K240701 · Brainlab AG · OLO · May 23, 2024 · Neurology
Device Facts
Record ID
K240701
Device Name
Drill Guide; Drill Bit; Spine & Trauma Navigation
Applicant
Brainlab AG
Product Code
OLO · Neurology
Decision Date
May 23, 2024
Decision
SESE
Submission Type
Special
Regulation
21 CFR 882.4560
Device Class
Class 2
Indications for Use
Spine & Trauma Navigation is intended as an intraoperative image-guided localization system to enable open and minimally invasive surgery. It links a freehand probe, tracked by a passive marker sensor system to virtual computer image space on a patient's preoperative or intraoperative 2D or 3D image data. Spine & Trauma Navigation enables computer-assisted navigation of medical image data, which can either be acquired preoperatively or intraoperatively by an appropriate image acquisition system. The software offers screw and interbody device planning and navigation with surgical instruments. The system is indicated for any medical condition in which the use of stereotactic surgery may be appropriate and where a reference to a rigid anatomical structure, such as the skull, the pelvis, a long bone or vertebra can be identified relative to the acquired image (CT, MR, 3D fluoroscopic image reconstruction or 2D fluoroscopic image) and/or an image databased model of the anatomy.
Device Story
Spine & Trauma Navigation is an intraoperative image-guided surgery system for spine and trauma procedures. It inputs preoperative or intraoperative 2D/3D image data (CT, MR, or fluoroscopy) and tracks surgical instruments (drill guides, drill bits, probes) using an optical passive marker sensor system. The software registers patient anatomy to virtual image space, enabling computer-assisted navigation, screw planning, and interbody device placement. Used in OR settings by surgeons; the system provides real-time visual feedback on instrument position relative to patient anatomy. Navigation accuracy is maintained via optical tracking; modified drill guides feature improved handle rotation, magnetic array attachment, and click-interface tube fixation. The system assists surgeons in precise localization and trajectory guidance, potentially reducing surgical invasiveness and improving procedural accuracy. Instruments are stainless steel, delivered non-sterile, and require steam sterilization.
Clinical Evidence
No clinical testing was required. Bench testing included assembly stability, accuracy (tip deviation ≤1.7mm, angular deviation ≤1.7° at 95th percentile), lifecycle assessment, skiving performance, mechanical failure testing for drill bits, and usability testing with 15 representative users. Biocompatibility evaluated per ISO 10993-1; reprocessing validated per ISO 11737-1, ANSI/AAMI ST98, and ISO 15883-5.
Technological Characteristics
Optical tracking system; stainless steel instruments. Drill guides include guide tubes, trocar inserts, handles, arrays, and depth controls. Connectivity via image acquisition systems (CT, MR, fluoroscopy). Sterilization: steam. Software-based navigation platform. Accuracy: ≤2mm positional, ≤2° angular.
Indications for Use
Indicated for patients requiring stereotactic surgery where a rigid anatomical structure (skull, pelvis, long bone, or vertebra) can be referenced to preoperative or intraoperative 2D/3D image data (CT, MR, or fluoroscopy) to enable open or minimally invasive surgical navigation.
Regulatory Classification
Identification
A stereotaxic instrument is a device consisting of a rigid frame with a calibrated guide mechanism for precisely positioning probes or other devices within a patient's brain, spinal cord, or other part of the nervous system.
{0}------------------------------------------------
May 23, 2024
Image /page/0/Picture/1 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
Brainlab AG Esther Moreno Garcia Manager Regulatory Affairs Olof-Palme-Str. 9 Munich, 81829 Germany
Re: K240701
Trade/Device Name: Drill Guide; Drill Bit; Spine & Trauma Navigation Regulation Number: 21 CFR 882.4560 Regulation Name: Stereotaxic Instrument Regulatory Class: Class II Product Code: OLO Dated: March 14, 2024 Received: April 23, 2024
## Dear Esther Moreno Garcia:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
{1}------------------------------------------------
Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Tejen D. Soni -S
For
Shumaya Ali, M.P.H. Assistant Director DHT6C: Division of Restorative, Repair and Trauma Devices OHT6: Office of Orthopedic Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K240701
Device Name Drill Guide; Drill Bit; Spine & Trauma Navigation
#### Indications for Use (Describe)
Spine & Trauma Navigation is intended as an intraoperative image-guided localization system to enable open and minimally invasive surgery. It links a freehand probe, tracked by a passive marker sensor system to virtual computer image space on a patient's preoperative or intraoperative 2D or 3D image data.
