COSINE™ Spacer

K234035 · Globus Medical, Inc. · MAX · Sep 6, 2024 · Orthopedic

Device Facts

Record IDK234035
Device NameCOSINE™ Spacer
ApplicantGlobus Medical, Inc.
Product CodeMAX · Orthopedic
Decision DateSep 6, 2024
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 888.3080
Device ClassClass 2
AttributesTherapeutic

Intended Use

The COSINE™ Spacer is an interbody fusion device intended for use at one or more levels of the thoracic spine (T1-T12), thoracolumbar junction (T12-L1), or lumbosacral spine (L1-S1) as an adjunct to fusion in patients with the following indications: degenerative disc disease (DDD), disc herniation (with myelopathy and/or radiculopathy), spondylolisthesis, deformity (degenerative scoliosis or kyphosis), spinal stenosis, and failed previous fusion (pseudarthrosis). DDD is defined as discogenic back pain with degeneration of the disc confirmed by history and radiographic studies. These patients should be skeletally mature and have had at least six (6) months of non-operative treatment. The COSINE™ Spacer is to be filled with autograft and/or allogenic bone graft comprised of cancellous and/or corticocancellous bone or a bone void filler as cleared by the FDA for use in intervertebral body fusion to facilitate fusion. This device is intended to be used with supplemental fixation systems that have been cleared for use in the thoracolumbosacral spine (e.g., posterior pedicle screw and rod systems, anterior plate systems, and anterior screw and rod systems).Hyperlordotic interbody devices (≥20° lordosis) must be used with at least anterior supplemental fixation.

Device Story

Expandable lateral lumbar interbody fusion device; provides structural stability post-discectomy. Device filled with autograft, allogenic bone graft, or FDA-cleared bone void filler to facilitate fusion. Used in thoracic, thoracolumbar, or lumbosacral spine. Requires supplemental fixation (e.g., pedicle screws, rods, plates); hyperlordotic versions (≥20°) require anterior fixation. Implanted by surgeons in clinical settings. Benefits patient by restoring disc height and providing stability to support spinal fusion.

Clinical Evidence

Bench testing only. Mechanical testing performed per ASTM F2077 and ASTM F2267, including static and dynamic compression, compression-shear, and subsidence testing, to demonstrate substantial equivalence.

Technological Characteristics

Expandable interbody spacer. Materials: Titanium alloy (ASTM F136/F1295), additively manufactured titanium powder endplates (ASTM F3001), and radiolucent PEEK internal component. Mechanical sensing/actuation not applicable. No software/connectivity.

Indications for Use

Indicated for skeletally mature patients with degenerative disc disease, disc herniation, spondylolisthesis, deformity, spinal stenosis, or failed previous fusion requiring interbody fusion at T1-S1 levels. Requires 6 months of failed non-operative treatment.

Regulatory Classification

Identification

An intervertebral body fusion device is an implanted single or multiple component spinal device made from a variety of materials, including titanium and polymers. The device is inserted into the intervertebral body space of the cervical or lumbosacral spine, and is intended for intervertebral body fusion.

Special Controls

*Classification.* (1) Class II (special controls) for intervertebral body fusion devices that contain bone grafting material. The special control is the FDA guidance document entitled “Class II Special Controls Guidance Document: Intervertebral Body Fusion Device.” See § 888.1(e) for the availability of this guidance document.(2) Class III (premarket approval) for intervertebral body fusion devices that include any therapeutic biologic (e.g., bone morphogenic protein). Intervertebral body fusion devices that contain any therapeutic biologic require premarket approval. (c) *Date premarket approval application (PMA) or notice of product development protocol (PDP) is required.* Devices described in paragraph (b)(2) of this section shall have an approved PMA or a declared completed PDP in effect before being placed in commercial distribution.

