K240206 · GE Medical Systems Ultrasound and Primary Care Diagnostics · IYN · Jun 10, 2024 · Radiology
Device Facts
Record ID
K240206
Device Name
Venue Sprint
Applicant
GE Medical Systems Ultrasound and Primary Care Diagnostics
Product Code
IYN · Radiology
Decision Date
Jun 10, 2024
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 892.1550
Device Class
Class 2
Attributes
Pediatric
AI Performance
Output
Algorithm
Acceptance
Observed
Dev DS
Dev Readers
Test DS
Test Readers
Bladder Volume Measurement
—
—
—
—
—
—
—
Inferior Vena Cava (IVC) Diameter Measurement
—
—
—
—
—
—
—
Lung B-Lines Detection
—
—
—
—
—
—
—
Ejection Fraction
—
—
—
—
—
—
—
Nerve Identification
—
—
—
—
—
—
—
Indications for Use
Venue Sprint is a general-purpose diagnostic ultrasound system for use by qualified and trained healthcare professionals or practitioners that are legally authorized or licensed by law in the country, state or other local municipality in which he or she practices, for ultrasound imaging, measurement, display and analysis of the human body and fluid. The users may or may not be working under supervision or authority of a physician. Users may also include medical students working under the supervision or authority of a physician during their education / training. The Venue Sprint is intended to be used in a hospital, medical clinic, home environment and road/air ambulance. Venue Sprint clinical applications include: abdominal (GYN and Urology), thoracic/pleural, ophthalmic, Fetal/OB, Small Organ (including breast, testes, thyroid), Vascular/Peripheral vascular, neonatal and adult cephalic, pediatric, musculoskeletal (conventional and superficial), cardiac (adults and pediatric, 40 kg and above) and interventional guidance (includes free hand tissue biopsy, fluid drainage, vascular and non-vascular access). Modes of operation include: B, M, PW Doppler, Color Doppler and Harmonic Imaging.
Device Story
Venue Sprint is a general-purpose diagnostic ultrasound system; utilizes Vscan Air CL and SL wireless probes for data acquisition; provides user interface for probe control, image analysis, and storage. System software runs on Windows-based off-the-shelf (OTS) tablet; can be mounted on mobile cart with wireless probe charging. Operated by qualified healthcare professionals or supervised medical students in hospitals, clinics, home, or emergency transport. Features include Bladder Volume Tool, MSK diagrams (knee, wrist/hand, hip, elbow, ankle/foot), and Venue Coach for reference images. Software updates performed by users or service personnel via GEHC website. Output consists of ultrasound images/videos for clinical evaluation; assists in interventional guidance and diagnostic assessment. Benefits include portable, versatile imaging across diverse clinical settings.
Clinical Evidence
No clinical studies were required to support substantial equivalence. Evidence is based on equipment design similarities, conformance to recognized performance standards (acoustic output, biocompatibility, electrical/mechanical safety, EMC), and system-level verification and validation testing.
Technological Characteristics
Diagnostic ultrasound system using wireless Vscan Air CL/SL probes. Software runs on Windows-based OTS tablet. Connectivity via wireless interface; supports DICOM. Complies with AAMI/ANSI ES60601-1, IEC 60601-1-2, IEC 60601-2-37, IEC 62359, ISO 10993-1, and ISO 14971. Biocompatible materials. Modes: B, M, PW Doppler, Color Doppler, Harmonic Imaging.
Indications for Use
Indicated for qualified healthcare professionals/practitioners for ultrasound imaging, measurement, display, and analysis of human body and fluid. Applications: abdominal (GYN/Urology), thoracic/pleural, ophthalmic, Fetal/OB, Small Organ (breast, testes, thyroid), Vascular/Peripheral, neonatal/adult cephalic, pediatric, musculoskeletal, cardiac (adult/pediatric ≥40kg), and interventional guidance (biopsy, drainage, vascular/non-vascular access). Used in hospitals, clinics, home, and road/air ambulances.
