The Wrap Accessory Electrodes are intended to be used with legally marketed electrical stimulating devices such as transcutaneous electrical nerve stimulators or powered muscle stimulators. The wrap accessory electrodes will deliver the stimulation signals generated by the stimulator to the body surface with which they are in contact. The wrap accessory electrodes, made up of conductive silicone rubber, can be used for the stimulation on the body areas such as abdomen, neck, shoulder, elbow, leg, Hand, wrist, ankle, knee and foot.
Device Story
Wrap Accessory Electrodes are conductive silicone rubber garments/accessories used with legally marketed TENS or powered muscle stimulators. They function as passive interfaces, delivering electrical stimulation signals from the stimulator to the patient's skin surface. Designed for OTC use, they replace traditional adhesive electrode patches. Available in various configurations (e.g., belt, wrist, elbow, leg, ankle, knee, foot pad, glove, shoulder, neck), they are applied to the corresponding body area by the user. The device does not generate stimulation itself; it acts as a conduit for the stimulator's output. By facilitating muscle stimulation or nerve modulation, the device aims to provide therapeutic benefits consistent with the primary stimulator's indications. Healthcare providers or patients select the appropriate accessory based on the treatment site.
Clinical Evidence
No clinical test data was used to support the decision of substantial equivalence. Bench testing only.
Technological Characteristics
Materials: Conductive silicone rubber. Sensing/Actuation: Passive conductive interface for electrical stimulation. Connectivity: Wired connection to external TENS/EMS stimulator. Standards: ISO 10993-1, ISO 10993-10, ISO 10993-23, IEC 60601-1-11, IEC 60601-2-10, IEC 60601-1-6. Impedance: 12.5 Ω/in². Form factor: Wearable garment/wrap accessories.
Indications for Use
Indicated for use with legally marketed TENS or powered muscle stimulators to deliver electrical stimulation to the body surface of adult patients for stimulation of the abdomen, neck, shoulder, elbow, leg, hand, wrist, ankle, knee, and foot.
Regulatory Classification
Identification
A cutaneous electrode is an electrode that is applied directly to a patient's skin either to record physiological signals (e.g., the electroencephalogram) or to apply electrical stimulation.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: a symbol on the left and the FDA name and title on the right. The symbol on the left is a stylized representation of a human figure, while the text on the right reads "FDA U.S. FOOD & DRUG ADMINISTRATION" in blue letters.
June 10, 2024
Gymmax Technology Shenzhen Co., Ltd. % Salon Chen System engineer IMD Medical & Drug technology service institutions Room 308, Building 11, No. 23 Jinqu Road, Wanjiang District Dongguan, Guangdong 523039 China
Re: K240108
Trade/Device Name: Wrap Accessory Electrodes (GMX-ABSBELT01, GMX-ABSBELT02, GMX-WRIST01, GMX-ELBOW01, GMX-LEG 01, GMX-ANKLE01, GMX-KNEE01, GMX-FOOT PAD 01, GMX-GLOVES 01, SHOULDER01, GMX-NCK01) Regulation Number: 21 CFR 882.1320 Regulation Name: Cutaneous Electrode Regulatory Class: Class II Product Code: GXY Dated: May 13, 2024 Received: May 14, 2024
Dear Salon Chen:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrb/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}------------------------------------------------
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30. Design controls; 21 CFR 820.90. Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely.
Heather L. Dean -S
Heather Dean, PhD Assistant Director DHT5B: Division of Neuromodulation and Rehabilitation Devices
{2}------------------------------------------------
OHT5: Office of Neurological and Physical Medicine Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{3}------------------------------------------------
# Indications for Use
Submission Number (if known)
K240108
Device Name
Wrap Accessory Electrodes (GMX-ABSBELT01、GMX-ABSBELT02、GMX-WRIST01、GMX-ELBOW01、GMX-LEG 01、GMX-ANKLE01、GMX-KNEE01、GMX-FOOT PAD 01、GMX-GLOVES 01、SHOULDER01、GMX-NCK01)
## Indications for Use (Describe)
The Wrap Accessory Electrodes are intended to be used with legally marketed electrical stimulating devices such as transcutaneous electrical nerve stimulators or powered muscle stimulators. The wrap accessory electrodes will deliver the stimulation signals generated by the stimulator to the body surface with which they are in contact. The wrap accessory electrodes, made up of conductive silicone rubber, can be used for the stimulation on the body areas such as abdomen, neck, shoulder, elbow, leg, hand, wrist, ankle, knee and foot .
