uNion® MAX Cervical Plate System

K240515 · Ulrich Medical USA, Inc. · KWQ · Mar 18, 2024 · Orthopedic

Device Facts

Record IDK240515
Device NameuNion® MAX Cervical Plate System
ApplicantUlrich Medical USA, Inc.
Product CodeKWQ · Orthopedic
Decision DateMar 18, 2024
DecisionSESE
Submission TypeSpecial
Regulation21 CFR 888.3060
Device ClassClass 2
AttributesTherapeutic

Intended Use

The uNion MAX Cervical Plate System is intended for anterior fixation of the cervical spine (C2 to T1). The system is to be used to provide stabilization of the anterior cervical spine as an adjunct to fusion for the treatment of degenerative disc disease (defined as neck pain of discogenic origin with the degeneration of the disc confirmed by history and radiographic studies), spondylolisthesis, trauma (i.e., fractures or dislocations), tumors, spinal stenosis, deformity (i.e., kyphosis, lordosis or scoliosis), pseudarthrosis or failed previous fusion.

Device Story

Anterior cervical plate and screw system; provides mechanical support/stabilization across cervical spine levels (C2-T1) until fusion achieved. Components: 1-4 level plates (8mm-84mm); Ø4.0mm/Ø4.5mm fixed and variable screws (self-tapping/self-drilling). Plates feature pivoting locking mechanism to prevent screw backout. Used in surgical setting by surgeons; implanted to provide rigid fixation. Output: mechanical stabilization of spinal segments. Benefit: facilitates spinal fusion; treats instability/deformity.

Clinical Evidence

Bench testing only. ASTM F1717 mechanical testing performed on worst-case constructs, including static and dynamic compression bending and static torsion. Results demonstrate performance equivalent to predicate devices.

Technological Characteristics

Materials: Ti-6Al-4V ELI titanium alloy (ASTM F136). Configuration: Plates (1-4 level) and screws (fixed/variable, self-tapping/self-drilling). Locking mechanism: Pivoting. Dimensions: 8mm-84mm plate lengths. Energy source: None (mechanical). Sterilization: Non-sterile.

Indications for Use

Indicated for anterior cervical spine (C2-T1) stabilization as an adjunct to fusion in patients with degenerative disc disease, spondylolisthesis, trauma (fractures/dislocations), tumors, spinal stenosis, deformity (kyphosis, lordosis, scoliosis), pseudarthrosis, or failed previous fusion.

Regulatory Classification

Identification

A spinal intervertebral body fixation orthosis is a device intended to be implanted made of titanium. It consists of various vertebral plates that are punched into each of a series of vertebral bodies. An eye-type screw is inserted in a hole in the center of each of the plates. A braided cable is threaded through each eye-type screw. The cable is tightened with a tension device and it is fastened or crimped at each eye-type screw. The device is used to apply force to a series of vertebrae to correct “sway back,” scoliosis (lateral curvature of the spine), or other conditions.

