IntraOp Alignment System

K240199 · Alphatec Spine, Inc. · OWB · May 22, 2024 · Radiology

Device Facts

Record IDK240199
Device NameIntraOp Alignment System
ApplicantAlphatec Spine, Inc.
Product CodeOWB · Radiology
Decision DateMay 22, 2024
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 892.1650
Device ClassClass 2

Indications for Use

The IntraOp Alignment System is intended for use in applications where a mobile C-arm fluoroscope is incorporated to aid in diagnosis and treatment during spinal surgery. The IntraOp Alignment System is also intended to assist healthcare professionals in viewing, storing and measuring spinal alignment assessment images at various time points during surgery as well as planning spinal surgical procedures. The IntraOp Alignment System is intended for use with patients aged 18 years and older.

Device Story

IntraOp Alignment System (IOA) is a software-based system for intraoperative spinal surgery; interfaces with mobile C-arm fluoroscope via video cable/frame grabber. Uses optical tracking sensor and reflective markers to track 6 DOF C-arm position/orientation in real-time. Transforms inputs by stitching multiple fluoroscopic images into long bi-planar radiographic images; performs distortion removal using C-arm calibration data. User identifies spine location in bi-planar views for scaling/depth correction. Provides measurement tools for spinal alignment assessment and surgical planning. Optional cloud connectivity for importing/exporting pre-operative/post-operative alignment parameters. Used in hospital operating rooms by trained healthcare professionals. Output displayed on monitor to assist intraoperative visualization, assessment, and surgical decision-making; benefits patient through improved alignment planning.

Clinical Evidence

No clinical data. Substantial equivalence based on non-clinical performance testing, including verification of system accuracy, workflow, image stitching, C-arm tracking, spinal alignment measurements, and EMC/electrical safety testing.

Technological Characteristics

System includes optical tracking sensor, touchscreen computer, mobile cart, extension arm, and electronics box with frame grabber. Connectivity via Wi-Fi for cloud database access. Software-based image processing and stitching. Passive display system; does not transmit signals to fluoroscope. Optical tracking principle.

Indications for Use

Indicated for patients aged 18 years and older undergoing spinal surgery where a mobile C-arm fluoroscope is used for diagnosis and treatment. Assists healthcare professionals in viewing, storing, and measuring spinal alignment images and planning surgical procedures.

Regulatory Classification

Identification

An image-intensified fluoroscopic x-ray system is a device intended to visualize anatomical structures by converting a pattern of x-radiation into a visible image through electronic amplification. This generic type of device may include signal analysis and display equipment, patient and equipment supports, component parts, and accessories.

Special Controls

*Classification.* Class II (special controls). An anthrogram tray or radiology dental tray intended for use with an image-intensified fluoroscopic x-ray system only is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 892.9. In addition, when intended as an accessory to the device described in paragraph (a) of this section, the fluoroscopic compression device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 892.9.

Predicate Devices

Reference Devices

Submission Summary (Full Text)

