← Product Code [OAB](/productcode/OAB) · K240297

# Canady Helios Cold Plasma™ XL-1000CP™ Ablation System (XL-1000CPSYS) (K240297)

_Us Medical Innovations, LLC · OAB · May 3, 2024 · General, Plastic Surgery · SESE_

**Canonical URL:** https://fda.innolitics.com/device/K240297

## Device Facts

- **Applicant:** Us Medical Innovations, LLC
- **Product Code:** [OAB](/productcode/OAB.md)
- **Decision Date:** May 3, 2024
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 878.4400
- **Device Class:** Class 2
- **Review Panel:** General, Plastic Surgery
- **Attributes:** Therapeutic

## Indications for Use

The Canady Helios Cold Plasma™ XL-1000CP™ Ablation System is indicated for the gas enhanced ablation of soft tissue using helium-based plasma.

## Device Story

The Canady Helios Cold Plasma™ XL-1000CP™ Ablation System is a transportable electrosurgical device for use in operating rooms by clinicians. It utilizes a SMART Electrosurgical Generator to create high-frequency, short-duration energy pulses combined with regulated helium gas. This energy is delivered to target soft tissue via sterile, single-use monopolar Canady Helios™ Cold Plasma Ablators to perform gas-enhanced ablation. The system includes a footswitch, power supply, and trolley cart. By delivering high-frequency, low-energy current via helium plasma, the device enables precise tissue ablation. The healthcare provider operates the system via the generator interface and footswitch to control energy and gas flow, facilitating surgical tissue management.

## Clinical Evidence

Bench testing only. Performance was verified through electrical safety (AAMI/ANSI ES 60601-1), electromagnetic compatibility (IEC 60601-1-2), high-frequency surgical equipment standards (IEC 60601-2-2), sterilization validation (ISO 11135, ISO 11737), software verification/validation (IEC 62304), and ablation effectiveness testing. No clinical data was provided.

## Technological Characteristics

High-frequency (390 kHz) electrosurgical generator; 4kV peak-to-peak voltage; helium gas-enhanced monopolar ablation; FPGA-based control; LED display; convection cooling; 100-240 VAC input. Sterile, single-use monopolar ablators. Complies with IEC 60601-1, IEC 60601-1-2, IEC 60601-2-2, and IEC 62304 standards.

## Regulatory Identification

An electrosurgical cutting and coagulation device and accessories is a device intended to remove tissue and control bleeding by use of high-frequency electrical current.

## Predicate Devices

- Canady Plasma SMART XL-1000 Electrosurgical Generator with Accessories ([K192124](/device/K192124.md))

## Reference Devices

- Canady Plasma Hybrid Scalpel ([K113500](/device/K113500.md))
- Canady Plasma Probes for Flexible Endoscopy ([K100669](/device/K100669.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA logo on the right. The FDA logo is in blue and includes the letters "FDA" in a square and the words "U.S. Food & Drug Administration".

May 3, 2024

US Medical Innovations, LLC % Audrey Swearingen Sr. Regulatory Consultant Emergo Global Consulting, LLC 2500 Bee Cave Road Building 1, Suite 300 Austin, Texas 78746

Re: K240297

Trade/Device Name: Canady Helios Cold Plasma™ XL-1000CP™ Ablation System (XL-1000CPSYS) Regulation Number: 21 CFR 878.4400 Regulation Name: Electrosurgical cutting and coagulation device and accessories Regulatory Class: Class II Product Code: GEI Dated: February 1, 2024 Received: February 1, 2024

Dear Audrey Swearingen:

We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device"

{1}------------------------------------------------

(https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).

Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30. Design controls; 21 CFR 820.90. Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review. the OS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely.

Digitally signed by Mark Mark Trumbore -S Trumbore -S Date: 2024.05.03 11:01:07 -04'00

Mark Trumbore, Ph.D. Assistant Director DHT4A: Division of General Surgery Devices OHT4: Office of Surgical and Infection Control Devices

{2}------------------------------------------------

Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{3}------------------------------------------------

# K240297

DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration

# Indications for Use

Submission Number (if known)

K240297

Device Name

Canady Helios Cold Plasma™ XL-1000CP™ Ablation System (XL-1000CPSYS)

Indications for Use (Describe)

The Canady Helios Cold Plasma™ XL-1000CP™ Ablation System is indicated for the gas enhanced ablation of soft tissue using helium-based plasma.

