PillCam™ Genius SB System with PillCam Software v9.7 and PillCam Cloud Reader Software
K240276 · Given Imaging Ltd. (D.B.A. Medtronic) · NEZ · May 10, 2024 · Gastroenterology, Urology
Device Facts
Record ID
K240276
Device Name
PillCam™ Genius SB System with PillCam Software v9.7 and PillCam Cloud Reader Software
Applicant
Given Imaging Ltd. (D.B.A. Medtronic)
Product Code
NEZ · Gastroenterology, Urology
Decision Date
May 10, 2024
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 876.1300
Device Class
Class 2
Indications for Use
The PillCam™ Genius SB Capsule Endoscopy Kit with the PillCam Genius SB Capsule is intended for visualization of the small bowel mucosa. It may be used in the visualization and monitoring of lesions that may indicate Crohn's disease not detected by upper and lower endoscopy. It may be used in the visualization and monitoring of lesions that may be a source of obscure bleeding (either overt or occult) not detected by upper and lower endoscopy. It may be used in the visualization and monitoring of lesions that may be potential causes of iron deficiency anemia (IDA) not detected by upper and lower endoscopy. The PillCam Genius SB Capsule may be used as a tool in the detection of abnormalities of the small bowel and is intended for use in adults.
Device Story
System captures small bowel images via ingestible telemetric capsule; transmits data to single-use Genius Link Device adhered to patient abdomen. Genius Sync Agent software manages procedure setup and converts Link Device data for workstation processing. Optional Real Time View Application allows HCPs to view live images during procedure to estimate capsule location. HCPs review compiled images on workstation/cloud software to detect abnormalities. System aids diagnosis of Crohn's, obscure bleeding, and iron deficiency anemia. Used in clinical settings; operated by healthcare providers.
Clinical Evidence
No clinical testing was conducted in support of this 510(k) submission. Evidence is based on bench testing, including pH resistance, mechanical integrity, battery reliability, field of view, communication performance, adhesion testing, and usability testing. System complies with IEC 60601-1 and IEC 60601-1-2.
Technological Characteristics
Ingestible telemetric capsule imaging system. Components: PillCam Genius SB Capsule, single-use Genius Link Device (adhered to abdomen), Genius Sync Agent software, and workstation software. Connectivity: Wireless capsule-to-link communication; workstation/cloud integration. Software: PillCam Software v9.7. Biocompatibility per ISO 10993-1. Electrical safety per IEC 60601-1; EMC per IEC 60601-1-2.
Indications for Use
Indicated for visualization of small bowel mucosa in adults; monitoring lesions indicating Crohn's disease, obscure GI bleeding, or iron deficiency anemia not detected by upper/lower endoscopy.
Regulatory Classification
Identification
An ingestible telemetric gastrointestinal capsule imaging system is used for visualization of the small bowel mucosa as an adjunctive tool in the detection of abnormalities of the small bowel. The device captures images of the small bowel with a wireless camera contained in a capsule. This device includes an ingestible capsule (containing a light source, camera, transmitter, and battery), an antenna array, a receiving/recording unit, a data storage device, computer software to process the images, and accessories.
Special Controls
*Classification.* Class II (special controls). The special control is FDA's guidance, “Class II Special Controls Guidance Document: Ingestible Telemetric Gastrointestinal Capsule Imaging Systems; Final Guidance for Industry and FDA.”
Predicate Devices
PillCam SB 3 capsule endoscopy system with PillCam Software 9.0E (including PillCam Cloud Reader Software) (K230991)
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logos of the Department of Health & Human Services and the U.S. Food & Drug Administration (FDA). The Department of Health & Human Services logo is on the left, featuring a stylized human figure. To the right is the FDA logo, with the acronym "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG" in a larger, bolder font, and "ADMINISTRATION" in a smaller font below.
May 10, 2024
Given Imaging Ltd. (d.b.a. Medtronic) Aaron Niklaus Sr. Regulatory Affairs Specialist 2 Hacarmel St. New Industrial Park PO Box 258 Yoqneam Northern, 20692 Israel
Re: K240276
Trade/Device Name: PillCam™ Genius SB System with PillCam Software v9.7 and PillCam Cloud Reader Software Regulation Number: 21 CFR 876.1300 Regulation Name: Ingestible Telemetric Gastrointestinal Capsule Imaging System Regulatory Class: Class II Product Code: NEZ Dated: January 31, 2024 Received: January 31, 2024
Dear Aaron Niklaus:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}------------------------------------------------
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality System (OS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn).