Regulatory Classification
Identification
An ultrasonic pulsed doppler imaging system is a device that combines the features of continuous wave doppler-effect technology with pulsed-echo effect technology and is intended to determine stationary body tissue characteristics, such as depth or location of tissue interfaces or dynamic tissue characteristics such as velocity of blood or tissue motion. This generic type of device may include signal analysis and display equipment, patient and equipment supports, component parts, and accessories.
{1}------------------------------------------------
June 10, 2024
Image /page/0/Picture/1 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left, there is a seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES-USA" surrounding a graphic. To the right, the FDA logo is displayed in blue, with the acronym "FDA" above the words "U.S. FOOD & DRUG" and "ADMINISTRATION" below.
GE Medical Systems Ultrasound and Primary Care Diagnostics % Karin Shimoni Regulatory Affairs Manager 9900 Innovation Drive WAUWATOSA WI 53226
Re: K240206
Trade/Device Name: Venue Sprint Regulation Number: 21 CFR 892.1550 Regulation Name: Ultrasonic Pulsed Doppler Imaging System Regulatory Class: Class II Product Code: IYN, IYO, ITX Dated: January 25, 2024 Received: May 9, 2024
Dear Karin Shimoni:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
{2}------------------------------------------------
Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
# Yanna S. Kang -S
Yanna Kang, Ph.D. Assistant Director Mammography and Ultrasound Team DHT8C: Division of Radiological Imaging and Radiation Therapy Devices OHT8: Office of Radiological Health Office of Product Evaluation and Quality Center for Devices and Radiological Health
{3}------------------------------------------------
## Indications for Use
Submission Number (if known)
K240206
Device Name
Venue Sprint
Indications for Use (Describe)
Venue Sprint is a general-purpose diagnostic ultrasound system for use by qualified and trained healthcare professionals or practitioners that are legally authorized or licensed by law in the country. state or other local municipality in which he or she practices, for ultrasound imaging, measurement. display and analysis of the human body and fluid. The users may or may not be working under supervision or authority of a physician. Users may also include medical students working under the supervision or authority of a physician during their education / training.
The Venue Sprint is intended to be used in a hospital, medical clinic, home environment and road/air ambulance.
Venue Sprint clinical applications include: abdominal (GYN and Urology), thoracic/pleural, ophthalmic, Fetal/OB, Small Organ (including breast, testes, thyroid), Vascular/Peripheral vascular, neonatal and adult cephalic, pediatric, musculoskeletal (conventional and superficial), cardiac (adults and pediatric, 40 kg and above) and interventional guidance (includes free hand tissue biopsy, fluid drainage, vascular access). Modes of operation include: B. M. PW Doppler, Color Doppler and Harmonic Imaging.
| Type of Use (Select one or both, as applicable) | |
|-------------------------------------------------|--|
|-------------------------------------------------|--|
> Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{4}------------------------------------------------
Image /page/3/Picture/1 description: The image shows the GE HealthCare logo. The logo consists of a circular emblem on the left and the text "GE HealthCare" on the right. The emblem is a stylized, circular design with swirling patterns inside. The text is in a sans-serif font and is colored in the same shade of purple as the emblem.
