← Product Code [PHR](/productcode/PHR) · K241041

# Silver Fluoride Hypersensitivity Varnish (K241041)

_Young Dental Manufacturing Co. 1, LLC · PHR · Feb 28, 2025 · Dental · SESE_

**Canonical URL:** https://fda.innolitics.com/device/K241041

## Device Facts

- **Applicant:** Young Dental Manufacturing Co. 1, LLC
- **Product Code:** [PHR](/productcode/PHR.md)
- **Decision Date:** Feb 28, 2025
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 872.3260
- **Device Class:** Class 2
- **Review Panel:** Dental
- **Attributes:** Therapeutic

## Indications for Use

Young Silver Fluoride Hypersensitivity Varnish is indicated for the treatment of dental hypersensitivity in adults. It is intended for professional use only.

## Device Story

Young Silver Fluoride Hypersensitivity Varnish is a dental varnish containing silver, sodium fluoride, chitosan, and acetic acid. It is used by dental professionals in a clinical setting to treat dental hypersensitivity. The device is applied topically to cleaned, isolated tooth surfaces using a brush applicator. The varnish works by physically occluding exposed dentinal tubules via its silver and fluoride components; chitosan and acetic acid facilitate even solution dispersion. The device is supplied in a 10mL bottle with a dropper tip. By sealing the tubules, the varnish reduces sensitivity, providing patient relief. The procedure involves cleaning the tooth, isolating the area, applying the solution, and allowing it to air dry. A second application may be administered after one week if necessary.

## Clinical Evidence

No clinical data. Bench testing only. Studies included dentinal occlusion (SEM analysis), ion release (fluoride and silver), hydraulic conductance (microfluidic flow reduction), device lifetime, transportation (ASTM D4169), and shelf-life stability. Biocompatibility testing performed per ISO 10993 (cytotoxicity, sensitization, irritation, oral toxicity, genotoxicity, subacute/subchronic toxicity, and implantation).

## Technological Characteristics

Composition: colloidal silver, silver cation, sodium fluoride, chitosan, acetic acid, purified water. Principle: physical occlusion of dentinal tubules via silver/fluoride ions. Form factor: 10mL bottle with dropper tip. Application: 1-step, brush-applied. Sterilization: not specified. Software: none.

## Regulatory Identification

Cavity varnish is a device that consists of a compound intended to coat a prepared cavity of a tooth before insertion of restorative materials. The device is intended to prevent penetration of restorative materials, such as amalgam, into the dentinal tissue.

## Special Controls

*Classification.* Class II (special controls). The device, when it is an external cleaning solution, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 872.9.

## Predicate Devices

- Elevate Advantage Arrest ([K102973](/device/K102973.md))

## Reference Devices

- Riva Star ([K172047](/device/K172047.md))
- Colgate PreviDent Varnish

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image contains the logos of the Department of Health & Human Services and the U.S. Food & Drug Administration (FDA). The Department of Health & Human Services logo is on the left, and the FDA logo is on the right. The FDA logo includes the text "FDA U.S. FOOD & DRUG ADMINISTRATION" in blue.

February 28, 2005

Young Dental Manufacturing Co. 1, LLC Brian Prange Director, Product Development 13705 Shoreline Ct. East Earth City, Missouri 63045

Re: K241041

Trade/Device Name: Silver Fluoride Hypersensitivity Varnish Regulation Number: 21 CFR 872.3260 Regulation Name: Cavity Varnish Regulatory Class: Class II Product Code: PHR Dated: January 27, 2025 Received: January 27, 2025

Dear Brian Prange:

We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{1}------------------------------------------------

Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).

Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30. Design controls; 21 CFR 820.90. Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.

All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-device-advicecomprehensive-regulatory-assistance/unique-device-identification-system-udi-system.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatory

{2}------------------------------------------------

assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

# MICHAEL E. ADJODHA -S

Michael E. Adjodha, MChE, RAC, CQIA Assistant Director DHT1B: Division of Dental and ENT Devices OHT1: Office of Ophthalmic, Anesthesia, Respiratory, ENT, and Dental Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{3}------------------------------------------------

## Indications for Use

Submission Number (if known)

K241041

Device Name

Silver Fluoride Hypersensitivity Varnish

Indications for Use (Describe)

Young Silver Fluoride Hypersensitivity Varnish is indicated for the treatment of dental hypersensitivity in adults.

