The Q Guidance System, when used with the Spine Guidance Software, is intended as a planning and intraoperative guidance system to enable open or percutaneous computer assisted surgery in adult and pediatric (adolescent) patients. The system is indicated for any surgical procedure on the spine in which the use of computer assisted planning and surgery may be appropriate. The system can be used for intraoperative guidance where a reference to a rigid anatomical structure such as the skull, pelvis or spine can be identified. The system assists in the positioning of instruments for procedures on the pelvis and spine, including: Screw and Needle placement in the spine or pelvis.
Device Story
System provides image-guided stereotaxic planning and intraoperative navigation for spinal/pelvic surgery. Inputs include medical imaging (CT, MR, PET, 2D DICOM). System uses infrared optical tracking (active/passive) to monitor patient anatomy and surgical instruments in real-time. Surgeon uses computer platform (touchscreens, console) to perform preoperative planning (segmentation, trajectory definition) and intraoperative registration (anatomical, surface, 3D-to-2D). CORE 2 console powers surgical handpieces (e.g., Electric System 6) for drilling/reaming/screw placement. Software provides visual guidance on monitors, assisting instrument positioning. Output is real-time 3D navigation display. Used in OR by surgeons to improve precision of instrument placement; reduces risk of malpositioning; benefits patient through minimally invasive or more accurate surgical outcomes.
Clinical Evidence
No clinical data. Bench testing only. Performance evaluated via 3D positional accuracy (mean error ≤ 2.0mm) and trajectory angle accuracy (mean error < 2.0 degrees) using anatomically representative phantoms. Software verified per IEC 62304; biocompatibility per ISO 10993-1; electrical safety/EMC per IEC 60601 series.
Technological Characteristics
Stereotaxic navigation system using infrared optical tracking (active/passive). Components: computer platform, touchscreen monitors, CORE 2 console, navigated instruments. Materials biocompatible per ISO 10993-1. Sterilization: steam (reusable instruments) and gamma (batteries). Connectivity: Ethernet, USB. Software: Spine Guidance 5.0.
Indications for Use
Indicated for adult and pediatric (adolescent) patients undergoing open or percutaneous computer-assisted spinal or pelvic surgery, including screw and needle placement. Used for intraoperative guidance referencing rigid anatomical structures (skull, pelvis, spine).
Regulatory Classification
Identification
A stereotaxic instrument is a device consisting of a rigid frame with a calibrated guide mechanism for precisely positioning probes or other devices within a patient's brain, spinal cord, or other part of the nervous system.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA text logo on the right. The FDA text logo is in blue and includes the acronym "FDA" followed by "U.S. FOOD & DRUG ADMINISTRATION" in two lines. The logo is simple and professional, reflecting the organization's role in regulating food and drugs.
Build Correspondence
June 6, 2024
Stryker Leibinger GmbH & Co. KG Andrea Wallen-Gerding Principal Regulatory Affairs Specialist Bötzinger Straße 41 Freiburg Baden-Wurttembe, D-79111 Germany
# Re: K240662
-
Trade/Device Name: Spine Guidance Software (version 5.0); Xia 3/ Serrato O Instruments: Everest Q Instruments; Mesa 2 Q Instruments; 0 Pedicle Preparation Instruments; 0 S2AI Drills; Q Handles; Consolidated Operating Room Equipment (CORE) 2 Console; Electric System 6 Dual Trigger Rotary Handpiece
Regulation Number: 21 CFR 882.4560 Regulation Name: Stereotaxic Instrument Regulatory Class: Class II Product Code: OLO Dated: March 8, 2024 Received: March 8, 2024
Dear Andrea Wallen-Gerding:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrb/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}------------------------------------------------
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products: and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Shumaya Alı, M Assistant Director DHT6C: Division of Restorative, Repair and Trauma Devices OHT6: Office of Orthopedic Devices
{2}------------------------------------------------
Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{3}------------------------------------------------
# Indications for Use
Form Approved: OMB No. 0910-0120 Expiration Date: 07/31/2026 See PRA Statement below.
