Sejoy Blood Glucose Monitoring System; Sejoy Advance Link Blood Glucose Monitoring System
Applicant
Sejoy Biomedical Co., Ltd.
Product Code
NBW · Clinical Chemistry
Decision Date
Oct 8, 2024
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 862.1345
Device Class
Class 2
Indications for Use
The Sejoy Blood Glucose Monitoring System is intended to be used for the quantitative measurement of glucose in fresh capillary whole blood collected from the fingertip. The Sejoy Blood Glucose Monitoring System is intended for self-testing outside the body (in-vitro diagnostic use), by individuals with diabetes at home as an aid to monitor the effectiveness of diabetes control. This system is intended to be used by a single person and should not be shared. The system should not be used for the diagnosis of, or screening for diabetes or for neonatal use. The Sejoy Advance Link Blood Glucose Monitoring System is intended to be used for the quantitative measurement of glucose in fresh capillary whole blood drawn from the fingertips. The Sejoy Advance Link Blood Glucose Monitoring System is intended for self-testing outside the body (in-vitro diagnostic use), by individuals with diabetes at home as an aid to monitor the effectiveness of diabetes control. This system is intended to be used by a single person and should not be shared. The system should not be used for the diagnosis of, or screening for diabetes or for neonatal use.
Device Story
System measures glucose in fresh capillary whole blood via quantitative amperometry; uses test strips with glucose dehydrogenase (FAD-GDH) enzyme; blood drawn into reaction cell via capillary action; reaction produces electric current proportional to glucose concentration. Meter displays plasma-equivalent results. Intended for home self-testing by individuals with diabetes. Sejoy Advance Link model includes Bluetooth for wireless data transmission to Glucojoy App; both models feature optional voice output for results and error codes. Users apply blood directly to strip; no manual coding required. System includes control solutions for quality checks. Provides aid in monitoring diabetes control effectiveness. Not for multi-patient use; requires specific disinfection protocols.
Clinical Evidence
Clinical user evaluation study with 352 lay users comparing device results to YSI 2300 laboratory reference method. Accuracy results: 100% of results within ±20% of reference. Additional testing on 100 contrived samples (50 low, 50 high glucose) confirmed 100% accuracy within ±20% of reference. Bench testing included repeatability, intermediate precision, hematocrit effect, and interference testing.
Technological Characteristics
Amperometric glucose test system using FAD-GDH enzyme. Measurement range 20-600 mg/dL. Minimum sample volume 0.6 µL. Bluetooth connectivity (Advance Link model). Powered by battery. Compliant with IEC 60601-1, IEC 60601-1-2, and IEC 60601-1-11. Disinfection validated for 4000 cycles using CaviWipes.
Indications for Use
Indicated for individuals with diabetes for self-testing of glucose in fresh capillary whole blood from the fingertip to monitor diabetes control. Contraindicated for neonatal use, diagnosis or screening of diabetes, critically ill patients, patients in shock, hyperglycemic-hyperosmolar state, severe hypotension, or severe dehydration.
Regulatory Classification
Identification
A glucose test system is a device intended to measure glucose quantitatively in blood and other body fluids. Glucose measurements are used in the diagnosis and treatment of carbohydrate metabolism disorders including diabetes mellitus, neonatal hypoglycemia, and idiopathic hypoglycemia, and of pancreatic islet cell carcinoma.
Special Controls
*Classification.* Class II (special controls). The device, when it is solely intended for use as a drink to test glucose tolerance, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 862.9.
