The TOV Dental Implant System are endosseous implants intended to be surgically placed in the upper or lower jaw arches to provide support for prosthetic devices, such as an artificial tooth, in order to restore patient's esthetics and chewing function. Implants are intended for single or multiple unit restorations on splinted or non-splinted applications. Maer, Ragil and TCX are intended for immediate loading when good primary stability is achieved, and with appropriate occlusive loading. These implants can also be used for loading after a conventional healing period. Ragil 3.3 implants are intended to replace a lateral incisor in the maxilla and/or a central or lateral incisor in the mandible. Mandibular central and lateral incisors must be splinted if using two or more 3.3 implants adjacent to one another.
Device Story
TOV Dental Implant System consists of endosseous dental implants (Maer, Ragil, TCX designs) and various abutments; surgically placed in jaw arches by dental professionals; provides support for prosthetic devices (artificial teeth) to restore chewing function and esthetics. Implants feature internal hex or conical connections; grit-blasted and acid-etched surface. Supports immediate or delayed loading protocols. Abutments include healing caps, straight, angled, UCLA, ball, and multi-unit attachments. System components are manufactured from Ti-6AL-4V ELI. Clinical benefit includes restoration of oral function and aesthetics. Device is used in dental clinics; output is the physical implant/abutment assembly integrated into the patient's jaw.
Clinical Evidence
Bench testing only. Evidence includes dynamic fatigue testing (ISO 14801), biocompatibility (cytotoxicity, skin sensitization, irritation per ISO 10993), sterilization validation (ISO 17665-1, ISO 11137-2), bacterial endotoxin testing (ANSI/AAMI ST72:2019, USP <161>), and shelf-life/package integrity testing (ASTM D999-08, F3039-13, D5276-98, F1929-12, F1980-07). Surface cleanliness validated via SEM/EDX and organic carbon analysis.
Technological Characteristics
Materials: Ti-6AL-4V ELI. Surface: Grit-blasted and acid-etched (SLA). Connection: Internal hex and conical. Components: Implants (Maer, Ragil, TCX), abutments (straight, angled, UCLA, ball, multi-unit), healing caps, cover screws. Sterilization: Steam (ISO 17665-1) and Gamma irradiation (ISO 11137-2).
Indications for Use
Indicated for patients requiring endosseous dental implants in upper or lower jaw arches to support prosthetic devices (artificial teeth) for restoration of esthetics and chewing function. Includes single/multiple unit restorations. Ragil 3.3 implants indicated for maxillary lateral incisors or mandibular central/lateral incisors; mandibular incisors require splinting if adjacent 3.3 implants are used.
Regulatory Classification
Identification
An endosseous dental implant is a prescription device made of a material such as titanium or titanium alloy that is intended to be surgically placed in the bone of the upper or lower jaw arches to provide support for prosthetic devices, such as artificial teeth, in order to restore a patient's chewing function.
Special Controls
*Classification.* (1) Class II (special controls). The device is classified as class II if it is a root-form endosseous dental implant. The root-form endosseous dental implant is characterized by four geometrically distinct types: Basket, screw, solid cylinder, and hollow cylinder. The guidance document entitled “Class II Special Controls Guidance Document: Root-Form Endosseous Dental Implants and Endosseous Dental Implant Abutments” will serve as the special control. (See § 872.1(e) for the availability of this guidance document.)(2)
*Classification.* Class II (special controls). The device is classified as class II if it is a blade-form endosseous dental implant. The special controls for this device are:(i) The design characteristics of the device must ensure that the geometry and material composition are consistent with the intended use;
(ii) Mechanical performance (fatigue) testing under simulated physiological conditions to demonstrate maximum load (endurance limit) when the device is subjected to compressive and shear loads;
(iii) Corrosion testing under simulated physiological conditions to demonstrate corrosion potential of each metal or alloy, couple potential for an assembled dissimilar metal implant system, and corrosion rate for an assembled dissimilar metal implant system;
(iv) The device must be demonstrated to be biocompatible;
(v) Sterility testing must demonstrate the sterility of the device;
(vi) Performance testing to evaluate the compatibility of the device in a magnetic resonance (MR) environment;
(vii) Labeling must include a clear description of the technological features, how the device should be used in patients, detailed surgical protocol and restoration procedures, relevant precautions and warnings based on the clinical use of the device, and qualifications and training requirements for device users including technicians and clinicians;
(viii) Patient labeling must contain a description of how the device works, how the device is placed, how the patient needs to care for the implant, possible adverse events and how to report any complications; and
(ix) Documented clinical experience must demonstrate safe and effective use and capture any adverse events observed during clinical use.
