ENPOLE is a single-use medical device to have a port to insert endoscopic instruments for endoscopic procedures.
Device Story
The ENPOLE Bladeless Trocar is a sterile, single-use surgical access device. It consists of a sleeve (5-12 mm diameter) and an obturator used to penetrate the abdominal wall. The sleeve features an integrated self-adjusting outer seal and an internal seal to maintain pneumoperitoneum by minimizing gas leakage during instrument insertion or withdrawal. A stopcock valve allows for gas insufflation and desufflation via standard luer lock fittings. Used in laparoscopic surgery, the device provides a sealed port for endoscopic instruments, enabling surgeons to perform procedures at a constant abdominal pressure. The device is intended for use by trained healthcare professionals in a clinical or surgical setting.
Clinical Evidence
Bench testing only. Performance data included appearance, dimension, air leakage, seal testing, and instrument compatibility. Non-clinical testing included packaging validation, EO sterilization validation, bioburden testing, EO gas residue testing, shelf-life/real-time sterility testing, transport validation, and biocompatibility.
Technological Characteristics
Sterile, single-use bladeless trocar. Sizes: 5, 10, 11, 12 mm diameters; 60, 75, 100 mm lengths. Features integrated self-adjusting outer seal and internal seal. Includes stopcock valve for luer lock gas insufflation. Sterilization: Ethylene Oxide (EO) to S.A.L. 10^-6.
Indications for Use
Indicated for patients undergoing endoscopic procedures requiring a port for the insertion of endoscopic instruments.
Regulatory Classification
Identification
An endoscope and accessories is a device used to provide access, illumination, and allow observation or manipulation of body cavities, hollow organs, and canals. The device consists of various rigid or flexible instruments that are inserted into body spaces and may include an optical system for conveying an image to the user's eye and their accessories may assist in gaining access or increase the versatility and augment the capabilities of the devices. Examples of devices that are within this generic type of device include cleaning accessories for endoscopes, photographic accessories for endoscopes, nonpowered anoscopes, binolcular attachments for endoscopes, pocket battery boxes, flexible or rigid choledochoscopes, colonoscopes, diagnostic cystoscopes, cystourethroscopes, enteroscopes, esophagogastroduodenoscopes, rigid esophagoscopes, fiberoptic illuminators for endoscopes, incandescent endoscope lamps, biliary pancreatoscopes, proctoscopes, resectoscopes, nephroscopes, sigmoidoscopes, ureteroscopes, urethroscopes, endomagnetic retrievers, cytology brushes for endoscopes, and lubricating jelly for transurethral surgical instruments. This section does not apply to endoscopes that have specialized uses in other medical specialty areas and that are covered by classification regulations in other parts of the device classification regulations.
Special Controls
*Classification* —(1)*Class II (special controls).* The device, when it is an endoscope disinfectant basin, which consists solely of a container that holds disinfectant and endoscopes and accessories; an endoscopic magnetic retriever intended for single use; sterile scissors for cystoscope intended for single use; a disposable, non-powered endoscopic grasping/cutting instrument intended for single use; a diagnostic incandescent light source; a fiberoptic photographic light source; a routine fiberoptic light source; an endoscopic sponge carrier; a xenon arc endoscope light source; an endoscope transformer; an LED light source; or a gastroenterology-urology endoscopic guidewire, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 876.9.(2) Class I for the photographic accessories for endoscope, miscellaneous bulb adapter for endoscope, binocular attachment for endoscope, eyepiece attachment for prescription lens, teaching attachment, inflation bulb, measuring device for panendoscope, photographic equipment for physiologic function monitor, special lens instrument for endoscope, smoke removal tube, rechargeable battery box, pocket battery box, bite block for endoscope, and cleaning brush for endoscope. The devices subject to this paragraph (b)(2) are exempt from the premarket notification procedures in subpart E of part 807of this chapter, subject to the limitations in § 876.9.
