← Product Code [HBL](/productcode/HBL) · K240319

# DORO LUCENT Skull Clamp (1101.001); DORO LUCENT Skull Clamp Pediatric Set (1101.040) (K240319)

_Pro Med Instruments GmbH · HBL · Apr 24, 2024 · Neurology · SESE_

**Canonical URL:** https://fda.innolitics.com/device/K240319

## Device Facts

- **Applicant:** Pro Med Instruments GmbH
- **Product Code:** [HBL](/productcode/HBL.md)
- **Decision Date:** Apr 24, 2024
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 882.4460
- **Device Class:** Class 2
- **Review Panel:** Neurology
- **Attributes:** Therapeutic, Pediatric

## Indications for Use

The DORO LUCENT Skull clamp is part of a mechanical support system, which is used during head or neck surgery. The product can be used in an intra-operative setting for CT and X-ray imaging. The DORO LUCENT Skull clamp Pediatric Set is part of a mechanical support system, which is used during head or neck surgery. The product can be used in an intra-operative setting for CT and X-ray imaging. The target patient population includes children over 5 years of age.

## Device Story

Mechanical support system for rigid cranial fixation during neurosurgery; provides stable head positioning for intra-operative CT and X-ray imaging. Device comprises fixed arm, adjustable arm, and modular pin holders; configurable for three-pin or four-pin setups (pediatric). Operated by neurosurgeons in OR settings. Provides rigid skeletal fixation; ensures patient head stability during procedures. Benefits include secure immobilization and compatibility with imaging modalities, reducing need for repositioning.

## Clinical Evidence

Bench testing only. Performance verified via maximum load, static load, creep, pin force accuracy, torque load resistance, and CT imaging artifact evaluation. Reprocessing validated per ANSI/AAMI ST98:2022, DIN EN ISO 17664-1:2021, DIN EN ISO 17664-2:2021, and AAMI TIR 12:2020.

## Technological Characteristics

Mechanical cranial fixation system. Components: fixed arm, adjustable arm, modular pin holders. Supports three- or four-point fixation. Materials: radiolucent materials suitable for CT/X-ray. Connectivity: None. Sterilization: Validated for automated cleaning and disinfection. Standards: ANSI/AAMI ST98:2022, DIN EN ISO 17664-1:2021, DIN EN ISO 17664-2:2021, AAMI TIR 12:2020.

## Regulatory Identification

A neurosurgical head holder (skull clamp) is a device used to clamp the patient's skull to hold head and neck in a particular position during surgical procedures.

## Predicate Devices

- DORO LUCENT iXI and iMRI Headrest System ([K191740](/device/K191740.md))

## Reference Devices

- DORO Headrest System ([K001808](/device/K001808.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

April 24, 2024

Image /page/0/Picture/1 description: The image contains two logos. On the left is the Department of Health and Human Services logo. On the right is the Food and Drug Administration logo. The FDA logo is a blue square with the letters "FDA" in white, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue.

Pro Med Instruments GmbH Rabel Talpur Regulatory Affairs Specialist Bötzinger Straße 86 Freiburg, Baden-württemberg 79111 Germany

Re: K240319

Trade/Device Name: DORO LUCENT Skull Clamp (1101.001); DORO LUCENT Skull Clamp Pediatric Set (1101.040) Regulation Number: 21 CFR 882.4460 Regulation Name: Neurosurgical Head Holder (Skull Clamp) Regulatory Class: Class II Product Code: HBL Dated: February 2, 2024 Received: February 2, 2024

Dear Rabel Talpur:

We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device"

{1}------------------------------------------------

(https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).

Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30. Design controls; 21 CFR 820.90. Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review. the OS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely.

Digitally signed by Adam D. Adam D. Pierce -S Date: 2024.04.24 Pierce -S 13:21:24 -04'00'

Adam D. Pierce, Ph.D. Assistant Director DHT5A: Division of Neurosurgical, Neurointerventional

{2}------------------------------------------------

and Neurodiagnostic Devices OHT5: Office of Neurological and Physical Medicine Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{3}------------------------------------------------

### Indications for Use

510(k) Number (if known) K240319

Device Name DORO LUCENT Skull Clamp (1101.001); DORO LUCENT Skull Clamp Pediatric Set (1101.040)

Indications for Use (Describe)

The DORO LUCENT Skull clamp is part of a mechanical support system, which is used during head or neck surgery. The product can be used in an intra-operative setting for CT and X-ray imaging.

The DORO LUCENT Skull clamp Pediatric Set is part of a mechanical support system, which is used during head or neck surgery. The product can be used in an intra-operative setting for CT and X-ray imaging. The target patient population includes children over 5 years of age.