Spine & Trauma Navigation enables computer-assisted navigation of medical image data, which can either be acquired preoperatively or intraoperatively by an appropriate image acquisition system. The software offers screw and interbody device planning and navigation with surgical instruments.
The system is indicated for any medical condition in which the use of stereotactic surgery may be appropriate and where a reference to a rigid anatomical structure, such as the skull, the pelvis, a long bone or vertebra can be identified relative to the acquired image (CT, MR, 3D fluoroscopic image reconstruction or 2D fluoroscopic image) and/or an image databased model of the anatomy.
| Type of Use (Select one or both, as applicable) | |
|-------------------------------------------------------------------------------------------------------|---------------------------------------------|
| <span style="text-decoration: underline;"><b></b></span> Prescription Use (Part 21 CFR 801 Subpart D) | Over-The-Counter Use (21 CFR 801 Subpart C) |
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
Image /page/3/Picture/0 description: The image shows the Brainlab logo. The logo consists of a stylized pink brain icon on the left, followed by the word "BRAINLAB" in pink, uppercase letters. The brain icon appears to be made up of several curved lines, giving it a modern and abstract look.
# 510(k) Summary
### March 14th, 2024
| General Information | |
|--------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------|
| Manufacturer | Brainlab AG; Olof-Palme-Str.9, 81829, Munich, Germany |
| Establishment Registration | 8043933 |
| Trade Name | Drill Guide<br>Drill Bit<br>Spine & Trauma Navigation |
| Classification Name | Orthopedic Stereotaxic Instrument |
| Product Code | OLO |
| Regulation Number | 882.4560 |
| Regulatory Class | II |
| Panel | Orthopedic |
| Predicate Device(s) | K221618 - Spine & Trauma Navigation |
| Contact Information | |
| Primary Contact | Alternate Contact |
| Esther Moreno Garcia<br>Manager Regulatory Affairs<br>Phone: +49 89 99 15 68 0<br>Email: regulatory.affairs@brainlab.com | Chiara Cunico<br>Phone: +49 89 99 15 68 0<br>Fax: +49 89 99 15 68 5033<br>Email: chiara.cunico@brainlab.com |
#### 1. Indications for Use
Spine & Trauma Navigation is intended as an intraoperative image-guided localization system to enable open and minimally invasive surgery. It links a freehand probe, tracked by a passive marker sensor system to virtual computer image space on a patient's preoperative or intraoperative 2D or 3D image data.
Spine & Trauma Navigation enables computer-assisted navigation of medical image data, which can either be acquired preoperatively or intraoperatively by an appropriate image acquisition system. The software offers screw and interbody device planning and navigation with surgical instruments.
The system is indicated for any medical condition in which the use of stereotactic surgery may be appropriate and where a reference to a rigid anatomical structure, such as the skull, the pelvis, a long bone or vertebra can be identified relative to the acquired image (CT, MR, 3D fluoroscopic image reconstruction or 2D fluoroscopic image) and/or an image databased model of the anatomy.
#### 2. Device Description
{4}------------------------------------------------
Image /page/4/Picture/0 description: The image shows the Brainlab logo. The logo consists of a stylized pink symbol on the left, resembling a brain or a network, followed by the word "BRAINLAB" in pink, block letters. The overall design is clean and modern, suggesting a focus on technology and innovation in the field of neuroscience or medical technology.
The Spine & Trauma Navigation is an image quided surgery system for navigated treatments in the fields of spine and trauma surgery, whereas the user may use image data based on CT, MR, 3D fluoroscopic image reconstruction (cone beam CT) or 2D fluoroscopic images. It offers different patient image registration methods and instrument calibrations to allow surgical navigation by using optical tracking technology. To fulfil this purpose, it consists of software, Image Guided Surgery platforms and surgical instruments.