{0}------------------------------------------------ May 22, 2024 Image /page/0/Picture/1 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left, there is a circular seal with an eagle emblem, representing the Department of Health & Human Services-USA. To the right, there is the FDA logo in blue, with the words "U.S. FOOD & DRUG" stacked on top of "ADMINISTRATION." Alphatec Spine, Inc. % Unnati Bhuptani Sr. Regulatory Affairs Specialist 1950 Camino Vida Roble CARLSBAD, CA 92008 Re: K240199 Trade/Device Name: IntraOp Alignment System Regulation Number: 21 CFR 892.1650 Regulation Name: Image-Intensified Fluoroscopic X-Ray System Regulatory Class: Class II Product Code: OWB, JAA, LLZ Dated: January 24, 2024 Received: April 22, 2024 Dear Unnati Bhuptani: We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download). {1}------------------------------------------------ Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181). Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems. For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100). Sincerely, Lu Jiang Lu Jiang, Ph.D. Assistant Director Diagnostic X-Ray Systems Team DHT8B: Division of Radiologic Imaging Devices and Electronic Products OHT8: Office of Radiological Health Office of Product Evaluation and Quality Center for Devices and Radiological Health {2}------------------------------------------------ ## Indications for Use 510(k) Number (if known) K240199 Device Name IntraOp Alignment System ### Indications for Use (Describe) The IntraOp Alignment System is intended for use in applications where a mobile C-arm fluoroscope is incorporated to aid in diagnosis and treatment during spinal surgery. The IntraOp Alignment System is also intended to assist healthcare professionals in viewing, storing and measuring spinal alignment assessment images at various time points during surgery as well as planning spinal surgical procedures. The IntraOp Alignment System is intended for use with patients aged 18 years and older. | Type of Use (Select one or both, as applicable) | | |-----------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------| | <div> <span style="text-decoration: overline;">Prescription Use (Part 21 CFR 801 Subpart D)</span> </div> | <div> <span>Over-The-Counter Use (21 CFR 801 Subpart C)</span> </div> | ### CONTINUE ON A SEPARATE PAGE IF NEEDED. This section applies only to requirements of the Paperwork Reduction Act of 1995. ### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: > Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." {3}------------------------------------------------ Image /page/3/Picture/0 description: The image shows the word "atec" in a stylized font. The "a" is green, and the rest of the letters are dark blue. There is a green swoosh above the letters "tec". The letters are all lowercase. # K240199 - 510(k) Summary This 510(k) summary of safety and effectiveness is being submitted in accordance with the requirements of 21 CFR 807.92. | A. | SUBMITTER: | Alphatec Spine, Inc. | |----|------------------------|-----------------------------------| | | | 1950 Camino Vida Roble | | | | Carlsbad, CA 92008 | | | | Phone: (760) 432-9286 | | | | Fax: (760) 431-0289 | | | Contact Person: | Unnati Bhuptani | | | | Sr. Regulatory Affairs Specialist | | | | Alphatec Spine, Inc. | | | | Contact Phone: (760) 356-6711 | | | Date Summary Prepared: | January 24, 2024 | #### B. DEVICE | Trade Name: | IntraOp Alignment System | |-----------------------|-------------------------------------------------------------------| | Common or Usual Name: | Interventional Fluoroscopic X-ray System | | Classification Name: | Image-Intensified Fluoroscopic X-ray System<br>(21 CFR §892.1650) | | Regulatory Class: | Class II | | Product Code: | OWB | | Subsequent Codes: | JAA. LLZ | #### C. LEGALLY MARKETED PREDICATE DEVICES | Primary Predicate | | |------------------------|-------------------------------------------------------------------| | Trade Name: | Nuvasive® LessRay® with Enhanced Tracking | | 510(k) number: | K170800 | | 510(k) Clearance Date: | August 18, 2017 | | Common or Usual Name: | Interventional Fluoroscopic X-ray System | | Classification Name: | Image-Intensified Fluoroscopic X-ray System<br>(21 CFR §892.1650) | | Regulatory Class: | Class II | | Product Code: | OWB | | Subsequent Codes: | JAA, LLZ | | Additional Predicate | | | Trade Name: | ATEC Alignment App | | 510(k) number: | K211987 | | 510(k) Clearance Date: | October 29, 2021 | | Common or Usual Name: | System, Image Processing, Radiological | {4}------------------------------------------------ Image /page/4/Picture/0 description: The image shows the logo for "atec". The "a" is green, while the "tec" is dark blue. There is a trademark symbol to the right of the "c". | Classification Name: | Medical image management and processing system<br>(21 CFR §892.2050) | |------------------------|----------------------------------------------------------------------| | Regulatory