In combination with the general controls of the FD&C Act, the integrated operating table-electromechanical surgical system is subject to the following special controls:
1. (1) Premarket clinical performance testing, or a combination of premarket clinical performance testing and postmarket surveillance (in accordance with special control (2)), must include the following:
1. (i) Objective performance measures (e.g., rate and number of conversions to other surgical modalities, rate of device related adverse events (including tissue injury, hematoma, and increased blood loss), and their severity, cause, and outcomes) must be reported with relevant descriptive comparator performance measures.
2. (ii) The data must demonstrate the performance of the device for providing accurate and precise control of attached surgical instruments in range of clinical conditions relevant to the device's intended use.
3. (iii) The test dataset must include data collected from a patient population representative of the intended patient population under anticipated conditions of use.
2. (2) Data obtained from postmarket surveillance must demonstrate, in consideration of the premarket data obtained in accordance with special control (1), that the device performs in accordance with special control (1), unless FDA determines, based on the totality of the premarket data, that data from postmarket surveillance is not required to demonstrate that the device performs as intended. Such postmarket surveillance must be conducted per a protocol determined appropriate by FDA to demonstrate that the device performs as intended (in consideration of the premarket data obtained in accordance with special control (1)), and must include initiation, enrollment, and reporting requirements to ensure timely periodic updates to FDA on post-market surveillance progress and outcomes.
3. (3) Animal performance testing must evaluate the extent of port site trauma due to repositioning of table during surgical procedures when utilizing robotic minimally invasive and laparoscopic approaches
4. (4) The device manufacturer must develop, and update as necessary, a device-specific use training program that ensures proper device setup/use/shutdown, accurate control of instruments to perform the intended surgical procedures, troubleshooting and handling during unexpected events or emergencies, and safe practices to mitigate use error.
5. (5) The device manufacturer may only distribute the device to facilities that implement and maintain the device-specific use training program and ensure that users of the device have completed the device-specific use training program.
(6) Human factors assessment must demonstrate that the user can correctly use the device system across all intended use environments with the provided instructions and training materials, including patient access during normal operating conditions and emergency situations, and effects arising from the integrated nature of the operating table and robotic surgical arms.
(7) Labeling must include:
(i) A detailed summary of clinical performance testing conducted with the device, including study population, results, adverse events, and comparisons to any comparator groups identified;
(ii) A statement in the labeling that the safety and effectiveness for the representative specific procedures was based on evaluation of the device as a surgical tool and did not include evaluation of outcomes related to the treatment of the patient's underlying disease or condition, unless FDA determines that it can be removed or modified based on clinical performance data submitted to FDA;
(iii) Identification of compatible devices;
(iv) The list of surgical procedures for which the device has been determined to be safe with clinical justification;
(v) Reprocessing instructions for reusable components;
(vi) A shelf life for any sterile components;
(vii) A description of the device-specific use training program;
(viii) A statement that the device is only for distribution to facilities that implement and maintain the device-specific use training program and ensure that users of the device have completed the device-specific use training program; and
(ix) A summary of any completed postmarket surveillance data collected as required by special control (2), including updated labeling to accurately reflect outcomes observed in postmarket surveillance.
(8) Non-clinical performance testing must demonstrate that the device performs as intended under anticipated conditions of use and must include:
(i) Device motion accuracy and repeatability;
(ii) System testing;
(iii) Instrument reliability;
(iv) Crosstalk;
(v) Table motion control;
(vi) Thermal effects on tissue;
(vii) User-device interface performance;
(viii) Workspace access testing; and
(ix) Performance testing with compatible devices.
(9) Software verification, validation, and hazard analysis must be performed.
(10) Electromagnetic compatibility and electrical, thermal, and mechanical safety testing must be performed.
(11) Performance data must demonstrate the sterility of all patient-contacting device components.
(12) Performance data must support the shelf life of the device components provided sterile by demonstrating continued sterility and package integrity over the labeled shelf life.
(13) Performance data must validate the reprocessing instructions for the reusable components of the device.
