The blue spectrum light (415nm) is intended to treat moderate inflammatory acne vulgaris. The red spectrum light (633nm) is generally indicated for treatment of superficial, benign, vascular and pigmented lesions. The near-infrared light (830nm) is generally indicated for the temporary relief of minor muscle and joint pain, arthritis and muscle spasm; relieving stiffness; promoting the relaxation of muscle tissue; and to temporarily increase local blood circulation where applied.
Device Story
Dermalux Tri-Wave MD MKII and Compact MD are LED-based phototherapy systems for aesthetic and dermatological treatment. Devices emit narrow-band light at 415nm (blue), 633nm (red), and 830nm (near-infrared). System comprises a main control unit with a capacitive touch screen and an articulated treatment head (3 or 4 panels) containing LEDs. Operated by clinicians in professional settings to deliver pre-set or customized protocols based on time or energy. Protective eyewear is required. Devices do not perform measurements or automated diagnostics; they function as therapeutic light sources to treat specific skin conditions and provide temporary pain relief. No internet connectivity; software updates via USB.
Clinical Evidence
No clinical data provided. Substantial equivalence is supported by non-clinical testing demonstrating that the devices are as safe and effective as the predicate.
Technological Characteristics
LED-based light therapy system. Wavelengths: 415nm, 633nm, 830nm. Power: 100-240Vac, 50/60Hz. Form factor: Floor-standing (4-panel) or desk-mounted (3-panel) with articulated arms. Interface: Capacitive touch screen. Connectivity: Standalone, no network/internet, USB for updates. Class II medical device.
Indications for Use
Indicated for patients with moderate inflammatory acne vulgaris (blue light), superficial benign vascular and pigmented lesions (red light), and minor muscle/joint pain, arthritis, muscle spasm, or stiffness requiring increased local blood circulation (near-infrared light).
Regulatory Classification
Identification
(1) A carbon dioxide laser for use in general surgery and in dermatology is a laser device intended to cut, destroy, or remove tissue by light energy emitted by carbon dioxide.(2) An argon laser for use in dermatology is a laser device intended to destroy or coagulate tissue by light energy emitted by argon.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health and Human Services logo on the left and the FDA acronym followed by the full name of the agency on the right. The FDA part of the logo is in blue, with the acronym in a square and the full name written out in a sans-serif font below it. The Department of Health and Human Services logo is a stylized human figure.
June 4, 2024
Aesthetic Technology Ltd Martin Hodgson Head of Regulaton Unit 211 Europa Boulevard Warrington, Cheshire WA5 7TN United Kingdom
Re: K240222
Trade/Device Name: Dermalux Tri-Wave MD MKII (1000021); Dermalux Compact MD (1000024) Regulation Number: 21 CFR 878.4810 Regulation Name: Laser Surgical Instrument For Use In General And Plastic Surgery And In Dermatology Regulatory Class: Class II Product Code: GEX, ILY Dated: April 23, 2024 Received: April 23, 2024
Dear Martin Hodgson:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrb/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}------------------------------------------------
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn).
{2}------------------------------------------------
Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/deviceadvice-comprehensive-regulatory-assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely.
Digitally signed by
Tanisha L. Hithe -S Tanisha L. Date: 2024.06.04 Hithe -S 17:25:56 -04'00'
Tanisha Hithe Assistant Director DHT4A: Division of General Surgery Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{3}------------------------------------------------
### Indications for Use
510(k) Number (if known) K240222
#### Device Name
1. Dermalux Tri-Wave MD MKII
2. Dermalux Compact MD
#### Indications for Use (Describe)
Indications for use include:
· The blue light (415nm wavelength) indicated to treat dermatological conditions and specifically indicated to treat moderate inflammatory acne vulgaris.
· The red light (633nm wavelength) indicated for treatment of superficial, benign, vascular and pigmented lesions.
• The near-infrared light (830nm wavelength) is generally indicated for the temporary relief of minor muscle and joint pain, arthritis and muscle spasm; relieving stiffness; promoting the relaxation of muscle tissue; and to temporarily increase local blood circulation where applied.
