← Product Code [FXX](/productcode/FXX) · K240258

# Disposable Medical Face Mask (M643BE) (K240258)

_Makrite Industries, Inc. · FXX · Apr 18, 2024 · General, Plastic Surgery · SESE_

**Canonical URL:** https://fda.innolitics.com/device/K240258

## Device Facts

- **Applicant:** Makrite Industries, Inc.
- **Product Code:** [FXX](/productcode/FXX.md)
- **Decision Date:** Apr 18, 2024
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 878.4040
- **Device Class:** Class 2
- **Review Panel:** General, Plastic Surgery

## Indications for Use

The Disposable Medical Face Mask is intended to be worn to protect both the patient and healthcare personnel from transfer of microorganisms, body fluids and particulate material. The Disposable Medical Face Mask is intended for use in infection control practices to reduce the potential exposure to blood and body fluids. This a single use, disposable device(s), provided non-sterile.

## Device Story

Disposable, non-sterile, flat-folded medical face mask; three-layer construction (spun-bond polypropylene inner/outer layers, melt-blown polypropylene filter layer); features iron-core/polypropylene nose piece and nylon/spandex ear loops. Intended for single-use by healthcare personnel and patients to provide barrier protection against microorganisms, body fluids, and particulates. Used in clinical infection control settings. Device is worn over the nose and mouth, secured by ear loops and adjusted via the nose piece. Benefits include reduced potential exposure to blood and body fluids during medical procedures.

## Clinical Evidence

No clinical data. Bench testing only. Performance verified against ASTM F2100-23, ASTM F2101 (BFE ≥95%), EN 14683 (Delta P <5.0 mmH2O/cm²), ASTM F1862 (synthetic blood penetration at 80 mmHg), 16 CFR 1610 (Flammability Class 1), and ASTM F3502 (PFE ≥80%). Biocompatibility confirmed per ISO 10993-5 and ISO 10993-10.

## Technological Characteristics

Three-layer mask: spun-bond polypropylene (outer/inner), melt-blown polypropylene (filter). Nose piece: iron core coated with polypropylene. Ear loops: nylon and spandex. Dimensions: 175x95mm. Non-sterile, single-use. Complies with ASTM F2100-23. Biocompatible per ISO 10993.

## Regulatory Identification

Surgical apparel are devices that are intended to be worn by operating room personnel during surgical procedures to protect both the surgical patient and the operating room personnel from transfer of microorganisms, body fluids, and particulate material. Examples include surgical caps, hoods, masks, gowns, operating room shoes and shoe covers, and isolation masks and gowns. Surgical suits and dresses, commonly known as scrub suits, are excluded.

## Special Controls

*Classification.* (1) Class II (special controls) for surgical gowns and surgical masks. A surgical N95 respirator or N95 filtering facepiece respirator is not exempt if it is intended to prevent specific diseases or infections, or it is labeled or otherwise represented as filtering surgical smoke or plumes, filtering specific amounts of viruses or bacteria, reducing the amount of and/or killing viruses, bacteria, or fungi, or affecting allergenicity, or it contains coating technologies unrelated to filtration (e.g., to reduce and or kill microorganisms). Surgical N95 respirators and N95 filtering facepiece respirators are exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to § 878.9, and the following conditions for exemption:(i) The user contacting components of the device must be demonstrated to be biocompatible.
(ii) Analysis and nonclinical testing must:
(A) Characterize flammability and be demonstrated to be appropriate for the intended environment of use; and
(B) Demonstrate the ability of the device to resist penetration by fluids, such as blood and body fluids, at a velocity consistent with the intended use of the device.
(iii) NIOSH approved under its regulation.
(2) Class I (general controls) for surgical apparel other than surgical gowns and surgical masks. The class I device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to § 878.9.

## Predicate Devices

- Medical Disposable Face Masks ([K222809](/device/K222809.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.