{1}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, with the letters "FDA" in a blue square. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
August 15, 2024
RDJ TOV Implant Ltd % Angela Blackwell Senior Consultant Blackwell Device Consulting P.O. Box 718 Gresham, Oregon 97030-0172
Re: K240837
Trade/Device Name: TOV Dental Implant System Regulation Number: 21 CFR 872.3640 Regulation Name: Endosseous Dental Implant Regulatory Class: Class II Product Code: DZE, NHA Dated: July 16, 2024 Received: July 17, 2024
Dear Angela Blackwell:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{2}------------------------------------------------
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30. Design controls; 21 CFR 820.90. Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
{3}------------------------------------------------
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
# Andrew I. Steen -S
Andrew I. Steen Assistant Director DHT1B: Division of Dental and ENT Devices OHT1: Office of Ophthalmic, Anesthesia, Respiratory, ENT, and Dental Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{4}------------------------------------------------
# Indications for Use
510(k) Number (if known) K240837
Device Name TOV Dental Implant System
#### Indications for Use (Describe)
The TOV Dental Implant System are endosseous implants intended to be surgically placed in the upper or lower jaw arches to provide support for prosthetic devices, such as an artificial tooth, in order to restore patient's esthetics and chewing function. Implants are intended for single or multiple unit restorations on splinted applications. Maer, Ragil and TCX are intended for immediate loading when good primary stability is achieved, and with appropriate occlusive loading. These implants can also be used for loading after a conventional healing period. Ragil 3.3 implants are intended to replace a lateral incisor in the maxilla and/or a central or lateral incisor in the mandible. Mandibular central and lateral incisors must be splinted if using two or more 3.3 implants adjacent to one another.
Type of Use (Select one or both, as applicable)> Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{5}------------------------------------------------
# 510k Summary August 12, 2024 TOV Dental Implant System
Name and address: RDJ TOV Implant Ltd Mutsafi Street 8/4, View Tower Jerusalem 96420 Israel Contact Person: David Elkouby Phone Number: +972 548397143 Email: contact@tov-implant.com Name of device: TOV Dental Implant System Classification Name: Endosseous dental implants CFR: 21 CFR 872.3640 Primary Product Code: DZE Secondary Product Code: NHA
Device Description: The TOV Dental Implant System contains 2 designs of internal hex implants and various types of abutments as described below as well as 1 design of conical implant with NP and RP platforms and abutments corresponding to both platforms. All devices are made from Ti-6AL-4V ELI unless otherwise noted. The implants have a grit blasted and acid etched surface. Conical abutments need to match the implant platform of narrow or regular. No abutments other than UCLA abutments are intended to be modified by the user.
Maer implants are tapered internal hex implants with more tightly spaced flat edge threads at the top and wider spaced flat edge threads in the lower tapered section. Maer comes in 3.5, 3.75, 4.2, 5.0 and 6.0 mm diameter with lengths of 8, 10, 11.5, 13, and 16mm in 6.0mm diameter).
Ragil implants are cylindrical internal hex implants with evenly spaced threads which are flat edged at the top and sharp edged in the lower section. Ragil comes in 3.3, 3.75, 4.2, 5.0, and 6.0 mm diameter with lengths of 8, 10, 11.5, 13, and 16mm in 5.0 or 6.0mm diameter). 3.3mm diameter implants are not for use with angled abutments in the posterior region of the mouth.
Internal Hex Healing Abutments come in 4.5mm diameter with cuff heights of 2, 3, 4, 5, 6,7 mm. They come in narrow with a diameter 3.8mm with cuff heights of 3, 4, 5,6, 7 mm and wide with a diameter 5.5mm with cuff heights of 2,3,4,5,6,7 mm. Extra wide healing caps are 6.3mm diameter and come in cuff heights of 2,3,4,5 mm. An internal hex cover screw is also available. A healing cap for multi-units is available.
