Maverick Diagnostic System TC1000; Maverick Test Panel A0.B0
Applicant
Genalyte, Inc.
Product Code
JLW · Clinical Chemistry
Decision Date
Jul 22, 2024
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 862.1690
Device Class
Class 2
Indications for Use
The Maverick Test Panel A0.B0 is an immunoassay for the quantitative determination of human thyroid stimulating hormone (thyrotropin, TSH) in human serum and K2EDTA plasma on the Maverick Diagnostic System TC1000. Measurements of thyroid stimulating hormone produced by the anterior pituitary are used in the diagnosis of thyroid or pituitary disorders. The Maverick Diagnostic System TC1000 is an automated immunoassay analyzer intended for in vitro diagnostic use to determine analytes in a clinical laboratory. The system’s assay applications utilize silicon photonics technology.
Device Story
Maverick Diagnostic System TC1000 is an automated immunoassay analyzer; utilizes silicon photonics technology for analyte detection. Input: human serum or K2EDTA plasma samples. Operation: system performs quantitative determination of TSH; results provided to clinical laboratory personnel. Output: quantitative TSH concentration values. Clinical utility: aids in diagnosis of thyroid or pituitary disorders. Used in clinical laboratory environments by trained professionals.
Clinical Evidence
Bench testing only. Analytical performance established via precision (repeatability/reproducibility), linearity (0.12–25.45 µIU/mL), interference, cross-reactivity, and limit of quantitation (0.20 µIU/mL). Method comparison against predicate (N=127/125) showed high correlation (R²=0.95–0.96). Reference range established using 120 healthy individuals.
Technological Characteristics
Silicon photonics ring resonance sensing; sandwich immunoassay. Instrument: automated analyzer with laser/photodiode optics, fluidics, and cloud-based data processing. Consumables: silicon chip with capture antibodies, pre-filled stripwells. Connectivity: cloud-based protocol retrieval and data processing. Sterilization: not specified. Standards: CLSI EP05-A3, EP06-Ed2, EP07-A3, EP09c, EP28-A3c, EP37; IEC 62304, 60601-1, 60601-1-2.
Indications for Use
Indicated for quantitative determination of human thyroid stimulating hormone (TSH) in human serum and K2EDTA plasma for the diagnosis of thyroid or pituitary disorders in clinical laboratory settings.
Regulatory Classification
Identification
A thyroid stimulating hormone test system is a device intended to measure thyroid stimulating hormone, also known as thyrotrophin and thyrotrophic hormone, in serum and plasma. Measurements of thyroid stimulating hormone produced by the anterior pituitary are used in the diagnosis of thyroid or pituitary disorders.
Submission Summary (Full Text)
{1}
FDA
U.S. FOOD & DRUG
ADMINISTRATION
# 510(k) SUBSTANTIAL EQUIVALENCE DETERMINATION DECISION SUMMARY
ASSAY AND INSTRUMENT
## I Background Information:
A 510(k) Number
K234091
B Applicant
Genalyte, Inc.
C Proprietary and Established Names
Maverick Diagnostic System TC1000; Maverick Test Panel A0.B0
D Regulatory Information
| Product Code(s) | Classification | Regulation Section | Panel |
| --- | --- | --- | --- |
| JLW | Class II | 21 CFR 862.1690 - Thyroid Stimulating Hormone Test System | CH - Clinical Chemistry |
| JJE | Class I | 21 CFR 862.2160 - Discrete photometric chemistry analyzer for clinical use | CH - Clinical Chemistry |
## II Submission/Device Overview:
A Purpose for Submission:
New Device
B Measurand:
Human thyroid stimulating hormone (thyrotropin, TSH)
C Type of Test:
Sandwich immunoassay
Food and Drug Administration
10903 New Hampshire Avenue
Silver Spring, MD 20993-0002
www.fda.gov
{2}
III Intended Use/Indications for Use:
A Intended Use(s):
See Indications for Use below.
B Indication(s) for Use:
The Maverick Test Panel A0.B0 is an immunoassay for the quantitative determination of human thyroid stimulating hormone (thyrotropin, TSH) in human serum and K2EDTA plasma on the Maverick Diagnostic System TC1000. Measurements of thyroid stimulating hormone produced by the anterior pituitary are used in the diagnosis of thyroid or pituitary disorders.
