The Circadia C200 System

K234003 · Circadia Technologies, Ltd. · DRT · May 30, 2024 · Cardiovascular

Device Facts

Record IDK234003
Device NameThe Circadia C200 System
ApplicantCircadia Technologies, Ltd.
Product CodeDRT · Cardiovascular
Decision DateMay 30, 2024
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 870.2300
Device ClassClass 2

Indications for Use

The Circadia C200 System is intended for the measurement of respiratory rate and heart rate, including spot measurement. The system is indicated for adult patients in clinical settings, such as skilled nursing and long-term care facilities. The system is not indicated for active patient monitoring, and does not provide alarms for timely response in acute lifethreatening situations. The system is not intended to monitor heart rate in patients with arrhythmias. The system is intended to be used by healthcare professionals (HCPs) and data are intended by HCPs to inform patient care. The system also monitors patient motion, and patient presence or absence near the device (exits).

Device Story

Contactless cardiorespiratory monitor using 6.4-7.8 GHz UWB radar to detect skin micromotions caused by ventilation and heartbeat. System comprises bedside Monitor, Cloud Service, and Pro App (Android tablet). Monitor processes data locally, streams to Cloud via Wi-Fi; App allows HCPs to review retrospective RR/HR data and real-time motion/presence status. Includes notification if no HR detected within 8 hours. Used in clinical settings (skilled nursing/long-term care) by HCPs to inform patient care. Benefits include long-term, unsupervised monitoring without patient-worn sensors or active HCP interaction.

Clinical Evidence

Clinical validation performed in two studies (n=49, n=41) comparing HR to gold standard (K182030). Agreement met pre-specified acceptance criteria of ± 5 BPM. RR performance validated against reference device (K200445) with identical hardware, demonstrating equivalence. Bench testing included software V&V (IEC 62304), electrical safety (IEC 60601-1), EMC (IEC 60601-1-2), and wireless testing (IEEE C63.27).

Technological Characteristics

Contactless radar-based sensor (6.4-7.8 GHz UWB). Measures skin micro-motions. Dimensions/form factor: bedside monitor. Connectivity: Wi-Fi to cloud, tablet-based app. Software: IEC 62304 compliant. Power: AC/DC. Standards: IEC 60601-1, IEC 60601-1-2, IEEE C63.27.

Indications for Use

Indicated for adult patients in clinical settings (e.g., skilled nursing, long-term care) for measurement of respiratory rate, heart rate, patient motion, and presence/absence. Not for active patient monitoring, acute life-threatening situations, or patients with arrhythmias.

Regulatory Classification

Identification

A cardiac monitor (including cardiotachometer and rate alarm) is a device used to measure the heart rate from an analog signal produced by an electrocardiograph, vectorcardiograph, or blood pressure monitor. This device may sound an alarm when the heart rate falls outside preset upper and lower limits.

Predicate Devices

Reference Devices

Submission Summary (Full Text)