Predicate Devices

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ March 18, 2024 Image /page/0/Picture/1 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA logo on the right. The FDA logo features the letters 'FDA' in a blue square, followed by the words 'U.S. FOOD & DRUG ADMINISTRATION' in blue text. Ulrich Medical USA, Inc. Eric Lucas Vice President of Technology 3700 East Plano Parkway Suite 200 Plano, Texas 75074 Re: K240515 Trade/Device Name: uNion® MAX Cervical Plate System Regulation Number: 21 CFR 888.3060 Regulation Name: Spinal Intervertebral Body Fixation Orthosis Regulatory Class: Class II Product Code: KWQ Dated: January 23, 2024 Received: February 23, 2024 Dear Eric Lucas: We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download). {1}------------------------------------------------ Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review. the OS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181). Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems. For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100). Sincerely, Image /page/1/Picture/7 description: The image contains the text "Colin O'neill -S" in a simple, sans-serif font. The text is arranged vertically, with "Colin" on the top line and "O'neill -S" on the bottom line. The text is black, and the background is white. There is a faint FDA logo in the background. Colin O'Neill, M.B.E. Assistant Director DHT6B: Division of Spinal Devices OHT6: Office of Orthopedic Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health {2}------------------------------------------------ # Indications for Use 510(k) Number (if known) K240515 Device Name uNion® MAX Cervical Plate System #### Indications for Use (Describe) The uNion MAX Cervical Plate System is intended for anterior fixation of the cervical spine (C2 to T1). The system is to be used to provide stabilization of the anterior cervical spine as an adjunct to fusion for the treatment of degenerative disc disease (defined as neck pain of discogenic origin with the degeneration of the disc confirmed by history and radiographic studies), spondylolisthesis, trauma (i.e., fractures or dislocations), tumors, spinal stenosis, deformity (i.e., kyphosis, lordosis or scoliosis), pseudarthrosis or failed previous fusion. | Type of Use (Select one or both, as applicable) | | |-------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------| | <span style="font-family: Arial, sans-serif;"> <input checked="true" type="checkbox"/> Prescription Use (Part 21 CFR 801 Subpart D) </span> | <span style="font-family: Arial, sans-serif;"> <input type="checkbox"/> Over-The-Counter Use (21 CFR 801 Subpart C) </span> | ## CONTINUE ON A SEPARATE PAGE IF NEEDED. This section applies only to requirements of the Paperwork Reduction Act of 1995. ### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: > Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." {3}------------------------------------------------ # 510(k) Summary Image /page/3/Picture/2 description: The image shows the logo for Ulrich medical USA. The logo consists of the word "Ulrich" in a bold, sans-serif font, with the "U" stylized as a black square with a white "u" inside. Below the word "Ulrich" is the phrase "medical USA" in a smaller, sans-serif font. | Date: | 23 January 2024 | |-------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------| | Sponsor: | ulrich medical USA, Inc.<br>3700 East Plano Pkwy. Suite 200 | | | Plano, TX 75074<br>regulatory@ulrichmedicalusa.com | | | Phone: 469.238.0800 | | Sponsor Contact: | Eric Lucas PhD, Vice President of Technology | | Proposed Trade<br>Name: | uNion® MAX Cervical Plate System | | Common Name: | Anterior cervical plate system | | Device Classification: | Class II | | Classification Name: | Spinal intervertebral body fixation orthosis | | Regulation: | 888.3060 | | Device Product Code: | KWQ | | Device Description: | uNion MAX is an anterior cervical plate and screw system used to provide<br>mechanical support across implanted level(s) in the cervical spine until fusion<br>is achieved. The components include Ø4.0mm and Ø4.5mm fixed and<br>variable screws having self-tapping, self-drilling or self-drilling/self-tapping<br>tips and one- through four-level plates having lengths from 8mm to 84mm.<br>The plates incorporate a pivoting locking mechanism which functions to block<br>screw backout. The devices are sold non-sterile. | | Indications for Use: | The uNion MAX Cervical Plate System is intended for anterior fixation of the<br>cervical spine (C2 to T1). The system is to be used to provide stabilization of<br>the anterior cervical spine as an adjunct to fusion for the treatment of<br>degenerative disc disease (defined as neck pain of discogenic origin with the<br>degeneration of the disc confirmed by history and radiographic studies),<br>spondylolisthesis, trauma (i.e., fractures or dislocations), tumors, spinal<br>stenosis, deformity (i.e., kyphosis, lordosis or scoliosis), pseudarthrosis or<br>failed previous fusion. | | Materials: | uNion MAX components are manufactured from Ti-6Al-4V ELI titanium alloy<br>(ASTM F136). | | Primary Predicate: | uNion Cervical Plate System (ulrich medical USA - K150666) | | Predicate/Reference: | MaxAn® Anterior Cervical Plate System (Biomet Spine - K133518) | | Performance Data: | ASTM F1717 mechanical testing of the worst case uNion MAX constructs<br>included static and dynamic compression bending and static torsion. The<br>mechanical test results demonstrate that uNion MAX performance is<br>substantially equivalent to the predicate devices. | {4}------------------------------------------------ | Technological<br>Characteristics: | uNion MAX Cervical Plate System possesses similar technological<br>characteristics as one or more of the predicate devices. These include:<br>basic design (plates and screws configuration), material (titanium alloy) and sizes (dimensions are comparable to those offered by the predicate systems) The fundamental scientific technology of the uNion MAX system is the same as previously cleared devices. | |-----------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------| | Conclusion: | The uNion MAX Cervical Plate System possesses the same intended use<br>and similar technological characteristics as the predicate devices. Therefore<br>uNion MAX is substantially equivalent for its intended use. |
Innolitics
510(k) Summary
Decision Summary
Classification Order
Enter a record ID and click Load to view the document.
100%