The Eminent Spine SI Screw System is intended for sacroiliac joint fusion for conditions including degenerative sacroiliitis and sacroiliac joint disruptions, to augment immobilization and stabilization of the sacroiliac joint in skeletally mature patients undergoing sacropelvic fixation as part of a lumbar or thoracolumbar fusion and acute, non-acute, and non-traumatic fractures involving the sacroiliac joint. This includes those whose symptoms began during pregnancy or in the period and have persisted postpartum for more than 6 months.
Device Story
The Eminent Spine SI Screw System consists of cannulated, metallic bone fixation screws designed for sacroiliac joint fusion. The system utilizes screws manufactured from Ti-6Al-4V ELI via traditional machining or 3D-printing. Screws feature open, porous graft windows to facilitate bone ingrowth and may include optional locking crowns or fangs to improve anatomical conformity. The device is used by surgeons in a clinical/OR setting to provide immobilization and stabilization of the sacroiliac joint. The surgeon selects appropriate screw lengths and diameters based on patient anatomy to achieve fixation. The device benefits patients by stabilizing the joint, addressing degenerative conditions or fractures, and supporting fusion in patients undergoing broader spinal stabilization procedures.
Clinical Evidence
Bench testing only. Mechanical testing included static torsion, driving torque, axial pullout, and static/dynamic cantilever bend testing performed in accordance with ASTM F3574. No clinical data was provided.
Technological Characteristics
Materials: Ti-6Al-4V ELI (ASTM F136 for machined, ASTM F3001 for 3D-printed). Form factor: Cannulated screws with open/porous graft windows; optional locking crowns/fangs. Dimensions: Lengths 20-90mm; Diameters Ø8.5, Ø10.0, Ø11.5, Ø13.0mm. Energy source: None (mechanical). Connectivity: None. Sterilization: Not specified.
Indications for Use
Indicated for skeletally mature patients requiring sacroiliac joint fusion due to degenerative sacroiliitis, sacroiliac joint disruptions, sacropelvic fixation (as part of lumbar/thoracolumbar fusion), or acute/non-acute/non-traumatic fractures. Includes patients with postpartum symptoms persisting >6 months.
Regulatory Classification
Identification
A smooth or threaded metallic bone fixation fastener is a device intended to be implanted that consists of a stiff wire segment or rod made of alloys, such as cobalt-chromium-molybdenum and stainless steel, and that may be smooth on the outside, fully or partially threaded, straight or U-shaped; and may be either blunt pointed, sharp pointed, or have a formed, slotted head on the end. It may be used for fixation of bone fractures, for bone reconstructions, as a guide pin for insertion of other implants, or it may be implanted through the skin so that a pulling force (traction) may be applied to the skeletal system.
Predicate Devices
Entasis 3D Dual-Lead Sacroiliac Implant System (K223708)
{0}------------------------------------------------
June 4, 2024
Image /page/0/Picture/1 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA text logo on the right. The FDA text logo is in blue and includes the FDA acronym in a blue square, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in a sans-serif font.
Eminent Spine, LLC % Chhavy Tep-Cullison Regulatory Affairs Specialist Jalex Medical 27865 Clemens Road, Suite 3 Westlake. Ohio 44145
Re: K240505
Trade/Device Name: Eminent Spine SI Screw System Regulation Number: 21 CFR 888.3040 Regulation Name: Smooth Or Threaded Metallic Bone Fixation Fastener Regulatory Class: Class II Product Code: OUR Dated: February 21, 2024 Received: May 13, 2024
Dear Chhavy Tep-Cullison:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
{1}------------------------------------------------
Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review. the OS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
> Sincerely, Colin O'neill -S
Colin O'Neill, MBE Assistant Director DHT6B: Division of Spinal Devices OHT6: Office of Orthopedic Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
Submission Number (if known)
K240505
Device Name
Eminent Spine SI Screw System
#### Indications for Use (Describe)
The Eminent Spine Sl Screw System is intended for sacroiliac joint fusion for conditions including degenerative sacroillitis and sacroiliac joint disruptions, to augment immobilization and stabilization of the sacroiliac joint in skeletally mature patients undergoing sacropelvic fixation as part of a lumbar or thoracolumbar fusion and acute, non-acute, and non-traumatic fractures involving the sacroiliac joint. This includes those whose symptoms began during pregnancy or in the period and have persisted postpartum for more than 6 months.
Type of Use (Select one or both, as applicable)
> Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
# CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
Image /page/3/Picture/0 description: The image shows a logo for "Eminent Spine". The logo features a golden caduceus, which is a symbol of medicine, with a sword running through it. The caduceus and sword are surrounded by a golden, curved shape, resembling a laurel wreath. The text "EMINENT" is in a dark green, sans-serif font, and the text "SPINE" is in a lighter green, sans-serif font, with a line above and below the word.