{0}------------------------------------------------ Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, with the letters "FDA" in a blue square. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue. May 30, 2024 Circadia Technologies, Ltd. Erhan Ilhan Head of Quality and Regulatory 20 St Thomas Street London, Greater London SE1 9RS United Kingdom Re: K234003 Trade/Device Name: The Circadia C200 System Regulation Number: 21 CFR 870.2300 Regulation Name: Cardiac Monitor (Including Cardiotachometer And Rate Alarm) Regulatory Class: Class II Product Code: DRT, BZQ Dated: May 1, 2024 Received: May 1, 2024 Dear Erhan Ilhan: We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download). {1}------------------------------------------------ Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review. the OS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181). Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems. For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100). Sincerely, # Jennifer W. Shih -S Jennifer Shih Kozen Assistant Director Division of Cardiac Electrophysiology, Diagnostics and Monitoring Devices Office of Cardiovascular Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health {2}------------------------------------------------ ## Indications for Use 510(k) Number (if known) K234003 Device Name The Circadia C200 System ### Indications for Use (Describe) The Circadia C200 System is intended for the measurement of respiratory rate and heart rate, including spot measurement. The system is indicated for adult patients in clinical settings, such as skilled nursing and long-term care facilities. The system is not indicated for active patient monitoring, and does not provide alarms for timely response in acute lifethreatening situations. The system is not intended to monitor heart rate in patients with arrhythmias. The system is intended to be used by healthcare professionals (HCPs) and data are intended by HCPs to inform patient care. The system also monitors patient motion, and patient presence or absence near the device (exits). | Type of Use (Select one or both, as applicable) | | |---------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------| | <div style="display:flex; align-items:center;"><div><span style="font-size:20px;">☑</span></div><div>Prescription Use (Part 21 CFR 801 Subpart D)</div></div> | <div style="display:flex; align-items:center;"><div><span>☐</span></div><div>Over-The-Counter Use (21 CFR 801 Subpart C)</div></div> | ## CONTINUE ON A SEPARATE PAGE IF NEEDED. This section applies only to requirements of the Paperwork Reduction Act of 1995. ## *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: > Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." {3}------------------------------------------------ Image /page/3/Picture/0 description: The image shows the word "CIRCADIA" in a sans-serif font. To the left of the word is a circular logo that is partially filled in with a dark blue color. The word "CIRCADIA" is also in a dark blue color. The background of the image is white. # 510(k) Summary ### 1.0 Submitter Information Name: Circadia Technologies, Ltd. Address: Runway East Borough Market 20 St. Thomas Street. London SE1 9RS United Kingdom Contact Name: Erhan Ilhan Head of Quality and Regulatory Phone: (800) 985-5596 Date Prepared May 25, 2024 #### 2.0 Subject Device Trade Name: Common Name: Regulation Number: Class: Product Code: Premarket Review: The Circadia C200 System Cardiac Monitor 21 CFR 870.2300 II DRT, BZQ OPEQ/OHT2A: Cardiac Electrophysiology, Diagnostics, and Monitoring Devices Cardiovascular Review Panel: #### 3.0 Predicate Device Trade Name: Common Name: Regulation Number: Class: Product Code: Premarket Review: Vital Sign Monitoring Sensor (K202464) Cardiac Monitor 21 CFR 870.2300 II DRT, BZQ OPEQ/OHT2A: Cardiac Electrophysiology, Diagnostics, and Monitoring Devices Cardiovascular #### 4.0 Device Description Review Panel: The Circadia C200 System is a contactless system that uses radar to monitor respiratory rate (RR), heart rate (HR), motion, and presence of a patient in its detection range. The System is designed to monitor a patient automatically, without the need for the patient to wear or do anything, or for a healthcare professional (HCP) to interact with the device. The System is thus suitable for long-term and unsupervised monitoring. {4}------------------------------------------------ Image /page/4/Picture/0 description: The image contains the word "CIRCADIA" in a sans-serif font. To the left of the word is a circular logo. The logo is a dark blue color and appears to be a stylized representation