← Product Code [MQV](/productcode/MQV) · K240459

# Cerament Bone Void Filler (K240459)

_Bonesupport AB · MQV · Mar 5, 2024 · Orthopedic · SESE_

**Canonical URL:** https://fda.innolitics.com/device/K240459

## Device Facts

- **Applicant:** Bonesupport AB
- **Product Code:** [MQV](/productcode/MQV.md)
- **Decision Date:** Mar 5, 2024
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 888.3045
- **Device Class:** Class 2
- **Review Panel:** Orthopedic
- **Attributes:** Therapeutic, Pediatric

## Indications for Use

CERAMENT BONE VOID FILLER is a ceramic bone graft substitute intended for orthopedic applications as a filler for gaps and voids that are not intrinsic to the stability of the bony structure. CERAMENT BONE VOID FILLER is indicated to be injected, or placed, into bony voids or gaps in the skeletal system, i.e. extremities, pelvis, posterolateral spine (only during open surgery in spine), and intervertebral disc space. These defects may be the result of benign bone cysts and tumors (in adults and pediatric patients ≥ 9 years old), or osseous defects created as a result of either surgery or traumatic injury to the bone. When used in intervertebral body fusion procedures, CERAMENT BONE VOID FILLER must be used on its own with an intervertebral body fusion device cleared by FDA for use with a bone void filler. CERAMENT BONE VOID FILLER resorbs and is replaced by bone during the healing process. CERAMENT BONE VOID FILLER can augment hardware and support bone fragments during the surgical procedure. The cured paste acts only as a temporary support media and is not intended to provide structural support during the healing process. CERAMENT BONE VOID FILLER can be drilled and screws can be placed through it.

## Device Story

Ceramic bone graft substitute; composed of hydroxyapatite, calcium sulfate hemihydrate, and calcium sulfate dihydrate. Injected or placed into bony voids/gaps by surgeons during orthopedic procedures. Acts as temporary support media; resorbs and is replaced by bone during healing. Can be drilled and used with screws. Intended for use in extremities, pelvis, spine, and intervertebral disc space. Provides no structural support during healing. Benefits patient by filling defects and supporting bone fragments during surgical procedures.

## Clinical Evidence

Bench testing only. Leverages data from K201535 for sterility, shelf-life, endotoxin, pyrogenicity, biocompatibility, and bench performance. Performance in intervertebral body space supported by analysis of bone grafting materials in prior posterolateral spine fusion studies.

## Technological Characteristics

Ceramic bone graft substitute composed of hydroxyapatite, calcium sulfate hemihydrate, and calcium sulfate dihydrate. Injectable paste form factor. Resorbable. No structural support provided. Complies with Class II Special Controls Guidance for Resorbable Calcium Salt Bone Void Filler Devices.

## Regulatory Identification

A resorbable calcium salt bone void filler device is a resorbable implant intended to fill bony voids or gaps of the extremities, spine, and pelvis that are caused by trauma or surgery and are not intrinsic to the stability of the bony structure.

## Special Controls

*Classification.* Class II (special controls). The special control for this device is the FDA guidance document entitled “Class II Special Controls Guidance: Resorbable Calcium Salt Bone Void Filler Device; Guidance for Industry and FDA.” See § 888.1(e) of this chapter for the availability of this guidance.

## Predicate Devices

- NuVasive - Attrax Putty ([K203714](/device/K203714.md))
- Prosidyan - FibergraftTM BG Putty ([K222276](/device/K222276.md))

## Reference Devices

- BONESUPPORT AB - CERAMENT® BONE VOID FILLER ([K201535](/device/K201535.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

March 5, 2024

Image /page/0/Picture/1 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.

BONESUPPORT AB % Scott Bruder, MD, PhD Founder and CEO Bruder Consulting & Venture Group 268 Glen Place Franklin Lakes, New Jersey 07417

Re: K240459

Trade/Device Name: Cerament Bone Void Filler Regulation Number: 21 CFR 888.3045 Regulation Name: Resorbable calcium salt bone void filler device Regulatory Class: Class II Product Code: MQV Dated: February 16, 2024 Received: February 16, 2024

Dear Dr. Bruder:

We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).

{1}------------------------------------------------

Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review. the OS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

# Sara S. Thompson -S

For

Jesse Muir, Ph.D. Assistant Director DHT6C: Division of Restorative, Repair and Trauma Devices OHT6: Office of Orthopedic Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health 

{2}------------------------------------------------

Enclosure

{3}------------------------------------------------

# Indications for Use

510(k) Number (if known) K240459

Device Name CERAMENT BONE VOID FILLER

#### Indications for Use (Describe)

CERAMENT BONE VOID FILLER is a ceramic bone graft substitute intended for orthopedic applications as a filler for gaps and voids that are not intrinsic to the stability of the bony structure.

CERAMENT BONE VOID FILLER is indicated to be injected, or placed, into bony voids or gaps in the skeletal system, i.e. extremities, pelvis, posterolateral spine (only during open surgery in spine), and intervertebral disc space. These defects may be the result of benign bone cysts and tumors (in adults and pediatric patients ≥ 9 years old), or osseous defects created as a result of either surgery or traumatic injury to the bone. When used in intervertebral body fusion procedures, CERAMENT BONE VOID FILLER must be used on its own with an intervertebral body fusion device cleared by FDA for use with a bone void filler.