April 18, 2024

Makrite Industries, Inc. % Ivy Wang Technical Manager Shanghai Sungo Management Consulting Company Limited 14th Floor, 1500# Century Avenue Shanghai. 200122 China

Re: K240258

Trade/Device Name: Disposable Medical Face Mask (M643BE) Regulation Number: 21 CFR 878.4040 Regulation Name: Surgical apparel Regulatory Class: Class II Product Code: FXX Dated: January 31, 2024 Received: January 31, 2024

Dear Ivy Wang:

We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device"

{1}------------------------------------------------

(https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).

Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30. Design controls; 21 CFR 820.90. Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review. the OS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely.

For Bifeng Qian, M.D., Ph.D. Assistant Director DHT4B: Division of Infection Control and Plastic Surgery Devices

{2}------------------------------------------------

OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{3}------------------------------------------------

# Indications for Use

Submission Number (if known)

K240258

Device Name

Disposable Medical Face Mask (M643BE)

Indications for Use (Describe)

The Disposable Medical Face Mask is intended to be worn to protect both the patient and healthcare personnel from transfer of microorganisms, body fluids and particulate material. The Disposable Medical Face Mask is intended for use in infection control practices to reduce the potential exposure to blood and body fluids. This a single use, disposable device(s), provided non-sterile.

Type of Use (Select one or both, as applicable)

Prescription Use (Part 21 CFR 801 Subpart D)

> Over-The-Counter Use (21 CFR 801 Subpart C)

## CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

## *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{4}------------------------------------------------

# 510(K) Summary

# K240258

(As required by 21 CFR 807.92)

Date prepared: April 9, 2024

#### A. Applicant:

Name: MAKRITE INDUSTRIES INC. Address: 11F-5., NO.79, Sec.1, Xintai 5th Rd., XiZhi Dist., New Taipei City 221, Taiwan Contact Person: Bob Wen Tel: +886-2-26982419 Fax: +886-2-26982423 Email: bobwen@makrite.com.tw

Submission Correspondent: Primary contact: Ms. Ivy Wang Shanghai SUNGO Management Consulting Co., Ltd. Room 1401, Dongfang Building, 1500# Century Ave., Shanghai 200122, China Tel: +86-21-58817802 Email: haiyu.wang@sungoglobal.com Secondary contact: Mr. Raymond Luo Room 1401, Dongfang Building, 1500# Century Ave., Shanghai 200122, China Tel: +86-21-68828050 Email: zxfda(@sungoglobal.com

#### B. Device:

Trade Name: Disposable Medical Face Mask Common Name: Disposable Medical Face Mask Model: M643BE

Regulatory Information Classification Name: Surgical Mask Classification: Class II Product code: FXX Regulation Number: 21 CFR 878.4040 Review Panel: Surgical Apparel

## C. Identification of Primary Predicate device: K22809

Anqing Mayfield Medical Limited. Trade/Device Name: Medical Disposable Face Masks

{5}------------------------------------------------

Regulatory Information Classification Name: Surgical Mask Classification: Class II Product code: FXX Regulation Regulation Number: 21 CFR 878.4040 Review Panel: Surgical Apparel

#### D. Indications for use:

The Disposable Medical Face Mask is intended to be worn to protect both the patient and healthcare personnel from transfer of microorganisms, body fluids and particulate material. The Disposable Medical Face Mask is intended for use in infection control practices to reduce the potential exposure to blood and body fluids. This a single use, disposable device(s), provided non-sterile.

#### E. Device Description:

The Disposable Medical Face Mask is blue color, single use, flat-folded masks with nose piece and ear loops. The blue colorant is polypropylene (PP) master batch.

The body of the mask is composed of three layers: the inner (3rd layer) and outer (1st layers are made of spun-bond polypropylene, the middle (2nd layer) is made of melt blown polypropylene. The nose piece is made of Iron core coated with polypropylene, ear loop is made of Nylon and Spandex.

Each mask contains ear loops to secure the mask over the user's face and mouth with a bendable nose piece to firmly fit over the nose. This device is not made with natural rubber latex.

The Disposable Medical Face Masks are sold and are intended to be single use, disposable devices.

The mask is designed and manufactured in accordance with ASTM F2100-23 Standard Specification for Performance of Materials Used in Medical Face Masks.

