NM-01/CPT neurometer (NM-01/CPT)

K240189 · Mde Orvosbiológiai Kutató, Fejleszto, · JXE · Mar 26, 2025 · Neurology

Device Facts

Record IDK240189
Device NameNM-01/CPT neurometer (NM-01/CPT)
ApplicantMde Orvosbiológiai Kutató, Fejleszto,
Product CodeJXE · Neurology
Decision DateMar 26, 2025
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 882.1550
Device ClassClass 2

Indications for Use

The NM-01/CPT neurometer is a noninvasive electrodiagnostic device intended for verifying nerve integrity in conscious adult patients, with intact skin surface. The population of subject for whom this device may be used for diagnostic purposes would include any individual capable of communicating their perception of cutaneous sensation. The NM-01/CPT neurometer may be conducted as a part of a routine neurological examination. The measured data can be utilized in evaluating patients suspected of having neuropathies. The measured data must be used in the context of other patient information and must be reviewed and interpreted by a physician. The device is intended for use on adults in medical clinics, healthcare practices and out-patient departments of hospitals.

Device Story

NM-01/CPT is a noninvasive electrodiagnostic device for assessing sensory nerve fiber function (small unmyelinated C, small myelinated A-delta, large myelinated A-beta fibers). It emits non-aversive transcutaneous electrical stimuli (sinusoidal waveform) via gold-coated electrodes at fixed frequencies (5Hz, 250Hz, 2000Hz). The device measures Current Perception Threshold (CPT) values, manually adjustable from 0.01 to 9.99 mA. Operated by trained assistants in clinics/hospitals under physician supervision. The system includes software for device control, data storage, and normative value comparison. Physicians review output to evaluate neuropathies. Benefits include painless, noninvasive assessment of nerve integrity at any cutaneous site, unaffected by skin temperature or edema.

Clinical Evidence

Bench testing only. Comparative analysis performed between NM-01/CPT and predicate device using resistors and CPT values. Two-sample TOST tests confirmed measurement equivalence within a 10% limit for resistance values (1, 5, 30 Kohm) and CPT values >10. CPT 10 values showed a mean ratio of 1.109, falling outside the 10% equivalence limit, but these values are not expected in clinical practice.

Technological Characteristics

Electrodiagnostic device; sinusoidal waveform stimulus; 0-10 mA constant AC output; 5Hz, 250Hz, 2000Hz frequencies. Permanent gold-coated electrodes. Rechargeable internal battery. Bluetooth connectivity. Software-based control (Windows, iOS, Android). Dimensions: 30 x 12 x 25.5 cm. Includes automatic sensor displacement detection and patient cable integrity checks.

Indications for Use

Indicated for verifying nerve integrity in conscious adult patients with intact skin capable of communicating cutaneous sensation. Used as part of routine neurological exams to evaluate suspected neuropathies.

Regulatory Classification

Identification

A nerve conduction velocity measurement device is a device which measures nerve conduction time by applying a stimulus, usually to a patient's peripheral nerve. This device includes the stimulator and the electronic processing equipment for measuring and displaying the nerve conduction time.

Predicate Devices

Submission Summary (Full Text)

