← Product Code [NKB](/productcode/NKB) · K240350

# AnyPlus II Spinal Fixation System (K240350)

_GS Medical Co., Ltd. · NKB · Apr 3, 2024 · Orthopedic · SESE_

**Canonical URL:** https://fda.innolitics.com/device/K240350

## Device Facts

- **Applicant:** GS Medical Co., Ltd.
- **Product Code:** [NKB](/productcode/NKB.md)
- **Decision Date:** Apr 3, 2024
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 888.3070
- **Device Class:** Class 2
- **Review Panel:** Orthopedic
- **Attributes:** Therapeutic

## Indications for Use

The GS Medical AnyPlus II Spinal Fixation System is composed of a non-cervical spinal fixation devices. It is intended for use as posterior pedicle screw fixation system (T1-S2), a posterior hook fixation system (T1-L5), or as an anterolateral fixation system (T8-L5). All components in the system are limited to skeletally mature patients. System components are to be used for immobilization and stabilization of the spine as an adjunct to fusion. These devices are indicated for all following indications regardless of intended use: degenerative disc disease (defined as discogenic back pain with degeneration of the disc confirmed by history and radiographic studies), spondylolisthesis, trauma (i.e., fracture or dislocation), deformities or curvatures (i.e., scoliosis, and/or lordosis, Scheuermann's Disease), tumor, stenosis, pseudoarthrosis, and failed previous fusion.

## Device Story

Non-cervical spinal fixation system; includes mono/polyaxial screws, reduction screws, cannulated screws, long arm screws, cortical screws, and set screws. Implants manufactured from Ti6Al4V ELI (ASTM F136); instruments from precipitation hardened stainless steel (ASTM F899). Used by surgeons in clinical settings for spinal immobilization and stabilization as an adjunct to fusion. System components provide mechanical support to the spine; intended to facilitate bone fusion in patients with degenerative, traumatic, or deformity-related spinal conditions. Implants are non-sterile, single-use; instruments are reusable. Performance verified via static/dynamic axial compression bending and static torsion testing per ASTM F1717.

## Clinical Evidence

Bench testing only. Biomechanical performance evaluated via static and dynamic axial compression bending and static torsion testing per ASTM F1717. No clinical data provided.

## Technological Characteristics

Materials: Ti6Al4V ELI (ASTM F136) for implants; precipitation hardened stainless steel (ASTM F899) for instruments. Components: mono/polyaxial, reduction, cannulated, long arm, and cortical screws. Diameters 4.5-10.5mm; lengths 20-100mm. Non-sterile, single-use implants; reusable instruments. Mechanical fixation system.

## Regulatory Identification

(1) Rigid pedicle screw systems are comprised of multiple components, made from a variety of materials that allow the surgeon to build an implant system to fit the patient's anatomical and physiological requirements. Such a spinal implant assembly consists of a combination of screws, longitudinal members (e.g., plates, rods including dual diameter rods, plate/rod combinations), transverse or cross connectors, and interconnection mechanisms (e.g., rod-to-rod connectors, offset connectors).(2) Semi-rigid systems are defined as systems that contain one or more of the following features (including but not limited to): Non-uniform longitudinal elements, or features that allow more motion or flexibility compared to rigid systems.