| Type of Use (Select one or both, as applicable) |
|-------------------------------------------------|
|-------------------------------------------------|
X Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{4}------------------------------------------------
## 510(k) Summary # K240222
Prepared on: 2024-05-23
| Contact Details | 21 CFR 807.92(a)(1) |
|-----------------------------|----------------------------------------------------------------------|
| Applicant Name | Aesthetic Technology Ltd |
| Applicant Address | Unit 211 Europa Boulevard Warrington Cheshire WA5 7TN United Kingdom |
| Applicant Contact Telephone | +44 (0) 1925 396721 |
| Applicant Contact | Mr. Martin Hodgson |
| Applicant Contact Email | martin.h@dermaluxled.com |
| Device Name | 21 CFR 807.92(a)(2) |
|---------------------|-------------------------------------------------------------------------------------|
| Device Trade Name | Dermalux Tri-Wave MD MKII (1000021); |
| | Dermalux Compact MD (1000024); |
| Common Name | Laser surgical instrument for use in general and plastic surgery and in dermatology |
| Classification Name | Powered Laser Surgical Instrument |
| Regulation Number | 878.4810 |
| Product Code | GEX, ILY |
| Legally Marketed Predicate Devices | 21 CFR 807.92(a)(3) |
|------------------------------------|---------------------|
|------------------------------------|---------------------|
| Predicate # | Predicate Trade Name (Primary Predicate is listed first) | Product |
|--------------|----------------------------------------------------------|---------|
| Code K200659 | Dermalux Tri-Wave MD | GEX |
# Device Description Summary
The Dermalux Tri-Wave MD MK II & Dermalux Compact MD is a Medical Device (MDR Class II) which emits specific wavelengths of low level, narrow band light for the treatment of certain aesthetic, dermatological and medical indications. The wavelengths used in the Tri-Wave MD MK II & Compact MD systems are Blue 415nm, Red 633nm and Near Infrared 830nm. The Dermalux professional use family consists of the Tri-Wave MD MKI; a floor standing unit with 4 pod treatment head, Compact MD; desk mounted unit with a 3 pod treatment head.
The system enables treatment of the body via an articulated 4 or 3 panel LED head. The panels can be easily adjusted for focused intensity treatments and application to larger body areas.
The system consists of a main body that contains the power supply (100-240Vac, 50/60Hz), power switch, touch screen control panel and a 4 or 3 panel adjustable treatment head that contains Light Emitting Diodes (LEDs).
The power supply is designed to power the internal electronics for the system and the Light Emitting Diodes. The power supply is connected to a suitable mains outlet. The device is switch located below the screen in the base unit.
The Tri-Wave MD MK II & Compact MD system is operated by a full colour capacitive touch screen which features pre-set and post treatment protocols that can be customised with three power settings. Bespoke protocols can be programmed based on delivered energy or time. The device cannot be networked or connect to the internet, software updates can be performed via a USB flash drive (thumb drive) if required.
## 21 CFR 807.92(a)(4)
{5}------------------------------------------------
The light is generated by Light Emitting Diodes (LED's) which a treatment head. The treatment head is connected to the main body by articulating arm mechanism. The arm allows the operator to correctly position the LED head ment area.
Protective eye wear is supplied with each Dermalux unit for patient and operator comfort.
The Dermalux systems are not used to make measurements of any sort, or to draw any conclusions regarding the indication to treat. The equipment does not require checks on the light output as the LEDs do not dim with age to any practical extent.
# Intended Use/Indications for Use
The blue spectrum light (415nm) is intended to treat moderate inflammatory acne vulgaris.
The red spectrum light (633nm) is generally indicated for treatment of superficial, benign, vascular and pigmented lesions.
The near-infrared light (830nm) is generally indicated for the temporary relief of minor muscle and ioint pain, arthritis and muscle spasm; relieving stiffness; promoting the relaxation of muscle tissue; and to temporarily increase local blood circulation where applied.
# Indications for Use Comparison
All devices described in this document have the same indications as those listed in the above section "Intended Use/ Indications for Use" and therefore Aesthetic Technology Ltd consider the Dermalux Tri-Wave MD MKI, and the Dermalux Compact MD to be equivalent to the Dermalux Tri-Wave MD (K200659) for purposes of Section 510(k) of the Federal Food, drug and Cosmetic Act.
# Technological Comparison
Each of the Dermalux devices (Dermalux Tri-Wave MD MKI) & Dernalux Compact MD) whether three panels have the same technological characteristicans for use as the Dermalux Tri-Wave MD placed in commercial distribution by Aesthetic Technology Ltd. on May 9, 2020 pursuant to 510(k) premarket notification K200659 and are therefore considered to be substantially equivalent to the Dermalux Tri-Wave MD.
The use of the blue, red and NIR regions of the spectrum are to emit energy to treat the following conditions:
Blue light (415nm wavelength): Treatment of dermatological conditions and specifically indicated to treat moderate inflammatory acne vulgaris.
Red light (633nm wavelength): Treatment of superficial, benign, vascular and pigmented lesions.
NIR (830m wavelength):Temporary relief of minor muscle and joint pain, arthritis and muscle spasm; relieving stiffness; promoting the relaxation of muscle tissue; and to temporarily increase local blood circulation where applied
# Non-Clinical and/or Clinical Tests Summary & Conclusions 21 CFR 807.92(b)
The conclusions drawn from the noncinical tests that demonstrate that the device is as safe, as effective, and performs as well as or better than the legally marketed device identified in paragraph (a)(3) of this section.
21 CFR 807.92(a)(5)
## 21 CFR 807.92(a)(5)
## 21 CFR 807.92(a)(6)
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Report an issue or suggest a feature
Spot a broken citation, a wrong extraction, a missing document, or something you wish this app did? Tell us — no account needed.
or drag and drop files here
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.