K240752 · El.En S.P.A. · GEX · May 16, 2024 · General, Plastic Surgery
Device Facts
Record ID
K240752
Device Name
DEKA TORO
Applicant
El.En S.P.A.
Product Code
GEX · General, Plastic Surgery
Decision Date
May 16, 2024
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 878.4810
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
785 nm laser source Intended for removal of tattoos for Fitzpatrick skin types II-IV to treat the following tattoo colors: green and blue. 1064 & 532 nm laser sources in Q-Switched, nanosecond mode Intended for treatment of benign vascular lesions, benign pigmented lesions, and for hair, tattoo removal and the incision, excision, ablation, vaporization of soft tissue for General dermatology such as, but not limited to: 532 nm (Q-Switched, nanosecond mode), including microbeam handpieces: - Removal of light ink (red, sky blue, green, tan, purple, and orange) tattoos - Treatment of benign vascular lesions including, but not limited to: - port wine birthmarks - telangiectasias - spider angioma - Cherry angioma - Spider nevi - Treatment of benign pigmented lesions including, but not limited to: - cafe-au-lait birthmarks - Ephalides, solar lentigines - senile lentigines - Becker's nevi - freckles - common nevi - nevus spilus - Ota Nevus - Treatment of seborrheic keratosis - Treatment of post inflammatory hyperpigmentation - Skin resurfacing procedures for the treatment of acne scars and wrinkles. 1064 nm (Q-Switched, nanosecond mode), including microbeam handpieces: - Removal of dark ink (black, blue and brown) tattoos - Removal of benign pigmented lesions including; - nevus of Ota - Café au lait spot - Ephalides, solar lentigo (lentigines) - Becker Nevus - Nevus spilus - Treatment of common nevi - Removal or lightening of unwanted hair - Skin resurfacing procedures for the treatment of acne scars and wrinkles 1064nm laser source in thermal mode (long-pulse) Intended for: - Treatment of wrinkles - Treatment of mild to moderate inflammatory acne vulgaris
Device Story
DEKA Toro is a multi-mode medical laser system used in clinical dermatology settings by trained professionals. It utilizes Nd:YAG (532 nm, 1064 nm) and Ti:Sapphire (785 nm) laser sources. The device operates in Q-switched nanosecond, picosecond, and long-pulse (thermal) modes. Laser energy is delivered to the patient via an articulated arm and handpiece. The clinician selects the wavelength and pulse mode based on the target condition (e.g., tattoo ink color, vascular lesion type, or skin resurfacing needs). The laser energy interacts with target chromophores (pigment, hemoglobin, or tissue) to achieve therapeutic effects such as tattoo fragmentation, lesion ablation, or tissue heating. The device output is controlled by the operator to achieve clinical goals like wrinkle reduction or acne treatment. It provides a versatile platform for various dermatological procedures, benefiting patients by offering non-invasive or minimally invasive options for skin rejuvenation and lesion removal.
Clinical Evidence
No clinical data provided; bench testing only.
Technological Characteristics
Laser system with Nd:YAG (532/1064 nm) and Ti:Sapphire (785 nm) sources. Modes: Q-switched (nanosecond), picosecond, and long-pulse (thermal). Beam delivery via articulated arm and handpieces. Standards: ANSI AAMI ES60601-1, IEC 60601-1-2, IEC 60601-2-22, IEC 60825-1.
Indications for Use
Indicated for tattoo removal (Fitzpatrick II-IV, specific colors), benign vascular/pigmented lesion treatment, hair removal, and soft tissue incision/excision/ablation/vaporization in general dermatology. Includes treatment of acne scars, wrinkles, and inflammatory acne vulgaris.
Regulatory Classification
Identification
(1) A carbon dioxide laser for use in general surgery and in dermatology is a laser device intended to cut, destroy, or remove tissue by light energy emitted by carbon dioxide.(2) An argon laser for use in dermatology is a laser device intended to destroy or coagulate tissue by light energy emitted by argon.
