Picosecond Nd:YAG Laser Systems (BL-C10)
K240386 · Shanghai Bele Medical Technology Co.,Ltd · GEX · May 16, 2024 · General, Plastic Surgery
Device Facts
| Record ID | K240386 |
| Device Name | Picosecond Nd:YAG Laser Systems (BL-C10) |
| Applicant | Shanghai Bele Medical Technology Co.,Ltd |
| Product Code | GEX · General, Plastic Surgery |
| Decision Date | May 16, 2024 |
| Decision | SESE |
| Submission Type | Traditional |
| Regulation | 21 CFR 878.4810 |
| Device Class | Class 2 |
| Attributes | Therapeutic |
Intended Use
The Picosecond Nd: YAG Laser System is intended for use in surgical and aesthetic application in the medical dermatology and general and plastic surgery as follows: 1064nm wavelength: - Removal of tattoos on all skin type l-VI) with the following tattoo colors: black, brown, green, blue and purple. - Treatment of benign pigmented lesions on Fitzpatrick skin types I-IV. 532nm wavelength: - Removal of tattoos on Fitzpatrick skin types I-III with the following tattoo colors: red, yellow and orange. - Treatment of benign pigmented lesions on Fitzpatrick skin types I-IV.
Device Story
BL-C10 Picosecond Nd:YAG Laser System; AC-powered laser device. Inputs: electrical energy; transforms via Nd:YAG crystal to generate ultra-short picosecond pulses (1064nm/532nm). Outputs: laser energy targeting pigments in tattoos or benign lesions. Used in medical dermatology, general, and plastic surgery clinics; operated by physicians. High peak power pulses interact with targets; short duration minimizes thermal damage to surrounding tissue. Healthcare providers use output to clear tattoos or treat lesions; clinical decision-making based on visual assessment of skin condition. Benefits: precise pigment removal with reduced thermal side effects.
Clinical Evidence
No clinical data. Substantial equivalence supported by bench testing, including electrical safety (IEC 60601-1), electromagnetic compatibility (IEC 60601-1-2, IEC TR 60601-4-2), and laser safety (IEC 60601-2-22, IEC 60825-1).
Technological Characteristics
Nd:YAG and frequency-doubled Nd:YAG laser; 1064nm and 532nm wavelengths; pulse duration 300-500ps. Mains AC-powered (110-240V). Dimensions: 1070mm x 600mm x 1250mm; weight: 120kg. Single handpiece for both wavelengths. Standards: IEC 60601-1, IEC 60601-1-2, IEC 60601-2-22, IEC 60825-1.
Indications for Use
Indicated for tattoo removal (1064nm: Fitzpatrick I-VI, black/brown/green/blue/purple; 532nm: Fitzpatrick I-III, red/yellow/orange) and treatment of benign pigmented lesions (1064nm/532nm: Fitzpatrick I-IV).
Regulatory Classification
Identification
(1) A carbon dioxide laser for use in general surgery and in dermatology is a laser device intended to cut, destroy, or remove tissue by light energy emitted by carbon dioxide.(2) An argon laser for use in dermatology is a laser device intended to destroy or coagulate tissue by light energy emitted by argon.
Predicate Devices
- PICOSECOND Nd: YAG Laser System (K233018)
- PicoWay Laser System (K191685)
Related Devices
- K240118 — Picosecond Nd:YAG Laser System · Smedtrum Medical Technology Co., Ltd. · Apr 30, 2024
- K241144 — Picosecond Nd:YAG Laser (PICOCAREMAJESTY) · Wontech Co., Ltd. · Jun 24, 2024
- K233018 — PICOSECOND Nd:YAG Laser System · Ids, Ltd. · Dec 21, 2023
- K200116 — PicoSecond Nd: YAG Laser System · Shanghai Apolo Medical Technology Co., Ltd. · May 28, 2020
- K193609 — Q Switched Nd: YAG Laser machine · Beijing Lead Beauty S & T Co., Ltd. · Mar 20, 2020
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
Shanghai Bele Medical Technology Co.,Ltd Jeremy Li General Manager Room 402, 4F, Building 1, NO.3255 Shengang Road, Songjiang District Shanghai, Shanghai 201600 China
Re: K240386
Trade/Device Name: Picosecond Nd: Y AG Laser Systems (BL-C10) Regulation Number: 21 CFR 878.4810 Regulation Name: Laser Surgical Instrument For Use In General And Plastic Surgery And In Dermatology Regulatory Class: Class II Product Code: GEX Dated: February 5, 2024 Received: February 8, 2024
Dear Jeremy Li:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device"
{1}------------------------------------------------
(https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30. Design controls; 21 CFR 820.90. Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review. the OS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely.
