Intended to produce oxygen for emergency use at 6 LPM flow rate for at least 15 minutes (90 liters) for individuals 17 years and older requiring emergency or supplemental oxygen.
Device Story
Chemical oxygen generator; single-use; provides supplemental oxygen for emergency, home, or commercial use. Operation: user opens lid, turns dial clockwise 2-3 times to initiate friction interaction between integrated phosphorous match and candle block; chemical reaction produces oxygen. Output: minimum 90 liters of oxygen at ≥96% purity, average flow rate ≥6 LPM for ≥15 minutes. Device is OTC; user-operated. Patient places face mask (user-supplied) over face to receive oxygen. Benefits: provides immediate supplemental oxygen in locations where compressed cylinders are unsuitable.
Clinical Evidence
Bench testing only. Performance testing included casing and gas temperature, shock and vibration, drop testing, sand and rain blow testing, and flow duration. Biocompatibility testing performed per ISO 18562-2 and ISO 18562-3. All results met pre-defined acceptance criteria.
Technological Characteristics
Chemical oxygen generator; single-use; disposable. Dimensions: 28.5 cm x 11.5 cm; weight: 1.5 kg. Principle: friction-initiated chemical reaction. Housing temperature <45°C; gas outlet temperature <40°C. Storage temperature: 0° to +40°C. Biocompatibility per ISO 18562-2/3.
Indications for Use
Indicated for individuals 17 years and older requiring emergency or supplemental oxygen.
Regulatory Classification
Identification
A portable oxygen generator is a device that is intended to release oxygen for respiratory therapy by means of either a chemical reaction or physical means (e.g., a molecular sieve).
{0}------------------------------------------------
December 6, 2024
Image /page/0/Picture/1 description: The image shows the logos of the Department of Health & Human Services and the U.S. Food & Drug Administration (FDA). The Department of Health & Human Services logo is on the left, featuring a stylized human figure. To the right is the FDA logo, with the letters "FDA" in a blue square and the words "U.S. FOOD & DRUG ADMINISTRATION" in blue text.
Molecular Products Ltd. % Paul Dryden President ProMedic. LLC 131 Bay Point Drive NE Saint Petersburg, Florida 33704
Re: K240561
Trade/Device Name: Rugged Oxygen Generator (ROG) Regulation Number: 21 CFR 868.5440 Regulation Name: Portable oxygen generator Regulatory Class: Class II Product Code: CAW Dated: November 7, 2024 Received: November 8, 2024
Dear Paul Dryden:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's
{1}------------------------------------------------
requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801; medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Bradley Q. Quinn -S
Bradley Quinn Assistant Director DHT1C: Division of Anesthesia. Respiratory, and Sleep Devices OHT1: Office of Ophthalmic, Anesthesia, Respiratory, ENT, and Dental Devices Office of Product Evaluation and Ouality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K240561
Device Name Rugged Oxygen Generator (ROG)
Indications for Use (Describe)
Intended to produce oxygen for emergency use at 6 LPM flow rate for at least 15 minutes (90 liters) for individuals 17 years and older requiring emergency or supplemental oxygen.
| Type of Use (Select one or both, as applicable) | |
|-------------------------------------------------|--|
|-------------------------------------------------|--|
Prescription Use (Part 21 CFR 801 Subpart D)
|X | Over-The-Counter Use (21 CFR 801 Subpart C)
# CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
# *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
# 510(k) Summarv Page 1 of 4
| Date Prepared: | December 4, 2024 |
|--------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------|
| Sponsor: | Molecular Products Ltd.<br>Parkway, Harlow Business Park, Harlow.<br>Essex, CM19 5FR, UK |
| Sponsor Contact: | Joe Atkinson - Technical Manager<br>Phone - +44 (0)1279 401236 |
| Submission Correspondent: | Paul Dryden<br>ProMedic, LLC |
| Proprietary or Trade Name:<br>Common/Usual Name:<br>Classification Name:<br>Product Code:<br>Regulation Number: | Rugged Oxygen Generator (ROG)<br>Oxygen Generator<br>Generator, Oxygen, Portable<br>CAW<br>21 CFR 868.5440 |
| Predicate Device:<br>Proprietary or Trade Name:<br>Common/Usual Name:<br>Classification Name:<br>Product Code:<br>Regulation Number: | ROGS - K131016<br>Rugged Oxygen Generator (ROG)<br>Oxygen Generator<br>Generator, Oxygen, Portable<br>CAW<br>21 CFR 868.5440 |
# Device Description
The Rugged Oxygen Generator (ROG) is intended for environment such as emergency use, home use, and commercial use and is OTC. The ROG is designed to provide supplemental oxygen and is a single use product. The ROG is a chemical oxygen generator that produces a minimum of 90 liters of oxygen at ≥96% purity for minimum of 15 minutes with an average flow rate of 6 liters per minute or greater. There is a need for oxygen in a variety of applications/locations where compressed oxygen cylinders are not suitable.
#### Principle of Operation
The subject device is initiated though a friction interaction between an integrated phosphorous match and the candle block. To initiate, open the lid and turn the dial clockwise 2-3 times. The turning of the dial will rotate the match and screw the match down into the candle block. This will start the chemical reaction, which will produce oxygen. Once oxygen is flowing, close the lid and place the face mask onto patient.