#### F. Comparison of technological characteristics with the predicate device

Provided below table 1 is a comparison of the proposed device with the predicate device.

| Table 1 Comparison of Proposed and Predicate Devices    |                              |                               |         |
|---------------------------------------------------------|------------------------------|-------------------------------|---------|
| Comparison between proposed device and predicate device |                              |                               |         |
| Comparison Items                                        | proposed device              | Predicate Device              | Result  |
| Device Name                                             | Disposable Medical Face Mask | Medical Disposable Face Masks | Similar |
| 510k Number                                             | K240258                      | K222809                       | ---     |
| Product Code                                            | FXX                          | FXX                           | Same    |
| Regulation Number                                       | 21 CFR 878.4040              | 21 CFR 878.4040               | Same    |
| Regulatory Class                                        | Class II                     | Class II                      | Same    |

|  |  | able 1 Comparison of Proposed and Predicate Devices |
|--|--|-----------------------------------------------------|
|--|--|-----------------------------------------------------|

{6}------------------------------------------------

|                                      | Indications for Use            | The Disposable Medical Face<br>Mask is intended to be worn to<br>protect both the patient and<br>healthcare personnel from<br>transfer of microorganisms,<br>body fluids and particulate<br>material. The Disposable<br>Medical Face Mask is intended<br>for use in infection control<br>practices to reduce the<br>potential exposure to blood<br>and body fluids. This a single<br>use, disposable device(s),<br>provided non-sterile. | The Medical Disposable Face<br>Mask is intended to be worn to<br>protect both the patient and<br>healthcare personnel from<br>transfer of microorganisms,<br>body fluids and particulate<br>material. The face mask is<br>intended for use in infection<br>control practices to reduce the<br>potential exposure to blood and<br>body fluids. This is a single<br>use, disposable device,<br>provided nonsterile. | Similar                                                                                                                  |  |
|--------------------------------------|--------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------|--|
| Size<br>specification<br>(Dimension) |                                | 175×95mm                                                                                                                                                                                                                                                                                                                                                                                                                                 | 17.5×9.5cm<br>14.5×9.5cm                                                                                                                                                                                                                                                                                                                                                                                          | Similar                                                                                                                  |  |
| Color                                |                                | Blue                                                                                                                                                                                                                                                                                                                                                                                                                                     | Blue                                                                                                                                                                                                                                                                                                                                                                                                              | Same                                                                                                                     |  |
| Design Feature                       |                                | Ear Loops, Flat-pleated, 3 layers                                                                                                                                                                                                                                                                                                                                                                                                        | Ear Loops, Flat-pleated, 3 layers                                                                                                                                                                                                                                                                                                                                                                                 | Same                                                                                                                     |  |
| Use                                  |                                | Single Use, Disposable                                                                                                                                                                                                                                                                                                                                                                                                                   | Single Use, Disposable                                                                                                                                                                                                                                                                                                                                                                                            | Same                                                                                                                     |  |
| Materials                            | Outer layer                    | Spun-bond polypropylene                                                                                                                                                                                                                                                                                                                                                                                                                  | Spunbond Polypropylene                                                                                                                                                                                                                                                                                                                                                                                            | Same                                                                                                                     |  |
|                                      | Inner layer                    | Spun-bond polypropylene                                                                                                                                                                                                                                                                                                                                                                                                                  | Spunbond Polypropylene                                                                                                                                                                                                                                                                                                                                                                                            | Same                                                                                                                     |  |
|                                      | Middle layer<br>(Filter layer) | Melt blown polypropylene filter                                                                                                                                                                                                                                                                                                                                                                                                          | Melt-blown Polypropylene                                                                                                                                                                                                                                                                                                                                                                                          | Same                                                                                                                     |  |
|                                      | Nose piece                     | Iron core coated with polypropylene                                                                                                                                                                                                                                                                                                                                                                                                      | Aluminum with Polypropylene<br>(PP) covering                                                                                                                                                                                                                                                                                                                                                                      | Different.<br>The proposed device<br>has different material<br>of nose piece and ear<br>loop to the predicate<br>device. |  |
|                                      | Ear loop                       | Nylon, Spandex                                                                                                                                                                                                                                                                                                                                                                                                                           | Polyester, Spandex                                                                                                                                                                                                                                                                                                                                                                                                |                                                                                                                          |  |
| OTC use                              |                                | Yes                                                                                                                                                                                                                                                                                                                                                                                                                                      | Yes                                                                                                                                                                                                                                                                                                                                                                                                               | Same                                                                                                                     |  |
| Sterility                            |                                | Non-Sterile                                                                                                                                                                                                                                                                                                                                                                                                                              | Non-Sterile                                                                                                                                                                                                                                                                                                                                                                                                       | Same                                                                                                                     |  |
|                                      | ASTM F2100 Level               | ASTM F2100-23 Level 1                                                                                                                                                                                                                                                                                                                                                                                                                    | ASTM F2100-19 Level 1                                                                                                                                                                                                                                                                                                                                                                                             | Similar                                                                                                                  |  |
| Fluid<br>Performance                 | Resistance                     | 32 out of 32 per lot pass at 80 mmHg 3 non-consecutive lots tested                                                                                                                                                                                                                                                                                                                                                                       | 32 out of 32 per lot pass at 80 mmHg 3 non-consecutive lots tested                                                                                                                                                                                                                                                                                                                                                | Same                                                                                                                     |  |
|                                      | ASTM F1862                     |                                                                                                                                                                                                                                                                                                                                                                                                                                          | tested                                                                                                                                                                                                                                                                                                                                                                                                            | tested                                                                                                                   |  |