Predicate Devices
OneTouch Verio Reflect Glucose Monitoring System (K193475)
Submission Summary (Full Text)
{0}
FDA
U.S. FOOD & DRUG
ADMINISTRATION
# 510(k) SUBSTANTIAL EQUIVALENCE DETERMINATION DECISION SUMMARY
ASSAY AND INSTRUMENT
## I Background Information:
A 510(k) Number
K240640
B Applicant
Sejoy Biomedical Co., Ltd
C Proprietary and Established Names
Sejoy blood glucose monitoring system and Sejoy advance link blood glucose monitoring system
D Regulatory Information
| Product Code(s) | Classification | Regulation Section | Panel |
| --- | --- | --- | --- |
| NBW | Class II | 21 CFR 862.1345 - Glucose Test System | CH - Clinical Chemistry |
## II Submission/Device Overview:
A Purpose for Submission:
New Devices
B Measurand:
Glucose in fingerstick capillary whole blood
C Type of Test:
Quantitative amperometry for blood, glucose dehydrogenase (FAD-GDH)
## III Intended Use/Indications for Use:
A Intended Use(s):
Food and Drug Administration
10903 New Hampshire Avenue
Silver Spring, MD 20993-0002
www.fda.gov
{1}
See Indications for Use below.
## B Indication(s) for Use:
The Sejoy Blood Glucose Monitoring System is composed of the Sejoy Blood Glucose Meter and Sejoy Blood Glucose Test Strips. The Sejoy Blood Glucose Monitoring System is intended to be used for the quantitative measurement of glucose in fresh capillary whole blood collected from the fingertip. The Sejoy Blood Glucose Monitoring System is intended for self-testing outside the body (in-vitro diagnostic use), by individuals with diabetes at home as an aid to monitor the effectiveness of diabetes control. This system is intended to be used by a single person and should not be shared. The system should not be used for the diagnosis of, or screening for diabetes or for neonatal use.
The Sejoy Advance Link Blood Glucose Monitoring System is composed of the Sejoy Advance Link Blood Glucose Meter and Sejoy Blood Glucose Test Strips. The Sejoy Advance Link Blood Glucose Monitoring System is intended to be used for the quantitative measurement of glucose in fresh capillary whole blood drawn from the fingertips. The Sejoy Advance Link Blood Glucose Monitoring System is intended for self-testing outside the body (in-vitro diagnostic use), by individuals with diabetes at home as an aid to monitor the effectiveness of diabetes control. This system is intended to be used by a single person and should not be shared. The system should not be used for the diagnosis of, or screening for diabetes or for neonatal use.
## C Special Conditions for Use Statement(s):
OTC - Over The Counter
- The meter and lancing device are for single patient use. Do not share them with anyone including other family members! Do not use on multiple patients.
- Not for use on critically ill patients, patients in shock or in a hyperglycemic-hyperosmolar state with or without ketosis.
- Not for neonatal use.
- Not for Alternative Site Testing (AST).
- Not for use with severe hypotension.
- Not for use with patients with severe dehydration.
- Not intended for diagnosis or screening of diabetes.
- Not for screening or diagnosis of diabetes mellitus.
- Do not use at altitudes above 10000ft (3048 meters) above sea level.
- Do not perform test during or soon after xylose absorption testing. High xylose level in the blood will cause inaccurate results.
- This meter is not intended for use in healthcare or assisted-use settings such as hospitals, physician offices, or long-term care facilities.
- Use of this meter on multiple patients may lead to transmission of Human Immunodeficiency Virus (HIV), Hepatitis C Virus (HCV), Hepatitis B Virus (HBV), or other bloodborne pathogens.
## D Special Instrument Requirements:
Sejoy Blood Glucose Meter
Sejoy Advance Link Blood Glucose Meter
K240640 - Page 2 of 13
{2}
K240640 - Page 3 of 13
## IV Device/System Characteristics:
### A Device Description:
Sejoy Blood Glucose Monitoring System consists of the Sejoy Blood Glucose Meter and the Sejoy Blood Glucose Test Strips and the Sejoy Advanced Link Blood Glucose Monitoring System consists of the Sejoy Advanced Link Blood Glucose Meter and the Sejoy Blood Glucose Test Strips. The test strips can be purchased in vials or individually wrapped foil packages.