The Maverick Diagnostic System TC1000 is an automated immunoassay analyzer intended for in vitro diagnostic use to determine analytes in a clinical laboratory. The system’s assay applications utilize silicon photonics technology.
C Special Conditions for Use Statement(s):
Rx - For Prescription Use Only
D Special Instrument Requirements:
Maverick Diagnostic System TC1000
IV Device/System Characteristics:
A Device Description:
The Maverick Test Panel A0.B0 consists of:
- A silicon chip spotted with capture antibody (anti-TSH mouse monoclonal antibody), and three quality control proteins housed in a carrier in foil pouch with desiccant.
- A sealed, pre-filled stripwell that contains all the reagents required to complete the test protocol. Each stripwell plate contains:
- High salt buffer (wells A1 & A2) containing phosphate buffer and salt
- Assay buffer (wells B1, C1, D1 & B2, C2, D2) containing PBS, Tween®-20, proprietary proteins, and preservative
- TSH Assay Detection Reagent buffer (wells F1 & F2) containing biotinylated anti-TSH mouse monoclonal antibody in assay buffer
- Assay pre-conditioning buffers (wells G1, H1 & G2, H2) containing borate and citrate buffers
- Amplification buffer (wells E1 & E2) contains streptavidin reagent in assay buffer
The Maverick Diagnostic System consists of the Maverick instrument, computer, software, peripherals including a bar code reader and a monitor, and MTP A0.B0 assay specific reagent kit. The Maverick Diagnostic System TC1000 instrument is provided with necessary electronics, optics, fluidics, software and mechanical hardware to perform assays for quantitative detection of analytes. The Maverick Diagnostic System TC1000 instrument has two independently controlled bays, each of which can accept separate chip and reagent consumables. The chip carrier contains two (2) discrete channels to accommodate two (2) different samples. Two samples can be run
K234091 - Page 2 of 11
{3}
simultaneously in each bay, or four samples total per instrument. Each bay contains mechanisms to independently accept the reagent stripwell and chip (in carrier) assay components (consumable interface); control their positioning (Y-Z Stage); provide laser input for scanning the chip and detect light frequencies output by the chip (photodiode in optics scanner). On the side of the instrument there is a fluidics component for each bay that consists of two electronically-driven syringe pumps fluid lines, manifold. The computer component of the MDS manages the user interface, the software which controls the instrument, and provides a gateway between instrument and the cloud. The system uses cloud-based retrieval of protocols for each kit; after the test is completed, the raw data is uploaded to the cloud where algorithms evaluate the controls and process the data into assay results. Test results are returned to the laboratory's information system.
## B Principle of Operation:
The assay relies on silicon photonics that uses ring resonance to measure binding of macromolecules to sensors on a miniature silicon chip. The Maverick Diagnostic System detects changes in resonance wavelength as macromolecules such as antibodies, proteins, or hormones bind to their respective antigens that are bound to the chip.
The silicon chips have two fluidic channels to run two independent assays. Each channel consists of 16 silicon clusters each with 4 rings. The chip is assembled into a chip carrier that is inserted into Maverick Diagnostic System to run the assay. The MTP A0.B0 assay consists of antibody and internal controls described below. Anti TSH antibody, Human Serum Albumin (HSA), goat anti-human IgG, and Human IgG are spotted on one cluster each per channel.
The Maverick TSH assay is a sandwich immunoassay. Patient sample is added to wells of a stripwell which already contains required buffers. The stripwell and the chip carrier are loaded into the Maverick instrument. When diluted sample is flowed over the chip, TSH in the sample binds to anti-TSH capture antibody. After a wash step detection reagent flows over the chip. TSH bound to the capture on rind sensors binds to the detection antibody and after a wash step, amplification reagent binds to the biotinylated antibody. The changes in resonant wavelength due to the binding are detected by the instrument. The raw results are converted to reportable units as $\mu \mathrm{IU} / \mathrm{mL}$ for TSH using the lot specific calibration.
## C Instrument Description Information:
1. Instrument Name:
Maverick Diagnostic System TC1000
2. Specimen Identification:
The specimen is held in a tube with a barcode label. Specimen identification is done by scanning the barcode.
3. Specimen Sampling and Handling:
Sample handling instructions are provided in the device labeling.