Class: | Class II | | Product Code: | LLZ | | Reference Devices: | | | Trade Name: | NuVasive® NuvaLine® | | 510(k) number: | K192435 | | 510(k) Clearance Date: | September 5, 2019 | | Common or Usual Name: | System, Image Processing, Radiological | | Classification Name: | Medical image management and processing system<br>(21 CFR §892.2050) | | Regulatory Class: | Class II | | Product Code: | LLZ | #### D. DEVICE DESCRIPTION The subject IntraOp Alignment (IOA) System consists of an optical tracking sensor, touchscreen computer, a mobile cart (stand) including an extension arm, and an electronics box with frame grabber, ports for video cables, power cables and external storage devices. The subject IntraOp Alignment System is a software-based system that from one end interfaces with a mobile C-arm fluoroscope through a video cable and a frame capture device to receive fluoroscopic images as they are acquired, and from the other end connects to an optical position tracking sensor to track the position and orientation of a C-arm in real-time. By using subject IntraOp Alignment System along the side of mobile C-arm fluoroscopy equipment, the user of the system can combine multiple fluoroscopic images, as they are being acquired, into stitched long bi-planar radiographic images for intraoperative visualization and assessment. The user is required to identify the location of the spine in both bi-planar views to scale the image content and provide information regarding the depth of the operative anatomy. The user can then use the measurement tools available on the subject IntraOp Alignment System to visualize and assess the spinal alignment for intraoperative planning of the surgical procedure. The subject IntraOp Alignment System has an optional capability to connect to a hospitalprovided Wi-Fi to import pre-operatively planned alignment parameters from an Alphatec cloud database. The subject IntraOp Alignment System also has the capability to export the post-operative alignment parameters to the cloud database. The subject IntraOp Alignment System sits outside the sterile field and is provided non-The subject device is intended for use by trained healthcare professionals, and sterile. appropriately trained clinical and non-clinical personnel. The subject device is intended for use in operating room environments of hospitals and surgical centers. System setup may be performed by trained non-clinical personnel. {5}------------------------------------------------ Image /page/5/Picture/0 description: The image shows the logo for "atec". The "a" is green and the "tec" is dark blue. There is a small trademark symbol in the upper right corner of the logo. The logo is simple and modern. This submission seeks to establish that the IntraOp Alignment System is substantially equivalent to the primary predicate, Nuvasive® LessRay® with Enhanced Tracking (K170800), and additional predicate ATEC Alignment App (K211987) with respect to indications for use (IFU), intended use. technology, and performance specifications. #### E. INDICATIONS FOR USE The IntraOp Alignment System is intended for use in applications where a mobile C-arm fluoroscope is incorporated to aid in diagnosis and treatment during spinal surgery. The IntraOp Alignment System is also intended to assist healthcare professionals in viewing, storing and measuring spinal alignment assessment images at various time points during surgery as well as planning spinal surgical procedures. The IntraOp Alignment System is intended for use with patients aged 18 years and older. #### F. TECHNOLOGICAL COMPARISON TO PREDICATES The subject device was compared to the predicate device in intended use, indications for use, design, function and technology and it was demonstrated that they are substantially equivalent. Any technological differences within this 510(k), between the subject device and the predicate device, does not impact substantially equivalence, or safety and effectiveness. Table 1 below provided a detailed analysis of the substantial equivalence of the following device characteristics: indications for use, device functionalities. C-arm tracking, algorithms and accuracy specifications. {6}------------------------------------------------ | Attribute | Predicate Devices | | Reference Device | Subject Device | SE Discussion | |----------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------| | | Nuvasive® LessRay® with Enhanced<br>Tracking (K170800) | ATEC Alignment App<br>(K211987) | NuVasive® NuvaLine®<br>(K192435) | IntraOp Alignment System | | | Indications for Use | NuVasive® LessRay® with Enhanced<br>Tracking is intended for use in any<br>application where a fluoroscope<br>incorporated to aid in diagnosis and<br>treatment of disease. | The ATEC Alignment App assists<br>healthcare professionals in viewing,<br>measuring, and storing images as well as<br>planning orthopedic surgeries. The app<br>allows the healthcare professional to<br>perform generic and