Type of Use (Select one or both, as applicable)

> Prescription Use (Part 21 CFR 801 Subpart D)

Over-The-Counter Use (21 CFR 801 Subpart C)

### CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{4}------------------------------------------------

# 510(k) Summary K240319

{5}------------------------------------------------

#### Device Information: 1

| Category                                         | Comments                                                                                                                                                                                                    |
|--------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Sponsor:                                         | pro med instruments GmbH<br>Bötzinger Straße 86<br>79111 Freiburg im Breisgau<br>Germany<br>Tel: + 49 (0) 761 384 222 10<br>Fax: +49 (0) 761 384 222 81<br>E-Mail: regulatoryaffairs@blackforestmedical.com |
| Correspondent Contact Information                | Name: Rabel Talpur<br>Position: Regulatory Affairs Specialist<br>Tel.: + 49 (0) 761 384 222 67<br>E-Mail: regulatoryaffairs@blackforestmedical.com                                                          |
| Device Common Name:                              | Neurosurgical head holder (skull clamp)                                                                                                                                                                     |
| Device Regulation & Name                         | 882.4460, Neurosurgical head holder (skull clamp)                                                                                                                                                           |
| Classification & Product Code:<br>510(K) Number: | Device Classification: II<br>Product Code: HBL<br>510(K) Number: K240319                                                                                                                                    |
| Device Proprietary Name:                         | DORO LUCENT® Skull Clamp                                                                                                                                                                                    |

#### 2 Predicate Device Information:

| Predicate Device                                | DORO LUCENT® iXI and iMRI Headrest System                     |
|-------------------------------------------------|---------------------------------------------------------------|
| Predicate Device Manufacturer:                  | pro med instruments GmbH                                      |
| Predicate Device Common Name:                   | Neurosurgical head holder (skull clamp)                       |
| Predicate Device Premarket Notification #       | K191740                                                       |
| Predicate Device Classification & Name          | Device Class: II<br>Holder, head, neurosurgical (Skull Clamp) |
| Predicate Device Classification & Product Code: | Device Class: II<br>Product Code: HBL                         |

#### ന Reference Device Information:

| Reference Device                          | DORO® Headrest System                   |
|-------------------------------------------|-----------------------------------------|
| Reference Device Manufacturer:            | pro med instruments GmbH                |
| Reference Device Common Name:             | Neurosurgical head holder (skull clamp) |
| Reference Device Premarket Notification # | K001808                                 |

{6}------------------------------------------------

### pro med instruments GmbH

DORO LUCENT® Skull Clamp - K240319

| Reference Device Classification & Name          | Device Class: II<br>Neurosurgical head holder (skull clamp) |
|-------------------------------------------------|-------------------------------------------------------------|
| Reference Device Classification & Product Code: | Device Class: II<br>Product Code: HBL                       |

#### Date Summary Prepared ব

24.04.2024

#### Device Description ട

The DORO LUCENT® Skull Clamp ensures an adequate positioning of a patient's head and provides rigid fixation for neurosurgery. The device is suitable for intra-operative X-ray and CT based imaging procedures.

Image /page/6/Picture/7 description: The image shows a medical device called a DORO head holder. The device is black and blue and has a U-shape. The DORO logo is visible on the bottom of the device.

Figure 1: DORO LUCENT® Skull Clamp (REF: 1101.001)

The DORO LUCENT® Skull Clamp consists of a fixed arm, an adjustable arm and modular pin holders. Thus, using the DORO LUCENT® Skull Clamp Pediatric Set, the skull clamp can be configured as three-pin or four-pin pediatric set up.

Image /page/6/Picture/10 description: The image shows three different views of a blue and black mechanical part. The first view shows the part from the side, with a cylindrical body and two curved black arms extending from the center. The second view shows a similar part, but with a different end and the arms in a slightly different position. The third view shows a bolt with a large, ridged head and a threaded shaft, with the word "Start" printed on the side.

Figure 2: DORO LUCENT® Skull Clamp pediatric Set (REF: 1101.040)

{7}------------------------------------------------

#### Indications for Use e

The DORO LUCENT® Skull clamp is part of a mechanical support system, which is used during head or neck surgery. The product can be used in an intra-operative setting for CT and X-ray imaging.