It is intended for professional use only.

Type of Use (Select one or both, as applicable)

< Prescription Use (Part 21 CFR 801 Subpart D)

Over-The-Counter Use (21 CFR 801 Subpart C)

#### CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{4}------------------------------------------------

# 510(k) Summary

## K241041

| Premarket Notification (510(k)) Number |  |
|----------------------------------------|--|
|----------------------------------------|--|

| Submitter's Name                  | Young Dental Manufacturing Co. 1, LLC<br>13705 Shoreline Ct. East<br>Earth City, MO 63045<br>Telephone: (314) 344-0010                       |
|-----------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------|
| Contact Person                    | Brian Prange<br>Director of Product Development<br>Young Dental Manufacturing Co. 1, LLC<br>13705 Shoreline Ct. East<br>Earth City, MO 63045 |
| Establishment Registration Number | 1941138                                                                                                                                      |
| Date Prepared                     | February 28, 2025                                                                                                                            |
| Device Trade Name                 | Young Silver Fluoride Hypersensitivity<br>Varnish                                                                                            |
| Device Common Name                | Cavity Varnish                                                                                                                               |
| Classification Name               | Cavity Varnish                                                                                                                               |
| Product Code                      | PHR                                                                                                                                          |
| Device Classification             | Class II                                                                                                                                     |
| Panel                             | Dental                                                                                                                                       |
| Regulatory Classification         | 21 CFR 872.3260                                                                                                                              |
| Type of 510(k) Submission         | Traditional                                                                                                                                  |
| Legally Marketed Predicate Device | Elevate Advantage Arrest (K102973)                                                                                                           |

{5}------------------------------------------------

## Description

Young Silver Fluoride Hypersensitivity Varnish is a formulation of silver, sodium fluoride, chitosan, acetic acid, and purified water intended for the treatment of dental hypersensitivity in adults. It is intended for professional use only.

Like other 510(k) cleared dental hypersensitivity varnishes including the predicate device, the proposed device works by physically occluding dentinal tubules through its silver and fluoride components. It contains small silver particles to facilitate their entry into the dentinal tubules, as well as chitosan and acetic acid to evenly disperse the solution. The product is supplied in a 10mL bottle with a dropper dispensing tip.

One to two drops of the varnish can treat up to 8 sites. If needed, a second application of the varnish may be administered after 1 week.

## Indications for Use

Young Silver Fluoride Hypersensitivity Varnish is indicated for the treatment of dental hypersensitivity in adults. It is intended for professional use only.

#### Predicate Device

Young Silver Fluoride Hypersensitivity Varnish is substantially equivalent to the Elevate Advantage Arrest Hypersensitivity Varnish cleared under K102973. Both devices are intended for the treatment of dental hypersensitivity in adults. Both the subject and predicate devices also have similar technological characteristics. In particular, both devices contain silver and fluoride to physically occlude dentinal tubules in the treatment of dental hypersensitivity.

Please refer to the Substantial Equivalence Summary Comparison Table below for a summary of the similarities and differences between the subject and predicate devices.

#### Description of Operation/Use

After cleaning the tooth/teeth to be treated and adjacent teeth with a non-fluoride prophylaxis paste and prophy cup, the condition of the teeth should be assessed. Only teeth with normal pulp status should be treated. The affected area of the tooth is isolated with cotton rolls and gingival tissue of the affected tooth is protected with petroleum jelly or similar. Alternatively, a rubber dam or saliva evacuator can be used to isolate the area.

{6}------------------------------------------------

One to two drops of solution (i.e., the subject device) can treat up to 8 sites per patient. The solution is dispensed from the device bottle's dropper tip into a disposable dappen dish. Ensuring that the affected tooth surface is dry, a small brush applicator is used to carefully transfer the solution directly to the affected tooth surface, which is allowed to air dry.