specialty<br>measurements of the images, and to plan<br>surgical procedures. The app includes<br>tools for measuring anatomical<br>components for implant selection and<br>offers the possibility to share data among<br>ATEC Alignment App users. The app is<br>for planning use only and is not intended<br>for primary diagnostic use. | NuVasive NuvaLine is a medical device<br>software application intended to assist<br>healthcare professionals in capturing,<br>viewing, measuring, and storage and<br>distribution of spinal alignment assessment<br>images at various time points in patient<br>care. Online synchronization of the<br>database allows healthcare professionals<br>and service providers to conveniently<br>perform and review spinal alignment<br>assessments of images by featuring<br>measurement tools on various platforms.<br>Clinical judgment and experience are<br>required to properly use the software. | The IntraOp Alignment System is intended<br>for use in applications where a mobile C-<br>arm fluoroscope is incorporated to aid in<br>diagnosis and treatment during spinal<br>surgery.<br>The IntraOp Alignment System is also<br>intended to assist healthcare professionals<br>in viewing, storing and measuring spinal<br>alignment assessment images at various<br>time points during surgery as well as<br>planning spinal surgical procedures.<br>The IntraOp Alignment System is intended<br>for use with patients aged 18 years and<br>older. | The subject device indications for use are<br>same as the primary predicate,<br>Nuvasive® LessRay® with Enhanced<br>Tracking (K170800) for the image<br>stitching functionality. The indications<br>for use for use are also identical to the<br>additional predicate ATEC Alignment<br>App (K211987) for the spinal alignment<br>assessments functionality. Completed<br>V&V testing successfully demonstrated<br>that the risks related to the differences in<br>the indications for use between the<br>subject device and primary predicate<br>have been mitigated and do not affect the<br>safety and effectiveness of the device.<br>Therefore, the subject device can be<br>considered substantially equivalent to the<br>primary predicate, Nuvasive® LessRay®<br>with Enhanced Tracking (K170800) and<br>additional predicate, ATEC Alignment<br>App (K211987). | | Regulation Number | §892.1650 | §892.2050 | §892.2050 | §892.1650; | Same | | Product Code | OWB, JAA, LLZ | LLZ | LLZ | OWB, JAA, LLZ | Same | | Device Class | II | II | II | II | Same | | Device Classification Name | Interventional Fluoroscopic X-Ray<br>System | System, Image Processing, Radiological | System, Image Processing, Radiological | Interventional Fluoroscopic X-Ray System | Same | | Device Functionalities | Image Acquisition, Enhancement and<br>Display :<br>• Software based device used to<br>provide computer display systems<br>interfaced to fluoroscope through a<br>video cable. The images produced<br>by the fluoroscope are transmitted<br>through a cable to a frame capture<br>board in the computer where the<br>images are enhanced and then<br>displayed on the monitor.<br>• Enhanced images are displayed on a<br>computer monitor at the same time<br>that the corresponding original | Measurement:<br>• Spinal alignment assessments of<br>images | Measurement:<br>• Spinal alignment assessments of<br>images | Image Acquisition, Processing,<br>Stitching, and Display:<br>• Software based system that from<br>one end interfaces with a mobile C-<br>arm fluoroscope through a video<br>cable and a frame capture device to<br>receive fluoroscopic images as they<br>are acquired.<br>• As the images are being acquired,<br>the user of the system can combine<br>multiple fluoroscopic images, into<br>stitched long bi-planar radiographic<br>images for intraoperative | Substantially Equivalent. The image<br>stitching feature of the subject device is<br>similar in function to the primary<br>predicate. Nuvasive® LessRay® with<br>Enhanced Tracking (K170800). The<br>visualization and spinal alignment<br>assessment feature of the subject device is<br>similar in function to the additional<br>predicate, ATEC Alignment App<br>(K211987). | | Attribute | Predicate Devices | | Reference Device | Subject Device | SE Discussion | | | Nuvasive® LessRay® with Enhanced<br>Tracking (K170800) | ATEC Alignment App<br>(K211987) | NuVasive® NuvaLine®<br>(K192435) | IntraOp Alignment System | | | | • image is displayed on the<br>fluoroscope monitor(s).<br>• Serves only as an image display<br>which is in addition to the<br>fluoroscope's standard image<br>display device. Device is passive, in<br>that the operation depends only on<br>the video output of the fluoroscope,<br>and it does not transmit any signals<br>or images to the fluoroscope. | | | visualization and spinal alignment<br>assessments.<br>• Serves only as an image display which<br>is in addition to the fluoroscope's<br>standard image display device. Device<br>is passive, in that the operation depends<br>only on the video output of the<br>fluoroscope, and it does not transmit<br>any signals or images to the<br>fluoroscope.