CERAMENT BONE VOID FILLER resorbs and is replaced by bone during the healing process.

CERAMENT BONE VOID FILLER can augment hardware and support bone fragments during the surgical procedure. The cured paste acts only as a temporary support media and is not intended to provide structural support during the healing process. CERAMENT BONE VOID FILLER can be drilled and screws can be placed through it.

| Type of Use (Select one or both, as applicable) |
|-------------------------------------------------|
|-------------------------------------------------|

| <span style="font-family: Arial;">☑</span> Prescription Use (Part 21 CFR 801 Subpart D) |
|-----------------------------------------------------------------------------------------|
| <span style="font-family: Arial;">☐</span> Over-The-Counter Use (21 CFR 801 Subpart C)  |

#### CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{4}------------------------------------------------

Image /page/4/Picture/0 description: The image shows the logo for BONESUPPORT. The logo consists of a blue geometric shape on the left and the word "BONESUPPORT" in black letters on the right. The geometric shape is made up of four blue triangles that are arranged to form a larger diamond shape. The word "BONESUPPORT" is written in a bold, sans-serif font, and the letters are all capitalized. There is a trademark symbol after the word "SUPPORT".

# 510(k) Summary

#### Submitter

Blerta Shuka Senior Regulatory Affairs Specialist BONESUPPORT AB Scheelevägen 19 SE-223 70 Lund Sweden

#### Correspondent

Scott Bruder, MD, PhD Founder and CEO Bruder Consulting & Venture Group Scott(a)BruderConsulting.com 201.874.9701

#### Date Prepared

February 16, 2024

#### Device

| Trade Name     | CERAMENT BONE VOID FILLER                                       |
|----------------|-----------------------------------------------------------------|
| Common Name    | Bone void filler                                                |
| Regulation     | 21 CFR 888.3045 Resorbable calcium salt bone void filler device |
| Classification | Class II                                                        |
| Product Code   | MQV                                                             |
| Panel          | Orthopedic                                                      |

#### Predicates

| Primary Predicate             | NuVasive - Attrax Putty (K203714)           |
|-------------------------------|---------------------------------------------|
| Secondary Predicate           | Prosidyan - FibergraftTM BG Putty (K222276) |
| Reference Device<br>(K201535) | BONESUPPORT AB - CERAMENT® BONE VOID FILLER |

#### Device Description

CERAMENT BONE VOID FILLER is composed of hydroxyapatite, calcium sulfate hemihydrate, and calcium sulfate dihydrate, and it identical to the device cleared in K201535. This submission expands the device's indication to include use in the intervertebral space.

#### Indications for Use Statement

CERAMENT BONE VOID FILLER is a ceramic bone graft substitute intended for orthopedic applications as a filler for gaps and voids that are not intrinsic to the stability of the bony structure.

CERAMENT BONE VOID FILLER is indicated to be injected, or placed, into bony voids or gaps in the skeletal system, i.e. extremities, pelvis, posterolateral spine (only during open surgery

{5}------------------------------------------------

Image /page/5/Picture/0 description: The image shows the logo for BONESUPPORT. The logo consists of a blue geometric shape resembling an abstract letter "S" on the left, followed by the word "BONESUPPORT" in black, sans-serif font. A small trademark symbol is placed to the upper right of the word "BONESUPPORT".

in spine), and intervertebral disc space. These defects may be the result of benign bone cysts and tumors (in adults and pediatric patients ≥ 9 years old), or osseous defects created as a result of either surgery or traumatic injury to the bone. When used in intervertebral body fusion procedures, CERAMENT BONE VOID FILLER must be used on its own with an intervertebral body fusion device cleared by FDA for use with a bone void filler.

CERAMENT BONE VOID FILLER resorbs and is replaced by bone during the healing process.

CERAMENT BONE VOID FILLER can augment hardware and support bone fragments during the surgical procedure. The cured paste acts only as a temporary support media and is not intended to provide structural support during the healing process. CERAMENT BONE VOID FILLER can be drilled and screws can be placed through it.

# Substantial Equivalence

The intended use is the same for subject and predicate devices. The subject device and predicates are all cleared for use as bone void fillers in the intervertebral body space. There are differences in the device's material composition, but these do not raise different questions of safety and effectiveness. The effect of differences in material composition is addressed in the prior functional animal model performance.

# Performance

The subject device has been previously cleared under K201535, which serves as a Reference Device. This submission is leveraged to support the device's sterility, shelf-life, endotoxin, pyrogenicity, biocompatibility, and characterizations/bench performance as recommended in FDA's Class II Special Controls Guidance Document for Resorbable Calcium Salt Bone Void Filler Devices. The device's performance in the intervertebral body space was supported by a robust analysis of bone grafting materials in the prior posterolateral spine fusion studies.

### Summarv

The subject device and predicates have the same intended use, and the same specific indications for use in the intervertebral body space. Any differences in technological characteristics between the subject device and predicate do not raise different questions of safety and effectiveness. Based on the clinical analysis, the device is substantially equivalent to the predicates.

---

**Source:** [https://fda.innolitics.com/device/K240459](https://fda.innolitics.com/device/K240459)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
