K234110 · Belun Technology Company Limited · DQA · Oct 11, 2024 · Cardiovascular
Device Facts
Record ID
K234110
Device Name
Belun Ring BLR-200 (BLR-200)
Applicant
Belun Technology Company Limited
Product Code
DQA · Cardiovascular
Decision Date
Oct 11, 2024
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 870.2700
Device Class
Class 2
Indications for Use
Belun Ring BLR-200 is a wireless, non-invasive and stand-alone pulse oximeter intended to be used for spot-checking and/or continuous data collection and recording of oxygen saturation of arterial hemoglobin (SpO2) and the pulse rate of adult patients through index finger in hospital and home environment for up to ten hours, during no motion conditions, and for patients who are well or poorly perfused. It is not intended for single-use and out-of-hospital transport use and does not have alarms.
Device Story
Wireless, non-invasive, stand-alone pulse oximeter worn on index finger; collects SpO2 and pulse rate data. Device consists of a ring and host program; ring performs signal acquisition, data processing, parameter calculation, and storage; transfers data via Bluetooth Low Energy to host program. Host program displays data as text and graphs for user interpretation. Used in home or hospital environments by patients or clinicians. Provides spot-check or continuous monitoring (up to 10 hours). Output assists in health status assessment. Benefits include convenient, comfortable, and accurate physiological monitoring without alarms.
Clinical Evidence
Bench testing only. SpO2 and pulse rate accuracy tested against functional testers under low perfusion and motion conditions. Accuracy also validated on healthy subjects compared to predicate device during hypoxia tests. Compliance with IEC 60601-1, IEC 60601-1-2, IEC 60601-1-11, and ISO 80601-2-61.
Technological Characteristics
Wireless pulse oximeter; Bluetooth Low Energy connectivity. Materials compliant with ISO 10993-1, -5, -10. Battery compliant with IEC 62133. Software includes embedded firmware in ring and host management program. Modular design with proprietary communication protocol.
Indications for Use
Indicated for adult patients, including those who are well or poorly perfused, for spot-checking and/or continuous monitoring of SpO2 and pulse rate in home or hospital settings. Not for single-use or out-of-hospital transport.
Regulatory Classification
Identification
An oximeter is a device used to transmit radiation at a known wavelength(s) through blood and to measure the blood oxygen saturation based on the amount of reflected or scattered radiation. It may be used alone or in conjunction with a fiberoptic oximeter catheter.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
October 11, 2024
Belun Technology Company Limited Lap Wai Lydia Leung Chief Executive Officer Flat/Rm 225B, 2/F, Building 1W, Phase One Hong Kong Science Park, Pak Shek Kok, N.T. Hong Kong. Hong Kong
Re: K234110
Trade/Device Name: Belun Ring BLR-200 (BLR-200) Regulation Number: 21 CFR 870.2700 Regulation Name: Oximeter Regulatory Class: Class II Product Code: DOA Dated: March 11, 2024 Received: March 11, 2024
Dear Lap Wai Lydia Leung:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}------------------------------------------------
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30. Design controls; 21 CFR 820.90. Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-device-advicecomprehensive-regulatory-assistance/unique-device-identification-system-udi-system.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatory
{2}------------------------------------------------
assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
# James J. Lee -S
for Bradley Quinn Assistant Director DHT1C: Division of Anesthesia, Respiratory, and Sleep Devices OHT1: Office of Ophthalmic, Anesthesia, Respiratory, ENT, and Dental Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{3}------------------------------------------------
### Indications for Use
Submission Number (if known)
K234110
Device Name
Belun Ring BLR-200 (BLR-200)
#### Indications for Use (Describe)
Belun Ring BLR-200 is a wireless, non-invasive and stand-alone pulse oximeter intended to be used for spot-checking and/or continuous data collection and recording of oxygen saturation of arterial hemoglobin (SpO2) and the pulse rate of adult patients through index finger in home environment for up to ten hours, during no motion and motion conditions, and for patients who are well or poorly perfused. It is not intended for single-use and out-of-hospital transport use and does not have alarms.
Type of Use (Select one or both, as applicable)
> Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{4}------------------------------------------------
| 510(k) #: | K234110 | 510(k) Summary | Prepared on: | 2024-09-05 |
|---------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------|----------------|--------------|---------------------|
| Contact Details | | | | 21 CFR 807.92(a)(1) |
| Applicant Name | Belun Technology Company Limited | | | |
| Applicant Address | Flat/Rm 225B, 2/F, Building 1W, Phase One Hong Kong Science Park,<br>Pak Shek Kok N.T. Hong Kong | | | |
| Applicant Contact Telephone | +85237065640 | | | |
| Applicant Contact | Dr. Lap Wai Lydia Leung | | | |
| Applicant Contact Email | belun_reg@beluntech.com | | | |
| Device Name | | | | 21 CFR 807.92(a)(2) |
| Device Trade Name | Belun Ring BLR-200 (BLR-200) | | | |
| Common Name | Oximeter | | | |
| Classification Name | Oximeter | | | |
| Regulation Number | 870.2700 | | | |
| Product Code | DQA | | | |
| Legally Marketed Predicate Devices | | | | 21 CFR 807.92(a)(3) |
| Predicate # | Predicate Trade Name (Primary Predicate is listed first) | | Product Code | |
| K211407 | Belun Ring BLR-100X | | DQA | |
| K221361 | circul™ pro Ring | | DQA | |
| Device Description Summary | | | | 21 CFR 807.92(a)(4) |
| Belun Ring BLR-200 consists of two parts: a Ring, and a host program. The Ring, which has smooth and light design and is easy to wear | | | | |
Belun Ring BLR-200 consists of two parts: a Ring, which has smooth and light design and iight design and is easy to wear and take off, is intended to be worn on the bottom of index finger providing comfortable and accurate measurements. To make the Optical module appropriately contact with the soft part of the Ring (hereinafter referred to as "the Ring arm") is designed to be changeable for fitting different sizes of fing transfers the collected data to host via Bluetooth low power technology. The host program translates the collected data into text and graph which can be easily interpreted by the user.
