K241091 · Inogen, Inc. · SDT · Dec 23, 2024 · Anesthesiology
Device Facts
Record ID
K241091
Device Name
SIMEOX 200 Airway Clearance Device
Applicant
Inogen, Inc.
Product Code
SDT · Anesthesiology
Decision Date
Dec 23, 2024
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 868.5665
Device Class
Class 2
Attributes
Therapeutic, Pediatric
Indications for Use
The SIMEOX 200 airway clearance device is intended to promote airway clearance and to improve bronchial drainage by enhancing mobilization of bronchial secretions via high frequency oscillatory vibrations and intermittent negative pressure to the airway during exhalation. The device is intended to be prescribed for use in patients capable of independently generating cough. SIMEOX 200 is intended for use in the home, hospital, and other healthcare institutions by individuals weighing at least 23 Kg.
Device Story
Portable airway clearance device; consists of generator unit and single-patient-use expiratory kit (mouthpiece, tubing, filter). Patient inhales ambient air; exhales into mouthpiece. Device applies high-frequency negative pressure oscillations (6 Hz or 12 Hz) during exhalation to loosen mucus and increase secretion mobility. Used in home or clinical settings by patients weighing ≥23 kg. Features manual or 'DETECT' trigger (starts upon sensing exhalation). Includes RFID-based usage tracking for expiratory kits. Output is pneumatic vibration; assists patient in clearing secretions via independent cough. Benefits include improved bronchial drainage and reduced mucus viscosity.
Clinical Evidence
Bench testing only. Includes performance verification, software verification/validation, biocompatibility (ISO 10993-1), human factors validation study (simulated use in adults and child/caregiver dyads), and cybersecurity/EMC/electrical/mechanical safety testing.
Technological Characteristics
ABS/PC housing; stainless-steel grid inlet. Powered by AC or Li-ion battery. Expiratory kits use PVC tubing, SBC filters, polypropylene mouthpieces. RFID-based usage lockout. Oscillation via pneumatic negative pressure pulses (0 to -60 cm H2O). Fixed frequencies (12 Hz/6 Hz).
Indications for Use
Indicated for patients weighing at least 23 kg with difficulty clearing bronchial secretions who are capable of independently generating a cough.
Regulatory Classification
Identification
A powered percussor is a device that is intended to transmit vibration through a patient's chest wall to aid in freeing mucus deposits in the lung in order to improve bronchial drainage and that may be powered by electricity or compressed gas.
{0}------------------------------------------------
December 23, 2024
Image /page/0/Picture/1 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, with the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG" in blue, and the word "ADMINISTRATION" in smaller blue letters below that.
Inogen, Inc. % Michael Nilo President and Principal Consultant Nilo Medical Consulting Group 3706 Butler Street #313 Pittsburgh, Pennsylvania 15201
Re: K241091
Trade/Device Name: SIMEOX 200 Airway Clearance Device Regulation Number: 21 CFR 868.5665 Regulation Name: Powered Percussor Regulatory Class: Class II Product Code: SDT Dated: November 27, 2024 Received: November 27, 2024
Dear Michael Nilo:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}------------------------------------------------
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30. Design controls; 21 CFR 820.90. Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-device-advicecomprehensive-regulatory-assistance/unique-device-identification-system-udi-system.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatory
{2}------------------------------------------------
assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
# Ethan L. Nyberg -S
Ethan Nyberg, Ph.D. Assistant Director, Respiratory Devices Team DHT1C: Division of Sleep Disordered Breathing, Respiratory and Anesthesia Devices OHT1: Office of Ophthalmic, Anesthesia, Respiratory, ENT and Dental Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{3}------------------------------------------------
# Indications for Use
510(k) Number (if known) K241091
Device Name SIMEOX 200 Airway Clearance Device
Indications for Use (Describe)
The SIMEOX 200 airway clearance device is intended to promote airway clearance and to improve bronchial drainage by enhancing mobilization of bronchial secretions via high frequency oscillatory vibrations and intermittent negative pressure to the airway during exhalation. The device is intended to be prescribed for use in patients capable of independently generating cough.