Submission Number (if known)
K240662
Device Name
- · Spine Guidance Software (version 5.0)
- · Q Pedicle Instruments
- o Xia 3 Q Taps
- o Serrato Q Taps
- o Xia 3/Serrato Q Screwdriver
- o ES2 Q Taps
- o ES2 Q Screwdriver
- o Everest Q Taps
- o Everest MI XT Q Taps
- o Everest Q Screwdriver
- o Everest MI XT Q Screwdriver
- o Mesa 2 Q Taps
- o Mesa 2 Q Screwdriver
- o Q S2Al Drills
- o Q Drills
- o Q Awl
- o Q Thoracic Probe
- o Q Lumbar Probe
- o Q Ratchet T-Handle
- o Q Ratchet Round Handle
- o Q Gear Shift Handle
- · Consolidated Operating Room Equipment (CORE) 2 Console
- · Electric System 6 Heavy Duty Dual Trigger Rotary Handpiece
Indications for Use (Describe)
Spine Guidance Software 5.0
The Q Guidance System, when used with the Spine Guidance Software, is intended as a planning and intraoperative guidance system to enable open or percutaneous computer assisted surgery in adult and pediatric (adolescent) patients. The system is indicated for any surgical procedure on the spine in which the use of computer assisted planning and surgery may be appropriate. The system can be used for intraoperative guidance where a reference to a rigid anatomical structure such as the skull, pelvis or spine can be identified. The system assists in the positioning of instruments for procedures on the pelvis and spine, including:
· Screw and Needle placement in the spine or pelvis.
Q Pedicle Instruments (QPI)
Stryker Spine Q Pedicle Instruments are intended to be used as accessories to the Stryker Spine Guidance System to facilitate placement of Stryker Spine implants. They can be navigated instruments or non-navigated manual instruments. When navigated, Stryker Spine Q Pedicle Instruments are specifically designed for use with the Spine Guidance Software. They are intended to be used with a Q Navigation Adaptor and associated trackers manually or under power with a Stryker Rotary Handpiece to facilitate pedicle preparation and placement of Stryker Spine implants in the non-cervical spine in adult and pediatric (adolescent) patients in accordance with the
{4}------------------------------------------------
indications and contraindications of the associated Stryker Spine Implant System.
Consolidated Operating Room Equipment (CORE) 2 Console
The Stryker Consolidated Operating Room Equipment (CORE) 2 Console is intended for use in the cutting, drilling, reaming, decorticating, shaping, and smoothing of bone, bone cement and teeth in a variety of surgical procedures, including but not limited to orthopedic, dental, ENT (Ear, Nose, Throat), neuro, spine, and endoscopic applications. The console is also usable in the placement or cutting of screws, metal, wires, pins, and other fixation devices. The CORE 2 Console is also indicated as an accessory to the Stryker Spine Guidance Software for stereotactic surgical procedures on the spine in adult and pediatric (adolescent) patients.
Electric System 6 Heavy Duty Dual Trigger Rotary Handpiece
The Stryker Electric System 6 Heavy Duty Dual Trigger Rotary Handpiece is used with the Consolidated Operating Room Equipment (CORE™) System. When used with a variety of Strykerapproved attachments, the handpiece is intended for orthopedic surgical procedures involving drilling, reaming, driving wire or pins, cutting bone and hard tissue. The Stryker Electric System 6 Heavy Duty Dual Trigger Rotary Handpiece, when used in conjunction with Stryker Spine Guidance Software, is intended to be used to facilitate preparation and placement of screws of Stryker Spine implant systems in adult and pediatric (adolescent) patients.
