← Product Code [KLA](/productcode/KLA) · K234107

# Solar™ Anorectal Manometry Catheter (Model Number: K1210AC-L-2212) (K234107)

_Laborie Medical Technologies, Corp. · KLA · Apr 23, 2024 · Gastroenterology, Urology · SESE_

**Canonical URL:** https://fda.innolitics.com/device/K234107

## Device Facts

- **Applicant:** Laborie Medical Technologies, Corp.
- **Product Code:** [KLA](/productcode/KLA.md)
- **Decision Date:** Apr 23, 2024
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 876.1725
- **Device Class:** Class 2
- **Review Panel:** Gastroenterology, Urology

## Indications for Use

The Solar anorectal manometry catheter is a non-sterile, single patient use catheter for use on patient requiring anorectal manometry testing.

## Device Story

High-resolution anorectal manometry (HRAM) catheter; disposable, non-sterile, single-patient use. Device features 10 air-charged sensing balloons and one distal rectal balloon. Operated by inserting into the anorectal anatomy; connects to a reusable Solar Anometry Charger. Charger utilizes a piston-and-cylinder mechanism to inject air into balloons; internal transducers measure pressure. Output provides clinicians with anorectal pressure data, sensation, and reflex assessments (e.g., recto anal inhibitory reflex). Used in clinical settings to diagnose anorectal motility disorders; assists in clinical decision-making regarding patient gastrointestinal function.

## Clinical Evidence

Bench testing only. Testing included design verification (visual, dimensional, functional, destructive) after accelerated aging and transit simulation, biocompatibility (cytotoxicity, irritation, sensitization), bioburden, and firmware compatibility testing. Formative usability evaluation conducted per EN IEC 62366-1.

## Technological Characteristics

HRAM catheter with 10 air-charged sensing balloons and 1 rectal balloon. Connects to external pressure measurement system (charger). Uses pneumatic sensing principle with piston-and-cylinder air injection. Disposable, non-sterile. Firmware used for compatibility with Solar GI system.

## Regulatory Identification

A gastrointestinal motility monitoring system is a device used to measure peristalic activity or pressure in the stomach or esophagus by means of a probe with transducers that is introduced through the mouth into the gastrointestinal tract. The device may include signal conditioning, amplifying, and recording equipment. This generic type of device includes the esophageal motility monitor and tube, the gastrointestinal motility (electrical) system, and certain accessories, such as a pressure transducer, amplifier, and external recorder.

## Predicate Devices

- Latitude Anorectal Pressure Catheter (Model: GIM-6000A) ([K022023](/device/K022023.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: a symbol on the left and the FDA name on the right. The symbol on the left is a stylized image of a human figure, while the FDA name on the right is written in blue letters. The words "U.S. FOOD & DRUG ADMINISTRATION" are written in a clear, sans-serif font.

April 23, 2024

Laborie Medical Technologies Corp. Arpit Sisodiya Sr. Regulatory Affairs Specialist 180 International Drive Portsmouth, New Hampshire 03801

# Re: K234107

Trade/Device Name: Solar™ Anorectal Manometry Catheter (Model Number: K1210AC-L-2212) Regulation Number: 21 CFR 876.1725 Regulation Name: Gastrointestinal Motility Monitoring System Regulatory Class: Class II Product Code: KLA Dated: March 21, 2024 Received: March 21, 2024

Dear Arpit Sisodiya:

We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device"

{1}------------------------------------------------

(https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).

Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30. Design controls; 21 CFR 820.90. Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review. the OS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

> Sincerely. Anthony Lee -S

Anthony C. Lee, Ph.D., M.B.A. Assistant Director DHT3A: Division of Renal, Gastrointestinal, Obesity and Transplant Devices OHT3: Office of Gastrorenal, ObGyn,

{2}------------------------------------------------

General Hospital, and Urology Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{3}------------------------------------------------

# Indications for Use

Form Approved: OMB No. 0910-0120 Expiration Date: 07/31/2026 See PRA Statement below.

Submission Number (if known)

K234107

Device Name

Solar™ Anorectal Manometry Catheter (Model Number: K1210AC-L-2212)

Indications for Use (Describe)

The Solar anorectal manometry catheter is a non-sterile, single patient use catheter for use on patient requiring anorectal manometry testing.

Type of Use (Select one or both, as applicable)

Prescription Use (Part 21 CFR 801 Subpart D)

Over-The-Counter Use (21 CFR 801 Subpart C)

# CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{4}------------------------------------------------

Image /page/4/Picture/1 description: The image contains the word "Laborie" in a teal, sans-serif font. The word is positioned horizontally and appears to be the primary focus of the image. To the right of the word, there is a light green geometric shape, possibly a square or rectangle, partially visible in the upper right corner of the image.

