RELIEEV Suction Curette (Flexible 3.0/ Standard 3.6)
K240434 · Li Medical Corporation , Ltd. · HHK · Sep 25, 2024 · Obstetrics/Gynecology
Device Facts
Record ID
K240434
Device Name
RELIEEV Suction Curette (Flexible 3.0/ Standard 3.6)
Applicant
Li Medical Corporation , Ltd.
Product Code
HHK · Obstetrics/Gynecology
Decision Date
Sep 25, 2024
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 884.1175
Device Class
Class 2
Indications for Use
The "RELIEEV" Suction Curette (Flexible 3.0/ Standard 3.6) is used to remove samples of materials from the uterus and from the mucosal lining of the uterus by scraping or vacuum suction. These devices obtain tissue samples for purposes of biopsy precancer screening or they can be used for menstrual extraction.
Device Story
The RELIEEV Suction Curette is a sterile, single-use, manual vacuum device used by clinicians to obtain endometrial tissue samples. It consists of a polypropylene tube and piston with a silicone O-ring. The device is inserted into the uterus; the clinician manually operates the piston to create vacuum suction or uses the device to scrape the mucosal lining. The device captures tissue samples within the tube for subsequent pathological analysis. It is intended for use in clinical settings for biopsy or menstrual extraction. The device provides a simple, manual method for tissue collection, aiding in the diagnosis of uterine conditions. It is a disposable, non-powered instrument.
Clinical Evidence
Bench testing only. No clinical data provided. Testing included dimensional analysis, suction force, marker readability, bending force, biocompatibility (ISO 10993-5, -10, -23), packaging integrity (ASTM F88, F1929, F1886), shipping validation (ASTM D4332, D4169), and sterilization validation (ISO 11135). Shelf life of 3 years supported by accelerated aging (ASTM F1980).
Technological Characteristics
Materials: Polypropylene tube/piston, silicone O-ring. Dimensions: 3.0mm and 3.6mm outer diameters. Principle: Manual vacuum suction/scraping. Connectivity: None. Sterilization: Ethylene Oxide (EtO) per ISO 11135. Packaging: Medical paper grade pouch. Shelf life: 3 years.
Indications for Use
Indicated for patients requiring uterine mucosal tissue sampling for biopsy precancer screening or menstrual extraction.
Regulatory Classification
Identification
An endometrial suction curette is a device used to remove material from the uterus and from the mucosal lining of the uterus by scraping and vacuum suction. This device is used to obtain tissue for biopsy or for menstrual extraction. This generic type of device may include catheters, syringes, and tissue filters or traps.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA logo on the right. The FDA logo is a blue square with the letters "FDA" in white, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue.
September 25, 2024
LI Medical Corporation Ltd. Jago Chen CEO 2F., No. 43, Zhongxing Rd. New Taipei City, Xizhi Dist. 221012 TAIWAN
Re: K240434
> Trade/Device Name: "RELIEEV" Suction Curette (Flexible 3.0/ Standard 3.6) Regulation Number: 21 CFR 884.1175 Regulation Name: Endometrial Suction Curette and accessories Regulatory Class: II Product Code: HHK Received: August 30, 2024
Dear Jago Chen:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
{1}------------------------------------------------
Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review. the OS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-device-advicecomprehensive-regulatory-assistance/unique-device-identification-system-udi-system.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
{2}------------------------------------------------
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatory-
assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
# Reginald K. Avery -S
for Jason R. Roberts, Ph.D. Assistant Director DHT3B: Division of Reproductive, Gynecology, and Urology Devices OHT3: Office of Gastrorenal, ObGyn, General Hospital, and Urology Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{3}------------------------------------------------
### DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration
### Indications for Use
510(k) Number (if known) K240434
Device Name
"RELIEEV" Suction Curette (Flexible 3.0/ Standard 3.6)
Indications for Use (Describe)
The "RELIEEV" Suction Curette (Flexible 3.0/ Standard 3.6) is used to remove samples of materials from the uterus and from the mucosal lining of the uterus by scraping or vacuum suction. These devices obtain tissue samples for purposes of biopsy precancer screening or they can be used for menstrual extraction.