The Sejoy Advanced Link Blood Glucose Meter uses Bluetooth to transmit glucose readings wirelessly to mobile devices to the Glucojoy App. Both meters have an optional voice feature that speaks certain messages such as the glucose results once testing is completed, announces when the battery is low, speaks the error codes if an error were to occur etc.
The Sejoy Blood Glucose Control Solutions (CTRL1, CTRL2, CTRL3), Sejoy Lancing Device (K222034) and Sejoy Disposable Safety Lancets (K222408) can be purchased separately for use with both systems.
### B Principle of Operation:
The Sejoy Blood Glucose Monitoring System and Sejoy Advance Link Blood Glucose Monitoring Systems quantitatively measure the amount of glucose in fresh capillary whole blood from the fingertip. Blood is applied to the end tip of the test strip and is then automatically pulled into the reaction cell through capillary action. The reaction area of the test strip includes the enzyme glucose dehydrogenase with cofactor flavin adenine dinucleotide (FAD-GDH) and mediator. Glucose in the sample reacts with the test strip chemistry to produce an electric current, which is proportional to the amount of glucose in the sample. The meter measures the strength of the current and displays the corresponding blood glucose concentration. The meter provides plasma-equivalent results.
### C Instrument Description Information:
1. Instrument Name:
Sejoy Blood Glucose Meter
Sejoy Advance Link Blood Glucose Meter
2. Specimen Identification:
There is no sample identification function with this device. Samples are applied directly to the test strip as they are collected.
3. Specimen Sampling and Handling:
The glucose system is intended to be used with capillary whole blood from the finger only. The whole blood sample is applied directly to the test strip by capillary action.
{3}
K240640 - Page 4 of 13
4. Calibration:
The meter does not require calibration or coding by the user. The meter is automatically coded.
5. Quality Control:
Three levels of glucose control solutions (CTRL1, CTRL2, CTRL3) are available with this system and can be purchased separately. Recommendations on when to test the control materials are provided in the labeling. Acceptable ranges for each level of control solution are printed on the test strip vial label and test strip foil packet. If quality control test results are out of range, the user is advised to repeat the test. If problems continue, the user is cautioned not to use the meter and to contact the distributor. The Sejoy Blood Glucose Meter and Sejoy Advance Link Blood Glucose Meter recognize the control solutions automatically and do not store the control solution results.
This medical device product has functions subject to FDA premarket review as well as functions that are not subject to FDA premarket review. For this application, if the product has functions that are not subject to FDA premarket review, FDA assessed those functions only to the extent that they either could adversely impact the safety and effectiveness of the functions subject to FDA premarket review or they are included as a labeled positive impact that was considered in the assessment of the functions subject to FDA premarket review.
V Substantial Equivalence Information:
A Predicate Device Name(s):
OneTouch Verio Reflect Blood Glucose Monitoring System
B Predicate 510(k) Number(s):
K193475
C Comparison with Predicate(s):
| Device & Predicate Device(s): | K240640 | K240640 | K193475 |
| --- | --- | --- | --- |
| Device Trade Name | Sejoy Advance Link Blood Glucose Monitoring System | Sejoy Blood Glucose Monitoring System | OneTouchVerio Reflect Blood Glucose Monitoring System |
| General Device Characteristic Similarities | | | |
| Intended Use/Indications For Use | To be used for the quantitative measurement of glucose in fresh capillary whole | Same | Same |
{4}
| Device & Predicate Device(s): | K240640 | K240640 | K193475 |
| --- | --- | --- | --- |
| | blood drawn from the fingertips
as an aid to monitor
the effectiveness of
diabetes control. | | |
| Enzyme | Glucose
Dehydrogenase | Same | Same |
| Measurement Range | 20 - 600 mg/dL | Same | Same |
| General Device
Characteristic
Differences | | | |
| Sample Volume | Minimum 0.6ul | Minimum 0.6ul | Minimum 0.4ul |
| Hematocrit Range | 15-60% | 15-60% | 20-60% |
| Data Transmission | Bluetooth | None | Via USB or Bluetooth |
VI Standards/Guidance Documents Referenced:
IEC 60601-1 Edition 3.2: Medical electrical equipment - Part 1: General requirements for basic safety and essential performance.