All protective/isolation materials used in the mouth are removed and the dappen dish and applicators are discarded in accordance with local regulations. If needed, a second application may be administered after 1 week.

Young Silver Fluoride Hypersensitivity Varnish is a prescription device intended for professional use only.

# Technological Characteristics Summary

Young Silver Fluoride Hypersensitivity Varnish is substantially equivalent to Elevate Advantage Arrest (K102973) with respect to intended use, principle of operation/use, technological characteristics, and performance.

Both devices are cavity varnishes with the same intended use and the same key functional ingredients (silver and fluoride) that physically occlude dentinal tubules in the treatment of hypersensitivity. While the diamine (ammonia component) in the predicate device is used to allow the silver and fluoride to be in an evenly dispersed solution, the subject device uses acetic acid and chitosan to evenly disperse the solution. The subject device includes small silver particles to facilitate their entry into the dentinal tubules. These differences do not affect the overall means by which the silver and fluoride in the subject and predicate devices are used to occlude dentinal tubules in the treatment of dental hypersensitivity and do not raise new questions of safety or effectiveness.

| Characteristic                       | Young Silver<br>Fluoride<br>Hypersensitivity<br>Varnish<br>(Subject Device)                                                                                                                                                                                                                                                                                                                                  | Elevate Advantage<br>Arrest (K102973)                                                                                                                                                                                                                                                                                                                                                                                | Comparison                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                        |
|--------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Classification Name                  | Cavity Varnish                                                                                                                                                                                                                                                                                                                                                                                               | Cavity Varnish                                                                                                                                                                                                                                                                                                                                                                                                       | Identical                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                         |
| CFR Citation                         | 21 CFR 872.3260                                                                                                                                                                                                                                                                                                                                                                                              | 21 CFR 872.3260                                                                                                                                                                                                                                                                                                                                                                                                      | Identical                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                         |
| Classification Panel                 | Dental                                                                                                                                                                                                                                                                                                                                                                                                       | Dental                                                                                                                                                                                                                                                                                                                                                                                                               | Identical                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                         |
| Classification                       | Class II                                                                                                                                                                                                                                                                                                                                                                                                     | Class II                                                                                                                                                                                                                                                                                                                                                                                                             | Identical                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                         |
| Product Code                         | PHR                                                                                                                                                                                                                                                                                                                                                                                                          | PHR                                                                                                                                                                                                                                                                                                                                                                                                                  | Identical                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                         |
| Characteristic                       | Young Silver<br>Fluoride<br>Hypersensitivity<br>Varnish<br>(Subject Device)                                                                                                                                                                                                                                                                                                                                  | Elevate Advantage<br>Arrest (K102973)                                                                                                                                                                                                                                                                                                                                                                                | Comparison                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                        |