Yan Fu -S Digitally signed by Yan Fu -S
Date: 2024.05.16 21:17:35
-04'00'
for
Tanisha Hithe Assistant Director DHT4A: Division of General Surgery Devices OHT4: Office of Surgical and Infection Control Devices
{2}------------------------------------------------
Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{3}------------------------------------------------
#### Indications for Use
Form Approved: OMB No. 0910-0120 Expiration Date: 07/31/2026 See PRA Statement below.
Submission Number (if known)
K240386
Device Name
Picosecond Nd:YAG Laser Systems (BL-C10)
Indications for Use (Describe)
The Picosecond Nd: YAG Laser System is intended for use in surgical and aesthetic application in the medical dermatology and general and plastic surgery as follows:
1064nm wavelength:
- Removal of tattoos on all skin type (Fitzpatrick skin types I-VI) with the following tattoo colors: black, brown, green, blue and purple.
- Treatment of benign pigmented lesions on Fitzpatrick skin types I-IV.
532nm wavelength:
- Removal of tattoos on Fitzpatrick skin tvpes I-III with the following tattoo colors: red, yellow and orange.
- Treatment of benign pigmented lesions on Fitzpatrick skin types I-IV.
Type of Use (Select one or both, as applicable)
Prescription Use (Part 21 CFR 801 Subpart D)
er-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{4}------------------------------------------------
| 510(k) #: K240386 | | 510(k) Summary | Prepared on: 2024-05-16 | |
|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------|-------------------------------------------------------------------------------------------------------|-------------------------|--|
| Contact Details | | | 21 CFR 807.92(a)(1) | |
| Applicant Name | | SHANGHAI BELE MEDICAL TECHNOLOGY CO.,LTD | | |
| Applicant Address | | Room 402, 4F, Building 1, NO.3255 Shengang Road, Songjiang District<br>Shanghai Shanghai 201600 China | | |
| Applicant Contact Telephone | | +86-21 51012882 | | |
| Applicant Contact | | Jeremy Li | | |
| Applicant Contact Email | | jeremy@belemed.com | | |
| Device Name | | | 21 CFR 807.92(a)(2) | |
| Device Trade Name | | Picosecond Nd:YAG Laser Systems (BL-C10) | | |
| Common Name | | Laser surgical instrument for use in general and plastic surgery and in<br>dermatology | | |
| Classification Name | | Powered Laser Surgical Instrument | | |
| Regulation Number | | 878.4810 | | |
| Product Code(s) | | GEX | | |
| Legally Marketed Predicate Devices | | | 21 CFR 807.92(a)(3) | |
| Predicate # | Predicate Trade Name (Primary Predicate is listed first) | | Product Code | |
| K233018 | PICOSECOND Nd: YAG Laser System | | GEX | |
| K191685 | PicoWay Laser System | | GEX | |
| Device Description Summary | | | 21 CFR 807.92(a)(4) | |
| The BL-C10 Picosecond ND:YAG Laser Systems is designed with a mains electricity (AC-powered) light amplification by stimulated<br>emission of radiation (LASER) device assembly based on neodymium-doped yttrium-aluminium-garnet (Nd:YAG) and frequency-<br>doubled Nd:YAG laser technology. The laser system generates ultra-short pulses in the picosecond range (1 picosecond = 10^-12 | | | | |
seconds). The pulses are then amplified within the Nd:YAG crystal. The system operates at two wavelengths: 1064nm and 532mm. These wavelengths are chosen to target different colors of pigments in tattoos and pigmented lesions.