GE HealthCare 510(k) Premarket Notification Submission
## 510(k) Summary - K240206
| | In accordance with 21 CFR 807.92 the following summary of information is provided: |
|--------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Date:<br>Submitter: | January 25, 2024<br>GE Medical Systems Ultrasound and Primary Care Diagnostics<br>9900 Innovation Drive<br>Wauwatosa, WI 53226 |
| Primary Contact Person: | Karin Shimoni<br>Regulatory Affairs Manager<br>GE HealthCare<br>T: (+972) 546347710 |
| Secondary Contact Person: | Lee Bush<br>Regulatory Affairs Director<br>GE HealthCare<br>T: (262) 3099429 |
| Device Trade Name:<br>Common/Usual Name:<br>Classification Names:<br>Product Code: | Venue Sprint<br>Diagnostic Ultrasound System<br>Class II<br>Ultrasonic Pulsed Doppler Imaging System. 21CFR 892.1550 90-<br>IYN; Ultrasonic Pulsed Echo Imaging System, 21CFR 892.1560,<br>90-IYO; Diagnostic Ultrasound Transducer, 21 CFR 892.1570,<br>90-ITX |
| Primary Predicate Device:<br>Classification Names:<br>Product Code: | K220848 Venue Fit, Diagnostic Ultrasound System<br>Class II<br>Ultrasonic Pulsed Doppler Imaging System. 21CFR 892.1550 90-<br>IYN; Ultrasonic Pulsed Echo Imaging System, 21CFR 892.1560,<br>90-IYO; Diagnostic Ultrasound Transducer, 21 CFR 892.1570,<br>90-ITX |
| Reference Device(s):<br>Classification Names:<br>Product Code: | K231301 Vscan Air<br>Class II<br>Ultrasonic Pulsed Doppler Imaging System. 21CFR 892.1550 90-<br>IYN; Ultrasonic Pulsed Echo Imaging System, 21CFR 892.1560,<br>90-IYO; Diagnostic Ultrasound Transducer, 21 CFR 892.1570,<br>90-ITX |
| <b>Reference Device(s):</b> | K170714 Venue |
| <b>Classification Names:</b> | Class II |
| <b>Product Code:</b> | Ultrasonic Pulsed Doppler Imaging System. 21CFR 892.1550 90-<br>IYN; Ultrasonic Pulsed Echo Imaging System, 21CFR 892.1560,<br>90-IYO; Diagnostic Ultrasound Transducer, 21 CFR 892.1570,<br>90-ITX |
| <b>Reference Device(s):</b> | K202035 Vscan Air |
| <b>Classification Names:</b> | Class II |
| <b>Product Code:</b> | Ultrasonic Pulsed Doppler Imaging System. 21CFR 892.1550 90-<br>IYN; Ultrasonic Pulsed Echo Imaging System, 21CFR 892.1560,<br>90-IYO; Diagnostic Ultrasound Transducer, 21 CFR 892.1570,<br>90-ITX |
| <b>Reference Device(s):</b> | K203137 Venue Fit |
| <b>Classification Names:</b> | Class II |
| <b>Product Code:</b> | Ultrasonic Pulsed Doppler Imaging System. 21CFR 892.1550 90-<br>IYN; Ultrasonic Pulsed Echo Imaging System, 21CFR 892.1560,<br>90-IYO; Diagnostic Ultrasound Transducer, 21 CFR 892.1570,<br>90-ITX |
| <b>Reference Device(s):</b> | K200851 Vivid T8, Vivid T9 |
| <b>Classification Names:</b> | Class II |
| <b>Product Code:</b> | Ultrasonic Pulsed Doppler Imaging System. 21CFR 892.1550 90-<br>IYN; Ultrasonic Pulsed Echo Imaging System, 21CFR 892.1560,<br>90-IYO; Diagnostic Ultrasound Transducer, 21 CFR 892.1570,<br>90-ITX |
| <b>Device Description:</b> | Venue Sprint is a general-purpose diagnostic ultrasound system<br>intended for use by qualified and trained healthcare professionals<br>to evaluate the body by ultrasound imaging and fluid flow<br>analysis.<br>The Venue Sprint is used together with the Vscan Air probes and<br>provides the user interface for control of the probes and the<br>needed software functionality for analysis of the ultrasound<br>images and saving/storage of the related images and videos. |
| <b>Intended Use:</b> | Venue Sprint is a general-purpose diagnostic ultrasound system<br>for use by qualified and trained healthcare professionals or<br>practitioners that are legally authorized or licensed by law in the<br>country, state or other local municipality in which he or she<br>practices, for ultrasound imaging, measurement, display and<br>analysis of the human body and fluid. The users may or may not |
{5}------------------------------------------------
Image /page/4/Picture/0 description: The image contains the logo for GE HealthCare. The logo consists of the GE monogram enclosed in a circular border on the left, followed by the text "GE HealthCare" on the right. The logo is rendered in a purple color.
# GE HealthCare 510(k) Premarket Notification Submission
{6}------------------------------------------------
Image /page/5/Picture/1 description: The image shows the GE HealthCare logo. The logo consists of a purple circular emblem with a stylized "GE" inside. To the right of the emblem, the words "GE HealthCare" are written in purple, with "GE" being slightly larger than "HealthCare".