Internal Hex Straight Abutments have a 3.75mm interface diameter and come in 4.5mm diameter with total heights above the platform of 5,7,9,11,13,15 mm, and in 5.5mm diameter with heights above the platform of 9, 11, and 13 mm.
{6}------------------------------------------------
Internal Hex Straight Narrow Abutments have a 3.75mm interface area and come 3.75mm diameter with heights above the platform of 5,7,9, 11 mm.
Internal Hex Straight Shouldered Abutments come in 3.75mm interface diameter in 4.5 and 5.4 mm diameter with gingival heights of 1,2,3,4 mm.
Internal Hex Angled Abutments come in 15° 3.75mm diameter with height above platform of 9, and 11.4mm, and 25° 3.75mm diameter with height above platform of 8.5 mm.
Internal Hex Angled Anatomic Abutments come in 15° 3.75mm diameter with cuff heights of 1,2,3,4 mm and height above low side of shoulder of 8,9,10,11 mm and 25° 3.75mm diameter with cuff heights of 1,2,3,4 mm and height above platform of 8.3, 9.2, 10.3, 10.3 mm.
Internal Hex UCLA abutment bases come in 3.75mm diameter and use a Delrin plastic sleeve. The UCLA are for making straight restorations which are 4mm above the gingival collar and have a post height of no more than 12mm. The minimum wall thickness of the cast abutments is 0.3mm. The angulation, wall thickness and diameter of the UCLA base component are not intended to be modified.
Internal Hex Ball attachments come in 4.0mm diameter with cuff heights of 2, 3, 4, 5, 6, or 7mm.The ball attachments snap into a stainless steel housing which has a polyamide or polyether retention cap. The Retention Caps come in the colors yellow, pink and clear which represent 0.5, 0.9, and 1.3 kg retention levels. The retention caps allow implants to be within 14° of vertical and still snap into place.
Internal Hex Double Loc Attachments come in 3.85mm diameter with cuff heights 1,2,3,4,5,6 mm. The Double Loc attachments snap into a Ti6AL4V ELI housing which has a polyamide, polyether or polyoxymethylene retention cap. The Retention Caps come in the colors yellow, pink, clear, purple which represent 0.6, 0.8, 1.0, 1.5 kg retention levels. The retention caps allow implants to be within 20° of vertical and still snap into place.
Internal Hex Straight Multi-Unit come in 4.8mm diameter with cuff heights of 1, 2, 3, 4, 5 mm.
Internal Hex Angled Multi-Units come in 17° and 30° with cuff heights of 1, or 2 mm.
TCX implants are very slightly tapered conical implants with more tightly spaced flat edge threads at the top and wider spaced sharp edge threads in the lower tapered section. TCX comes in 3.5 (NP), 4.3 (RP) and, 5.0 (RP) mm diameter with lengths of 8, 10, 11.5, 13, and 16mm.
Conical healing caps come in NP and RP with cuff heights of 2,3,4,5 mm. NP and RP conical cover screws are also available. A healing cap for multi-units is available.
Conical straight abutments come in NP and RP with heights above platform of 9 or 13 mm.
Conical shoulder abutments come in NP and RP with gingival heights of 1,2,3 mm.
{7}------------------------------------------------
Conical angled shoulder abutments 15° come in NP and RP with cuff heights of 1,2,3 mm and total heights of 10.7, 12.2, 13.7 mm for NP and 11, 12, 13 mm for RP.
Conical angled shoulder abutments 25° come in NP and RP with cuff heights of 1,2,3 mm and total heights of 10.7, 12.2, 13.7 mm for NP and 11, 12, 13 mm for RP.
Conical ball attachments come in 3.5mm diameter for NP and 5.0mm diameter for RP with cuff heights of 1,2,3,4,5,6 mm. The ball attachments snap into a stainless steel housing which has a polyamide or polyether retention cap. The Retention Caps come in the colors yellow, pink and clear which represent 0.5, 0.9, and 1.3 kg retention levels. The retention caps allow implants to be within 14° of vertical and still snap into place.