SIMEOX 200 is intended for use in the home, hospital, and other healthcare institutions by individuals weighing at least 23 Kg.
| Type of Use (Select one or both, as applicable) | |
|------------------------------------------------------------------------------------------|--|
| <span style="font-size: 10pt;"> Prescription Use (Part 21 CFR 801 Subpart D) </span> | |
| <span style="font-size: 10pt;">Over-The-Counter Use (21 CFR 801 Subpart C)</span> | |
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{4}------------------------------------------------
#### 510(K) SUMMARY
| DATE OF PREPARATION | December 20, 2024 |
|-------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| APPLICANT | Inogen, Inc.<br>859 Ward Drive<br>Suite 200<br>Goleta, California 93111 |
| OFFICIAL CORRESPONDENT | Michael Nilo, MS BME, President<br>Nilo Medical Consulting Group<br>3706 Butler Street, #313<br>Pittsburgh, PA 15201<br>Phone: (717) 421-4396<br>E-mail: michael.nilo@nilomed.com |
| DEVICE NAME | SIMEOX 200 Airway Clearance Device |
| DEVICE CLASSIFICATION | Common Name: Powered percussor<br>Regulation: 21 CFR 868.5665<br>Class II |
| PRIMARY PREDICATE DEVICE | AirPhysio Positive Expiratory Pressure (PEP) Device<br>(K203209) |
| SECONDARY PREDICATE<br>DEVICE | Philips Respironics SimplyClear Device (K122111) |
#### DEVICE DESCRIPTION
The SIMEOX 200 Airway Clearance Device consists of a portable generator and a single patient use expiratory kit. The patient inhales from the ambient environment, naturally and independent of the device, and then exhales into the device mouthpiece. During the patient's exhalation, the SIMEOX 200 generates a pneumatic vibration throughout the airway. This vibration, applied at 6 Hz or 12 Hz via negative pressure pulses, disseminates throughout the bronchial tree in order to decrease mucus viscosity and increase secretion mobility. During or after therapy, the patient can clear the secretions by coughing independently.
{5}------------------------------------------------
The SIMEOX 200 device is housed within an acrylonitrile butadiene styrene polycarbonate (ABS/PC) unit, with control buttons located on the front and top panels. The "Go" button, which is located on the front panel of the device, is used to start the pneumatic vibration signal of SIMEOX 200. Beneath the "Go" button is a connection inlet for the expiratory kit and mouthpiece. This inlet is equipped with a stainless-steel grid to prevent large particles from penetrating inside the air circuit. The remote controller has a redundant "Go" button for ease of use. The device also has an external power supply with an AC cord, and an optional Lithium-Ion battery.
Expiratory kits, which consist of the mouthpiece and respiratory tubing, are available in two different versions and are supplied separately from the SIMEOX 200 device. TUB10 expiratory kits come with one polyvinyl chloride (PVC) respiratory tube, one styrene-butadiene copolymer (SBC) filter with a polypropylene membrane, and 10 single-use polypropylene mouthpieces, each intended to be disposed of after each use. Similarly, TUB25 expiratory kits come with one PVC respiratory tube, one SBC filter with a polypropylene membrane, and 3 reusable polypropylene mouthpieces, each intended to be used up to ten times. The filters supplied in both kits each come with an RFID tag, which tracks the number of uses of the filter and tube. The RFID tag ensures that the device will not operate upon connection to an expiratory kit that has expended its maximal number of uses (10 maximal uses for TUB10, and 25 maximal uses for TUB25).
## INDICATIONS FOR USE
The SIMEOX 200 airway clearance device is intended to promote airway clearance and to improve bronchial drainage by enhancing mobilization of bronchial secretions via high frequency oscillatory vibrations and intermittent negative pressure to the airway during exhalation. The device is intended to be prescribed for use in patients capable of independently generating cough.