of a circle with a crescent shape cut out of it. The System may also be used to obtain on-demand spot measurements of heart rate and respiratory rate. This allows HCPs to control the frequency and timing of these measurements using the C200 System. The System consists of the Circadia Contactless Cardiorespiratory Monitor (the "Monitor"), the Circadia Cloud Service (the "Cloud Service"), and the Circadia Pro App (the "App"). The Monitor may be installed next to a patient's bedside. It uses a radar-based motion sensor to detect micromotions caused by ventilation and heartbeat, to measure a patient's RR and HR while the patient is in its detection range at rest. Data is processed continuously on the Monitor, and streamed to the Cloud Service over a Wi-Fi network. The Cloud Service offers a set of Application Programming Interfaces (APIs) that allows the Monitor to connect to the server and send data over a secure channel. In addition, it allows for patient data to be retrieved from the App. The App allows a healthcare professional to retrospectively review RR and HR data from multiple connected Monitors. Motion and presence/exit data are available in real time. The App operates from an Android tablet (not supplied, not included in the System). The App includes a functionality to notify a user if no HR has been obtained within the most recent 8 hours. #### 5.0 Indications for Use The Circadia C200 System is intended for the measurement of respiratory rate and heart rate, including spot measurement. The system is indicated for adult patients in clinical settings, such as skilled nursing and long-term care facilities. The system is not indicated for active patient monitoring, and does not provide alarms for timely response in acute life-threatening situations. The system is not intended to monitor heart rate in patients with arrhythmias. The system is intended to be used by healthcare professionals (HCPs) and data are intended to be reviewed by HCPs to inform patient care. The system also monitors patient motion, and patient presence or absence near the device (exits). #### 6.0 Comparison to Predicate Device The Circadia C200 System is substantially equivalent to the Vital Sign Monitoring Sensor (Model: XK300) predicate devices (K202464), based on similar intended use and {5}------------------------------------------------ Image /page/5/Picture/0 description: The image shows the logo for Circadia. The logo consists of a stylized circle on the left and the word "CIRCADIA" on the right. The circle is made up of two semi-circles, one filled in and one outlined. The word "CIRCADIA" is written in a sans-serif font. technological characteristics. The subject device has the same intended use as the predicate device cleared in K202464 which is to contactlessly measure heart rate and respiration rate in adult patients in a clinical setting, such as skilled nursing and long-term care facilities. Additionally, the subject and predicate devices are intended to monitor motion in the detection range, and to monitor the presence or absence of a patient in the detection range. The subject device has similar technological characteristics to the device cleared in K202464. These technological characteristics have undergone testing to ensure the device is as safe and effective as the predicates. Table 6-1 compares the indications for use and characteristics of the Circadia C200 system to the predicate. | Feature | Subject Device:<br>Circadia C200<br>System | Predicate Device:<br>Vital Sign<br>Monitoring<br>Sensor (Model<br>XK300) | Reference Device:<br>Circadia C100<br>System | Comments | |-----------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------| | 510(k) | K234003 | K202464 | K200445 | N/A | | Decision date | N/A | 26-APR-2021 | 24-JUN-2020 | N/A | | Product Name | Cardiac Monitor<br>(Including<br>Cardiotachometer And<br>Rate Alarm) | Cardiac Monitor<br>(Including<br>Cardiotachometer<br>And Rate Alarm) | Breathing<br>Frequency Monitor | Identical to Predicate | | Product Code | DRT, BZQ | DRT, BZQ | BZQ | Identical to Predicate | | Intended Use | The intended use is to<br>contactlessly measure<br>heart rate and<br>respiration rate and<br>detect patient motion. | The intended use<br>is to contactlessly<br>measure heart rate<br>and respiration<br>rate and detect<br>patient motion. | The intended use is<br>to contactlessly<br>measure respiratory<br>rate. | Identical to Predicate | | Indications<br>for Use | The Circadia C200<br>System is intended for<br>the measurement of<br>respiratory rate and<br>heart rate, including<br>spot measurement.