# GE HealthCare 510(k) Premarket Notification Submission
be working under supervision or authority of a physician. Users may also include medical students working under the supervision or authority of a physician during their education / training. The Venue Sprint is intended to be used in a hospital, medical clinic, home environment and road/air ambulance. Venue Sprint clinical applications include: abdominal (GYN and Urology), thoracic/pleural, ophthalmic, Fetal/OB, Small Organ (including breast, testes, thyroid), Vascular/Peripheral vascular, neonatal and adult cephalic, pediatric, musculoskeletal (conventional and superficial), cardiac (adults and pediatric, 40 kg and above) and interventional guidance (includes free hand tissue biopsy, fluid drainage, vascular and non-vascular access). Modes of operation include: B, M, PW Doppler, Color Doppler and Harmonic Imaging.
The Venue Sprint employs the same fundamental scientific Re. Technology: technology as its predicate and reference devices.
Determination of Comparison to Predicate Device
Substantial Equivalence: The Venue Sprint system is substantially equivalent to the predicate device with regards to intended use, imaging capabilities, technological characteristics and safety and effectiveness. All probes used with the proposed Venue Sprint system are used unchanged from the cleared predicate. They are made of the same materials and their shape is unchanged.
> The following is an overview of the differences between the proposed Venue Sprint and the predicate Venue Fit:
Indications for Use:
- The proposed Venue Sprint and predicate Venue Fit ● (K220848) have similar clinical indications for use, however the IFU statement is updated to be consistent with the wording cleared for predicate Vscan Air K231301:
- Add details to the operator qualification/profile O
- Add "free hand", as the method for tissue biopsy O
- Add "40 kg and above" for cardiac (pediatric) O application
- The proposed Venue Sprint and predicate Venue Fit (K220848) have similar use-environments, however Venue Sprint is to be used also in home environment and road/air ambulance. No impact to safe or effective use. This is consistent with the use-environments of the cleared predicate Vscan Air K231301
{7}------------------------------------------------
Image /page/6/Picture/1 description: The image shows the GE HealthCare logo. The logo consists of a purple circular emblem with a stylized "GE" monogram on the left. To the right of the emblem, the words "GE HealthCare" are written in a simple, sans-serif font, also in purple. The logo is clean and modern, reflecting the company's focus on healthcare technology and innovation.
## GE HealthCare
510(k) Premarket Notification Submission
- . The proposed Venue Sprint and predicate Venue Fit (K220848) have similar imaging modes.
#### Transducers:
Vscan Air CL and SL probes on proposed Venue Sprint: The Vscan Air CL was first cleared in K202035 and the Vscan Air SL was first cleared in K231301. The clinical indications of the Vscan Air CL and SL are identical as on the reference device Vscan Air K231301.
#### Features/Functionality:
- Vscan Air CL and Vscan Air SL probes: The proposed ● Venue Sprint system supports the Vscan Air CL and Vscan Air SL cleared in K231301.
- . Bladder Volume Tool: Bladder Volume Tool provides a simplified workflow for bladder volume measurement.
- . MSK diagrams The MSK Diagram enables documenting and summarizing MSK exam in an anatomical view. On predicate Venue Fit K220848, the only diagram that was available for MSK preset was the shoulder diagram. On the proposed Venue Sprint, diagrams are also added for the knee, wrist/hand, hip, elbow and ankle/foot.
- Venue Coach: Venue Coach contains all the diagrams and ● reference images available in the proposed Venue Sprint system. From the Venue Coach tab, user can configure a diagram, a reference image or both.
- Electronic delivery of SW: The proposed Venue Sprint . system allows the user, in addition to service personnel, to update the SW by logging into a GEHC website to download SW available to them and install it on the system.
- Expanding cleared Auto Inferior Vena Cava (IVC), Auto ● B-Lines, Real Time ejection fraction (RT-EF) and cNerve for Vscan Air CL and SL probes. These probes are already cleared on predicate Vscan Air K231301. The Vscan Air CL probe is cleared also on reference device LOGIQ E10s/LOGIQ Fortis K231989.