{1}------------------------------------------------ Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the FDA logo is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue. March 26, 2025 MDE Orvosbiológiai Kutató, Fejleszto, Aba Tomjanovich CEO Gyártó Korlátolt Felelosségu Társaság Podmaniczky u. 87 Budapest, 1064 Hungary Re: K240189 Trade/Device Name: NM-01/CPT neurometer (NM-01/CPT) Regulation Number: 21 CFR 21 CFR 882.1550 Regulation Name: Nerve Conduction Velocity Measurement Device Regulatory Class: Class II Product Code: JXE Dated: February 24, 2025 Received: February 24, 2025 Dear Aba Tomjanovich: We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. {2}------------------------------------------------ Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download). Your device is also subject to, among other requirements, the Quality System (OS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181). Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050. All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-device-advicecomprehensive-regulatory-assistance/unique-device-identification-system-udi-system. Also, please note the regulation entitled. "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems. For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatory {3}------------------------------------------------ assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100). Sincerely, Image /page/2/Picture/3 description: The image shows the FDA logo on the left and the name "Patrick Antkowiak -S" on the right. The FDA logo is in a light blue color. The name is in black and is written in a simple, sans-serif font. for Jay Gupta Assistant Director DHT5A: Division of Neurosurgical, Neurointerventional, and Neurodiagnostic Devices OHT5: Office of Neurological and Physical Medicine Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health Enclosure {4}------------------------------------------------ ## Indications for Use Submission Number (if known) K240189 Device Name NM-01/CPT neurometer (NM-01/CPT) Indications for Use (Describe) The NM-01/CPT neurometer is a noninvasive electrodiagnostic device intended for verifying nerve integrity in conscious adult patients, with intact skin surface. The population of subject for whom this device may be used for diagnostic purposes would include any individual capable of communicating their perception of cutaneous sensation. The NM-01/CPT neurometer may be conducted as a part of a routine neurological examination. The measured data can be utilized in evaluating patients suspected of having neuropathies. The measured data must be used in the context of other patient information and must be reviewed and interpreted by a physician. The device is intended for use on adults in medical clinics, healthcare practices and out-patient departments of hospitals. Type of Use (Select one or both, as applicable) > Prescription Use (Part 21 CFR 801 Subpart D) Over-The-Counter Use (21 CFR 801 Subpart C) #### CONTINUE ON A SEPARATE PAGE IF NEEDED. This section applies only to requirements of the Paperwork Reduction Act of 1995. #### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* The burden time for this collection of information is estimated to average 79 hours be response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: > Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." {5}------------------------------------------------ | 510(k) #: | K240189 | 510(k) Summary | Prepared on: | 2025-02-23 | |-----------|---------|----------------|--------------|------------| |-----------|---------|----------------|--------------|------------| | Contact Details | | 21 CFR 807.92(a)(1) | |-----------------------------|-----------------------------------------------------------------------------|---------------------| | Applicant Name | MDE Orvosbiológiai Kutató, Fejlesztő, Gyártó Korlátolt Felelősségű Társaság | | | Applicant Address | Podmaniczky u. 87 Budapest 1064 Hungary | | | Applicant Contact Telephone | +36305968065 | | | Applicant Contact | Mr. Aba Tomjanovich | | | Applicant Contact Email | qara@mdegmbh.eu | | | Device Name | | 21 CFR 807.92(a)(2) | |---------------------|-----------------------------------------------|---------------------| | Device Trade Name | NM-01/CPT neurometer (NM-01/CPT) | | | Common Name | Nerve conduction velocity measurement device | | | Classification Name | Device, Nerve Conduction Velocity Measurement | | | Regulation Number | 882.1550 | | | Product Code(s) | JXE | | | Legally Marketed Predicate Devices | | 21 CFR 807.92(a)(3) | |------------------------------------|--|---------------------| |------------------------------------|--|---------------------| | Predicate # | Predicate Trade Name (Primary Predicate is listed first) | Product Code | |-------------|----------------------------------------------------------|--------------| | K843924 | Digital Electroneurometer S-100 | JXE | | K190536 | Mediracer | JXE | | Device Description Summary | | 21 CFR 807.92(a)(4) | |----------------------------|--|---------------------| |----------------------------|--|---------------------| The NM-01/CPT neurometer is a noninvasive device intentity in conscious adult patients, with intact skin surface. The population of subject for whom this device may