| Product Code<br>Description          | Diamine Silver<br>Fluoride Dental<br>Hypersensitivity<br>Varnish                                                                                                                                                                                                                                                                                                                                             | Diamine Silver<br>Fluoride Dental<br>Hypersensitivity<br>Varnish                                                                                                                                                                                                                                                                                                                                                     | Although the proposed<br>product does not<br>include diamine silver,<br>the diamine (ammonia<br>component) in the<br>predicate device is<br>used to allow the<br>silver and fluoride to<br>be in an evenly<br>dispersed solution,<br>while the proposed<br>device uses acetic acid<br>and chitosan to evenly<br>disperse the solution.<br>Both devices contain<br>silver and fluoride to<br>physically occlude the<br>dentinal tubules in the<br>treatment of dental<br>hypersensitivity,<br>therefore these<br>differences do not<br>raise different<br>questions of safety or<br>effectiveness.<br>Performance testing<br>demonstrates the<br>subject device<br>occludes dentinal<br>tubules and releases<br>silver and fluoride<br>ions comparably to the<br>predicate device. |
| Active Ingredients                   | 0.20% silver<br>3.00% sodium fluoride                                                                                                                                                                                                                                                                                                                                                                        | 38% silver diamine<br>fluoride                                                                                                                                                                                                                                                                                                                                                                                       | Both formulations<br>contain silver and                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                           |
| Characteristic                       | Young Silver<br>Fluoride<br>Hypersensitivity<br>Varnish<br>(Subject Device)                                                                                                                                                                                                                                                                                                                                  | Elevate Advantage<br>Arrest (K102973)                                                                                                                                                                                                                                                                                                                                                                                | Comparison                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                        |
| Device Description                   | Varnish containing<br>silver and fluoride<br>applied directly to<br>affected tooth or teeth<br>in the treatment of<br>dental<br>hypersensitivity.                                                                                                                                                                                                                                                            | Varnish containing<br>silver and fluoride<br>applied directly to<br>affected tooth or teeth<br>in the treatment of<br>dental<br>hypersensitivity.                                                                                                                                                                                                                                                                    | fluoride to achieve the<br>intended effect of<br>occluding dentinal<br>tubules in the<br>treatment of<br>hypersensitivity. The<br>subject product<br>includes silver and<br>fluoride, but not silver<br>diamine fluoride (i.e.,<br>ammonia).<br>Biocompatibility<br>testing has been<br>successfully<br>completed for the<br>subject device.<br>Similar. Both devices<br>have the same<br>intended use and the<br>same functional<br>ingredients (silver and<br>fluoride) to physically<br>occlude dentinal<br>tubules in the<br>treatment of dental<br>hypersensitivity. The<br>subject device<br>additionally includes<br>chitosan and acetic<br>acid to evenly disperse<br>the solution and omits<br>the diamine<br>(ammonia) used in the<br>predicate.                        |
| Characteristic                       | Young Silver<br>Fluoride<br>Hypersensitivity<br>Varnish<br>(Subject Device)                                                                                                                                                                                                                                                                                                                                  | Elevate Advantage<br>Arrest (K102973)                                                                                                                                                                                                                                                                                                                                                                                | Comparison                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                        |
| Intended Use/<br>Indications for Use | Treatment of dental<br>hypersensitivity in<br>adults                                                                                                                                                                                                                                                                                                                                                         | Treatment of dental<br>hypersensitivity in<br>adults                                                                                                                                                                                                                                                                                                                                                                 | Identical                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                         |
| Technological<br>Characteristics     | Composition:<br>Colloidal silver<br>and silver cation<br>mixture Sodium fluoride Chitosan Acetic acid Purified water                                                                                                                                                                                                                                                                                         | Composition:<br>38% silver<br>diamine fluoride              Note: beyond the<br>public disclosure of<br>38% silver diamine<br>fluoride, the exact<br>formulation of Elevate<br>Advantage Arrest is<br>proprietary. In<br>dentistry practice,<br>silver diamine fluoride<br>is commonly prepared<br>by mixing powdered<br>silver fluoride into an<br>ammonia/water<br>mixture as follows:<br>25% silver<br>8% ammonia | Similar. Both devices<br>contain silver and<br>fluoride to occlude<br>dentinal tubules in the<br>treatment of dental<br>hypersensitivity. The<br>proposed device<br>contains small silver<br>particles to facilitate<br>their entry into the<br>dentinal tubules as<br>well as chitosan and<br>acetic acid to evenly<br>disperse the solution.<br>These differences do<br>not affect the overall<br>means by which the<br>dentinal tubules are<br>occluded in the<br>treatment of                                                                                                                                                                                                                                                                                                 |
| Characteristic                       | Young Silver<br>Fluoride<br>Hypersensitivity<br>Varnish<br>(Subject Device)                                                                                                                                                                                                                                                                                                                                  | Elevate Advantage<br>Arrest (K102973)                                                                                                                                                                                                                                                                                                                                                                                | Comparison                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                        |
