← Product Code [QXM](/productcode/QXM) · K240037

# Revi™ System (K240037)

_Bluewind Medical , Ltd. · QXM · May 2, 2024 · Gastroenterology, Urology · SESE_

**Canonical URL:** https://fda.innolitics.com/device/K240037

## Device Facts

- **Applicant:** Bluewind Medical , Ltd.
- **Product Code:** [QXM](/productcode/QXM.md)
- **Decision Date:** May 2, 2024
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 876.5305
- **Device Class:** Class 2
- **Review Panel:** Gastroenterology, Urology
- **Attributes:** Therapeutic

## Indications for Use

The Revi System is indicated for the treatment of patients with symptoms of urgency incontinence alone or in combination with urinary urgency.

## Device Story

Revi System is an implantable tibial electrical urinary continence device; treats urge urinary incontinence. System consists of an implanted wireless neurostimulation component and a non-implanted external wearable unit. Wearable unit transmits power wirelessly to the implant; implant delivers electrical pulses to the tibial nerve near the ankle. Stimulation travels along the leg to the sacral plexus and spinal cord; modulates nerve function to relieve symptoms. Device is used by patients; wearable unit provides control. Subject device (v4.3 software) increases treatment program storage capacity from 2 to 4 compared to predicate. Clinical benefit is symptom relief for urinary urgency/incontinence.

## Clinical Evidence

Bench testing only. Testing included stimulation frequency control, charging mode protection, system performance verification, and software/cybersecurity verification. Human factors/usability evaluation confirmed no new critical tasks were introduced.

## Technological Characteristics

Implantable tibial neurostimulator; wireless power transfer from external wearable unit. Software v4.3 supports storage of up to 4 treatment programs. Connectivity via wireless power/data link between wearable and implant.

## Regulatory Identification

An implanted tibial electrical urinary continence device is an implanted prescription device that receives power from a non-implanted external power source to provide electrical stimulation of the tibial nerve in proximity to the ankle. The device is intended for the treatment of overactive bladder related symptoms of urge urinary incontinence, urinary urgency, urinary frequency and nocturia.

## Special Controls

In combination with the general controls of the FD&C Act, an implanted tibial electrical urinary continence device is subject to the following special controls:

## Predicate Devices

- Revi System ([DEN220073](/device/DEN220073.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the symbol of the Department of Health & Human Services on the left and the FDA acronym along with the full name of the agency on the right. The FDA part of the logo is in blue, with the acronym in a square and the full name written out next to it.

May 2, 2024

BlueWind Medical Ltd. Jason Woehrle Director Regulatory Affairs 6 Maskit Street Herzliya, 4614002 ISRAEL

Re: K240037

> Trade/Device Name: Revi™ System Regulation Number: 21 CFR 876.5305 Regulation Name: Implanted Tibial Electrical Urinary Continence Device Regulatory Class: II Product Code: QXM Dated: January 4, 2024 Received: January 5, 2024

Dear Jason Woehrle:

We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{1}------------------------------------------------

Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).

Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30. Design controls; 21 CFR 820.90. Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See 

{2}------------------------------------------------

the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

# Sharon M. Andrews -S

Sharon M. Andrews Director DHT3B: Division of Reproductive, Gynecology and Urology Devices OHT3: Office of Gastrorenal, ObGyn, General Hospital, and Urology Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{3}------------------------------------------------

# Indications for Use

Submission Number (if known)

K240037

Device Name

Revi™ System

Indications for Use (Describe)

The Revi System is indicated for the treatment of patients with symptoms of urgency incontinence alone or in combination with urinary urgency.

Type of Use (Select one or both, as applicable)

Prescription Use (Part 21 CFR 801 Subpart D)

Over-The-Counter Use (21 CFR 801 Subpart C)

### CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{4}------------------------------------------------

Image /page/4/Picture/0 description: The image shows the logo for Blue Wind Medical. The logo consists of the words "Blue Wind" stacked on top of each other, with the word "MEDICAL" underneath. To the right of the text is a curved line of blue circles that get progressively smaller.