<br>The system is<br>indicated for adult<br>patients in clinical<br>settings, such as<br>skilled nursing and<br>long-term care<br>facilities.<br>The system is not<br>indicated for active<br>patient monitoring,<br>and does not provide<br>alarms for timely<br>response in acute<br>life-threatening<br>situations. The system<br>is not intended to<br>monitor heart rate in<br>patients with<br>arrhythmias.<br>The system is intended<br>to be used by<br>healthcare<br>professionals (HCPs)<br>and data are intended<br>to be reviewed by<br>HCPs to inform patient<br>care.<br>The system also<br>monitors patient<br>motion, and patient<br>presence or absence<br>near the device (exits). | The Vital Sign<br>Monitoring Sensor<br>(Model XK300) is<br>intended to<br>measure heart rate<br>and respiration<br>rate in adult<br>patients in a<br>general care<br>hospital<br>environment<br>including nursing<br>homes. The Vital<br>Sign Monitoring<br>Sensor can be<br>used for home<br>healthcare for data<br>collection to<br>inform patient<br>care but not to<br>acutely treat a<br>patient. XK300<br>monitors the<br>presence or<br>absence of a<br>patient in a<br>detection area of<br>within 7 meters.<br>The XK300 also<br>monitors the<br>length of<br>continuous patient<br>motion or absence<br>of patient motion. | The Circadia C100<br>System is indicated<br>for both contactless<br>spot checking and<br>continuous<br>measurement of<br>respiratory rate data<br>as part of a vital<br>signs assessment.<br>The system records,<br>transmits, and<br>displays respiratory<br>rate from multiple<br>connected devices<br>for retrospective<br>analysis only. The<br>system is intended<br>to be used under the<br>care of clinicians<br>and medically<br>qualified personnel.<br>The system is<br>indicated for use in<br>adult patients during<br>no-motion<br>conditions, for<br>patients in health<br>care facilities. It is<br>available for sale<br>only upon the order<br>of a physician or<br>licensed health care<br>provider.<br>The Circadia C100<br>System is not<br>indicated for active<br>patient monitoring,<br>as it does not<br>provide alarms for<br>timely response in<br>life-threatening<br>situations. It is not<br>intended to monitor<br>vital signs. This<br>system is not an | Similar to Predicate<br>The Subject device is<br>not intended for use in<br>the home environment. | | | | | apnea monitor. | | | Patient Type | Adult | Adult | Adult | Identical to Predicate | | User<br>Population | Healthcare Providers | Healthcare<br>Providers | Healthcare<br>Providers | Identical to Predicate | | Environment | Healthcare facilities,<br>including skilled<br>nursing and<br>long-term care<br>facilities. | General Care<br>Hospital<br>Environment<br>including Nursing<br>Homes | Healthcare facilities | Similar to Predicate | | Technology | Contactless,<br>radar-based<br>measurement of<br>micro-motions | Contactless,<br>radar-based<br>measurement of<br>micro-motions | Contactless,<br>radar-based<br>measurement of<br>micro-motions | Identical to Predicate | | Physiological<br>phenomenon<br>measured | Micro-motions of the<br>skin, caused by<br>ventilation and<br>heartbeat | Micro-motions of<br>the skin, caused<br>by ventilation and<br>heartbeat | Micro-motions of<br>the skin, caused by<br>ventilation and<br>heartbeat | Identical to Predicate | | Radar sensor | 6.4 - 7.8 GHz UWB | 6.4 - 7.8 GHz<br>UWB | 6.4 - 7.8 GHz UWB | Identical to Predicate | | Contactless<br>Monitoring | Yes | Yes | Yes | Identical to Predicate | | Heart Range<br>Measurement<br>Range | 40 - 140 beats per<br>minute | 60 - 120 beats per<br>minute | N/A | Similar to Predicate | | Respiratory<br>Rate<br>Measurement<br>Range | 7 - 38 breaths per<br>minute | 6 - 55 breaths per<br>minute | 7 – 38 breaths per<br>minute | Similar to Predicate<br>and Identical to C100<br>Reference | | Detects<br>Patient<br>Motion, and<br>Presence/<br>Absence | Yes | Yes | No | Identical to Predicate | ## Table 6-1: Comparison to the Predicate Device {6}------------------------------------------------ Image /page/6/Picture/0 description: The image shows the logo for Circadia. The logo consists of a circular symbol on the left and the word "CIRCADIA" on the right. The circular symbol is made up of two crescent shapes, one inside the other. The text is in a sans-serif font and is a dark blue color. {7}------------------------------------------------ Image /page/7/Picture/0 description: The image shows the word "CIRCADIA" in a sans-serif font. To the left of the word is a circular logo that is split into two halves. The left half of the logo is white, and the right half is dark blue. #### 7.0 Performance Data (Non-Clinical Testing) The