IEC 60601-1-2 Edition 4.1: Medical electrical equipment - Part 1-2: General requirements for basic safety and essential performance - Collateral Standard: Electromagnetic disturbances - Requirements and tests
CLSI EP06 2nd Edition: Evaluation of the Linearity of Quantitative Measurement Procedures
CLSI EP07 3rd Edition: Interference Testing in Clinical Chemistry.
CLSI EP05-A3 Evaluation of Precision of Quantitative Measurement Procedures
CLSI EP25-A Evaluation of Stability of In Vitro Diagnostic Reagents
ANSI AAMI SW96:2023 Standard for medical device security - Security risk management for device manufacturers
IEC 60601-1-11 Edition 2.1 2020-07 Consolidated Version: Medical electrical equipment - Part 1-11: General requirements for basic safety and essential performance - Collateral Standard: Requirements for medical electrical equipment and medical electrical systems used in the home healthcare environment
IEC TR 60601-4-2 Edition 1.0 2016-05 Medical electrical equipment - Part 4-2: Guidance and interpretation - Electromagnetic immunity: performance of medical electrical equipment and medical electrical systems
K240640 - Page 5 of 13
{5}
FDA Guidance Document: Self-Monitoring Blood Glucose Test Systems for Over-the-Counter Use. Guidance for Industry and Food and Drug Administration Staff. Issued on September 29, 2020.
## VII Performance Characteristics (if/when applicable):
The Sejoy Advance Link Blood Glucose Monitoring System was used as a representative model for the performance evaluations below to support both the Sejoy Advance Link Blood Glucose Monitoring System and the Sejoy Blood Glucose Monitoring System. The only difference between the systems is the presence of Bluetooth technology in the Sejoy Advance Link Blood Glucose Blood Glucose Meter, which does not impact the glucose measurement.
## A Analytical Performance:
### 1. Precision/Reproducibility:
#### Within-Run Precision (Repeatability)
Within-run precision studies were performed using venous whole blood samples adjusted to 5 glucose concentration levels (30 to 50, 51 to 110, 111 to 150, 151 to 250, 251 to 400 mg/dL). Each sample was tested in replicates of 10 with 3 lots of test strips and 10 meters for a total of 300 tests per glucose level. Results are summarized below:
| Glucose Level (mg/dL) | Strip Lot | N | Mean (mg/dL) | SD (mg/dL) | %CV |
| --- | --- | --- | --- | --- | --- |
| 30 to 50 | 1 | 100 | 40.14 | 1.76 | 4.39% |
| | 2 | 100 | 40.34 | 1.74 | 4.30% |
| | 3 | 100 | 40.22 | 1.66 | 4.13% |
| | Combined | 300 | 40.23 | 1.72 | 4.27% |
| 51 to 110 | 1 | 100 | 90.11 | 2.71 | 3.00% |
| | 2 | 100 | 90.10 | 2.88 | 3.19% |
| | 3 | 100 | 90.02 | 2.52 | 2.80% |
| | Combined | 300 | 90.08 | 2.70 | 2.99% |
| 111 to 150 | 1 | 100 | 130.10 | 3.24 | 2.49% |
| | 2 | 100 | 130.45 | 3.43 | 2.63% |
| | 3 | 100 | 129.58 | 3.14 | 2.42% |