SIMEOX 200 is intended for use in the home, hospital, and other healthcare institutions by individuals weighing at least 23 Kg.
## PRINCIPLES OF OPERATION / MECHANISM OF ACTION
The SIMEOX 200 Airway Clearance Device promotes airway clearance by loosening and mobilizing secretions. Therapy is delivered via exhalation into the device plastic mouthpiece. During exhalation, a high frequency negative pressure oscillation (pneumatic vibration) is applied to the airways which loosens mucus from the bronchial walls, increasing its mobilization.
Airway oscillation (vibration) during exhalation is an established mechanism of action used to loosen mucus and mobilize secretions from the airways for expectoration.
{6}------------------------------------------------
## KEY TECHNICAL SPECIFICATIONS
The SIMEOX 200 has two different Therapy settings. The first is Standard Mode, which can be set to 25%. 50%. 75%, or 100% power. Standard Mode gives the same amount of air flow while you breathe out into the device for your entire exhalation. The other type of Therapy mode is ADAPT Mode. With ADAPT Mode, the device will decrease the amount of airflow during your exhalation, which may make breathing into the device more comfortable.
The SIMEOX 200 has two different Trigger settings, also. The first is a Manual Trigger, which means the device will not start the airflow unless you press the GO button, either the button on the device on the one on the remote control. The other Trigger is the DETECT Trigger. When the DETECT Trigger is chosen, the GO button must be pressed, but the device will not start air flow therapy until you begin breathing out into the mouthpiece. This may help you if it is difficult to coordinate the GO button with your exhalations.
Both Therapy Modes can be used with both Trigger Modes, so any combination of Therapy and Trigger may be chosen.
A summary of key specifications for device operation are as follows:
- Power Modes: Standard Mode (adjustable at 25%, 50%, 75%, or 100% total power)* and . Adapt Mode (power is automatically regulated by the device, consisting of decreased power over the course of each exhalation).
- Cycle Duration: 10 breaths (exhalations). The device generates vibration during exhalation . into the device mouthpiece.
- . Oscillation Frequency: 12 Hz for the first eight (8) exhalations of each 10-breath cycle and 6 Hz for the last two (2) exhalations of each 10-breath cycle, to enable secretion transport.
- Duration of treatment: The device will perform a maximum of 10 cycles (i.e., 100 exhalations) ● or 60 minutes of therapy, whichever occurs first, prior to shutting down, for each mode of operation.
- Activation of Therapy (Detect and Manual Trigger): In Detect Trigger, after the patient . presses the 'Go' button, the device does not deliver a signal until it detects patient exhalation has begun. With the device set to Manual Trigger, therapy is initiated as soon as the 'Go' button is activated.
## SUBSTANTIAL EQUIVALENCE
The SIMEOX 200 Airway Clearance Device is substantially equivalent in intended use and technological characteristics to two predicate devices. The primary predicate device for the SIMEOX 200 is the AirPhysio device (K203209). The secondary predicate is the SimplyClear device (K122111). All three devices, the subject device and the two predicates, are intended to improve the clearance of mucosal secretions from the airways by vibrating the airway walls to assist in loosening mucus. All three devices apply this vibration during exhalation into the device via a patient mouthpiece. All three devices may be used with therapy being delivered during
{7}------------------------------------------------
exhalation-only, with inhalation occurring at an ambient/atmospheric pressure: this is inherent to the SIMEOX 200 and AirPhysio devices, and this is one available operating mode for the SimplyClear Device, namely, exsufflation-only mode.