K234091 - Page 3 of 11
{4}
K234091 - Page 4 of 11
4. Calibration:
The lot specific coefficients are entered in the MDS Cloud. The barcode for each kit contains the information about lot coefficients in the cloud. When a sample is tested, the raw Genalyte Response units (GRU) results are converted to the reportable TSH units $\mu \mathrm{IU} / \mathrm{mL}$ using the lot specific coefficients.
5. Quality Control:
External controls are not provided with the assay kit. The sponsor recommends that users run external controls as daily controls for the instrument. Internal controls are incorporated into the assay test chip: a negative control to assess non-specific binding; a positive control that verifies specimen has been added; and a system control that assesses the system function.
V Substantial Equivalence Information:
A Predicate Device Name(s):
Access TSH (3RD IS) Assay
Unicel DxI 800 Access Immunoassay System
B Predicate 510(k) Number(s):
K153651
K023764
C Comparison with Predicate(s):
| Device & Predicate Device(s): | K234091 | K153651 |
| --- | --- | --- |
| Device Trade Name | Maverick Test Panel A0.B0 | Access TSH (3rd IS) Assay |
| General Device Characteristic Similarities | | |
| Intended Use/Indications For Use | For the quantitative determination of human thyroid stimulating hormone (TSH, thyrotropin) in human serum and plasma. | Same |
| Assay Type | Sandwich immunoassay | Same |
| General Device Characteristic Differences | | |
| Assay Principle | Photonic Ring Immunoassay | Chemiluminescent Immunoassay |
| Solid phase | Silicone rings fabricated on a chip | Paramagnetic microparticles (beads) |
{5}
| Assay Measuring Range | 0.2– 20.0 μIU/mL | 0.01 – 50.0 μIU/mL |
| --- | --- | --- |
| Device & Predicate Device(s): | K234091 | K023764 |
| --- | --- | --- |
| Device Trade Name | Maverick Diagnostic System TC1000 | Unicel DxI 800 Access Immunoassay System |
| General Device Characteristic Similarities | | |
| Intended Use/Indications For Use | Intended for in vitro diagnostic use to determine analytes in a clinical laboratory | Same |
| General Device Characteristic Differences | | |
| Detection/Operating Principle | Photonic Ring Immunoassay | Chemiluminescent Immunoassay |
VI Standards/Guidance Documents Referenced:
CLSI EP05-A3: Evaluation of Precision of Qualitative Measurement Methods Procedures; Approved Guideline – Third Edition
CLSI EP06-Ed2: Evaluation of the Linearity of Quantitative Measurement Procedures – 2nd Edition
CLSI EP07-A3: Interference Testing in Clinical Chemistry; Approved Guideline– Third Edition
CLSI EP09c: Measurement Procedure Comparison and Bias Estimation Using Patient Samples; Approved Guideline 3rd Edition
CLSI EP28-A3c: Defining, Establishing, and Verifying Reference Intervals in the Clinical Laboratory; Approved Guideline
CLSI EP37: Supplemental Tables for Interference Testing in Clinical Chemistry– 1st Edition
IEC 62304 Edition 1.1 2015-06: Medical device software - Software life cycle processes
IEC 60601-1-2 Edition 4.1 2020-09: Medical electrical equipment - Part 1-2: General requirements for basic safety and essential performance - Collateral Standard: Electromagnetic disturbances - Requirements and tests
IEC 60601-1 Edition 3.2 2020-08: Medical Electrical Equipment - Part 1: General Requirements For Basic Safety And Essential Performance
K234091 - Page 5 of 11
{6}
VII Performance Characteristics (if/when applicable):
A Analytical Performance:
1. Precision/Reproducibility:
Precision studies were conducted to estimate repeatability and within-laboratory precision. Five serum samples spanning the assay measuring range were tested in 2 runs per day, per bay of the instrument, for 20 days on each bay of the instrument, for a total of 80 results per sample tested per bay of the instrument. There are two bays in an instrument and one chip per bay.
