The UniSpace® Stand-Alone C Cage is a standalone anterior cervical interbody fusion device indicated for skeletally mature patients with cervical disc disease at one level from the C2-C3 disc to the C7-T1 disc. Cervical disc disease is defined as intractable radiculopathy and/or myelopathy with herniated disc and/or osteophyte formation on posterior vertebral endplates producing symptomatic nerve root and/or spinal cord compression confirmed by radiographic studies. The UniSpace® Stand-Alone C Cage implants are to be used with either autogenous bone graff and/or allogenic bone graft comprised of cancellous and/or corticocancellous bone graft and implanted via an open, anterior approach. The Cages of the UniSpace® Stand-Alone C Cage must be used with the two internal fixation screws provided. The Cages with ≥ 20° lordosis must be used with an additional supplemental fixation system that has been cleared by the FDA for use in the cervical spine, in addition to the two integrated fixation screws provided. This cervical device is to be used in patients who have had six weeks of non-operative treatment.
Device Story
Standalone anterior cervical interbody fusion device; provides spinal column stability. Inputs: autogenous or allogenic bone graft. Components: 3D-printed cages (additive manufacturing), variable screws, locking plates. Operation: implanted via open anterior approach; screws provide fixation; locking plate/screw mechanism prevents screw back-out. Used in clinical/surgical settings by surgeons. Output: structural support for fusion. Benefits: stabilization of cervical spine segments; relief of nerve root/spinal cord compression. Cages with ≥ 20° lordosis require supplemental fixation.
Clinical Evidence
Bench testing only. No clinical data. Testing included static/dynamic compression, torsion, and compression-shear (ASTM F2077); subsidence (ASTM F2267); and axial pullout, torsional properties, driving torque, and locking mechanism strength (ASTM F543).
Indicated for skeletally mature patients with cervical disc disease (intractable radiculopathy/myelopathy, herniated disc, osteophyte formation, nerve root/spinal cord compression) at one level from C2-C3 to C7-T1, following six weeks of failed non-operative treatment.
Regulatory Classification
Identification
An intervertebral body fusion device is an implanted single or multiple component spinal device made from a variety of materials, including titanium and polymers. The device is inserted into the intervertebral body space of the cervical or lumbosacral spine, and is intended for intervertebral body fusion.
Special Controls
*Classification.* (1) Class II (special controls) for intervertebral body fusion devices that contain bone grafting material. The special control is the FDA guidance document entitled “Class II Special Controls Guidance Document: Intervertebral Body Fusion Device.” See § 888.1(e) for the availability of this guidance document.(2) Class III (premarket approval) for intervertebral body fusion devices that include any therapeutic biologic (e.g., bone morphogenic protein). Intervertebral body fusion devices that contain any therapeutic biologic require premarket approval.
(c)
*Date premarket approval application (PMA) or notice of product development protocol (PDP) is required.* Devices described in paragraph (b)(2) of this section shall have an approved PMA or a declared completed PDP in effect before being placed in commercial distribution.
{0}------------------------------------------------
April 19, 2024
Image /page/0/Picture/1 description: The image shows the logos of the Department of Health & Human Services and the Food and Drug Administration (FDA). The Department of Health & Human Services logo is on the left and features a stylized image of a human figure. The FDA logo is on the right and features the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue text.
Innosys Co., Ltd. Chae-Hyeon Kim Regulatory Affairs Specialist 20, Sandan-ro 76beon-gil(Rd) Uijeongbu-si. Gyeonggi-do 11781 South Korea
Re: K234119
Trade/Device Name: UniSpace® Stand-Alone C Cage Regulation Number: 21 CFR 888.3080 Regulation Name: Intervertebral Body Fusion Device Regulatory Class: Class II Product Code: OVE Dated: March 28, 2024 Received: March 28, 2024
Dear Chae-Hyeon Kim:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
{1}------------------------------------------------
Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
# Katherine D. Kavlock -S
for
Brent Showalter, Ph.D. Assistant Director DHT6B: Division of Spinal Devices OHT6: Office of Orthopedic Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
{2}------------------------------------------------
## Indications for Use
Submission Number (if known)
K234119
Device Name
UniSpace® Stand-Alone C Cage
#### Indications for Use (Describe)
The UniSpace® Stand-Alone C Cage is a standalone anterior cervical interbody fusion device indicated for skeletally mature patients with cervical disc disease at one level from the C2-C3 disc to the C7-T1 disc. Cervical disc disease is defined as intractable radiculopathy and/or myelopathy with herniated disc and/or osteophyte formation on posterior vertebral endplates producing symptomatic nerve root and/or spinal cord compression confirmed by radiographic studies. The UniSpace® Stand-Alone C Cage implants are to be used with either autogenous bone graff and/or allogenic bone graft comprised of cancellous and/or corticocancellous bone graft and implanted via an open, anterior approach. The Cages of the UniSpace® Stand-Alone C Cage must be used with the two internal fixation screws provided. The Cages with ≥ 20° lordosis must be used with an additional supplemental fixation system that has been cleared by the FDA for use in the cervical spine, in addition to the two integrated fixation screws provided.
This cervical device is to be used in patients who have had six weeks of non-operative treatment.