Modified Drill Guides and Drill Bits have been introduced as part of the Subject Device. The Drill Guide instruments are navigated instruments which support the surgeon in guiding drill bits and K-wires during spinal procedures. They consist of a guide tube, a trocar insert (both available in five different diameters), a body with two available handles, an array and a depth control (available in two different sizes for various drilling depths). The Drill Guide Tubes and Drill Guide Trocar Inserts have patient contact. All instruments are delivered unsterile and require end user sterilization.
The Drill Bits are used for drilling of bone. They are made of stainless steel and are delivered non-sterile. They require steam sterilization onsite before use. There are several variants in terms of diameter, length, and presence of a depth stop feature.
### 3. Substantial Equivalence
For the Substantial Equivalence determination, comparison of the Subject Device features with the following predicate device(s) was carried out:
K221618 - Spine & Trauma Navigation (product code: OLO)
The predicate device was chosen since it's the predecessor device and therefore very similar to the subject device w.r.t the indications for use, technological characteristics and use cases.
The main difference compared to the predicate device are modified instruments.
| Feature | Predicate Device K221618 | Subject Device |
|--------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Indications<br>for use | Spine & Trauma Navigation is<br>intended as an intraoperative image-<br>guided localization system to enable<br>open and minimally invasive surgery.<br>It links a freehand probe, tracked by<br>a passive marker sensor system to<br>virtual computer image space on a<br>patient's preoperative or<br>intraoperative 2D or 3D image data.<br><br>Spine & Trauma Navigation enables<br>computer-assisted navigation of<br>medical image data, which can either | Spine & Trauma Navigation is<br>intended as an intraoperative image-<br>guided localization system to enable<br>open and minimally invasive surgery.<br>It links a freehand probe, tracked by<br>a passive marker sensor system to<br>virtual computer image space on a<br>patient's preoperative or<br>intraoperative 2D or 3D image data.<br><br>Spine & Trauma Navigation enables<br>computer-assisted navigation of<br>medical image data, which can either |
| Feature | Predicate Device K221618 | Subject Device |
| | be acquired preoperatively or<br>intraoperatively by an appropriate<br>image acquisition system. The<br>software offers screw and interbody<br>device planning and navigation with<br>surgical instruments. | be acquired preoperatively or<br>intraoperatively by an appropriate<br>image acquisition system. The<br>software offers screw and interbody<br>device planning and navigation with<br>surgical instruments. |
| | The system is indicated for any<br>medical condition in which the use of<br>stereotactic surgery may be<br>appropriate and where a reference to<br>a rigid anatomical structure, such as<br>the skull, the pelvis, a long bone or<br>vertebra can be identified relative to<br>the acquired image (CT, MR, 3D<br>fluoroscopic image reconstruction or<br>2D fluoroscopic image) and/or an<br>image databased model of the<br>anatomy. | The system is indicated for any<br>medical condition in which the use<br>of stereotactic surgery may be<br>appropriate and where a reference<br>to a rigid anatomical structure, such<br>as the skull, the pelvis, a long bone<br>or vertebra can be identified relative<br>to the acquired image (CT, MR, 3D<br>fluoroscopic image reconstruction or<br>2D fluoroscopic image) and/or an<br>image databased model of the<br>anatomy. |
| Navigation<br>accuracy | Mean Positional Error of the placed<br>instrument's tip ≤ 2 mm<br>Mean Angular Error of the placed<br>instrument's axis ≤ 2° | Mean Positional Error of the placed<br>instrument's tip ≤ 2 mm<br>Mean Angular Error of the placed<br>instrument's axis ≤ 2° |
| Tracking<br>technology | Optical tracking | Optical tracking |
| Drill guide<br>components | -Drill Guide Handles<br>-Drill Guide Compact Handle<br>-Drill Guide Tubes<br>-Drill Guide Trocars<br>-Tissue Protection Sleeve<br>-Drill Guide Depth Controls | -Drill Guide Handle<br>-Drill Guide Handle Compact<br>-Drill Guide Tubes<br>-Drill Guide Trocar Inserts<br>-Drill Guide Depth Controls<br>-Drill Guide Array |
| Adjustment<br>and<br>handling<br>features of<br>Drill Guides | • Adjusting position of handle<br>requires loosening of the<br>counter nut.<br>• Array is fixed to the handle body<br>• The guide tubes are attached to<br>the main drill guide body by<br>means of a left-handed thread.<br>• Depth control is attached to drill | • Rotatable handle allows to<br>adjust position relative to array<br>without loosening any fixation<br>elements.<br>• Detachable array with integrated<br>magnet to prevent it from<br>dropping. |