Conical Retentor attachments in 3.5mm diameter for NP and 5.0mm diameter for RP with cuff heights of 1,2,3,4 mm. The Retentor attachments snap into a Ti6AL4V ELI housing which has a polyamide, polyether or polyoxymethylene retention cap. The Retention Caps come in the colors yellow, pink, clear, purple which represent 0.6, 0.8, 1.0, 1.5 kg retention levels. The retention caps allow implants to be within 20° of vertical and still snap into place.
Conical straight multi-units in NP and RP are 4.8mm in diameter with cuff heights of 1,2,3,4,5 mm.
Conical NP and RP Angled Multi-Units come in 17° and 30° with cuff heights of 1, or 2 mm.
### Indications for Use:
The TOV Dental Implant System are endosseous implants intended to be surgically placed in the upper or lower jaw arches to provide support for prosthetic devices, such as an artificial tooth, in order to restore patient's esthetics and chewing function. Implants are intended for single or multiple unit restorations on splinted or non-splinted applications. Maer, Ragil and TCX are intended for immediate loading when good primary stability is achieved, and with appropriate occlusive loading. These implants can also be used for loading after a conventional healing period. Ragil 3.3 implants are intended to replace a lateral incisor in the maxilla and/or a central or lateral incisor in the mandible. Mandibular central and lateral incisors must be splinted if using two or more 3.3 implants adjacent to one another.
Testing Summary: Dynamic fatigue testing according to ISO 14801 was conducted to determine the abutments and implants are strong enough for their intended use. Cytotoxicity testing according to ISO 10993 was done on both implants and abutments. Skin sensitization testing to ISO 10993-10:2021 was conducted on implants. Irritation testing according to ISO 10993-23:2021 was conducted on implants. Steam sterilization was conducted according to ISO 17665-1. Bacterial endotoxin testing was conducted according to ANSI/AAMI ST72:2019 and USP <161>. Gamma irradiation was conducted according to ISO 11137-2. Package testing was conducted according to ASTM D999-08, ASTM F3039-13, and ASTM D5276-98(2009) and then shelf life testing was conducted according to ASTM F1929-12, and ASTM F1980-07. Testing of the modified surface included testing for organic carbon, hydrocarbons, and SEM/EDX. All were within the limits based on relevant standards set in the cleaning validation protocol.
{8}------------------------------------------------
#### MR Environment Condition
Non-clinical worst-case MRI review was performed to evaluate the metallic TOV Dental Implant System devices in the MRI environment using scientific rationale and published literature (e.g., Woods, Terry O., Jana G. Delfino, and Sunder Rajan. "Assessment of Magnetically Induced Displacement Force and Torque on Metal Alloys Used in Medical Devices." Journal of Testing and Evaluation 49.2 (2019): 783-795), based on the entire system including all variations (all compatible implant bodies, dental abutments, and fixation screws) and material composition. Rationale addressed parameters per the FDA guidance "Testing and Labeling Medical Devices for Safety in the Magnetic Resonance (MR) Environment," including magnetically induced displacement force and torque.
## Primary Predicate Device: OsseOne Dental Implant System K182293
Reference Predicates: Surgikor Dental Implant System K182615 Cortex Dental Implants K163385 healing abutment 3.8mm with 2mm platform height and angled abutment anatomic/shouldered 3.75mm with 4mm collar, Straumann K171784 NC 3.3 and 3.6mm healing abutment and straight multiunit in gingival height of 5.5mm
### Substantial Equivalence:
The TOV Dental Implant System is substantially equivalent to OsseOne Dental Implants in indications for use, materials, design, and fatigue performance. Slight differences in implant and abutment design between the subject devices and the predicate devices were addressed by showing both had adequate fatigue performance. Some abutment types and sizes needed a reference device so Surgikor Dental Implant System is used. Some individual devices needed a reference device for a specific size or dimension so these are listed in the right column.