Bay 1
| Sample | N | Mean (μIU/mL) | Repeatability | | Between Run | | Within Day | | Between Day | | Within Laboratory | |
| --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- |
| | | | SD | CV (%) | SD | CV (%) | SD | CV (%) | SD | CV (%) | SD | CV (%) |
| 1 | 80 | 0.28 | 0.02 | 6.7 | 0.09 | 3.4 | 0.02 | 6.8 | 0.01 | 2.0 | 0.02 | 7.0 |
| 2 | 80 | 2.53 | 0.08 | 3.1 | 0.00 | 0.0 | 0.12 | 4.6 | 0.00 | 0.0 | 0.12 | 4.6 |
| 3 | 80 | 4.76 | 0.24 | 5.0 | 0.14 | 1.4 | 0.36 | 7.5 | 0.00 | 0.0 | 0.36 | 7.5 |
| 4 | 80 | 9.94 | 0.47 | 4.8 | 0.00 | 0.7 | 0.49 | 5.0 | 0.00 | 0.0 | 0.49 | 5.0 |
| 5 | 80 | 16.70 | 0.69 | 4.1 | 0.27 | 5.6 | 0.97 | 5.8 | 0.00 | 0.0 | 0.97 | 5.8 |
Bay 2
| Sample | N | Mean (μIU/mL) | Repeatability | | Between Run | | Within Day | | Between Day | | Within Laboratory | |
| --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- |
| | | | SD | CV (%) | SD | CV (%) | SD | CV (%) | SD | CV (%) | SD | CV (%) |
| 1 | 80 | 0.28 | 0.02 | 7.0 | 0.08 | 3.2 | 0.11 | 4.4 | 0.00 | 0.0 | 0.02 | 7.3 |
| 2 | 80 | 2.57 | 0.08 | 3.1 | 0.00 | 0.0 | 0.01 | 7.3 | 0.01 | 7.4 | 0.11 | 4.4 |
| 3 | 80 | 4.90 | 0.26 | 5.2 | 0.14 | 1.4 | 0.40 | 4.1 | 0.45 | 4.5 | 0.34 | 7.0 |
| 4 | 80 | 9.94 | 0.38 | 3.8 | 0.00 | 0.0 | 0.02 | 7.0 | 0.02 | 7.3 | 0.45 | 4.5 |
| 5 | 80 | 17.41 | 0.71 | 4.1 | 0.23 | 4.7 | 0.34 | 7.0 | 0.00 | 0.0 | 1.12 | 6.4 |
A reproducibility study was conducted to assess between lot variability. Five serum samples with different TSH concentrations spanning the assay measuring range were tested across 3 reagent lots. Each sample was tested in duplicate per run, per bay of the instrument, 3 runs per day, per reagent lot, for 5 days. The within-run (repeatability), within lot, between-lot, and reproducibility was calculated for each reagent lot. The result summary is shown below:
| Sample | N | Mean | Repeatability | | Within Reagent Lot | | Between Reagent Lot | | Reproducibility | |
| --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- |
| | | | SD | CV (%) | SD | CV (%) | SD | CV (%) | SD | CV (%) |
| 1 | 180 | 0.22 | 0.02 | 7.8 | 0.02 | 7.9 | 0.00 | 0.3 | 0.02 | 7.9 |
| 2 | 180 | 1.94 | 0.09 | 4.6 | 0.09 | 4.7 | 0.03 | 1.8 | 0.10 | 5.0 |
| 3 | 180 | 4.42 | 0.23 | 5.2 | 0.23 | 5.2 | 0.09 | 2.0 | 0.25 | 5.6 |
| 4 | 180 | 12.11 | 0.68 | 5.6 | 0.68 | 5.6 | 0.07 | 0.6 | 0.69 | 5.7 |
K234091 - Page 6 of 11
{7}
K234091 - Page 7 of 11
| 5 | 180 | 18.11 | 0.85 | 4.7 | 0.85 | 4.7 | 0.33 | 1.8 | 0.91 | 5.0 |
| --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- |
A reproducibility study was conducted to assess between instrument variability. Four serum samples spanning assay measuring range were tested in duplicates per run, 3 runs per day per bay of the instrument on 3 instruments (total 6 bays) for 5 days. The result summary is shown below:
| Sample | N | Mean | Repeatability | | Within Instrument | | Between Instrument | | Reproducibility | |
| --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- |
| | | | SD | CV (%) | SD | CV (%) | SD | CV (%) | SD | CV (%) |
| 1 | 180 | 0.58 | 0.04 | 7.3 | 0.04 | 7.3 | 0.00 | 0.6 | 0.04 | 7.4 |
| 2 | 180 | 5.02 | 0.38 | 7.6 | 0.39 | 7.8 | 0.00 | 0.0 | 0.39 | 7.8 |
| 3 | 180 | 13.14 | 1.17 | 8.9 | 1.16 | 8.8 | 0.47 | 3.6 | 1.25 | 9.5 |
| 4 | 180 | 15.78 | 0.75 | 4.8 | 0.74 | 4.7 | 0.24 | 1.5 | 0.76 | 4.8 |
2. Linearity:
A high TSH serum pool was created using two native samples with known high TSH values. The high TSH serum pool was serially diluted with a low serum sample TSH pool to create 16 TSH serum samples distributed across the analytical measuring interval. Each level was tested in replicates of 4 using 2 instruments and one reagent kit lot. Linearity study data were analyzed by weighted linear regression and the observed values (mean of the replicates) were compared to the best fitted straight line (predicted values). The linearity regression results are shown below. The results of the linearity study were used by the sponsor to support the claimed measuring interval of 0.2 μIU/mL to 20.0 μIU/mL.
