K241113
Device Name VIVIX-M
Indications for Use (Describe)
VIVIX-M detectors are flat panel detectors used in mammographic applications to acquire digital images for screening and diagnosis.
| Type of Use (Select one or both, as applicable) |
|-------------------------------------------------|
|-------------------------------------------------|
X Prescription Use (Part 21 CFR 801 Subpart D)
| Over-The-Counter Use (21 CFR 801 Subpart C)
### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{5}------------------------------------------------
# 510(k) Summary (K241113)
This 510(k) summary information is prepared in accordance with 21 CFR807.92
### 1. Date Prepared [21 CFR 807.92(a) (1)]
1/15/2025
### 2. Submitter's Information [21 CFR 807.92(a) (1)]
| Name of Sponsor: | Vieworks Co., Ltd. |
|-----------------------|---------------------------------------------------------------------------------------------------------|
| Address: | (Gwanyang-dong) 41-3, Burim-ro 170beon-gil,Dongan-gu<br>Anyang-si, Gyeonggi-do, 14055 Republic of Korea |
| Contact Name: | Oh, Kevin / Regulatory Affairs Assistant Manager |
| Contact Number | +82-70-4466-3220 |
| Registration Number: | 3006013411, 3017585294 |
| Name of Manufacturer: | Same as Sponsor |
#### 3. Trade Name, Common Name, Classification [21 CFR 807.92(a) (2)]
| Trade Name: | VIVIX-M |
|----------------------------|-------------------------------------------------|
| Model Name: | FXMD-2430S, FXMD-1008S |
| Common Name: | Digital Mammography System |
| Classification Name: | Full Field Digital, System, X-Ray, Mammographic |
| Classification Panel: | Radiology |
| Classification Regulation: | 21 CFR 892.1715 |
| Product Code: | MUE |
| Device Class: | 2 |
#### 4. Identification of Predicate Device(s) [21 CFR 807.92(a) (3)]
| <b>510(k) Number:</b> K170930 |
|-----------------------------------------------------------------------------|
| <b>Applicant:</b> DRTECT Corporation |
| <b>Trade Name:</b> RSM 2430C |
| <b>Common Name:</b> Digital Mammography System |
| <b>Classification Name:</b> Full Field Digital, System, X-Ray, Mammographic |
| <b>Classification Panel:</b> Radiology |
| <b>Classification Regulation:</b> 21 CFR 892. 1715 |
| <b>Product Code:</b> MUE |
| <b>Device Class:</b> 2 |
| <b>Decision Date:</b> 04/17/2017 |
| <b>Type:</b> Traditional |
{6}------------------------------------------------
### 5. Description of the Device [21 CFR 807.92(a) (4)]
#### General Description O
VIVIX-M, a series of flat panel detectors models named; FXMD-2430S, with imaging areas of 24cm x 30cm, 18cm x 24cm, respectively. The device intercepts x-ray photons and the scintillator emits visible spectrum photons that illuminate an array of photo (a-SI)-detectors that create electrical signals. After the electrical signals are generated, it is converted to digital value, and the Software acquires and processes the data values from the detector. The resulting digital images will be displayed on monitors. These devices should be integrated with an operating PC and an X-Ray generator. It can be utilized to digitalize x-ray images and transfer for radiography diagnostic.
VXvue Mammo is a digital X-ray imaging software designed for mammography applications. It acquires, processes, and transmits digital images from VIVIX-M detectors (FXMD-1008S and FXMD-2430S) while ensuring compliance with DICOM standards for seamless integration with PACS and other medical systems.
### 6. Indications for Use [21 CFR 807.92(a)(5)]
VIVIX-M detectors are flat panel detectors used in mammographic applications to acquire digital images for screening and diagnosis.
#### 7. Technological Characteristics [21 CFR 807.92(a) (6)]
Compared with the predicate, devices show the technological characteristics of the proposed VIVIX-M device to be substantially equivalent to the predicate device. The proposed devices are functionally identical to the predicate device.
| Contents | Requirements |
|------------|---------------------------------|
| Tube Type | Rotating anode or Stationary |
| Target | Molybdenum, Rhodium or Tungsten |
| Bucky Size | 18cm x 24cm or 24cm x 30cm |
| Grid | Moving |
| Filter | Mo, Rh |
| kVp | 20 ~ 35 (Mo, W), 28 ~ 39 (Rh) |
| mAs | 3 ~ 500 |
#### 8. Integrated System Requirements
- Software Requirement
: These detectors are compatible with VXvue Mammo (V 1.0)
### 9. Substantial Equivalence [21 CFR 807.92(b)(1) and 807.92]
When compared to the predicate device (K170930), the VIVIX-M presented in this submission is substantially equivalent:
- Intended Use
- Technological characteristics
- Operating principle
- Design features
- Communication Method
- Scintillator Materials
- Resolution
{7}------------------------------------------------
There is similar performance as follow.