{1}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
May 16, 2024
El.En. S.p.A. Paolo Peruzzi Regulatory Affairs Manager Via Baldanzese 17 Calenzano, FI 50141 Italy
Re: K240752
Trade/Device Name: Deka Toro Regulation Number: 21 CFR 878.4810 Regulation Name: Laser Surgical Instrument For Use In General And Plastic Surgery And In Dermatology Regulatory Class: Class II Product Code: GEX Dated: March 19, 2024 Received: March 19, 2024
Dear Paolo Peruzzi:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{2}------------------------------------------------
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn).
{3}------------------------------------------------
Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/deviceadvice-comprehensive-regulatory-assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely.
Digitally signed Tanisha L. by Tanisha L. Hithe -S Hithe -S Date: 2024.05.16 12:49:41 -04'00'
Tanisha Hithe Assistant Director DHT4A: Division of General Surgery Devices OHT4: Office of Surgical & Infection Control Devices, Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{4}------------------------------------------------
# Indications for Use
510(k) Number (if known) K240752
Device Name Deka Toro
# Indications for Use (Describe)
785 nm laser source
Intended for removal of tattoos for Fitzpatrick skin types II-IV to treat the following tattoo colors: green and blue.
1064 & 532 nm laser sources in Q-Switched, nanosecond mode
Intended for treatment of benign vascular lesions, benign pigmented lesions, and for hair, tattoo removal and the incision, excision, ablation, vaporization of soft tissue for General dermatology such as, but not limited to:
532 nm (Q-Switched, nanosecond mode), including microbeam handpieces:
- · Removal of light ink (red, sky blue, green, tan, purple, and orange) tattoos
- · Treatment of benign vascular lesions including, but not limited to:
- port wine birthmarks
- telangiectasias
- spider angioma
- Cherry angioma
- Spider nevi
- · Treatment of benign pigmented lesions including, but not limited to:
- cafe-au-Iait birthmarks
- Ephalides, solar lentigines
- senile lentigines
- Becker's nevi
- freckles
- common nevi
- nevus spilus
- Ota Nevus
- · Treatment of seborrheic keratosis
- · Treatment of post inflammatory hyperpigmentation
- · Skin resurfacing procedures for the treatment of acne scars and wrinkles.
1064 nm (Q-Switched, nanosecond mode), including microbeam handpieces:
- · Removal of dark ink (black, blue and brown) tattoos
- · Removal of benign pigmented lesions including;
- nevus of Ota
- Café au lait spot
- Ephalides, solar lentigo (lentigines)
- Becker Nevus
- Nevus spilus
- · Treatment of common nevi
- · Removal or lightening of unwanted hair
- · Skin resurfacing procedures for the treatment of acne scars and wrinkles
1064nm laser source in thermal mode (long-pulse) Intended for:
- Treatment of wrinkles
- · Treatment of mild to moderate inflammatory acne vulgaris
{5}------------------------------------------------
X Prescription Use (Part 21 CFR 801 Subpart D)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{6}------------------------------------------------
# 510(k) Summary # K240752
## DEKA TORO
### Submitter:
El.En. S.p.A. Via Baldanzese, 17 50041 Calenzano (FI), Italy
### Contact:
Paolo Peruzzi Regulatory Affairs Manager & Official Correspondent Phone: +39.055.8826807 E-mail: p.peruzzi@elen.it
### Date Summary Prepared:
March 19, 2024
### Device Trade Name:
DEKA TORO
### Common Name:
Medical Laser system
#### Regulation Name:
Laser Surgical Instrument For Use In General And Plastic Surgery And In Dermatology
#### Product Code:
GEX
#### Regulatory Class:
Class II
#### Regulation Number:
21 CFR 878.4810
## Predicate Devices:
QUANTA SYSTYEM CHROME (K202503)
LUTRONIC HOLLYWOOD SPECTRA (K213569)
LASEROPTEK HELIOS 785 Pico (K230373)
{7}------------------------------------------------
# Device Description:
TORO device is a DEKA laser system with a wide choice of emission modes, ranging from picosecond to hundreds of microseconds of pulse duration.