| | Combined | 300 | 130.04 | 3.28 | 2.52% |
| 151 to 250 | 1 | 100 | 205.18 | 5.09 | 2.48% |
| | 2 | 100 | 204.49 | 4.40 | 2.15% |
| | 3 | 100 | 205.01 | 4.91 | 2.39% |
| | Combined | 300 | 204.89 | 4.80 | 2.34% |
| 251 to 400 | 1 | 100 | 348.67 | 7.95 | 2.28% |
| | 2 | 100 | 348.18 | 7.99 | 2.30% |
| | 3 | 100 | 348.52 | 7.08 | 2.03% |
| | Combined | 300 | 348.52 | 7.66 | 2.20% |
K240640 - Page 6 of 13
{6}
Intermediate Precision (Between Run)
Intermediate (between run) precision was evaluated using 5 levels of glucose control solutions (30 to 50, 51 to 110, 111 to 150, 151 to 250, 251 to 400 mg/dL) with 3 test strip lots and 10 meters. Each control solution level was measured once a day for 10 days with each meter and test strip lot, for a total of 100 replicates per control solution level per test strip lot for a total of 300 replicates for each glucose control solution level. Results are summarized below:
| Glucose Level (mg/dL) | Strip Lot | N | Mean (mg/dL) | SD (mg/dL) | %CV |
| --- | --- | --- | --- | --- | --- |
| 30 to 50 | 1 | 100 | 40.07 | 1.84 | 4.59% |
| | 2 | 100 | 40.22 | 1.99 | 4.94% |
| | 3 | 100 | 40.26 | 1.97 | 4.89% |
| | Combined | 300 | 40.18 | 1.93 | 4.80% |
| 51 to 110 | 1 | 100 | 92.2 | 3.09 | 3.35% |
| | 2 | 100 | 91.63 | 2.67 | 2.92% |
| | 3 | 100 | 92.1 | 3 | 3.26% |
| | Combined | 300 | 91.98 | 2.93 | 3.18% |
| 111 to 150 | 1 | 100 | 134.87 | 3.21 | 2.38% |
| | 2 | 100 | 134.88 | 3.22 | 2.39% |
| | 3 | 100 | 134.48 | 3.21 | 2.39% |
| | Combined | 300 | 134.74 | 3.21 | 2.38% |
| 151 to 250 | 1 | 100 | 195.47 | 4.83 | 2.47% |
| | 2 | 100 | 195.39 | 4.64 | 2.38% |
| | 3 | 100 | 195.01 | 4.73 | 2.42% |
| | Combined | 300 | 195.29 | 4.72 | 2.42% |
| 251 to 400 | 1 | 100 | 325.52 | 7.62 | 2.34% |
| | 2 | 100 | 324.13 | 7.35 | 2.27% |
| | 3 | 100 | 326.18 | 7.97 | 2.44% |
| | Combined | 300 | 325.35 | 7.68 | 2.36% |
2. Linearity:
The linearity was evaluated using venous whole blood samples adjusted to 11 glucose levels ranging from 17 to 619 mg/dL (17, 74, 139, 195, 253, 317, 382, 433, 492, 551 and 619 mg/dL as measured by comparator method YSI 2300 Analyzer) with 3 lots of test strips. Test procedures were performed in 4 replicates per lot per concentration level. Glucose results obtained using the Sejoy Advance Link Blood Glucose Monitoring System were compared to those obtained using comparator method YSI 2300 analyzer. The results of the regression analysis are: $y = 1.0063x + 0.9607$ ; $R^2 = 0.99667$ .
The results of the study support the sponsor's claimed glucose measuring range of 20 to 600 mg/dL. If a sample result is less than 20 mg/dL glucose, the result is flagged by the meter as "Lo". If a sample result exceeds 600 mg/dL glucose, the result is flagged by the meter as "Hi". The "Lo" and "Hi" functions were validated by the sponsor and were demonstrated to function as intended.