{7}------------------------------------------------

| Particulate Filtration<br>Efficiency                     | ≥ 80%<br>ASTM F3502-2022a 8.1                                                                                | ≥ 95%<br>ASTM F2299                                                                                          | Similar. Both masks<br>meet the particulate<br>filtration efficiency<br>requirements for their<br>respective versions of<br>ASTM F2100 |
|----------------------------------------------------------|--------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------|
| Bacterial Filtration<br>Efficiency<br>ASTM F2101         | ≥ 95%                                                                                                        | ≥ 95%                                                                                                        | Same                                                                                                                                   |
| Differential<br>Pressure<br>(Delta P)<br>EN14683 Annex C | < 5.0mmH₂O/cm²                                                                                               | < 5.0mmH₂O/cm²                                                                                               | Same                                                                                                                                   |
| Flammability<br>CFR 1610                                 | Class 1                                                                                                      | Class 1                                                                                                      | Same                                                                                                                                   |
| Biocompatibility                                         | ISO 10993<br>Under the conditions of the<br>studies, non-cytotoxic, not an<br>irritant, and not a sensitizer | ISO 10993<br>Under the conditions of the<br>studies, non-cytotoxic, not an<br>irritant, and not a sensitizer | Same.                                                                                                                                  |

#### Difference Analysis:

The proposed device has same intended use and similar structure, parameters, and performance as the predicate device. The materials of inner layer and outer layer of the proposed device are also similar to those of the predicate device.

The proposed device has different materials for the nose piece and ear loops than the predicate device. However, the proposed device meets the requirements of ISO 10993. The results do not show any adverse effect. Thus, this difference will not affect the safety and effectiveness between the proposed device and the predicate device.

The Proposed device has different Sub-micron particulate filtration efficiency because the testing of proposed device has different standards and testing methods than the predicate device. The proposed device has been tested and the test results shown that the proposed device meets of Particulate Filtration Efficiency in ASTM F2100-23.

#### G. Summary of Non-Clinical Testing

- > Performance testing
Non-clinical tests were conducted to verify that the proposed device met all design specifications. The test results demonstrated that the proposed device complies with the following standards and the requirements stated in the Guidance for Industry and FDA Staff: Surgical Mask - Premarket Notification [510(K)] Submission issued on March 5, 2004:

· ASTM F2100, Standard Specification for Performance of Materials Used in Medical Face Masks.