The available sources are Nd:YAG lasers at 532 nm and 1064 nm, operating in Q-switched mode, and Ti:Sapphire laser at 785 nm operating in the Picosecond pulse mode.
The Q-switched laser at 1064 nm can also operate in Thermal mode, delivering pulses of hundreds of microseconds.
The laser beam generated by the source is then optically coupled to the articulated arm with a handpiece plugged at its end and delivered to the Patient.
## Indications for Use:
#### 785 nm laser source
Intended for removal of tattoos for Fitzpatrick skin types II-IV to treat the following tattoo colors: green and blue.
#### 1064 & 532 nm laser sources in Q-Switched, nanosecond mode
Intended for treatment of benign vascular lesions, benign pigmented lesions, and for hair, tattoo removal and the incision, excision, ablation, vaporization of soft tissue for General dermatology such as, but not limited to:
532 nm (Q-Switched, nanosecond mode), including microbeam handpieces:
- Removal of light ink (red, sky blue, green, tan, purple, and orange) tattoos
- . Treatment of benign vascular lesions including, but not limited to:
- । port wine birthmarks
- telangiectasias ।
- । spider angioma
- Cherry angioma ー
- l Spider nevi
- Treatment of benign pigmented lesions including, but not limited to: ●
- cafe-au-lait birthmarks
- Ephalides, solar lentigines —
- senile lentigines
- -Becker's nevi
- freckles
- common nevi
- nevus spilus
- Ota Nevus l
- Treatment of seborrheic keratosis
- Treatment of post inflammatory hyperpigmentation
- Skin resurfacing procedures for the treatment of acne scars and wrinkles.
1064 nm (Q-Switched, nanosecond mode), including microbeam handpieces:
- . Removal of dark ink (black, blue and brown) tattoos
{8}------------------------------------------------
- Removal of benign pigmented lesions including; ●
- । nevus of Ota
- Café au lait spot ।
- । Ephalides, solar lentigo (lentigines)
- | Becker Nevus
- । Nevus spilus
- Treatment of common nevi
- . Removal or lightening of unwanted hair
- Skin resurfacing procedures for the treatment of acne scars and wrinkles
## 1064nm laser source in thermal mode (long-pulse)
Intended for:
- Treatment of wrinkles
- Treatment of mild to moderate inflammatory acne vulgaris
# Comparison with The Predicate Device:
### 785nm laser source
| | Proposed device<br>DEKA Toro | K230373<br>HELIOS 785 Pico<br>Ti:Sapphire laser (785nm) | COMMENTS |
|--------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------|
| INDICATIONS FOR USE | Removal of tattoos for Fitzpatrick<br>skin types II-IV to treat the<br>following tattoo colors: green and<br>blue. | Removal of tattoos for Fitzpatrick<br>skin types II-IV to treat the<br>following tattoo colors: green and<br>blue. | Identical |
| Product Code | GEX | GEX | Identical |
| Classification | Class II | Class II | Identical |
| Regulation Number | 21 CFR 878.4810 | 21 CFR 878.4810 | Identical |
| Wavelength (nm) | 785nm ± 25 nm | 785nm ±20% | Identical |
| Pulse Duration | 600 - 900 ps | 600 ps | Similar |
| Pulse Energy (max) | 0.2 J | 0.2 J | Identical |
| Fluence (max) | 4 J/cm² | 4 J/cm² | Identical |
| Max Peak Power | 0.33 GW | 0.33 GW | Identical |
| Max Average Power | 1 W | 2 W | Subset, within the<br>range of the predicate |
| Spot Size (mm) | 2-6 mm | 1-8 mm | Subset, within the<br>range of the predicate |
| Repetition Rate (Hz) | 1-5 Hz | 1-10 Hz | Subset, within the<br>range of the predicate |
| Beam Delivery System | Articulated Arm with Handpiece | Articulated Arm with Handpiece | Identical |