{2}------------------------------------------------
Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/deviceadvice-comprehensive-regulatory-assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
# Sivakami Venkatachalam -S
for
Shanil P. Haugen, Ph.D. Assistant Director DHT3A: Division of Renal, Gastrointestinal, Obesity and Transplant Devices OHT3: Office of GastroRenal, ObGyn, General Hospital and Urology Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{3}------------------------------------------------
#### Indications for Use
Submission Number (if known)
K240276
Device Name
PillCam™ Genius SB System with PillCam Software v9.7 and PillCam Cloud Reader Software
#### Indications for Use (Describe)
The PillCam™ Genius SB Capsule Endoscopy Kit with the PillCam Genius SB Capsule is intended for visualization of the small bowel mucosa.
• It may be used in the visualization and monitoring of lesions that may indicate Crohn's disease not detected by upper and lower endoscopy.
· It may be used in the visualization and monitoring of lesions that may be a source of obscure bleeding (either overt or occult) not detected by upper and lower endoscopy.
• It may be used in the visualization and monitoring of lesions that may be potential causes of iron deficiency anemia (IDA) not detected by upper and lower endoscopy.
The PillCam Genius SB Capsule may be used as a tool in the detection of abnormalities of the small bowel and is intended for use in adults.
Type of Use (Select one or both, as applicable)
Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{4}------------------------------------------------
#### Given Imaging Ltd. (dba Medtronic Inc.)
#### PillCam™ Genius SB System with PillCam Software v9.7 and
#### PillCam Cloud Reader Software
Date Prepared: April 9, 2024
- Submitter: Given Imaging Ltd. (d.b.a. Medtronic) 2 Hacarmel St. New Industrial Park, PO Box 258 Yoqneam Northern, Israel 20692 Registration Number: 9710107 Phone: (763) 398-7010
- Contact Person: Aaron Niklaus Sr. Regulatory Affairs Specialist Medtronic Endoscopy Phone: 763-294-1241 Email: aaron.j.niklaus(@medtronic.com
- Alternate Contact: Monika McDole-Russell Sr. Regulatory Affairs Director Medtronic Endoscopy Phone: 763-294-1241 Email: monika.mcdole-russell(@medtronic.com
#### Name of Device:
PillCam Genius SB System with PillCam Software v9.7 and PillCam Cloud Reader Software
#### Common or Usual Name:
Ingestible telemetric gastrointestinal capsule imaging system
#### Classification:
System, Imaging, Gastrointestinal, Wireless, Capsule
#### Regulatory Class:
Class II, 21 CFR 876.1300
{5}------------------------------------------------
#### Product Code:
NEZ
#### Predicate Device:
PillCam SB 3 capsule endoscopy system with PillCam Software 9.0E (including PillCam Cloud Reader Software) (K230991)
# Device Description
The subject device introduces the PillCam Genius Capsule Endoscopy Kit (PillCam Genius SB Capsule and the PillCam Genius Link Device), the PillCam Genius Sync Agent, and the Real Time View Application.
The PillCam Genius Link Device is a single use device that is adhered to the patient's abdomen throughout the PillCam Genius SB procedure that connects to the PillCam Genius SB Capsule. Each Genius Link Device is paired and packaged with a single PillCam Genius SB Capsule. At the end of the procedure, the Link Device is removed and returned to the clinic.
The PillCam Genius Sync Agent is software used by the healthcare provider to setup and manage PillCam Genius SB endoscopy procedures and converts data from the PillCam Genius Link Device to the workstation.
The optional Real Time View Application allows the healthcare provider to view real time images from the capsule for a short period of time during the procedure to estimate the location of the capsule in the gastrointestinal tract.
The PillCam Desktop Software and HCP Software Application have been updated to display PillCam Genius SB Capsule procedures. The HCP Software Application has also been updated to be compatible with the PillCam Genius SB System and allow for uploading multiple procedures.