Type of Use (Select one or both, as applicable)
Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
## 510(k) Summary
| Manufacturer: | Innosys Co., Ltd.<br>20, Sandan-ro 76beon-gil(Rd), Uijeongbu-si, Gyeonggi-do,<br>11781, Korea |
|----------------------|-----------------------------------------------------------------------------------------------|
| Sponsor: | Innosys Co., Ltd.<br>20, Sandan-ro 76beon-gil(Rd), Uijeongbu-si, Gyeonggi-do,<br>11781, Korea |
| Sponsor Contact: | Chae-Hyeon Kim, Regulatory Affairs Specialist<br>+82 31 860 6836<br>ch.kim@inno-sys.net |
| Date Prepared: | March 28, 2024 |
| Device Trade Name: | UniSpace® Stand-Alone C Cage |
| Classification Name: | Intervertebral Fusion Device with Bone Graft, Cervical<br>, per 21 CFR 888.3080 |
| Common Name: | Intervertebral Body Fusion Device, IBF Device |
| Product Code: | OVE |
| Predicate Devices: | UniSpace™ SA Cervical Cage (K213791) |
| Reference Device: | Aspiron™ S ACP System (K220147)<br>Velofix™ SA Cervical Cage (K172424) |
## Description of Device:
The UniSpace® Stand-Alone C Cage is product for cervical spinal column stability. The implants of the UniSpace® Stand-Alone C Cage consist of the Cages manufactured through additive manufacturing (ASTM F3001) and the machined Variable screws and Locking plates (ASTM F136). The Locking plate consists of a Blocking plate and a Locking screw and is an additional device for the anti-backout of the Variable screw. This optional implant can help users who feel uncomfortable with the primary self-locking mechanism or when rescue screws are utilized. The Cages and Variable screws are provided in a variety of sizes to accommodate each
Image /page/3/Picture/6 description: The image shows the logo for Innosys. The word "innosys" is written in a bold, sans-serif font. The "inno" part of the word is in blue, while the "sys" part is in orange. The logo is simple and modern.
{4}------------------------------------------------
patient's individual clinical case. The Cages are designed to be used with either autogenous bone graft and/or allogenic bone graft comprised of cancellous and/or corticocancellous bone graft. The Cages of the UniSpace® Stand-Alone C Cage are provided as a sterile pack. The UniSpace® Stand-Alone C Cage is implanted by using the instruments manufactured from stainless steel material that conform to ASTM F899.
### Indications For Use:
The UniSpace® Stand-Alone C Cage is a standalone anterior cervical interbody fusion device indicated for skeletally mature patients with cervical disc disease at one level from the C2-C3 disc to the C7-T1 disc. Cervical disc disease is defined as intractable radiculopathy and/or myelopathy with herniated disc and/or osteophyte formation on posterior vertebral endplates producing symptomatic nerve root and/or spinal cord compression confirmed by radiographic studies. The UniSpace® Stand-Alone C Cage implants are to be used with either autogenous bone graft and/or allogenic bone graft comprised of cancellous and/or corticocancellous bone graft and implanted via an open, anterior approach. The Cages of the UniSpace® Stand-Alone C Cage must be used with the two internal fixation screws provided. The Cages with ≥ 20° lordosis must be used with an additional supplemental fixation system that has been cleared by the FDA for use in the cervical spine, in addition to the two integrated fixation screws provided. This cervical device is to be used in patients who have had six weeks of non-operative treatment.
## Substantial Equivalence:
UniSpace® Stand-Alone C Cage is substantially equivalent to UniSpace™ SA Cervical Cage (K213791) in design, material, mechanical performance, function and intended use.
The mechanical performance of UniSpace® Stand-Alone C Caqe falls within the acceptance criteria which have been established from the predicate devices.
#### 1. Comparison Technological Characteristics
The predicate and proposed devices have similar intended use and basic fundamental scientific technology and share the following similarities;
- The same indications for use
- The similar design features
- The same materials
- The equivalent mechanical performance
Image /page/4/Picture/14 description: The image shows the logo for Innosys. The word "innosys" is written in a sans-serif font. The "inno" portion of the word is in blue, while the "sys" portion is in orange. There is a small blue circle above the "i" in "inno."
{5}------------------------------------------------
## 2. Performance Testing
The UniSpace® Stand-Alone C Cage was tested in a non-clinical setting (bench testing) to assess that no new safety and effectiveness issues were raised with this device. The testing meets all acceptance criteria and verifies that performance of the UniSpace® Stand-Alone C Cage is substantially equivalent to the predicate devices and reference devices.
The following tests were performed:
- 1) UniSpace® Stand-Alone C Cage
- (1) Static compression test according to ASTM F2077
- (2) Static torsion test according to ASTM F2077
- (3) Static compression shear to ASTM F2077
- (4) Dynamic compression test to ASTM F2077
- (5) Dynamic torsion test to ASTM F2077
- (6) Dynamic compression shear to ASTM F2077
- (7) Subsidence test to ASTM F2267
- (8) Axial pullout test to ASTM F543
- (9) Torsional properties test to ASTM F543
- (10) Driving Torque test to ASTM F543
- (11) Locking band strength test to ASTM F04-25-02-02, ASTM F543
- (12) Locking plate strength test to ASTM F04-25-02-02, ASTM F543
- (13) Torsional strength test for Locking screw to ASTM F543
#### 3. Conclusion
The data and information provided in this submission support the conclusion that the UniSpace® Stand-Alone C Cage is substantially equivalent to its predicate devices with respect to indications for use and technological characteristics.
Image /page/5/Picture/20 description: The image shows the logo for "innosys". The word "innosys" is written in a sans-serif font, with the "inno" part in blue and the "sys" part in orange. The "o" in "inno" is slightly larger and positioned above the rest of the word.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.