The following table summarizes the key technical characteristics between the SIMEOX 200 Airway Clearance Device, the SimplyClear Device, and the AirPhysio Device.
| | SUBJECT DEVICE<br>(SIMEOX-200) | PRIMARY<br>PREDICATE<br>(AIRPHYSIO) | SECONDARY<br>PREDICATE<br>(SIMPLYCLEAR) | SUMMARY/<br>DISCUSSION |
|--------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Product<br>Name | SIMEOX-200<br>Airway Clearance<br>Device | AirPhysio Positive<br>Expiratory<br>Pressure (PEP)<br>Device | SimplyClear<br>Device | N/A |
| Manufacturer | Inogen, Inc. | AirPhysio Pty Ltd | Philips Respironics<br>Inc., USA | N/A |
| 510(k)<br>number | K241091 | K203209 | K122111 | N/A |
| Regulatory<br>Classification | 21 CFR 868.5665 | 21 CFR 868.5665 | 21 CFR 868.5665 | Identical |
| Regulatory<br>Identification | A powered<br>percussor is a<br>device that is<br>intended to transmit<br>vibration through a<br>patient's chest wall<br>to aid in freeing<br>mucus deposits in<br>the lung in order to<br>improve bronchial<br>drainage and that<br>may be powered by<br>electricity or<br>compressed gas. | A powered<br>percussor is a device<br>that is intended to<br>transmit vibration<br>through a patient's<br>chest wall to aid in<br>freeing mucus<br>deposits in the lung<br>in order to improve<br>bronchial drainage<br>and that may be<br>powered by<br>electricity or<br>compressed gas. | A powered<br>percussor is a<br>device that is<br>intended to<br>transmit vibration<br>through a patient's<br>chest wall to aid<br>in freeing mucus<br>deposits in the<br>lung in order to<br>improve bronchial<br>drainage and that<br>may be powered<br>by electricity or<br>compressed gas. | Identical |
| Classification | Class II | Class II | Class II | Identical |
| Indications<br>for Use | The SIMEOX 200<br>airway clearance<br>device is intended to | AirPhysio is<br>indicated for use as a | Intended Use: The<br>Philips Respironics<br>SimplyClear device | Identical: All 3<br>devices are |
| | | | | |
| promote airway | Pressure (PEP)<br>Device for | assists patients in<br>loosening,<br>mobilizing and<br>clearing secretions<br>by providing high<br>frequency<br>oscillatory<br>vibrations while<br>gradually applying a<br>positive pressure to<br>the airway, then<br>rapidly shifting to a<br>negative pressure.<br>The oscillatory<br>vibrations assist in<br>loosening and<br>mobilizing the<br>secretions while the<br>rapid shift in<br>pressure produces a<br>high expiratory<br>flow rate from the<br>lungs, which<br>promotes the<br>clearance of<br>secretions. | patients in<br>loosening,<br>mobilizing and<br>clearing<br>secretions by<br>providing high<br>frequency<br>oscillatory<br>vibrations to the<br>airways.<br>Different: Only<br>SimplyClear is<br>intended for<br>patients with<br>"inability to<br>cough", therefore<br>SIMEOX-200<br>treats a subset of<br>SimplyClear<br>intended use.<br>Only<br>SimplyClear is<br>intended for<br>patients with<br>endotracheal or<br>tracheostomy<br>tube, therefore<br>SIMEOX 200<br>treats a subset of<br>the SimplyClear<br>intended use.<br>SIMEOX 200<br>and the<br>SimplyClear in<br>exsufflation-only<br>mode deliver<br>vibration<br>negative pressure<br>during<br>exhalation.<br>AirPhysio<br>delivers vibration | |
| clearance and to | improving the | | | |
| improve bronchial | clearance of mucus | | | |
| drainage by | secretions from the | | | |
| enhancing | airways. | | | |
| mobilization of | | | | |
| bronchial secretions | AirPhysio vibrates | | | |
| via high frequency | the airway walls to | | | |
| oscillatory | assist in loosening | | | |
| vibrations and | mucus and | | | |
| intermittent negative | intermittently | | | |
| pressure to the | increases | | | |
| airway during | endobronchial | | | |
| exhalation. The | pressure so that the | | | |
| device is intended to | airway patency can | | | |
| be prescribed for use | be maintained during | | | |
| in patients capable | exhalation, reducing…
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.