- Performance (MTF)
- Performance (DQE)
There are no significant difference between the VIVIX-M and the predicate device that would adversely affect the use of the product. It is substantially equivalent to the predicate device in design, function, materials, operational principles and intended use.
| Parameter | Predicate Device | Subject Device | Equivalence |
|--------------------------------------|-------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------|
| 510(k) Number | K170930 | K241113 | - |
| Manufacturer | DRTECH Corporation | Vieworks Co., Ltd. | - |
| Model Name | RSM 2430C | FXMD-2430S | - |
| Classification Name | Full Field Digital, System, X-ray, Mammographic | | Equivalent |
| Classification Panel | Radiology | | Equivalent |
| Classification<br>Regulation | 21 CFR 892.1715 | | Equivalent |
| Product Code | MUE | | Equivalent |
| Device Class | 2 | | Equivalent |
| Indications for Use | The RSM 2430C is a detector<br>indicated for use in screening<br>and diagnostic<br>mammography. | VIVIX-M detectors are flat<br>panel detectors used in<br>mammographic applications<br>to acquire digital images for<br>screening and diagnosis. | Substantially<br>Equivalent |
| Scintillating Material | Cesium lodide | Thallium doped Cesium<br>lodide | Substantially<br>Equivalent |
| Pixel Pitch (µm) | 76 | 75 | Substantially<br>Equivalent |
| Pixel Matrix | 3072x3840 | 3072×3840 | Equivalent |
| Dimensions (H x W x D) | 327.15mm x 253.75mm x<br>14.2mm | 254.6 mm × 327.6 mm ×<br>15.0 mm | Substantially<br>Equivalent |
| Weight (kg) | 1.5 | FXMD-1008S: 1.0<br>FXMD-2430S: 1.2 | - |
| Rated Power Supply | DC +12V | DC +24V, Max. 0.75A | - |
| Power Consumption | 30W | Max. 18W | - |
| Data Output and Software<br>Features | Rconsole 1 | VXvue Mammo | - |
| MTF | 70% at 2lp/mm<br>30% at 5lp/mm | 54.9% at 3lp/mm<br>33.1% at 5lp/mm | Substantially<br>Equivalent |
| DQE | 43% at 2lp/mm<br>30% at 5lp/mm | 59.0% at 3lp/mm<br>42.5% at 5lp/mm | Substantially<br>Equivalent |
{8}------------------------------------------------
| Parameter | Predicate Device | Subject Device | Equivalence |
|--------------------------------------|-------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------|
| 510(k) Number | K170930 | K241113 | - |
| Manufacturer | DRTECH Corporation | Vieworks Co., Ltd. | - |
| Model Name | RSM 2430C | FXMD-1008S | - |
| Common Name | Digital Flat Panel X-ray Detector | | Equivalent |
| Classification Name | Solid State X-Ray Imager (Flat Panel/Digital Imager) | | Equivalent |
| Classification Panel | Radiology | | Equivalent |
| Classification<br>Regulation | 21 CFR 892.1715 | | Equivalent |
| Product Code | MUE | | Equivalent |
| Device Class | 2 | | Equivalent |
| Indications for Use | The RSM 2430C is a detector<br>indicated for use in<br>screening and diagnostic<br>mammography. | VIVIX-M detectors are flat<br>panel detectors used in<br>mammographic applications<br>to acquire digital images for<br>screening and diagnosis. | Substantially<br>Equivalent |
| Scintillating Material | Cesium lodide | Thallium doped Cesium<br>lodide | Substantially<br>Equivalent |
| Pixel Pitch (μm) | 76 | 75 | Substantially<br>Equivalent |
| Pixel Matrix | 3072x3840 | 3072×2304 | Substantially<br>Equivalent |
| Dimensions (H x W x D) | 327.15mm x 253.75mm x<br>14.2mm | 254.6 mm × 327.6 mm ×<br>15.0 mm | Substantially<br>Equivalent |
| Weight (kg) | 1.5 | FXMD-1008S: 1.0<br>FXMD-2430S: 1.2 | - |
| Rated Power Supply | DC +12V | DC +24V, Max. 0.75A | - |
| Data Output and Software<br>Features | Rconsole 1 | VXvue Mammo | - |
| MTF | 70% at 2lp/mm<br>30% at 5lp/mm | 52.7% at 3lp/mm<br>32.7% at 5lp/mm | Substantially<br>Equivalent |
| DQE | 43% at 2lp/mm<br>30% at 5lp/mm | 65.8% at 3lp/mm<br>46.9% at 5lp/mm | Substantially<br>Equivalent |
{9}------------------------------------------------
### 10. Summary of Non-Clinical Data
This section provides a high-level summary of the non-clinical laboratory testing conducted to demonstrate substantial equivalence between the subject devices (VIVIX-M models FXMD-1008S and FXMD-2430S) and the predicate device (K170930). The testing complies with FDA-recognized standards and the recommendations outlined in the Full Field Digital Mammography System - Class II Special Controls Guidance for Industry and FDA Staff (March 27, 2012).