Type of Use (Select one or both, as applicable)
Prescription Use (Part 21 CFR 801 Subpart D)
he-Counter Use (21 CFR 801 Subpart C)
# CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
# *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{5}------------------------------------------------
# ar
| 807.92(a)(1) - Submitter Information | |
|--------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| 510(k) Submitter: | Stryker Leibinger GmbH & Co. KG<br>Bötzinger Straße 41<br>79111 Freiburg, Germany |
| Contact Information: | Primary:<br>Andrea N. Wallen-Gerding<br>Principal Regulatory Affairs Specialist<br>(269) 491-9234<br>andrea.wallengerding@stryker.com<br><br>Secondary:<br>Kirsten Reinhold<br>Senior Manager, Regulatory Affairs<br>(201) 419-9044<br>kirsten.reinhold@stryker.com |
| Date Summary Prepared: | March 7, 2024 |
# 510(k) Summary - K240662
| 807.92(a)(2) – Name of Device | |
|-------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Trade Name(s): | • Spine Guidance Software (version 5.0) • Q Pedicle Instruments ○ Xia 3 Q Taps ○ Serrato Q Taps ○ Xia 3/Serrato Q Screwdriver ○ ES2 Q Taps ○ ES2 Q Screwdriver ○ Everest Q Taps ○ Everest MI XT Q Taps ○ Everest Q Screwdriver ○ Everest MI XT Q Screwdriver ○ Mesa 2 Q Taps ○ Mesa 2 Q Screwdriver ○ Q S2AI Drills ○ Q Drills ○ Q Awl ○ Q Thoracic Probe ○ Q Lumbar Probe ○ Q Ratchet T-Handle ○ Q Ratchet Round Handle ○ Q Gear Shift Handle • Consolidated Operating Room Equipment (CORE) 2 Console • Electric System 6 Heavy Duty Dual Trigger Rotary Handpiece |
| Classification Name: | Orthopedic Stereotaxic Instrument |
| Device Class: | Class II (21 CFR 882.4560) |
| Product Code(s): | OLO |
# Stryker Leibinger GmbH & Co. KG
{6}------------------------------------------------
| 807.92(a)(3) - Legally marketed device(s) to which equivalence is claimed | |
|---------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------|
| Spine Guidance Software, v5.0 | Spine Guidance Software, v4.0, K220593 |
| Q Pedicle Instruments | Navigated Spine Instruments, K203205, K172034 |
| Consolidated Operating Room Equipment<br>(CORE) 2 Console | Stryker Consolidated Operating Room Equipment<br>(CORE) 2 Console, K171840 |
| Electric System 6 Heavy Duty Dual Trigger<br>Rotary Handpiece | FDA Product Code: HWE (Instrument, Surgical,<br>Orthopedic, AC-Powered Motor and Accessory/<br>Attachment)<br>Regulation Number: 878.4820 |
# 007 02/~\/2)
# 807.92(a)(4) - Device Description
# Spine Guidance 5.0 System Overview
The Spine Guidance 5.0 system is an image guided stereotaxic, planning, and intraoperative guidance system intended to enable open or percutaneous computer-assisted surgery. It assists the surgeon in precisely positioning manual and powered instruments and locating patient anatomy during spinal surgery.
The system is comprised of the Spine Guidance 5.0 Software, the O Guidance System (computer platform), navigated accessories/ instruments (e.g., powered drills, pointers), and various system components (e.g., calibration devices, navigation adaptors, patient/ instrument trackers, etc.). The system provides intraoperative guidance to the surgeon using passive and active wireless optical tracking technologies. The computer platform consists of a computer, camera, big touchscreen monitor, and a small touchscreen monitor. The Spine Guidance 5.0 software functionality is described in terms of its capabilities that feature planning, registration, and navigation of medical devices,
The following new instrumentation can be used with the system:
- . The O Pedicle Instruments are intended to be used to facilitate pedicle preparation and screw placement of Stryker Spine implant systems.