The system consists of two main platforms, namely Ring and host. It includes one embedded software and one host program, namely Ring firmware embedded in Ring Management (BRM) executed in host respectively. The Ring is responsible for signal acquisition, data processing, parameters calculation (SpO2/PR algorithm), sensor interfacing data storage. The host is for data display, data export and user interface. The system is modularized, and the communication protocol is proprietary. The system is secured with cybersecurity measures.
## Intended Use/Indications for Use
Belun Ring BLR-200 is a wireless, non-invasive and stand-alone pulse oximeter intended to be used for spot-checking and/or continuous data collection and recording of oxygen saturation of arterial hemoglobin (SpO2) and the pulse rate of adult patients through index finger in hospital and home environment for up to ten hours, during no motion conditions, and for patients who are well or poorly perfused. It is not intended for single-use and out-of-hospital transport use and does not have alarms.
## Indications for Use Comparison
21 CFR 807.92(a)(5)
21 CFR 807.92(a)(5)
{5}------------------------------------------------
The difference of intended use / indications for use of the proposed and predicate devices:
- The proposed device has an additional option of spot-checking display of measurement data from Ring on host program via Bluetooth communication, while the predicate device does not have, and the real-time measurement data displaying on host device via Bluetooth communication.
Bench testing was performed to support the substantial equivalence.
### Technological Comparison
### 21 CFR 807.92(a)(6)
The difference of intended use / indications for use of the proposed and predicate devices:
- The proposed device has an additional option of spot-checking display of measurement data from Ring on host program via Bluetooth communication, while the predicate device does not have and the spot-check measurement data displaying on mobile application via Bluetooth communication.
#### Non-Clinical and/or Clinical Tests Summary & Conclusions 21 CFR 807.92(b)
The proposed device Belun Ring BLR-200 is in compliance with both mandatory and voluntary standards and testings, including:
Electrical, EMC, Mechanical and Thermal Safety Testing:
- IEC 60601-1 Edition 3.2 2020-08 CONSOLIDATED VERSION - Medical electrical equipment - Part 1: General requirements for basic safety and essential performance
- IEC 60601-1-2 Edition 4.1 2020-09 CONSOLIDATED VERSION - Medical electrical equipment - Part 1-2: General requirements for basic safety and essential performance - Collateral Standard: Electromagnetic disturbances - Requirements and tests
- IEC 60601-1-11 Edition 2.1 2020-07 CONSOLIDATED VERSION - Medical electrical equipment - Part 1-11: General requirements for basic safety and essential performance - Collateral Requirements for medical electrical equipment and medical electrical systems used in the home healthcare environment
- ISO 80601-2-61 Second edition 2017-12 (Corrected version 2018-02) - Medical electrical equipment -- Part 2-61: Particular requirements for basic safety and essential performance of pulse oximeter equipment
- BQB Bluetooth qualification
- FCC Part 15B & FCC Part 18 certification
Battery Safety Testing:
The battery of the proposed device is identical to that of predicate with the standard-IEC 62133 Edition 2.0 2012-12
Biocompatibility Testing:
The patient intact skin contacting materials of the subject device are ldentical as those of predicate with the standards, including:
- ISO 10993-1:2018, Biological evaluation of medical devices -- Part 1: Evaluation and testing within a risk management process
- ISO 10993-5:2009 Biological evaluation of medical devices -- Part 5: Tests for in vitro cytotoxicity
- ISO 10993-10:2010 Biological evaluation of medical devices -- Part 10: Tests for irritation and skin sensitization
Software and Cybersecurity Testing:
The Software Validation is in compliance with FDA Guidance for the Content of Premarket Submissions for Software in Medical Devices. The Management of Cybersecurity is in compliance for the Content of Premarket Submissions for Management of Cybersecurity in Medical Devices.
Bench Testina:
The SpO2 and pulse rate accuracy of proposed device under low perfusion and motion conditions have been tested against functional tester. The SpO2 and pulse rate accuracy of proposed device have also been tested on healthy subjects and compared with the predicate device in hypoxia tests.
As the proposed devices use the same measurement technology as the previous clearance K21 1407, no additional clinical testing was required to support the substantial equivalence.
Based on the tests and information provided in this submission, the proposed device, Belun Ring BLR-200, is considered to be substantially equivalent to the predicate device.