### Testing Summary
The following non-clinical tests were performed to evaluate the performance of the subject devices:
#### 1) Modulation Transfer Function (MTF):
- Purpose: To assess spatial resolution. ●
- . Protocol: Conducted per IEC 62220-1-2:2007 standards using a tungsten edge phantom.
- Results: Both models demonstrated spatial resolution equivalent to the predicate device, ● with MTF values exceeding specified thresholds.
### 2) Detective Quantum Efficiency (DQE):
- Purpose: To evaluate signal-to-noise ratio transfer efficiency. ●
- Protocol: Measurements followed IEC 62220-1-2:2007 guidelines using standard clinical exposure conditions.
- . Results: DQE values at all spatial frequencies were comparable to the predicate device, confirming equivalent imaging performance.
#### Noise Power Spectrum (NPS): 3)
- Purpose: To analyze noise characteristics.
- Protocol: Conducted under seven exposure levels as per IEC 62220-1-2.
- Results: NPS results confirmed consistent noise performance across tested exposure levels.
#### 4) Dynamic Range Testing:
- Purpose: To determine the range of exposure levels without saturation artifacts. ●
- Protocol: Based on sensitometric response data from IEC 62220-1-2.
- Results: Both models exhibited a wide dynamic range suitable for mammographic imaging.
#### Image Erasure and Fading Test (Ghosting): ട്)
- Purpose: To assess residual image effects after repeated exposures.
- Protocol: Evaluated using a Pb phantom under specified conditions.
- . Results: No significant ghosting or residual artifacts were observed.
#### ର) Automatic Exposure Control (AEC) Performance:
- Purpose: To verify compatibility with AEC systems for varying breast thicknesses.
- . Protocol: Tested across PMMA thicknesses from 20mm to 70mm in standard and magnification modes.
- Results: AEC performance met manufacturer-defined limits, ensuring consistent image quality.
#### 7) Phantom Testing (ACR, TE, CDMAM Phantoms):
- Purpose: To evaluate image quality metrics such as detectability of specks, fibers, and masses.
- . Protocols:
- ACR Phantom tested for visibility of structures under automatic exposure settings. .
- TE Phantom assessed line pair resolution and mass detection. ●
- CDMAM Phantom analyzed contrast-detail thresholds.
- . Results: All tests demonstrated diagnostic image quality equivalent to or better than the predicate device.
- 8) Patient Radiation Dose Measurement (Average Glandular Dose):
- Purpose: To ensure compliance with dose safety limits while maintaining image quality. ●
- Protocol: Conducted under MQSA-compliant conditions simulating compressed breast thicknesses (30%, 50%, and 70% glandular tissue compositions).
{10}------------------------------------------------
- Results: Average glandular dose was within FDA-recommended limits for all scenarios.
#### Standards and Guidance Referenced
The following standards and guidance documents were utilized:
- IEC 62220-1-2:2007 (Medical Electrical Equipment – Characteristics of Digital X-Ray Imaging Devices).
- Full Field Digital Mammography System – Class II Special Controls Guidance for Industry and FDA Staff, March 27, 2012.
- MQSA requirements for average glandular dose limits.
#### Conclusion
The non-clinical testing results demonstrate that the VIVIX-M models FXMD-1008S and FXMD-2430S are substantially equivalent to the predicate device in terms of safety, effectiveness, and performance. The devices comply with applicable FDA-recognized consensus standards, ensuring suitability for their intended use in mammographic imaging.
### 11. Summary of Clinical Data
A clinical image evaluation according to CDRH's 'Guidance for Industry and FDA Staff - Class II Special Controls Guidance Document: Full-Field Digital Mammography System' was conducted, and the study confirmed that the new x-ray detectors VIVIX-M provide images of equivalent diagnostic capability to the predicate device, RSM 2430C and its results demonstrate substantial equivalence.
### 12. Conclusion [21 CFR 807.92(b) (3)]
The VIVIX-M Digital X-ray detectors are substantially equivalent to the currently marketed and predicate devices (K170930) in terms of design features, fundamental scientific technology, indications for use, and safety and effectiveness.
Additionally, Substantial equivalence was demonstrated through the non-clinical performance, which complied with the requirements specified in the international and FDA-recognized consensus standards, IEC60601-1, IEC 60601-1-2, and the clinical test, which complied with the requirements specified in the CDRH's 'Class II Special Controls Guidance Document: Full-Field Digital Mammography System'.
The results of these tests demonstrate that VIVIX-M Digital X-ray detectors meets the acceptance criteria and is adequate for this intended use. The comparison of technological characteristics, non-clinical performance data, safety testing, and clinical image concurrence data demonstrates that the device is as safe, as effective, and performance is substantially equivalent to the predicate device as well.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.