- AppAPI functionality: to facilitate future integration of ● multiple software components.
#### Hardware:
The Venue Sprint software can be installed on a Windows-based off-the-shelf (OTS) consumer display device (tablet). Similar to predicate Venue Fit K220848, the tablet can be attached to a
{8}------------------------------------------------
Image /page/7/Picture/1 description: The image shows the GE HealthCare logo. The logo consists of a purple circular emblem with the letters "GE" inside, followed by the text "GE HealthCare" in purple. The text is written in a clean, sans-serif font.
# GE HealthCare
510(k) Premarket Notification Submission
mobile cart. The cart of proposed Venue Sprint provides power for the wireless charger of Vscan Air CL and SL probes.
#### Summary of Non-Clinical Tests:
The proposed Venue Sprint has been evaluated for acoustic output, biocompatibility, cleaning and disinfection effectiveness as well as thermal, electrical, electromagnetic, and mechanical safety, and has been found to comply with applicable medical device safety standards. The Venue Sprint complies with voluntary standards:
- Marketing Clearance of Diagnostic Ultrasound Systems ● and Transducers - Guidance for Industry and Food and Drug Administration Staff, issued on February 21, 2023
- AAMI/ANSI ES60601-1, Medical Electrical Equipment -● Part 1: General Requirements for Safety, 2005/ A2:2021
- IEC 60601-1-2, Medical Electrical Equipment Part 1-2: ● General Requirements for Safety - Collateral Standard: Electromagnetic Compatibility Requirements and Tests, 2020
- IEC 60601-2-37, Medical Electrical Equipment Part 2-● 37: Particular Requirements for the Safety of Ultrasonic Medical Diagnostic and Monitoring Equipment, 2015
- IEC 62359, Ultrasonics Field characterization Test ● methods for the determination of thermal and mechanical indices related to medical diagnostic ultrasonic fields, 2017
- . ISO 10993-1, Biological Evaluation of Medical Devices-Part 1: Evaluation and Testing Within a Risk Management Process, Fifth edition, 2018
- ISO 14971, Application of risk management to medical ● devices. 2019
- NEMA PS 3.1 3.20e, Digital Imaging and ● Communications in Medicine (DICOM) Set, 2021
- AAMI TIR69, Technical Information Report Risk ● management of radio-frequency wireless coexistence for medical devices and systems, 2020
- IEC 60601-1-11, Medical Electrical Equipment Part 1-● 11: General Requirements For Basic Safety And Essential Performance - Collateral Standard: Requirements For Medical Electrical Equipment And Medical Electrical Systems Used In The Home Healthcare Environment, 2020
{9}------------------------------------------------
Image /page/8/Picture/1 description: The image contains the GE HealthCare logo. The logo consists of a purple circular emblem with a stylized "GE" inside. To the right of the emblem, the words "GE HealthCare" are written in purple sans-serif font. The text is aligned horizontally with the emblem.
## GE HealthCare
510(k) Premarket Notification Submission
- IEC 60601-1-12, Medical Electrical Equipment Part 1-. 12: General Requirements For Basic Safety And Essential Performance - Collateral Standard: Requirements For Medical Electrical Equipment And Medical Electrical Systems Intended For Use In The Emergency Medical Services Environment, 2020
The following quality assurance measures are applied to the development of the system:
- Risk Analysis
- Requirements Reviews ●
- Design Reviews
- Testing on unit level (Module verification) ●
- Integration testing (System verification) ●
- Performance testing (Verification & Validation) ●
- Safety testing (Verification) ●
Transducer material and are biocompatible.
#### Summary of Clinical Tests:
The subject of this premarket submission, Venue Sprint, did not require clinical studies to support substantial equivalence.
- Conclusion: Based on the equipment design similarities, conformance to recognized performance standards, and performance testing, GE HealthCare considers the proposed Venue Sprint to be as safe, Effective, and performs in a substantially equivalent manner as the predicate Venue Fit (K220848).
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.