K240640 - Page 7 of 13
{7}
# 3. Analytical Specificity/Interference:
To assess potential interference, studies were performed by spiking 33 exogenous and endogenous substances into venous whole blood adjusted to three glucose levels ( $\sim 60 \mathrm{mg} / \mathrm{dL}$ , $\sim 125 \mathrm{mg} / \mathrm{dL}$ , and $\sim 240 \mathrm{mg} / \mathrm{dL}$ ). Each of these samples was divided into a test pool and a control pool, with the potential endogenous and exogenous interfering substances added to the test pool. The relative bias ( $\mathrm{mg} / \mathrm{dL}$ ) and $\%$ bias between the test sample and the control sample was calculated using the mean of 10 replicates for each of 3 strip lots tested. The highest tested concentration for each substance at which no significant interference was observed (defined by the sponsor as less than $\pm 10\%$ ) is summarized in the following table:
| Test Substance | Highest concentration tested with no significant interference |
| --- | --- |
| Acetaminophen | 20 mg/dL |
| Ascorbic acid | 6 mg/dL |
| Cholesterol | 625 mg/dL |
| Conjugated Bilirubin | 50 mg/dL |
| Creatinine | 15 mg/dL |
| Dopamine | 0.09 mg/dL |
| EDTA | 0.1 mg/dL |
| Galactose | 60 mg/dL |
| Gentisic acid | 1.8 mg/dL |
| Glipizide Tablets | 0.1 mg/dL |
| Hemoglobin | 1000 mg/dL |
| Heparin | 300 IU/dL |
| Ibuprofen | 50 mg/dL |
| Icodextrin | 1094 mg/dL |
| Isomalt | 0.09 mg/dL |
| Lactitol | 0.09 mg/dL |
| L-DOPA | 0.75 mg/dL |
| Maltitol | 0.09 mg/dL |
| Maltose | 900 mg/dL |
| Mannitol | 1800 mg/dL |
| Metformin Hydrochloride Tablets | 0.4 mg/dL |
| Methyldopa | 2 mg/dL |
| Reduced Glutathione | 4.6 mg/dL |
| Salicylic acid | 60 mg/dL |
| Sodium chloride | 180 mmol/L |
| Sorbitol | 0.09 mg/dL |
K240640 - Page 8 of 13
{8}
| Test Substance | Highest concentration tested with no significant interference |
| --- | --- |
| Tolazamide | 9 mg/dL |
| Tolbutamide | 72 mg/dL |
| Triglyceride | 1500 mg/dL |
| Unconjugated Bilirubin | 40 mg/dL |
| Uric acid | 23.5 mg/dL |
| Xylitol | 0.09 mg/dL |
| Xylose | 50 mg/dL |
The sponsor includes the following statement in their labeling:
- Do not use during or soon after xylose absorption testing since xylose may cause inaccurate glucose results. Ask your doctor how long to wait before performing a glucose test.
4. Assay Reportable Range:
20 to 600 mg/dL glucose
5. Traceability, Stability, Expected Values (Controls, Calibrators, or Methods):
Traceability
The glucose measurement functionality of the Sejoy Blood Glucose Monitoring System and Sejoy Advance Link Blood Glucose Monitoring System is traceable to the NIST SRM 917c glucose reference material. The method comparison/lay-user study was performed using the YSI 2300 STAT Plus Glucose Analyzer as the comparator method (see section VII.C.3).
Test Strip Stability
Test strip stability was assessed using real time stability and accelerated stability studies. Protocols and acceptance criteria were reviewed and found acceptable to support the labeling claims that the test strips are stable for 6 months after first being opened, and that closed vials and individual foil wrapped test strips are stable for 24 months when stored at the recommended storage temperatures 33.8°F - 86°F (1°C - 30°C) and 10 - 90% relative humidity. The labeling instructs the users not to freeze the test strips.
6. Detection Limit:
The reportable range is 20 to 600 mg/dL.
7. Assay Cut-Off:
Not applicable
K240640 - Page 9 of 13
{9}
8. Accuracy (Instrument):
Not applicable
9. Carry-Over:
Not applicable
## B Comparison Studies:
1. Method Comparison with Predicate Device:
Please refer to lay user study below in section VII.C3.