| Company &<br>Device Name | TOV Dental<br>Implant System | OsseOne Dental<br>Implant System<br>K182293<br>Predicate Device | Surgikor's Dental<br>Implant System<br>K182615<br>Reference Device |
|----------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Indications for<br>Use | The TOV Dental<br>Implant System<br>are endosseous<br>implants intended<br>to be surgically<br>placed in the<br>upper or lower<br>jaw arches to<br>provide support<br>for prosthetic<br>devices, such as an<br>artificial tooth, in<br>order to restore<br>patient's esthetics<br>and chewing | OsseOne Dental<br>Implants are<br>endosseous<br>implants intended<br>to be surgically<br>placed in the<br>upper or lower<br>jaw arches to<br>provide support<br>for prosthetic<br>devices, such as an<br>artificial tooth, in<br>order to restore<br>patient's esthetics<br>and chewing | Surgikor's Dental<br>Implant System is<br>indicated for use<br>in surgical and<br>restorative<br>applications for<br>placement in the<br>bone of the upper<br>or lower jaw to<br>provide support<br>for prosthetic<br>devices, such as<br>artificial teeth, in<br>order to restore<br>the patient's |
| | function. Implanet<br>implants are<br>intended for single<br>or multiple unit<br>restorations on<br>splinted or non-<br>splinted<br>applications.<br>Maer, Ragil and<br>TCX are intended<br>for immediate<br>loading when<br>good primary<br>stability is<br>achieved, and with<br>appropriate<br>occlusive loading.<br>These implants<br>can also be used<br>for loading after a<br>conventional<br>healing period.<br>Ragil 3.3 implants<br>are intended to<br>replace a lateral<br>incisor in the<br>maxilla and/or a<br>central or lateral<br>incisor in the<br>mandible.<br>Mandibular<br>central and lateral<br>incisors must be<br>splinted if using<br>two or more 3.3<br>implants adjacent<br>to one another. | function. OsseOne<br>implants are<br>intended for single<br>or multiple unit<br>restorations on<br>splinted or non-<br>splinted<br>applications.<br>OsseoPlus and<br>OsseoLock are<br>intended for<br>immediate loading<br>when good<br>primary stability is<br>achieved, and with<br>appropriate<br>occlusive loading.<br>These implants<br>can also be used<br>for loading after a<br>conventional<br>healing period.<br>OsseoLock 3.3<br>implants are<br>intended to<br>replace a lateral<br>incisor in the<br>maxilla and/or a<br>central or lateral<br>incisor in the<br>mandible.<br>Mandibular<br>central and lateral<br>incisors must be<br>splinted if using<br>two or more 3.3<br>implants adjacent<br>to one another. | chewing function.<br>The Dental<br>Implant System is<br>indicated also for<br>immediate loading<br>when good<br>primary stability is<br>achieved and with<br>appropriate<br>occlusal loading.<br>The 7mm implants<br>are intended to be<br>used in the molar<br>region. |
| | | | |
| Implant<br>Diameters | Maer 3.5, 3.75,<br>4.2, 5.0, 6.0 mm<br>Ragil 3.3, 3.75, 4.2,<br>5.0, 6.0 mm<br>TCX 3.5, 4.3, 5.0<br>mm NP = 3.5 RP = | OsseoLock 3.3,<br>3.75, 4.2, 5.0,<br>6.0mm<br>OsseoPlus 3.5,<br>3.75, 4.2, 5.0,<br>6.0mm | Versatile Hex 3.5,<br>3.75, 4.2, 4.5, 5.0,<br>6.0mm<br>Immediate Hex<br>3.5, 3.75,4.2, 4.5,<br>5.0, 6.0, 7.0mm |
| | | | Fixation Narrow<br>Platform 3.0mm<br>Fixation Regular<br>Platform 3.5,<br>3.9mm<br>Fixation Wide<br>Platform 4.3,<br>5.0mm |
