CustomizedBone Service (CustomizedBone) is intended to replace bony voids in the cranial and/or craniofacial skeleton (frontal bone including the brow ridge). This device is indicated for both adult and pediatric use (for children 7 years of age and above). CustomizedBone implants are suitable for reconstructing craniofacial defects resulting from: - · trauma and vascular pathologies, either associated to cranial decompression, - · removal of tumours: - · reabsorption of autologous bone; - · rejection of other prosthetic materials; - · congenital malformations.
Device Story
CustomizedBone Service provides patient-specific implants for cranial/craniofacial reconstruction. Input: patient CT scan data. Process: design and production of porous biomimetic hydroxyapatite implants based on surgeon specifications and patient anatomy. Output: custom-shaped hydroxyapatite implant with trabecular structure (macro, micro, and interconnecting pores). Usage: clinical setting; operated by surgeons. Fixation: utilizes PEEK OPTIMA™ cranial bone fixation clamps (Cranial LOOP (L)). Benefit: restores bony voids with biocompatible material resembling human bone mineral component.
Clinical Evidence
Bench testing only. Validation included stability verification of the implant when stabilized with PEEK OPTIMA™ clamps and usability testing of the alternative fixation method.
Technological Characteristics
Material: porous biomimetic hydroxyapatite (resembles human bone mineral). Structure: trabecular with macro, micro, and interconnecting pores. Fixation: PEEK OPTIMA™ cranial bone fixation clamps. Customization: patient-specific design based on CT scan data.
Indications for Use
Indicated for adult and pediatric patients (≥7 years) requiring replacement of bony voids in the cranial or craniofacial skeleton (including frontal bone/brow ridge) due to trauma, vascular pathologies, tumor removal, autologous bone reabsorption, prosthetic rejection, or congenital malformations.
Regulatory Classification
Identification
A preformed nonalterable cranioplasty plate is a device that is implanted in a patient to repair a skull defect and is constructed of a material, e.g., stainless steel or vitallium, that cannot be altered or reshaped at the time of surgery without changing the chemical behavior of the material.
{0}------------------------------------------------
March 28, 2024
Image /page/0/Picture/1 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
Fin-ceramica faenza s.p.a. % Stephanie Perryman Fortrea 5353 Wayzata Boulevard, Suite 505 Minneapolis, Minnesota 55416-1334
Re: K240567
Trade/Device Name: CustomizedBone Service Regulation Number: 21 CFR 882.5330 Regulation Name: Preformed Nonalterable Cranioplasty Plate Regulatory Class: Class II Product Code: GXN, PJN Dated: February 26, 2024 Received: February 29, 2024
Dear Stephanie Perryman:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
{1}------------------------------------------------
Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review. the OS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Adam D. Digitally signed by Adam D. Pierce -S Digitally signed by Pierce -S 16:45:07-04'00'
Adam D. Pierce, Ph.D. Assistant Director DHT5A: Division of Neurosurgical, Neurointerventional and Neurodiagnostic Devices OHT5: Office of Neurological
{2}------------------------------------------------
and Physical Medicine Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{3}------------------------------------------------
#### Indications for Use
Submission Number (if known)
K240567
Device Name
CustomizedBone Service
Indications for Use (Describe)
CustomizedBone Service (CustomizedBone) is intended to replace bony voids in the cranial and/or craniofacial skeleton (frontal bone including the brow ridge). This device is indicated for both adult and pediatric use (for children 7 years of age and above).
CustomizedBone implants are suitable for reconstructing craniofacial defects resulting from:
- · trauma and vascular pathologies, either associated to cranial decompression,
- · removal of tumours:
- · reabsorption of autologous bone;
- · rejection of other prosthetic materials;
- · congenital malformations.