Type of Use (Select one or both, as applicable)

> Prescription Use (Part 21 CFR 801 Subpart D)

Over-The-Counter Use (21 CFR 801 Subpart C)

# CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

# *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{4}------------------------------------------------

# 510(k) Summary Canady Helios Cold Plasma™ XL-1000CP™ Ablation System

The following information is provided in accordance with 21 CFR 807.92.

#### Submission Sponsor:

US Medical Innovations LLC 6930 Carroll Avenue, Suite 1000 Takoma Park, MD 20912 Contact: Jerome Canady, MD., FACS Title: CEO Phone: (301) 270-0147

#### Submission Correspondent:

Emergo by UL 2500 Bee Cave Road, Building 1, Suite 300 Austin, TX 78746 Contact: Ms. Audrey Swearingen Senior Regulatory Consultant Title: Phone: (512) 327-9997 LST.AUS.ProjectManagement@ul.com Email:

# Date: April 18, 2024

#### Device Identification:

| Trade or Proprietary Name: | Canady Helios Cold Plasma™ XL-1000CP™ Ablation System (XL-1000CPSYS |
|----------------------------|---------------------------------------------------------------------|
| Common or Usual Name:      | Surgical ablation system                                            |
| Regulation Number:         | 21 CFR 878.4400                                                     |
| Regulation Name:           | Electrosurgical cutting and coagulation device and accessories      |
| Product Code:              | GEI                                                                 |
| Device Class:              | II                                                                  |
| Panel:                     | General & Plastic Surgery                                           |

#### Legally Marketed Predicate Device:

| Trade Name:        | Canady Plasma SMART XL-1000 Electrosurgical Generator with Accessories |
|--------------------|------------------------------------------------------------------------|
| 510(k) Number:     | K192124                                                                |
| Manufacturer:      | US Medical Innovations LLC                                             |
| Regulation Number: | 21 CFR 878.4400                                                        |
| Regulation Name:   | Electrosurgical cutting and coagulation device and accessories         |
| Product Code:      | GEI                                                                    |

# Intended Use/Indications for Use:

The Canady Helios Cold Plasma™ XL-1000CP™ Ablation System is indicated for the gas enhanced ablation of soft tissue using helium-based plasma.

{5}------------------------------------------------

# Device Description:

The Canady Helios Cold Plasma™ XL-1000CP™ Ablation System produces a Helium-based cold plasma spray to deliver high frequency low energy current intended for the ablation of soft tissues.

The Canady Helios Cold Plasma™ XL-1000CP Ablation System includes the following items:

- Canady Helios Cold Plasma™ XL-1000CP SMART Electrosurgical Generator
- Canady Helios™ Cold Plasma Ablators
- Foot Pedal
- Trolley Cart

The Canady Helios Cold Plasma™ XL-1000CP™ SMART Electrosurgical Generator is a transportable nonsterile electrical device compatible for use in the operating room environment. The Canady Helios Cold Plasma™ XL-1000CP SMART Electrosurgical Generator creates high frequency, short duration energy pulses with regulated Helium gas, which is delivered to the target tissue via the Canady Helios™ Cold Plasma Ablators. The Canady Helios™ Cold Plasma Ablators are sterile monopolar electrosurgical instruments intended for single use.