## Special Controls

*Classification.* (1) Class II (special controls), when intended to provide immobilization and stabilization of spinal segments in skeletally mature patients as an adjunct to fusion in the treatment of the following acute and chronic instabilities or deformities of the thoracic, lumbar, and sacral spine: severe spondylolisthesis (grades 3 and 4) of the L5-S1 vertebra; degenerative spondylolisthesis with objective evidence of neurologic impairment; fracture; dislocation; scoliosis; kyphosis; spinal tumor; and failed previous fusion (pseudarthrosis). These pedicle screw spinal systems must comply with the following special controls:(i) Compliance with material standards;
(ii) Compliance with mechanical testing standards;
(iii) Compliance with biocompatibility standards; and
(iv) Labeling that contains these two statements in addition to other appropriate labeling information:
 “Warning: The safety and effectiveness of pedicle screw spinal systems have been established only for spinal conditions with significant mechanical instability or deformity requiring fusion with instrumentation. These conditions are significant mechanical instability or deformity of the thoracic, lumbar, and sacral spine secondary to severe spondylolisthesis (grades 3 and 4) of the L5-S1 vertebra, degenerative spondylolisthesis with objective evidence of neurologic impairment, fracture, dislocation, scoliosis, kyphosis, spinal tumor, and failed previous fusion (pseudarthrosis). The safety and effectiveness of these devices for any other conditions are unknown.”
“Precaution: The implantation of pedicle screw spinal systems should be performed only by experienced spinal surgeons with specific training in the use of this pedicle screw spinal system because this is a technically demanding procedure presenting a risk of serious injury to the patient.”
(2) Class II (special controls), when a rigid pedicle screw system is intended to provide immobilization and stabilization of spinal segments in the thoracic, lumbar, and sacral spine as an adjunct to fusion in the treatment of degenerative disc disease and spondylolisthesis other than either severe spondylolisthesis (grades 3 and 4) at L5-S1 or degenerative spondylolisthesis with objective evidence of neurologic impairment. These pedicle screw systems must comply with the following special controls:
(i) The design characteristics of the device, including engineering schematics, must ensure that the geometry and material composition are consistent with the intended use.
(ii) Non-clinical performance testing must demonstrate the mechanical function and durability of the implant.
(iii) Device components must be demonstrated to be biocompatible.
(iv) Validation testing must demonstrate the cleanliness and sterility of, or the ability to clean and sterilize, the device components and device-specific instruments.
(v) Labeling must include the following:
(A) A clear description of the technological features of the device including identification of device materials and the principles of device operation;
(B) Intended use and indications for use, including levels of fixation;
(C) Identification of magnetic resonance (MR) compatibility status;
(D) Cleaning and sterilization instructions for devices and instruments that are provided non-sterile to the end user; and
(E) Detailed instructions of each surgical step, including device removal.
(3) Class II (special controls), when a semi-rigid system is intended to provide immobilization and stabilization of spinal segments in the thoracic, lumbar, and sacral spine as an adjunct to fusion for any indication. In addition to complying with the special controls in paragraphs (b)(2)(i) through (v) of this section, these pedicle screw systems must comply with the following special controls:
(i) Demonstration that clinical performance characteristics of the device support the intended use of the product, including assessment of fusion compared to a clinically acceptable fusion rate.
(ii) Semi-rigid systems marketed prior to the effective date of this reclassification must submit an amendment to their previously cleared premarket notification (510(k)) demonstrating compliance with the special controls in paragraphs (b)(2)(i) through (v) and paragraph (b)(3)(i) of this section.

## Predicate Devices

- Anyplus Spinal Fixation System ([K172546](/device/K172546.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

April 3, 2024

Image /page/0/Picture/1 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services USA seal. To the right of the seal is the FDA logo, with the letters "FDA" in a blue box, followed by the words "U.S. FOOD & DRUG" in blue, and the word "ADMINISTRATION" in a smaller font below.

GS Medical Co. Ltd. % Barry Sands President and Founder RQMIS, Inc 110 Haverhill Rd Suite 524 Amesbury, Massachusetts 01913

Re: K240350

Trade/Device Name: AnyPlus II Spinal Fixation System Regulation Number: 21 CFR 888.3070 Regulation Name: Thoracolumbosacral Pedicle Screw System Regulatory Class: Class II Product Code: NKB, KWP, KWQ Dated: February 1, 2024 Received: February 5, 2024

Dear Barry Sands:

We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).

{1}------------------------------------------------

2

Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

Eileen
Digitally signed
by Eileen Cadel
S
Cadel -S Date: 2024.04.03
10:12:50-04'00'
for

Colin O'Neill, MBE Assistant Director DHT6B: Division of Spinal Devices OHT6: Office of Orthopedic Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

{2}------------------------------------------------

# Indications for Use

510(k) Number (if known) K240350

### Device Name AnyPlus II Spinal Fixation System

### Indications for Use (Describe)

The GS Medical AnyPlus II Spinal Fixation System is composed of a non-cervical spinal fixation devices. It is intended for use as posterior pedicle screw fixation system (T1-S2), a posterior hook fixation system (T1-L5), or as an anterolateral fixation system (T8-L5). All components in the system are limited to skeletally mature patients. System components are to be used for immobilization and stabilization of the spine as an adjunct to fusion. These devices are indicated for all following indications regardless of intended use: degenerative disc disease (defined as discogenic back pain with degeneration of the disc confirmed by history and radiographic studies), spondylolisthesis, trauma (i.e., fracture or dislocation), deformities or curvatures (i.e., scoliosis, and/or lordosis, Scheuermann's Disease), tumor, stenosis, pseudoarthrosis, and failed previous fusion.

| Type of Use (Select one or both, as applicable) |  |
|-------------------------------------------------|--|
|-------------------------------------------------|--|