Type of Use (Select one or both, as applicable)
Prescription Use (Part 21 CFR 801 Subpart D)
X Over-The-Counter Use (21 CFR 801 Subpart C)
## CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
## *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{4}------------------------------------------------
#### 510(k) Summary
### 1. Submitter's Identification:
- > Company Name: Gymmax Technology Shenzhen Co., Ltd.
- > Establishment Registration Number: 3007689716
- > Address:5F 4th Bldg., Heping Complex (4th Bldg. Huimingsheng Technology Park) Fuhai Subdistrict, Baoan District Shenzhen City, Guangdong Province 518103 P.R.C.
- > Phone: +86-755-2912-4050
- A Fax: +86-755-2912-4050
- > Contact Person (Title): Benson Wang (General Manager)
- > E-mail: benson gmx@gq.com
- > Date of Preparation: May 04, 2024
- 2. Common Name and Classification:
- > Regulation Name: Cutaneous Electrode
- > Classification Product Code: GXY
- > Regulation Number: 21 CFR 882.1320
- > Regulatory Class: Class II
- > Review Panel: Neurology
- > Trade Name: Wrap Accessory Electrodes
- Predicate Device Information1: 3.
- > 510(k) Number: K210383
- > Predicate Device Name: Wrap Accessory Electrodes
- > Manufacturer: Hi-Dow International Inc.
- A Intended Use:
The Wrap Accessory Electrodes are intended to be used with legally marketed electrical stimulating devices such as transcutaneous electrical nerve stimulators or powered muscle stimulators. The wrap accessory electrodes will deliver the stimulation signals generated by the stimulator to the body surface with which they are in contact. The wrap accessory electrodes,
{5}------------------------------------------------
made up of conductive silicone rubber, can be used for the stimulation on the body areas such as back, neck, shoulder, elbow, wrist, knee and foot.
- 4. Predicate Device Information2:
- > 510(k) Number: K190617
- > Predicate Device Name: Wrap accessory electrodes
- > Manufacturer: Hi-Dow International Inc.
#### > Intended Use:
Wrap accessory electrodes, available in a glove style, are cutaneous electrodes intended to be used with legally marketed TENS or EMS stimulating devices. The Wrap accessory electrodes are non-sterile reusable OTC conductive garments that are intended to deliver the stimulation signals generated by the stimulator to the body surface with which they are in contact. These body parts can include the hands (gloves) and feet (socks). It is intended for adult patients.
### 5. Application Correspondent
- > Company Name: IMD Medical & Drug technology service institutions
- > Phone: +86-18613190779
- > Fax: +86-755-62809168
- > Contact Person (Title): Salon Chen (System engineer)
- > E-mail: 33999439@gq.com
- > Address: Room 308, Building 11, No. 23 Jinqu Road, Wanjiang District, Dongguan City, Guangdong Province, China
#### 6. Device Description
Wrap Accessory Electrodes with GMX-ABSBELT01、GMX-ABSBELT02、GMX-WRIST01、 GMX-ELBOW01、GMX-LEG 01、GMX-ANKLE01、GMX-KNEE01、GMX-FOOT PAD 01、 GMX-GLOVES 01、SHOULDER01、GMX-NCK01, are intended to be used with legally Marketed transcutaneous electrical nerve stimulators or powered muscle stimulators which could be used in place of traditional cutaneous electrode patches. The different series of model means different placement of the Wrap Accessory Electrodes. The wrap accessory electrodes will deliver the stimulation signals generated by the stimulator to the body surface with which they are in contact.
{6}------------------------------------------------
The wrap accessory electrodes, made up of conductive silicone rubber, can be used for the stimulation on the body areas such as abdomen, neck, shoulder, elbow, leg, Hand, wrist, ankle, knee and foot.
## 7. Indications for Use
The Wrap Accessory Electrodes are intended to be used with legally marketed electrical stimulating devices such as transcutaneous electrical nerve stimulators or powered muscle stimulators. The wrap accessory electrodes will deliver the stimulation signals generated by the stimulator to the body surface with which they are in contact. The wrap accessory electrodes, made up of conductive silicone rubber, can be used for the stimulation on the body areas such as abdomen, neck, shoulder, elbow, leg, Hand, wrist, ankle, knee and foot.