The DORO LUCENT® Skull clamp Pediatric Set is part of a mechanical support system, which is used during head or neck surgery. The product can be used in an intra-operative setting for CT and X-ray imaging. The target patient population includes children over 5 years of age.

| Characteristic                        | pro med instruments GmbH<br>DORO LUCENT® Skull Clamp<br>(K240319)                                                                                                                                                                                                                                                                                                                                                                                                                                                            | pro med instruments GmbH<br>DORO LUCENT® iXI and iMRI<br>Headrest System<br>(K191740)                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                    | Impact on<br>Substantial<br>Equivalence |
|---------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------|
| Company                               | pro med instruments GmbH                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                     | pro med instruments GmbH                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                 | Substantially<br>Equivalent             |
| Regulation<br>Number                  | 882.4460                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                     | 882.4460                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                 | Substantially<br>Equivalent             |
| Product Code                          | HBL                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                          | HBL                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                      | Substantially<br>Equivalent             |
| Intended Use                          | The DORO LUCENT® Skull clamp<br>is part of a mechanical support<br>system, which is used during head<br>or neck surgery. The product can<br>be used in an intra-operative<br>setting for CT and X-ray imaging.                                                                                                                                                                                                                                                                                                               | The DORO LUCENT radiolucent/MRI<br>Compatible Headrest System with<br>Skull Pins are components of a<br>mechanical support system, which is<br>used in cranial and spine surgery<br>when rigid skeletal fixation is required<br>for cranial stabilization and when<br>intra-operative<br>imaging is used.<br>The DORO LUCENT radiolucent/MRI<br>Compatible Headrest System with<br>noninvasive head<br>positioning or noninvasive cranial<br>stabilization device are components of<br>a mechanical support system, which is<br>used in cranial and spine surgery<br>when noninvasive head positioning or<br>noninvasive cranial stabilization is<br>required and when intra-operative<br>imaging is used.<br>The DORO LUCENT Headrest System<br>provides an interface for accessories<br>like retractor systems, navigation<br>adaptors or other items. | Substantially<br>Equivalent             |
|                                       |                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                              |                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                          |                                         |
| Characteristic                        | pro med instruments GmbH<br>DORO LUCENT® Skull Clamp<br>(K240319)                                                                                                                                                                                                                                                                                                                                                                                                                                                            | pro med instruments GmbH<br>DORO LUCENT® iXI and iMRI<br>Headrest System<br>(K191740)                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                    | Impact on<br>Substantial<br>Equivalence |
| Indications for<br>Use                | The DORO LUCENT® Skull clamp<br>is part of a mechanical support<br>system, which is used during head<br>or neck surgery. The product can<br>be used in an intra-operative<br>setting for CT and X-ray imaging.<br>The DORO LUCENT® Skull clamp<br>Pediatric Set is part of a<br>mechanical support system, which<br>is used during head or neck<br>surgery. The product can be used<br>in an intra-operative setting for CT<br>and X-ray imaging. The target<br>patient population includes<br>children over 5 years of age. | The DORO LUCENT radiolucent/MRI<br>Compatible Headrest System with<br>Skull Pins are components of a<br>mechanical support system, which is<br>used in cranial and spine surgery<br>when rigid skeletal fixation is required<br>for cranial stabilization and when<br>intra-operative imaging is used.<br>The DORO LUCENT radiolucent/MRI<br>Compatible Headrest System with<br>noninvasive head positioning or<br>noninvasive cranial stabilization<br>device are components of a<br>mechanical support system, which is<br>used in cranial and spine surgery<br>when noninvasive head positioning or<br>noninvasive cranial stabilization is<br>required and when intra-operative<br>imaging is used.<br>The DORO LUCENT Headrest System<br>provides an interface for accessories<br>Like retractor systems, navigation<br>adaptors or other items.    | Substantially<br>Equivalent             |
| Technology                            | Cranial fixation                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                             | Cranial fixation                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                         | Substantially<br>Equivalent             |
| Features:                             |                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                              |                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                          |                                         |
| Principles of<br>operation            | Mechanical support system                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                    | Mechanical support system                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                | Substantially<br>Equivalent             |
| Method of<br>cranial<br>stabilization | Rigid skeletal fixation                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                      | Rigid skeletal fixation or secured<br>noninvasive stabilization                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                          | Substantially<br>Equivalent             |
| Method of<br>cranial<br>fixation      | Three- and Four-point fixation in<br>prone, supine or lateral position                                                                                                                                                                                                                                                                                                                                                                                                                                                       | Three-point fixation in prone, supine,<br>lateral or sitting position                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                    | Substantially<br>Equivalent             |
| Intra-<br>operative<br>imaging        | CT-, X-Ray Imaging                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                           | MR-, CT-, X-Ray- and Angiography-<br>Imaging                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                             | Substantially<br>Equivalent             |
| reprocessing                          | Automated cleaning and<br>disinfection between uses                                                                                                                                                                                                                                                                                                                                                                                                                                                                          | Manual cleaning and disinfection<br>between uses                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                         | Substantially<br>Equivalent             |

#### 7 Comparison of the Technological Characteristics

{8}------------------------------------------------

### pro med instruments GmbH

DORO LUCENT® Skull Clamp - K240319

## 7.1 Summary of Supporting Data

The above-listed Technological Characteristics show that the DORO LUCENT® Skull Clamp, and the DORO LUCENT® iXI and iMRI Headrest System are substantially equivalent.