Predicate Devices

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ September 6, 2024 Image /page/0/Picture/1 description: The image contains the logos of the Department of Health & Human Services and the Food and Drug Administration (FDA). The Department of Health & Human Services logo is on the left, and the FDA logo is on the right. The FDA logo includes the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue text. Globus Medical, Inc. Jennifer Antonacci, Ph.D. Senior Group Manager, Regulatory Affairs 2560 General Armistead Ave. Audubon, Pennsylvania 19403 # Re: K234035 Trade/Device Name: COSINE™ Spacer Regulation Number: 21 CFR 888.3080 Regulation Name: Intervertebral Body Fusion Device Regulatory Class: Class II Product Code: MAX, PHM Dated: August 15, 2024 Received: August 15, 2024 # Dear Dr. Antonacci: We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download). {1}------------------------------------------------ Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review. the OS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181). Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050. All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-device-advicecomprehensive-regulatory-assistance/unique-device-identification-system-udi-system. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems. For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100). {2}------------------------------------------------ Sincerely, # Katherine D. Kavlock -S for Brent Showalter, Ph.D. Assistant Director DHT6B: Division of Spinal Devices OHT6: Office of Orthopedic Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health Enclosure {3}------------------------------------------------ # Indications for Use Submission Number (if known) #### K234035 Device Name COSINE™ Spacer #### Indications for Use (Describe) The COSINE™ Spacer is an interbody fusion device intended for use at one or more levels of the thoracic spine (T1-T12), thoracolumbar junction (T12-L1), or lumbosacral spine (L1-S1) as an adjunct to fusion in patients with the following indications: degenerative disc disease (DDD), disc herniation (with myelopathy and/or radiculopathy), spondylolisthesis, deformity (degenerative scoliosis or kyphosis), spinal stenosis, and failed previous fusion (pseudarthrosis). DDD is defined as discogenic back pain with degeneration of the disc confirmed by history and radiographic studies. These patients should be skeletally mature and have had at least six (6) months of non-operative treatment. The COSINE™ Spacer is to be filled with autograft and/or allogenic bone graft comprised of cancellous and/or corticocancellous bone or a bone void filler as cleared by the FDA for use in intervertebral body fusion to facilitate fusion. This device is intended to be used with supplemental fixation systems that have been cleared for use in the thoracolumbosacral spine (e.g., posterior pedicle screw and rod systems, anterior plate systems, and anterior screw and rod systems).Hyperlordotic interbody devices (≥20° lordosis) must be used with at least anterior supplemental fixation. Type of Use (Select one or both, as applicable) > Prescription Use (Part 21 CFR 801 Subpart D) Over-The-Counter Use (21 CFR 801 Subpart C) # CONTINUE ON A SEPARATE PAGE IF NEEDED. This section applies only to requirements of the Paperwork Reduction Act of 1995. #### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: > Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." Form Approved: OMB No. 0910-0120 Expiration Date: 06/30/2023 See PRA Statement below. > K234035 Page 1 of 1 {4}------------------------------------------------ # 510(k) Summary: COSINE™ Spacer | Company: | Globus Medical Inc.<br>2560 General Armistead Ave.<br>Audubon, PA 19403<br>610-930-1800 | |---------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------| | Contact: | Jennifer Antonacci, Ph.D.<br>Senior Group Manager, Regulatory Affairs | | Date Prepared: | August 15, 2024 | | Device Name: | COSINE™ Spacer | | Common Name: | Intervertebral Body Fusion Device | | Classification: | Per 21 CFR as follows:<br>§888.3080 Intervertebral Body Fusion Device<br>Product Code(s): MAX, PHM<br>Regulatory Class: II, Panel Code: 87 | | Primary Predicate: | RISE® Spacer (K221894) | | Additional<br>Predicates: | RISE® Spacer (K113447)<br>ELSA® Spacers (K161379, K203278)<br>SABLE® Expandable Spacer (K192115, K222270)<br>Modulus-XLIF Interbody System and Attrax Putty (K203714)<br>MOD-FX XLIF Interbody System (K210439, K220478) | # Purpose: The purpose of this submission is to request clearance for the COSINE™ Spacer. # Device Description: COSINE™ Spacers are expandable lateral lumbar interbody fusion devices used to provide structural stability in skeletally mature individuals following discectomy. The devices are available in various heights and geometric options to fit the anatomical needs of a wide variety of patients. COSINE™ Spacers are manufactured from titanium alloy per ASTM F136 or ASTM F1295. The endplates are additively manufactured from titanium powder, as specified in ASTM F3001, and an internal component is manufactured from radiolucent PEEK polymer. {5}------------------------------------------------ # Indications for Use: The COSINE™ Spacer is an interbody fusion device intended for use at one or more levels of the thoracic spine (T1-T12), thoracolumbar junction (T12-L1), or lumbosacral spine (L1-S1) as an adjunct to fusion in patients with the following indications: degenerative disc disease (DDD), disc herniation (with myelopathy and/or radiculopathy), spondylolisthesis, deformity (degenerative scoliosis or kyphosis), spinal stenosis, and failed previous fusion (pseudarthrosis). DDD is defined as discogenic back pain with degeneration of the disc confirmed by history and radiographic studies. These patients should be skeletally mature and have had at least six (6) months of non-operative treatment. The COSINE™ Spacer is to be filled with autograft and/or allogenic bone graft comprised of cancellous and/or corticocancellous bone or a bone void filler as cleared by the FDA for use in intervertebral body fusion to facilitate fusion. This device is intended to be used with supplemental fixation systems that have been cleared for use in the thoracolumbosacral spine (e.g., posterior pedicle screw and rod systems, anterior plate systems, and anterior screw and rod systems). Hyperlordotic interbody devices (≥20° lordosis) must be used with at least anterior supplemental fixation. # Performance Data: Mechanical testing (static and dynamic compression and compression-shear, and subsidence) was conducted in accordance with the "Guidance for Industry and FDA Staff, Class II Special Controls Guidance Document: Intervertebral Fusion Device." June 12. 2007. ASTM F2077. and ASTM F2267 to demonstrate substantial equivalence to the predicate devices. # Technological Characteristics: Subject implants have the same technological characteristics as the predicate devices including design, intended use, material composition, and range of sizes. # Basis of Substantial Equivalence: Subject interbody devices have been found to be substantially equivalent to the predicate devices with respect to technical characteristics, performance, and intended use. The information provided within this premarket notification supports substantial equivalence of the subject spacers to the predicate devices.
Innolitics
510(k) Summary
Decision Summary
Classification Order
Enter a record ID and click Load to view the document.
100%