· ASTM F2101, Standard Test Method for Evaluating the Bacterial Filtration Efficiency (BFE) Of Medical Face Mask Materials, Using A Biological Aerosol of Staphylococcus Aureus;

· EN 14683, Medical Face Masks- Requirements and Test Methods Annex C;

{8}------------------------------------------------

• ASTM F1862/F 1862M, Standard Test Method for Resistance of Medical Face Masks to Penetration by Synthetic Blood (Horizontal Projection of Fixed Volume at A Known Velocity)

- · 16 CFR 1610, Standard for the Flammability of clothing textiles;
· ASTM F3502, Standard Specification for Barrier Face Coverings.

Performance testing are summarized in below table 2.

| Test Methodology                                                              | Purpose                                                                                      | Acceptance Criteria<br>for Level 3 Barrier | Result                                                                                                                                                                                                           |
|-------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------|--------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Bacterial Filtration<br>Efficiency<br>ASTM F2101                              | Measure bacterial<br>filtration efficiency                                                   | ≥95%                                       | Passed<br>3 non-consecutive lots tested, using a<br>sample size of 32/lot<br>Lot 0923: ≥95%<br>Lot 1023: ≥95%<br>Lot 1123: ≥95%                                                                                  |
| Differential Pressure<br>(mmH2O/cm²)<br>EN 14683:2019<br>Annex C              | Determine<br>breathability of the<br>mask                                                    | <5.0 mmH2O/cm²                             | Passed<br>3 non-consecutive lots tested, using a<br>sample size of 32/lot<br>Lot 0923: <5.0<br>Lot 1023: <5.0<br>Lot 1123: <5.0                                                                                  |
| Sub-micron<br>Particulate Filtration<br>Efficiency<br>ASTM F3502-2022a<br>8.1 | Measure initial particle<br>filtration efficiency                                            | ≥80%                                       | Passed<br>3 non-consecutive lots tested, using a<br>sample size of 32/lot<br>Lot 0923: ≥80%<br>Lot 1023: ≥80%<br>Lot 1123: ≥80%                                                                                  |
| Resistance to<br>Penetration by<br>Synthetic Blood<br>ASTM F1862-17           | Evaluate the resistance<br>to penetration by<br>impact of small volume<br>of synthetic blood | 32 out of 32 pass at 80<br>mmHg            | Passed<br>3 non-consecutive lots tested, using a<br>sample size of 32/lot<br>Lot 0923: 32 out of 32 pass at 80<br>mmHg<br>Lot 1023: 32 out of 32 pass at 80<br>mmHg<br>Lot 1123: 32 out of 32 pass at 80<br>mmHg |
| Flammability<br>16 CFR Part<br>1610-2008                                      | Response of materials<br>to heat and flame                                                   | Class I                                    | Passed<br>3 non-consecutive lots tested, using a<br>sample size of 32/lot<br>Lot 0923: Class 1<br>Lot 1023: Class 1<br>Lot 1123: Class 1                                                                         |
| ISO 10993-5                                                                   | Evaluate potential for                                                                       | Under the conditions                       | Passed                                                                                                                                                                                                           |

|  | Table 2: Summary of Performance Testing |  |
|--|-----------------------------------------|--|

{9}------------------------------------------------

| Tests for in vitro<br>cytotoxicity                                | cytotoxicity                                 | of the study,<br>non-cytotoxic                            |        |
|-------------------------------------------------------------------|----------------------------------------------|-----------------------------------------------------------|--------|
| ISO 10993-10<br>Tests for irritation<br>and skin<br>sensitization | Evaluate potential for<br>irritation         | Under the conditions<br>of the study, not an<br>irritant  | Passed |
|                                                                   | Evaluate potential for<br>skin sensitization | Under the conditions<br>of the study, not a<br>sensitizer | Passed |

>> Biocompatibility of patient-contacting material

The biocompatibility of the subject device was determined via the evaluation according to the ISO 10993-5: 2009 and ISO 10993-10:2010 standards.

## H. Summary of Clinical Testing

No clinical study is included in this submission.

#### I. Conclusion

The conclusions drawn from the nonclinical tests demonstrate that the subject device in this 510(K) submission, the Disposable Medical Face Mask, is as safe, as effective, and performs as well as or better than the legally marketed predicate device cleared under K222809.

---

**Source:** [https://fda.innolitics.com/device/K240258](https://fda.innolitics.com/device/K240258)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