(14) Performance data must demonstrate that all patient-contacting components of the device are biocompatible.
(15) Performance data must demonstrate that all patient-contacting components of the device are non-pyrogenic.
(16) The device manufacturer must submit a report to the FDA annually on the anniversary of initial marketing authorization for the device, until such time as FDA may terminate such reporting, which comprises the following information:
(i) Cumulative summary, by year, of complaints and adverse events since date of initial marketing authorization; and
(ii) Identification and rationale for changes made to the device, labeling, or device specific use training program, which did not require submission of a premarket notification during the reporting period.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: a symbol on the left and the FDA name on the right. The symbol on the left is a stylized image of a human figure, while the FDA name on the right is written in blue letters. The words "U.S. FOOD & DRUG ADMINISTRATION" are written in a clear, sans-serif font.
June 18, 2024
Yuwonmeditech Co., Ltd. Bae Ji Hyeon Manager, Regulatory Affairs Rm 103, 104, 108, 113, 114, 117, Donghwa Medical Instrument 147-1, Donghwagongdan-ro, Munmak-eup Wonju-si, Gangwon-do 26365 Korea. South
Re: K240196
Trade/Device Name: Bladeless Trocar (E05, E10, E11, E12, ES05, ES10, ES12) Regulation Number: 21 CFR 876.1500 Regulation Name: Endoscope And Accessories Regulatory Class: Class II Product Code: GCJ Dated: May 20, 2024 Received: May 20, 2024
Dear Bae Ji Hyeon :
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device"
{1}------------------------------------------------
(https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30. Design controls; 21 CFR 820.90. Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the OS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely.
Digitally signed by Mark Mark Trumbore -S Trumbore -S Date: 2024.06.18 08:49:59 -04'00' Mark Trumbore, Ph.D. Assistant Director DHT4A: Division of General Surgery Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Quality
{2}------------------------------------------------
Enclosure
{3}------------------------------------------------
# Indications for Use
Submission Number (if known)
K240196
Device Name
Bladeless Trocar (E05, E10, E11, E12, ES05, ES10, ES12)
Indications for Use (Describe)
ENPOLE is a single-use medical device to have a port to insert endoscopic instruments for endoscopic procedures.
Type of Use (Select one or both, as applicable)
> Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{4}------------------------------------------------
# 510(k) Summary
(ENPOLE® Bladeless Trocar)
This summary of 510(k) safety and effectiveness information is being submitted in accordance with requirements of 21 CFR Part 807.92.
Date 510k summary prepared: 18/01/2024
# 1. SUBMITTER
Submitter's Name: YUWONMEDITECH CO., LTD. Submitter's Address: 103, 104, 108, 113, 114, 117, Donghwa Medical Instrument Complex, 147-1, Donghwagongdan, Munmak-eup, Wonju-si, Gangwon-do, Korea Telephone Number: +82-33-743-0081 Fax: +82-33-743-0081
# 2. CONTACT PERSON
Byung-kyu, Park Quality Management Representative
Ji-Hyeon, Bae Manager, Regulatory Affairs
TEL. +82-33-743-0081 E-Mail: bkpark@yuwonmeditech.com, jhbae@yuwonmeditech.com Address: 103, 104, 108, 113, 114, 117, Donghwa Medical Instrument Complex, 147-1, Donghwagongdan-ro, Munmak-eup, Wonju-si, Gangwon-do, Korea
# 3. DEVICE INFORMATION
Trade Name: ENPOLE® Bladeless Trocar Common Name: Bladeless Trocar Product Code: GCJ Classification Regulation: 21 CFR 876.1500 Device Class: Class II Classification Panel: General and Plastic Surgery
#### 4. PREDICATE DEVICE
The proposed ENPOLE® Bladeless trocars are substantially equivalent to the predicate devices:
| Predicate Device | Manufacturer | 510(k) No. | Decision Date |
|------------------|-----------------|------------|---------------|
| CORE-Trocar | INCORE CO.,LTD. | K211148 | 05/19/2022 |
1
{5}------------------------------------------------
#### 5. DEVICE DESCRIPTION
The Bladeless Trocar is a sterile single patient use instrument consisting of a sleeve in sizes from 5 mm to 12 mm diameters. The trocar sleeve for the devices contains two seal, an outer integrated self-adjusting seal that accommodates instruments ranging from 5 mm to 12.5 mm in diameter where indicated and an internal seal. Together, these seal minimize gas leakage when instruments are inserted or withdrawn through the trocar. The 5 mm trocar sleeve accommodates only 5 mm instruments. A stopcock valve is compatible with standard luer lock fittings and provides attachment for gas insufflation and desufflation.