| Implant Lengths | Maer 8, 10, 11.5,<br>13,16mm (no<br>16mm in 6.0mm<br>diameter)<br><br>Ragil 8, 10, 11.5,<br>13, and 16mm (no<br>16mm in 5.0 or<br>6.0mm diameter)<br><br>TCX 8, 10, 11.5,<br>13,16mm | 8, 10, 11.5, 13, 16<br>(OsseoLock 5.0<br>and 6.0 not in<br>16mm) | Versatile Hex 8,<br>10, 11.5, 13, 16,<br>18 (4.2 only), 20<br>(4.2 only)mm<br>Immediate Hex 8<br>(no 3.5 diameter)<br>10, 11.5, 13,<br>16mm 7.00<br>diameter not in<br>11.5, 13 or 16mm.<br><br>Fixation Narrow<br>Platform 10, 11.5,<br>13, 15mm<br>Fixation Regular<br>Platform 8.5, 10,<br>11.5, 13, 15,<br>18mm<br>Fixation Wide<br>Platform 8.5, 10,<br>11.5, 13, 15,<br>18mm<br>Solution5 Narrow<br>Platform 10,11.5,<br>13, 15mm<br>Solution5 Regular<br>Platform 3.5,<br>4.0mm<br>Solution5 Wide<br>Platform 4.5, 5.0,<br>5.5, 6.0mm<br>Solution2 3.25mm |
| | | | |
| | | | diameter only),<br>8.5, 10, 11.5, 13,<br>15mm<br>Solution5 Wide<br>Platform 7.0, 8.5,<br>10, 11.5, 13,<br>15mm<br>Solution2 10,<br>11.5,13 16mm |
| Material of<br>devices included<br>in the submission | Ti-6AL-4V ELI | Ti-6AL-4V ELI | Ti-6AL-4V ELI |
| Type of abutment<br>and maximum<br>angulation | Pre-manufactured<br>of no more than<br>30° | Pre-manufactured<br>of no more than<br>30° | Pre-manufactured<br>of no more than<br>30° |
| Interface<br>type/shape | Internal hex,<br>conical | Internal hex | Internal hex,<br>conical |
| ISO 14801 Fatigue<br>Testing | Sufficient run out<br>load for their<br>intended use | Sufficient run out<br>load for their<br>intended use | Sufficient run out<br>load for their<br>intended use |
| Surface<br>Treatment | SLA | SLA | HA blasted and<br>double acid<br>etched |
| Post Surface<br>Treatment<br>Cleanliness<br>Demonstrated | Yes | Yes | Yes |
{9}------------------------------------------------
{10}------------------------------------------------
{11}------------------------------------------------
| Cover screw | Cover screw for IH<br>NP and RP | Cover Screw for<br>IH | Cover screws NP<br>and RP | Reference Device for<br>individual devices |
|--------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------|
| Multi-Unit<br>Abutments in<br>IH, NP and RP | 4.8 mm diameter<br>multi-units in IH,<br>NP and RP with<br>cuff heights of<br>1,2,3,4,5mm | 4.5 mm multi-unit<br>abutments in<br>heights of 1,2,3<br>and 4 mm | 5.00 mm multi-<br>unit abutments in<br>heights of 1,2,3<br>and 4 mm | Straumann K171784<br>straight multi-unit in<br>gingival height of<br>5.5mm |
| 17° and 30°<br>Angled Multi-<br>Unit<br>Abutments<br>IH, NP and RP | 17° and 30°<br>Angled Multi-Unit<br>Abutments IH, NP<br>and RP with<br>platform heights<br>of 1,2 mm | 17° and 30°<br>overdenture<br>abutments in<br>heights of 1,2,<br>and 3 mm | 18° and 30° Multi-<br>unit abutments in<br>platform heights<br>of 1,2,3 and 4 mm | |
| Locator<br>Abutments<br>with metal | Double Loc /<br>Conical Retentor<br>attachments IH | 3.85mm diameter<br>Flat abutment in | 3.85mm diameter<br>Locator<br>abutments in | |
| housing and<br>retention cap<br>allowing 20°<br>divergence of<br>the implants | (3.85mm), NP and<br>RP (3.9mm)<br>Cuff heights of 1,<br>2, 3, 4, 5, 6 mm (5<br>and 6 only in IH) | heights of 1,2,3,4,<br>5, and 6 mm | heights of<br>1,2,3,4,5, and<br>6mm | |
| | Ti6AL4V ELI<br>housing | Ti6AL4V ELI<br>housing | Ti6AL4V ELI<br>housing | |
| | Polyamide,<br>polyether or<br>polyoxymethylene<br>retention cap with<br>retention levels<br>from 0.6to 1.5 kg | Polyamide,<br>polyether or<br>polyoxymethylene<br>retention cap with<br>retention levels<br>from 0.6to 1.5 kg | Polyamide,<br>polyether or<br>polyoxymethylene<br>retention cap with<br>retention levels<br>from 0.6to 1.5 kg | |