# Substantial Equivalence:

The following table compares the proposed Canady Helios Cold Plasma™ XL-1000CP™ Ablation System to the Canady Plasma SMART XL-1000 Electrosurgical Generator with Accessories predicate device, with respect to intended use and technological characteristics.

| Trade Name                         | Canady Helios Cold Plasma™ XL-<br>1000CP™ SMART Ablation<br>System                                                                                                                                      | Canady Plasma SMART XL-1000<br>Electrosurgical Generator with<br>Accessories System (K192124)                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                          | Significant Difference                                                                                    |
|------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------|
| Manufacturer                       | US Medical Innovations, LLC                                                                                                                                                                             | US Medical Innovations, LLC                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                            | N/A                                                                                                       |
| Product code(s)/<br>Regulation No. | GEI /<br>878.4400                                                                                                                                                                                       | GEI /<br>878.4400                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                      | Both are under the<br>same regulation                                                                     |
| Intended Use                       | Gas enhanced ablation of soft<br>tissue.                                                                                                                                                                | Gas enhanced ablation of soft<br>tissue.                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                               | None                                                                                                      |
| Indications for Use                | The Canady Helios Cold Plasma™<br>XL-1000CP™ SMART Ablation<br>System is indicated for gas<br>enhanced ablation of soft tissue<br>using helium based plasma.                                            | The Canady Plasma® SMART XL-<br>1000™ Electrosurgical Generator<br>with Accessories is intended to<br>provide gas-enhanced<br>coagulation during general<br>surgery procedures. The Canady<br>Plasma® SMART XL-1000™<br>Electrosurgical Generator with<br>Accessories is designed for gas<br>enhanced coagulation when<br>used only with the Canady<br>Plasma Electrosurgery Unit and<br>compatible monopolar and/or<br>bipolar handpieces.<br>The Canady Electrosurgery Unit<br>Generators are intended to cut<br>and/or coagulate tissue when<br>used with compatible<br>monopolar and/or bipolar RF | The XL-1000CP uses a<br>subset of the<br>Indications for use<br>defined for the<br>predicate device.      |
| System Components                  | • Canady Helios Cold Plasma™<br>XL-1000CP™ SMART<br>Electrosurgical Generator<br>• Canady Helios™ Cold Plasma<br>Ablators<br>• Footswitch<br>• Power Supply Cord<br>• XL-1000/XL-1000CP Trolley<br>Cart | handpieces. The Canady<br>Electrosurgery Unit Generators<br>are capable of monopolar<br>argon gas enhanced<br>coagulation when used with<br>Canady Plasma Coagulator and<br>probes.<br>• Canady Plasma® SMART XL-<br>1000™ Electrosurgical<br>Generator<br>• Canady Plasma Hybrid Scalpel<br>(K113500)<br>• Canady Plasma Probes for<br>Flexible Endoscopy<br>(K100669)<br>• Footswitch<br>• Power Supply Cord<br>• XL-1000/XL-1000CP Trolley<br>Cart                                                                                                                                                  | Same. The subject<br>Ablators are identical<br>to the predicate's<br>handpieces.                          |
| Generator Type                     | High Frequency                                                                                                                                                                                          | High Frequency                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                         | None                                                                                                      |
| Inert Gas Type                     | Helium                                                                                                                                                                                                  | Argon                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                  | The XL-1000CP uses<br>the inert gas Helium<br>instead of Argon.                                           |
| Frequency                          | 390 kHz                                                                                                                                                                                                 | 390 kHz                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                | None                                                                                                      |
| Peak to Peak Voltage               | 4kV (1000 Ω)                                                                                                                                                                                            | 4kV (1000 Ω)                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                           | None                                                                                                      |
| Main CPU Type                      | FPGA                                                                                                                                                                                                    | FPGA                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                   | None                                                                                                      |
| Display Type                       | LED with backlight                                                                                                                                                                                      | LED with backlight                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                     | None                                                                                                      |
| Monopolar                          | Yes, insulated                                                                                                                                                                                          | Yes, insulated                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                         | None                                                                                                      |
| Bipolar                            | No                                                                                                                                                                                                      | Yes                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                    | The XL-1000CP<br>generator only<br>supports monopolar<br>ablation.                                        |
| Monopolar Modes                    | Canady Helios Cold Plasma<br>Ablation Mode (equivalent to<br>Fulgurate High on the XL-1000)                                                                                                             | Argon Plasma - Cut (Pure, Blend<br>I,II,III, Pure High Cut, Blend High<br>Cut I,II), E-Cut, Argon Plasma<br>(Spray, Fulgurate High,<br>Fulgurate Low, Desiccate I,II,III)                                                                                                                                                                                                                                                                                                                                                                                                                              | Cold Plasma ablation<br>mode is the same as<br>the Argon Plasma<br>Fulgurate High mode<br>of the XL-1000. |
| Gas Flow Rate                      | 0.0 - 5.0 L/min                                                                                                                                                                                         | 0.0 - 5.0 L/min                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                        | None                                                                                                      |
| Power Setting Shown                | Yes                                                                                                                                                                                                     | Yes                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                    | None                                                                                                      |
| Cooling Type                       | Convection                                                                                                                                                                                              | Convection                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                             | None                                                                                                      |
| Input Voltage, VAC                 | 100 – 240                                                                                                                                                                                               | 100 - 240                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                              | None                                                                                                      |
| Performance                        | Shown to ablate soft tissue at<br>specified parameters.                                                                                                                                                 | Shown to ablate soft tissue at<br>specified parameters.                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                | Both perform as<br>intended to ablate<br>soft tissue.                                                     |