be used for diagnostic purposes would include any individual capable of communicating their perception of cutaneous sensation. The NM-01/CPT neurometer may be conducted as a part of a routine neurological examination. The measured data can be utilized in evaluating patients suspected of having neuropathies. The measured data must be used in the context of other patient information and interpreted by a physician. The device is intended for use on adults in medical clinics, healthcare practices and out-patient departments of hospitals. The NEUROMETER NM-01/CPT is a unique neurodiagnostic device that painlessly evaluates the functioning of small unmyelinated (C, small myelinated (A) and large myelinated (A) sensory nerve fibers at any cutaneous site by determining neuro selective sensory nerve conduction threshold. The electronic unit emits non-aversive transcutaneous electrical stimuli through a pair of special noninvasive electrodes in three fixed frequency ranges (thick myelincoated fiber 250Hz, thin myelin-coated fiber 2000Hz) with a manually adjustable current value. The emitted stimulus pulses canbe changed between 0.01 and 9.99 mA and their values qiven by the software are in CPT (Current Precipitation Threshold) values accepted in clinical practice (1CPT = 0.01mA). The test method is sensitive to both hyperesthetic abnormalities, and has more than 800 peer reviewed research publications documenting its diagnostic capabilities. The device can measure sensory nerve function at any cutaneous site including mucosal surfaces such as the bladder and the measures are not affected by skin temperature, edema or electromag-netic interference unlike traditional electro-diagnostic and biopsy procedures. The conducted tests are painless, noninvasive and non-aversive procedure is easy to perform and safe and harmless however it is only for profession clinical use. The evaluation of the tests result could only be approved by a medical doctor but the tests could be conducted by a trained assistant. {6}------------------------------------------------ # Intended Use/Indications for Use ### The NM-01/CPT neurometer is a noninyasive electrodiagnostic device intended for verifying nerve integrity in conscious adult patients, with intact skin surface. The population of subject for whom this device may be used for diagnostic purposes would include any individual capable of communicating their perception of cutaneous sensation. The NM-01/CPT neurometer may be conducted as a part of a routine neurological examination. The measured data can be utilized in evaluating patients suspected of having neuropathies. The measured data must be used in the patient information and must be reviewed and interpreted by a physician. The device is intended for use on adults in medical clinics, healthcare practices and out-patient departments of hospitals. # Indications for Use Comparison # 21 CFR 807.92(a)(5) ## Neurometer CPT - K843924 "electrodiagnostic devices perform automated neuroselective sensory Nerve Conduction Threshold(sNCT) evaluations by determining Current Perception Threshold (CPT) measures" ## NM-01 CPT - K240189 "electrodiagnostic devices perform automated neuroselection Threshold (sNCT) evaluations by determining Current Perception Threshold (CPT) measures" The two devices are substantially equivalent. # Technological Comparison The Neurometer CPT (K843924) and NM-01 CPT (K240189) are highly similar in their technological approach. Both are electrodiagnostic devices designed for automated nerve testing and evaluating peripheral neuropathy. They utilize a nusoidal waveform stimulus and detect nerve responses using verbal and remote-based methods. Additionally, both devices are powered by a rechargeable internal battery, making them portable and convenient for clinical use. Due to these similarities, NM-01 CPT is considered substantially equivalent to the Neurometer CPT. ln contrast, the Mediracer (K190536), while also used for evaluating peripheral neuropathy, employs a different technological approach. Instead of a sinusoidal waveform stimulus, it uses a dual-phase, rectangular waveform. #### Non-Clinical and/or Clinical Tests Summary & Conclusions 21 CFR 807.92(b) Bench testing was performed on the NM-01/CPT neurometer measured values with the results of the Neurometer used in clinical practice. For the comparative analysis, we calculated the ratio of the two measurements. We examined the equivalence of the ratios per resistor and per CPT value. For the analysis, we used two-sample TOST tests to check whether the (90%; 111%) interval included the 90% confidence interval of the quotient. The ratio of the measured result equivalence was above 1 for each resistance (1, 5, 30 Kohm). The confidence intervals do not include a value of 1 in any of the cases, however in all three cases they show a difference within 10% of the confidence intervals (90%; 11%), According to the TOST tests performed the two measurements can be considered within the 10% limit. For the CPT breakdown, the ratios are above 1 for CPT values between 10 and 250 and below 1 for CPT values above that . The confidence intervals of the ratios for values greater than 10 CPTs are (90%; 1 1%), and t he TOST tests performed indicate that the two measurements are equivalent