- . The CORE 2 Console has three handpiece ports. two footswitch ports, and one irrigation cassette port. The console supplies power to devices for a variety of surgical procedures as described in its indications for use. It is comprised of a power supply, central processing unit, motor controller, irrigation pump controller, RFID module, liquid crystal display (LCD) screen and speaker for output, touch screen for user input, USB port, and an Ethernet port. The CORE 2 software contains an optional depth-stop feature driven by Spine Guidance 5 Software.
- . The ES 6 Heavy Duty Dual Trigger Rotary Handpiece is used in drilling, reaming and decorticating of bone and other bone related tissue in a variety of surgical procedures. It is also usable in the placement of screws, wires, pins, and other fixation devices. This handpiece supports the depth-stop feature.
This system is for professional use only within a professional healthcare environment.
### Stryker Leibinger GmbH & Co. KG
{7}------------------------------------------------
# 807.92(a)(5) - Intended Use of the Device Indications for Use:
# Spine Guidance Software 5.0
The Q Guidance System, when used with the Spine Guidance Software, is intended as a planning and intraoperative guidance system to enable open or percutaneous computer assisted surgery in adult and pediatric (adolescent) patients.
The system is indicated for any surgical procedure on the spine in which the use of computer assisted planning and surgery may be appropriate. The system can be used for intraoperative guidance where a reference to a rigid anatomical structure such as the skull, pelvis or spine can be identified.
The system assists in the positioning of instruments for procedures on the pelvis and spine, including:
- Screw and Needle placement in the spine or pelvis.
# Q Pedicle Instruments (QPI)
Stryker Spine Q Pedicle Instruments are intended to be used as accessories to the Stryker Spine Guidance System to facilitate placement of Stryker Spine implants. They can be navigated instruments or nonnavigated manual instruments. When navigated, Stryker Spine Q Pedicle Instruments are specifically designed for use with the Spine Guidance Software. They are intended to be used with a Q Navigation Adaptor and associated trackers manually or under power with a Stryker Rotary Handpiece to facilitate pedicle preparation and placement of Stryker Spine implants in the non-cervical spine in adult and pediatric (adolescent) patients in accordance with the indications and contraindications of the associated Stryker Spine Implant System.
# Consolidated Operating Room Equipment (CORE) 2 Console
The Stryker Consolidated Operating Room Equipment (CORE) 2 Console is intended for use in the cutting, drilling, reaming, decorticating, shaping, and smoothing of bone, bone cement and teeth in a variety of surgical procedures, including but not limited to orthopedic, dental, ENT (Ear, Nose, Throat), neuro, spine, and endoscopic applications. The console is also usable in the placement or cutting of screws, metal, wires, pins, and other fixation devices.
The CORE 2 Console is also indicated as an accessory to the Stryker Spine Guidance Software for stereotactic surgical procedures on the spine in adult and pediatric (adolescent) patients.
# Electric System 6 Heavy Duty Dual Trigger Rotary Handpiece
The Stryker Electric System 6 Heavy Duty Dual Trigger Rotary Handpiece is used with the Consolidated Operating Room Equipment (CORE™) System. When used with a variety of Strykerapproved attachments, the handpiece is intended for orthopedic surgical procedures involving drilling. reaming, driving wire or pins, cutting bone and hard tissue.
The Stryker Electric System 6 Heavy Duty Dual Trigger Rotary Handpiece, when used in conjunction with Stryker Spine Guidance Software, is intended to be used to facilitate preparation and placement of screws of Stryker Spine implant systems in adult and pediatric (adolescent) patients.