# 510(k) Summary
| Submitted By: | Eminent Spine, LLC<br>2004 Ventura Dr. Suite #100<br>Plano, TX 75093 |
|-----------------------------|----------------------------------------------------------------------|
| Date: | 02/21/2024 |
| Contact Person: | Chhavy Tep-Cullison, Regulatory Affairs Specialist |
| Contact Telephone: | (440) 320-4338 |
| Contact Fax: | (440) 933-7839 |
| Device Trade Name: | Eminent Spine SI Screw System |
| Device Classification Name: | Smooth or threaded metallic bone fixation fastener (21 CFR 888.3040) |
| Device Classification: | Class II |
| Reviewing Panel: | Orthopedic |
| Product Code: | OUR |
| Predicate Device: | Entasis 3D Dual-Lead Sacroiliac Implant System (K223708) |
| Reference Device: | T-FIX® 3DSI Joint Fusion System (K214123) |
#### Device Description:
The Eminent Spine Sacroiliac (SI) Screw System is intended for sacroiliac joint fixation for conditions including degenerative sacroiliac joint disruption. The Eminent Spine SI Screw System consists of screws manufactured from Ti-6Al-4V ELI per ASTM F136 using traditional machining methods, as well as 3D-printed titanium alloy Ti-6AI-4V ELI per ASTM F3001. The screws are available in a variety of lengths and diameters to accommodate varying patient anatomy. The cannulated screws have open and porous graft windows for the Ø8.5, Ø10.0, Ø11.5, and Ø13.0mm options to allow bone growth through the implant. Optional locking crowns/locking fangs are included for all of the available screw diameters to aid in conforming to patient anatomy.
#### Indications for Use:
The Eminent Spine SI Screw System is intended for sacroiliac joint fusion for conditions including degenerative sacroilitis and sacroiliac joint disruptions, to augment immobilization of the sacroiliac joint in skeletally mature patients undergoing sacropelvic fixation as part of a lumbar or thoracolumbar fusion and acute, and non-traumatic fractures involving the sacroiliac joint. This includes those whose symptoms began during pregnancy or in the period and have persisted postpartum for more than 6 months.
{4}------------------------------------------------
Image /page/4/Picture/0 description: The image is a logo for Eminent Spine. The logo features a golden caduceus with a sword through it, surrounded by a golden circle with green accents. Below the image is the word "EMINENT" in green, with the word "SPINE" in a smaller font below it.
## Summary of Technological Characteristics:
The Eminent Spine SI Screw System and the predicate system both have the same intended use and fundamental scientific technology. A comparison table of the subject device and predicate device technological characteristics is provided in this submission in Section 005 Substantial Equivalence. A condensed comparison table is also presented below. There are no differences in technological characteristics that raise questions of safety and efficacy.
| Item | Subject Device:<br>Eminent Spine SI Screw System | Predicate Device:<br>Entasis 3D Dual-Lead Sacroiliac<br>Implant System (K223708) | Comparison |
|------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------|
| Classification<br>Name | Smooth or threaded metallic bone<br>fixation fastener | Smooth or threaded metallic bone<br>fixation fastener | Same |
| Regulation | 21 CFR 888.3040 | 21 CFR 888.3040 | Same |
| Product Code | OUR | OUR, HWC | Same |
| Indications for<br>Use | The Eminent Spine SI System is<br>intended for sacroiliac joint fusion for<br>conditions including degenerative<br>sacroiliitis and sacroiliac joint<br>disruptions, to augment<br>immobilization and stabilization of the<br>sacroiliac joint in skeletally mature<br>patients undergoing sacropelvic<br>fixation as part of a lumbar or<br>thoracolumbar fusion and acute, non-<br>acute, and non-traumatic fractures<br>involving the sacroiliac joint. This<br>includes those whose symptoms began<br>during pregnancy or in the peripartum<br>period and have persisted postpartum<br>for more than 6 months. | The Entasis 3D Dual-Lead Sacroiliac<br>Implant System is intended for<br>sacroiliac joint fusion for conditions<br>including degenerative sacroiliitis and<br>sacroiliac joint disruptions, to<br>augment immobilization and<br>stabilization of the sacroiliac joint in<br>skeletally mature patients undergoing<br>sacropelvic fixation as part of a<br>lumbar or thoracolumbar fusion and<br>acute, non-acute, and non-traumatic<br>fractures involving the sacroiliac joint.<br>This includes those whose symptoms<br>began during pregnancy or in the<br>peripartum period and have persisted<br>postpartum for more than 6 months. | Same |
| Screw<br>Lengths | 20 - 90 mm in 5 mm increments | Machined: 30-70mm in 5mm<br>increments<br>3D-Printed: 30-110mm in 5mm<br>increments | Substantially<br>Equivalent |
| Screw<br>Diameters | Ø8.5, Ø10.0, Ø11.5, Ø13.0 mm | Machined: Ø7.0, Ø9.5, Ø11.5 mm<br>3D-Printed: Ø9.5, Ø11.5, Ø14.5 mm | Substantially<br>Equivalent |
| Material | Ti-6Al-4V ELI per ASTM F136<br>Ti-6Al-4V per ASTM F3001 | Ti-6Al-4V ELI per ASTM F136<br>Ti-6Al-4V per ASTM F3001 | Same |
# Table 1: Dimensions and Technological Characteristics Comparison
{5}------------------------------------------------
Image /page/5/Picture/0 description: The image shows a logo for Eminent Spine. The logo features a golden sword with a caduceus wrapped around it, set within a golden oval. Below the emblem, the word "EMINENT" is written in a serif font, with the word "SPINE" underneath in a smaller font.
# Non-Clinical Testing:
Substantial equivalence is supported by the results of mechanical testing which includes static torsion, driving torque, and axial pullout testing per ASTM F3574, and static and dynamic cantilever bend testing per ASTM F3574. Results support that the strength of the subject SI Screw System is sufficient for its intended use and is substantially equivalent to legally marketed predicate devices.
## Conclusion:
Based on the indications for use, technological characteristics, and comparison with the predicate device, the subject device has demonstrated substantial equivalence.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.