| Ball<br>attachments<br>with metal<br>housing and<br>retention cap<br>allowing 14°<br>divergence of<br>the implants | ball attachments<br>IH (4.00mm<br>diameter), NP<br>(3.5mm diameter)<br>and RP (5.0mm<br>diameter)<br>Cuff heights of 2,<br>3, 4, 5, 6, 7 mm<br>for IH and<br>1.2.3.4.5.6 mm<br>for NP and RP | 4.00 mm Ball<br>attachments in<br>platform heights<br>of 2,3,4,5, and<br>6mm | 4.1 mm Ball<br>attachments in<br>heights of<br>1,2,3,4,5, and<br>6mm | |
| | Stainless steel 316<br>housing | Stainless steel 316<br>housing | Polyamide or<br>polyether<br>retention cap with<br>retention levels<br>from 0.5 to 2.7 kg | |
| | Polyamide or<br>polyether<br>retention cap with<br>retention levels<br>from 0.5 to 1.3 kg | Polyamide or<br>polyether<br>retention cap with<br>retention levels<br>from 0.5 to 2.7 kg | | |
| Healing Caps<br>3.8mm | IH 3.8mm<br>diameter healing<br>cap in 3,4,5,6,7<br>mm cuff height<br><br>3.75 diameter NP<br>healing cap in<br>2,3,4,5 mm cuff<br>heights | 3.8mm diameter<br>Healing Cap in<br>3,4,5,6,7 mm<br>platform height | Cortex Dental<br>Implants K163385<br>healing abutment<br>3.8mm with 2mm<br>platform height | |
| Healing Caps<br>4.6 diameter<br>standard | IH 4.5mm<br>diameter healing<br>cap in 2,3,4,5,6,7<br>mm cuff height | 4.5mm diameter<br>Healing cap in<br>2,3,4,5,6, and<br>mm cuff height | 4.5mm diameter<br>Healing cap in<br>2,3,4,5,6mm | |
| | | | | |
| Healing Caps<br>5.5 diameter wide | 4.5 mm diameter<br>RP healing cap<br>with cuff heights<br>2,3,4,5 mm. | 7mm platform height | | |
| | IH 5.5mm<br>diameter healing<br>abutment in<br>2,3,4,5,6,7 mm<br>cuff height | 5.5mm diameter<br>Healing cap in<br>2,3,4,5,6 and<br>7mm platform height | | |
| Healing Cap<br>6.3mm diameter | IH 6.3mm<br>diameter healing<br>abutment in<br>2,3,4,5 mm cuff<br>height | | Wide emergence<br>healing cap in<br>2,3,4,5,6mm<br>height | |
| Straight<br>Abutment | IH 4.5mm<br>diameter<br>abutment with<br>heights above<br>platform of 5,<br>7,9,11,13,15 mm<br><br>NP and RP 4.5mm<br>diameter with<br>heights above<br>platform of 9 and<br>13 mm | | 4.9mm Non-<br>shouldered<br>Standard Titanium<br>Abutment with<br>heights of 5,7, 9,<br>12, and 15 mm | |
| Straight<br>Narrow<br>Abutment | IH 3.75mm<br>diameter<br>abutment with<br>total heights of<br>5,7,9, 11 mm | IH 3.75mm<br>diameter<br>abutment with<br>total heights of<br>5,7,9, 11 mm | | |
| Straight Wide<br>Abutment | IH 5.5mm<br>diameter<br>abutment with<br>total heights of<br>9,11 or 13 mm | 5.5mm diameter<br>abutment with<br>total heights of<br>9,11 or 13 mm | | |
| Shoulder<br>Abutment | 4.5 diameter IH,<br>NP, RP shoulder<br>abutment with<br>shoulder heights<br>of 1,2,3,4 mm (4<br>only in IH)<br>IH total heights of<br>10.9, 11.9, 12.9, | 4.5mm diameter<br>Standard Wide<br>Shoulder<br>Abutment with<br>heights of 1,2,3,4<br>mm and total<br>heights of 10.9,<br>11.9, 12.9, 13.9 | 4.3mm diameter<br>Standard Wide<br>Shoulder<br>Abutment with<br>heights of 1,2,3,<br>and 4mm and<br>total heights of<br>9.5, 10.5, 11.5, | |
| | NP and RP total<br>heights of 11.8,<br>12.6, and 13.6mm | 5.4mm diameter<br>Standard Wide<br>shoulder<br>abutment with<br>shoulder heights<br>of 1.2.3 mm and<br>total heights of<br>10.7, 11.7 and<br>12.7mm | 6.9mm diameter<br>wide emergence<br>abutment total<br>heights of<br>5,7,9,12,15 mm | |