Type of Use (Select one or both, as applicable)
Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours be response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{4}------------------------------------------------
Prepared on: 2024-03-27
21 CFR 807.92(a)(4)
| Contact Details | | | 21 CFR 807.92(a)(1) |
|------------------------------------|---------------------------------------------------------------------------|--------------|---------------------|
| Applicant Name | Fin-ceramica faenza s.p.a. | | |
| Applicant Address | Via Ravegnana 186 Faenza RA 48018 Italy | | |
| Applicant Contact Telephone | +39 0546 607311 | | |
| Applicant Contact | Marina Monticelli | | |
| Applicant Contact Email | mmonticelli@finceramica.it | | |
| Correspondent Name | Fortrea | | |
| Correspondent Address | 5353 Wayzata Boulevard, Suite 505 Minneapolis MN 55416-1334 United States | | |
| Correspondent Contact Telephone | 612-476-4371 | | |
| Correspondent Contact | Stephanie Perryman | | |
| Correspondent Contact Email | stephanie.perryman@fortrea.com | | |
| Device Name | | | 21 CFR 807.92(a)(2) |
| Device Trade Name | CustomizedBone Service | | |
| Common Name | Preformed nonalterable cranioplasty plate | | |
| Classification Name | Plate, Cranioplasty, Preformed, Non-Alterable | | |
| Regulation Number | 882.5330 | | |
| Product Code(s) | GXN, PJN | | |
| Legally Marketed Predicate Devices | | | 21 CFR 807.92(a)(3) |
| Predicate # | Predicate Trade Name (Primary Predicate is listed first) | Product Code | |
| K193547 | CustomizedBone Service | GXN | |
510(k) Sun
arv
## Device Description Summary
Hydroxyapatite is contained in human bones in percentage close to 70% and is one of the most important elements of human bone structure. CustomizedBone Service patient specific implants for the reconstruction of cranial/craniofacial defects are made of porous biomimetic hydroxyapatite with a chemical composition and structure that resembles the mineral component of human bones. This biomaterial is highly porous with trabecular structure and is composed of pores with the following characteristics:
· macro-pores,
• interconnecting pores,
• micro-pores.
This material is biocompatible.
The implants are designed and produced by Fin-Ceramica Faenza according to the surgeon's specifications and based on the patient's CT scan data, obtained through a standardized protocol. During the pre-operative planning phase, the final
{5}------------------------------------------------
implant design. All the implants are accompanied by the patient's identification code.
# Intended Use/Indications for Use
CustomizedBone Service (CustomizedBone) is intended to replace bony voids in the cranial and/or craniofacial skeleton (frontal bone including the brow ridge). This device is indicated for both adult and pediatric use (for children 7 years of above). CustomizedBone implants are suitable for reconstructing craniofacial defects resulting from:
- · trauma and vascular pathologies, either associated or non-associated to cranial decompression;
• removal of tumours;
- · reabsorption of autologous bone;
- · rejection of other prosthetic materials;
- · congenital malformations.
# Indications for Use Comparison
The indications for use of the subject device remain the same as the predicate device.
## Technological Comparison
The design and materials of the implant remain the same as the predicate device. The only change between the subject device and predicate device is the introduction of an alternative fixation method (Cranial bone fixation clamp made of biocompatible PEEK OPTIMA™ - Cranial LOOP (L) manufactured by NEOS Surgery S.L. or equivalent). The new fixation method has been validated and the subject device can be considered substantially equivalent to the predicate device.
#### Non-Clinical and/or Clinical Tests Summary & Conclusions 21 CFR 807.92(b)
The following tests were completed to validate the effectiveness of the implantation with PEEK OPTIMA™ clamps:
· Stability verification of CustomizedBone Service stabilized with Cranial LOOP (L)
• Usability of an alternative method of fixation for CustomizedBone Service product
Based on the above testing, fixation of the CustomizedBone Service with PEEK OPTIMA™ clamps (Cranial LOOP (L) manufactured by NEOS Surgery S.L. or equivalent) has been validated and the can be considered substantially equivalent to the predicate device.
Image /page/5/Picture/17 description: The image is a blank white square. The square is uniformly white, with no discernible features or patterns. A thin gray line is present on the right edge of the square, possibly indicating a border or edge of the image.
### 21 CFR 807.92(a)(5)
### 21 CFR 807.92(a)(6)
21 CFR 807.92(a)(5)
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.