Bladeless trocar is access into the abdomen of laparoscopy that is particular to the insertion of surgical instruments through small incisions. The sleeve penetrated into the intra- abdominal by the obturator is inserted and fixed in the abdominal wall, which is a principle that allows laparoscopy to be performed at a constant abdominal pressure state by keeping it sealed to the outside of the human body.
#### 6. INDICATIONS FOR USE
ENPOLE Trocar is a single-use medical device to have a port to insert endoscopic instruments for endoscopic procedures.
## 7. SUBSTANTIAL EQUIVALENCE
The proposed ENPOLE® Bladeless Trocars have the same technology and functional characteristics as the predicate systems. The intended use, operating principles, performance specifications, sterilization and expiration dating of the subject device are the same as the predicate device.
| Descriptive Information | Subject Device | Predicate Device (K211148) |
|-------------------------------|-------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------|
| Manufacturer | YUWONMEDITECH CO., LTD. | INCORE CO., LTD. |
| Model Name | ENPOLE | CORE-TROCAR |
| Classification<br>Regulation | 21 CFR 876.1500 | 21 CFR 876.1500 |
| Product Code | GCJ | GCJ |
| Classification | Class II | Class II |
| Device Classification<br>Name | Laparoscope, General & Plastic<br>Surgery | Laparoscope, General & Plastic<br>Surgery |
| Indications for Use | ENPOLE Trocar is a single-use<br>medical device to have a port<br>to insert endoscopic<br>instruments for endoscopic<br>procedures. | A single-use medical device to<br>have a port to insert<br>endoscopic instruments for<br>endoscopic procedures. |
| Compared elements | Cannula<br>Tip of obturator | Cannula<br>Tip of obturator |
# 7.1 Comparison Table
{6}------------------------------------------------
| Picture | Image: Medical device | Image: Medical device |
|---------------------|--------------------------------|--------------------------------|
| Dimension | Diameter: 5, 10, 11, 12<br>mm | Diameter: 2 ~ 12 mm |
| | Length: 60, 75,100 mm | Length: 100 ~ 150 mm |
| Sterilization | EO Sterilization (S.A.L. 10-6) | EO Sterilization (S.A.L. 10-6) |
| Shelf Life | 5 years | 3 years |
| Single use / Re-use | Single-use | Single-use |
## 8. PERFORMANCE DATA
| Test | Criteria | Standard | Result |
|-------------------------------|-------------------|-------------------|--------|
| Appearance | Internal standard | Internal standard | Pass |
| Dimension | Internal standard | Internal standard | Pass |
| Air leakage | Internal standard | Internal standard | Pass |
| Seal test | Internal standard | Internal standard | Pass |
| Instrument compatibility test | Internal standard | Internal standard | Pass |
# 9. NON-CLINICAL TEST
- Packaging Validation -
- -E.O sterilization Validation
- -Bioburden Test, E.O gas residue Test
- -Shelf-Life Test/Real-time Sterility Test
- Transport Validation -
- Biocompatibility -
#### 10. Conclusion
Based upon the test results, the proposed ENPOLE® Bladeless Trocars are substantially equivalent in performance to the predicate devices cleared to market via 510(k) K211148. The proposed ENPOLE® Bladeless Trocars do not introduce any new issues of safety and effectiveness.
-