|                                      |                                                                                                                                                                                                                                                                                                                                                                                                              | 5% fluoride<br>62% water                                                                                                                                                                                                                                                                                                                                                                                             | hypersensitivity nor do<br>they raise new<br>questions of safety and<br>effectiveness.<br>Performance testing<br>demonstrates the<br>subject device<br>occludes dentinal<br>tubules and releases<br>silver and fluoride<br>ions comparably to the<br>predicate device.                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                            |
| pH                                   | Acid                                                                                                                                                                                                                                                                                                                                                                                                         | Strong Base                                                                                                                                                                                                                                                                                                                                                                                                          | While the proposed<br>product has an acid pH<br>and the predicate is a<br>strong base<br>(ammonia), these<br>differences do not<br>affect safety or<br>effectiveness. The<br>ammonia in the<br>predicate device is<br>used to allow the<br>silver and fluoride to<br>be in an evenly<br>dispersed solution,<br>while the proposed<br>device uses acetic acid<br>and chitosan to evenly<br>disperse the solution.<br>In addition, the subject<br>device is used no more<br>frequently, and<br>potentially less<br>frequently than the                                                                                                                                                                                                                                              |
| Characteristic                       | Young Silver<br>Fluoride<br>Hypersensitivity<br>Varnish<br>(Subject Device)                                                                                                                                                                                                                                                                                                                                  | Elevate Advantage<br>Arrest (K102973)                                                                                                                                                                                                                                                                                                                                                                                | Comparison                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                        |
| Mechanism of Action                  | Like other dental<br>hypersensitivity<br>devices, the<br>formulation works by<br>physically occluding<br>dentinal tubules. Like<br>other silver fluoride<br>formulations cleared<br>under the 510(k)<br>process, treatment of<br>dental hypersensitivity<br>is achieved by the<br>silver and fluoride<br>components of the<br>formulation, which<br>physically plug or<br>close exposed dentinal<br>tubules. | Like other dental<br>hypersensitivity<br>devices, the<br>formulation works by<br>physically occluding<br>dentinal tubules. Like<br>other silver fluoride<br>formulations cleared<br>under the 510(k)<br>process, treatment of<br>dental hypersensitivity<br>is achieved by the<br>silver and fluoride<br>components of the<br>formulation, which<br>physically plug or<br>close exposed dentinal<br>tubules.         | predicate device, since<br>the subject and<br>predicate devices may<br>both be applied once<br>or twice, while the<br>predicate device may<br>additionally be applied<br>a third time.<br>Similar – both devices<br>include silver and<br>fluoride and treat<br>dental hypersensitivity<br>by physically<br>occluding dentinal<br>tubules. Performance<br>testing demonstrates<br>the subject device<br>occludes dentinal<br>tubules and releases<br>silver and fluoride<br>ions comparably to the<br>predicate device.                                                                                                                                                                                                                                                           |
| Number of Application<br>Steps       | 1-step                                                                                                                                                                                                                                                                                                                                                                                                       | 1-step                                                                                                                                                                                                                                                                                                                                                                                                               | Same                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                              |
| Method of Application                | One drop treats up to 8<br>sites                                                                                                                                                                                                                                                                                                                                                                             | One drop treats 5 sites                                                                                                                                                                                                                                                                                                                                                                                              | Similar                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                           |
| Number of<br>Applications            | Up to two                                                                                                                                                                                                                                                                                                                                                                                                    | Up to three                                                                                                                                                                                                                                                                                                                                                                                                          | Similar. Subject<br>device has lower<br>maximum frequency<br>of use                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                               |