2. Matrix Comparison:
Not applicable. The device is only intended for use with fresh capillary whole blood drawn from the fingertip.
## C Clinical Studies:
1. Clinical Sensitivity:
Not applicable
2. Clinical Specificity:
Not applicable
3. Other Clinical Supportive Data (When 1. and 2. Are Not Applicable):
## Method Comparison/Lay User Performance Study
To assess the performance of the Sejoy Blood Glucose Monitoring System and Sejoy Advanced Link Blood Glucose Monitoring System in the hands of the intended lay users, the sponsor conducted a study involving 352 lay user participants who collected and tested samples from their own fingertip using only the instructions from the product labeling in English. This study used the Sejoy Advanced Link Blood Glucose Monitoring System as the more complicated of the two candidate systems and 3 test strip lots. Results were analyzed by comparing blood glucose results obtained by the lay users with the Sejoy Advanced Link System against results obtained using a laboratory comparator method (YSI 2300 analyzer). Glucose concentrations in the samples ranged from approximately 64.2 to 491.5 mg/dL as measured by the comparator method, which included 45 unaltered samples below 80 mg/dL and 51 samples above 250 mg/dL. Results are summarized in the tables below:
| Within ± 5% | Within ± 10% | Within ± 15% | Within ± 20% |
| --- | --- | --- | --- |
| 200/352
(56.8%) | 318/352
(90.3%) | 346/352
(98.3%) | 352/352
(100.0%) |
K240640 - Page 10 of 13
{10}
K240640 - Page 11 of 13
| Regression Analysis | | |
| --- | --- | --- |
| Slope | y-intercept | R²-value |
| 1.024 | 1.204 | 0.987 |
## Usability
At the end of the lay-user study, each participant was asked to complete a usability questionnaire regarding ease of understanding of information in the user manual and the ease of use when performing a blood glucose test. The sponsor's analysis of the questionnaire responses demonstrated that the participants were satisfied with the ease of operation by following the instructions for use in the user's manual and with the overall performance of the Sejoy Advanced Link Blood Glucose Monitoring system.
## Labeling Readability
A Flesch-Kincaid readability assessment was conducted on the user's manual, test strip insert, and quick reference guide that demonstrated a readability level at an 8th grade level or lower.
## Extreme Glucose Study
An accuracy study was performed to evaluate the performance of the Sejoy Advanced Link Blood Glucose Monitoring System with 100 capillary blood samples containing extreme glucose concentrations at the extreme lower and upper ends of the claimed glucose measuring range. Glucose concentrations below 80 mg/dL were achieved with 50 samples altered by glycolysis and glucose concentrations greater than 250 mg/dL were achieved with 50 samples spiked with glucose. Results on the candidate device using 3 test strip lots were compared to the results obtained using the comparator method (YSI 2300 analyzer) and are summarized below:
| Glucose Range | Within ± 5% | Within ± 10% | Within ± 15% | Within ± 20% |
| --- | --- | --- | --- | --- |
| Extreme Low (<80 mg/dL) | 27/50 (54.0%) | 46/50 (92.0%) | 50/50 (100%) | 50/50 (100%) |
| Extreme High (>250 mg/dL) | 30/50 (60.0%) | 48/50 (96.0%) | 50/50 (100%) | 50/50 (100%) |
## D Clinical Cut-Off:
Not applicable.
## E Expected Values/Reference Range:
The sponsor includes the following expected blood glucose values for people without diabetes in the labeling:
< 100 mg/dL fasting
<140 mg/dL 2 hours after a meal
Reference: American Diabetes Association; Standards of Care in Diabetes-2023
{11}
F Other Supportive Instrument Performance Characteristics Data:
1. Hematocrit Study
The effect of different hematocrit levels was evaluated using venous whole blood samples with hematocrit levels of 15 to 60% (15%, 20%, 25%, 30%, 35%, 42%, 45%, 50%, 55%, and 60%) at 5 levels of glucose (30 – 50, 51 – 110, 111 – 150, 151 – 250, and 251 – 400 mg/dL). Each sample was tested in replicates of 10 using 3 test strip lots. The results obtained with the candidate system were compared to those obtained with the comparator method (YSI 2300 analyzer). The results demonstrated acceptable performance of the candidate blood glucose monitoring systems across the claimed hematocrit range of 15 to 60%.
