K240743 · Ultradent Products, Inc. · KLE · Mar 20, 2024 · Dental
Device Facts
Record ID
K240743
Device Name
Peak Universal Bond
Applicant
Ultradent Products, Inc.
Product Code
KLE · Dental
Decision Date
Mar 20, 2024
Decision
SESE
Submission Type
Special
Regulation
21 CFR 872.3200
Device Class
Class 2
Attributes
Therapeutic, Pediatric, 3rd-Party Reviewed
Indications for Use
Use for virtually all etching and bonding needs in restorative dentistry for patients of all ages. Peak Universal Bond is used for desensitizing/sealing of tooth structure. Peak Universal Bond adhesive bonds to the following materials: - Dentin and Enamel - Porcelain, Zirconia - Metal - Composite
Device Story
Peak Universal Bond is a resin bonding agent delivered via bottle, syringe, or unit dose. Used in restorative dentistry by dental professionals; supports self-etch or total-etch techniques. Contains 7.5% filler, ethyl alcohol solvent, and 0.2% chlorhexidine. Cures via high-intensity light (e.g., LEDs). Mechanism involves mechanical occlusion of dentin tubules by resins to reduce tooth hypersensitivity. Provides adhesion to dentin, enamel, porcelain, zirconia, metal, and composite. Clinical benefit includes long-term bond strength and reduction of tooth sensitivity.
Clinical Evidence
Bench testing only. Verification testing performed per ISO 29022:2013. SEM imaging used to confirm dentinal tubule blockage. Simulated use and validation testing conducted by dental professionals confirmed device meets user needs and indications for use.
Technological Characteristics
Resin bonding agent; 7.5% filled with ethyl alcohol solvent; contains 0.2% chlorhexidine. Cures via high-intensity light. Tested to ISO 29022:2013. Delivered in bottle, syringe, or unit dose.
Indications for Use
Indicated for patients of all ages requiring etching and bonding in restorative dentistry, including desensitizing and sealing of tooth structure. Bonds to dentin, enamel, porcelain, zirconia, metal, and composite.
Regulatory Classification
Identification
A resin tooth bonding agent is a device material, such as methylmethacrylate, intended to be painted on the interior of a prepared cavity of a tooth to improve retention of a restoration, such as a filling.
{1}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
March 20, 2024
Ultradent Products, Inc. % Dave Yungvirt CEO Third Party Review Group, LLC 25 Independence Blvd Warren, New Jersey 07059
Re: K240743
Trade/Device Name: Peak Universal Bond Regulation Number: 21 CFR 872.3200 Regulation Name: Resin tooth bonding agent Regulatory Class: Class II Product Code: KLE Dated: March 18, 2024 Received: March 18, 2024
Dear Dave Yungvirt:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{2}------------------------------------------------
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely.
Michael E. Adjodha -S
Michael E. Adjodha, MChE, RAC, COIA Assistant Director DHT1B: Division of Dental and ENT Devices OHT1: Office of Ophthalmic. Anesthesia. Respiratory, ENT and Dental Devices Office of Product Evaluation and Ouality Center for Devices and Radiological Health
Enclosure
{3}------------------------------------------------
# Indications for Use
510(k) Number (if known) K240743
Device Name Peak Universal Bond
Indications for Use (Describe)
Use for virtually all etching and bonding needs in restorative dentistry for patients of all ages. Peak Universal Bond is used for desensitizing/sealing of tooth structure.
Peak Universal Bond adhesive bonds to the following materials:
- · Dentin and Enamel
- Porcelain, Zirconia
- · Metal
- · Composite
| Type of Use (Select one or both, as applicable) | |
|-------------------------------------------------|--|
|-------------------------------------------------|--|
X Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{4}------------------------------------------------
# 510(k) Summary
## Contact Details
Applicant Name: Ultradent Products, Inc. Applicant Address: 505 West Ultradent Drive, South Jordan, UT, 84095 Applicant Contact Telephone: 801-557-1125 Applicant Contact: Dr. Carolyn Taylor Applicant Contact Email: Carolyn.taylor(@ultradent.com
### Device Name
Device Trade Name: Peak Universal Bond Common Name: Resin tooth bonding agent Classification Name: Agent, Tooth Bonding, Resin Regulation Number: 872.3200 Product Code(s): KLE
#### Legally Marketed Predicate Devices
K100752 Peak Universal Bond - KLE K230465 Hi-Bond Universal - KLE
#### Device Description Summary
Peak™M Universal Bond is a bottle, syringe, or unit dose delivered bonding resin. It can be used with selfetch or with total etch techniques. It is 7.5% filled with an ethyl alcohol solvent carrier and will cure with most high intensity lights including LEDs. Peak Universal Bond adhesive contains 0.2% chlorhexidine which may ensure long term bond strengths. Its mode of action produces a mechanical occlusion of dentin tubules by the resins, thus reducing tooth hypersensitivity.
#### Intended Use/Indications for Use
Use for virtually all etching and bonding needs in restorative dentistry for patients of all ages. Peak Universal Bond is used for desensitizing/sealing of tooth structure.
Peak Universal Bond adhesive bonds to the following materials:
- · Dentin and Enamel
- · Porcelain, Zirconia
- Metal
- · Composite
### Indications for Use Comparison
The indications for use between the subject device and primary predicate are identical with the addition of zirconia and desensitizing/sealing of tooth structure. The secondary predicate contains both of these indications for use.
#### Technological Comparison
The subject device and primary predicate are identical structure and technological characteristics. The only change in this Special 510(k) application between the primary predicate and the subject device is in the intended use. The subject device includes the indications for use of bonding to zirconia as well as desensitizing/sealing the tooth structure. The secondary predications
{5}------------------------------------------------
for use and is tested to many of the same standards as the subject device. There are some minor differences in formulation between the subject device and secondary predicate, but the standard of testing remains the same. The subject device shear bond testing has been performed to ISO 29022 and adheres to this standard. The shear bond testing of the secondary predicate was performed to ISO 11405. This standard has been withdrawn, but follows the same principles as ISO 29022 which is the updated standard for testing dental adhesives. The secondary predicate was tested for dentinal tubule blockage. The subject device was also tested for this using SEM imaging of the device applied to the tooth. The physical properties of the subject device, predicate devices, and secondary predicate device are all the same. None of the differences in the devices impact safety or efficacy of the device.
Overall, these slight differences do not introduce new questions related to product risk, nor do they impact the safety or efficacy of the subject device as compared to the predicate product. As such, the subject device can be considered substantially equivalent to the predicate device.
#### Non-Clinical and/or Clinical Tests Summary & Conclusions
Verification and validation testing was performed to show that the subject device is substantially equivalent to the currently marketed predicate device, Peak Universal Bond. Full verification testing to ISO 29022:2013 and SEM imaging were completed to support the additional indications for use of this device to the primary predicate in accordance with the same testing performed to support these indications in the secondary predicate, Hi-Bond Universal K230465.
All additional verification testing was performed in accordance with Risk Management Procedures outline in SOP QA 0028 which was prepared in accordance with ISO 14971. Simulated use and validation testing was conducted by dental professionals to confirm that all aspects from usability assessments and risk management were implemented in the labeling and the design of the device itself. The conclusion of this validation testing confirmed that the subject device met the predefined user needs and the indications for use, which are identical to the primary predicate device and/or the secondary predicate. Thus the verification and validation support the conclusion that the subject device, Peak Universal Bond, is substantially equivalent to the predicate device, Peak Universal Bond (K100752).
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.