following tests were performed to demonstrate safety based on current industry standards: {8}------------------------------------------------ Image /page/8/Picture/0 description: The image shows the word "CIRCADIA" in bold, dark blue letters. To the left of the word is a dark blue symbol that looks like two crescent moons facing each other. The symbol is slightly smaller than the word "CIRCADIA". The overall design is simple and modern. - . Software Verification and Validation (per IEC 62304:2006/A1:2015) - . Electrical Safety (per IEC 60601-1:2005) - . EMC Testing (per IEC 60601-1-2:2014) - . Wireless Testing (per IEEE C63.27-2017) - Motion Validation - . Presence Validation - . HR Detectable Range Verification - . HR Acquisition Rate Verification - Circadia Monitor Drop Test . - Circadia Monitor Power Cord/Button Fatigue Test ● The results of these tests indicate that the Circadia C200 System is substantially equivalent to the predicate device. #### 8.0 Performance Data (Clinical Testing) To evaluate the performance of the Circadia C200 System, heart rate (HR) and respiratory rate (RR) clinical validation was performed. Clinical testing was performed to evaluate RR monitoring performance, using the Circadia C100 System (reference device, K200445), which has identical hardware and RR monitoring functionality as the subject device. Clinical testing demonstrated that RR monitoring performance of the subject device was equivalent to the reference device. Clinical testing was performed to evaluate HR monitoring performance, in two separate studies, by comparing subject device HR to gold standard reference HR (obtained using reference device K182030). Performance was assessed in clinical populations (n=49 and n=41 patients, for studies 1 and 2. respectively) and use cases representative of the subject device indications for use. Agreement with the gold standard was found to meet the pre-specified acceptance criteria of ± 5 BPM. Clinical testing demonstrated that HR monitoring performance of the subject device was equivalent to the predicate device. Subject demographics and baseline characteristics for both studies can be found in Table 8-1. {9}------------------------------------------------ Image /page/9/Picture/0 description: The image shows the logo for Circadia. The logo consists of a stylized circle on the left and the word "CIRCADIA" on the right. The circle is dark blue and has a crescent shape cut out of it. The text is also dark blue and in all caps. Table 8-1: Subject baseline demographics | Characteristic | Study 1, N=49 | Study 2, N=41 | Total, N=90 | |-------------------------------------|---------------|---------------|-------------| | Age [years] | | | | | Range | 33 - 92 | 27 - 88 | 27 - 92 | | Median (interquartile range) | 75 (11) | 72 (12) | 73 (14) | | Gender [N (%)] | | | | | Female | 27 (55%) | 21 (51%) | 48 (53%) | | Male | 22 (45%) | 20 (49%) | 42 (47%) | | Body-mass index [kg/m²] | | | | | Range | 17 - 73 | 15 - 56 | 15 - 73 | | Median (interquartile range) | 32 (11) | 27 (8) | 28 (12) | | Race [N (%)] | | | | | American Indian or Alaska<br>Native | - | 1 (2%) | 1 (1%) | | Asian | 2 (4%) | 1 (2%) | 3 (3%) | | Black or African American | 7 (14%) | 13 (32%) | 20 (22%) | | White | 40 (82%) | 23 (56%) | 63 (70%) | | More than one race | - | 3 (7%) | 3 (3%) | | Ethnicity [N (%)] | | | | | Hispanic or Latino | 6 (12%) | 6 (15%) | 12 (13%) | | Not Hispanic or Latino | 43 (88%) | 35 (85%) | 78 (87%) | N, number; kg, kilogram; m, meter. #### 9.0 Warnings and Important Use Conditions The C200 System is not intended to monitor heart rate in patients with arrhythmias, such as atrial fibrillation. Use within this patient population will likely result in reduced performance. For optimal heart rate monitoring performance, the C200 System Monitor is intended to be used to the side of the patient, and should be positioned 1.0 meter (3.0 feet) or less from the patient. {10}------------------------------------------------ #### 10.0 Conclusion The Circadia C200 System is substantially equivalent to the predicate device based on the testing performed, the same intended use and similar indications, technological characteristics and principles of operation. The differences between the devices have been properly evaluated through the software verification and validation, bench test validation, and clinical validation, and do not present any new issues of safety or effectiveness. Thus, the Circadia C200 System was determined to be substantially equivalent to the predicate devices.
Innolitics
Guidance Recalls

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...