- 1064nm Wavelength: Used for treating tattoos with colors like black, brown, green, blue, and purple, as well as for treating benign pigmented lesions.
- 532nm Wavelength: Used for removing tattoos with colors such as red, yellow, and for treating benign pigmented lesions. The ultra-short pulses have very high peak power, allowing them to interact with target (tattoos or pigmented lesions) in a precise and controlled manner. The short duration of the pulses helps minimize heat buildup in surrounding tissues.
## Intended Use/Indications for Use
## 21 CFR 807.92(a)(5)
The Picosecond Nd: YAG Laser System is intended for use in surgical and aesthetic application in the medical dermatology and general and plastic surgery as follows:
1064nm wavelength:
- Removal of tattoos on all skin type l-VI) with the following tattoo colors: black, brown, green, blue and purple. - Treatment of benign pigmented lesions on Fitzpatrick skin types I-IV.
{5}------------------------------------------------
532nm wavelength:
- Removal of tattoos on Fitzpatrick skin types I-III with the following tattoo colors: red, yellow and orange.
- Treatment of benign pigmented lesions on Fitzpatrick skin types I-IV.
## Indications for Use Comparison
The indication of the proposed device is covered by the predicate.
## Technological Comparison
The technical characteristics of the subject device are the predicate (including wavelengths, maximum pulse energy, repetition rate, spot size, beam delivery), except for the following parameters:
- Difference 1: The pulse duration of the subject device is 300-500p; the predicate device is 300ps - both modes. The pulse duration differs slightly, however, the pulsed device is within the range of the reference, K191685, which is 240-500ps.
- Difference 2: The subject device has a one handpiece (532nm and 1064 nm); the primary predicate has a one handpiece for 532nm and 1064 nm : the secondary predicate has three of hand pieces Resolve (532 nm high energy, 1064 nm) indicated for wrinkle treatment, Resolve Fusion (532nm and 1064nm) indicated for treatment of benign pigmented lesion and 1064nm) for tattoo removal and benign pigmented lesion treatment. The subject device han the secondary predicate, however, the handpiece in the subject device is the same as two handpieces of the predicates.
- Difference 3&4&5: system dimension, system weight, and electrical power supply. The subject device's dimension (1070mm x 600mm x 1250mm), weight (120kg) and power supply (110-240V AC, 10A, 50/60 Hz) are slightly different from the reference device K191685, dimension (1070 mm x 46mm x 690 mm), weight 12kg, power supply (220-240 V AC, 30A, 50/600 Hz) (the prameters of the predicate are unknown). However, the difference does not affect the device function. The subject has passed the IEC 60601-1-2 tests, and the test results show that the device can function as intended.
#### Non-Clinical and/or Clinical Tests Summary & Conclusions 21 CFR 807.92(b)
Non-Clinical Testing:
A battery of tests was performed to verify that the proposed device met all design specification. The test result demonstrated that the proposed device complies with the following standards:
Electrical safety and electromagnetic compatibility
IEC 60601-1: 2005+cor.1:2006+Cor.2.2007+A1:2012 Medical electrical equipment - Part 1: General requirements for basic safety and essential performance
IEC 60601-1-2:2014 Medical electrical equipments for basic safety and essential performance-Collateral Standard: Electromagnetic disturbances - Requirements and tests
IEC TR 60601-4-2: 2016 Medical electrical equipment - Part 4-2: Guidance and interpretation - Electromagnetic immunity: performance of medical electrical equipment and medical electrical systems
IEC 60601-2-22:2007+A1: 2012 Medical electrical equipments for the safety and essettial performance of surgical, cosmetic, therapeutic and diagnostic laser equipment
IEC 60825-1:2014 Safety of Laser products-Part 1: Equipment classification and requirements
Clinical Testing: lt is not applicable.
Conclusions:
Base on the performance testing and validation studiect device is substantially equivalent to the predicate device.
# 21 CFR 807.92(a)(5)
### 21 CFR 807.92(a)(6)