## Substantial Equivalence Summary Comparison Table

{7}------------------------------------------------

{8}------------------------------------------------

{9}------------------------------------------------

{10}------------------------------------------------

{11}------------------------------------------------

{12}------------------------------------------------

## Performance Testing

As described below, comprehensive bench and biocompatibility was submitted to confirm product conformance with device requirements and substantially equivalent performance as the predicate device. Performance data demonstrated dentinal occlusion and silver and fluoride ion release were comparable to the predicate device. Additional testing included transportation, stability/shelf life, device lifetime testing, and simulated use and functional testing. The key studies summarized below demonstrate that the device is as safe, as effective, and performs as well or better than the legally marketed predicate device.

In addition, predicate device characterization testing was also performed.

## Biocompatibility Testing

Biocompatibility testing was conducted in accordance with ISO 10993. Testing included cytotoxicity, sensitization, irritation, oral toxicity, and genotoxicity. Other endpoints evaluated in the Biological Evaluation Report included subacute/subchronic toxicity and implantation.

## Key Bench Testing

Dentinal Occlusion Testing: Dentinal occlusion testing was conducted to compare the subject and predicate devices with respect to their ability to occlude human dentine tubules. A scanning electron microscope (SEM) was used to capture images of the dentine samples and trained assessors graded the occlusion achieved by the test varnishes on a 5-point occlusion scale ranging from 1 (occluded) to 5 (unoccluded). SEM images of the dentine tubules of each sample were captured at 3000X magnification. One representative SEM image of the dentine tubules of each sample was additionally captured at 500X magnification. The study demonstrated that the subject device performed as well or better than the predicate device in occluding dentinal tubules.

Ion Release Testing: The subject and predicate devices were evaluated and compared with respect to the release of free fluoride and silver ions into deionized water solutions for 7 days. Both the subject and predicate devices were effective at releasing free fluoride ions and free silver ions once they were inserted into deionized water. Both devices had similar release profiles, with the majority of fluoride and silver ions released within the first 15 minutes. The percentages of silver released over 7 days was comparable between the subject and predicate devices.

{13}------------------------------------------------

Hydraulic Conductance Testing: The study used human dentine samples as the test substrate to evaluate the ability of a single application of four varnish treatments versus a negative control to occlude dentine tubules when assessed via hydraulic conductance. A worst-case formulation of the subject device with 2.72% sodium fluoride (vs. 3% in the final formulation) was compared against the predicate device (Elevate Advantage Arrest), Riva Star (K172047), Colgate PreviDent Varnish, and deionized water as a control. The mean percentage reduction in microfluidic flow was highest for Elevate Advantage Arrest, followed by Young formulation. The Young Varnish performed significantly better than the PreviDent Varnish and deionized water control. The Young Varnish also had a much larger reduction in flow than the Riva Star varnish, though the difference was not statistically significant.

Device Lifetime Testing: This study confirmed a variety of device attributes in a simulated device lifetime evaluation where the samples were subjected to approximately three times the exposure to air and sunlight as expected during normal use. The study demonstrated that the device and packaging configuration can withstand a typical device lifetime without experiencing any significant damage or product degradation.

Transportation Testing: Packaged product was exposed to hot, cold and ambient conditions and subjected to simulated shipping conditions according to ASTM D4169 Schedule 13 guidelines. Shipper boxes, unit boxes and product bottles were visually assessed for damage, legibility of labeling, evidence of leaking, and other attributes. All testing was successfully completed, demonstrating that the packaging configuration can withstand a distribution evaluation.

#### Shelf Life Testing

Accelerated and real-time stability studies of the subject device were conducted to support the proposed shelf life. The products were packaged in the intended packaging (including bottle with dispensing tip and cap) and subjected to accelerated conditions and evaluated at relevant time points. Product was evaluated for a range of functional and specification attributes after both accelerated and real-time conditions. All measures passed all tests for all test conditions to support the labeled shelf life.

#### Conclusions

The bench, biocompatibility and stability/shelf life studies demonstrated the subject device is (1) biocompatible; (2) maintains device and package integrity over the labeled shelf and usage life; and (3) has substantially equivalent performance to the predicate device with respect to dentinal tubule occlusion and silver and fluoride ion release. These results demonstrate that Young Silver 

{14}------------------------------------------------

Fluoride Hypersensitivity Varnish is as safe, as effective, and performs as well or better than the legally marketed predicate device.

# Summary

In summary, Young Silver Fluoride Hypersensitivity Varnish is substantially equivalent to the predicate Elevate Advantage Arrest varnish with respect to intended use, technological characteristics and performance. The biocompatibility and bench studies concluded that the device is as safe, as effective, and performs as well or better than the legally marketed predicate device.

---

**Source:** [https://fda.innolitics.com/device/K241041](https://fda.innolitics.com/device/K241041)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