| | 5.4mm diameter<br>IH shoulder<br>abutment with<br>shoulder heights<br>of 1.2.3.4 mm and<br>total heights of<br>10.7, 11.7, 12.17<br>and 13.7 mm | | | |
| 15° Angled<br>Abutment | 3.75mm diameter<br>IH 15° angled<br>abutment with<br>total heights of 9<br>or 11.4 mm | 3.5mm diameter<br>Standard 15°<br>Abutment with<br>heights of 1,2,3<br>mm | 3.75mm diameter<br>Standard 15°<br>Abutment with<br>lengths of<br>9,11,13mm | |
| 25° Angled<br>Abutment | 3.75mm diameter<br>IH 25° angled<br>abutment with<br>total height of 8.5<br>mm | 3.5mm diameter<br>Standard 25°<br>Abutment with<br>heights of 1,2,3<br>mm | 3.75mm diameter<br>Standard 25°<br>Abutment with<br>lengths of<br>9,11,13mm | |
| Shouldered<br>15° Angled<br>Abutment<br>Hex, Conical<br>NP, and RP | IH 3.75mm<br>diameter 15°<br>Angled Shoulder<br>Abutment with<br>cuff heights of<br>1,2,3,4 mm and<br>height above low<br>side of shoulder<br>8,9,10,11 mm<br><br>NP and RP 15°<br>Angled Shoulder<br>Abutment with<br>cuff heights of<br>1,2,3 mm and<br>total heights of<br>10.7, 12.2 and<br>13.7 mm & 11, 12,<br>13 mm | 3.75mm diameter<br>15° Angled<br>Shoulder<br>Abutment with<br>cuff heights of<br>1,2,3 mm and<br>height above low<br>side of shoulder<br>8,9,10 mm | 3.75mm diameter<br>Anatomic<br>Shouldered<br>Standard 15°<br>Abutment with<br>collar heights of<br>1,2, or 3mm | Cortex Dental<br>Implants K163385<br>angled abutment<br>anatomic/shouldered<br>3.75mm with 4mm<br>collar |
| Shouldered<br>25° Angled<br>Abutment<br>Hex, Conical<br>NP and RP | IH 3.75mm<br>diameter 25°<br>Angled Shoulder<br>Abutment with<br>cuff heights of<br>1,2,3,4 mm with | 3.75mm diameter<br>25° Angled<br>Shoulder<br>Abutment with<br>cuff heights of<br>1,2,3 mm with | 3.75mm diameter<br>Anatomic<br>Shouldered<br>Standard 25°<br>Abutment with | Cortex Dental<br>Implants K163385<br>angled abutment<br>anatomic/shouldered<br>3.75mm with 4mm<br>collar |
| | heights above platform of<br>8.3,9.2,10.3,10.3 mm<br><br>NP and RP 25°<br>Angled Shoulder<br>Abutment with<br>cuff heights of<br>1,2,3 mm with<br>total heights of<br>10.7, 12.2, and<br>13.7 mm & 11, 12,<br>13mm | heights above platform of<br>8.3,9.2,10.3 mm | collar heights of<br>1,2, or 3mm | |
| UCLA Base Ti | 4.5mm diameter<br>UCLA base in<br>hexed | 4.5 mm diameter<br>UCLA base hexed | 4.5mm hexed and<br>non-hexed | |
{12}------------------------------------------------
{13}------------------------------------------------
{14}------------------------------------------------
{15}------------------------------------------------
## Conclusion:
TOV Dental Implant System is substantially equivalent to OsseOne Dental Implant System. They both have the same indications for use, are of the same material, and have internal hex connections. The abutments, healing caps, and angled abutments are offered in similar designs and heights. Any abutment designs not found within the OsseOne Dental Implant System were found in the reference devices which have the same materials, similar indications for use and same internal hex and conical connections as the TOV Dental Implant System. TOV and OsseOne have the same types of performance testing with similar results. Performance testing demonstrates substantial equivalence to the identified predicate devices.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.