| Test Range (μIU/mL) | Linear regression | Claimed Measuring Range (μIU/mL) |
| --- | --- | --- |
| 0.12 – 25.45 | y = 1.01x – 0.004, R2 = 1.00 | 0.2 – 20.0 |
3. Analytical Specificity/Interference:
i. Interference:
Two serum samples containing hTSH concentrations of approximately 0.45 μIU/mL and 4.5 μIU/mL were tested by spiking with potentially interfering compounds at levels listed in the table below. Significant interference was defined as more than ±10% bias as compared to the control results. The results of the interference study are summarized in the table below:
{8}
| Interferent | Concentration ≤ 10% Bias | Units |
| --- | --- | --- |
| Acetaminophen | 15.6 | mg/dL |
| Acetylcysteine | 15.0 | mg/dL |
| Acetylsalicylic acid | 3.00 | mg/dL |
| Ampicillin - Na | 7.50 | mg/dL |
| Ascorbic acid | 5.25 | mg/dL |
| Biotin | 0.35 | mg/dL |
| Cefoxitin | 660 | mg/dL |
| Conjugated bilirubin | 45 | mg/dL |
| Cyclosporine | 0.18 | mg/dL |
| Hemoglobin | 500 | mg/dL |
| Heparin | 330 | Units/dL |
| Ibuprofen | 500 | mg/dL |
| Levothyroxine | 0.0429 | mg/dL |
| Liothyronine (Triiodothyronine) | 450 | pg/dL |
| Multivitamin | 1.0 | %V/V |
| Prednisolone | 0.0099 | mg/dL |
| Rheumatoid factor | 40.0 | IU/mL |
| Rifampicin | 4.80 | mg/dL |
| Total protein/Albumin | 6000 | mg/dL |
| Triglycerides | 3300 | mg/dL |
| Unconjugated bilirubin | 45 | mg/dL |
The assay package insert instructs users not to use hemolyzed, lipemic or icteric samples.
ii. Cross-reactivity
Cross-reactivity studies were performed to evaluate the susceptibility of the Maverick Diagnostic System to cross-reactivity with the following endogenous structural analogs to human TSH: human follicle stimulating hormone (FSH), human luteinizing hormone (LH), and human chorionic gonadotropin (hCG). Four replicates of two serum samples containing hTSH concentrations of approximately 0.45 µIU/mL and 4.5 µIU/mL were tested. Significant cross reactivity was defined as ≤10% bias in expected results compared to the control. The table below shows the highest concentrations tested with no cross reactivity.
| Substance | Highest concentration tested (mIU/mL) |
| --- | --- |
| Human Chorionic Gonadotropin | 100,000 |
| Follicle Stimulating Hormone | 1000 |
| Luteinizing Hormone | 3000 |
iii. High Dose Hook Effect
The high-dose hook effect of the Maverick Test Panel A0.B0 assay on the MDS TC1000 system was assessed by testing 16 serial dilutions of serum pool human with high TSH
K234091 - Page 8 of 11
{9}
level. Each sample was tested in replicates of four using five instruments and one reagent kit lot. No high dose hook effect was observed up to 5,000 μIU/mL of TSH. Samples with TSH concentrations above the measuring interval and as high as 5,000 μIU/mL will report > 20.0 μIU/mL.
K234091 - Page 9 of 11
4. Assay Reportable Range:
See section VII.A.2 Linearity.
5. Traceability, Stability, Expected Values (Controls, Calibrators, or Methods):
The measurand (TSH) in the Maverick Test Panel A0.B0 Calibrator is traceable to WHO 3rd International Reference Preparation, (NIBSC Coded 81/565).