| Feature | Predicate Device K221618 | Subject Device |
| | depth setting is secured with a<br>counter nut. | The Guide Tube is attached to<br>the Handle by simple linear<br>insertion and held in place by a<br>click interface. Depth control is attached and<br>depth setting is secured with a<br>snapping mechanism. |
| Drill Bits | Drill bits with diameters between<br>2.4mm to 4.5mm. | Four new drill bits are introduced,<br>three within the same range of<br>diameters (3.2 and 4.5mm). A new<br>worst case drill bit with a diameter of<br>2mm. |
| Cleaning<br>and<br>sterilization | Automated cleaning and automated<br>thermal disinfection<br>Steam sterilization | Automated cleaning and automated<br>thermal disinfection<br>Steam sterilization |
{5}------------------------------------------------
Image /page/5/Picture/0 description: The image shows the Brainlab logo. The logo consists of a stylized pink brain icon on the left, followed by the word "BRAINLAB" in pink, sans-serif capital letters. The brain icon is abstract and appears to be made up of several curved lines.
{6}------------------------------------------------
Image /page/6/Picture/0 description: The image shows the Brainlab logo. The logo consists of a stylized brain icon on the left and the word "BRAINLAB" in capital letters on the right. The color of the logo is pink.
### 4. Performance Data
The following testing was conducted on the Subject Device to establish substantial equivalence with the predicate device:
- Assembly stability: In order to verify that the stiffness and fixation of the instruments can resist unintended loads. The Drill Guide was able to withstand the forces without losing accuracy or function.
- . Accuracy: Accuracy testing was performed with acceptance criteria: tip position deviation must be equal or below 1.7 mm (95th percentile), and the angular deviation must be equal or below 1.7° (95th percentile). Testing showed the modified Drill Guide fulfills the same accuracy requirements as the predecessor one.
- Lifecyle assessment: To evaluate the influence of wear, reprocessing and aging on . selected aspects of product performance throughout the product lifetime. Accuracy as well as label readability requirements were met.
- . Skiving: Skiving testing showed the modified Drill Guide provides a more stable hold with the improved teeth design compared to the predicate Guide Tubes.
- . Handling and interface analysis: The Drill Guide design changes in terms of depth control and array attachment concept were evaluated and acceptance criteria were met in all cases.
- . Usability: Summative usability testing was performed in order to validate the changed design of the Drill Guides. The evaluation was conducted with 15 representative users
{7}------------------------------------------------
Image /page/7/Picture/0 description: The image shows the Brainlab logo. The logo consists of a stylized brain icon on the left and the word "BRAINLAB" in capital letters on the right. The color of the logo is pink.
which had to assemble and use the Drill Guide in a simulated scenario. The final design was proven safe and effective for use in the defined use scenarios.
- . Mechanical failure testing: The new worst case drill bit (diameter 2 mm) was tested under different scenarios in order to ensure it is strong enough to withstand the expected torques and possible bending under worst case conditions.
- . Biocompatibility: Was evaluated according to ISO 10993-1:2018 "Biological evaluation of medical devices – Part 1: Evaluation and testing within a risk management process.
- . Reprocessing validation: Was evaluated according to ISO 11737-1:2018, ANSI/AAMI ST98:2022 and ISO 15883-5:2021-07.
No clinical testing was required for the subject device.
### 5. Conclusion
The comparison of the Subject Device with the predicate device shows that the modified Spine & Trauma Navigation including modified instruments has similar functionality, intended use and technological characteristics as the predicate device. Based on the comparison to the predicate and the performance testing conducted, the Subject Device is considered substantially equivalent to the predicate device.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Report an issue or suggest a feature
Spot a broken citation, a wrong extraction, a missing document, or something you wish this app did? Tell us — no account needed.
or drag and drop files here
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.