## 510(k) Summary

- I. Company: BlueWind Medical Ltd. 6 Maskit Street Herzliya, 4614002 Israel
	- Jason Woehrle (Primary) Contact: Director, Regulatory Affairs Telephone Number: (1) 952-240-2229 Email:Jason.Woehrle@bluewindmedical.com

Roni Diaz (Alternate) VP Regulatorv and Clinical Affairs Telephone Number: (1) 888-715-2080 Email: roni.diaz(a)bluewindmedical.com

Date Prepared: May 1, 2024

- II. Proprietary Trade Name: Revi™ System
- III. Common or Usual Name: Implantable tibial electrical urinary continence device
- Classification Name: Implantable tibial electrical urinary continence device (21 CFR IV. 876.5305)
- Classification: Class II V.
- VI. Product Code: QXM

#### VII. Product Description:

The Revi™ System is an implanted tibial electrical urinary continence device that wirelessly receives power from a non-implanted external wearable unit to provide electrical stimulation of the tibial nerve in proximity to the ankle. The device is intended for the treatment of urge urinary incontinence, alone or in combination with urinary urgency. The implantable device is implanted in the vicinity of the tibial neurovascular bundle. The treatment effect of the system is achieved by the implantable wireless neurostimulation component, which sends pulses to the tibial nerve when energized by the wearable unit transmitted power. The electrical pulses stimulate the nerve along the leg, reaching the sacral plexus and entering the spinal cord. This stimulation has the power to modulate nerve function, relieving symptoms.

#### VIII. Indications for Use

The Revi System is indicated for the treatment of patients with symptoms of urgency incontinence alone or in combination with urinary urgency.

{5}------------------------------------------------

Image /page/5/Picture/0 description: The image shows the logo for Blue Wind Medical. The logo features the words "Blue Wind" in a bold, sans-serif font, with "Blue" stacked above "Wind". Below "Wind" and slightly offset to the left is the word "MEDICAL" in a smaller font. Above and below the words "Blue Wind" are a series of blue dots arranged in a curved line, creating a visual representation of wind or movement.

| Description             | Subject Device<br>Revi System with v4.3 software<br>(K240037)                                                                                                    | Predicate Device<br>Revi<br>System with v4.1.2.5<br>software<br>(DEN220073)                                                                                            |
|-------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Indications for Use     | The Revi System is indicated for the<br>treatment of patients with symptoms<br>of urgency incontinence alone or in<br>combination with urinary urgency.          | The Revi System is<br>indicated for the<br>treatment of patients<br>with symptoms of<br>urgency incontinence<br>alone or in combination<br>with urinary urgency.       |
| Principles of Operation | Implanted device received power<br>from non- implanted external<br>wearable device providing electrical<br>stimulation to tibial nerve in<br>proximity to ankle. | Implanted device<br>received power from<br>non- implanted external<br>wearable device<br>providing electrical<br>stimulation to tibial nerve<br>in proximity to ankle. |
| Treatment Programs      | Ability to store up to 4 treatment<br>programs                                                                                                                   | Ability to store up to 2<br>treatment programs                                                                                                                         |

#### Summary of the Technological Characteristics IX.

The modified Revi System with v4.3 software is identical to the predicate Revi System with v4.1.2.5 software in terms of indications for use, intended use, operational principle, design and is equivalent in terms of software function. The differences between the modified Revi System and predicate device do not raise different questions of safety and effectiveness.

#### Identification of Legally Marketed Devices X.

Revi System (DEN220073) Note: The predicate device has not been subject to a design-related recall.

- XI. Discussion of the Performance Testing
Non-clinical benchtop testing was conducted to evaluate the performance characteristics of the modified Revi System.

{6}------------------------------------------------

Image /page/6/Picture/1 description: The image shows the logo for Blue Wind Medical. The logo features the words "Blue Wind" in a bold, sans-serif font, with the word "Blue" in a darker shade than the word "Wind." Below "Blue Wind" is the word "MEDICAL" in a smaller, sans-serif font. To the right of the text is a curved line of blue circles, starting with larger circles and gradually decreasing in size.

Verification testing to ensure identical specification to predicate included the following:

- stimulation frequency control to ensure range and accuracy compliance with ● specification
- charging mode protection to ensure protection operates in compliance with . specification
- system performance to ensure system performance in compliance with technical . specification
- software verification to verify software requirements, including cybersecurity ● testing

A human factors/usability evaluation of the Revi™ System with v4.3 software and a gap analysis with the predicate device found that no new critical tasks were introduced by the subject device.

#### Conclusion XII.

The Revi™ System with v4.3 software has been shown through testing and comparison to be substantially equivalent to the identified predicate device.

---

**Source:** [https://fda.innolitics.com/device/K240037](https://fda.innolitics.com/device/K240037)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