{6}------------------------------------------------

# Performance Testing:

The following tests were conducted to confirm that the Canady Helios Cold Plasma™ XL-1000CP™ Ablation System meets specifications and that it performs equivalently to the predicate Canady Plasma SMART XL-1000 Electrosurgical Generator with Accessories.

{7}------------------------------------------------

| Electrical Safety Testing                     | Met criteria |
|-----------------------------------------------|--------------|
| High Frequency Testing                        | Met criteria |
| Electromagnetic Compatibility Testing         | Met criteria |
| Electromechanical Safety Testing              | Met criteria |
| Package Integrity                             | Met criteria |
| Shelf Life                                    | Met criteria |
| Sterilization                                 | Met criteria |
| Software and System Verification / Validation | Met criteria |
| Cybersecurity                                 | Met criteria |
| Biological Safety                             | Met criteria |
| Ablation Effectiveness                        | Met criteria |
| Thermal Effect                                | Met criteria |

The device complies with the following applicable test standards:

| ANSI/AAMI/ISO 11135:<br>2014/AMD 2018        | Sterilization of health-care products — Ethylene oxide — Requirements for the<br>development, validation and routine control of a sterilization process for medical<br>devices                     |
|----------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| ANSI/AAMI/ISO 11737-<br>1:2018               | Sterilization of health care products — Microbiological methods — Part 1:<br>Determination of a population of microorganisms on products                                                           |
| ANSI/AAMI/ISO 11737-2<br>2019                | Sterilization of medical devices — Microbiological methods — Part 2: Tests of<br>sterility performed in the definition, validation and maintenance of a sterilization<br>process                   |
| AAMI/ANSI ES 60601-<br>1:2005/2012           | Medical Electrical Equipment - Part 1 General Requirements for Basic Safety and<br>Essential Performance                                                                                           |
| IEC 60601-1-2:2020                           | Medical Electrical Equipment - Part 1-2: General Requirements for Basic Satety and<br>Essential Performance - Collateral Standard: Electromagnetic Compatibility -<br>Requirements and Tests       |
| IEC 60601-2-2:2017<br>Edition 6.0            | Medical electrical equipment - Part 2-2: Particular requirements for the basic safety<br>and essential performance of high frequency surgical equipment and high frequency<br>surgical accessories |
| IEC 60601-1-<br>6:2010+AMD1:2013             | Medical electrical equipment - Part 1-6: General requirements for basic safety and<br>essential performance - Collateral standard: Usability                                                       |
| IEC 62304 Ed 1.1 2015-06,<br>Consol. version | Medical device software - Software life cycle processes                                                                                                                                            |

# Conclusion:

The Canady Helios Cold Plasma™ XL-1000CP™ Ablation System has the same intended use and the same or similar technological characteristics as the predicate device. Testing to recognized consensus standards and internal performance testing has demonstrated the device is as safe and effective for requested intended use, as the predicate device. The minor technological differences compared to the predicate do not raise different types of questions of safety and effectiveness. Therefore, the Canady Helios Cold Plasma™ XL-1000CP™ Ablation System is substantially equivalent to the predicate device.

---

**Source:** [https://fda.innolitics.com/device/K240297](https://fda.innolitics.com/device/K240297)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