<br>Spinal Alignment Measurements:<br>• After bi-planar images are acquired,<br>they are calibrated by the user using<br>the established tools available on<br>IntraOp Alignment System<br>software.<br>• The user can place measurement<br>tools for spinal alignment<br>assessments. | | | C-arm Tracking | • When tracking is enabled, will<br>automatically choose the Baseline<br>when the fluoroscope is near the<br>location and orientation that the<br>Baseline was initially taken.<br>• When tracking is enabled, requires<br>hardware components in order to<br>mount the off-the-shelf tracking<br>hardware to the C-arm and to the<br>operating table.<br>• When tracking is enabled, requires<br>the use of an off-the-shelf tracking<br>system in order to track the 6 DOF<br>location of the C-arm relative to the<br>operating table.<br>• When tracking is enabled, visual<br>cues are provided which help guide<br>the user in positioning the C-arm<br>back to where a prior Baseline was<br>taken. | N/A | N/A | • IntraOp Alignment System uses off-<br>the-shelf tracking sensor and reflective<br>tracking markers to track the 6 DOF<br>location of the C-arm relative to<br>tracking sensor.<br>• The reflective tracking markers are<br>attached on the mobile C-arm and are<br>registered and calibrated on the IntraOp<br>Alignment System after its initial<br>application on the C-arm.<br>• The IntraOp Alignment System GUI<br>guides the user regarding the<br>placement of the C-arm over the<br>operative anatomy and tracks the C-<br>arm position relative to the tracking<br>sensor using the reflective markers. | Substantially Equivalent. | | Tracking Options | • Electromagnetic or optical | N/A | N/A | • Optical | Substantially Equivalent | | Algorithms | • Image quality improvement using<br>averaging algorithm | Patient parameters and calculations based<br>on published literature | Various spinal assessment algorithms | • Image quality improvement by<br>distortion removal using C-Arm<br>calibration data | Substantially Equivalent. The image<br>stitching algorithm of the subject device is<br>similar in function to the primary | | Attribute | Predicate Devices | | Reference Device | Subject Device | | | | Nuvasive® LessRay® with Enhanced<br>Tracking (K170800) | ATEC Alignment App<br>(K211987) | NuVasive® NuvaLine®<br>(K192435) | IntraOp Alignment System | SE Discussion | | | • Contrast and brightness<br>enhancement with simultaneous<br>reduction of random noise | | | • Image stitching using C-Arm position<br>coordinates<br>• Depth correction based on user input<br>• Various spinal assessment algorithms | predicate, Nuvasive® LessRay® with<br>Enhanced Tracking (K170800). The<br>spinal alignment assessment algorithm of<br>the subject device is similar in function to<br>the additional predicate, ATEC<br>Alignment App<br>(K211987). | ## Table 1: Comparison for Substantial Equivalence {7}------------------------------------------------ {8}------------------------------------------------ Image /page/8/Picture/0 description: The image shows the word "atec" in a stylized font. The "a" is green and has a swoosh-like extension that goes over the "t". The rest of the letters are in a dark blue color. There is a trademark symbol in the upper right corner of the "c". {9}------------------------------------------------ Image /page/9/Picture/0 description: The image shows the logo for "atec". The "a" is green, and the rest of the letters are black. There is a trademark symbol in the upper right corner of the logo. There is a black line underneath the logo. #### PERFORMANCE DATA G. Nonclinical performance testing demonstrates that the subject IntraOp Alignment System meets the functional, system, and software requirements. The following testing was conducted: - Verification testing to confirm that the system meets the accuracy specifications ● - Verification of the workflow to confirm that the system performs according to ● specifications - Verification testing to verify that the following features: image stitching, C-arm tracking and spinal alignment measurements meet performance specifications EMC and Electrical Safety Testing of the IntraOp Alignment System was performed to ensure all functions of the system and its accessories are electrically safe and comply with recognized electrical safety standards. ## Clinical Information Determination of substantial equivalence is not based on an assessment of clinical performance data. #### H. CONCLUSION The information provided in this submission demonstrates that the subject device does not pose additional risk to safety and effectiveness when compared to the predicate device. The subject device, IntraOp Alignment System, is substantially equivalent to the primary predicate, Nuvasive® with Enhanced Tracking (K170800), and additional predicate ATEC Alignment App (K211987).
Innolitics

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...