### Stryker Leibinger GmbH & Co. KG
{8}------------------------------------------------
| Characteristic | Subject Device | Predicate Device |
|---------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Indication(s) | The Q Guidance System, when used<br>with the Spine Guidance Software, is<br>intended as a planning and<br>intraoperative guidance system to<br>enable open or percutaneous<br>computer assisted surgery in adult<br>and pediatric (adolescent) patients.<br><br>The system is indicated for any<br>surgical procedure on the spine in<br>which the use of computer assisted<br>planning and surgery may be<br>appropriate. The system can be used<br>for intraoperative guidance where a<br>reference to a rigid anatomical<br>structure such as the skull, pelvis or<br>spine can be identified.<br><br>The system assists in the positioning<br>of instruments for procedures on the<br>pelvis and spine, including:<br>• Screw and needle placement in<br>the spine or pelvis | The Stryker Q Guidance System,<br>when used with the Spine<br>Guidance Software, is intended a<br>a planning and intraoperative<br>guidance system to enable open<br>or percutaneous computer<br>assisted surgery in adult and<br>pediatric patients.<br><br>The system is indicated for any<br>surgical procedure on the spine in<br>which the use of computer<br>assisted planning and surgery<br>may be appropriate. The system<br>can be used for intraoperative<br>guidance where a reference to a<br>rigid anatomical structure such as<br>the spine, pelvis or skull can be<br>identified.<br><br>The system assists in the<br>positioning of instruments for<br>procedures on the spine and<br>pelvis, including:<br>• Screw Placement in the spine<br>or pelvis |
| Main System Components | • Q Guidance System<br>• Navigated Instruments<br>• System accessories and<br>components | • Q Guidance System<br>• Navigated Instruments<br>• System accessories and<br>components |
| Modes of Operation | • Patient Preparation<br>• System Set-up<br>• Image Import<br>• Planning<br>• Patient Registration<br>• Navigation | • Patient Preparation<br>• System Set-up<br>• Image Import<br>• Planning<br>• Patient Registration<br>• Navigation |
| Localizing and Tracking | Infrared Optical Active and Passive<br>Tracking | Infrared Optical Active and<br>Passive Tracking |
| System Accuracy | The system has a mean accuracy of 2<br>mm for positional displacement and<br>2° for trajectory angle displacement.<br>Accuracy values apply to tracking in<br>the workspace. | The system has a mean accuracy<br>of 2 mm for positional<br>displacement and 2° for trajectory<br>angle displacement. Accuracy<br>values apply to tracking in the<br>workspace. |
| Supported Imaging<br>Modalities | Computed tomography (CT) Magnetic resonance (MR) Position emission tomography (PET) 2D DICOM images (SCOUT only) | Computed tomography (CT) Magnetic resonance (MR) Position emission tomography (PET) |
| Planning Features | Screws Measurements Trajectories Segmentations (Manual and automatic) Merge Levels (local correlation) | Screws Measurements Trajectories Segmentations (Manual and automatic) Merge Levels (local correlation) Image merge |
| Registration Features | Anatomical (Point-to-Point) Registration Surface Registration 3D CT/ C-Arm Registration Automatic Intraoperative Mask Registration | Anatomical (Point-to-Point) Registration Surface Registration 3D CT/ C-Arm Registration Automatic Intraoperative Mask Registration Mask Registration |
| Intended Use Environment | Operating Room (OR) | Operating Room (OR) |
| Output | 3D image Analog and Digital Video Images | 3D image Analog and Digital Video Images |
| Graphical User<br>Interface | Black-style graphical user interface 16:9 screen ratio Case Dashboard to access all operation modes One tab per task concept from left to right on top of screen Image box with image tools Current task panel on the right or bottom Image settings task panel on the left | Black-style graphical user interface 16:9 screen ratio Case Dashboard to access all operation modes Image box with image tools Current task panel on the right Image settings task panel on the left |
| User Interface | Large touch monitor Small touch monitor Keyboard Mouse Buttons on active optical instruments | Large touch monitor Small touch monitor Keyboard Mouse Buttons on active optical instruments |
# Stryker Leibinger GmbH & Co. KG
{9}------------------------------------------------
# Technological Comparison between the Navigated Q Pedicle Instruments and their predicate devices
The navigated Q Pedicle Instruments in scope of this traditional 510(k) have similar designs, materials, intended use, sterilization, and fundamental scientific technology to their predicate devices and system
# Stryker Leibinger GmbH & Co. KG
{10}------------------------------------------------
Image /page/10/Picture/0 description: The image shows the word "stryker" in a bold, sans-serif font. The word is all lowercase and appears to be a logo or brand name. The letters are closely spaced, giving the word a compact appearance. The color of the text is black, and it is set against a white background.
components. The modifications to the instruments do not adversely impact the technological characteristics of the predicate devices.