# Intended Use/Indications for Use
The Subject Device has the same intended use and indications for use as the Predicate Device except for the removal of pediatric indications and removal of the optional SBI feature for PillCam Genius SB Capsule procedures.
# Intended Use
The PillCam Genius SB Capsule Endoscopy Kit with the PillCam Genius SB Capsule is intended for visualization of the small bowel mucosa.
#### Indications for Use:
The PillCam Genius SB Capsule Endoscopy Kit with the PillCam Genius SB Capsule is intended for visualization of the small bowel mucosa.
- It may be used in the visualization and monitoring of lesions that may indicate ● Crohn's disease not detected by upper and lower endoscopy.
- It may be used in the visualization and monitoring of lesions that may be a source of obscure bleeding (either overt or occult) not detected by upper and lower endoscopy.
{6}------------------------------------------------
- It may be used in the visualization and monitoring of lesions that may be potential . causes of iron deficiency anemia (IDA) not detected by upper and lower endoscopy.
The PillCam Genius SB Capsule may be used as a tool in the detection of abnormalities of the small bowel and is intended for use in adults.
# Technological Characteristics
The Subject Device has similar technological characteristics when compared to the Predicate Device and any technological differences between the devices do not raise different questions of safety and effectiveness. The Subject Device has two main subsystems and one optional subsystem.
The introduction of new components replaces the functionality of some of the existing components, but the new components share similar technological characteristics.
The PillCam Genius Capsule is now sold with the PillCam Genius Link Device as the PillCam Genius SB Capsule Endoscopy Kit. The Genius Link Device has the receiving and recording functions of the Predicate Device's DR3 recorder. An optional Real Time View Application has been added to allow HCPs to view live images of the endoscopy procedure. The Genius Sync Agent converts capsule procedure data into a compiled format for the Desktop Software which had previously been done by the DR3 recorder. The Genius Sync Agent replaces the PillCam Sync Agent when used to synchronize between the PillCam Workstation and the PillCam Cloud Reader Software.
The Software Workstation and PillCam Cloud Reader Software retain the same functionality as the Predicate Device.
# Principles of Operation
Both the Subject Device and the Predicate Device have the same principles of operation, where the receiver and capsule are paired, the patient places a receiver on their abdomen, ingests the capsule, and real-time images of the small bowel are obtained, which are then reviewed by an HCP.
# Performance Testing
# Biocompatibility Testing
Biocompatibility testing was conducted on the PillCam Genius Link Device in accordance with Guidance for Industry and FDA Staff "Use of International Standard ISO 10993-1, "Biological evaluation of medical devices - Part 1: Evaluation and testing within a risk management process".
{7}------------------------------------------------
#### Bench Testing
Bench testing was conducted to demonstrate the safety and efficacy of the PillCam Genius Capsule Endoscopy Kit according to Guidance for Industry and FDA Staff "Ingestible Telemetric Gastrointestinal Capsule Imaging System - Final Class II Special Controls".
The PillCam Genius Capsule Endoscopy Kit passed the following tests:
- . PH resistance
- Mechanical integrity ●
- Batterv reliability
- Field of View (FOV) and Depth of Field ●
- Communication performance between capsule and recorder
- Adhesion on skin testing
- Environmental testing
- Usability testing ●
#### Animal Studies
No animal studies were conducted in support of this 510(k) submission.
# Clinical Testing
No clinical testing was conducted in support of this 510(k) submission.
# Electrical Safety and EMC
Electrical safety and EMC testing were performed on the PillCam Genius Capsule Endoscopy Kit. The system complies with IEC 60601-1 for safety and IEC 60601-1-2 for EMC.
# Software Documentation
Documentation was provided as recommended by FDA's Guidance for Industry and FDA Staff, "Content of Premarket Submissions for Device Software Functions". The software documentation level for the Subject Device is considered basic as a failure or latent flaw of the device software function(s) would not present a hazardous situation with a probable risk of death or serious injury to either a patient, user of the device, or others in the environment of use, prior to the implementation of risk control measures.
# Conclusion
The conclusions drawn from the nonclinical testing demonstrates that the Subject Device is as safe, as effective, and performs as well as or better than the legally marketed Predicate Device.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.