2. System Operating Conditions Study
To evaluate device performance over varying operating temperature and humidity conditions, venous whole blood samples were adjusted to 5 glucose levels (20-50, 51-110, 111-150, 151-250, and 251-600 mg/dL). There were four extreme temperature and humidity combinations tested (high temperature/low humidity 45°C/10%, low temperature/high humidity 5°C/90%, high temperature/high humidity 45°C/90%, and low temperature/low humidity 5°C/10%). The results obtained with the candidate system were compared to those obtained with the comparator method (YSI 2300 analyzer). The results support the claims in the labeling that the candidate blood glucose monitoring systems can be used in conditions of 41 to 113°F (5 to 45°C) with relative humidity of 10 to 90%.
3. Altitude Effects Study
To evaluate the effects of altitude, a study was conducted using fingerstick whole blood samples altered to achieve 5 glucose concentrations ranging from 40.0 to 319.2 mg/dL. Each sample was tested using 3 test strip lots and 2 glucose in a pressurized chamber with corresponding pressure and oxygen content at 8 simulated altitudes from 0 to 11,482.9 ft. The results of obtained with the candidate system were compared to those obtained with the comparator method (YSI 2300 analyzer). The results support the claims that the candidate blood glucose monitoring systems function as intended at altitudes up to the claimed altitude of 10,000 feet.
4. Sample Volume Study
Venous blood samples were prepared at 3 glucose range intervals (0-65 mg/dL, 100-120 mg/dL, and 200-250 mg/dL) and were tested at different sample volumes starting at 1.0 µL and incrementally decreased by 0.1 µL until an error code appeared. The results of obtained with the candidate system were compared to those obtained with the comparator method (YSI 2300 analyzer). The sponsor provided validation studies demonstrating that with blood volumes below 0.6 µL, the insufficient sample volume error message functioned as intended. The results support the claimed minimum sample volume of 0.6 µL for the candidate blood glucose monitoring systems.
5. Flex Studies
The following additional flex studies were performed with the candidate system: drop/shock, early test strip removal, intermittent sampling, sample outside measuring range, sample perturbation, shipping, used test strips and vibration testing. The results demonstrated that the performance of the candidate blood glucose monitoring systems are robust under these conditions.
K240640 - Page 12 of 13
{12}
6. Electrical Safety and Electromagnetic Compatibility (EMC) Testing
The sponsor provided documentation certifying that acceptable electrical safety and (EMC) testing had been performed. The Sejoy blood glucose monitoring system was found to be compliant.
7. Infection Control Testing
The device systems are intended for single-patient use only. Disinfection efficacy studies were performed on the external meter materials by an outside commercial testing laboratory demonstrating complete inactivation of hepatitis B virus (HBV) with the chosen disinfectant, CaviWipes Disinfecting Towelettes (EPA Registration # 46781-17). A robustness study was also conducted by the sponsor demonstrating that there was no change in performance nor in the external materials of the meter after 4000 cleaning and disinfection cycles using the CaviWipes Disinfecting Towelettes. The robustness studies were designed to simulate 3 years of single-patient device use. Labeling was reviewed for adequate instructions for the validated cleaning and disinfection procedures.
8. Glucose Test Strip Lot Release Protocol
The test strip lot release protocol and criteria were reviewed and found to be acceptable.
VIII Proposed Labeling:
The labeling supports the finding of substantial equivalence for this device.
IX Conclusion:
The submitted information in this premarket notification is complete and supports a substantial equivalence decision.
K240640 - Page 13 of 13
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.