# Technological Comparison between the CORE 2 Console and its predicate device
There have been no changes made to the intended use, materials, or fundamental scientific technology of the CORE 2 Console. The device has undergone a software modification to enable communication with the Spine Guidance 5.0 Software System. The indications for use have also been updated to allow use with the Spine Guidance 5 Software.
## Technological Comparison between the Electric System 6 and its predicate device
There have been no changes to the intended use, design, materials, fundamental scientific technology, or sterilization process. The only update is to the indication for use to allow it to be used with the Spine Guidance 5 Software.
# 807.92(b)(1) – Nonclinical Testing to Support Submission
The function and performance of the subject devices have been evaluated through non-clinical design verification and validation testing. Additional testing was performed on the subject devices to ensure they met their design requirements. The results of the evaluation tests demonstrate that the subject devices successfully meet the requirements of their intended use.
# Accuracy
System demonstrated performance in 3D positional accuracy with a mean error ≤ 2.0mm and in trajectory angle accuracy with a mean error < 2.0 degrees. This performance was determined using anatomically representative phantoms and utilizing a subset of system components and features that represent the worst-case combinations of all potential system components.
# Software
Software verification and validation testing was conducted as required by IEC 62304 and FDA Guidance on General Principles of Software Validation, January 11, 2002. All requirements were met, and no new issues of safety or effectiveness were raised.
# Biocompatibility
The biocompatibility of all patient contact materials was verified according to ISO 10993-1:2018 and the FDA guidance Use of International Standard ISO 10993-1, "Biological evaluation of medical devices - Part 1: Evaluation and testing within a risk management process, September 2023. No new issues of safety or effectiveness were raised.
# Electrical Safety and Electromagnetic Compatibility
Verified conformance to IEC 60601-1:2005. IEC 60601-1:2005/AMD1:2012. IEC 60601-1:2005/AMD2:2020.
Verified conformance to IEC 60601-1-2:2014, IEC 60601-1-2:2014/AMD1:2020, CISPR 11 Group 1, Class A requirements as well as additional testing to verify compatibility with RFID devices operating in the 125 - 134 kHz and 13.56 MHz frequency band.
### Stryker Leibinger GmbH & Co. KG
{11}------------------------------------------------
# Sterilization
The reusable subject devices underwent a steam sterilization to demonstrate that they can be expected to be sterile and have a sterility assurance level (SAL) of 10-6 or greater after processing. The Tracking Instrument Battery underwent gamma sterilization validation per ISO 11137-2: 2013/(R) 2019 to demonstrate that they can be expected to be sterile and have a sterility assurance level (SAL) of 10-6
# 807.92(b)(2) - Clinical Testing
No clinical testing was required to support this submission.
# 807.92(b)(3) - Conclusions Drawn from Testing Performed
Results of performance testing were determined successful for all protocols and demonstrate that the functionality, integrity, and safety and effectiveness of the subject devices are sufficient for their intended use. Performance test results for the subject devices demonstrate substantial equivalent to the predicate devices.
# Conclusion/ Substantial Equivalence (SE) Rationale:
The subject device, in comparison with the legally marketed predicate, has the same intended use, indications for use, operating principles, energy source, and functional outputs. Performance testing and risk analysis demonstrate that the device is as safe and effective as the predicate device and supports the determination of substantial equivalence.
### Stryker Leibinger GmbH & Co. KG
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.