| | Proposed device<br>DEKA TORO | K202503 CHROME<br>1064 & 532 nm (Q-Switched, nanosecond<br>mode) | COMMENTS |
| INDICATIONS<br>FOR USE | Treatment of benign vascular lesions,<br>benign pigmented lesions, and for hair,<br>tattoo removal and the incision, excision,<br>ablation, vaporization of soft tissue for<br>General dermatology such as, but not<br>limited to:<br>532 nm including microbeam handpieces:<br>Removal of light ink (red, sky blue, green,<br>tan, purple, and orange) tattoos<br>Treatment of benign vascular lesions<br>including, but not limited to:<br>- port wine birthmarks<br>- telangiectasias<br>- spider angioma<br>- Cherry angioma<br>- Spider nevi<br>Treatment of benign pigmented lesions<br>including, but not limited to:<br>- cafe-au-lait birthmarks<br>- Ephalides, solar lentigines<br>- senile lentigines<br>- Becker's nevi<br>- freckles<br>- common nevi<br>- nevus spilus<br>- Ota Nevus<br>Treatment of seborrheic keratosis<br>Treatment of post inflammatory<br>hyperpigmentation<br>Skin resurfacing procedures for the<br>treatment of acne scars and wrinkles.<br>1064 nm including microbeam<br>handpieces:<br>Removal of dark ink (black, blue and<br>brown) tattoos<br>Removal of benign pigmented lesions<br>including;<br>- nevus of Ota<br>- Café au lait spot<br>- Ephalides, solar lentigo (lentigines)<br>- Becker Nevus<br>- Nevus spilus<br>Treatment of common nevi<br>Removal or lightening of unwanted hair<br>Skin resurfacing procedures for the<br>treatment of acne scars and wrinkles | Treatment of benign vascular lesions,<br>benign pigmented lesions, and for hair,<br>tattoo removal and the incision, excision,<br>ablation, vaporization of soft tissue for<br>General dermatology such as, but not<br>limited to treatment of:<br>532 nm including microbeam handpieces:<br>Removal of light ink (red, sky blue, green,<br>tan, purple, and orange) tattoos<br>Treatment of benign vascular lesions<br>including, but not limited to:<br>- port wine birthmarks<br>- telangiectasias<br>- spider angioma<br>- Cherry angioma<br>- Spider nevi<br>Treatment of benign pigmented lesions<br>including, but not limited to:<br>- cafe-au-lait birthmarks<br>- Ephalides, solar lentigines<br>- senile lentigines<br>- Becker's nevi<br>- freckles<br>- common nevi<br>- nevus spilus<br>- Ota Nevus<br>Treatment of seborrheic keratosis<br>Treatment of post inflammatory<br>hyperpigmentation<br>Skin resurfacing procedures for the<br>treatment of acne scars and wrinkles.<br>1064 nm including microbeam<br>handpieces:<br>Removal of dark ink (black, blue and<br>brown) tattoos<br>Removal of benign pigmented lesions<br>including;<br>- nevus of Ota<br>- Café au lait spot<br>- Ephalides, solar lentigo (lentigines)<br>- Becker Nevus<br>- Nevus spilus<br>Treatment of common nevi<br>Removal or lightening of unwanted hair<br>Skin resurfacing procedures for the<br>treatment of acne scars and wrinkles | Identical |
| | Proposed device<br>DEKA TORO | K202503 CHROME<br>1064 & 532 nm (Q-Switched, nanosecond<br>mode) | COMMENTS |
| Product Code | GEX | GEX | Identical |
| Classification | Class II | Class II | Identical |
| Regulation<br>Number | 21 CFR 878.4810 | 21 CFR 878.4810 | Identical |
| Wavelength | 1064, 532 nm | 1064, 532 nm | Identical |
| Pulse Duration | 5 - 12 ns @1064 nm<br>5 - 12 ns @532 nm | 6 - 12 ns @1064<br>6 - 12 ns @532nm | Similar |
| Pulse Energy<br>(max) | 1.0 J @ 1064 nm<br>0.5 J @ 532 nm | 1.0 J @1064<br>0.5 J @532…
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.