| Type of Use (Select one or both, as applicable) | |
|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------|
| <span style="font-size: 12px;"> Prescription Use (Part 21 CFR 801 Subpart D) </span> | <span style="font-size: 12px;"> Over-The-Counter Use (21 CFR 801 Subpart C) </span> |
| CONTINUE ON A SEPARATE PAGE IF NEEDED. | |
| | This section applies only to requirements of the Paperwork Reduction Act of 1995. |
| | <b>*DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*</b> |
| The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: | |
| Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff | PRAStaff@fda.hhs.gov |
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
EF
PSC Publishing Services (301) 443-6740
==============================================================================================================================================================================
{4}------------------------------------------------
# 510(k) Summary
510(K) Number: K240434
# I. SUBMITTER
LI Medical Corporation LTD. 2F., No. 43, Zhongxing Rd., Xizhi Dist., New Taipei City 221012, Taiwan Phone: +886-2-86461999 Fax: +886-2-26471999 Contact Person: Jago Chen, CEO Email: jago.chen@li-med.com.tw Date Prepared: Sep. 24, 2024
### II. DEVICE
| Device Name : | "RELIEEV" Suction Curette (Flexible 3.0/ Standard 3.6) |
|--------------------------|--------------------------------------------------------|
| Trade/Proprietary Name : | RELIEEV |
| Common Name : | Endometrial suction curette |
| Regulation Number : | 21 CFR 884.1175 |
| Regulation Name : | Endometrial suction curette and accessories |
| Regulatory Class : | Class II |
| Product Code : | HHK |
{5}------------------------------------------------
### PREDICATE DEVICES REFERENCED IN THE SUBMISSION lll.
| Predicate Device : | The PANPAC SUCTION CURETTE |
|--------------------|----------------------------|
| Company Name : | PANPAC |
| 510(k) number: | K092982 |
The predicate device has not been subject to a design-related recall.
### DEVICE DESCRIPTION IV.
The "RELIEEV" Suction Curette is a sterile, disposable, single-use device used to obtain a sample from the uterine mucosal lining. It is available in two models – Flexible 3.0 and Standard 3.6 with outer diameters of 3.0 mm and 3.6 mm respectively. The device consists of a tube and piston made of polypropylene (PP) with a silicone O-ring. The subject device is packaged in medical paper grade pouch (60g).
# V. INDICATIONS FOR USE
The "RELIEEV" Suction Curette (Flexible 3.0/ Standard 3.6) is used to remove samples of materials from the uterus and from the mucosal lining of the uterus by scraping or vacuum suction. These devices obtain tissue samples for purposes of biopsy precancer screening or they can be used for menstrual extraction.
{6}------------------------------------------------
# VII. COMPARISON TO PREDICATE DEVICES
| | Subject Device (K240434) | Predicate Device (K092982) | |
|--------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------|
| Manufacture | LI Medical Corporation LTD. | PANPAC Medical Corporation | COMPARISON |
| Device name | "RELIEEV" Suction Curette (Flexible 3.0/<br>Standard 3.6) | The PANPAC Suction Curette | |
| Model | Model 1: ESC30A1 (Flexible 3.0)<br>Model 2: ESC36A1 (Standard 3.6) | The Preferred Curette<br>(Standard 3.0- Model M0014) | |
| Common name | Endometrial Suction Curette | Endometrial Suction Curette | Same |
| Regulation number | 21 CFR 884.1175 | 21 CFR 884.1175 | Same |
| Product code | HHK | HHK | Same |
| Classification | Class II | Class II | Same |
| Prescription medical<br>device | Yes | Yes | Same |
| Single patient use | Yes | Yes | Same |
| Sterile device | Yes | Yes | Same |
| Disposable device | Yes | Yes | Same |
| Sterilization method | Ethylene Oxide (EtO) | EtO | Same |
| Hole on the tube | YES<br>(1 hole on the side of the tube) | YES<br>(1 hole on the side of the tube) | Same |
| Indications for Use | The "RELIEEV" Suction Curette (Flexible<br>3.0/ Standard 3.6) is used to remove<br>samples of materials from the uterus and<br>from the mucosal lining of the uterus by | An endometrial suction curette,<br>such as the curette and preferred<br>curette (available in 2.5 and 3.0<br>mm outer diameters) are devices | Same |
| | | | |
| | scraping or vacuum suction. These devices<br>obtain tissue samples for purposes of<br>biopsy precancer screening or they can be<br>used for menstrual extraction. | used to remove samples of<br>materials from the uterus and<br>from the mucosal lining of the<br>uterus by scraping or vacuum<br>suction. These devices obtain<br>tissue samples for purposes of<br>biopsy precancer screening or<br>they can be used for menstrual<br>extraction. | |