## 8. Comparison to the predicate device
| Elements of<br>Comparison | Proposed Device<br>K240108 | Predicate Device 1<br>K210383 | Predicate Device 2<br>K190617 | Judgment |
|-----------------------------------|-----------------------------------------|-------------------------------|-------------------------------|----------|
| Company<br>Name | Gymmax Technology<br>Shenzhen Co., Ltd. | Hi-Dow International Inc. | Hi-Dow International Inc. | / |
| Device Name | Wrap Accessory<br>Electrodes | Wrap Accessory<br>Electrodes | Wrap Accessory<br>Electrodes | / |
| Classification<br>Product<br>Code | GXY | GXY | GXY | SE |
| Regulation | 21 CFR 882.1320 | 21 CFR 882.1320 | 21 CFR 882.1320 | SE |
| Classification<br>Name | Electrode, Cutaneous | Electrode, Cutaneous | Electrode, Cutaneous | SE |
| Class | 2 | 2 | 2 | SE |
| Prescription<br>or OTC | OTC | OTC | OTC | SE |
### 8.1 Table 1 General Comparison
{7}------------------------------------------------
| Intended Use | The Wrap Accessory Electrodes are intended to be used with legally marketed electrical stimulating devices such as transcutaneous electrical nerve stimulators or powered muscle stimulators. The wrap accessory electrodes will deliver the stimulation signals generated by the stimulator to the body surface with which they are in contact. The wrap accessory electrodes, made up of conductive silicone rubber, can be used for the stimulation on the body areas such as abdomen, neck, shoulder, elbow, leg, Hand, wrist, ankle, knee and foot. | The Wrap Accessory Electrodes are intended to be used with legally marketed electrical stimulating devices such as transcutaneous electrical nerve stimulators or powered muscle stimulators. The wrap accessory electrodes will deliver the stimulation signals generated by the stimulator to the body surface with which they are in contact. The wrap accessory electrodes, made up of conductive silicone rubber, can be used for the stimulation on the body areas such as back, neck, shoulder, elbow, wrist, knee and foot. | Wrap accessory electrodes, available in a glove style or socks style, are cutaneous electrodes intended to be used with legally marketed TENS or EMS stimulating devices. The Wrap accessory electrodes are non-sterile reusable OTC conductive garments that are intended to deliver the stimulation signals generated by the stimulator to the body surface with which they are in contact. These body parts can include the hands (gloves) and feet (socks). It is intended for adult patients. | Note1 |
|--------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------|
|--------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------|
Table 1 General Comparison
Note 1: As shown in the above comparison Table,Proposed Device adds and can be used for the stimulation on the body areas such as abdomen, leg, ankle and foot. Electronic pulses are sent through electrode pads to passively exercise the affected muscle. EMS sends comfortable impulses through your skin that stimulates the nerves in the treatment area, in the same way, TENS or EMS is effect in other areas of the body areas, they are all used to deliver the stimulation signals generated by electrical stimulator to the body surface, the difference does not raise new questions of safety and effectiveness as compared to the predicate devices.
{8}------------------------------------------------
8.2Table 2 Safety factor & Performance Comparison
| Safety factor &<br>Performance | Proposed Device | Predicate Device1 | Predicate Device2 | Judgment |
|--------------------------------------------------------------------|--------------------------------|--------------------------------|--------------------------------|----------|
| impedance per area<br>( $Ω/in.²$ ) as a<br>worst-case<br>condition | 12.5 | / | / | Note 2 |
| Biocompatiblity | Compliance with ISO<br>10993-1 | Compliance with ISO<br>10993-1 | Compliance with ISO<br>10993-1 | SE |
Table 2 Safety factor & Performance Comparison
Note 2: Electrode conductive area and Impedance Parameters Although the subject device has 12.5 Ω per inch, which is different from the predicate device, "Maximum Average Current Density" and "Maximum Peak Power Density" property are similar and do not raise new questions of safety or effectiveness. The standard of "Maximum Average Current Density" refers to "IEC60601-2-10 Clause 201.4.2" and the standard of "Maximum Peak Power Density" refer to "Guidance Document for Powered Muscle Stimulator 510K Section 3." Indicating similar performance and safety between subject device and predicate device. Otherwise, the subject devices are complying with ISO 10993 requirements. So, the differences of the function specifications will not raise any safety or effectiveness issue. For detailed measurement data, please refer to the above table. The subject device is Substantially Equivalent to the predicate device.
Electrical Stimulator System has applied for K232439 with Models: GMX-007 and GMX-006, Electrical Stimulator System with Wrap Accessory Electrodes have been tested with below:
#### 1) Basic Safety and Essential Performance _
IEC 60601-1-11:2020: General Requirements For Basic Safety And Essential Performance -Collateral Standard: Requirements For Medical Electrical Equipment And Medical Electrical Systems Used In The Home Healthcare Environment.
IEC 60601-2-10:2012 +A1:2016+A2:2023medical electrical equipment - part 2-10: particular requirements for the basic safety and essential performance of nerve and muscle stimulators. IEC 60601-1-6 :2010+A1:2013+A2:2020 Medical Electrical Equipment - Part 1-6: General Requirements For Basic Safety And Essential Performance - Collateral Standard: Usability.