# Indications for Use:
Intended to produce oxygen for emergency use at 6 LPM flow rate for at least 15 minutes (90 liters) for individuals 17 years and older requiring emergency or supplemental oxygen.
#### Patient Population:
For individuals 17 years and older requiring emergency or supplemental oxygen.
#### Environments of use:
Emergency use, home use, and commercial use
{4}------------------------------------------------
# 510(k) Summary Page 2 of 4
| | Subject Device | Predicate device | Comparison |
|---------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| | Molecular Products | Molecular Products ROG | |
| | Rugged Oxygen Generator (ROG) | | |
| K# | K240561 | K131016 | |
| Product Code | CAW | CAW | Identical |
| CFR | 21 CFR 868.5440 | 21 CFR 868.5440 | Identical |
| Indications for Use | Intended to produce oxygen for emergency use at 6<br>LPM flow rate for at least 15 minutes (90 liters) for<br>individuals 17 years and older requiring emergency or<br>supplemental oxygen | Intended to produce oxygen for emergency use at 6<br>LPM flow rate for at least 15 minutes (90 liters) | Similar<br>Population was not specified |
| Patient Population | For individuals 17 years and older requiring<br>emergency or supplemental oxygen. | Individuals requiring emergency or supplemental<br>oxygen. | Similar<br>Population was not specified |
| Environment of Use | Emergency, OTC, home and commercial settings | Emergency, OTC, home and commercial settings | Identical |
| Contraindications | Same | Same | Identical |
| Method for oxygen<br>generation | Chemical reaction | Chemical reaction | Identical |
| Patient Interface | Standard oxygen mask | Standard oxygen mask or nasal Standard oxygen<br>mask or nasal cannula | The subject device does not<br>offer a nasal cannula for use,<br>which does not raise different<br>safety or efficacy as a mask is<br>still available for use like the<br>predicate. |
| Specifications | % Oxygen - 96%<br>Flow rate minimum - 6 LPM<br>Duration at least 15 minutes<br>Initiation of oxygen flow - 5 seconds | % Oxygen - 99%<br>Flow rate minimum - 6 LPM<br>Duration at least 15 minutes<br>Initiation of oxygen flow - 5 seconds | Oxygen purity is not a FDA<br>guidance specification. USP<br>defines medical oxygen purity<br>as > 93%. This change in the<br>specification does not<br>Raise different risk concerns. |
| Single Use, Disposable | Yes | Yes | Identical |
| OTC | Yes | Yes | Identical |
| VOC / PM Testing | Yes | Yes | Identical |
| Housing temperature | <45°C Maximum | <45°C Maximum | Identical |
| Temperature of gas at<br>outlet | <40°C Maximum | <40°C Maximum | Identical |
| Storage temperatures | 0° to + 40°C / 32° to +104°F | 0° to + 40°C / 32° to +104°F | Identical |
| Dimensions | 28.5 cm x 11.5 cm | 28.5 cm x 11.5 cm | Similar |
| Gross weight | 1.5kg | 1.5kg | Similar |
| Accessories | None - User supplied | Mask and oxygen cannula | Mask and tubing are user<br>supplied |
{5}------------------------------------------------
### Substantial Equivalence Discussion
The table above compares the key features of the proposed device with the identified predicate -ROG - K131016. The comparison demonstrates that the subject device can be found to be substantially equivalent to the predicate device.
# Indications for Use -
The indications for use are similar for the subject device when compared to the predicate device.
Discussion - The indications for use are similar.
#### Technology and construction -
The technology and principle of operation is similar for the subject device when compared to the predicate device. Discussion - Both the subject and predicate device have identical principles of Operation.
#### Environment of Use -
The environments of use are identical to predicate, which are Emergency, OTC, home and commercial settings.
Discussion - The environments of use are the identical.
#### Patient Population -
The patient population is similar for individuals requiring emergency oxygen. Discussion - The subject and predicate device patient populations are similar.
# Non-Clinical Testing Summarv
#### Bench testing -
Performance testing demonstrated that the subject device met its acceptance criteria. Testing included:
- Casing, Gas at patient temperature testing .
- Shock and Vibration
- Drop Testing .
- Sand and Rain Blow Testing ●
- Flow Duration .
Discussion - The test results are similar to the predicate and within pre-defined acceptance criteria.
#### Biocompatibility -
The sponsor performed ISO 18562-2 and ISO 18562-3 testing.
Discussion - The subject device was found to meet the applicable requirements for biocompatibility safety for the intended population
#### Discussion of Differences
The subject does not include oxygen tubing and mask as was included by the predicate device. The oxygen percent is 96% as compared to the predicate, which was 99%. The flow rate minimum, duration, and initiation of oxygen remain unchanged. The subject device also incorporates design updates from Molecular Product's own device - the predicate ROG - K131016.
{6}------------------------------------------------
The changes incorporated in the subject device do not raise any new concerns of safety of effectiveness.
# Substantial Equivalence Conclusion
Through testing outlined above, the sponsor has demonstrated that the subject device comparably performs to the predicate, and the subject device do not raise any new concerns of safety of effectiveness when compared to the predicate ROG - K131016.
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.