# 510(k) Summary

| Applicant:      | Laborie Medical Technologies Corp.<br>180 International Drive,<br>Portsmouth, NH 03801                       |
|-----------------|--------------------------------------------------------------------------------------------------------------|
| Contact Person: | Arpit Sisodiya<br>Sr. Regulatory Affairs Specialist<br>Phone: (313) 420-8956<br>Email: asisodiya@laborie.com |

### Device Name

| Device Trade Name:     | Solar™ Anorectal Manometry Catheter (K1210AC-L-2212) |
|------------------------|------------------------------------------------------|
| Common Name:           | Monitor, Anorectal Motility, And Tube                |
| Classification Name:   | Gastrointestinal Motility Monitoring System          |
| Regulation Number:     | 21 CFR 876.1725                                      |
| Product Code:          | KLA                                                  |
| Device Classification: | Class II                                             |

### Legally Marketed Predicate Device

| Predicate 510(k) Number: | K022023                                                 |
|--------------------------|---------------------------------------------------------|
| Predicate Trade Name:    | Latitude Anorectal Pressure Catheter (Model: GIM-6000A) |
| Product Code:            | KLA                                                     |

#### Device Description Summary

The Solar™ Anorectal Manometry Catheter is a high-resolution anorectal manometry (HRAM) catheter. It is a diagnostic device used to assess the function of anal and rectum muscles. The device consists of 10 air-charged sensing balloons which can sense pressures during anorectal manometry, along with a distal balloon used for the assessment of recto anal inhibitory reflex (RAIR). The catheter is disposable and non-sterile. Through manual insertion in the device assesses pressures within the anorectal anatomy as well as assessing sensation and reflexes.

This product is non-sterile, single patient use catheter intended for use with a reusable Solar™ Anometry Charger on patients requiring lower gastrointestinal pressure measurements.

### Intended Use/Indications for Use

The Solar anorectal manometry catheter is a non-sterile, single patient use on patient requiring anorectal manometry testing.

#### Indications for Use Comparison

The indications for use of the subject device (Solar™ Anorectal Manometry Catheter) and the predicate device (Latitude anorectal pressure catheter) are the same. The Solar™ Anorectal Manometry Catheter for use on patient 

{5}------------------------------------------------

requiring anorectal manometry testing. Similarly, the predicate, Latitude anorectal pressure catheter (GIM-6000A) is indicated for use in anorectal pressure studies.

### Technological Comparison

The basic technology for the subject device (Solar™ Anometry Catheter) and the predicate device (Latitude anorectal pressure catheter) is the same. There are 10 sensing balloons and one rectal balloon on the Solar™ Anorectal Manometry Catheter, whereas the Latitude anorectal pressure catheter has 4 sensing balloons and one rectal balloon. Both the Solar™ Anorectal Manometry Catheters and Latitude anorectal pressure catheters connect to ancillary pressure measurement systems (charger) that use a piston-and-cylinder charging mechanism to inject a small amount of air into the sensing balloons and internal transducers to read the pressure. The Latitude catheter uses individual charging cables for each sensing balloon while the Solar™ Anorectal Manometry Charger has a single catheter connection and charges all 10 balloons simultaneously.

# Non-Clinical and/or Clinical Tests Summary and Conclusions

A series of nonclinical bench performance testing was conducted on the Solar™ Anorectal Manometry Catheter and Solar™ Anorectal Manometry charger to demonstrate Substantial Equivalence. Following tests were performed on Solar™ Anorectal Manometry Catheter to demonstrate the Substantial Equivalence:

- Design Verification testing (visual, dimensional, performance/ functional, destructive) after 6 Months Accelerated Aging
- Design Verification testing (visual, dimensional, performance/ functional, destructive) after transit simulation
- Biocompatibility (Cytotoxicity, irritation, sensitization)
- Bioburden

Following tests were performed on Solar™ Anorectal Manometry Charger:

- . Functionality Testing
- Biological Safety Assessment
- Distribution Testing
- . Firmware Testing (Compatibility with the Solar GI system)

Additionally, a formative evaluation was conducted to assess the usability of the device by applying evaluation process specified in EN IEC 62366-1, Medical devices-Part 1: Application of usability engineering to medical devices.

# Conclusion

Performance testing was conducted on all key performance attributes of the device. All samples met their acceptance criteria, demonstrating that when manufactured to specification, the device functions as intended and can be found substantially equivalent to the predicate device.

---

**Source:** [https://fda.innolitics.com/device/K234107](https://fda.innolitics.com/device/K234107)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com