X Prescription Use (Part 21 CFR 801 Subpart D)

Over-The-Counter Use (21 CFR 801 Subpart C)

### CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

#### 510(K) SUMMARY l.

# GS Medical's AnyPlus II Spinal Fixation System

#### SUBMITTER'S ADDRESS, TELEPHONE NUMBER, CONTACT PERSON A.

Seon Yeon Kim RA Manager GS Medical Co. Ltd. 90 Osongsaengmyong 4-ro Osong-eup, Heungdeok-gu, Gheongjusi, Chungcheongbuk-do 28161 Korea Tel.: 405-535-2719 Email: sykim@gsmedi.com

# Author:

Arunkumar Prabhakaran US RA Submission Manager Phone: 978-358-7307 Email: regulatorysubmissions@rqmis.com

# Primary Contact:

Barry E. Sands RQMIS Inc. 110 Haverhill Road, Suite 524 Amesbury, MA 01913 Phone: (978) 358-7307 Email: regulatorysubmissions@rqmis.com

Date Prepared: January 30, 2024

#### B. SUBJECT DEVICE

| Trade/Proprietary Name of Device: | Anyplus II Spinal Fixation System       |
|-----------------------------------|-----------------------------------------|
| Common Or Usual Name:             | Pedicle Screw System                    |
| Classification Name:              | Thoracolumbosacral Pedicle Screw System |
| Regulation Number:                | 21 CFR 888.3070                         |
| Classification:                   | Class 2                                 |
| Product Code:                     | NKB, KWP, KWQ                           |

{4}------------------------------------------------

### C. PREDICATE DEVICES

# Primary Predicate:

| Trade/Proprietary Name of Device: | Anyplus Spinal Fixation System          |
|-----------------------------------|-----------------------------------------|
| 510(k) number                     | K172546                                 |
| Common Or Usual Name:             | Pedicle Screw System                    |
| Classification Name:              | Thoracolumbosacral Pedicle Screw System |
| Regulation Number:                | 21 CFR 888.3070                         |
| Classification:                   | Class 2                                 |
| Product Code:                     | NKB, KWP, KWQ, MNH, MNI                 |

### D. INDICATIONS FOR USE

The GS Medical AnyPlus II Spinal Fixation System is composed of a non-cervical spinal fixation devices. It is intended for use as posterior pedicle screw fixation system (T1-S2), a posterior hook fixation system (T1-L5), or as an anterolateral fixation system (T8-L5). All components in the system are limited to skeletally mature patients. System components are to be used for immobilization and stabilization of the spine as an adjunct to fusion. These devices are indicated for all following indications regardless of intended use: degenerative disc disease (defined as discogenic back pain with degeneration of the disc confirmed by history and radiographic studies), spondylolisthesis, trauma (i.e., fracture or dislocation), deformities or curvatures (i.e., scoliosis, kyphosis, and/or lordosis, Scheuermann's Disease), tumor, stenosis, pseudoarthrosis, and failed previous fusion.

### E. DEVICE DESCRIPTION

The AnyPlus II Spinal Fixation System consists of various screws (mono and polyaxial screws, reduction screw, canulated screws, long arm screws, cortical screws, and a set screw). The screws are available with diameters ranging from 4.5 to 10.5mm, and lengths ranging from 20-100mm in 5mm increments. They are available in noncannulated, cannulated, and longarm geometries with both mono and poly types for each geometry option. All AnyPlus II Spinal Fixation System implant components are manufactured from Ti6AI4V ELI according to ASTM F136.

These implants are supplied non-sterile and are not reusable. The instruments in this system are supplied non-sterile, are reusable, and patient contacting materials are fabricated from Precipitation Hardened Stainless Steels (STS630 and STS555) that complies with ASTM F899.

{5}------------------------------------------------

### PERFORMANCE DATA F.

The worst-case construct of the AnyPlus II Spinal Fixation System underwent testing according to ASTM F1717, specifically: static and dynamic axial compression bending testing and static torsion testing. The results demonstrate that the subject device is substantially equivalent to the primary predicate's (K091717) biomechanical performance.

### G. CONCLUSION

The technological differences between the subject device and the predicates (K172546) do not raise new questions of safety and effectiveness. Any differences in technological characteristics have been tested and documented. The subject device and predicate(s) (K172546) have been determined to be substantially equivalent in terms of indications for use, materials, performance, sterility, and biocompatibility.

---

**Source:** [https://fda.innolitics.com/device/K240350](https://fda.innolitics.com/device/K240350)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