| Length | Flexible 3.0<br>PP tube: $240.0 \pm 3.0$ mm<br>Piston: $264.0 \pm 3.0$ mm<br>Standard 3.6<br>PP tube: $240.0 \pm 3.0$ mm<br>Piston: $264.0 \pm 3.0$ mm | M0014<br>(Model M0014) Curette Length-<br>240 mm | Same |
| Dimensions | Flexible 3.0<br>PP tube O.D: $3.0 + 0.10/-0.03$ mm<br>Piston cover with O-ring O.D.: $2.7 \pm 0.05$ mm<br>Standard 3.6<br>PP tube O.D: $3.6 +0.10/-0.03$ mm<br>Piston cover with O-ring O.D.: $2.7 \pm 0.05$ mm | M0014<br>PP tube O.D: 3.0 mm<br>Piston O.D: 2.2 mm | Different |
| Material | Flexible 3.0 and Standard 3.6<br>Tube: Polypropylene | M0014<br>Tube: Polypropylene | Different |
| | Piston: Polypropylene<br>O-ring: Silicone (Purple pigment) | Piston: Polypropylene<br>O-ring: Silicone (blue color) | |
| Scale marked on the<br>tube | Laser engraving | Ink printed | Different |
| Packaging | Medical paper grade pouch (60g) | Tyvek pouch | Different |
| Shelf life | 3 years | 3 years | Same |
{7}------------------------------------------------
{8}------------------------------------------------
K240434
### LI Medical Corporation LTD. 510(k): K240434
The subject and predicate device differ in the dimensions, materials, scale marking and packaging. These technological differences do not raise different questions of safety and effectiveness and are addressed with performance testing.
{9}------------------------------------------------
#### PERFORMANCE DATA VIII.
The following performance data were provided in support of the substantial equivalence determination.
# Bench Testing
Physical bench testing was performed to confirm that the device performs according to the product specifications. Device evaluation consisted of physical and functional testing including:
- Dimensional Analysis,
- Suction Force,
- Marker readability after bending, and
- Bending Force
The subject device test results were compared with the predicate device and passed all functional bench testing.
# Biocompatibility Testing
The biocompatibility testing was conducted in accordance with the 2023 FDA guidance document, Use of International Standard ISO10993-1, "Biological evaluation of medical devices - Part 1: Evaluation and testing within a risk management process". The "RELIEEV" Suction Curette is considered mucosal tissue contacting for less than 24 hours and the testing included the following tests with passing results.
| Test name | Results | Standards |
|---------------|----------------|-------------------|
| Cytotoxicity | Non-cytotoxic | ISO 10993-5:2009 |
| Sensitization | Non-sensitizer | ISO 10993-10:2021 |
| Irritation | Non-irritant | ISO 10993-23:2021 |
# Packaging and Shipping Validation
Packaging and shipping validation were conducted on the subject device in order to confirm that the packaging materials are capable of withstanding the distribution environment during the rigors of shipping and handling.
For the simulated shipping distribution testing the environmental conditions were selected in accordance with ASTM D4332-22 and then the simulated shipping distribution was conducted per ASTM D4169-22, DC 13.
{10}------------------------------------------------
Packaging integrity testing consisting of seal peel test per ASTM F88/F88M-23, dye penetration per ASTM F1929-15, and visual inspection per ASTM F1886/F1886M-16 were completed with passing results.
# Sterilization validation
The subject device is sterilized via ethylene oxide. Sterilization validation was conducted as per ISO 11135:2014 Sterilization of health-care products – Ethylene oxide requirement for the development, validation, and routine control of a sterilization process for medical devices. Sterility testing and documentation were provided as outlined in the 2024 FDA guidance document, Submission and Review of Sterility Information in Premarket Notification (510(k)) Submissions for Devices Labeled as Sterile.
# Shelf Life
Shelf life testing was performed to support a shelf life of 3 years. Accelerated aging was conducted as per ASTM F1980-21 and package integrity testing for seal integrity and seal strength was completed using the methods under ASTM F1929-15. Bench testing, as described above, were completed on aged devices to support device functionality after aging. The results of these tests support the 3-year shelf-life claim.
#### CONCLUSIONS IX.
Subject and predicate devices share the same intended use and fundamental scientific technology, including principles of operation and mechanism of action. Differences in design between the subject and predicate devices do not raise different questions of safety and effectiveness. The results of performance testing demonstrate that the subject device is as safe and effective as the predicate devices to support a substantial equivalence determination.
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.