2) Biocompatibility testing
{9}------------------------------------------------
ISO 10993-1:2009 Biological evaluation of medical devices-Parts 1: Evaluation and testing.
ISO 10993-10:2021 Biological evaluation of medical devices-Parts 10: Tests for skin sensitization
ISO 10993-23:2021 Biological evaluation of medical devices-Parts 10: Tests for irritation
- 3) Maximum output voltage (volts+/-20%) at 500Ω、Maximum output voltage (volts+/-20%) at 2KΩ、 Maximum output voltage (volts+/-20%) at 10KΩ、Maximum output current (mA+/-20%) at 500Ω、 Maximum output current (mA+/-20%) at 2KΩ、Maximum output current (mA+/-20%) at 10KΩ
| Safety factor &<br>Performance | Proposed Device |
|---------------------------------------------------------|------------------------------------------------------------------------------------|
| Maximum<br>output voltage<br>(volts+/-20%) at<br>$500Ω$ | TENS 1:112<br>TENS 2: 114<br>STIM1:116<br>STIM2:116<br>STIM3:76<br>STIM4: 70 |
| Maximum<br>output voltage<br>(volts+/-20%) at<br>$2kΩ$ | TENS 1: 250<br>TENS 2: 240<br>STIM1: 250<br>STIM2: 250<br>STIM3: 158<br>STIM4: 146 |
| Maximum<br>output voltage<br>(volts+/-20%) at<br>$10kΩ$ | TENS 1: 300<br>TENS 2: 286<br>STIM1: 310<br>STIM2: 308<br>STIM3: 278<br>STIM4: 276 |
| Maximum<br>output current<br>(mA+/-20%) at<br>$500Ω$ | TENS 1: 224<br>TENS 2: 228<br>STIM1: 232<br>STIM2: 232<br>STIM3: 152<br>STIM4: 140 |
| Maximum<br>output current<br>(mA+/-20%) at<br>$2kΩ$ | TENS 1:125<br>TENS 2: 120<br>STIM1: 125<br>STIM2: 125<br>STIM3:79<br>STIM4: 73 |
{10}------------------------------------------------
| Maximum<br>output current<br>(mA+/-20%) at<br>10KΩ | TENS 1: 30<br>TENS 2: 28.6<br>STIM1: 31<br>STIM2:30.8<br>STIM3: 27.8<br>STIM4: 27.6 |
|------------------------------------------------------------|----------------------------------------------------------------------------------------------|
| Pulse period<br>(mSec) | 20~200 |
| Frequency (Hz) | TENS 1: 33<br>TENS 2: 5<br>STIM1:37<br>STIM2:25<br>STIM3: 5<br>STIM4: 50 |
| Maximum Phase<br>charge (µC) at<br>500Ω | TENS 1: 11<br>TENS 2: 11<br>STIM1: 11.4<br>STIM2: 10.6<br>STIM3: 22.8<br>STIM4: 20.4 |
| Maximum<br>charge density<br>(mA/cm²) at<br>500Ω | TENS 1: 0.024<br>TENS 2: 0.01<br>STIM1:0.027<br>STIM2: 0.021<br>STIM3: 0.021<br>STIM4: 0.059 |
| Maximum<br>average power<br>density<br>(mW/cm²) at<br>500Ω | TENS 1: 1.33<br>TENS 2: 0.554<br>STIM1: 1.54<br>STIM2: 1.13<br>STIM3: 0.793<br>STIM4: 2.01 |
# 9. Discussion of Non-clinical Tests Performed for Determination of Substantial Equivalence are as follows:
The proposed device is the same as the predicate device but the two devices are different. A
series of safety and performance tests were conducted on the subject device:
Biocompatibility
Electromagnetic compatibility and electrical safety
Function test with Maximum resistance value、Minimum resistance value、Maximum output
{11}------------------------------------------------
voltage、Maximum output current、Maximum Phase charge、Maximum charge density、 Maximum average power density、hot spot test and impedance per area. All the test results demonstrate Wrap Accessory Electrodes meets the requirements of its pre-defined acceptance criteria and intended uses and is substantially equivalent to the predicate devices.
#### 10. Clinical Tests Performed
No clinical test data was used to support the decision of substantial equivalence.
### 11. Conclusion
The subject devices have all features of the predicate devices. The few differences do not affect the safety and effectiveness of the subject devices.
Thus, the subject devices are substantially equivalent to the predicate devices.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Report an issue or suggest a feature
Spot a broken citation, a wrong extraction, a missing document, or something you wish this app did? Tell us — no account needed.
or drag and drop files here
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.