6. Detection Limit
i. Limit of Blank (LoB)
Limit of Blank (LoB) was determined by assaying each of four TSH depleted analyte-free serum samples in replicates of six, per reagent lot on each instrument bay (two bays) with two reagent lots for three days. The LoB was determined to be 0.08 μIU/mL using the 95% non-parametric upper limit of the 144 replicates.
ii. Limit of Detection (LoD)
Limit of Detection (LoD) was determined by assaying four serum samples with low TSH concentrations in replicates of six, per reagent lot on each instrument bay (two bays) with two reagent lots for three days. The LoD was determined to be 0.10 μIU/mL based on the precision model, multiplied by the 95th percentile of the standard normal distribution and added to the LoB to calculate the LoD.
iii. Limit of Quantitation (LoQ)
Four serum samples with low levels of TSH were tested. The samples were tested in replicates of 12, per reagent lot on each instrument bay (two bays) with two reagent lots for three days. The limit of quantitation was defined as the value at which the within laboratory CV is ≤ 20% and determined to be 0.20 μIU/mL.
| LoB | LoD | LoQ |
| --- | --- | --- |
| 0.08 μIU/mL | 0.10 μIU/mL | 0.20 μIU/mL |
7. Assay Cut-Off:
Not applicable.
8. Accuracy (Instrument):
See Section VII B.1. below.
{10}
9. Carry-Over:
The chip and chip carrier are single use. Therefore, after running a test each chip will be disposed. A new chip will be used for the next assay. The flow during an assay run is one way through sippers and to the waste. There is no possibility for sample carry-over between runs.
B Comparison Studies:
1. Method Comparison with Predicate Device:
A method comparison study was performed comparing the Maverick Test Panel A0.B0 assay on the Maverick Diagnostic System TC1000 to a legally marketed comparator device. Native human serum samples were tested in single replicates on each bay of the Maverick Diagnostic System TC1000. The study was run on seven Maverick Diagnostic System TC1000 instruments with seven kit lots by three operators on three different days. A Passing Bablok regression analysis was performed to demonstrate equivalency between the MTP A0.B0 assay and the comparator device. The results of the regression analysis are summarized below:
| Bay | N | Sample Range (μIU/mL) | Slope (95% CI) | Intercept (95% CI) | Correlation Coefficient (R2) |
| --- | --- | --- | --- | --- | --- |
| 1 | 127 | Test: 0.24 – 19.40
Predicate: 0.25 – 24.47 | 0.96 (0.93 to 0.99) | 0.023 (-0.017 to 0.09) | 0.96 |
| 2 | 125 | Test: 0.20 – 19.67
Predicate: 0.21 – 24.47 | 0.95 (0.93 to 0.99) | 0.038 (-0.018 to 0.098) | 0.95 |
2. Matrix Comparison:
A matrix comparison study was conducted utilizing 55 matched K2EDTA plasma and serum samples. Five altered samples were prepared by spiking known concentrations of human TSH into both sample types. All samples were assayed in single replicates for each sample matrix on two lots of reagent, using five instruments. Weighted Deming regression analysis results are shown below:
| Comparison | N | Slope | Intercept | Correlation Coefficient (r) | Range (μIU/mL) |
| --- | --- | --- | --- | --- | --- |
| Serum vs. K2EDTA plasma | 55 | 0.97 | 0.004 | 0.99 | Serum: 0.19 – 17.08
Plasma: 0.20 – 16.86 |
K234091 - Page 10 of 11
{11}
K234091 - Page 11 of 11
C Clinical Studies:
1. Clinical Sensitivity:
Not applicable
2. Clinical Specificity:
Not applicable
3. Other Clinical Supportive Data (When 1. and 2. Are Not Applicable):
Not applicable
D Clinical Cut-Off:
Not applicable.
E Expected Values/Reference Range:
To establish a normal range for the candidate assay, a total of 120 serum samples were tested from presumptively healthy individuals 21 years and older (60 males and 60 females). Non-parametric analysis was performed. The sponsor claimed the established reference range is 0.36 – 4.34 µIU/mL.
F Other Supportive Instrument Performance Characteristics Data:
Not applicable.
VIII Proposed Labeling:
The labeling supports the finding of substantial equivalence for this device.
IX Conclusion:
The submitted information in this premarket notification is complete and supports a substantial equivalence decision.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.