{9}------------------------------------------------

#### Discussion of Performance Testing ထ

Tests were performed and the results are shown in the table below.

| Test                                                                                                                                                                        | Result                                                                                                                                                        |
|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------|
| DORO LUCENT® Skull Clamp                                                                                                                                                    |                                                                                                                                                               |
| Maximum Load Test (System Test)                                                                                                                                             | Pass                                                                                                                                                          |
| Verifies the skull clamp can resist the forces imposed by the<br>patient and the surgeon when in use and in locked position.                                                | The skull clamp is capable to withstand given<br>force without movement in any connection,<br>joint and/or plastic deformation.                               |
| Usability                                                                                                                                                                   | Pass                                                                                                                                                          |
| Verifies if the usability of the skull clamp is given.                                                                                                                      | The usability of skull Clamp is given                                                                                                                         |
| Static load (Latching teeth mechanism)                                                                                                                                      | Pass                                                                                                                                                          |
| Verifies the ability of the skull clamp to withstand maximum<br>static load.                                                                                                | The skull clamp withstands a static load for<br>given duration without signs of permanent<br>deformation, fracture, or breakage.                              |
| Creep                                                                                                                                                                       | Pass                                                                                                                                                          |
| Verifies the mechanical integrity of the Skull Clamp and its<br>ability to withstand constant loading over time without a<br>significant loss of clamping force.            | The skull clamp can maintain the applied<br>maximum force without a force deviation from<br>the initially applied load by a defined value.                    |
| Pin force accuracy                                                                                                                                                          | Pass                                                                                                                                                          |
| Verifies the force delivery component to ensure the force<br>readings are accurate and depict the actual force applied.                                                     | The skull clamp force delivery component is<br>verified at each force level throughout its range<br>to deliver the stated force within the actual<br>setting. |
| Torque load resistance                                                                                                                                                      | Pass                                                                                                                                                          |
| Verifies skull clamp can withstand rotational movement when<br>in use and locked position without impacting the locking<br>mechanism.                                       | The skull clamp withstands a defined torque for<br>given amount of time without deformation or<br>any structural failures.                                    |
| CT Evaluation                                                                                                                                                               | Pass                                                                                                                                                          |
| Verifies that the skull clamp does not affects the CT images in<br>a negative way and produce less artifacts according to the set<br>benchmark for adequate imaging result. | The skull Clamp produces less artefacts than<br>the set benchmark as defined in the test report.                                                              |

{10}------------------------------------------------

| Reprocessing                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                      | Pass                                                                                                                                    |
|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------|
| The reprocessing is done according to specifications that are defined in the instruction for use (IFU) of the product. The accumulation of residue evaluation, Cleaning effectiveness validation and Disinfection effectiveness validation is done according to:<br>• ANSI / AAMI ST98:2022 - Cleaning validation of health care products - Requirements for development and validation of cleaning process for medical devices<br>• DIN EN ISO 17664-1:2021 - Processing of health care products - Information to be provided by the medical device manufacturer for the processing of medical devices - Part 1: Critical and semi-critical medical devices<br>• DIN EN ISO 17664-2:2021 Processing of health care products - Information to be provided by the medical device manufacturer for the processing of medical devices - Part 2: Non-critical medical devices.<br>• AAMI TIR 12: 2020, Designing, testing, and labeling reusable medical devices for reprocessing in health care facilities: A guide for medical device manufacturers | The tests data shows the reprocessing of DORO LUCENT® Skull Clamp is validated and test report in support to the statement is provided. |

Testing confirmed that the performance of the DORO LUCENT® Skull Clamp meets the product specification of the device.

#### 9 Conclusion

The DORO LUCENT® Skull Clamp is used as a head and neck support to stabilize a patient's head during neurosurgical operative procedures.

This device is comparable in design, construction, intended use and performance characteristics to the predicate device and reference devices.

Based on available 510(k) information herein provided, DORO LUCENT® Skull Clamp is considered substantially equivalent to the predicate device in terms of intended use, technology and performance specifications, additional characteristic such as four-pin fixation is comparable to the reference device K001808.

There are no differences between the devices which may raise new issues concerning safety or effectiveness.

---

**Source:** [https://fda.innolitics.com/device/K240319](https://fda.innolitics.com/device/K240319)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