within the 10% limit. For CPT 10, the mean of the ratio is 1.109, with a 90% confidence interval of (1.0531; 1.1648), i.e. in this case the two measurements cannot be considered equivalent within the 10% limit. Note that such values are not expected in the population under study. To summarise the above, since the measurements were equivalent in all cases except for the 10 CPT values not used in clinical practice, the results of the two instruments can be considered equivalent in the measurement domain to be applied. # 21 CFR 807.92(a)(6) {7}------------------------------------------------ ## Table 1 - Technological Comparison (21 CFR 807.92(a)(6)) | | Substantially<br>equivalent? | Nerometer CPT - K843924 | NM-01 CPT - K240189 | Mediracer - K190536 | Substantially equivalent? | |------------------------------------|---------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------| | Diseases to be Diagnosed | Yes | Peripheral Neuropathy | Peripheral Neuropathy | Peripheral Neuropathy | Yes | | Intended use | Yes | Performs automated<br>neuroselective sensory Nerve<br>Conduction Threshold<br>(sNCT) evaluations by determining<br>Current Perception Threshold<br>(CPT) measures | Performs automated<br>neuroselective sensory Nerve<br>Conduction Threshold<br>(sNCT) evaluations by determining<br>Current Perception Threshold<br>(CPT) measures | Device for stimulate and<br>measure neuromuscular<br>that are useful in signals<br>that are useful in<br>diagnosing and evaluating<br>systemic and entrapment<br>neuropathies. | Yes | | Power supply | Yes | Rechargeable Internal Battery | Rechargeable Internal Battery | Rechargeable Internal Batter | Yes | | Simulation detection | Yes | Verbal and remote based | Verbal and remote based | Muscle response by<br>electronical measurement | No | | Stimulus Type | Yes | Sinusoid | Sinusoid | Dual phase, rectangular<br>wave | No | | Resolution | Yes | 0.001 mA | 0.001 mA | 10 bits | No | | Output Current | Yes | 0-10 mA Constant AC | 0-10 mA Constant AC | NA | No | | Stimulation Frequency | Yes | 5Hz, 250Hz, 2000Hz | 5Hz, 250Hz, 2000Hz | 2 Hz | No | | Recording Channels | Yes | 1 channel | 1 channel | 1 channel | Yes | | Connectivity | No | Wire connection | Bluetooth connection | Bluetooth connection | Yes | | Dimensions | Yes | 11,43 x 8,89 x 6,35 cm | 30 x 12 x 25,5 cm | 18.99x6.971 | Yes | | Electrodes | Yes | Disposable Goldtrode (golden) | Permanent gold coated electrodes | Stimulation Ring Electrode | Yes | | Electrode safety | Yes | Cabel chack before each usage | automatic sensor displacement<br>detection built in testing protocol<br>for the checking the patient cable<br>damages | NA | No | | Controlling the device | Yes | Manual control by buttons of the<br>device | Software based controlling the<br>device that could run on different<br>operating system (Win, los,<br>Android) | Software based controlling<br>the device that could run<br>on different operating<br>system (Win) | Yes | | Nerves Tested | Yes | small unmylenated (C), small<br>myelinated (Aδ) and large<br>myelinated (Aβ) sensory nerve<br>fibers | small unmylenated (C), small<br>myelinated (Aδ) and large<br>myelinated (Aβ) sensory nerve<br>fibers | Median and ulnar nerves | No | | Normative values | Yes | Normative values established for<br>more<br>than 30 multiple testing sites | Normative values included to the<br>software evaluation for more than<br>30 multiple testing sites | NA | No | | Assessment time | Yes | 10 - 20 minutes depending on the<br>number of the analyzed arrea | 10 - 18 minutes depending on the<br>number of the analyzed arrea | 15 minutes | Yes | | Access to test data adjustment | Yes | Using the device remote to export<br>the stored measurement values<br>from memory | Cursor placing adjustment | Cursor placing adjustment | Yes | | Patient Data Handling and Storage | Neurometer does<br>not have patient<br>handling system. | The device has a built in memory<br>for saving the measured CPT<br>values, data could be saved to a<br>disk for export but it does not<br>have patient management<br>systemvsince it is an analog device | On the software included part of<br>the system. All patient and test<br>data could be stored on the device. | On the software included<br>part of the system. All<br>patient and test data could<br>be stored on the device. | Yes | | Operator Skill Requirements | Yes | Trained technician supervised by a<br>physician or a physician | Trained technician supervised by a<br>physician or a physician | Trained technician<br>supervised by a physician<br>or a physician | Yes | | Interpretation of the Test Results | Yes | Physician or specialist locally | Physician or specialist locally | Physician or specialist<br>locally | Yes | | Examination areas | Yes | Examination of all skin/mucous<br>membrane areas | Examination of all skin areas | forarm and foot | Yes | | Alternative measurement protocols | Yes | Suitable to measure a patient's<br>Pain Tolerance Threshold (PTT) | Suitable to measure a patient's<br>Pain Tolerance Threshold (PTT) | NA | No | | Regulations | NM CPT is under<br>FDA Approval | FDA and CE mark | CE - marked | FDA and CE mark | Yes | {8}------------------------------------------------
Innolitics
Guidance Recalls

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...